Basic Fundamental Parameters in X-Ray Fluorescence

Similar documents
Lectures about XRF (X-Ray Fluorescence)

Coating Thickness and Composition Analysis by Micro-EDXRF

EDXRF of Used Automotive Catalytic Converters

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Atomic and Nuclear Physics Laboratory (Physics 4780)

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Instrumentation. (Figure 2)

Atomic Calculations. 2.1 Composition of the Atom. number of protons + number of neutrons = mass number

X Ray Flourescence (XRF)

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

The Three Heat Transfer Modes in Reflow Soldering

Quality Control on Aerospace Components Using Handheld X-ray Fluorescence (XRF)

Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE

Introduction to the Monte Carlo method

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, Bologna, Italy

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

where h = J s

Spectral distribution from end window X-ray tubes

Introduction to Elemental Analysis by ED-XRF Justin Masone Product Specialist 3 June 2015

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

- thus, the total number of atoms per second that absorb a photon is

C-DEBI Research Coordination Network

Copyright by Mark Brandt, Ph.D. 12

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse

5.33 Lecture Notes: Introduction to Spectroscopy

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün

Introduction to Powder X-Ray Diffraction History Basic Principles

GRID AND PRISM SPECTROMETERS

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

Introduction to Fourier Transform Infrared Spectrometry

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

O6: The Diffraction Grating Spectrometer

Time out states and transitions

X-ray Techniques: Overview. Ron Jenkins

FTIR Instrumentation

Rate Equations and Detailed Balance

Field portable XRF analysis of environmental samples

GAMMA-RAY SPECTRA REFERENCES

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

How To Understand Light And Color

OLIVÉR BÁNHIDI 1. Introduction

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and

Universal Data Acquisition (UDA)

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

Introduction to Geiger Counters

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Tutorial 4.6 Gamma Spectrum Analysis

Radioactivity III: Measurement of Half Life.

- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier

Raman Scattering Theory David W. Hahn Department of Mechanical and Aerospace Engineering University of Florida

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Experiment #5: Qualitative Absorption Spectroscopy

CHEM 1411 Chapter 5 Homework Answers

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Generation of X-Rays (prepared by James R. Connolly, for EPS , Introduction to X-Ray Powder Diffraction, Spring 2005)

MODERN ATOMIC THEORY AND THE PERIODIC TABLE

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Production of X-rays and Interactions of X-rays with Matter

Polarization Dependence in X-ray Spectroscopy and Scattering. S P Collins et al Diamond Light Source UK

Using the Spectrophotometer

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

Nuclear Physics. Nuclear Physics comprises the study of:

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

Energy Dispersive Spectroscopy on the SEM: A Primer

INFRARED SPECTROSCOPY (IR)

13C NMR Spectroscopy

Copyright 2011 Casa Software Ltd. Centre of Mass

3 - Atomic Absorption Spectroscopy

Main properties of atoms and nucleus

X-Ray Diffraction HOW IT WORKS WHAT IT CAN AND WHAT IT CANNOT TELL US. Hanno zur Loye

COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS

Limitations of Handheld XRF Instruments as Quantitative Tools for Analyzing Heavy Metal Pesticides on Organic Art Objects

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

X-Ray Fluorescence Analytical Techniques

Name Date Class ELECTRONS IN ATOMS. Standard Curriculum Core content Extension topics

Cross section, Flux, Luminosity, Scattering Rates

2 Absorbing Solar Energy

Vacuum Evaporation Recap

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)

Infrared Spectroscopy: Theory

Introduction to the EXAFS data analysis

Homework #10 (749508)

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

Molecular Spectroscopy

Transcription:

46 Spectroscopy 22(5) May 2007 Basic Fundamental Parameters in X-Ray Fluorescence The fundamental parameters approach to calibration in X-ray fluorescence is unique because it is based upon the theoretical relationship between measured X-ray intensities and the concentrations of elements in the sample. This theoretical relationship is based upon X-ray physics and the measured values of fundamental atomic parameters in the X- ray region of the electromagnetic spectrum. In this tutorial, an introduction to the means of calibration is provided based upon a simplified instrument sample geometry, thus eliminating some of the mathematical details of the traditional derivations. Volker Thomsen In X-ray fluorescence (XRF) spectrometry, the characteristic spectral line radiation emitted by the analyte is measured to determine the concentration. The theoretical relationship between these two quantities was first developed in the mid-1950s (1) and is the basis of the fundamental parameters approach to calibration in XRF, based upon the X-ray properties of the elements. Primary fluorescence is fluorescence that results from the exciting beam of X-ray photons, producing the characteristic radiation of an element or elements (Figure 1a). Some of the XRF of these elements returns to the detector, as indicated by the direction of the thin arrows. Secondary fluorescence occurs when the characteristic radiation produced in turn induces the characteristic radiation of another element in the sample (Figure 1b). Tertiary fluorescence results when the characteristic radiation of the secondary fluorescence results in the production of the characteristic radiation of a third element (Figure 1c). In this tutorial, we consider only primary fluorescence and present a somewhat simplified derivation of the relationship between the primary fluorescence intensity and the associated analyte concentration. Nevertheless, it basically follows the treatments available in standard texts on quantitative X- ray spectrometry (2 4). These works are also the sources for the data provided in the tables below. Consider the geometry of Figure 2 and note the following simplifying assumptions: A beam of monochromatic and parallel X-rays of energy (E 0 ) and intensity (I 0 ) is incident on a sample of thickness h. The angle of incidence is 90 and the detector is on the same side of the sample, adjacent to the source. (Note: This eliminates the trigonometric functions that clutter the mathematics of the derivation.) We will focus on the K X-ray emission of element i with concentration C i. The problem is approached in steps, first examining the attenuation of the incident beam and the absorption of X-ray photons in the sample volume under consideration by element i. Next we determine the production of K fluorescent radiation of this element. Finally, we examine the attenuation of this K radiation on its way back to the detector. Incident Radiation Absorption Effects The radiation intensity that reaches the sample volume element (highlighted in Figure 2) is attenuated in traveling a pathlength x. Therefore, the intensity that reaches the sample volume under consideration is as follows: I = I 0 exp ( s x) [1] Here s = mass absorption coefficient of the sample as a whole at the excitation energy (E 0 ), and = the sample density. This was addressed in a previous tutorial (5). Some of the radiation that reaches this volume element is absorbed. The fraction of radiation intensity (di) absorbed is just ( s dx). However, only some of this incoming radiation intensity is absorbed by element i. This fraction is the concentration of the element multiplied by the ratio of its mass absorption coefficient to that of the sample (C i [ i / s ]). Therefore, the absorption by element i in the volume element under consideration is given by [C i ( i / s )] ( s dx) = C i i dx [2] Then the incoming intensity that is absorbed by element i in the volume element under consideration and thus available to

May 2007 22(5) Spectroscopy 47 r K = max / min [5] Table I: Some K shell absorption jump ratios (a) (b) (c) These jump ratios have been measured and are shown in Table I. Transition Probability The probability that K radiation will be emitted rather than that of another K line is called g K and is given by g K = I(K )/[I(K ) + I(K )] or g K = [I(K )/I(K )]/{[I(K )/I(K )] + 1} [6] Some values for the K /K intensity ratio are given in Table II. Element Atomic Number J K Ti 22 0.888 V 23 0.885 Cr 24 0.882 Mn 25 0.880 Fe 26 0.877 Co 27 0.874 Ni 28 0.872 Cu 29 0.869 Zn 30 0.867 Mo 42 0.842 mined both experimentally and theoretically (Table III). Excitation Factor The product of these three probabilities gives us the excitation factor Q: Figure 1: X-ray fluorescence: (a) primary, (b) secondary, and (c) tertiary. produce K fluorescence is found by multiplying this fraction [2] by equation 1: di = I 0 (C i i dx) exp ( s x) [3] Photoelectric Emission The K fluorescence of element i is determined by the number of absorbed photons, given by equation [3], and the probabilities associated with three atomic events: The probability that a K-shell electron will be ejected rather than an L- or M- shell electron. The probability of K emission in preference to other K lines. The probability of K radiation rather than an Auger electron. Absorption Jump Ratio The first factor, the probability that a K- shell electron will be ejected rather than an L- or M-shell electron, is given by what is called the K-shell absorption jump ratio (J K ): J K = (r K 1)/r K [4] Fluorescence Yield The last factor, the probability of K radiation rather than the production of an Auger electron, is called the fluorescence yield K. Figure 4 shows the difference in fluorescence yield between the K and L spectral lines. Fluorescence yields have been deter- Q = J K g K K = [(r K 1)/r K ] g K K [7] Fluorescence Radiation Absorption The fluorescent radiation (E i ) from element i in the volume element is attenuated in traveling the path length x to the Here, r K is the K-shell absorption jump defined as the ratio of the mass absorption coefficients at the K absorption edge (Figure 3): Circle 36

48 Spectroscopy 22(5) May 2007 Table II: Some measured relative intensities x dx 600 Maximum 500 Element Atomic Number K /K Incident beam Detector Ti 22 7.52 Cr 24 7.52 Fe 26 7.46 Ni 28 7.41 Zn 30 7.41 Zr 40 5.41 Mo 42 5.18 400 300 200 100 Minimum 0 3 4 5 6 7 8 9 10 11 12 Figure 2: Excitation-sample-detector geometry. detector. The fraction transmitted is given by exp ( s,ei x) [8] Geometric Considerations The fluorescence radiation is emitted uniformly in every direction. The fraction that enters the detector is given by ( /4 ) [9] where is the solid angle defined by the detector collimator, and 4 is the total solid angle. In the geometry as shown in Figure 5, the detector (radius, a) is a distance (d) away from the source of radiation, the fluorescence produced by the element under consideration. Then the solid angle is given by = 2 [1 d/ (d 2 + a 2 )] [10] If d >> a, this is approximately = a 2 /d 2 Putting It All Together The contribution to the intensity of the primary fluorescence di i (x) of element i is given by the product of the above factors in equations 3, 7, 8, and 9. After integration, the result is I i = {[(I 0 C i i Q) ( /4 )]/( s + s,ei )}{1 exp [ h ( s + s,ei )]} [11] The steps to equation 11 are provided in the appendix. The form of I i (h) is shown in Figure 6. The primary fluorescence radiation intensity rapidly approaches a maximum h X-ray energy (kev) Figure 3: K absorption edge for chromium. with increasing sample thickness. Application 1: The Thick Sample For a thick sample, where h becomes sufficiently large, the exponential term of equation [11] becomes negligible compared to 1. Then we are left with I i = [(I 0 C i i Q) ( /4 )]/( s + s,ei ) Substituting from equation 7 for Q and rearranging terms, we get I i = (I 0 C i J K g K K )( /4 )[ i /( s + s,ei )] [12] Example: A Pure Metal Consider the example of a piece of pure iron excited by 10 mci 109 Cd radioisotope (E = 22.1 kev). (Note: The curie [Ci] is the unit of nuclear activity and equals 3.7 10 10 disintegrations per second.) Then the numbers for equation [12] are I 0 ~ 2.5 10 7 photons/s per steradian C i = 1.0 (100%) J K = 0.877 g K = 0.882 K = 0.347 i = s (at 22.1 kev) = 18.55 cm 2 /g s,ei (at 6.4 kev) = 70.46 cm 2 /g The solid angle, as computed in this example, is based upon a detector area of 25 mm 2 (a ~ 2.8 mm) and sampledetector distance of d ~ 22 mm. Then, using equation 10, we find = 0.05 steradian. Making the substitutions, we find I i = 5570 counts per second (cps). An experimental setup with geometry postulated in this derivation was not available. However, measurements made on a handheld energy-dispersive XRF spectrometer with quite different geometry, but otherwise identical to the parameters provided earlier, gave I i = 1335 cps. This number is certainly in the ballpark. Application 2: The Thin-Film Approximation For a thin film, h is very small. Now note the mathematical relationship that for x << 1, 1 exp ( x) = x. Therefore, the term {1 exp [ h ( s + s,ei )]} of equation 11 reduces to just [ h ( s + s,ei )]. We can now rewrite this equation as I i = {[(I 0 C i i Q)( / 4 )]/( s + s,ei )}[ h ( s + s,ei )]. Simplifying I i = (I 0 C i i Q) ( /4 )( h) [13] for a given experimental geometry and excitation conditions, equation 13 can be rewritten as I i = kc i. That is, the fluorescent intensity is directly proportional to the concentration of the fluorescing element. Discussion Relative Rates The result of this fundamental parameters computation as expressed in equations 11 13 is seldom used. Thus, the exercise of the pure metal example is not common. Instead, one generally creates an intensity ratio, referencing the analyte intensity to that of the pure element. In this way, many factors cancel out. Consider equation 12 for analyte element i: I i = (I 0 C i J K g K K )( /4 )[ i /( s + s,ei )] Now we write equation 12 for the same element but with concentration 100%, so that C i = 1:

May 2007 22(5) Spectroscopy 49 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 0 10 20 30 40 50 60 70 80 90 100 Atomic number (Z) K alpha L alpha Figure 4: Fluorescence yield vs. atomic number. I 100 = (I 0 J K g K K )( /4 )[ i /( s + s,ei )] The relative rate (R i = I i /I 100 ) then reduces to the simple form R i = kc i. Secondary and Tertiary Fluorescence The previous discussion considers only primary fluorescence. Matrix effect corrections come into play when we consider secondary and tertiary fluorescence. It should be noted that secondary fluorescence accounts for 50% of the observed emission in extreme cases, while tertiary fluorescence rarely reaches 2 3% and is generally considered negligible (4). While the principles remain the same, the mathematics of secondary and tertiary fluorescence corrections becomes increasingly difficult. retical calibrations in optical emission spectroscopy (OES)? Certainly a theoretical framework exists for OES (7). However, the richness of the spectra coupled with the complexity of the excitachromatic. The modeling of this excitation spectrum is empirical and provides an additional source of uncertainty. The derivation presented here is about as simplified as possible while remaining true to the basics. More advanced fundamental parameter models (as noted in the references) attempt to account for some of these sources of error. Possibility of Fundamental Parameters in Optical Emission Spectroscopy Will we ever be able to compute theo- Source Solid angle Figure 5: Source detector geometry. Detector d a Sources of Error There are several possible sources of error in this fundamental parameters approach to calibration in XRF (6): The completeness of the characteristic X-ray model The degree of uncertainty in the fundamental parameters themselves, and The model of the excitation source. The first factor refers to whether secondary and tertiary fluorescence effects have been included in the computation. (See comments in the previous section.) A significant contribution is the uncertainty in the basic X-ray physics constants used in the computation, such as the fluorescence yield and transition probabilities. In this derivation, we have assumed a monochromatic and parallel beam of incident X-rays. The monochromatic part can be approximated by radioisotope sources of X-rays. However, with X- ray tubes, the excitation is decidedly poly- Circle 37

50 Spectroscopy 22(5) May 2007 Fluorescent intensity 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 Sample thickness (arbitrary units) Figure 6: Primary fluorescence intensity as a function of sample thickness. tion sources has made progress difficult. Nevertheless, fairly recently, some work has been done using numerical modeling methods (8,9). The referenced work models the glow-discharge source, perhaps the least unwieldy of the optical emission excitation sources. The second reference provides the first explicit calculation of the optical emission spectra of argon and copper atoms. Even more recently, this same Belgian research group has done some work on modeling the laser ablation process (10). Some progress has been made in this direction. It is a development that the spectrochemist surely will watch with Table III: Some K fluorescence yields Element Atomic Number K Ti 22 0.219 V 23 0.250 Cr 24 0.282 Mn 25 0.314 Fe 26 0.347 Co 27 0.381 Ni 28 0.414 Cu 29 0.445 Zn 30 0.479 Mo 42 0.765 interest in the coming years. Appendix: Derivation of Equation 11 Combining equations 3 6, we get di i = di Q ( /4 ) exp ( s,ei x) di i = I 0 (C i i dx) exp ( s x) Q ( /4 ) exp ( s,ei x) Combining the exponential terms gives di i = (I 0 C i i Q) ( / 4 )exp [ x ( s + s,ei )]dx Integrating from x = 0 to x = h gives the primary fluorescence intensity from the incident radiation. Now, let A = (I 0 C i i Q) ( /4 ) and B = ( s + s,ei ). Then, di i = A exp( Bx)dx,or I i = A exp( Bx)dx Because exp(ax) = exp(ax)/a, we have I i = (A/B) [1 exp ( Bh)]. Therefore, I i = {[(I 0 C i i Q) ( /4 )]/[ ( s + s,ei )]}{1 exp [ h ( s + s,ei )]} I i = {[(I 0 C i i Q) ( /4 )]/( s + s,ei )}{1 exp [ h ( s + s,ei )]} [11] Acknowledgments Portions of this article were written while the author was with NITON LLC. Thanks to Debbie Schatzlein for the illustrations of Figure 1. References (1) J. Sherman, Spectrochim. Acta 7, 283 (1955). (2) E.P. Bertin, Introduction to X-Ray Spectrometric Analysis (Plenum Press, New York, 1978). (3) R. Jenkins, R.W. Gould, and D. Gedcke, Quantitative X-Ray Spectrometry, 2nd edition (Marcel Dekker, Inc., New York, 1995). (4) R. Tertian and F. Claisse, Principles of Quantitative X-Ray Fluorescence Analysis (Heyden, London, 1982). (5) V. Thomsen, D. Schatzlein, and D. Mercuro, Spectroscopy 20(9), 22 25 (2005). (6) V.Y. Borkhodoev, X-Ray Spectrometry 31(3), 209 218 (2002). (7) W.C. Martin and W.L. Wiese, Atomic Spectroscopy: A Compendium of Basic Ideas, Notation, Data, and Formulas ( National Institute of Standards and Technology, Gaithersburg, Maryland, 1999). Online at http://physics.nist.gov/pybs/atspec/. (8) A. Bogaerts, R. Gijbels, and J. Vleck, Spectrochim. Acta, Part B 53, 1517 1526 (1998). (9) A. Bogaerts and R. Gijbels, J. Anal. At. Spectrom. 13, 721 726 (1998). (10) A. Bogaerts, Z. Chen, R. Gijbels, and A. Vertes, Spectrochim. Acta, Part B 58, 1867 1893 (2003). Volker Thomsen provides consulting services from his home in Atibaia, São Paulo, Brazil. Circle 32