A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples
|
|
|
- Ross Gregory
- 9 years ago
- Views:
Transcription
1 A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples Mary Lame, Hua Yang, Sherri Naughton, and Erin Chambers Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Simple, standardized approach to protein quantification; broadly applicable optimized digest kit eliminates method development; samples are ready for LC-MS analysis in 46 hours WATERS SOLUTIONS ProteinWorks express Direct Digest Kit (p/n ) Intact mab Check Standard (p/n ) ACQUITY UPLC Peptide BEH C 18, 3Å 1.7 μm, 2.1 mm x 15 mm Column (p/n ) ACQUITY UPLC Xevo TQ-S Mass Spectrometer ProteinWorks µelution SPE Clean-up Kit (p/n ) KEY WORDS Monoclonal antibody, infliximab, Remicade, protein quantification, express Direct Digest, ProteinWorks, adalimumab, Humira, bevacizumab, Avastin, trastuzumab, Herceptin INTRODUCTION Over the past 51 years, there has been a significant shift towards a greater of biologics in pharmaceutical pipelines. 1 However, the industry finds itself in the middle of patent expiry for many of the critical monoclonal antibody and other protein-based drugs, with patent expiration dates ranging from ~ This has resulted in a focus on protein quantification in bioanalytical labs, innovator pharma and CRO s as well as biomarker research labs. While immunoassay (IA) methods are sensitive and simple to execute, poor reagent reproducibility, lack of standardization, cross-reactivity, limited linear dynamic range, and other short-comings have led the drive to convert to LC-MS. These LC-MS workflows however, encompass a multitude of sub-segments, each having many steps. Those that are common to most workflows may include affinity purification, denaturation, reduction, alkylation, digestion, and SPE clean-up (each requiring optimization). Such traditional protein quantification protocols often require as much as a day and half for completion. Furthermore, the margin and possibility of error is significant within each individual step. There is a strong need for simpler, more standardized workflows which enable scientists to complete sample preparation and start an analytical run by mid-day. At the same time, ideally using generic, kitted methods, assay sensitivity must be high enough to accurately and precisely quantify low enough levels of the target protein to make critical decisions in discovery. The typical workflow complexity as shown in Figure 1, often leads to errors and poor reproducibility or sensitivity. In this application note, we have used the ProteinWorks express Direct Digest Kit to simplify and streamline the workflow process using the same universal protocol and reagents for all monoclonal antibody drugs tested. Infliximab, bevacizumab, trastuzumab, and adalimumab (Figures 25) in plasma were directly digested, and peptides extracted using SPE in under 4 hours total time. This enabled data to begin to be acquired the same day, with several 96-well plates being run by the following morning. 1
2 E X P E R IM E N TA L Sample description Infliximab, adalimumab, bevacizumab, or trastuzumab were spiked into human plasma. 35 µl plasma samples were then prepared for LC-MS analysis using the ProteinWorks express Direct Digest Kit and Protocol. After digestion, signature peptides were cleaned-up using the ProteinWorks µelution SPE Clean-up Kit and Protocol. Method conditions LC System: Detection: ACQUITY UPLC Xevo TQ-S Mass Spectrometer, ESI+ Column: ACQUITY UPLC Peptide BEH C 18, 3Å, 1.7 µm, 2.1 mm x 15 mm Column temp.: 55 C Sample temp.: 1 C Injection volume: 1 µl Mobile phase A:.1 formic acid in water Mobile phase B:.1 formic acid in acetonitrile Gradient: Gradient: Flow rate Time Profile (ml/min) (min) A B Curve Capillary (kv): 3 Cone (V): 3 Source offset (V): 5 Source temp. ( C): 15 Desolvation temp. ( C): 6 Cone gas flow (L/hr): 15 Desolvation gas flow (L/hr): 1 Collision gas flow (ml/min):.15 Nebulizer gas flow (Bar): 7 Protein Peptide MRM transition Cone voltage (V) Collision energy (ev) Infliximab SINSATHYAESVK 469.6> Bevacizumab FTFSLDTSK 523.3> Adalimumab APYTFGQGTK 535.3> Trastuzumab FTISADTSK 485.2> murine mab MNSLQTDDTAK (ISTD) 612.3> Table 1. MRM conditions for infliximab, adalimumab, trastuzumab, bevacizumab, and the murine mab internal standard. 2
3 Figure 1. Typical protein bioanalysis workflow. RESULTS AND DISCUSSION In a pre-clinical setting, there is an emphasis on simple, broadly applicable, generic protocols as method development time and expertise are at a premium. Multiple signature peptides were used to quantify 4 different monoclonal antibody drugs in human plasma using direct digestion and the ProteinWorks express Direct Digest Kit. For each protein, sensitivity, linearity, accuracy and precision data all met typical method validation requirements using the same broadly applicable ProteinWorks Kit. Through a direct digest of a 35 µl plasma sample, quantification limits ranged from 25 ng/ml2.5 µg/ml for the 4 monoclonal antibody-based drugs. Standard curves were linear over 3.54 orders of magnitude with the average accuracies for standard curve points typically within Summary statistics from standard curves for infliximab, adalimumab, trastuzumab, and bevacizumab are shown in Table 2 below. Protein Peptide Std. curve range Weighting Linear fit (r 2 ) Mean accuracy of all points Infliximab SINSATHYAESVK /X Bevacizumab FTFSLDTSK.55 1/X Adalimumab APYTFGQGTK /X Trastuzumab FTISADTSK /X Table 2. Linear dynamic range, weighting, and average accuracy for standard curves for infliximab, adalimumab, trastuzumab, and bevacizumab in plasma digested and extracted using the ProteinWorks express Direct Digest Kit. 3
4 Remicade Light chain [2]: DILLTQSPAILSVSPGERVSFSCRASQFVGSSIHWYQQRTNGSPRLLIKYASESMSGIPSRFSGSGSGTDFTLSINTVESEDIADYYCQQSH SWPFTFGSGTNLEVKTVAAPSVFIFPPSDEQLKSGTASVVCLLNNFYPREAKVQWKVDNALQSGNSQESVTEQDSKDSTYSLSSTLTLSKAD YEKHKVYACEVTHQGLSSPVTKSFNRGEC Remicade Heavy chain [2]: EVKLEESGGGLVQPGGSMKLSCVASGFIFSNHWMNWVRQSPEKGLEWVAEIRSKSINSATHYAESVKGRFTISRDDSKSAVYLQMNSLRTED TGVYYCSRNYYGSTYDYGQGTTLTVSXASTKGPSVFPLAPSSKSTSGGTAALGCLVKDYFPEPVTVSWNSGALTSGVHTFPAVLQSSGLYSL SSVVTVPSSSLGTQTYICNVNHKPSNTKVDKRVEPKSPKSCDKTHTCPPCPAPELLGGPSVFLFPPKPKDTLMISRTPEVTCVVVDVSHEDP EVKFNWYVDGVEVHNAKTKPREEQYNSTYRVVSVLTVLHQDWLNGKEYKCKVSNKALPAPIEKTISKAKGQPREPQVYTLPPSRDELTKNQV SLTCLVKGFYPSDIAVEWESNGQPENNYKTTPPVLDSDGSFFLYSKLTVDKSRWQQGNVFSCSVMHEALHNHYTQKSLSLSPGK Conserved region: blue variable regions: red CDR regions: green Unique signature Generic signature Van Dongen et al. 61st ASMS, MP525 Minneapolis, Minnesota, USA 9-13 June 213. Formula: C 6428 H 9912 N 1694 O 1987 S 46 Molecular Weight: ~ kd Figure 2. Structures of the monoclonal antibody-based drug infliximab (Remicade). Conserved region Surrogate Peptides Anti-HER2 Light chain DIQMTQSPSSLSASVGDRVTITCRASQDVNTAVAWYQQKPGKAPKLLIYSASFLYSGVPS RFSGSRSGTDFTLTISSLQPEDFATYYCQQHYTTPPTFGQGTKVEIKRTVAAPSVFIFPP SDEQLKSGTASVVCLLNNFYPREAKVQWKVDNALQSGNSQESVTEQDSKDSTYSLSSTLTLSKADYEKHKVYACEVTHQGLSSPVTKSFNRGEC Anti-HER2 Heavy chain EVQLVESGGGLVQPGGSLRLSCAASGFNIKDTYIHWVRQAPGKGLEWVARIYPTNGYTRY ADSVKGRFTISADTSKNTAYLQMNSLRAEDTAVYYCSRWGGDGFYAMDYWGQGTLVTVSS ASTKGPSVFPLAPSSKSTSGGTAALGCLVKDYFPEPVTVSWNSGALTSGVHTFPAVLQSS GLYSLSSVVTVPSSSLGTQTYICNVNHKPSNTKVDKKVEPPKSCDKTHTCPPCPAPELLG GPSVFLFPPKPKDTLMISRTPEVTCVVVDVSHEDPEVKFNWYVDGVEVHNAKTKPREEQY NSTYRVVSVLTVLHQDWLNGKEYKCKVSNKALPAPIEKTISKAKGQPREPQVYTLPPSRD ELTKNQVSLTCLVKGFYPSDIAVEWESNGQPENNYKTTPPVLDSDGSFFLYSKLTVDKSRWQQGNVFSCSVMHEALHNHYTQKSLSLSPGK Unique Signature Generic Signature Formula: C 647 H 112 N 1726 O 213 S 42 Molecular Weight: ~ kda (claims are 148 package insert) Figure 3. Structures of the monoclonal antibody-based drug trastuzumab (Herceptin). 4
5 Bevacizumab light chain DIQMTQSPSSLSASVGDRVTITCSASQDISNYLNWYQQKPGKAPKVLIYFTSSLHSGVPSRFSGSGSGTDFTLTIS SLQPEDFATYYCQQYSTVPWTFGQGTKVEIKRTVAAPSVFIFPPSDEQLKSGTASVVCLLNNFYPREAKVQWKVDN ALQSGNSQESVTEQDSKDSTYSLSSTLTLSKADYEKHKVYACEVTHQGLSSPVTKSFNRGEC Bevacizumab heavy chain EVQLVESGGGLVQPGGSLRLSCAASGYTFTNYGMNWVRQAPGKGLEWVGWINTYTGEPTYAADFKRRFTFSLDTSK STAYLQMNSLRAEDTAVYYCAKYPHYYGSSHWYFDVWGQGTLVTVSSASTKGPSVFPLAPSSKSTSGGTALGCLVK DYFPEPVTVSWNSGALTSGVHTFPAVLQSSGLYSLSSVVTVPSSSLGTQTYICNVNHKPSNTKVDKVEPKSCDKTH TCPPCPAPELLGGPSVFLFPPKPKDTLMISRTPEVTCVVVDVSHEDPEVKFNWYVDGVEVHNAKTKPREEQYNSTY RVVSVLTVLHQDWLNGKEYKCKVSNKALPAPIEKTISKAKGQPREPQVYTLPPS REEMTKNQVSLTCLVKGFYPSDIAVEWESNGQPENNYKTTPPVLDSDGSFFLYSKLTVDKSRWQQGNVFSCSVMHE ALHNHYTQKSLSLSPGK Unique Signature Generic Signature Drug Bank Formula: C 6538 H 134 N 1716 O 233 S 44 Molecular Weight: ~ 149. kd Figure 4. Structures of the monoclonal antibody-based drug bevacizumab (Avastin). Light chain: DIQMTQSPSSLSASVGDRVTITCRASQGIRNYLAWYQQKPGKAPKLLIYAASTLQSGVPS RFSGSGSGTDFTLTISSLQPEDVATYYCQRYNRAPYTFGQGTKVEIKRTVAAPSVFIFPP SDEQLKSGTASVVCLLNNFYPREAKVQWKVDNALQSGNSQESVTEQDSKDSTYSLSSTLTLSK ADYEKHKVYACEVTHQGLSSPVTKSFNRGEC Heavy chain: EVQLVESGGGLVQPGRSLRLSCAASGFTFDDYAMHWVRQAPGKGLEWVSAITWNSGHIDY ADSVEGRFTISRDNAKNSLYLQMNSLRAEDTAVYYCAKVSYLSTASSLDYWGQGTLVTVS SASTKGPSVFPLAPSSKSTSGGTAALGCLVKDYFPEPVTVSWNSGALTSGVHTFPAVLQS SGLYSLSSVVTVPSSSLGTQTYICNVNHKPSNTKVDKKVEPKSC DKTHTCPPCPAPELLG GPSVFLFPPKPKDTLMISRTPEVTCVVVDVSHEDPEVKFNWYVDGVEVHNAKTKPREEQY NSTYRVVSVLTVLHQDWLNGKEYKCKVSNKALPAPIEKTISKAKGQPREPQVYTLPPSRD ELTKNQVSLTCLVKGFYPSDIAVEWESNGQPENNYKTTPPVLDSDGSFFLYSKLTVDKSR WQQGNVFSCSVMHEALHNHYTQKSLSLSPGK Unique Signature Generic Signature C 6428 H 9912 N 1694 O 1987 S 46 Protein average weight Drug Bank Figure 5. Structures of the monoclonal antibody-based drug adalimumab (Humira). 5
6 At the same time, QC statistics (summarized in Table 3 below) also easily met regulatory guidelines, 3 with average precision values well under 15, and averaging in the single digits. For a typical discovery study, detection limits ~1 µg/ml are common for monoclonal antibody type drugs. Using the ProteinWorks express Direct Digest kit, these limits are easily obtained for all 4 drugs evaluated. Low QC chromatograms are shown in Figure 6 which demonstrate that adequate sensitivity is achieved with a single universal protocol and the kit. In this study, the single universal digest protocol and SPE method designed for tryptic peptides eliminated the need for discovery-stage method development. The fact that the kit was able to accurately and precisely quantify 4 monoclonal antibody drugs in plasma without the need for optimization demonstrates its broad applicability and utility in an environment where time is critical and experience with protein bioanalysis may be limited. Furthermore, the application of a kit with lot-traceable, pre-measured reagents ensures that methods may be seamlessly transferred across sites and labs, or analysts. Protein Peptide Infliximab SINSATHYAESVK Bevacizumab FTFSLDTSK Adalimumab APYTFGQGTK Trastuzumab FTISADTSK Table 3. Statistics for QC samples of infliximab, adalimumab, trastuzumab, and bevacizumab in plasma digested and extracted using the ProteinWorks express Direct Digest Kit. 6
7 1 APYTFGQGTK >91.44 (Humira APYTFGQGTK LC) 2.45e4 1 FTFSLDTSK > (Avastin FTFSLDTSK HC) 5.1e5 1 SINSATHYAESVK >63.8 (Remicade SINSATHYAESVK) 1.1e6 1 FTISADTSK >721.4 (Herceptin FTISADTSK HC) 2.79e4 Time Figure 6. Low QC chromatograms (3.5 µg/ml) for adalimumab, bevacizumab, infliximab, and trastuzumab. CONCLUSIONS The ProteinWorks express Direct Digest Kit was successfully used to purify and simultaneously quantify infliximab, adalimumab, bevacizumab, and trastuzumab from a typical set of standard curve and QC samples in human plasma. Quantification limits of 25 ng/ml to 2.5 µg/ml for each antibody were readily achieved, while maintaining excellent linearity and precision. The total sample prep time including digestion and SPE was just over 3 hours. The standardized, kit-based approach enables inexperienced users to immediately obtain meaningful data in discovery studies in order to make time sensitive and critical project decisions. References 1. Dalzell, Managed Care, October McKinsey and Company; Data Source: Evaluate Pharma, US Patent Expiration Dates. 3. FDA Guidance for Industry for Bioanalytical Method Validation, CDER. Waters, The Science of What s Possible, ACQUITY UPLC, Xevo, and MassLynx are registered trademarks of Waters Corporation. ProteinWorks is a trademark of Waters Corporation. All other trademarks are the property of their respective owners. 215 Waters Corporation. Produced in the U.S.A. November EN AG-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: F:
An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System
An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System Mary E. Lame, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA,
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
LC-MS/MS STRATEGIES FOR QUANTIFICATION OF THERAPEUTIC ANTIBODIES TO SUPPORT CLINICAL
LC-MS/MS STRATEGIES FOR QUANTIFICATION OF THERAPEUTIC ANTIBODIES TO SUPPORT CLINICAL AND PRECLINICAL STUDIES Matthew F. Ewles Bioanalytical Services Copyright 2015 Covance. All Rights Reserved Presentation
Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS
Development of Integrated Informatics Workflows for the Automated Assessment of Comparability for Antibody Drug Conjugates (ADCs) Using LC-UV and LC-UV/MS Henry Shion, Robert Birdsall, Steve Cubbedge,
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns Jonathan P. Danaceau, Kenneth J. Fountain, and Erin E. Chambers Waters
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3
Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3 Hongwei Xie and Weibin Chen Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
Discovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS
Automating the Determination of Drug-to-Antibody Ratio (DAR) of Antibody Drug Conjugates (ADCs) Based on Separation by Hydrophobic Interaction Chromatography (HIC) Robert Birdsall, Eoin Cosgrave, Henry
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
ASMS Regulated Bioanalysis Interest Group (RBIG) Workshop. Antibody-Drug Conjugates (ADC) A Complex Problem in Regulated Bioanalysis.
ASMS Regulated Bioanalysis Interest Group (RBIG) Workshop Antibody-Drug Conjugates (ADC) A Complex Problem in Regulated Bioanalysis June 17, 2014 Presiding: Dr. Keyang Xu (Genentech) and Dr. Fabio Garofolo
NUVISAN Pharma Services
NUVISAN Pharma Services CESI MS Now available! 1st CRO in Europe! At the highest levels of quality. LABORATORY SERVICES Equipment update STATE OF THE ART AT NUVISAN CESI MS Now available! 1st CRO in Europe!
Quantification of Multiple Therapeutic mabs in Serum Using microlc-esi-q-tof Mass Spectrometry
Quantification of Multiple Therapeutic mabs in Serum Using microlc-esi-q-tof Mass Spectrometry Paula Ladwig, David Barnidge, Mindy Kohlhagen, John Mills, Maria Willrich, Melissa R. Snyder and David Murray
Protein Separation Technology
[ ] Protein Separation Technology Protein Separation Technology The analysis and characterization of protein samples requires the detection of small chemical differences between large molecules. Most often
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
Salman Azimi, 1 Nayan S. Mistry, 2 and Michelle Wood 2. Drug Quality Control Laboratory, Supreme Council of Health, Doha, Qatar 2
UPLC/MS/MS for the Screening, Confirmation, and Quantification of Drugs Illegally Added to Herbal/Dietary Supplements for the Enhancement of Male Sexual Performance Salman Azimi, 1 Nayan S. Mistry, 2 and
[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS. Simplifed
[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS Simplifed The QuEChERS method (an acronym for Quick, Easy, Cheap, Effective, Rugged, and Safe) followed by dispersive solid-phase extraction
Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
How To Use An Acquity Qda Detector
Mass-Directed Isolation of a Synthetic Peptide Using the ACQUITY QDa Detector Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY QDa Detector
SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS
SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS Russell Watts 1, Allan Traynor 2, Michelle Wood 1. 1 Waters Corporation, MS Technologies Centre,
Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna
Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna Paula Hong and Pat McConville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E
MEDICINAL CHEMISTRY APPLICATIONS BOOK
MEDICINAL CHEMISTRY APPLICATIONS BOOK Introduction...3 The Role of LC and MS in Medicinal Chemistry...5 SCREENING System Management for a High Throughput Open Access UPLC/MS System Used During the Analysis
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
ApplicationNOTE Introduction
Introduction Honey, like other foods, is subject to strict quality control measures before it can be sold commercially. It is a natural product, produced solely by bees from flower and tree pollen. Human
Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )
Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 ) Andrew Aubin Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S Fast analysis of
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
Thermo Scientific Mass Spectrometric Immunoassay (MSIA)
INSTRUCTIONS FOR USE OF PRODUCTS Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Demonstration Protocol for Affinity Purification and Analysis of Human Insulin from Human Plasma Utilizing MSIA
Advances in LC-MS bioanalysis for drug research and development 液 相 质 谱 生 物 样 品 分 析 技 术 在 新 药 研 发 中 的 新 进 展
Advances in LC-MS bioanalysis for drug research and development 液 相 质 谱 生 物 样 品 分 析 技 术 在 新 药 研 发 中 的 新 进 展 张 杰 DMPK, Novartis Institutes for BioMedical Research ( 诺 华 ), New Jersey 2014 南 京 国 际 药 代 会
Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary
Application Note Increasing the activity of monoclonal antibody isoforms by MCSGP Category Matrix Method Keywords Countercurrent chromatography, FPLC Antibodies MCSGP FPLC, Biobetters, MCSGP, countercurrent
EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS
EKSIGENT EKSPERT NANOLC 4 Unmatched flexibility for low flow LC/MS EKSIGENT EKSPERT NANOLC 4 Focused ekspertise In 23, Eksigent changed the way scientists looked at nanoscale chromatography with the introduction
High Throughput Assays for Mouse Metabolic Markers: Insulin, Leptin and Adiponectin
Selen A.Stromgren Michael Tsionsky Tatiana Plissova Eli N.Glezer and Jacob N.Wohlstadter 9238 Gaither Road, Gaithersburg, MD 20877 Phone: 240.63.2522 Fax: 240.632.229 www.meso-scale.com Meso Scale Discovery,
Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS
Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS Solfrid Hegstad, PhD Department of Clinical Pharmacology St.Olavs Hospital St.Olavs Hospital Department of Clinical Pharmacology
Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
RapidFire High-throughput MS technology. Enhancing Drug Discovery Turning Mass Specs into Plate Readers
RapidFire High-throughput MS technology Enhancing Drug Discovery Turning Mass Specs into Plate Readers 1 Drug Discovery with RapidFire Target Selection High-throughput Target Screening (HTS) Lead Optimization
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.
DEVELOPMENT OF PFCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS
DEVELOPMENT O PCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS Tatsuya Ezaki 1, Nobutake Sato 1, Jun Yonekubo 1, Takeshi Nakano 2 1 Nihon Waters K.K,.1-3-12 Kitashinagawa,
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
XEVO TQ MS PHARMACEUTICAL APPLICATION NOTEBOOK
XEV TQ MS PHARMACEUTICAL APPLICATIN NTEBK Xevo TQ MS: Pharmaceutical Applications The Changing Face of LC/MS: From Experts to Users...3 Improving MS/MS Sensitivity using Xevo TQ MS with ScanWave...11 Simultaneous
Investigating Biological Variation of Liver Enzymes in Human Hepatocytes
Investigating Biological Variation of Liver Enzymes in Human Hepatocytes MS/MS ALL with SWATH Acquisition on the TripleTOF Systems Xu Wang 1, Hui Zhang 2, Christie Hunter 1 1 AB SCIEX, USA, 2 Pfizer, USA
DMPK: Experimentation & Data
DMPK: Experimentation & Data Interpretation Mingshe Zhu, Mike S. Lee, Naidong Weng, and Mark Hayward Prerequisite: Entry-level scientists with hands on experience in LC/MS as well as advanced students
HRMS in Clinical Research: from Targeted Quantification to Metabolomics
A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS Mikaël LEVI Shimadzu France Welcome Seminar Barcelona 2013 Interest in Steroids assay by LC-MS/MS Steroids measurement is essential
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity
AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS
AppNote 1/2012 Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS Oscar G. Cabrices, Fred D. Foster, John R. Stuff, Edward
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH Cristiana C. Leandro 1, Peter Hancock 1, Christian Soulier 2, Françoise Aime 2 1 Waters Corporation, Manchester,
Testing Services for Large Molecule Drug Development
Testing Services for Large Molecule Drug Development Our mission is to extend our clients capabilities by combining scientific knowledge, capacity, regulatory expertise and flexibility to provide the trusted,
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM
DRUG METABLISM Drug discovery & development solutions FR DRUG METABLISM Fast and efficient metabolite identification is critical in today s drug discovery pipeline. The goal is to achieve rapid structural
Analysis of Vegetable Oils by High Performance Liquid Chromatography
Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2 1 OpAns LLC, Durham, NC, USA; 2 Ionics Mass Spectrometry
High-Throughput 3-D Chromatography Through Ion Exchange SPE
High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit Fast, simple and cost effective analysis Many areas of biochemical research and
Innovative vs Traditional
And in the red corner, introducing the challenger NASDAQ April 4 2011 243.47 $ 1.38 $ 0.58 % 0.07 $ 0.02 $ 39.4 % Technology DVD Digital Streamline Instant Play Internet Access Unlimited Selection No Late
United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science
CLG-MG/CV2.00 Page 1 of 15 Contents A. INTRODUCTION... 2 B. EQUIPMENT... 3 C. REAGENTS AND SOLUTIONS... 4 D. STANDARDS... 5 E. SAMPLE PREPARATION... 6 F. ANALYTICAL PROCEDURE... 6 G. CONFIRMATION... 8
Patients with COPD were recruited according to the inclusion criteria; aged 40-75, current or
Online Data Supplement Materials and Methods The ECLIPSE Cohort Patients with COPD were recruited according to the inclusion criteria; aged 40-75, current or ex-smokers with >10 pack-year history; a post
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
CORESTA RECOMMENDED METHOD N 72
CORESTA RECOMMENDED METHOD N 72 DETERMINATION OF TOBACCO-SPECIFIC NITROSAMINES IN SMOKELESS TOBACCO PRODUCTS BY LC-MS/MS (February 216). INTRODUCTION The CORESTA Smokeless Tobacco Sub-Group studied various
aurora ICP-MS Innovation with Integrity The Science of Sensitivity ICP-MS
aurora ICP-MS The Science of Sensitivity Innovation with Integrity ICP-MS Powers the Business of Quantification aurora ICP-MS enhanced value package Bruker s ICP-MS is famous for its innovative and unique
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs
AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF
LC-MS/MS Solutions for the Pharmaceutical and CRO Industries TSQ Series -based Solutions for Drug Discovery and Development
m a s s s p e c t r o m e t r y LC-MS/MS Solutions for the Pharmaceutical and CRO Industries TSQ Series -based Solutions for Drug Discovery and Development Part of Thermo Fisher Scientific ACCELERATING
Mass Spectrometry Signal Calibration for Protein Quantitation
Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of
Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
A Novel Approach to Low Volume Sample Preparation
A Novel Approach to Low Volume Sample Preparation Tony Edge, Jon Bardsley, Ken Meadows, Luisa Pereira International Symposium on Chromatography Salzburg, Austria September 2014 1 The world leader in serving
Introduction. Methods. Sample Processing
PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System Olaf Scheibner, Maciej Bromirski, Thermo Fisher Scientific, Bremen, Germany
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics
MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process
- Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples
Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry
Monoclonal ntibody Fragment Separation and haracterization Using Size Exclusion hromatography oupled with Mass Spectrometry uthors Haiying hen Katherine McLaughlin Sepax Technologies, Inc. 5 Innovation
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Lean Process Improvements for the GMP Laboratory
Lean Process Improvements for the GMP Laboratory ITRODUCTIO Pharmaceutical manufacturing costs the industry approximately $90 billion each year and represents twice the expenditure of research and development.
STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080
Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien [email protected] 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary
Application Note Purifying common light-chain bispecific antibodies using MCSGP Category Matrix Method Keywords Analytes ID Continuous chromatography, biochromatography Antibodies MCSGP Bispecific antibody,
PrepTip. Reverse Phase PrepTip User Guide
PrepTip Reverse Phase PrepTip User Guide All text, photographs and illustrations are copyrighted by Harvard Apparatus, Inc. 2004. PrepTip is a trademark of Harvard Apparatus, Inc. Harvard Apparatus 84
AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT
AppNote 1/2014 Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine Fred D. Foster, John R. Stuff, Edward A. Pfannkoch GERSTEL, Inc., 701 Digital Dr. Suite J,