Conversion of Agilent EI GC/MSD Systems To Hydrogen Carrier Gas. Agilent Technologies

Similar documents
Addressing the World Helium Shortage For Gas Chromatography

Analysis of Drugs by using the Ultra Inert Inlet Liners with Wool

The Use of Hydrogen Carrier Gas for GC/MS

23 The Thermal Conductivity Detector

Tuning & Mass Calibration

5973 inert Mass Selective Detector

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Hydrogen as an alternative to Helium for gas chromatography.

Agilent G1701DA MSD ChemStation

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

The Electron Capture Detector (ECD)

Expectations for GC-MS Lab

Background Information

Analysis of Liquid Samples on the Agilent GC-MS

Gas Chromatography Liner Selection Guide

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

Electrospray Ion Trap Mass Spectrometry. Introduction

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine

Maintaining Your GC/MS System. Operate your Agilent GC/MS System with maximum efficiency. Diagnose tough MSD problems. Discover the best GC/MS columns

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

CHROMOCTANE O N LINE ANALYZER

GC Troubleshooting 10. How to use this Troubleshooting Help system GC Troubleshooting - Top Level Generic problem descriptions...

Split / Splitless Injection for Capillary GC

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Agilent 7697A Headspace Sampler Data Sheet

Pesticide Analysis by Mass Spectrometry

Gas Chromatograph GC Consumables. p/n:

Transformer Oil Gas Analysis

Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS

Clarus 580 Gas Chromatograph

Thermal Conductivity Detector

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

How To Use Gc-Ms

PLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative

SCION SQ Series GC-MS

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Micro GC Fusion Pre-Installation Checklist

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

13C NMR Spectroscopy

Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS

Fractional Distillation and Gas Chromatography

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# RFQ)

Appendix 5 Overview of requirements in English

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report

Opiates in Urine by SAMHSA GC/MS

Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder

Analyzing Small Molecules by EI and GC-MS. July 2014

Validation of process gas systems

Mass Spectrometry. Fundamental GC-MS. GC Considerations

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA ABSTRACT

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

TECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES. Randall Messman SW Regional Sales Manager

Operating Hints for Using Split/Splitless Injectors

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

Micromass LCT User s Guide

How To Test For Contamination In Large Volume Water

Paper Chromatography: Separation and Identification of Five Metal Cations

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

AMD Analysis & Technology AG

Chromatograph Maintenance and Troubleshooting Class # 5050

Setting up a Quantitative Analysis MS ChemStation

TechNote 3/2002. The Use of Different PTV Inlet Liner Types for Trapping Alkanes, Aromatics and Oxygenated Compounds During Thermal Desorption

New drugs of abuse? A case study on Phenazepam & Methoxetamine

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits

Chapter 28: High-Performance Liquid Chromatography (HPLC)

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

Micro Cryo-Trap Model 971

Agilent 5975 Series MSD

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS

EQuan MAX Getting Connected Field Service Guide

Protocol: HPLC (amino acids)

With proper calibration, the upper limit may be extended as needed.

EnAlgae SOP: Biogas analysis

AppNote 6/2011. Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS KEYWORDS ABSTRACT

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Application. Gas Chromatography September 1997

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Process Instrumentation

Analysis of Phthalate Esters in Children's Toys Using GC-MS

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Cryogenic Vacuum Pumps

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Thermo Scientific ConFlo IV Universal Interface

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

ChromQuest 5.0 Chromatography Data System

Transcription:

Conversion of Agilent EI GC/MSD Systems To Hydrogen Carrier Gas Agilent Technologies 1

Introduction: Converting From He To H 2 Many GC/MS users are considering changing from helium to hydrogen carrier gas due to price/availability problems with helium. This talk describes the steps recommended for converting EI GC/MS methods. It is important to recognize the differences with using hydrogen carrier. Time should be allotted for adapting the method, optimization, and resolving potential problems. Areas that will need attention include: choice of supply of H 2 Remove/bypass carrier gas filter (may be re-installed later if needed) GC/MSD hardware changes choosing new chromatographic conditions a 2-5 fold reduction in signal-to-noise ratio changes in spectra and abundance ratios for some compounds activity and reactivity with some analytes Page 2

Introduction: Converting From He To H 2 Methods that will generally require less optimization include analytes that: are durable compounds are at higher concentrations are analyzed with split injections are derivatized Methods that will generally require more optimization include analytes that: are fragile compounds are at trace concentrations Allow time for necessary updates to SOPs and validation. Page 3

4 H 2 Safety

Letter From Agilent Safety Engineer. Describes H 2 Safety Features of Agilent GCs 5

Designed for Reliability H 2 Safety Safety Shutdown When gas pressure set points are not met, the valve and heater are shut off to prevent explosion Flow Limiting Frit If valve fails in open position, inlet frit limits the flow Oven ON/OFF Sequence Fan purges the oven before turning on heater to remove any collected H 2 Explosion Test GC designed to contain parts in case of explosion Page 6

Safety First: Read This Before You Start G3170-90010 Page 7

Source of H 2 Carrier and Plumbing 8

Source of Hydrogen: Use A Hydrogen Generator Higher initial expense than cylinders, much lower cost over time Very clean H 2, >99.9999% available More consistant purity Safety considerations H 2 is only generated at needed pressure (like 40 psig) Flow is limited (like 250 ml/min) Auto-shutdown if setpoint pressure cannot be maintained Minimal stored gas (like 50 ml at 40 psig) of H 2 at any one time Make sure to buy a good one with a low spec for water and oxygen 9

10 This is an example of the type of H 2 generator to use. Look for generators with the > 99.9999% specification and low individual specs on water and oxygen.

11

Plumbing The Instrument Chromatographic quality stainless steel tubing is often recommended for H 2 plumbing and is probably the best choice if available. Users may have to follow local codes or internal company guidelines. We have also used new 1/8 th copper that has been cleaned for GC use. Dirty tubing will cause huge contamination problems, as H 2 appears to carry dirt out of metal more than He does. Don t use really old copper tubing, as it becomes brittle and can break. Note that MSD leak checks will not always find big outgoing leaks. Leak check when complete with electronic leak detector. When plumbing a H 2 generator, start out with no traps. Only add traps if needed. Make sure the water and oxygen levels are low enough. 12 Confidentiality Label September 6, 2012

Split Vent and Septum Purge Vent Should Be Connected to Exhaust 13

MS Components Needed Before Conversion: Magnet and Drawout Lens 14

Check The Magnet In The 5975 If the magnet in the 5975 does NOT have a serial number along the left edge (as viewed through the source window), it is suitable for use with Helium, but not Hydrogen. You should contact your Agilent service engineer to change it to a Helium and Hydrogen compatible magnet before converting the system to H 2 carrier. Note: all 5973 magnets are compatible with both He and H2 15

Replace The Standard 3 mm Drawout Lens With The Optional 6 mm When Using H 2 Carrier G2589-20045 6 mm Drawout Plate, Inert Standard 3 mm Drawout Plate, Inert 16 Confidentiality Label September 6, 2012

Choosing A Column And Conditions 17 Confidentiality Label September 6, 2012

MSD Pumping Capacity For H 2 Is Less Than For He This limits column choice Complete info on H 2 pumping specs are not yet available These are approximate values for the 5975. Max flow of H 2 to maintain reasonable source pressure: Performance turbo: 2 ml/min Standard turbo: 1 ml/min Diffusion pump: 0.75 ml/min Pressure pulsing: turbos <= 3 ml/min, diffusion <= 2.5 ml/min It is very helpful to have an ion gauge on the MS to monitor the vacuum vs column flow. Try to avoid flows that produce pressures higher than 5 x 10-5 torr. You can get useful data above this pressure, but performance starts to degrade rapidly 18

Guidelines For Choosing Column Dimensions And Flows For Method Converted From He To H 2 Determine max flow of H 2 into MS that will give source pressure of 5 x 10-5 torr or less source pressure. This is your max column flow. Choose column dimensions at initial oven temp of method to give: A flow <= max column flow for vacuum pump A flow >= min column flow for efficiency An inlet pressure of at least 5 psig Keeping a temperature ramp of the same number of C/void time will give similar elution order. Use the method translator for this. These are only approximate guidelines. Sometimes you have to deviate from them 19

Van Deemter Curves If the flow of He or H 2 is too low, you loose efficiency MUCH FASTER than if it is too high. This is why operating at or above minimum linear velocity is important. Use 35 cm/sec as minimum for H 2 20

Column Choice When in doubt (or a hurry), use: Use 20m x 180 um column and set constant flow at 0.3-0.5 ml/min. Try to get a linear velocity of at least 35 cm/sec Try to use the same phase and phase ratio to get similar elution order. Smaller bore columns have lower capacity. You may have overloading on high level cals or samples You may need to use pulsed injection to load sample into small bore column with low flow rate. The flow range for your H 2 setup is limited on the low side by the flow to get 35 cm/sec and on the high side by the vacuum pump capacity 21

Using The Flow Calculator. Set Linear Velocity to 35 For H 2 Carrier. If InletPressure Too Low, Raise Flow Inlet pressure is too low Set inlet pressure to 5 psig 22

Example: 30m x 0.25 mm id x 0.25 um Checkout Column With Perf Turbo Meets flow requirements for pump and column efficiency with 2 x speed gain 23

Example: 20m x 0.18 mm id x 0.18 um Column With Perf Turbo and Std Turbo Meets flow requirements for std turbo pump. Flow here is chosen to give a 2.5 x speed gain. 24

Status Of Calculators For Column Selection HP FlowCalc 2.0 and HP Method Translator used in earlier slides DO NOT WORK on Win 7 Newer versions of these can also be found in: GC/GCMS Instrument Utilities Kit. This CD ships with every Agilent GC, GC/MS, and autosampler. It also has manuals for GC and a firmware updating tool. 25

Initial Startup With Hydrogen 26

Initial Problems With Switching GC/MS Methods to H 2 High background that looks like hydrocarbons Reduced signal to noise (worse MDL) Significant tailing for many compounds Page 27

Switch to H 2. Early problem- High Background You will see this throughout work with H 2. It will get much smaller with time and use. It used to takes days or weeks for this to fall to acceptable levels. Agilent has a new conditioning protocol to clean up the background much faster (see later slide). Page 28

Tox Checkout Sample Abundance Time--> 2000000 1800000 1600000 1400000 1200000 1000000 800000 600000 400000 200000 0 Immediately after conversion to H 2, peakshapes on TIC are poor and tailing badly. 20m x 0.18mm id x 0.18 um DB-5MSUI 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 Hydrogen Abundance 3000000 2800000 2600000 2400000 2200000 2000000 1800000 1600000 1400000 1200000 1000000 800000 600000 400000 200000 0 Time--> 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 Helium Page 29

GC/MS Toxicology Checkout Mixture 5190-0471 1 Amphetamine 2 Phentermine 3 Methamphetamine 4 Nicotine 5 Methylenedioxyamphetamine(MDA) 6 Methylenedioxymethamphetamine(MDMA) 7 Methylenedioxyethylamphetamine 8 Meperidine 9 Phencyclidine 10 Methadone 11 Cocaine 12 SKF-525a (RTL Compound) 13 Oxazepam 9 14 Codeine 5 ng/ul each compound 17 15 Lorazepam 16 Diazepam 17 Hydrocodone 18 Tetrahydrocannabinol 19 Oxycodone 20 Temazepam 21 Flunitrazepam 22 Diacetylmorphine 23 Nitrazepam 24 Clonazepam 25 Alprazolam 26 Verapamil 27 Strychnine 28 Trazodone 1 2 3 4 6 7 8 10 12 11 14 16 15 18 21 20 19 22 23 24 25 26 27 28 5 13 2 4 6 8 10 12 14 16 18 30

Tailing Peaks: There is More Going On Here Than Simple Tailing Cocaine Abundance 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 Peak Spectrum Abundance Tail Spectrum 24000 Average of 8.018 to 8.036 min.: ToxCKO_H2_4.D\data.ms Average of 8.048 to 8.085 min.: ToxCKO_H2_4.D\data.ms 82 91 22000 182 14000 20000 18000 12000 16000 10000 14000 Cocaine 12000 8000 Not Cocaine 42 10000 6000 78 105 8000 41 6000 4000 64 55 4000 68 2000 115 303 128 167 2000 122 152 167 198 272 141 154 191 207221 135 242 215 265 281 300 0 230244 259 288 327341 0 314327341 40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 m/z--> m/z--> Page 31

Main Ions Of Cocaine Do Not Tail. Tail Made Up of Other Ions Tail is different than peak. May be formed from catalytically shredded main component. This problem is reduced with time Cocaine Tail is not cocaine ions Page 32

Minimizing The Time For Chromatography and Background To Stabilize Have a good source of clean hydrogen like a >99.9999% purity hydrogen generator. Keep H 2 flow limited to value suitable for your pumping system Use 6 mm drawout lens After setup, purging and pumpdown: Set the source to max temp for your source (check to see what your s is) Reduce the EMV to <= 800V Leave the FILAMENT ON overnight. This cleans up background rapidly. Peakshape will be much better and background will be much lower in the morning. Lower the source temp to method value, retune, and run some samples It is a good idea to have an extra set of filaments on hand in case one burns out. This hasn t been a problem, but a good idea. Page 33

H 2 Method With 6 mm Drawout And Overnite Cleaned Source. Peakshape Much Better and Background is Lower. Abundance 2600000 2400000 2200000 2000000 1800000 1600000 1400000 1200000 1000000 800000 600000 400000 200000 Time--> 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 Page 34

Performance Expectations Signal-to-noise ratio is usually worse by about 2-5 x. This obviously varies from compound to compound While most spectra remain same, there are always exceptions. Users should check the reference spectra for important targets to make sure they have not changed Same comment for target/qualifier ratios Some compounds may disappear at low levels. Examples include: some nitro and oxygen containing compounds (alcohols, aldehydes, ketones) like those found in flavor samples 35

Other Considerations: H 2 is not an inert gas, and therefore inertness problems will still exist or be worse. Some things to consider: Use the lowest inlet temp that works (to reduce reactions with H 2 ) Use pulsed injection, especially with small bore columns Consider using an MMI in cold splitless mode for fragile compounds Using a deactivated S/SL weldments might help (but don t ever scrub it) PN:G3452-60570 (bottom) PN:G3452-60586 (top) Avoid using methylene chloride as a solvent (especially wet). At higher inlet temps (like 300C), HCl is formed and causes problems. It can remove the gold from the seal and corrode the weldment. If DCM must be used, use lowest inlet temp and maybe a deactivated S/SL weldment (or MMI) Also avoid carbon dissulfide as a solvent Use liners with a taper at bottom to minimize sample contact with gold seal Use Agilent Ultra Inert Liners. 36

Hydrogen Carrier Summary 1. Get a GOOD hydrogen generator and plumb it properly 2. Pick a column that meets the flow needs of method and MSD pump 3. Do magnet update (if necessary) and install 6 mm drawout lens 4. After purging and pumpdown, set source to max temp and quad to 200C. Leave filaments on all night. 5. Reset source temp, retune, and run samples. Run practice samples for a few days to let system stabilize