5.4 CT Number Accuracy and Noise

Similar documents
How To Improve Your Ct Image Quality

Purchasing a cardiac CT scanner: What the radiologist needs to know

Fundamentals of Cone-Beam CT Imaging

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

American College of Radiology CT Accreditation Program. Testing Instructions

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Multi-slice Helical CT Scanning of the Chest

Lecture 14. Point Spread Function (PSF)

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

An iterative approach to the beam hardening correction in cone beam CT

Physics testing of image detectors

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Assessing Radiation Dose: How to Do It Right

GE Healthcare. Revolution EVO. More than just high tech. Higher purpose.

MEDICAL IMAGING 2nd Part Computed Tomography

State-of-the-Art Technology in Cardiac CT

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

GE Healthcare. Great treasures in small places. BrightSpeed Elite

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA.

Musculoskeletal MRI Technical Considerations

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

DENTAL Cone beam 3D X-RAY SYSTEM with

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.

IMAGE PROCESSING AND DATA ANALYSIS IN COMPUTED TOMOGRAPHY

Patient Prep Information

High Quality Image Magnification using Cross-Scale Self-Similarity

A d v a n c e d 1 2 8

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

Digital Mammography Update: Design and Characteristics of Current Systems

DICOM Correction Item

The Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO

Data. microcat +SPECT

Phantom Test Guidance

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan

Digital image processing

3D Accuitomo FPD XYZ Slice View Tomograph. Super-High Resolution Images of Region of Interest

5 Factors Affecting the Signal-to-Noise Ratio

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

The disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved.

Measuring Line Edge Roughness: Fluctuations in Uncertainty

SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems

Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction

ENG4BF3 Medical Image Processing. Image Visualization

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Spectral Line II. G ij (t) are calibrated as in chapter 5. To calibrated B ij (ν), observe a bright source that is known to be spectrally flat

IAEA HUMAN HEALTH SERIES

Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota

Canny Edge Detection

The File-Card-Browser View for Breast DCE-MRI Data

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

Proton tracking for medical imaging and dosimetry

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

Using Optech LMS to Calibrate Survey Data Without Ground Control Points

INTRODUCTION. A. Purpose

3D Scanner using Line Laser. 1. Introduction. 2. Theory

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

GE Healthcare. Discovery * CT750 HD Great care by design

High Performance GPU-based Preprocessing for Time-of-Flight Imaging in Medical Applications

SITE IMAGING MANUAL ACRIN 6698

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Introduction to CCDs and CCD Data Calibration

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!

ECE 533 Project Report Ashish Dhawan Aditi R. Ganesan

Endoscope Optics. Chapter Introduction

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

Automation of a CT Acquisition: A System-based User-Support. 1. Introduction. Bärbel KRATZ 1, Frank HEROLD 1, Malte KURFISS 1

Concepts for High-Resolution Low-Dose CT of the Breast

C1 Medical Imaging Modalities & Characteristics Linwei Wang

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Transcription:

186 Chapter 5 5.4 CT Number Accuracy and Noise In many clinical practices, radiologists rely on the value of measured CT numbers to differentiate healthy tissue from disease pathology. Although this practice is not recommended by most CT manufacturers (unless the difference between the healthy and diseased tissues is large), it underlines the importance of producing accurate CT numbers. Two aspects factor into CT number accuracy: CT number consistency and uniformity. CT number consistency dictates that if the same phantom is scanned with different slice thicknesses, at different times, or in the presence of other objects, the CT numbers of the reconstructed phantom should not be affected. CT number uniformity dictates that for a uniform phantom, the CT number measurement should not change with the location of the selected ROI or with the phantom position relative to the iso-center of the scanner. For illustration, Fig. 5.30 shows a reconstructed 20-cm water phantom. The average CT numbers in two ROI locations should be identical. Because of the effects of beam hardening, scatter, CT system stability, and many other factors, both CT number consistency and uniformity can be maintained within only a reasonable range. As long as radiologists understand the system limitations and the factors that influence the performance, the pitfalls of using absolute CT numbers for diagnosis can be avoided. It is important to point out that the CT number may change significantly with different reconstruction algorithms. Most kernels used for reconstruction Figure 5.30 Water phantom for CT number uniformity and noise measurement.

Key Performance Parameters of the CT Scanner 187 are designed for specific clinical applications and should not be used indiscriminately for all applications. For example, the bone or lung algorithm on HiSpeed or LightSpeed scanners is designed to enhance the visibility of fine structures of bony objects. It was designed with inherent edge enhancement characteristics and not intended to maintain the CT number accuracy of small objects. Consequently, some of the lung nodules reconstructed with the bone or lung algorithm appear to be falsely calcified, since the CT number of the nodules is artificially elevated. Figures 5.31(a) (c) depict a chest scan reconstructed with, respectively, the standard algorithm, the lung algorithm, and the difference image. The CT number of many lung nodules and even some large vessels is significantly higher for the lung algorithm, as indicated by the bright structures in the difference image. With the recent advancements in CT technology (to be discussed in a later chapter), the CT number uniformity requirement needs to be extended from 2D axial images (x-y plane) to volumetric 3D images (x-y-z), since state-ofthe-art CT scanners cover a large object volume in a single rotation. Similar to the CT number uniformity and accuracy test for the x-y plane where a uniform phantom, such as 20-cm water, is used, CT number along the z dimension can be tested with phantoms that are constant in z. As an illustration, Fig. 5.32(a) depicts a cylindrical phantom with uniform iodinerod inserts. To test the CT number uniformity, coronal images are generated, and the average CT numbers are measured at different ROI locations along the phantom. Because the phantom is constant in z, the CT number difference reflects the scanner s performance in maintaining CT number accuracy. Figure 5.32(b) depicts coronal/sagittal images formed across the diameters of Figure 5.31 Patient chest scan reconstructed with different algorithms (WW ¼ 1000): (a) standard algorithm, (b) lung algorithm, and (c) difference image.

188 Chapter 5 Figure 5.32 (a) Cylindrical phantom with different concentrations of iodine inserts. (b) Coronal images acquired on a wide-cone CT scanner with different calibrations. (Top): Coronal image of the phantom at the location shown by the dashed lines acquired on a widecone CT scanner without proper calibration. The CT number difference between the top and bottom ROIs is about 20 HU. (Bottom): Coronal image of the phantom at the location shown by the dashed lines acquired on a wide-cone CT scanner with proper calibration. The CT number difference between the top and bottom ROIs is about 2 HU. the rods, with the top portion representing reformatted images without the proper calibration, and the bottom portion representing reformatted images with the proper calibration. It is clear that the bottom image performs better than the top image in terms of CT number uniformity. Another important performance parameter is image noise, which is typically measured on uniform phantoms. To perform noise measurement, the standard deviation s within an ROI of a reconstructed image f i,j is calculated by sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi Pi, j ROI s ¼ ðf i,j f Þ 2, (5.7) N 1 where i and j are indexes of the 2D image, N is the total number of pixels inside the ROI, and f is the average pixel intensity that is calculated by f ¼ 1 N X i, j ROI f i, j : (5.8)

Key Performance Parameters of the CT Scanner 189 Note that in both equations, the summation is 2D over the ROI. For reliability, several ROIs are often used, and the average value of the measured standard deviations is reported. Ideally, noise measurements should be performed over an ensemble of reconstructed CT images of the same phantom under identical scan conditions to remove the impact of the correlated nature of the CT image noise and the potential biases due to image artifacts resulting from nonperfect system calibration. In such calculations, the summation over ROI shown in Eqs. (5.7) and (5.8) is replaced by the summation over the corresponding ROI of the difference image. Since the difference image is generated by subtracting two images acquired and reconstructed under identical conditions, image artifacts are subtracted out. The resulting standard deviation, of course, needs to be scaled down to reflect the noise level in the original image. In general, three major sources contribute to the noise in an image. The first source is quantum noise, determined by the x-ray flux or the number of x-ray photons that are detected. This source is influenced by two main factors: the scanning technique (x-ray tube voltage, tube current, slice thickness, scan speed, helical pitch, etc.), and the scanner efficiency [detector quantum detection efficiency (QDE), detector geometric detection efficiency (GDE), umbra-penumbra ratio, etc.]. The scanning technique determines the number of x-ray photons that reach the patient, and the scanner efficiency determines the percentage of the x-ray photons exiting the patient that convert to useful signals. For CT operators, the choices are limited to the scanning protocols. To reduce noise in an image, one can increase the x-ray tube current, x-ray tube voltage, or slice thickness, or reduce the scan speed or helical pitch. The operator must understand the tradeoffs for each choice. For example, although an increased tube voltage helps to reduce noise under an equivalent kw (the product of the tube current and the tube voltage) condition, LCD is generally reduced. Similarly, increased slice thickness may result in a degraded 3D image quality and increased partial-volume effect. A slower scan speed could lead to increased patient motion artifacts and reduced organ coverage. An increased tube current leads to increased patient dose and increased tube loading. As long as the tradeoffs are well understood, these options can be used effectively to combat noise. The second source that contributes to the noise in an image is the inherent physical limitations of the system. These include the electronic noise in the detector photodiode, electronic noise in the data acquisition system, x-ray translucency of the scanned object, scattered radiation, and many other factors. For CT operators, options to reduce noise in this category are limited. The third noise-contributing factor is the image generation process. This process can be further divided into different areas: reconstruction algorithms, reconstruction parameters, and effectiveness of calibration. The impact of different classes of reconstruction algorithms on noise was discussed in

190 Chapter 5 Chapter 3 (e.g., see Fig. 3.50) and will not be repeated here. For the same reconstruction algorithm, such as FBP, the reconstruction parameters include the selection of different reconstruction filter kernels, reconstruction FOV, image matrix size, and postprocessing techniques. In general, a high-resolution reconstruction kernel produces an increased noise level. This is mainly because these kernels preserve or enhance high-frequency contents in the projection. Unfortunately, most noise presents itself as high-frequency signals. An example of the filter kernel selection on noise is shown in Fig. 5.15. A few words should be said about postprocessing techniques. Many image-filtering techniques have been developed over the past few decades for noise suppression. To be effective, these techniques need to not only reduce noise and preserve fine structures in the original image, but also to maintain natural-looking noise texture. Often, techniques are rejected by radiologists because the filtered images are too artificial. In recent years, many advances have been made in this area; a more detailed discussion can be found in Chapter 11 (e.g., Fig. 11.26). As was discussed in Chapter 3, the calibration or preprocessing techniques used in CT to condition the collected data are not perfect. The residual error often manifests itself as artifacts of small magnitude. These artifacts sometimes cannot be visually detected. However, they do influence the standard deviation measurement and consequently should be considered as part of the noise source. Although standard deviation is the most straightforward way of measuring noise in an image and correlates fairly well with visual observation, it also has many limitations. As an illustration, Fig. 5.33 shows three images with different types of noise. Figure 5.33(a) shows the noise image of a reconstructed water phantom with the standard kernel. In a similar fashion, the noise image with the bone kernel was scaled, as shown in Fig. 5.33(b). The scaling factor was selected so that Figs. 5.33(a) and (b) have an identical Figure 5.33 Different types of noises with identical standard deviation (s ¼ 6.8). (a) Noise in the reconstructed water phantom with the standard algorithm. (b) Noise in the reconstructed water phantom with the bone algorithm. (c) Additive noise with normal distribution.