Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants
|
|
|
- Brook Fields
- 9 years ago
- Views:
Transcription
1 Metabolomic Profiling of Accurate Mass LC- Data to Identify Unexpected Environmental Pollutants André Schreiber 1, David Cox 1, Nadia Pace 1, Christopher Borton 2 1 AB SCIEX, Concord, ntario, Canada; 2 AB SCIEX, Golden, Colorado, USA verview There is growing concern that commonly used Pharmaceuticals and Personal Care Products (PPCP) are entering and contaminating the drinking water supply. The use of targeted quantitation of PPCP has been well established but there is an emerging trend to also screen for and identify unexpected environmental pollutants. The new AB SCIEX TripleTF 5600 LC/ was used to profile environmental samples for unexpected pollutants, to identify and characterize the chemical composition and structure of the pollutants, and to quantify the concentration in collected water samples. Introduction Pharmaceuticals and Personal Care Products (PPCP) are environmental contaminants of growing concern. In order to properly assess the effects of such compounds on our environment it is necessary to accurately monitor their presence. 1 The diversity of chemical properties of these compounds makes method development challenging. Liquid Chromatography coupled to tandem Mass Spectrometry (LC- ) is able to analyze polar, semi-volatile, and thermally labile compounds covering a wide molecular weight range, such as pesticides, antibiotics, drugs of abuse, x-ray contrast agents, drinking water disinfection by-products etc. In addition, state-of-the-art LC- instruments operated in selective Multiple Reaction Monitoring (MRM) mode, offer unmatched selectivity and sensitivity to quantify PPCP reproducibly at trace levels without time consuming and extensive sample preparation. 2 More recently there is a growing interest from environmental researchers to also screen for and identify non-targeted compounds in environmental samples, including metabolites and degradates, but also completely unexpected pollutants. The new AB SCIEX TripleTF 5600 LC/ system is capable of performing highly sensitive and fast scanning experiments to search for unknown molecular ions while also performing selective and characteristic scanning for further compound identification and, therefore, is the instrument of choice for this challenging task. General unknown screening workflows do not use a target analyte list and compound detection is not based on any a priori knowledge, including retention times and information on possible molecular and fragment ions. Therefore, acquired chromatograms are very rich in information and can easily contain thousands of ions from both any compounds present in the sample as well as from the sample matrix itself. Thus, powerful software tools are needed to explore such data to identify the unexpected compound. In metabolomic investigation the change (greek: meta ) of the fingerprint of a system is systematically studied. riginally developed to study cellular processes, such as smallmolecule metabolite profiles, statistical data analysis tools
2 are suitable to also study changes in environmental systems and, thus, to identify unexpected environmental pollutants. This article describes the use of the new AB SCIEX TripleTF 5600 system for the screening for unexpected environmental pollutants applying statistical data analysis using MarkerView Software. The detected compounds were identified based on empirical calculation of the molecular formula, interpretation of fragments and mass spectral library searching. The acquired high resolution and accurate mass spectra were further used to quantify the amounts of detected and identified environmental pollutants in a large set of environmental water samples. of Clomipramine resolution of at 315 Da Δmass < 0.5ppm of Clomipramine at 58 Da at 86 Da at 270 Da at 315 Da Method Details More than 70 water samples in different cities and countries from different type of waters, including drinking water, creeks, rivers, lakes, sea etc were collected by different scientist LC- analysis in triplicates and in a randomized order Ultra High Pressure Liquid Chromatography (UHPLC) using a Shimadzu UFLC XR system with a Phenomenex Kinetex 2.6µm (100x2.1mm) column and a gradient of water and acetonitrile with 0.1% formic acid Flow rate of 0.5mL/min and injection volume of 50µL Total run time of 15min AB SCIEX TripleTF 5600 System with Accelerator TF Analyzer and Electrospray Ionization source Continuous recalibration for EasyMass Accuracy IDA experiment with a TF- survey (accumulation time of 100ms) and up to 10 dependent scans (accumulation time of 50ms) Standardized collision energy (CE) = 35V with collision energy spread (CES) = ±15V Statistical data processing using MarkerView Software Compound identification and quantitation using PeakView Software Results and Discussion The new AB SCIEX TripleTF 5600 system with Accelerator TF Analyzer and continuous recalibration for EasyMass Accuracy delivers high resolution mass information and stable mass accuracy of ~1 ppm at fast acquisition speed in and mode. Figure 1 shows an example of and spectra of the tricyclic antidepressant Clomipramine highlighting the excellent resolution and mass accuracy over the entire mass range. Figure 1. Resolution and mass accuracy of Clomipramine in and mode using the AB SCIEX TripleTF 5600 System The metabolomic data profiling tool is Principal Components Analysis (PCA). LC- data was processed using PCA in MarkerView Software. PCA finds combinations of variables, in this case based on retention times, mass signals and intensities, that explain most of the variance present in a data set. For each principal component every sample has a score and every variable has a loading that represents its contribution to the combination. It is common to plot the scores and loadings of two principal components to visualize analytical results. Figure 2 shows the scores plot of all >70 water samples. Water samples are color coded based on sample type (i.e. green samples from wilderness areas, blue drinking water, and red samples from urban areas). The scores plot clearly differentiates samples originating from different water types. Samples from wilderness areas with extensive recreational use Drinking water produced from a river bank filtrate during flood Increasing urbanity of samples Figure 2. Scores plot of PCA of more than 70 water samples indicating trends by sample comparison
3 For example, river water samples arranged depending on urbanity at the bottom of the plot. ne drinking water sample was collected during flood season. The position of this sample in the scores plot suggests higher pollution of that drinking water produced from river bank filtrate and asks for further investigation. XIC of the unexpected 192 ion Samples collected in remote and wilderness areas arrange in the top part of the scores plot. The corresponding loading plot assists to find characteristic marker ions for this grouping. The loading plot in Figure 3 indicates mass 192 Da at 7.7min. The information of the unexpected ion characteristic for all water samples from wilderness areas was used to further explore a samples collected from Lake Zürich in Switzerland. The high resolution and accurate mass spectra of the detected molecular ion and the automatically collected spectrum is shown in Figure 4. Both spectra were further processed using the Formula Finder of PeakView Software to calculate empirical formulas for the molecular ion and detected fragment ions. The Formula Finder software shown in Figure 5 uses accurate mass information, the isotopic pattern and logic on distribution of chemical elements for this task. Figure 4. XIC of Da +/- 50m Da with high resolution accurate mass (left) and spectra (right)? Figure 3. Loadings plot of PCA of characteristic mass-retention timepairs to separate studied water samples Figure 5. Formula Finder software to empirically calculate the formula of the molecular ion and fragment ions of the unexpected environmental pollutant (192 Da at 7.7min)
4 The information gained from data processing was used to characterize the structure of the identified environmental pollutant (Figure 6) and identify the molecule as DEET (N,N- Diethyl-m-toluamide). H 3 C DEET N C 12 H 17 N+H + 192amu 2.4µg/L 2.6µg/L 3.3µg/L 2.5µg/L 3.6µg/L H C 3 C + H 3 C C + NH H N CH 2 C 8 H amu C 7 H amu C 6 H amu C 8 H 8 N + 118amu CH + CH + + C 5 H amu N CH 2 C 5 H 9 N+H + 192amu C 3 H 5 N+H + 72amu Figure 6. Structural characterization of fragments of the unexpected environmental pollutant After purchase of an analytical standard the identified environmental pollutant was quantified in the collected environmental samples using the XIC Manager in PeakView Software. The profile plot of DEET (Figure 8) shows that this compound is widespread in the environment and can even be found in remote areas with high recreational use, such as Algonquin Provincial Park, a popular canoeing destination in ntario (Canada), in the Rocky Mountains of Colorado (US), and in the Alps (Switzerland). Figure 8. Quantitation of DEET in environmental water sample Conclusion The new AB SCIEX TripleTF 5600 LC/ System and powerful software tools were used to screen for nontargeted environmental pollutants. More than 70 samples collected all over the world were processed in MarkerView Software using metabolomic profiling and statistical analysis (Principal Components Analysis PCA) to detect unexpected compounds. High resolution and accurate mass information was used to empirically calculate the formula of molecular ions and fragment ions using Formula Finder of PeakView Software. After compound identification the unexpected pollutant was quantified in all collected water samples. Finally, the collected full scan data can be further processed retrospectively to identify more unexpected environmental pollutants. TIC XIC of DEET Figure 7. Quantitation of DEET in environmental water samples using the web browser of the XIC Manager of PeakView Software
5 Acknowledgements The authors would like to thank colleagues of AB SCIEX for collecting water samples all over the world during their vacation or while attending conferences. References 1 D. Kolpin etal.: Pharmaceuticals, hormones, and other organic wastewater contaminants in U.S. streams, : A national reconnaissance Environ. Sci. Technol. 36 (2002) A. Schreiber et al.: Accelerated LC/ for the Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Water Samples Pittcon (March 2009) Chicago, IL (USA) For Research Use nly. Not for use in diagnostic procedures AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Publication Headquarters International Sales 353 Hatch Drive Foster City CA USA For our office locations please call the division Phone headquarters or refer to our website at
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM
DRUG METABLISM Drug discovery & development solutions FR DRUG METABLISM Fast and efficient metabolite identification is critical in today s drug discovery pipeline. The goal is to achieve rapid structural
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
Authenticity Assessment of Fruit Juices using LC-MS/MS and Metabolomic Data Processing
Authenticity Assessment of Fruit Juices using LC-MS/MS and Metabolomic Data Processing Lukas Vaclavik 1, Ondrej Lacina 1, André Schreiber 2, and Jana Hajslova 1 1 Institute of Chemical Technology, Prague
A Streamlined Workflow for Untargeted Metabolomics
A Streamlined Workflow for Untargeted Metabolomics Employing XCMS plus, a Simultaneous Data Processing and Metabolite Identification Software Package for Rapid Untargeted Metabolite Screening Baljit K.
MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis
MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis Overview MarkerView software is a novel program designed for metabolomics applications and biomarker profiling workflows 1. Using
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP and TripleTOF 5600 System Technology Targeted peptide quantification is a rapidly growing application
Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
ProteinPilot Report for ProteinPilot Software
ProteinPilot Report for ProteinPilot Software Detailed Analysis of Protein Identification / Quantitation Results Automatically Sean L Seymour, Christie Hunter SCIEX, USA Pow erful mass spectrometers like
Investigating Biological Variation of Liver Enzymes in Human Hepatocytes
Investigating Biological Variation of Liver Enzymes in Human Hepatocytes MS/MS ALL with SWATH Acquisition on the TripleTOF Systems Xu Wang 1, Hui Zhang 2, Christie Hunter 1 1 AB SCIEX, USA, 2 Pfizer, USA
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS
Application note Explosives by Direct Injection LC/MS/MS Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS API 5000 LC/MS/MS System
HRMS in Clinical Research: from Targeted Quantification to Metabolomics
A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier
AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs
AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report
PRODUCT BULLETIN Cliquid ChemoView 3.0 Software for Routine Screening and Quantitation Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report KEY FEATURES Secure user login that
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
NUVISAN Pharma Services
NUVISAN Pharma Services CESI MS Now available! 1st CRO in Europe! At the highest levels of quality. LABORATORY SERVICES Equipment update STATE OF THE ART AT NUVISAN CESI MS Now available! 1st CRO in Europe!
Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform
Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform Remco Swart Pittcon March 8-12, 2015 1 The world leader in serving science Outline Thermo Scientific Vanquish UHPLC
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity
Discovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Waters Core Chromatography Training (2 Days)
2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.
Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC
Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC Abstract: High performance liquid chromatography (HPLC) was used to determine the identity and concentrations of food dyes
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests AB SCIEX TOF/TOF 5800 System with DynamicExit Algorithm and ProteinPilot Software for Robust Protein
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS
amazon SL Setting New Standards in Performance, Simplicity and Value Innovation with Integrity Ion Trap Best-In-Class Ion Trap Mass Spectrometer for Routine Analysis The amazon SL entry-level system is
Authenticity Testing of Liquor Samples using LC-MS/MS and Statistical Data Processing
Authenticit Testing of Liquor Samples using LC-MS/MS and Statistical Data Processing André Schreiber 1, Christopher Borton 2 ; and Anna Marques 3 1 AB SCIEX Concord, ON (Canada), 2 AB SCIEX Golden, CO
Mass Frontier 7.0 Quick Start Guide
Mass Frontier 7.0 Quick Start Guide The topics in this guide briefly step you through key features of the Mass Frontier application. Editing a Structure Working with Spectral Trees Building a Library Predicting
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
Thermo Scientific ExactFinder Software
mass spectrometry Thermo Scientific ExactFinder Software Workflow software for routine targeted and general unknown screening Unifies qualitative and quantitative high-resolution accurate mass workflows
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow Abstract There is high demand in clinical research and forensic toxicology for comprehensive,
Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
Bruker ToxScreener TM A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS ToxScreener - Get the Complete Picture Forensic laboratories are frequently required
AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT
AppNote 6/2003 Analysis of Flavors using a Mass Spectral Based Chemical Sensor Vanessa R. Kinton Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA Kevin L. Goodner USDA, Citrus & Subtropical
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH Cristiana C. Leandro 1, Peter Hancock 1, Christian Soulier 2, Françoise Aime 2 1 Waters Corporation, Manchester,
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
S olid Ph ase Extraction (SP E) Method:
Solid Phase Extraction of US EPA 535 for Chloroacetanilide and Acetamide Degradates in Drinking Water Samples Prior to Liquid Chromatography/ Tandem Mass Spectrometry Toni Hofhine, Horizon Technology,
Analyst 1.6.1 Software
MAX cover image size 4.8 inches X 7.8 inches In Body page mode, draw a graphic frame to match this section and then go to the next step. Delete this text from the Front Cover Master page. View>Master Pages,
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
Metabolomics Software Tools. Xiuxia Du, Paul Benton, Stephen Barnes
Metabolomics Software Tools Xiuxia Du, Paul Benton, Stephen Barnes Outline 2 Introduction Software Tools for LC-MS metabolomics Software Tools for GC-MS metabolomics Software Tools for Statistical Analysis
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Analyst 1.6 Software. Getting Started Guide
Analyst 1.6 Software Getting Started Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment. This
New drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
A Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS
Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS AGILENT 6000 SERIES LC/MS SOLUTIONS UNRIVALED ACHIEVEMENT AT EVERY LEVEL The Agilent 6000 Series of LC/MS instruments
Analyst 1.6 Software. Software Reference Guide
Analyst 1.6 Software Software Reference Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment.
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
MassHunter for Agilent GC/MS & GC/MS/MS
MassHunter for Agilent GC/MS & GC/MS/MS Next Generation Data Analysis Software Presented by : Terry Harper GC/MS Product Specialist 1 Outline of Topics Topic 1: Introduction to MassHunter Topic 2: Data
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS
EKSIGENT EKSPERT NANOLC 4 Unmatched flexibility for low flow LC/MS EKSIGENT EKSPERT NANOLC 4 Focused ekspertise In 23, Eksigent changed the way scientists looked at nanoscale chromatography with the introduction
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns Jonathan P. Danaceau, Kenneth J. Fountain, and Erin E. Chambers Waters
Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference
Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas
High Resolution LC-MS Data Output and Analysis
High Resolution LC-MS Data Output and Analysis Your Sample LC/MS Result Sample Sample Efficient Separation Preparation Introduction Gradient (column) Today Data (computer) LC- MS Ions Detection Ions Separation
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model?
What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model? authors M. David Ho 1, Bob Xiong 1, S. Stellar 2, J. Proctor 2, J. Silva 2, H.K. Lim 2, Patrick Bennett 1, and Lily Li 1 Tandem Labs,
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
High Sensitivity MS Determination of Carryover in a New Autosampler Design
High Sensitivity MS Determination of Carryover in a New Autosampler Design 1 William Hedgepeth*, Sue Steinike Shimadzu Scientific Instruments, Columbia, MD USA Mauro Aiello, Robert Ellis, Andre Schreiber,
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System Olaf Scheibner, Maciej Bromirski, Thermo Fisher Scientific, Bremen, Germany
Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.
Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class
Agilent LCMS products and the application strategies in veterinary drug analysis
Agilent LCMS products and the application strategies in veterinary drug analysis Experience A New Level of Confidence Ma Xin 马 鑫 LCMS Application Engineer Internal Sales Training - The new 6495 QQQ LC/MS
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
Lipid analysis by multiple "omic" mass spectrometry platforms Where we ve been, where we are, and where we re going
Lipid analysis by multiple "omic" mass spectrometry platforms Where we ve been, where we are, and where we re going Jeff McDonald Department of Molecular Genetics UT Southwestern Medical Center Dallas,
Alignment and Preprocessing for Data Analysis
Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS
Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS Thorsten Bernsmann, Peter Fürst Chemical and Veterinary Analytical Institute Münsterland-Emscher-Lippe Joseph-König-Str. 40,
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method
Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Introduction During the last decade, the complexity of samples
Integrated Data Mining Strategy for Effective Metabolomic Data Analysis
The First International Symposium on Optimization and Systems Biology (OSB 07) Beijing, China, August 8 10, 2007 Copyright 2007 ORSC & APORC pp. 45 51 Integrated Data Mining Strategy for Effective Metabolomic
Urine Drug Testing. Why drug test? Set the Standard!
Urine Drug Testing Urine drug testing is a valuable tool for the physician when making treatment decisions. Patients and physicians benefit from safe and effective treatment of pain but there are always
Introduction. Methods. Sample Processing
PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko
Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories
Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories Matthew Clabaugh, Market Development AACC Workshop St. Louis, MO September 17,18 2013 Factors Influencing LCMSMS Moving into
