SYNAPT HIGH DEFINITION MASS SPECTROMETRY SYSTEM ANALYSIS OF STEROIDS

Similar documents
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Pesticide Analysis by Mass Spectrometry

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Salman Azimi, 1 Nayan S. Mistry, 2 and Michelle Wood 2. Drug Quality Control Laboratory, Supreme Council of Health, Doha, Qatar 2

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

L. Yu et al. Correspondence to: Q. Zhang

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

WATERS QUANTITATIVE ANALYSIS solutions

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

13C NMR Spectroscopy

Electrospray Ion Trap Mass Spectrometry. Introduction

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3

Correlation of the Mass Spectrometric Analysis of Heat-Treated Glutaraldehyde Preparations to Their 235nm / 280 nm UV Absorbance Ratio

Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data

CHE334 Identification of an Unknown Compound By NMR/IR/MS

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

[ Care and Use Manual ]

Micromass LCT User s Guide

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

Nuclear Magnetic Resonance Spectroscopy

XEVO TQ MS PHARMACEUTICAL APPLICATION NOTEBOOK

High Resolution LC-MS Data Output and Analysis

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

AMD Analysis & Technology AG

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

LC-MS/MS for Chromatographers

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Mass Frontier 7.0 Quick Start Guide

Protein Hit1, a novel box C/D snornp assembly factor, controls cellular concentration of protein Rsa1p by direct interaction

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Workshop IIc. Manual interpretation of MS/MS spectra. Ebbing de Jong. Center for Mass Spectrometry and Proteomics Phone (612) (612)

Waters Core Chromatography Training (2 Days)

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

ProteinPilot Report for ProteinPilot Software

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

How To Use An Acquity Qda Detector

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Nuclear Magnetic Resonance notes

ApplicationNOTE Introduction

MarkerView Software for Metabolomic and Biomarker Profiling Analysis

MS/MS analysis of Polyphenols

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

Xcalibur Versions

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry

Analyzing Small Molecules by EI and GC-MS. July 2014

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Mass Spectrometry. Overview

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Mass Frontier Version 7.0

MEDICINAL CHEMISTRY APPLICATIONS BOOK

Lipid analysis by multiple "omic" mass spectrometry platforms Where we ve been, where we are, and where we re going

PROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR)

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

SCREENING FOR DRUGS IN SERUM AND URINE BY LC/ESI/CID-MS AND MS/MS WITH LIBRARY SEARCHING *

1 st day Basic Training Course

What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model?

Overview. Purpose. Methods. Results

Proton Nuclear Magnetic Resonance Spectroscopy

A LEVEL. Type of resource H433 CHEMISTRY B. Theme: Carbon-13 MMR. October 2015

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)

Transcription:

SYNAPT HIGH DEFINITION MASS SPECTROMETRY SYSTEM ANALYSIS OF STEROIDS Jayne Kirk and Diana Uría Waters Corporation, Manchester, UK. AIM To show the utility of high efficiency ion mobility based separations combined with high resolution tandem mass spectrometry for the structure elucidation of analytically challenging compounds, such as steroids. INTRODUCTION The structural elucidation of natural and synthetic compounds is key for many scientific developments. Identifying the chemical structure that confers special properties to a substance opens the door to a meaningful product design. The application areas can range from the development of new drugs to the discovery of new flavors for foods. This is no trivial task and usually requires a high degree of expertise by the analyst to elucidate the structure. The combination of several information-rich techniques, such as NMR, X-Ray Crystallography and mass spectrometry is usually required. The normal MS approach typically starts with the proposal of elemental compositions. Using high resolution mass spectrometry, mass accuracy and isotopic pattern information can be obtained. A second step, consisting of the acquisition of fragmentation data, is then performed. Fragmentation data contain structural information and enable the assignment of a chemical structure to the analyte. While this approach can be relatively straightforward for certain compounds, some compound classes are known for being especially reluctant to fragmentation. This technical note demonstrates the utility of the Waters SYNAPT High Definition Mass Spectrometry (HDMS ) System, combined with MassFragment, a specific fragment prediction software package, for the structure elucidation of traditionally challenging compounds, such as steroids. SYNAPT HDMS System. EXPERIMENTAL Sample preparation A methyl testosterone standard solution was prepared at a 1ng μl -1 level in a mixture of acetonitrile water (50:50 acidified with 0.1% of formic acid. The sample was infused at a rate of 5 µl/minutes. MS conditions MS system: Synapt HDMS System Ionization Mode: ESI + Capillary Voltage: 3200 V Cone Voltage: 30 V Source Temp: 120 C Desolvation Temp: 200 C Desolvation Gas: 400 L/Hr V-optics: Enabled IMS Pressure: 0.54 mbar IMS wave: 5 to 15 V Trap CE: 15 V Transfer CE: 25 V Mass range: 50-1000 m/z

Acquisition and processing methods The data were acquired using Waters MassLynx TM Software, v. 4.1. This data was processed using DriftScope Software. This software package is designed to interpret four-dimensional datasets containing the retention time, drift-time, m/z, and intensity obtained from the SYNAPT High Definition Mass Spectrometry System. RESULTS AND DISCUSSION Fragmentation of steroids is frequently applied as part of a screening method, especially in regulatory analysis where confirmation of identity is generally based on three or four diagnostic product ions. 1 The identity of those diagnostic ions has been the subject of numerous scientific publications, mainly because of the complexity of the fragmentation reactions involved. 1,2 When subjected to collision induced fragmentation (CID) at low collision energy, steroids tend to fragment by the loss of small neutral molecules and at one or two specific ring positions via chargemediated reactions. 2 The fragments created in this manner are not always specific enough to allow the unambiguous identification of the compound, as there can be several isomeric species for steroids. More energetic conditions tend to produce over-fragmentation of the steroids when only small non specific fragments are visible. 3 The SYNAPT HDMS System with its patented TriWave technology provides the ability to perform Time Aligned Parallel (TAP) fragmentation. 4 In TAP analysis, the ion of interest (selected by the quadrupole) is fragmented in the TRAP T-Wave. These first generation product ions are then separated in the IMS T-Wave using high-efficiency ion mobility (diagonal series of ions in Figure 1). The separated packets of ions are then subjected to a second stage of CID in the TRANSFER T-wave to produce second generation product ions (vertical series of ions in Figure 1). The second generation product ions share the same drift-time as their respective first generation precursors, which simplifies the interpretation of these data. TAP fragmentation provides comprehensive pseudo-ms 3 information with high mass accuracy in a single experiment without the compromises in sensitivity and accuracy associated with MS 3 experiments on linear ion trap instruments. By combining this information-rich technique with the excellent mass accuracy for both parent and fragments, along with structure elucidation software tools such as MassFragment, the user can achieve detailed structural information with high confidence in a single experiment. TAP fragmentation experiments on methyl testosterone The m/z versus drift-time plot for the TAP fragmentation experiment of methyl testosterone shows clear regions of data related to the first and second generation of fragment ions (Figure 1). The signals on the diagonal contain information about the intensity and drifttime for the first generation products and the regions on the vertical axis contain the second generation fragments aligned with their respective precursor. Figure 1. TAP fragmentation for methyl testosterone. TriWave - the enabling technology within the SYNAPT HDMS System.

By selecting the appropriate regions and extracting the mass spectrometric information, independent (in Driftscope software) MS/MS spectra can be reconstructed for methyl testosterone. Each of these spectra contains key information for distinctive structural regions. (Figure 2). Section 1: Ring opening information Section 2: Block information Section 3: Neutral loss information Figure 2. Specific MS/MS spectra for methyl testosterone. The ions with longer drift-times that have been directly created from the precursor ion tend to reveal information about the possible neutral losses of the molecule. Water losses are especially easy to observe. The ions with shorter drift-times relate to the ring opening reactions and usually require high collision energy to be observed. The third region, corresponding to drift-times between 80 and 120 drift-time units relate to the sequential fragmenting of the assembly moieties of the steroid. The mass accuracy provided by the SYNAPT HDMS System can be used for the proposal of elemental compositions for the fragment ions. The proposed results are summarized in Table 1. These calculations were performed by taking the elemental composition of methyl testosterone to limit the possible elements present in the fragments and considering a maximum error of 3 ppm.

Empirical Proposed Theoretical Error mass elemental mass (mda) composition 303.2324 C 20 H 31 O 2 303.3024 0 285.2223 C 20 H 29 O 285.2218 0.5 267.2119 C 20 H 27 267.2113 0.6 227.1802 C 17 H 23 227.1800 0.3 215.1802 C 16 H 23 215.1800 0.2 211.1487 C 16 H 19 211.1487 0.1 201.1648 C 15 H 21 201.1643 0.5 189.1649 C 14 H 21 189.1643 0.6 185.1334 C 14 H 17 185.1330 0.4 109.0653 C 7 O 109.0653 0 105.0708 C 8 105.0704 0.4 97.0655 C 6 O 97.0653 0.2 91.0551 C 7 H 7 91.0548 0.3 81.0709 C 6 81.0709 0.5 9.0551 C 6 H 7 79.0548 0.3 77.0398 C 6 H 5 77.0391 0.7 MassFragment Software Once Fragment data have been acquired, the characterization of fragment ions needs to be performed. The MassFragment Software tool, integrated within the MassLynx Software significantly aids this process by automatically identifying fragment ions using a series of novel, chemically-intelligent algorithms. MassFragment Software proposes structures to the observed fragment ions by assigning a score value to each proposed structure based on the likelihood of breaking certain types of bonds. The analyst confirms the most appropriate fragment structures based on their score assignment and number of hydrogen added or removed from the structure and the exact mass error. The proposed structures for several of the most intense fragment ions for methyl testosterone are displayed in Figure 3. Table1. Mass accuracy for the proposed elemental compositions. 100 303.2324 285.2221 % 267.2117 109.0657 97.0658 227.1802 170.0970 189.1647 245.1902 304.2357 159.1176 171.1152 211.1482 91.0552 210.1289 77.0399 123.0814 185.1335 286.2261 133.1024145.1022 215.1798 229.1590 268.2141 65.0394 252.1791 274.1606 303.1630 305.4223 0 70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260 270 280 290 300 310 m/z Figure 3. MS/MS spectrum for methyl testosterone with assignments proposed by MassFragment Software.

CONCLUSION This works demonstrates that the SYNAPT HDMS System combined with the MassFragment Software package provides a powerful platform to overcome complex analytical challenges. It offers a deeper understanding of the structural elucidation of natural and synthetic compounds and fragmentation pathways and proposes valid structures for the fragment ions. Even with traditionally challenging compounds, such as steroids this approach works provides a powerful, efficient route to comprehensive structural characterization.. References 1. Williams, Kind, Houghton, and Hill. J. Mass Spectrom. 34, 2006-216, 1999. 2. Gua, Soma, and Luo. J. Am. Soc. Mass Spectrom. 17, 477-479, 2006. 3. Griffiths, Mass Spectrom Rev. 22, 81-152, 2003. 4. Yu, Castro-Perez, and Shockcor, Waters Application Note No. 720002542en. This total Waters solution includes: n SYNAPT HDMS Extra dimension of separation and high-efficiency ion mobility separation Time Aligned Parallel Fragmentation for advanced structural studies Exact mass and isotopic pattern information that facilitates the successful identification of unknowns n MassFragment Software Chemically intelligent software tool for advanced fragment structure prediction Waters is a registered trademark of Waters Corporation. SYNAPT, High Definition MS, HDMS, MassFragment, MassLynx, DriftScope, Triwave, T-Wave, and The Science of What s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2008 Waters Corporation Produced in the U.S.A. October 2008 720002802en AG-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com