DIGITAL + ANALOGUE HIGH VOLTAGE INSULATION TESTER

Similar documents
UT202A Operating Manual. Contents

Model UT33A: OPERATING MANUAL Table of Contents (1)

Model UT201/202: OPERATING MANUAL. Table of Contents

User Guide. 4 wire Earth Resistance Tester Model GRT300

User Guide. Model Insulation Tester / Megohmmeter

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual

Ground Resistance Clamp On Tester

Manual Ranging MultiMeter

Model UT33B/C/D: OPERATING MANUAL Table of Contents (1)

AutoRanging Digital MultiMeter

HVAC Clamp Meter. Users Manual

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

Ground Resistance Clamp-On Tester

10 Multimeter. Users Manual

MEGGER Insulation & Low Resistance Tester

323/324/325. Calibration Manual. Clamp Meter

User s manual FLIR VP50/VP52 Non-contact AC voltage detector

Hygro-Thermometer + InfraRed Thermometer Model RH101

162 CB CABLE TRACER. Filter Probe & Tone Generator INSTRUCTION MANUAL

User's Guide. True RMS Industrial Multimeter

i410/i1010 AC/DC Current Clamp

Model UT58A/B/C: OPERATING MANUAL

Table of Contents. Page

Model UT713 OPERATING MANUAL

Auto-ranging Digital Multimeter INSTRUCTION MANUAL

Model 1756 Test Lead Kit

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

Instructions A Amp AC/DC Current Probe

User s Manual Before using the inverter, you need to read and save the safety instructions.

Drayton Digistat +2RF/+3RF

Model UT60A: OPERATING MANUAL. Table of Contents. Page

7 FUNCTION DIGITAL MULTIMETER

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN

12 Volt 30 Amp Digital Solar Charge Controller

DIGITAL MULTIMETER OPERATION MANUAL ITEM DIGITAL MULTIMETER CONTENTS

Model UT61A/61B/61C/61D/61E: OPERATING MANUAL. Table of Contents

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

Seven function digital multimeter

EzFlex Combustible Gas Leak Detector Model EZ40

1. SAFETY INFORMATION

T+ and T+ PRO Electrical Tester

TMS TANK MANAGEMENT SYSTEM

Instructions: Part I. Charge the Power Bank. Method 1. Charge the Power Bank via Computer USB Terminal.

INSTRUCTION MANUAL. NC-500 NETcat Pro Wiring Tester

USER S MANUAL. MaxPower UPS. Uninterruptible Power System 28-2MAXPO0018

INSTRUCTION MANUAL IPM Phase Power Analyser EN FR IT DE ES

Programmable Room Thermostat 7 Day (5-2 Day) Models: / Power Supply: Battery / Mains

Digital Ground Resistance Tester. Models 4620 & Digital Ground Resistance Tester

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102

RIGOL. Quick Guide. DS1000CA Series Oscilloscope. Aug RIGOL Technologies, Inc.

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

Microwave Meter. Instruction Manual

How To Use A Cdm250 Digital Multimeter

Operating Manual for the Electronic Built-in Interval Timer. Micro II (Countdown Timer)

How To Use A Power Supply Unit (Upu)

Intelligent Digital Multimeters

Thermistor Thermometer MODEL NO

Multi Media Cable Tester TM-901N/TM-902/TM-903

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12

Integrating Sound Level Meter and Datalogger

RS DIGITAL THERMOMETER

MIT510/2, MIT520/2 and MIT1020/2

19 LED Tube Controller ORDERCODE 41003

SOUND LEVEL METER 94405

T0118 T2118 T3118. Instruction Manual

FUJI DIGITAL SOUND DETECTOR FSB-8D INSTRUCTION MANUAL

NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

Online UPS. PowerWalker VFI 1000R/1U

DSM155 Digital LED Light Meter User s Manual DSM155

Guidance for upgrading to GB (IEC : 2005, MOD)

RS Stock No Instruction Manual RS-1340 Hot Wire Anemometer

Installation. Smart-UPS VT MGE Galaxy Maintenance Bypass Panel kva 400 V

Thermocouple Thermometer Models

When you switch off your system, or mute the sound, the red indicator light appears immediately, indicating that the subwoofer is not in use.

User s Manual AURORA 1.2K/2.2K

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

Light Meter (Foot Candles)

T7560A,B,C Digital Wall Module

PowlVac Vacuum Integrity Tester

TIG INVERTER INSTRUCTION MANUAL

INSTRUCTION MANUAL PLEASE READ ALL THE INSTRUCTIONS COMPLETELY BEFORE USE AND SAVE THIS MANUAL FOR FUTURE REFERENCE

AIR-CONDITIONER SPLIT TYPE

Cable Tester Pro ORDERCODE D1909

EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM

BC-5000 OPERATIONS MANUAL BATTERY CAPACITY TESTER COFKO LLC.

ADM1TE 5/30A DIN rail single phase two wire energy meter

Infrarot-Bewegungsmelder IP44

How To Read A Pipe From A Sensor

MOUNTING AND OPERATING INSTRUCTIONS

Part I - Installation

DRM75A 230V 20/100A DIN rail single phase two wire energy meter

Alarm Clock USER GUIDE

Detector transparent with Color Inserts. FAA 500 TR P Trim Ring transparent with Color Inserts. FCA 500 / FCA 500 E Detector Bases

Mini Power Bank and Charger

Model UT71A/B OPERATING MANUAL

What is a multimeter?

5000V Graphical Megohmmeter Model 5070

Transcription:

DIGITAL + ANALOGUE HIGH VOLTAGE INSULATION TESTER

CONTENTS CONTENTS 1. SAFETY INFORMATION 1 1.1 PRELIMINARY 1 1.2 DURING USE 2 1.3 SYMBOLS 3 1.4 MAINTENANCE 4 2. DESCRIPTION 6 2.1 NAMES OF COMPONENTS 7 2.2 SWITCH AND BUTTONS ELUCIDATE 9 3. SPECIFICATIONS 10 3.1 GENERAL SPECIFICATIONS 10 3.2 ELECTRICAL SPECIFICATIONS 11 4. OPERATING INSTRUCTION 12

CONTENTS 4.1 ALARM 12 4.2 PREPARATION FOR MEASUREMENT 12 4.3 ENCLOSED BATTERY CHECK 13 4.4 INSULATION RESISTANCE MEASURING 14 4.5 CONTINUING MEASURING 16 4.6 USE OF GUARD TERMINAL 18 5. MAINTENANCE 20 5.1 BATTERY REPLACEMENT 20 5.2 TEST LINE AND TEST CLIPS REPLACEMENT 21 6. ACCESSORIES 22

SAFETY INFORMATION WARNING Read the operating instruction thoroughly and completely before using. Ray particular attention to warning, follow all safety operating procedures in order to avoid electrical shock and any damage. 1. SAFETY INFORMATION This meter has been designed in general to comply with IEC 348 and according to IEC-1010 concerning about electronic measuring instruments with an overvoltage category (5000V CAT ) and pollution 2. Follow all safety and operating instructions to ensure that the meter is used safely and is kept in good operating condition. Pay attention to cautions and warnings, which will inform you of potentially dangerous procedures. 1.1 PRELIMINARY WARNING To ensure safe operation, please follow the directions in this section carefully. - 1 -

SAFETY INFORMATION 1.1.1 When using the meter, the user must observe all normal safety rules concerning: Protection against the dangers of electrical current Protection of the meter against misuse 1.1.2 When the meter is delivered, check that it has not been damaged in transit. 1.1.3 When poor condition under harsh preservation or shipping conditions caused, inspect and confirm this meter without delay. 1.1.4 test line or test clips must be in good condition. Before using verify that the insulation on test line or test clips is not damaged and/or the leads wire is not exposed. 1.1.5 Full compliance with safety standards can be guaranteed only if used with test line or test clips supplied. If necessary, they must be replaced with the same model or same electric ratings. 1.1.6 This meter must only be used by a competent trained person and in strict accordance with the operator s manual. 1.2 DURING USE 1.2.1 Before use, ensuring that there is no any components damaged by high voltage in the measured circuit. - 2 -

SAFETY INFORMATION 1.2.2 During operating, once press the measuring button, you should be more careful to the existence of high voltage between the line and the other two connections terminal. 1.2.3 Press the button for measurement after the meter is connected with the circuit. 1.2.4 Keep fingers always behind the probe and the safeguard of the lead while measuring. 1.2.5 Do not attempt to take any measurement if the meter, the probe, the lead or the user s hand is wet. 1.2.6 Never use the meter under the condition of the explosive air, steam or dirt. 1.2.7 Never connect the meter leads across on live circuits. Ensure all circuits are de-energized before measuring. 1.2.8 If any faults or abnormalities are observed, the meter can not be used any more and it has to be checked out. 1.2.9 Never use the meter unless the rear case is in place and fastened fully. 1.2.10 Please do not store or use meter in areas exposed to direct sunlight, high temperature, humidity or condensation. 1.3 SYMBOLS Important safety information, refer to the operating manual. - 3 -

SAFETY INFORMATION risky voltage Double insulation(protection class ). LINE Test Line Terminal GUARD Guard Terminal EARTH Earth Terminal Low Battery 1.4 MAINTENANCE 1.4.1 Please do not attempt to adjust or repair the meter by removing the rear case while voltage is being applied. A technician who fully understands danger involved should only carry out such actions. 1.4.2 Before opening the battery cover of the meter, ensuring that the function switch is on the OFF position, the measuring button is loosened, and always disconnect test line or test clips from all sources of electric current. 1.4.3 In the course of keeping and using, the meter can not be excessively vibrated. 1.4.4 Always set the function switch to the OFF position when the meter is not in use. 1.4.5 If the meter is to be stored for a long period of time, the batteries should be removed to prevent - 4 -

SAFETY INFORMATION damage to the unit. 1.4.6 Do not use abrasives or solvents on the meter, use a damp cloth and mild detergent only. - 5 -

DESCRIPTION 2. DESCRIPTION - This meter is a portable professional measuring instrument, which can be used in the test of high voltage insulated for the maintenance and service of the industry equipment, such as the cables, transformers, generators, switch, etc. - This meter has of auto range, the indicator with a dual scales point needle, In the measuring range between 0 to 100000 M, the meter can auto-change the high or low range according to the value of the resistance. The flash of LED will indicate the user to get the reading from the scale line with the same color. - Using 8pcs of AA battery as the insulated resistance of the power measurement, its max. current is about 140mA, then the meter can be continuously worked for 4 hours. - When the measured resistance is larger than 100 M, the meter can be kept in a rated and high measured voltage (approx. 2500V, with the indicator of LCD), which could be used for precise measurement. - Dual measuring switch will help the user to avoid error operation. Test button also has the lock function. - In insulating measurement, the test button is pressed, the meter s alarming system will give out a humming sound, and the red high voltage output indicator is flashing, which warns the operator to pay - 6 -

DESCRIPTION attention to the high output voltage, and avoid getting electric shock. - Special airproof design will keep the inner circuit of the meter away from the humidity and the invasion of rotten air. - Having the function of battery test, which can expediently judge the battery s quality. - When using, the needle indicates the value of the measured insulated resistance, the reading of the measured voltage is shown in LCD indicator. - A rotary switch is used to select functions and ranges. - Releasing the test button automatically will discharge the capacitance of a circuit under test. 2.1 NAMES OF COMPONENTS (See Fig 1) ⑴ Test Line Probe ⑵ Guard Terminal ⑶ Earth Terminal ⑷ Green LED ⑸ Green Scale ⑹ Red LED - 7 -

DESCRIPTION 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Fig 1-8 -

DESCRIPTION ⑺ Red Scale ⑻ LCD Display ⑼ Indicator pointer ⑽ Indicator zero corrector ⑾ Panel ⑿ Test Button ⒀ Function Switch ⒁ Battery Cover 2.2 SWITCH AND BUTTONS ELUCIDATE Function Switch This switch is used to select measure mode and the switch of power. Test Button This Button is used to enter and keep the insulation resistance measurements. - 9 -

SPECIFICATIONS 3 SPECIFICATIONS Accuracy is specified for a period of year after calibration and at 18 to 28 (64 F to 82 F) with relative humidity to 70%. 3.1 GENERAL SPECIFICATIONS 3.1.1 Reading: see the scale with different color 3.1.2 Test Voltage Display: LCD 3.1.3 Low Battery Indication: displayed 3.1.4 Measuring method: Dual-slope integration A/D converter 3.1.5 Sampling Time: approx. 0.4 second 3.1.6 Power Supply: 12V: 1.5V 8 (size AA) battery (recommended to use the alkaline or heavy duty battery.), 9V: 9V 1 (6F22, 1604) battery 3.1.7 Power Consumption: Approx. 140mA(12V),Approx. 2mA(9V) 3.1.8 Operating Temperature: 0 to 40 (32 F to 104 F) 3.1.9 Operating Humidity: 85%RH 3.1.10 Storage Temperature: -10 to 50 (10 F to 122 F) - 10 -

SPECIFICATIONS 3.1.11 Dimension: 225 138 81 mm 3.1.12 Weight: Approx. 1200g(including battery) 3.2 ELECTRICAL SPECIFICATIONS Circumstance Temperature: 23 5 Relative Humidity: < 70% Measuring voltage DC Voltage Range 0-2000 M / 1000-100000 M (Auto-range) Accuracy Insulated 0-100M 3.0 % of arc Surroundings temperature:23 5 resistance 100-50000 M 2.5 % of arc Relative humidity:< 70 % 50000-100000M 3.0 % of arc 0-100M 6.0 % of arc Surroundings temperature:0 ~ + 40 100-50000 M 5.0 % of arc Relative humidity:< 85 % 50000-100000M 6.0 % of arc Output voltage 2500V 10 % 100-100000 M Insulated resistance 1000 M max. / 1000V Between the circuit and the case Voltage-resistant 5000V AC per minute Between the circuit and the case Remark: length of green scale:120mm, length of red scale:114mm - 11 -

OPERATING INSTRUCTION 4. OPERATING INSTRUCTION WARNING Once press the measured button, be careful to the high voltage between test line and other two connections. Earth terminal must be always connected with the ground for safe and reliable measurement. Before measure, ensure there are no any components damaged by high test voltage in the circuit. When measure the insulated resistance, the inside buzzer will send out humming sound continuously for reminding the user of safe operation. 4.1 ALARM In insulation range, the test button is pressed, the inside alarming system will send out a humming sound. 4.2 PREPARATION FOR MEASUREMENT 4.2.1 Set the function switch to OFF position. - 12 -

OPERATING INSTRUCTION 4.2.2 Use the screwdriver to adjust the zero corrector in the middle of the panel, make the needle to the left point of the green scale. 4.3 BATTERY CHECK 4.3.1 Set the function switch to BATT. position. 4.3.2 Press the measure button, the battery is good if the pointer is in the scale of BATT.OK ; if it is not in the scale, the battery is exhausted and should be replaced in time, otherwise, it will effect the measuring accuracy of the insulated resistance, or leak the liquid and then cause damage to the meter. ATTENTION While check battery, do not lock the measure button or press for long time to avoid excessive power consumption (the output current of which is about 150mA). - 13 -

OPERATING INSTRUCTION 4.4 INSULATION RESISTANCE MEASURING WARNING Insulation tests should be conducted on circuits that are out of energy. Ensure circuits are not live before commencing testing. Ensure that there is no any components damaged by high measured voltage. Using test line and test clips connect the meter and the tested circuit, then press test button. When testing, there is 2500V between the test line and other two connections terminal. 4.4.1 Set the function switch to OFF position. 4.4.2 Connect the tested circuit to the earth ground, then use the test clip to connect the earth and the earth terminal of the tested circuit, ensure better connection. 4.4.3 Set the function switch to M position. 4.4.4 Touch the probe of the test line to the tested circuit, then press the button. 4.4.5 LED on the dial will flash. The reading can be gotten from the green scale after the flash of green LED; or from the red scale after the flash of red LED. - 14 -

OPERATING INSTRUCTION 4.4.6 Loosen the button after the measurement, wait several seconds and then disconnect the test line from the tested circuit, which is to release the charge stored in the circuit tested. ATTENTION Due to huge power consumption in the measurement, continuous measurements for long time should be avoided. WARNING Never touch the circuit under test during insulation measuring. Never rotate the range switch while the test button is pressed. This may damage the meter. When measuring is complete ensure that the test button is released before the test clips are disconnected. This is because the system may be charged up and it must be allowed to discharge through the tester's internal discharge resistor. - 15 -

OPERATING INSTRUCTION 4.5 CONTINUOUS MEASUREMENT WARNING Insulation tests should be conducted on circuits that are de-energized. Ensure circuits are not live before commencing testing. Before measure, ensure there are no any components damaged by high test voltage in the circuit. Using test line and test clips connect the meter and the tested circuit, then press test button for testing. When testing, there is 2500V between the test line and other two connections terminal. 4.5.1 Set the function switch to OFF position. 4.5.2 Connect the tested circuit to the earth ground, then use the test clip to connect the earth and the earth terminal of the tested circuit, ensure better connection 4.5.3 Set the function switch to M position. 4.5.4 Touch the probe of the test line to the tested circuit. - 16 -

OPERATING INSTRUCTION 4.5.5 Press the button and clockwise rotate it to LOCK position for locking the measured button, then continuous measurement can be realized. 4.5.6 LED on the dial will flash. The reading can be gotten from the green scale after the flash of green LED; or from the red scale after the flash of red LED. 4.5.7 Loosen the button after the measurement, wait several seconds and then disconnect the test line from the tested circuit, which is to release the charge stored in the circuit. ATTENTION Due to huge power consumption in the measurement, continuous measurements for long time should be avoided. Warning Never touch the circuit under test during insulation measuring. Never rotate the range switch while the test button is pressed. This may damage the meter. When measuring is complete ensure that the test button is released before the test clips are disconnected. This is because the system may be charged up and it must be allowed to discharge through the tester's internal discharge resistor. - 17 -

OPERATING INSTRUCTION 4.6 USE OF GUARD TERMINAL Illustrated in Fig. 2 is an example of the insulation resistance measurement of an electric wire. If the line probe is simply connected to the wire conductor and the earth lead to the immersion liquid container as shown, a measurement error will be introduced as this results in the measurement of the combined resistance of insulation resistance and the surface leakage resistance at the cut end of the electric wire. In order to remove this surface leakage current, wind a guard wire around the cut end of the conductor and connect it to the guard terminal of the instrument using the guard lead. Then, the surface leakage current will bypass the indicating meter of the insulation resistance tester. - 18

OPERATING INSTRUCTION Fig 2-19 -

MAINTENANCE 5. MAINTENANCE 5.1 BATTERY REPLACEMENT WARNING Before attempting to remove the battery cover, be sure that test line or test clips have been disconnected from measurement circuit to avoid electric shock hazard. 5.1.1 12V(AA)BATTERY REPLACEMENT 5.1.1.1 If the battery test found its quality is poor, then replace with a new 12V battery. 5.1.1.2 Select OFF position. 5.1.1.3 Ensure that the test button is loosened. 5.1.1.4 Use the screwdriver to remove the screws on the battery drawer, open the drawer and replace the battery. 5.1.1.5 Put the battery cover as its origin, then fix it up by the screws. ATTENTION If the meter is used for a short time, then the battery should be removed to avoid damaging the meter - 20 -

MAINTENANCE because of its leakage. Replacing all of the old battery with new one, never intermix them. Had better use the Alkaline battery 5.1.2 9V(6F22)BATTERY REPLACEMENT 5.1.2.1 If the sign appears on the LCD display, it indicates that the battery should be replaced. 5.1.2.2 Loosen the screw fixing the battery cover and remove it. 5.1.2.3 Replace the exhausted battery with a new one. 5.1.2.4 Put the battery cover as its origin. 5.2 TEST LINE AND TEST CLIPS REPLACEMENT WARNING Full in compliance with safety standards can be guaranteed only if used with test line or test clips supplied. If necessary, they must be replaced with the same model or same electric ratings. Electric ratings of the test line or test clips: 2500V 1A You must replace the test line or test clips if the lead is exposed. - 21 -

ACCESSORIES 6. ACCESSORIES ⑴ Test line: Electric Ratings 2500V 1A One piece ⑵ Test Clips: Electric Ratings 2500V 1A One piece ⑶ Battery: 1.5V SIZE AA Eight pieces ⑷ Battery: 9V 6F22 1604 One piece ⑸ Operator s Manual One piece ⑹ Wrapping Bag One piece - 22 -

HYS004971 A0