1 Technical Update TAA.TU Rev. 1 Effective: 05 APR 2013 CO# Product: Subject: From: Phoenix Dialysis System Required Electrical Safety Inspections: VI, PET, ELT, PLT Gary Willis Abstract: This update provides a detailed description of and frequency for, the Electrical Safety Inspections (ESI) that the manufacturer requires to be performed on the Phoenix Dialysis System. The ESI procedures include the Visual Inspection (VI), the Protective Earth Test (PET), the Earth Leakage Test (ELT), and the Patient Leakage Test (PLT). NOTE: TO AVOID ELECTRICAL SHOCK HAZARD, ALL ESI TESTING MUST BE PERFORMED BY A TRAINED AND QUALIFIED TECHNICIAN USING PROPER ELECTROSTATIC SAFETY DEVICES (i.e. wrist grounding strap). NOTE: Please read and understand this entire document before performing the ESI procedures. ELECTRICAL SAFETY INSPECTION TEST DESCRIPTIONS 1. VI Visual Inspection detects potential faults or conditions that could be related to the electrical safety of the machine. 2. PET Protective Earth Test verifies that three grounded parts of the machine have a safe low resistance electrical path to earth ground in case of insulation failure. 3. ELT Earth Leakage Test verifies, with four test conditions, that the AC leakage current in the ground wire of the main power cord of the Phoenix machine is within safe limits, using normal and reverse AC polarity, and with neutral closed (Normal Condition - NC) and open (Single Fault Condition SFC). 4. PLT Patient Leakage Test verifies, with eight test conditions, that the AC and DC leakage current in the hydraulic fluid path of the Phoenix machine is within safe limits, using normal and reverse AC polarity, and with earth ground intact (Normal Condition - NC) and open (Single Fault Condition SFC). ELECTRICAL SAFETY INSPECTION TEST STANDARDS 1. The ESI tests described in this update are in accordance with the following standards: 1.1. IEC : Medical Electrical Equipment - Part 1: General requirements for safety IEC 62353: Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment. 2. Recognizing their enhanced patient safety features, Gambro has adopted these international standards, as have other biomedical equipment manufacturers worldwide. All ESI testing performed by Gambro Technical Services (GTS) employees on any Gambro equipment will be conducted in accordance with these standards. Page 1 of 11

2 ELECTRICAL SAFETY INSPECTION TEST FREQUENCY ESI testing should be performed: 1. After first installation. (VI, PET, ELT, PLT) 2. After repair / replacement of components that are connected to ground. * 3. After repair / replacement of components with a direct or indirect electrical contact with the dialysis fluid. * 4. After repair / replacement of components connected to AC power parts. * 5. After a component has been replaced that has been identified in its Instructions Sheet as requiring an ESI test following installation. * 6. After unexpected electrical events affecting the AC line voltage. 7. After accidental leakage of fluid onto internal components from external or internal sources. 8. After transportation (other than what is normal within the clinical environment). 9. During a Preventive Maintenance (PM). (VI, PET, ELT, PLT) 10. After an ESI test fails, and subsequent machine repairs are made, then the failed ESI test must be performed again to verify the success of the repair. * Due to the nature of the part that is replaced, ESI testing may include: 1. VI and PET 2. VI and PLT 3. VI, PET, ELT, and PLT Refer to the instruction sheet provided with the spare part for the ESI testing requirements, or contact Gambro Technical Assistance Services (TAS) at for further information. EQUIPMENT REQUIRED 1. Fluke ESA612 Electrical Safety Analyzer (ESA) and accessories Gambro Technical Services has standardized the ESI testing procedures described below around this analyzer. ESA devices from other manufacturers may be used, as long as they are able to meet the test requirements of the IEC standards described above Patient Leakage Test (PLT) Fixture Hexagon Wrench P2 Pump Calibration Tool A/V Test Cartridge 6. Condo/Temp Calibration Tool (included in Calibration Tool Kit) 7. Graduated Cylinder 8. Approx. 10 length of blood cartridge tubing filled with water ELECTRICAL SAFETY INSPECTION TEST PROCEDURES 1. VISUAL INSPECTION (VI) The VI procedure involves verifying and correcting any hardware condition of the Phoenix machine that does not conform to specifications in the following areas: 1.1. Verify exterior parts, panels, and covers for damage, cracks, etc Verify placement and condition of safety related markings and labeling Verify integrity of mechanical parts (damage, obstructions, misalignments, contamination, etc.) Verify electrical components and cabling (cuts, incorrect or loose connections and fuses, etc.). Page 2 of 11

3 2. PROTECTIVE EARTH TEST (PET) 2.1. NOTE: The PET procedure is performed with the PHOENIX MACHINE POWERED OFF and DISCONNECTED FROM MAIN AC POWER Turn on the ESA612 main power switch The ESA612 will display the message: Select a test or SETUP 2.4. Press the SETUP key Press the More (F4) key Press the Instrument (F2) key The Standard (F1) on the display should indicate (If not, press the Standard [F1] key and scroll to using the and arrow keys and then press the Standard [F1] key again.) 2.8. The GFCI Limit should indicate 10 ma. (If not, press the GFCI Limit [F2] key and scroll to 10 ma and press the GFCI Limit [F2] key again.) 2.9. The Polarity Delay should indicate 5 s. (If not, press the Polarity Delay [F3] key and scroll to 5 s and press the Polarity Delay [F3] key again.) The ESA612 display should appear as shown in FIGURE 1. FIGURE Press the Done (F4) key Connect the Phoenix AC power cord into the Equipment Outlet on the right side of the ESA Connect one end of the red test lead to the red V/Ω/A jack of the ESA612 and the other end to the red alligator clip Insert the 3 cm long silver Null Post Adapter into the green Ø / NULL jack of the ESA Connect the red alligator clip to the Null Post Adapter. See FIGURE Press the Ω key and press the Zero Leads (F4) key to zero out the resistance of the test lead Remove the red alligator clip from the adapter post and remove and store the Null Post Adapter NOTE: To avoid electrical shock, remove the Null Post Adapter from the Ø / NULL jack after zeroing the test lead. The adapter is grounded and could become potentially hazardous during some of the later ESI test conditions. Page 3 of 11

4 2.19. PET Test Point #1 (Left Door Hinge ) Remove the top cover and open the Phoenix machine using the Hexagonal Wrench or equivalent Connect the red alligator clip to the ground connections near the left door hinge of the Phoenix machine, as shown in FIGURE NOTE: During the measurement, the power cord shall be flexed along its length in order to detect possible variations of resistance due to deterioration Verify that the indicated resistance is < Ω (or < Ω if during installation or if the power cord has just been replaced). If not, find and correct the problem before NOTE: All requirements for maximum measured values indicated in these ESI procedures have been reduced to allow for the operating characteristics of the EAS NOTE: The test lead connection to the door hinge is subject to the contact resistance of the alligator clip. Wiggle the alligator clip on the hinge to obtain the lowest resistance indication on the ESA612 display. Use this same technique on the other two test locations to minimize the contact resistance RECORD the maximum observed resistance for the Left Door Hinge Remove the red alligator clip from the hinge. FIGURE 2 FIGURE PET Test Point #2 (Bracket Ethernet Board) Connect the red alligator clip to the Bracket of the Ethernet Board of the Phoenix machine, as shown in FIGURE 4A. If the Ethernet Board and bracket are not installed, connect the red alligator clip to the External Interface Plate grounding stud, as shown in FIGURE 4B Verify that the indicated resistance is < Ω (or < Ω if during installation or if the power cord has just been replaced). If not, find and correct the problem before RECORD the maximum observed resistance for the Bracket Ethernet Board Remove the red alligator clip from the test point. Page 4 of 11

5 FIGURE 4A FIGURE 4B FIGURE PET Test Point #3 (Hydraulic Metal Rack) Connect the red alligator clip to the hydraulic metal rack of the Phoenix machine, as shown in FIGURE Verify that the indicated resistance is < Ω (or < Ω if during installation or if the power cord has just been replaced). If not, find and correct the problem before RECORD the maximum observed resistance for the Hydraulic Metal Rack Remove the red alligator clip from the metal rack Remove the red test lead from the V/Ω/A jack of the ESA Remove the Phoenix power cord from the Equipment Outlet of the ESA EARTH LEAKAGE TEST (ELT) 3.1. NOTE: Do not attempt the ELT unless the above PET has been completed successfully Install the BJ2ECG Input Jack Adapter accessory on the top of the ESA612, as shown in FIGURE 6. FIGURE 6 FIGURE Install a red alligator clip at one end of a red test lead and a black alligator clip at the other end of the lead. Page 5 of 11

6 3.4. Connect the black alligator clip to the RL terminal of the Input Jack Adapter mounted on the ESA612, as shown in Figure Gather the supplies needed to setup the Phoenix for a simulated patient treatment Plug the Phoenix machine into the wall receptacle, turn it ON, and put the machine into a simulated treatment, with blood detection, and increase the Blood Pump speed until it reaches maximum NOTE: During ESI testing, the simulated patient treatment should utilize mid-range dialysate conductivities and temperature. Suggested values are: Acid = 15 ms/cm, Bicarb = 3.2 ms/cm and Temperature = 37 ºC. The leakage test results are not overly sensitive to conductivity variations, as long as, the conductivity is in the normal range for patient treatments Attach the PLT Test Fixture in-line with the blue dialyzer line and the blue connector of the P2 Pump Test Tool using the Condo/Temp Test Tool. Connect the red alligator clip from the ESA612 to the metal terminal of the PLT Fixture. See FIGURE NOTE: The PLT Fixture will not be used during the ELT procedures, but should be connected now to facilitate the PLT procedures later. The presence of the connection between the ESA612 and the PLT Fixture will not alter the ELT results since the ESA612 is configured internally based upon the selected test The general configuration of equipment when beginning the ELT (and PLT) measurements is represented in the diagram in FIGURE NOTE: The AC power cable connection between the Phoenix machine and the ESA612, indicated in FIGURE 8, will be accomplished later in this procedure NOTE: During ELT and PLT measurements, insure that the drain tube is not connected to an earth grounded drain system, which might affect leakage current readings. If unsure, allow the Phoenix machine to drain into an insulated container (plastic bucket, etc.) while taking measurements ELT Test Condition #1 (AC, Normal Polarity, Closed Neutral) Press the μa key on the ESA Press the Earth (F1) key Press the or arrow keys until the ESA612 display indicates AC only There will be no AC power applied to the ESA612 Equipment Outlet since the ESA612 display indicates OFF for POLARITY on the right side of the display Press the POLARITY key and the ESA612 display should indicate NORMAL POLARITY and CLOSED NEUTRAL. The EARTH key will have no options displayed. Power will be applied to the ESA612 receptacle at this time When the Phoenix is in full dialysis mode, unplug the Phoenix from the wall and plug it into the ESA612 Equipment Outlet. Page 6 of 11

7 FIGURE NOTE: ALTERNATELY, the Phoenix setup can be accomplished after plugging the machine into the ESA612 that has been configured as in steps to A Phoenix power failure alarm should be displayed (unless equipped with a second 12V Backup Battery). Press the Override key to clear it, and any other alarm caused by the power interruption. When the treatment has resumed proceed to the next step NOTE: All ELT / PLT leakage measurements must be made when the Phoenix machine is in hydraulic Phase 19. Failure to observe this condition could result in leakage currents that are lower than the correct values Verify that the Phoenix machine is in hydraulic Phase 19. Verify that the displayed current on the ESA612 is < 296 μa. If not, find and correct the problem before NOTE: All requirements for maximum measured values indicated in these ESI procedures have been reduced to allow for the operating characteristics of the EAS NOTE: During each of the four ELT test conditions, note the maximum leakage current observed. Record only one leakage current value when all four ELT tests are complete, which is the maximum value measured during any of the four ELT Test Conditions ELT Test Condition #2 (AC, Reverse Polarity, Closed Neutral) Press the POLARITY key and the ESA612 display should indicate OFF POLARITY. The NEUTRAL and EARTH keys will have no options displayed. The Phoenix machine will shut down, since power to the machine has been removed Wait a minimum of 5 seconds before proceeding Press the POLARITY key again and the ESA612 display should indicate REVERSE POLARITY and CLOSED NEUTRAL. The EARTH key will have no option displayed. Page 7 of 11

8 Power should now be applied to the Phoenix machine Resume the simulated patient treatment and clear any alarms until the Phoenix machine is again in hydraulic Phase Verify that the indicated current is < 296 μa. If not, find and correct the problem before ELT Test Condition #3 (AC, Reverse Polarity, Open Neutral) Press the NEUTRAL key and the EAS612 display should indicate REVERSE POLARITY and OPEN NEUTRAL. The EARTH key will have no options displayed The Phoenix machine will shut down, since power to the machine has been removed Verify that the indicated current is < 296 μa. If not, find and correct the problem before ELT Test Condition #4 (AC, Normal Polarity, Open Neutral) Press the POLARITY key and the ESA612 display should indicate OFF POLARITY. The NEUTRAL and the EARTH keys will have no options displayed Wait a minimum of 5 seconds Press the POLARITY key again and the EAS612 display should indicate NORMAL POLARITY and OPEN NEUTRAL. The EARTH key will have no options displayed The Phoenix machine will remain shut down, since power has not yet been restored Verify that the indicated current is < 296 μa. If not, find and correct the problem before NOTE: RECORD only one current value for the ELT, which is the maximum value measured at any of the ELT Test Conditions #1 through #4 in the field marked for ELT 115V: NC; SFC Press the NEUTRAL key and the EAS612 display should indicate NORMAL POLARITY and CLOSED NEUTRAL. The EARTH key will have no options displayed Power should now be applied to the Phoenix machine. Resume the simulated patient treatment and clear any alarms until the machine is again in hydraulic Phase PATIENT LEAKAGE TEST (PLT) 4.1. NOTE: Do not attempt the PLT unless the VI has been completed successfully NOTE: If the PET and ELT are required, they must be completed successfully before proceeding with the PLT Configure the ESA612 and setup the Phoenix machine as indicated in steps 3.2 through 3.12 above. (Skip this step if you have just completed the ELT procedure.) 4.4. NOTE: Ensure that all the air is removed from the fluid path of the PLT Fixture by holding the fixture so the blue dialysate connector from the Phoenix is at the bottom allowing dialysate to flow up through the fixture and remove any trapped air. Air remaining in the fixture will cause inaccurate leakage current measurements Turn the ESA612 ON and press the μa key Press the More (F4) key Press the Select (F1) key Verify the EAS612 display indicates AC only. Use the or keys if necessary to change Press the POLARITY key and the ESA612 display should indicate: AC only, NORMAL POLARITY, CLOSED NEUTRAL, CLOSED EARTH When the Phoenix is in full dialysis mode, unplug the Phoenix from the wall and plug it into the ESA612 Equipment Outlet A Phoenix power failure alarm should be displayed (unless equipped with a second 12V Backup Battery). Press the Override key to clear it, and any other alarm caused by the power interruption. When the treatment has resumed proceed to the next step. Page 8 of 11

9 4.12. NOTE: All ELT / PLT leakage measurements must be made when the Phoenix machine is in hydraulic Phase 19. Failure to observe this condition could result in leakage currents that are lower than the correct values The PLT procedure will test for leakage current in the dialysate under eight test conditions PLT Test Condition #1 (AC, Normal Polarity, Closed Neutral, Closed Earth) Before beginning the PLT measurements, verify the ESA612 display indicates: AC only, NORMAL POLARITY, CLOSED NEUTRAL, CLOSED EARTH. See FIGURE 9. (NOTE: the measured μa value may differ from the example shown in FIGURE 9.) FIGURE Verify the Phoenix machine is in a simulated treatment and in hydraulic Phase Verify that the indicated current is < 8.9 μa. If not, find and correct the problem before NOTE: All requirements for maximum measured values indicated in these ESI procedures have been reduced to allow for the operating characteristics of the EAS PLT Test Condition #2 (DC, Normal Polarity, Closed Neutral, Closed Earth) Press the or arrow keys until the Analyzer display indicates DC only Verify that the ESA612 display indicates: DC only, NORMAL POLARITY, CLOSED NEUTRAL, CLOSED EARTH Verify the Phoenix is in hydraulic Phase Verify that the indicated current is < 8.9 μa. If not, find and correct the problem before PLT Test Condition #3 (DC, Reverse Polarity, Closed Neutral, Closed Earth) Press the POLARITY key and the ESA612 display should indicate OFF POLARITY and CLOSED EARTH. The NEUTRAL will have no options displayed The Phoenix machine will shut down, since power to the machine has been removed Wait a minimum of 5 seconds Press the POLARITY key again and the ESA612 display should indicate: DC only, REVERSE POLARITY, CLOSED NEUTRAL, CLOSED EARTH Power should now be applied to the Phoenix machine. Page 9 of 11

10 Resume the simulated patient treatment and clear any alarms until the machine is again in hydraulic Phase Verify that the indicated current is < 8.9 μa. If not, find and correct the problem before PLT Test Condition #4 (AC, Reverse Polarity, Closed Neutral, Closed Earth) Press the or arrow keys until the ESA612 display indicates AC only Verify that the ESA612 display indicates: AC only, REVERSE POLARITY, CLOSED NEUTRAL, CLOSED EARTH Verify the Phoenix machine is in hydraulic Phase Verify that the indicated current is < 8.9 μa. If not, find and correct the problem before NOTE: RECORD the maximum value measured at either of PLT Test Conditions #1 & #4 in the field marked for the PLT NC 10 μa AC NOTE: RECORD the maximum value measured at either of PLT Test Conditions #2 & #3 in the field marked for the PLT NC 10 μa DC PLT Test Condition #5 (AC, Reverse Polarity, Closed Neutral, Open Earth) Press the EARTH key and the ESA612 display should indicate: AC only, REVERSE POLARITY, CLOSED NEUTRAL, OPEN EARTH Verify the Phoenix machine is in hydraulic Phase Verify that the indicated current is < 395 μa. If not, find and correct the problem before PLT Test Condition #6 (DC, Reverse Polarity, Closed Neutral, Open Earth) Press the or arrow keys until the ESA612 display indicates DC only Verify that the ESA612 display indicates: DC only, REVERSE POLARITY, CLOSED NEUTRAL, OPEN EARTH Verify the Phoenix machine is in hydraulic Phase Verify that the indicated current is < 38.6 μa. If not, find and correct the problem before PLT Test Condition #7 (DC, Normal Polarity, Closed Neutral, Open Earth) Press the POLARITY key and the ESA612 display should indicate: DC only, OFF POLARITY, OPEN EARTH. The NEUTRAL key will have no options displayed The Phoenix machine will shut down, since power to the machine has been removed Wait a minimum of 5 seconds Press the POLARITY key again and the ESA612 display should indicate: DC only, NORMAL POLARITY, CLOSED NEUTRAL, OPEN EARTH Resume the simulated patient treatment and clear any alarms until the Phoenix machine is again in hydraulic Phase Verify that the indicated current is < 38.6 μa. If not, find and correct the problem before PLT Test Condition #8 (AC, Normal Polarity, Closed Neutral, Open Earth) Press the or arrow keys until the ESA612 display indicates AC only Verify that the ESA612 display indicates: AC only, NORMAL POLARITY, CLOSED NEUTRAL, OPEN EARTH Verify the Phoenix machine is in hydraulic Phase Verify that the indicated current is < 395 μa. If not, find and correct the problem before NOTE: RECORD the maximum value measured at either of PLT Test Conditions #5 & #8 in the field marked for the PLT SFC 400 μa AC. Page 10 of 11

11 NOTE: RECORD the maximum value measured at either of PLT Test Conditions #6 & #7 in the field marked for the PLT SFC 40 μa DC. 5. This completes the ESI testing procedures. The simulated patient treatment may be continued for the length of time necessary to complete any other required testing or terminated at the Phoenix technician s discretion. 6. Power OFF the Phoenix machine and remove the ESI test equipment. 7. Power OFF the ESA Remove the Phoenix machine AC power cord from the ESA612 Equipment Outlet. 9. Remove the PLT Fixture from the dialysate flow path. 10. Rinse the PLT Fixture internal flow path with RO water to flush out any remaining dialysate. NOTE: Questions regarding these ESI procedures should be directed to Gambro Technical Assistance Services (TAS) at (800) Page 11 of 11

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011 Instruction Manual 2in1 LAN Tester & Multimeter Model: LA-1011 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.

More information

Model 1756 Test Lead Kit

Model 1756 Test Lead Kit Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Model 1756 Test Lead Kit Gerneral Purpose Test Lead Information Description These test leads allow you

More information

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05 RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,

More information

HP UPS R1500 Generation 3

HP UPS R1500 Generation 3 HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications

More information

INSTALLATION GUIDE. Card Reader & Controller with KIM Swipe Reader for Solitaire 850 / 950 / 850L Learnlok PK2930

INSTALLATION GUIDE. Card Reader & Controller with KIM Swipe Reader for Solitaire 850 / 950 / 850L Learnlok PK2930 INSTALLATION GUIDE Card Reader & Controller with KIM Swipe Reader for Solitaire 850 / 950 / 850L Learnlok PK2930 Card Reader and Controller Model 3.5 with KIM Swipe Reader Table of Contents 1. Features..................................

More information

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual. i800 AC Current Clamp Instruction Sheet Introduction The i800 AC Current Clamp, the Clamp, has been designed for use with multimeters, recorders, power analyzers, safety testers, etc., for accurate non-intrusive

More information

SERVICE MANUAL 12VDC WALL THERMOSTAT AIR CONDITIONING SYSTEMS ROOFTOP UNITS ONLY

SERVICE MANUAL 12VDC WALL THERMOSTAT AIR CONDITIONING SYSTEMS ROOFTOP UNITS ONLY SERVICE MANUAL 12VDC WALL THERMOSTAT AIR CONDITIONING SYSTEMS ROOFTOP UNITS ONLY! WARNING - SHOCK HAZARD! TO PREVENT THE POSSIBILITY OF SEVERE PERSONAL INJURY, DEATH, OR EQUIPMENT DAMAGE DUE TO ELECTRICAL

More information

CHAPTER 4 UTILITY SYSTEMS ELECTRICAL. Utility Systems Electrical. Main Panel

CHAPTER 4 UTILITY SYSTEMS ELECTRICAL. Utility Systems Electrical. Main Panel CHAPTER 4 UTILITY SYSTEMS ELECTRICAL Utility Systems Electrical The electrical supply to your home begins outside, where you will see either an overhead feed and piping down the side of your home or (if

More information

MCR1900 Media Converter 19-Slot Chassis

MCR1900 Media Converter 19-Slot Chassis MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the

More information

INSTALLATION & SERVICE MANUAL. Display Panel

INSTALLATION & SERVICE MANUAL. Display Panel INSTALLATION & SERVICE MANUAL Display Panel The PowerLine EMS TM is a specialized power distribution and energy management system intended to be used in recreational vehicles. The Control Module is housed

More information

HP R12000 and R18000 DirectFlow UPS User Guide

HP R12000 and R18000 DirectFlow UPS User Guide HP R12000 and R18000 DirectFlow UPS User Guide Abstract This document includes installation, configuration, and operation information for the HP R12000 and R18000 DirectFlow UPS. This document is for the

More information

CM705B - Universal Expander Module CM707B - Plug On Zone Expander Security Systems

CM705B - Universal Expander Module CM707B - Plug On Zone Expander Security Systems CM705B - Universal Expander Module CM707B - Plug On Zone Expander Security Systems EN Security System CM705B CM705B - Universal Expander Module The CM705B universal expander provides a cost effective way

More information

Argus Technologies Pathfinder10kW to Cordex3.6kW Shelf Replacement Procedure

Argus Technologies Pathfinder10kW to Cordex3.6kW Shelf Replacement Procedure Summary Argus Technologies Pathfinder10kW to Cordex3.6kW Shelf Replacement Procedure This is a general procedure on how to replace Pathfinder 48-10kW shelf with a Cordex 3.6kW shelf. Part Numbers 010-567-20

More information

Co-location Interconnect Power Panel (CIPP) Installation Guide

Co-location Interconnect Power Panel (CIPP) Installation Guide Co-location Interconnect Power Panel (CIPP) Installation Guide Document INS-97xxxxxxx This guide covers the following part numbers- Trimm 97xxxxxxx Series Table of Contents Section 1 - General Information...

More information

PC Tab INSTALLATION MANUAL SECURE-IT, INC. EAST LONGMEADOW, MA 01028. 800-451-7592 / 413-525-7039 (fax) 413-525-8807 http://www.secure-it.

PC Tab INSTALLATION MANUAL SECURE-IT, INC. EAST LONGMEADOW, MA 01028. 800-451-7592 / 413-525-7039 (fax) 413-525-8807 http://www.secure-it. PC Tab INSTALLATION MANUAL SECURE-IT, INC. EAST LONGMEADOW, MA 01028 800-451-7592 / 413-525-7039 (fax) 413-525-8807 http://www.secure-it.com IMPORTANT READ ENTIRE MANUAL BEFORE STARTING!!! System Overview

More information

How To Use A Power Supply Unit (Upu)

How To Use A Power Supply Unit (Upu) BRAVER UPS (Uninterruptible Power System) User s Manual Safety CAUTION! This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user replaceable parts.

More information

Troubleshooting Salt Generators

Troubleshooting Salt Generators Troubleshooting Salt Generators NOTE Turn off power to unit prior to attempting service or repair. Problems and Corrective Action Problem Possible Cause Corrective Action Low or no chlorine. Low stabilizer

More information

INSTRUCTION MANUAL FOR RECREATIONAL REFRIGERATOR/FREEZER MODEL

INSTRUCTION MANUAL FOR RECREATIONAL REFRIGERATOR/FREEZER MODEL INSTRUCTION MANUAL FOR RECREATIONAL REFRIGERATOR/FREEZER MODEL 15-LITER, 20-LITER, 35-LITER, 45-LITER, 60-LITER & 100-LITER SECTION 1 Basic Operation SECTION 2 Cleaning and Storing SECTION 3 Basic Trouble

More information

SERVICE MANUAL RESIDENTIAL ELECTRIC AND LIGHT DUTY COMMERCIAL ELECTRIC WATER HEATERS. Troubleshooting Guide and Instructions for Service

SERVICE MANUAL RESIDENTIAL ELECTRIC AND LIGHT DUTY COMMERCIAL ELECTRIC WATER HEATERS. Troubleshooting Guide and Instructions for Service RESIDENTIAL ELECTRIC AND LIGHT DUTY COMMERCIAL ELECTRIC WATER HEATERS SERVICE MANUAL Troubleshooting Guide and Instructions for Service (To be performed ONLY by qualified service providers) Models Covered

More information

IntelliBrite Controller (For IntelliBrite Pool, Spa and Landscape Lighting Fixtures) Installation and User s Guide

IntelliBrite Controller (For IntelliBrite Pool, Spa and Landscape Lighting Fixtures) Installation and User s Guide IntelliBrite Controller (For IntelliBrite Pool, Spa and Landscape Lighting Fixtures) Installation and User s Guide *619751* P/N 619751 - Rev C IMPORTANT SAFETY INSTRUCTIONS READ AND FOLLOW ALL INSTRUCTIONS

More information

Rain+Birdt. Simple To Set Timer (SST) Setup & Operation Instructions. English. 1-800- RAIN BIRD (800-724-6247) or visit www.rainbird.

Rain+Birdt. Simple To Set Timer (SST) Setup & Operation Instructions. English. 1-800- RAIN BIRD (800-724-6247) or visit www.rainbird. Rain+Birdt Simple To Set r (SST) Setup & Operation Instructions English Installation...2 Tools and Supplies Needed...2 Step 1. Mount r...2 Step 2. Connect Power...2 Indoor r...2 Outdoor r...2 Step 3. Connect

More information

Model UT713 OPERATING MANUAL

Model UT713 OPERATING MANUAL Model UT713 OPERATING MANUAL TABLE OF CONTENTS TITLE PAGE Introduction Unpacking Inspection Safety Information Turning the Calibrator On Simulating a Thermocouple Measuring a Thermocouple Simulating TC

More information

Yoke harness color code: Horizontal: Red & Blue. Vertical: Yellow & Brown, or Yellow & Green, or Violet and Brown.

Yoke harness color code: Horizontal: Red & Blue. Vertical: Yellow & Brown, or Yellow & Green, or Violet and Brown. CRT Safety and Engineering Notice YOKE PLUG ORIENTATION Note: Your CRT has been provided with a Factory mounted yoke. It has been designed to work with a variety of different chassis configurations, thus

More information

Part Name/Description Part Number Quantity. Power Cable 4000950-5 1

Part Name/Description Part Number Quantity. Power Cable 4000950-5 1 Note: Indented items indicate parts included in an assembly listed above Part Name/Description Part Number Quantity Power Cable 4000950-5 1 Raven Harness Adapter Kit 4100525 1 Installation Instructions

More information

BODi rs BD004 Series Bandwidth-on-Demand Internet with Reliability and Survivability. Quick Start Guide NOTES

BODi rs BD004 Series Bandwidth-on-Demand Internet with Reliability and Survivability. Quick Start Guide NOTES NOTES BODi rs BD004 Series Bandwidth-on-Demand Internet with Reliability and Survivability Quick Start Guide This is a Class A device and is not intended for use in a residential environment. Important

More information

The Child Reminder System Installation Manual

The Child Reminder System Installation Manual The Child Reminder System Installation Manual Revised June, 2006 Detailed installation information can be found at www.childreminder.com. Get through your installation quickly and easily by calling 1-888-330-6786

More information

Oil and Coolant Circulating Heating System. Model - OCSM

Oil and Coolant Circulating Heating System. Model - OCSM Oil and Coolant Circulating Heating System Model - OCSM Installation & Operation Manual 216280-000 REV 2 Identifying Your System The HOTSTART heating system is designed to heat fluids for use in marine

More information

StorTrends 3400 Hardware Guide for Onsite Support

StorTrends 3400 Hardware Guide for Onsite Support StorTrends 3400 Hardware Guide for Onsite Support MAN-3400-SS 11/21/2012 Copyright 1985-2012 American Megatrends, Inc. All rights reserved. American Megatrends, Inc. 5555 Oakbrook Parkway, Building 200

More information

Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001

Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001 Express5800/120Ed Rack Mount Kit Installation Procedures PN: 455-01607-001 Proprietary Notice and Liability Disclaimer The information disclosed in this document, including all designs and related materials,

More information

11.4.5 Optional Lab: Install, Configure, and Partition a Second Hard Drive in Windows Vista

11.4.5 Optional Lab: Install, Configure, and Partition a Second Hard Drive in Windows Vista 11.4.5 Optional Lab: Install, Configure, and Partition a Second Hard Drive in Windows Vista Introduction Print and complete this lab. In this lab, you will change the boot order, install a second hard

More information

Troubleshooting Guide for Jacks Down LED Lights

Troubleshooting Guide for Jacks Down LED Lights Troubleshooting Guide for Jacks Down LED Lights Equalizer Systems Auto-Level systems manufactured after 2005 feature a pressure switch system that monitors the retracted state of leveling jacks and any

More information

USER MANUAL CHARGING STATIONS FOR ELECTRIC VEHICLES

USER MANUAL CHARGING STATIONS FOR ELECTRIC VEHICLES USER MANUAL CHARGING STATIONS FOR ELECTRIC VEHICLES 204.CAxxx 204.CBxxx 204.UBxxx 204.WBxxx MP36289 1 ZP90856-GB-6 INDICE 1 SYSTEM DESCRIPTION... 4 1.1 MODES OF OPERATION... 4 2 USER INTERFACE... 6 2.1

More information

DR System Maintenance

DR System Maintenance 2015 DR System Maintenance Checklist and Guide This document will provide detailed instructions on how to effectively and properly maintain and backup your system for prolonged lifespan and reduced down

More information

B150 Series Commercial Brewer. Direct Plumb Kit

B150 Series Commercial Brewer. Direct Plumb Kit B150 Series Commercial Brewer Welcome Keurig Authorized Dealer! Direct Plumb Kit This manual provides: Installation of the Direct Plumb Kit (DPK) that allows you to connect your customer s B150 brewer

More information

Vertical Display and Storage B1350-2. SKOPE Gen2: Three Door Chiller

Vertical Display and Storage B1350-2. SKOPE Gen2: Three Door Chiller Vertical Display and Storage User Manual MAN1227 Rev. 3.0 March 2008 edition CONTACT ADDRESSES Designed and Manufactured by New Zealand SKOPE INDUSTRIES LIMITED PO Box 1091, Christchurch New Zealand Freephone:

More information

USER S MANUAL. MaxPower 400-600 UPS. Uninterruptible Power System 28-2MAXPO0018

USER S MANUAL. MaxPower 400-600 UPS. Uninterruptible Power System 28-2MAXPO0018 USER S MANUAL MaxPower 400-600 UPS Uninterruptible Power System 28-2MAXPO0018 IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS This manual contains important instructions for models MaxPower 400 and

More information

Advantium 2 Plus Alarm

Advantium 2 Plus Alarm ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia

More information

Electrical Safety Tester Verification

Electrical Safety Tester Verification Ensuring Validity of Regulatory Tests Verification of electrical safety testing equipment is a procedure that is often overlooked by manufacturers. Running test verification is crucial to ensuring that

More information

Pro Logic Version 4.10. Diagnostics Manual

Pro Logic Version 4.10. Diagnostics Manual Pro Logic Version 4.10 Diagnostics Manual Turbo Cell & Control Electronics 2009 Hayward Industries Table of Contents Important safety instructions Pg. 1 No Cell Power 1 & No Cell Power 2 Pg. 2 No Cell

More information

Residential Gateway (RG) Battery Backup

Residential Gateway (RG) Battery Backup Residential Gateway (RG) Battery Backup BU3DC000-12V User Manual Table of Contents Page Introduction... 3 Safety Information... 3 System Requirements... 3 Package Contents... 4 Recycling... 4 Product Layout

More information

Solar Home System. User Manual. AEH-SHS01-10W2L Solar Home System 2 Lamps

Solar Home System. User Manual. AEH-SHS01-10W2L Solar Home System 2 Lamps Solar Home System User Manual AEHSHS0110W2L Solar Home System 2 Lamps All rights reserved Specifications subject to change without prior notice 2 Dear Customer, Thank you for purchasing Schneider Electric

More information

Mobile Data Power Model: MDP-25

Mobile Data Power Model: MDP-25 Mobile Data Power Model: MDP-25 Topic Section Features... 2 Operational Features Summary... 2 Back-up Battery Power Internal Charger Voltage Spike Protection RF Noise Filtering Warning of Imminent Loss

More information

Electronic Pressure Switch EDS 300

Electronic Pressure Switch EDS 300 Electronic Pressure Switch EDS 300 User manual Page 2 of 16 Content 1. Functions of the EDS 300...3 2. Mounting...3 3. Operating keys on the membrane keypad...4 4. Digital display...4 5. Output function...5

More information

HM-W536 Install Guide

HM-W536 Install Guide HM-W536 Install Guide 9/13/2013 IMPORTANT SAFETY INSTRUCTIONS Warning - When using electrical devices, basic safety precautions should be followed to reduce the risk of fire, electrical shock or injury.

More information

MODEL 5010 DUAL CHANNEL SMOKE/FIRE DETECTION MODULE

MODEL 5010 DUAL CHANNEL SMOKE/FIRE DETECTION MODULE DESCRIPTION MODEL 5010 DUAL CHANNEL SMOKE/FIRE DETECTION MODULE DESCRIPTION The SST Model 5010 Two Channel Smoke/Fire Detection Module provides two independent detection input channels for the NOVA-5000

More information

VIEW. SLX300 SpeakerLinX IP Zone. Amplifier Installation and Setup Guide. AVoIP

VIEW. SLX300 SpeakerLinX IP Zone. Amplifier Installation and Setup Guide. AVoIP VIEW SLX300 SpeakerLinX IP Zone Amplifier Installation and Setup Guide TM AVoIP ClearOne 5225 Wiley Post Way Suite 500 Salt Lake City, UT 84116 Telephone 1.800.283.5936 1.801.974.3760 Tech Sales 1.800.705.2103

More information

CROWN BOILER COMPANY BWF SERIES BOILER CATEGORY I VENT KIT INSTALLATION AND OPERATING INSTRUCTIONS

CROWN BOILER COMPANY BWF SERIES BOILER CATEGORY I VENT KIT INSTALLATION AND OPERATING INSTRUCTIONS CROWN BOILER COMPANY BWF SERIES BOILER CATEGORY I VENT KIT INSTALLATION AND OPERATING INSTRUCTIONS WARNING Improper installation, adjustment, alteration, service, or maintenance of this product can cause

More information

3M Wrist Strap and Footwear Tester 740. User s Guide

3M Wrist Strap and Footwear Tester 740. User s Guide 3M Wrist Strap and Footwear Tester 740 User s Guide Table of Contents Section Page Safety Information...3 1.0 General...4 2.0 Description...4 3.0 Operation...5 4.0 Wall Mounting...5 5.0 Wrist Strap Test...6

More information

INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE SAFETY BULLETIN #23

INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE SAFETY BULLETIN #23 INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE SAFETY BULLETIN #23 GUIDELINES FOR WORKING WITH PORTABLE POWER DISTRIBUTION SYSTEMS AND OTHER ELECTRICAL EQUIPMENT "ADDENDUM C" WORKING WITH 480 VOLT SYSTEMS

More information

42U/45U 28" Wide Rack Installation & Service Guide

42U/45U 28 Wide Rack Installation & Service Guide 42U/45U 28" Wide Rack Installation & Service Guide 96-00171-005 Rev B Important Information Information in this document is subject to change without notice and does not represent a commitment on the part

More information

Networkfleet 3500 Product Line Installation Guide

Networkfleet 3500 Product Line Installation Guide Networkfleet 3500 Product Line Installation Guide Light/Medium Duty (L3500) Heavy Duty (H3500) Universal (U3500) www.networkcar.com/fleet Customer Care: (866) 227-7323 customercare@networkcar.com Table

More information

AUTODIALLER / QUICKDIALLER - SA132

AUTODIALLER / QUICKDIALLER - SA132 AUTODIALLER / QUICKDIALLER - SA132 INSTRUCTION LEAFLET ENGLISH www.thermomax-group.com CONTENTS 1 SETUP AT A GLANCE... 2 2 FOREWORD....... 3 3 INSTALLATION...... 4 4 KEYPAD AND INDICATORS...... 5 SETTING

More information

SERVICE MANUAL FOR 12 VDC WALL THERMOSTAT AIR CONDITIONING SYSTEMS ROOF TOP UNITS ONLY

SERVICE MANUAL FOR 12 VDC WALL THERMOSTAT AIR CONDITIONING SYSTEMS ROOF TOP UNITS ONLY RV Products Division SERVICE MANUAL FOR 12 VDC WALL THERMOSTAT AIR CONDITIONING SYSTEMS ROOF TOP UNITS ONLY Airxcel, Inc. RV Products Division P.O. Box 4020 Wichita, KS 67204 1976A376 (1-11) TABLE OF CONTENTS

More information

VOYAGER 570G. 744A Sprayer Control

VOYAGER 570G. 744A Sprayer Control VOYAGER 570G 744A Sprayer Control U S E R M A N U A L U S E R M A N U A L Table of Contents CHAPTER 1 - INTRODUCTION...1 SYSTEM CONFIGURATIONS...1 KIT CONTENTS...3 CONTROL HOUSING ASSEMBLY...5 CHAPTER

More information

LINE POWERED ADA TELEPHONE USER S MANUAL Use With Part Numbers 11-580, 11-581, 11-582, 11-583, 11-585,11-586 and 11-589

LINE POWERED ADA TELEPHONE USER S MANUAL Use With Part Numbers 11-580, 11-581, 11-582, 11-583, 11-585,11-586 and 11-589 LINE POWERED ADA TELEPHONE USER S MANUAL Use With Part Numbers 11-580, 11-581, 11-582, 11-583, 11-585,11-586 and 11-589 6200 Brent Drive, Toledo, Ohio 43611 Phone: 800-837-1066 Fax: 419-729-5764 Email:

More information

ATL Fuel Level Sender Probes

ATL Fuel Level Sender Probes T E C H N I C A L S P E C I F I C A T I O N The ATL EL-AD-151 (Resistance Output) and EL-AD-152 (Voltage Output) Fuel Level Senders are highly advanced sensors for continuously measuring the contents of

More information

OPL BASIC. Dosing System for Professional Laundry machines. Contents

OPL BASIC. Dosing System for Professional Laundry machines. Contents OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page

More information

MultiVOIP FX SIP Gateways

MultiVOIP FX SIP Gateways MultiVOIP FX SIP Gateways MVPFX2-2 MVPFX2-4 MVPFX2-8 Cabling Guide Cabling Guide MultiVOIP FX SIP Gateways, Models MVPFX2-2, MVPFX2-4, MVPFX2-8 82100090L Rev. A Copyright This publication may not be reproduced,

More information

162 CB CABLE TRACER. Filter Probe & Tone Generator INSTRUCTION MANUAL

162 CB CABLE TRACER. Filter Probe & Tone Generator INSTRUCTION MANUAL 162 CB CABLE TRACER Filter Probe & Tone Generator INSTRUCTION MANUAL INDEX PAGE 1. INTRODUCTION... 1 2. FILTER PROBE... 1-3 3. TONE GENERATOR... 3-6 4. SPECIFICATION... 7-8 5. MAINTENANCE... 8 1. INTRODUCTION

More information

tidesmarine Smart Seal Temperature Alarm System Generation II Installation Instructions Starboard side cable

tidesmarine Smart Seal Temperature Alarm System Generation II Installation Instructions Starboard side cable tidesmarine Smart Seal Temperature Alarm System Generation II Installation Instructions Starboard side cable Port side cable (with black cable tie attached) Power cable Preparing for Installation 1 Overall

More information

Portable Air Conditioner

Portable Air Conditioner Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE

More information

Installation Guide (No Router)

Installation Guide (No Router) Installation Guide (No Router) This installation guide will show you how to get your voip phone service working. This installation guide should be used if you have a standard DSL or cable modem and no

More information

466-1936 Rev E October 2004 ZZZ*(6HFXULW\FRP. Part No: 60-883-95R. CareGard. User Guide

466-1936 Rev E October 2004 ZZZ*(6HFXULW\FRP. Part No: 60-883-95R. CareGard. User Guide ) *(6HFXULW\ 466-1936 Rev E October 2004 ZZZ*(6HFXULW\FRP Part No: 60-883-95R CareGard User Guide FCC Notices FCC Part 15 Information to the User Changes or modifications not expressly approved by GE Security

More information

i410/i1010 AC/DC Current Clamp

i410/i1010 AC/DC Current Clamp /i1010 AC/DC Current Clamp Instruction Sheet Safety Information XW Read First: Safety Information To ensure safe operation and service of the current clamp, follow these instructions: Read all operating

More information

Owner s Manual. Not suitable for mobile applications. Important Safety Instructions 2. Quick Installation 3. Basic Operation 4. Storage & Service 7

Owner s Manual. Not suitable for mobile applications. Important Safety Instructions 2. Quick Installation 3. Basic Operation 4. Storage & Service 7 Register online today for a chance to win a FREE Tripp Lite product! www.tripplite.com/warranty Owner s Manual Internet Office & BC Personal UPS Systems Not suitable for mobile applications. Important

More information

INFRARED QUARTZ WALL HEATER

INFRARED QUARTZ WALL HEATER INFRARED QUARTZ WALL HEATER MODEL NO: IQ2000 PART NO: 6939004 MOUNTING & OPERATION INSTRUCTIONS GC0715 INTRODUCTION Thank you for purchasing this CLARKE Infrared Wall Heater. Before attempting to use this

More information

Drayton Digistat +2RF/+3RF

Drayton Digistat +2RF/+3RF /+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer

More information

Safety & Environmental Protection Services Guidance Note

Safety & Environmental Protection Services Guidance Note Safety & Environmental Protection Services Guidance Note A STEP-BY-STEP GUIDE TO VISUAL INSPECTION OF ELECTRICAL EQUIPMENT Why Inspect? The Health and Safety at Work etc Act 1974 and the Electricity at

More information

Installation and Operation Guide PD4600 Series Converter Replacement

Installation and Operation Guide PD4600 Series Converter Replacement Installation and Operation Guide PD4600 Series Converter Replacement Extended warranties are available for purchase at www.progressivedyn.com Member Thank you for selecting Progressive Dynamics as your

More information

MIC-WKT and MIC-WKT-IR

MIC-WKT and MIC-WKT-IR MIC-WKT and MIC-WKT-IR Installation Manual Bosch Security Systems EN Installation and Operation Manual MIC-WKTI and MIC-WKT-IR Installation Manual EN 2 MIC-WKT and MIC-WKT-IR Washer Pump Drive Card Kits

More information

PowerFlex 700H and 700S AC Drives Frame 11 Main Fan Capacitor Replacement Kit

PowerFlex 700H and 700S AC Drives Frame 11 Main Fan Capacitor Replacement Kit Installation Instructions PowerFlex 700H and 700S AC Drives Frame 11 Main Fan Capacitor Replacement Kit ATTENTION: The sheet metal cover and mounting screws on the ASIC Board located on the power structure

More information

Manual for Fire Suppression & Methane Detection System

Manual for Fire Suppression & Methane Detection System Manual for Fire Suppression & Methane Detection System Fogmaker North America Post address: 150 Gordon Dr Exton, PA 19341 Delivery address: 150 Gordon Dr Exton, PA 19341 Tel: 610-265-3610 Fax: 610-265-8327

More information

IP DSLAM IDL-2402. Quick Installation Guide

IP DSLAM IDL-2402. Quick Installation Guide IP DSLAM IDL-2402 Quick Installation Guide Table of Contents Package Contents... 3 Overview... 4 Setup the IDL series IP DSLAM... 5 Safety Instruction... 5 Hardware Installation... 6 WEB Configuration...

More information

ReadyNet Easy Jack 2 Voice/Data and Data Only Owner s Manual PX-211d and PX-211v

ReadyNet Easy Jack 2 Voice/Data and Data Only Owner s Manual PX-211d and PX-211v ReadyNet Easy Jack 2 Voice/Data and Data Only Owner s Manual PX-211d and PX-211v Phonex Broadband Corporation dba ReadyNet 6952 High Tech Drive Midvale, Utah 84047 801.566.0100 Phone 801.566.0880 Fax www.readynetsolutions.com

More information

Troubleshooting Guide, Freedom and Fleet Power Inverter/Chargers

Troubleshooting Guide, Freedom and Fleet Power Inverter/Chargers Technical Note Freedom/Fleet Power 512-0084-01-01 Rev 1 Troubleshooting Guide, Freedom and Fleet Power Inverter/Chargers Overview This document is a guide for troubleshooting inverters, battery chargers,

More information

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting

More information

Battery Charger For Nickel Cadmium and Nickel-Metal Hydride Rechargeable Batteries Model PSN Series

Battery Charger For Nickel Cadmium and Nickel-Metal Hydride Rechargeable Batteries Model PSN Series Battery Charger For Nickel Cadmium and Nickel-Metal Hydride Rechargeable Batteries Model PSN Series Operating Instructions WARNING CONCERNING THE REMOVAL OF COVER: CAUTION: TO PREVENT THE RISK OF ELECTRIC

More information

Name of Equipment Silver King Model SKMCD1P/C1. This equipment chapter is to be inserted in the appropriate section of the Equipment Manual.

Name of Equipment Silver King Model SKMCD1P/C1. This equipment chapter is to be inserted in the appropriate section of the Equipment Manual. Name of Equipment Silver King Model SKMCD1P/C1 This equipment chapter is to be inserted in the appropriate section of the Equipment Manual. Manufactured exclusively for McDonald s By Silver King Refrigeration,

More information

Automatic Phone-Out Home Monitoring Systems

Automatic Phone-Out Home Monitoring Systems Automatic Phone-Out Home Monitoring Systems Power Outage and Freeze Alarm Model Number: THP202 Power Outage, Freeze and Flood Alarm Product Description Model Number: THP201 These monitoring systems are

More information

BX7000 Multi-Access Gateway Getting Started Guide

BX7000 Multi-Access Gateway Getting Started Guide BX7000 Multi-Access Gateway Getting Started Guide This guide provides instructions on connecting the Juniper Networks BX Series 7000 Multi-Access Gateway to a network. For more information, see the BX7000

More information

OASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE

OASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE OASIS-PLUS 120V INFORMATION & OPERATING INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE 56041852 WARNING:

More information

MultiVOIP FX FXS-Only SIP Gateways

MultiVOIP FX FXS-Only SIP Gateways MultiVOIP FX FXS-Only SIP Gateways MVPFXS-8 MVPFXS-16 MVPFXS-24 Cabling Guide MultiVOIP MVPFXS-8/16/24 Cabling Guide Cabling Guide MultiVOIP FX SIP FXS-Only Gateways Models MVPFXS-8, MVPFXS-16, MVPFXS-24

More information

Intersection Preassembled 19-inch Rack

Intersection Preassembled 19-inch Rack Intersection Preassembled 19-inch Rack The intersection preassembled 19-inch rack comes fully assembled, wired and ready to be installed in a server rack. The rack provides communications, power conversion

More information

FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE. Battery operation instructions:

FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE. Battery operation instructions: FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE Battery operation instructions: 1. The removing and inserting of batteries is to be conducted by the adults only. Unscrew the screw holding the battery pack s cover

More information

Union County Public Schools. Facilities Department. Electrical. Safe Work Practices

Union County Public Schools. Facilities Department. Electrical. Safe Work Practices 1 Union County Public Schools Facilities Department Electrical Safe Work Practices 2 Purpose In accordance with OSHA Standards 1910.331-335, safety-related work practices shall be used by Union County

More information

User s Manual AURORA 1.2K/2.2K

User s Manual AURORA 1.2K/2.2K User s Manual AURORA 1.2K/2.2K Uninterruptible Power System Safety CAUTION This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user serviceable

More information

Inspection and testing of portable electrical equipment

Inspection and testing of portable electrical equipment February 2008 Inspection and testing of portable electrical equipment This is a guidance document for Schools and Support Services. Units should use the guidance to help them develop their own local health

More information

Installation and Operation Guide for PD4000 Series Power Control Center

Installation and Operation Guide for PD4000 Series Power Control Center Extended warranties are available for purchase at www.progressivedyn.com Installation and Operation Guide for PD4000 Series Power Control Center Member Thank you for selecting Progressive Dynamics as your

More information

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS ELECTRICAL SAFETY This appliance is for indoor use only. Always turn off the unit and unplug

More information

EKOS Cart. Instructions for Use

EKOS Cart. Instructions for Use EKOS Cart Instructions for Use EKOS Corporation 11911 North Creek Parkway South Bothell, WA 98011 USA (425) 415-3100 (tel) (425) 415-3102 (fax) info@ekoscorp.com (e-mail) - 1-4913-002 REV E Intended Use

More information

Evolution Digital HD Set-Top Box Important Safety Instructions

Evolution Digital HD Set-Top Box Important Safety Instructions Evolution Digital HD Set-Top Box Important Safety Instructions 1. Read these instructions. 2. Keep these instructions. 3. Heed all warnings. 4. Follow all instructions. 5. Do not use this apparatus near

More information

CAUTION RISK OF ELECTRIC SHOCK NO NOT OPEN

CAUTION RISK OF ELECTRIC SHOCK NO NOT OPEN Evolution Digital HD Set-Top Box Important Safety Instructions 1. Read these instructions. 2. Keep these instructions. 3. Heed all warnings. 4. Follow all instructions. 5. Do not use this apparatus near

More information

LESTRONIC DV AUTOMATIC 12 OR 24 VOLT DUAL OUTPUT DUAL MODE BATTERY CHARGER MODEL 16350 TYPE 12/24EL40-10ET

LESTRONIC DV AUTOMATIC 12 OR 24 VOLT DUAL OUTPUT DUAL MODE BATTERY CHARGER MODEL 16350 TYPE 12/24EL40-10ET LESTRONIC DV AUTOMATIC 12 OR 2 VOLT DUAL OUTPUT DUAL MODE BATTERY CHARGER MODEL 16350 TYPE 12/2EL0-10ET PLEASE SAVE THESE IMPORTANT SAFETY AND OPERATING INSTRUCTIONS For correct operation of the equipment,

More information

100 Series Maxi Internal Filters. Instruction Manual

100 Series Maxi Internal Filters. Instruction Manual 100 Series Maxi Internal Filters Instruction Manual Important Safeguards This appliance is not intended for use by children or infirm persons without supervision. Young children should be supervised to

More information

No Phone Line II-48 System

No Phone Line II-48 System Used in conjunction with an EntraGuard Telephone Entry System, the -48 provides No Phone Line communication and door or gate control through existing building telephone lines without the need to purchase

More information

DC400 Dispensing Cutoff System

DC400 Dispensing Cutoff System DC400 Dispensing Cutoff System DC404 and DC406 Installation Instructions Franklin Fueling Systems 3760 Marsh Rd. Madison, WI 53718 USA Tel: +1 608 838 8786 800 225 9787 Fax: +1 608 838 6433 www.franklinfueling.com

More information

Pet hair clipper. Model 96822. Diagrams within this manual may not be drawn proportionally.

Pet hair clipper. Model 96822. Diagrams within this manual may not be drawn proportionally. Pet hair clipper Model 96822 Cleaning And Operation Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly from the

More information

2013 VTech Printed in China 91-009656-000 US

2013 VTech Printed in China 91-009656-000 US Rechargeable Power Pack User s Manual 2013 VTech Printed in China 91-009656-000 US INTRODUCTION The Rechargeable Power Pack makes it easier than ever to keep the InnoTab 3 or InnoTab 3S charged and ready

More information

Online UPS. PowerWalker VFI 1000R/1U

Online UPS. PowerWalker VFI 1000R/1U Online UPS PowerWalker VFI 1000R/1U Manual Uninterruptible Power Supply System Table of Contents 1. Important Safety Warning... 1 1-1. Transportation... 1 1-2. Preparation... 1 1-3. Installation... 1 1-4.

More information

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &

More information