Checklist for Insertion and Setup of an ICP Monitoring Device

Similar documents
SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

INTERDISCIPLINARY CLINICAL MANUAL Policy and Procedure

ATI Skills Modules Checklist for Urinary Catheter Care

NORTH CAROLINA BOARD OF NURSING NURSE AIDE II TRAINING MODULE

COLLEGE OF MEDICINE. Sony Portable Video Conference Systems. Operations Manual

N26 Chest Tubes 5/9/2012

Arterial pressure monitoring Direct arterial pressure monitoring permits continuous measurement of systolic, diastolic, and mean pressures and allows

PATIENT GUIDE. Understand and care for your peripherally inserted central venous catheter (PICC). MEDICAL

Optional Lab: Install, Configure, and Partition a Second Hard Drive in Windows Vista

X-Plain Subclavian Inserted Central Catheter (SICC Line) Reference Summary

How to Change a Foley Catheter Step-by-step instructions for the caregiver

TRANSPORT PROTOCOLS FOR PATIENTS ON AN INTRA-AORTIC BALLOON PUMP

Medications or therapeutic solutions may be injected directly into the bloodstream

URINARY CATHETER INSERTION - STRAIGHT OR INDWELLING CATHETER

OM2260VW2 USER MANUAL VERIZON WIRELESS HOME PHONE CONNECT

Your Guide to Peritoneal Dialysis Module 3: Doing Peritoneal Dialysis at Home

BARD MEDICAL DIVISION UROLOGICAL DRAINAGE. Foley Catheter Care & Maintenance. Patient Education Guide

How To Use A U.S. Cell Phone At Home

Direct Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003

This Document Contains:

All About Your Peripherally Inserted Central Catheter (PICC)

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

Patient Transport. with the Automated Impella Controller

Technical Note. TN: Technical note for phone access of HMS system. Rev. a

GJ TUBE (GASTROJEJUNOSTOMY TUBE) Daily Care

Caring for a Tenckhoff Catheter

BROOKFIELD MODEL VTE250 ELECTRIC VISCOSEL PROCESS VISCOMETER. Installation, Operation, and Maintenance Instructions. Manual No.

MISSISSIPPI BOARD OF NURSING IV THERAPY COURSE FOR THE EXPANDED ROLE LICENSED PRACTICAL NURSE COURSE OUTLINE

510 / 511 Cardiorespiratory Monitor Preventative Maintenance Test Procedure

CABLE ONE ALL DIGITAL

Fact Sheet. Caring for and Changing your Supra-Pubic Catheter (SPC) Queensland Spinal Cord Injuries Service

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS

Section Two: Arterial Pressure Monitoring

RACEAIR REMOTE PAGER SYSTEM

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Installation Guide (No Router)

User Guide VERIZON WIRELESS HOME PHONE CONNECT

5CHAPTER. Setting Up the Polling Place. IMPORTANT: The electronic poll book MUST be set up first and you MUST begin the update by 5:15 am.

Located On IV Pole Front of cart- Arrest Board

Your Digital Adapter Easy Setup Guide

CareSentinel Set Up Guide for Android Devices

Model PBX 416+ Programming & User Guide

Troubleshooting the Verizon MI424WR Router

Welcome to life on. Get started with this easy Self-Installation Guide.

ReSound Unite TV FREQUENTLY ASKED QUESTIONS. Setup & Configuration. Use & Operation. Troubleshooting

Technical Update TAA.TU Rev. 1

910 Series Photo Jetprinter. Setup Solutions

PICC & Midline Catheters Patient Information Guide

Comfort Contego User Manual

PA500II Enterprise Mobile Computer

Biliary Drain. What is a biliary drain?

User Installation Guide

Digital Adapter Easy Setup Guide

4. Do you have a VGA splitter ( Y Cable)? a document camera?

Upgrade Instructions. View Manager 96 v Software Upgrade

FiberOptix TM. AP Sensor Zeroing and Calibration Guide INTRA-AORTIC BALLOON PRODUCT SUPPORT HOTLINE. U.S. and Canada: IABP (4227)

Installation Guide. Wyse D Class 4-GB RAM Upgrade Option Kit. Products: D90D7, D90DW. Issue: PN: L Rev. A

OPL BASIC. Dosing System for Professional Laundry machines. Contents

Safety FIRST: Infection Prevention Tips

Fluid Management DYONICS 25. DYONICS 25 Tube Sets. Fluid Management System. Easy set up and use. High flow. Consistent performance

Ward 29 guide to the safe preparation and administration of intravenous (IV) antibiotics at home

OPERATOR S MANUAL Vet Pump

Model PBX 308 Plus System Administration Guide

HP Advanced Wireless Docking Station. User Guide

2016 Frontier Communications. Here s your guide to easy installation. Get the most out of Frontier High-Speed Internet.

Troubleshooting the Actiontec M1424WR Router

9.0 Components Description SENTIO MMG System Quick Setup Plugging in Control Unit Attaching Sensor Harness

PG-01instruction manual

Central Line Blood Draw

ibidi PumpControl Instruction Manual Version 1.5.2

Actiontec Router Troubleshooting

Computer Lane Monitor for Windows

Regents of the University of Michigan 5/7/2015. Review pump functionality, sets and accessories Understand key process steps for

SMART Board TM SBID 6052i Interactive Display Troubleshooting Guide

Verizon 9100EM Router Troubleshooting

Quick Start Guide. Vonage VWR Device

Paediatric Intensive Care Unit Nursing Procedure: Care of Arterial Lines.

TROUBLESHOOTING FLOW CHART

Chester County Hospital E134 Conference Room. Control for AV system

Step by step guide to installing your own Ku Band satellite dish

Policies & Procedures. I.D. Number: 1073

with FLASH TIPS PLUM A+ INFUSION SYSTEM OPERATING INSTRUCTIONS REFER TO SYSTEM OPERATING MANUAL FOR COMPLETE OPERATING INSTRUCTIONS

CutMaster Major External Replacement Parts

Patient Information Guide Morpheus CT Peripherally Inserted Central Catheter

Nexus 7 Troubleshooting HINTS, TIPS & TROUBLESHOOTING

Back-UPS Pro 1300/1500 Installation and Operation

Easy Connect Guide Digital Adapter. Say hello to your TV s new best friend.

Transmitter Interface Program

2. Does the patient have one of the following appropriate indications for placing indwelling urinary catheters?

Patient Information Sheet

Chapter 1 Connecting the Router to the Internet

Model PBX816ex. Programming & User Guide

8 Channel Status Input Panel model SIP-8

ADVANCED USER S GUIDE

Hemodynamic Monitoring: Principles to Practice M. L. Cheatham, MD, FACS, FCCM

Red rubber all-purpose urethral. Catheters. Bard Clean-Cath Intermittent catheter: Fr. 100/case

Quick Start Guide 3G Router from Vodafone

PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly

CADD-legacy PCA pump Model ml Cassette

PRODUCT SHEET OUT1 SPECIFICATIONS

Transcription:

Checklist for Insertion and Setup of an ICP Monitoring Device Before proceeding, contact neurosurgery to identify the type of catheter to be inserted. The catheter type will define the equipment requirements and preparation needed. 1. Obtain the Catheter (choose one of 3 types below): A. Codman Microsensors (two types): Codman Microsensors are catheters that measure intracranial pressure via a strain guage microchip located at the tip of the catheter. Both Microsensor catheters below are connected to the Codman Express for pressure monitoring. Unlike fluid filled pressure monitoring, pressure transmission is unaffected by obstructed fluid paths (e.g., from clot or tissue). I. Codman Microsensor Intraparenchymal Catheter: This catheter is inserted into superficial brain tissue. It measures pressure only, it does not drain CSF. It may be used for monitoring when ventricles are too small to access, or in patients with normal intracranial pressure who are at risk to develop elevated pressures. If ICP rises, it may be changed to an EVD. II. Codman Microsensor Intraventricular Drainage Catheter: This is an intraventicular catheter. It is used to measure ICP and to drain CSF. 1

B. Standard Intraventricular Drainage Catheter: The standard ventricular catheter is used to measure ICP and to drain CSF. Pressure is measured using a standard tranducer, picking up pressure that is transmitted through the fluid filled channel of the catheter. 2. Obtain supplies to insert ICP Catheter: Ventricular Drainage Tray (ensure drill is packaged with tray) Hair clipper with clipper head Line cart Sterile gowns Large sterile drape and sterile towels Local anaesthetic Tegaderm dressing 3. Do you need a CSF drainage unit (Codman EDS 3)? Both the Codman Microsensor venticular drainage catheter and the standard ventricular drainage catheter require a CSF drainage unit. There is no drainage of CSF with an intraparenchymal catheter. Supplies Needed: Codman EDS 3 Standard pressure transducer kit (used to flush drainage unit) 500 ml bag of normal saline Codman laser level Portable IV pole 2

4. Obtain supplies for pressure monitoring: Microsensor catheters (intraparenchymal or intraventricular): Codman Express monitor Codman Express cable (should be attached to Codman Express ) Codman -Datex pressure cable for transferring pressure from the Codman Express to the Datex monitor (Datex end of cable is red) Datex pressure module (e.g., P3, P4 or an unused P2) Standard Intraventricular Drainage Catheter: Datex pressure module (e.g., P3, P4 or an unused P2) Datex regular pressure cable Pressure transducer (also used to prime Codman EDS3 drainage unit) PROCEDURE FOR PRIMING CODMAN EDS3 (CSF collection unit): For Codman Microsensor ventricular drainage catheter and standard ventricular drainage catheter. Perform hand hygiene Hang the Codman EDS3 on IV pole Prime pressure tubing with normal saline (do not pressurize saline bag) Connect pressure tubing to the Codman EDS3 stopcock that is closest to the patient catheter Flush drainage tubing toward patient catheter connection, then into CSF collection chamber Turn patient stopcock to 45 degrees (off in all directions) until ready to hook up Using laser level, adjust position of drainage unit until 0 on drainage panel is level with external auditory canal Position the drip chamber to the ordered level (be sure the orders clearly identify whether the order is for cmh20 or mmhg 3

4

PROCEDURE FOR SETUP UP OF CODMAN EXPRESS For both Codman Microsensor Catheters Perform hand hygiene Place Codman Express on IV pole and plug into red electrical outlet Turn Codman Express on and listen for audio beep Assist physician to prepare sterile field and open Codman Microsensor catheter 5

Connect Microsensor to Codman Express cable As soon as cable is connected, Transducer Detected will appear on the screen Provide physician with a syringe of sterile saline. This will be used to fill the catheter tray and submerge the tip of the catheter during zeroing. With catheter in the horizontal position and submerged in 3 inches of sterile saline, press the Zero button when prompted 6

When zeroing is completed, a 3 digit reference number will display Document the 3 digit reference number in the Kardex and AI record. Record the number on the Microsensor with a Sharpie Press Menu/Enter Catheter can now be inserted; the pressure will automatically display on the Codman Express PROCEDURE FOR CONNECTING CODMAN EXPRESS TO DATEX MONITOR For both Codman Microsensor Catheters ICP is measured by the Codman Express. This pressure can be duplicated on the Datex monitor for the purpose of generating an ICP waveform. Perform hand hygiene Connect the Codman -Datex cable to the back of the Codman Express and to any spare pressure module (e.g., P3, P4 or an unused P2) 7

Go to Invasive Pressures (Datex ) and change the module label to ICP When prompted (on the Codman Express screen), press the Zero button on the Datex module. (Note that you do not need to open any stopcocks or adjust the position of any tubing or devices). Confirm that the reference line for the ICP waveform is displaying zero 8

When prompted (on the Codman Express screen), confirm that the monitor is reading zero by selecting the menu/enter button When prompted to choose the calibration signal, select the 20 button Confirm that the bedside monitor has a reference line at 20 mmhg 9

Press the menu/enter button to verify that the bedside monitor is correct; if it is not, use the arrow keys to adjust the reference line PROCEDURE FOR MONITORING ICP WITH STANDARD VENTRICULAR CATHETER: ICP is measured with standard transducer and fluid filled circuit. Perform hand hygiene After priming the Codman EDS3, connect the pressure transducer cable to a bedside Datex module (e.g., P3, P4 or a spare P2) Change the Module label to ICP Tape the transducer to the back of the Codman EDS3 aligning the stopcock from the transducer with the zero reference stopcock on the Codman EDS3. Confirm that the stopcocks are level with the patient s external auditory canal using a laser level Zero the transducer Please note the following: An LCD switch is located on the back of the ICP Express to see screen in the dark Alarms are automatically off when ICP express is turned on. To activate alarms go to menu/enter button, choose on/off alarm limits option and activate alarm option by arrowing up or down for required limits. Troubleshooting When zeroing the ICP Express monitor, if the value -99 appears this means the grey cable that connects to the front of the ICP Express monitor needs to be replaced If a zero reference number appears before zeroing, turn ICP Express off and restart steps Traveling Disconnect catheter from grey cable attached to ICP Express When reconnecting catheter to ICP Express, the reference number on the monitor must match the number written on the catheter. Either accept the displayed reference number (if they match) or adjust. Go to menu/enter and use arrow key to choose the correct reference number The Codman Microsensor is MRI compatible Judy Hackett RN BScN CNCC(c) Clinical Educator, CCTC June 12, 2008. Revised April 7, 2010 (BM); Revised May 28, 2010 (BM) 10