GC Troubleshooting 10. How to use this Troubleshooting Help system GC Troubleshooting - Top Level Generic problem descriptions...
|
|
|
- Elfreda Collins
- 9 years ago
- Views:
Transcription
1 Table of Contents GC Troubleshooting 10 How to use this Troubleshooting Help system GC Troubleshooting - Top Level Generic problem descriptions Baseline related problems Peak related problems Results related problems System-component related problems Flame Ionization Detector (FID) Troubleshooting Electron Capture Detector (ECD) Troubleshooting Nitrogen Phosphorous Detector (NPD) Troubleshooting13 Flame Photometric Detector (FPD) Troubleshooting. 13 Photoionization Detector (PID) Troubleshooting Thermal Conductivity Detector (TCD) Troubleshooting15 Mass Spectrometer (MS) Troubleshooting PTV Injector troubleshooting Gas Sampling Valve (GSV) troubleshooting Leak Checking 17 Checking for Leaks Leak-checking DPFC-equipped systems Leak-checking non-dpfc systems Performing an Automatic Leak Check Performing an Automatic Column Evaluation Leak testing the lines from the Gas Source to the GC
2 2 Identifying and Removing Leaks... 20
3 Baseline related problems with specific system components 20 Baseline drifting 20 Baseline falling 21 Baseline falling away slowly after a high initial value 21 Baseline rising 22 Baseline rising under temperature program control 22 Baseline - high standing current 22 Baseline irregular shape: dip after solvent peak23 Baseline irregular shape: S-shaped 23 Baseline irregular shape: square waves 24 Baseline high frequency noise 24 Baseline spiking 25 Peak related problems with specific system components 25 Peaks broadening 26 Peaks clipping at bottom 26 Peaks clipping at top 27 Double Peaks 27 3
4 Peak Fronting (excessive slope on left side) 27 Ghost Peaks 28 Broad Ghost Peaks 28 Irregular, Chair-shaped Peaks 29 Negative Peaks 29 No Peaks after solvent peak 29 No Peaks at all 30 Sample Peak Tailing (excessive slope on right side) 31 Solvent Peak Tailing (excessive slope on right side) 31 Unresolved Peaks 32 Discrete high-intensity contaminant peaks 32 Results related problems with specific system components 33 Low reproducibility of peaks area 33 Poor sensitivity with increased retention time 34 Poor sensitivity with normal retention time 34 Retention times decreasing 35 Retention times increasing 35 4
5 Low reproducibility of retention times 36 Retention times are inconsistent 36 FID Sensitivity problems 37 FID gases 37 FID detector gas flow rates 37 FID operating conditions 38 FID Flame Ignition problems 38 FID Contamination problems 39 ECD Contamination problems 39 Warning: Radioactivity 40 ECD gases 40 ECD Sensitivity problems 40 High base frequency (ECD) 41 Negative dips after peaks (ECD) 41 Baseline drift with changing pulse voltage (ECD)42 NPD gases 43 NPD Sensitivity problems 42 NPD operating conditions 43 No NPD detector response 43 NPD detector response lower than expected 43 5
6 High background level (NPD) 44 NPD shows FID-like response for solvent and other carbon based compounds 44 Solvent quenching effect (NPD) 45 Unstable baseline (NPD) 45 Low carbon rejection (NPD) 45 FPD gases 46 FPD Sensitivity problems 45 FPD gas flow rates 46 High standing current and noise (FPD) 46 No standing current (FPD) 47 Unstable and excessively noisy baseline (FPD) 47 FPD temperature does not reach the set point 47 Low sensitivity (FPD) 47 Low sensitivity and water droplets generate between the heat shields (FPD) 49 UV lamps for PID 49 PID Sensitivity problems 50 PID Contamination problems 50 UV lamp does not light immediately (PID) 50 6
7 UV lamp does not light at all (PID) 50 No standing current (PID) 51 Tailing of solvent peak (PID) 52 Tailing of sample peaks (PID) 52 Low sensitivity (PID) 52 TCD Sensitivity problems 53 Negative Peaks (TCD) 53 Baseline Fluctuation (TCD) 53 Baseline Drift (TCD) 54 Low Sensitivity (TCD) 55 TCD does not work 56 High Baseline (MS) 56 Peak Tailing (MS) 56 Low Sensitivity (MS) 57 Poor Reproducibility of Results (MS) 57 PTV Sensitivity problems 58 PTV Discrimination problems 58 7
8 Discrimination of heavy compounds in splitless mode (PTV) 58 Discrimination of volatile compounds in splitless mode (PTV) 60 Discrimination of volatile compounds in solvent split mode (PTV) 61 Discrimination in split mode (PTV) 61 Excessive broadening of solvent peak (PTV) 62 Sample degradation (PTV) 62 PTV Large Volume - Poor reproducibility of results 63 PTV Large Volume - Discrimination of volatile compounds 64 PTV Large Volume - Sample degradation 65 Analyses results are not reproducible (GSV) 65 No signal at all (GSV) 66 GSV does not work 66 Checking for leaks - Capillary Columns, DPFC, DGFC 67 Checking for leaks - Capillary Columns, DPFC, Non-DGFC 67 Checking for leaks - Capillary Columns, Non-DPFC, Non- DGFC 68 8
9 Checking for leaks - Packed Columns, DPFC, DGFC 69 Checking for leaks - Packed Columns, DPFC, Non-DGFC 69 Checking for leaks - Packed Columns, Non-DPFC, Non- DGFC 70 Multi GC system 71 9
10 GC Troubleshooting How to use this Troubleshooting Help system This Help system is designed to help you diagnose and resolve problems that may occur with your Gas Chromatograph system. Because there are a number of possible variations in the configuration of your system, the problem descriptions have been grouped partly by the general type of problem, and partly by the specific hardware components. The main general problem type groups are Baseline related problems, Peak related problems, and Results related problems. All the suggested possible causes and required remedies within these general problems areas are applicable whatever your system configuration. However, you will also find that there may be some additional possible causes that are applicable only to one particular system configuration. For example, the Baseline related problems area lists a number of generic possible causes. However, if your system utilises a Nitrogen Phosphorous Detector (NPD), you will find an additional possible cause listed under the NPD Troubleshooting section - Unstable Baseline (NPD). To make sure that you don t overlook these additional problem descriptions, there is a See also list in each of the general problem areas, which lists the problems in the area that relate to specific system components. The suggested method of use, therefore, is to start with the general problem list for the type of problem you are experiencing. If this doesn t resolve the problem, then try the See also list for any additional problem descriptions that apply to your system configuration. Finally, try the troubleshooting sections that refer specifically to the components in your system. You can also use the Index and Search tabs on the Navigation panel to search for keywords that describe your problem. If you encounter a problem that is not described here, or have a problem that is not resolved by the remedy suggested, contact your local service representative who will be able to provide additional advice. Click here to go to the top level troubleshooting list. GC Troubleshooting - Top Level This Help file is designed to help you diagnose and resolve problems that may occur from time-to-time with your GC system. Click here for advice on how to use this help system. Choose from the following list the area that matches your category of problem. The first three problem areas describe general problem types that are independent of the particular system configuration being used. The remainder of the list takes you to troubleshooting topics for specific system components. To move through the problem areas, and also to move through the separate problem descriptions within each problem area, you can use the Browse buttons and. Baseline related problems, such as rising or falling levels of Baseline, drifting, or Baseline noise. Peak related problems, such as Ghost Peaks, Tailing, or No Peaks at all. Results related problems, such as varying Retention Times, Poor Sensitivity, or Poor Reproducibility. Flame Ionization Detector (FID) Troubleshooting Electron Capture Detector (ECD) Troubleshooting Nitrogen Phosphorous Detector (NPD) Troubleshooting 10
11 Flame Photometric Detector (FPD) Troubleshooting Photoionization Detector (PID) Troubleshooting Thermal Conductivity Detector (TCD) Troubleshooting Mass Spectrometer (MS) Troubleshooting PTV Injector troubleshooting Gas Sampling Valve troubleshooting Within each of these main problem-type areas, specific problem descriptions are listed. When you select the problem description that best matches your problem, you will see a list of the possible causes of the problem, with suggestions as to how each cause can be remedied. If you encounter a problem that is not described here, or have a problem that is not resolved by the remedy suggested, contact your local service representative who will be able to provide additional advice. Generic problem descriptions Baseline related problems Select the problem description that best matches your problem: Baseline drifting Baseline falling Baseline falling away slowly after a high initial value Baseline rising Baseline rising under temperature program control Baseline - high standing current Baseline irregular shape: dip after solvent peak Baseline irregular shape: S-shaped Baseline irregular shape: square waves Baseline high frequency noise Baseline spiking See also: Baseline related problems with specific system components Back to top level troubleshooting list. Peak related problems Select the problem description that best matches your problem: Peaks broadening Peaks clipping at bottom Peaks clipping at top Double Peaks Peak Fronting (excessive slope on left side) Ghost Peaks Broad Ghost Peaks Irregular, Chair-shaped Peaks 11
12 Negative Peaks No Peaks after solvent peak No Peaks at all Sample Peak Tailing (excessive slope on right side) Solvent Peak Tailing (excessive slope on right side) Unresolved Peaks Discrete high-intensity contaminant peaks See also: Peak related problems with specific system components Back to top level troubleshooting list. Results related problems Select the problem description that best matches your problem: Low reproducibility of peaks area Poor sensitivity with increased retention time Poor sensitivity with normal retention time Retention times decreasing Retention times increasing Low reproducibility of retention times Retention times are inconsistent See also: Results related problems with specific system components Back to top level troubleshooting list. System-component related problems Flame Ionization Detector (FID) Troubleshooting To ensure optimum performance of the FID, you must keep it clean and free of dust and deposits. Symptoms such as reduced sensitivity and increased noise indicate that detector needs cleaning. To properly maintain the FID, you should perform the following cleaning or replacement sequences: cleaning the jet replacing the jet cleaning the collecting electrode replacing the collecting electrode replacing the ignition assembly These maintenance activities are described in Chapter 9 of the Maintenance and troubleshooting Manual. Select the problem description that best matches your problem: FID Sensitivity problems FID Flame Ignition problems FID Contamination problems Back to top level troubleshooting list. 12
13 Electron Capture Detector (ECD) Troubleshooting Select the problem description that best matches your problem: ECD Contamination problems ECD Sensitivity problems High base frequency Negative dips after peaks Baseline drift with changing pulse voltage Back to top level troubleshooting list. Nitrogen Phosphorous Detector (NPD) Troubleshooting To ensure optimum performance of the NPD, you must keep it clean and free of dust and deposits. Symptoms such as reduced sensitivity and increased noise indicate that detector cleaning or thermionic source replacement could be necessary. To properly maintain the Nitrogen-Phosphorus Detector, you should perform the following sequences: cleaning the jet. replacing the jet. cleaning the collecting electrode. replacing the collecting electrode. replacing the thermionic source. These maintenance activities are described in Chapter 11 of the Maintenance and Troubleshooting Manual. Select the problem description that best matches your problem: NPD Sensitivity problems No NPD detector response NPD detector response lower than expected High background level (NPD) NPD shows FID-like response for solvent and other carbon based compounds Solvent quenching effect (NPD) Unstable baseline (NPD) Low carbon rejection (NPD) Back to top level troubleshooting list. Flame Photometric Detector (FPD) Troubleshooting To ensure optimum performance of the FPD, you must keep it clean and free of dust and deposits. Symptoms such as reduced sensitivity and increased noise indicate that detector needs cleaning. To properly maintain the FPD, you should perform the following cleaning or replacement sequences: cleaning the jet cleaning the interferential filter cleaning the mirror metal plug 13
14 cleaning the flame-side heat shield replacing the jet replacing the interferential filter replacing the heat shields These maintenance activities are described in Chapter 12 of the Maintenance and Troubleshooting Manual. Select the problem description that best matches your problem: FPD Sensitivity problems High standing current and noise (FPD) No standing current (FPD) Unstable and excessively noisy baseline (FPD) FPD temperature does not reach the set point Low sensitivity (FPD) Low sensitivity and water droplets generate between the heat shields (FPD) Back to top level troubleshooting list. Photoionization Detector (PID) Troubleshooting The Photoionization Detector requires relatively little maintenance. To properly maintaining the detector, you should perform the following sequences: UV lamp replacement Cleaning the UV lamp window These maintenance activities are described in Chapter 13 of the Maintenance and Troubleshooting Manual. The UV Lamp must be replaced when exhausted, faulty or when a different class of compounds is to be analyzed. Click here for the available UV lamps and their application field. Note: Keep always a spare UV lamp in your laboratory because the lamp could collapse suddenly, without warning signs. CAUTION: Make sure to have appropriate filters installed on the carrier and make-up gas lines, and replace them regularly. Select the problem description that best matches your problem: PID Sensitivity problems PID Contamination problems UV lamp does not light immediately (PID) UV lamp does not light at all (PID) No standing current (PID) Tailing of solvent peak (PID) Tailing of sample peaks (PID) Low sensitivity (PID) Back to top level troubleshooting list. 14
15 Thermal Conductivity Detector (TCD) Troubleshooting The TCD detector does not usually need current maintenance. Nevertheless, if you follow a few simple sequences, you will avoid troubles and prolong the detector s lifetime. Pay special attention to avoid contamination or damage of the filaments. Follow these simple rules: Avoid activating/deactivating the bridge when not necessary. This operation would considerably reduce the filaments lifetime. Avoid injecting samples that contain halogenated or acid compounds at high concentrations. Ensure that oxygen (air) could not enter into the filaments cells. Oxidation would irreversibly damage the filaments. Install traps for moisture and oxygen on the gas lines to reduce the hazard. WARNING! Set Filament power to Off before disconnecting the column from the detector. When the column is disconnected, air will enter into the cell and the filament, if powered, will burn. For the same reason set Filament power to On only if the column has been connected. It is a good practice to let the reference and make-up gases flow through the cells for minutes before powering the filaments. Select the problem description that best matches your problem: TCD Sensitivity problems Negative Peaks (TCD) Baseline Fluctuation (TCD) Baseline Drift (TCD) Low Sensitivity (TCD) TCD does not work Back to top level troubleshooting list. Mass Spectrometer (MS) Troubleshooting This section covers only chromatography-related problems that may be observed when using a GC/MS system, where the cause may be with the MS rather than the GC. Refer to your MS troubleshooting documentation and help for specific help with MS problems. Select the problem description that best matches your problem: High Baseline (MS) Peak Tailing (MS) Low Sensitivity (MS) Poor Reproducibility of Results (MS) Back to top level troubleshooting list. PTV Injector troubleshooting The PTV injector will normally be serviced by ThermoFinnigan authorized technical personnel. In order to operate at peak performances, the injector requires periodic maintenance from the user. This maintenance includes: the replacement of the septum 15
16 the cleaning or replacement of the liner. These maintenance activities are described in Chapter 6 of the Maintenance and Troubleshooting Manual. In addition, you will find information on how to adjust the quartz wool packing inside the liner used in the Large Volume version of the PTV injector (LV PTV). Select the problem description that best matches your problem: PTV Sensitivity problems PTV Discrimination problems Discrimination of heavy compounds in splitless mode (PTV) Discrimination of volatile compounds in splitless mode (PTV) Discrimination of volatile compounds in solvent split mode (PTV) Discrimination in split mode (PTV) Excessive broadening of solvent peak (PTV) (also with volatile compounds obscured and lack of sensitivity) Sample degradation (PTV) PTV Large Volume - Poor reproducibility of results PTV Large Volume - Discrimination of volatile compounds PTV Large Volume - Sample degradation Back to top level troubleshooting list. Gas Sampling Valve (GSV) troubleshooting The Gas Sampling Valve is available in two versions: Manual and Automatic. Both types will normally be serviced by ThermoFinnigan authorized technical personnel. The Manual Gas Sampling Valve does not require periodic maintenance. To operate effectively, the Automatic Gas Sampling Valve requires periodic maintenance from the user. This maintenance includes: the replacement of the diaphragm; the cleaning or replacement of the removable frit of the filter. For either version, you could be required to replace the valve loop, if troubleshooting indicates that it needs to be cleaned or replaced. These maintenance activities are described in Chapter 7 of the Maintenance and Troubleshooting Manual. Incorrect operation of the gas sampling valves lowers the reproducibility of the injected volumes of sample causing unreproducible results. Select the problem description that best matches your problem: Analyses results are not reproducible (GSV) No signal at all (GSV) GSV does not work Back to top level troubleshooting list. 16
17 Leak Checking Checking for Leaks There are various automatic and manual operations you can use to keep under control the tightness of your chromatographic system. The sequences of operations you can use to find and correct possible leaks in the system are organized according to the instrument configuration. Specifically, this means the type of column installed, and whether the system is equipped with the Digital Pressure Flow Control (DPFC) and Digital Gas Flow Control systems. Click here for information on leak-checking DPFC-equipped systems. Click here for information on leak-checking non-dpfc systems. Select the relevant leak-checking procedure from the list below: Performing an Automatic Leak Check Performing an Automatic Column Evaluation Checking for leaks - Capillary Columns, DPFC, DGFC Checking for leaks - Capillary Columns, DPFC, Non-DGFC Checking for leaks - Capillary Columns, Non-DPFC, Non-DGFC Checking for leaks - Packed Columns, DPFC, DGFC Checking for leaks - Packed Columns, DPFC, Non-DGFC Checking for leaks - Packed Columns, Non-DPFC, Non-DGFC Leak testing the lines from the Gas Source to the GC Identifying and Removing Leaks Back to top level troubleshooting list. Leak-checking DPFC-equipped systems With DPFC (Digital Pressure Flow Control), the circular sequence Automatic Leak Check - Manual Check for Leaks - Column Evaluation - Automatic Leak Check is the key for minimizing troubles related to leaks. Automatic Leak Check When you perform an Automatic Leak Check, the GC measures the column flow with a true mass flow sensor and compares it to the calculated flow from the original column constant to see if the numbers match. The instrument assumes a gas leak exists if there is a change and notifies you that the system is not leak tight. When the Automatic Leak Check highlights leaks, you should perform a manual leak check, to determine the location of the leak. You should execute the Automatic Leak Check: once a day every time you have disconnected the gas lines for any reason (e.g. to clean or replace a component of the system or to install a new column). Note that only a previous Column Evaluation, performed in a condition of true tightness, can ensure the validity of the subsequent Automatic Leak Check responses. Column Evaluation 17
18 Once the leak has been removed and the tightness of the system is reasonably sure, you should perform the Column Evaluation automatic control and compare the response with the K Factor values reported in the K Factor Quick Reference card supplied with the Trace GC manuals. If the value obtained does not agree with the one reported on the card, the leaks have not been repaired. Performing the Column Evaluation is the necessary condition for the success of any subsequent Automatic Leak Check. Click here for detailed instructions on Performing an Automatic Column Evaluation. Leak-checking Select the relevant procedure for leak-checking: Performing an Automatic Leak Check Performing an Automatic Column Evaluation Checking for leaks - Capillary Columns, DPFC, DGFC Checking for leaks - Capillary Columns, DPFC, Non-DGFC Checking for leaks - Packed Columns, DPFC, DGFC Checking for leaks - Packed Columns, DPFC, Non-DGFC Leak testing the lines from the Gas Source to the GC Identifying and Removing Leaks Back to Leak-checking. Back to top level troubleshooting list. Leak-checking non-dpfc systems To keep under control the tightness of a Non-DPFC equipped gas chromatograph you should check the system for leaks regularly, at least once a week. In addition, you should also check for leaks every time you disconnect the gas lines for any reason (e.g. to clean or replace a component of the system or to install a new column). Select the relevant procedure for leak-checking: Checking for leaks - Capillary Columns, Non-DPFC, Non-DGFC Checking for leaks - Packed Columns, Non-DPFC, Non-DGFC Leak testing the lines from the Gas Source to the GC Identifying and Removing Leaks Back to Leak-checking. Back to top level troubleshooting list. Performing an Automatic Leak Check (DPFC systems only) 1. Press LEAK CHECK to access the Leak Check control table. 2. Select the carrier gas channel you want to check and press ENTER. The channel selected is automatically pressurized with carrier gas and sealed to perform the leak check. The following message will be displayed: Please wait Leak check in progress 3. If the system is free of leaks, the following message is displayed: 18
19 Leak check: passed 4. If leaks are found, an error message will be displayed. In this case, eliminate leaks and repeat the leak check procedure. Back to Leak-checking. Back to top level troubleshooting list. Performing an Automatic Column Evaluation (DPFC systems only) Note: Before you begin this procedure, make sure the GC is in stand-by condition at the initial temperature. 1. Press COLUMN EVAL to access the relevant control table. 2. Scroll to Right column or Left column (as required) and press ENTER. The applicable control table is displayed (parameters not editable). The GC automatically performs the column evaluation. 3. After a few minutes the following message will be displayed: L COL EVALUATION COMPLETED SUCCESSFULLY K = X.XX 4. Make a note of the displayed K Factor value. Back to Leak-checking. Back to top level troubleshooting list. Leak testing the lines from the Gas Source to the GC Single GC system 1. Switch off the gas chromatograph power supply. 2. Turn on all gas supply lines to the GC, setting the input pressure to 420 kpa (60 psi). Allow the gas lines to pressurize for several seconds. 3. Turn off the input gas source. The pressure value should not change more than 5% in 10 minutes. If the value drops down immediately, one or more leaks are present. In this case, carry on with step Check the connections with a handheld electronic leak detector to find the leaks. Multiple GC systems If more than one system is attached to the same gas source, you will need to isolate each GC system. 1. Install a cut-off valve and pressure gauge with an 1/8 inch input tee in the gas line leading to the GC that you are leak testing, as shown in the figure. Be sure the pressure gauge is connected in line between the on/off valve and the GC. 2. Turn on the gas supply and allow the system to pressurize. 3. Turn off the input gas source. The pressure value should not change more than 5% in 10 minutes. If the value drops down immediately, one or more leaks are present in the plumbing from the cut-off valve to the GC. In this case, carry on with step Check the connections of the line you are testing with a handheld electronic leak detector to find the leaks. Back to Leak-checking. 19
20 Back to top level troubleshooting list. Identifying and Removing Leaks If the Automatic Leak Check (or any significant symptom) has notified a possible leak in the system, you should: Check the whole system for leaks, from the carrier gas inlet to the detector base body. Check the accessible, critical connections (column to injector, column to detector, split and purge valves, septum caps). A possible source of leaks may be the gas bottles/gas chromatograph connections. Check these lines before the others, if specific symptoms indicate that the leak is outside the gas chromatograph. Back to Leak-checking. Back to top level troubleshooting list. Baseline related problems with specific system components Note: Use the Back button to return to this list after checking one of the possible causes below. Baseline drift with changing pulse voltage (ECD) Unstable baseline (NPD) High standing current and noise (FPD) No standing current (FPD) Unstable and excessively noisy baseline (FPD) No standing current (PID) Baseline Fluctuation (TCD) Baseline Drift (TCD) High Baseline (MS) Back to main Baseline problem list. Baseline drifting Accumulation of stationary phase. Remove the end section of the column. Carrier gas cylinder pressure too low to allow control. Replace the carrier gas cylinder, or increase the pressure. Drifting carrier gas or combustion gas flows. Check the gas controllers. 20
21 Accumulation of impurities in the column. Check impurity levels in the gas source. Use correct gas purity. Back to main Baseline problem list. Baseline falling Carrier gas leak in the system. Perform a leak test and check the tightness of the connections on the carrier gas line. Column is baking out. Allow enough time for the column to stabilize. Back to main Baseline problem list. Baseline falling away slowly after a high initial value Purge valve left closed during acquisition. Alter the GC program. See your GC user manual for details. Inadequate purge flow rate. Increase the purge flow rate. Purge valve left closed for too long. Shorten the purge time. Solvent tail peak. Increase the solvent delay or shorten the purge time. Pre-filters are dirty. (When using a quadropole MS detector) Contact your service representative. 21
22 Back to main Baseline problem list. Baseline rising Accumulation of impurities in the column. Check impurity levels in the gas source. Use correct gas purity. Contaminated detector. Check the detector and clean it. There is bleeding from the GC column. Condition or change the column. Air is leaking into the system. See Note 1. Trace and repair the leak. Note 1 With an air leak, the Baseline can also fall after an initial rise, as the GC ferrule expands with oven temperature and seals off the air. Back to main Baseline problem list. Baseline rising under temperature program control Column contaminated. Recondition the column. Back to main Baseline problem list. Baseline - high standing current Carrier gas flow rate too high. Reduce the carrier gas flow. Column contaminated. Recondition the column. 22
23 Contaminated gases. Replace gas cylinders or gas filters. Excessive column stationary phase bleeding. Check the oven temperature, ensuring that it doesn t exceed the column upper limit. Recondition the column. Replace the column. Loose connections. Ensure that all interconnections and screw connections are tight. Back to main Baseline problem list. Baseline irregular shape: dip after solvent peak Detector contaminated. Bake out or clean the detector. Back to main Baseline problem list. Baseline irregular shape: S-shaped Excessive column bleed during column temperature programming. Reduce the upper column temperature. Bake out the column. Install a high temperature column. Oxygen contamination is decomposing the stationary phase. Install oxygen filters in the carrier gas line. Check the pneumatic and inlet systems for leaks. Use correct gas purity with low oxygen content. Back to main Baseline problem list. 23
24 Baseline irregular shape: square waves Large AC power fluctuations; heavy equipment on the same line. Use a dedicated clean AC line of sufficient amperage. Back to main Baseline problem list. Baseline high frequency noise Contaminated detector. Isolate the detector from the electronics. If noise disappears, clean the collector. Combustion gas flow too low or too high. Check the detector gas flows. Column contaminated. Condition the column. Contaminated detector gas supply. Check the gas purity and install appropriate filters. Defective electrometer. Replace electrometer. Detector temperature higher than column maximum temperature. Reduce the detector temperature to the column temperature upper limit. External electrical interference. Attach an AC line monitor and check the AC supply for interference. Loose column fittings. Tighten fittings accordingly. 24
25 Loose detector electrical connections. Make sure the leads are properly connected. Back to main Baseline problem list. Baseline spiking Defective electrometer or amplifier. Replace the electrometer or amplifier. Column too close to flame. (When using an FID) Lower the column to the correct position (2-3 mm below the tip of the jet). Dirty jet or detector. Isolate the detector from the electronics. If the spiking disappears, clean the jet and the collector. FID temperature too low. (When using an FID) Increase the FID temperature to at least 150 C. Back to main Baseline problem list. Peak related problems with specific system components Note: Use the Back button to return to this list after checking one of the possible causes below. Negative dips after peaks (ECD) Tailing of solvent peak (PID) Tailing of sample peaks (PID) Negative Peaks (TCD) Peak Tailing (MS) Excessive broadening of solvent peak (PTV) Back to main Peaks related problem list. 25
26 Peaks broadening Column flow too high. Reduce the flow to slightly above optimum. Column flow too low. Increase the flow to slightly above optimum. Split flow too low in split injection. Increase the flow to ml/min. Column performances degraded. Test the column at the optimum flow rate. Dirty injector. Clean or replace the liner. Stationary phase accumulated in the outlet. Remove the last two coils from the column. Detector base body temperature too low. Increase the temperature to 5 C below the column maximum. The sample is overloading the column. Reduce the amount and/or concentration of the sample. Back to main Peaks related problem list. Peaks clipping at bottom Detector or integrator zero set too low. Set the zero correctly. Back to main Peaks related problem list. 26
27 Peaks clipping at top Data system zoomed in too close. Zoom out to view the entire chromatogram. Detector or integrator attenuation set too low. Set the attenuation higher. Detector range too sensitive. Set a less-sensitive detector range. Incorrect input to recording unit. Correct and check the recording unit. Back to main Peaks related problem list. Double Peaks Injection speed too low. Inject more rapidly in a smooth motion. Wrong autosampler injection speed or mode. Use a higher speed. Back to main Peaks related problem list. Peak Fronting (excessive slope on left side) Column or detector overloaded. Decrease the injected amount and/or analyte concentrations. Increase the split ratio. Column temperature too low. Increase the temperature. Stationary phase too Use a thicker-film 27
28 thin. column. Poor injection technique. Repeat, with better injection technique. Back to main Peaks related problem list. Ghost Peaks Contaminated carrier gas. Replace the cylinder or replace the filter. Contamination from laboratory glassware. Ensure the glassware is clean and contamination-free. Decomposition of injected sample. Decrease the injection port temperature. Use the on-column injection technique. Dirty injection solution. Carry out adequate clean-up of sample prior to injection. Back to main Peaks related problem list. Broad Ghost Peaks Contaminated inlet or pneumatics. Remove the column and bake out the inlet. Use a high-quality septum. Replace the split vent filter. Install an in-line filter between the pneumatics and the inlet. Incomplete elution of previous sample. Increase the final oven program temperature or total run time. Increase the column 28
29 flow rate. Back to main Peaks related problem list. Irregular, Chair-shaped Peaks Solvent flooding of column. Increase the initial oven temperature. Reduce the injection volume (OC). Install a retention gap (OC). Back to main Peaks related problem list. Negative Peaks All Peaks negative Integrator wires reversed. Some Peaks negative This symptom can be normal. Correct the connections. None required. Back to main Peaks related problem list. No Peaks after solvent peak Carrier gas flow too high. Reduce the carrier gas flow rate. Combustion gas flow incorrect. Check the combustion gas flow. Detector contaminated. Bake out or clean the detector. FID flame Check the detector 29
30 extinguished by solvent peak. temperature. Too much sample injected. Inject less sample. Incorrect column position in S/SL injector (too high). Check the column position. Back to main Peaks related problem list. No Peaks at all Clogged syringe needle. Replace or repair the syringe. Column broken or disconnected. Check the column and connections. Defective electrometer or amplifier. Replace the electrometer or amplifier. Defective recording device. Replace the recording device. FID flame is out. Light the flame. Poor or missing electrical connection. Check the cable connections. Incorrect column position in S/SL injector (too high). Check the column position. Back to main Peaks related problem list. 30
31 Sample Peak Tailing (excessive slope on right side) Column degradation causing activity. Inject a test mixture and evaluate the column. Column/oven temperature too low. Increase the column/oven temperature. Do not exceed the recommended maximum temperature for the stationary phase. Dirty liner. Clean or replace the liner. Glass wool or inlet liner causing activity. Replace with fresh silanized wool and a clean inlet liner. Inlet temperature too low. Increase the inlet temperature. Poor or obstructed column connections. Remake the column inlet connection. Wrong stationary phase. Replace the column according to the column manufacturer s literature. Back to main Peaks related problem list. Solvent Peak Tailing (excessive slope on right side) Incorrect column position in inlet. Reinstall the column. 31
32 Initial oven temperature too high (On Column). Reduce the initial oven temperature. Septum purge flow too low and/or split/splitless vent flow too low. Check and adjust the septum purge and vent flows. Too large injection size. Reduce the injection size. Back to main Peaks related problem list. Unresolved Peaks Carrier gas flow rate too high. Reduce the carrier gas flow rate. Column deteriorated. Replace the column. Column temperature too high. Lower the column oven temperature. Column too short. Use a longer column. Incorrect column choice. Install a suitable column. Injection technique is not adequate. Choose a correct injection technique. Back to main Peaks related problem list. Discrete high-intensity contaminant peaks Bleed from the GC column. Condition or change the column. Bleed from the septum. Replace the septum. 32
33 Sample vial septa are contaminating the sample. Discard sample. Store samples upright, in a refrigerator. Use teflon faced septa, with the teflon facing downwards (i.e. towards the sample). Back to main Peaks related problem list. Results related problems with specific system components Note: Use the Back button to return to this list after checking one of the possible causes below. FID Sensitivity problems ECD Sensitivity problems NPD Sensitivity problems FPD Sensitivity problems Low sensitivity (FPD) Low sensitivity and water droplets generate between the heat shields (FPD) PID Sensitivity problems Low sensitivity (PID) TCD Sensitivity problems Low Sensitivity (TCD) Low Sensitivity (MS) Poor Reproducibility of Results (MS) PTV Sensitivity problems PTV Large Volume - Poor reproducibility of results Analyses results are not reproducible (GSV) Back to main Results related problem list. Low reproducibility of peaks area Concentration not compatible with the dynamic range of the detection system. Ensure that the sample concentration is suitable for the detection system. Inappropriate injection technique. Try a different injection technique. 33
34 Injection parameters inappropriate. Check the injection temperature and flow rates. Non reproducible sample injection technique. Evaluate the sample preparation sequences. Compare the results with a series of standard injections. Leaking syringe or septum. Check and replace the syringe and/or septum at regular intervals. Leaks at the injection. Check the column connections. Run a leak check. Poor injection technique. Carefully meter the injected amount. Use a clean, good-quality syringe. Poor split flow or ratio control. Monitor the flow. Replace the in-line filter. Back to main Results related problem list. Poor sensitivity with increased retention time Carrier gas flow rate too low. Increase the carrier gas flow rate. Locate and remove possible obstructions in the carrier gas line. Check the injector/column ferrules. Back to main Results related problem list. Poor sensitivity with normal retention time 34
35 Oven or injector parameters are not optimised. Adjust the oven or injector parameters. Leaks in the GC carrier gas line. Run a leak test and correct leaks. Syringe leaks during injection. Replace syringe or piston seals, if applicable. Split injection temperature too low. Increase the temperature of the injector. Column is in poor condition, or wrong column type used. Condition or change the column. Back to main Results related problem list. Retention times decreasing Stationary phase deteriorated by oxygen and/or water. Use a carrier gas free of oxygen and water. Stationary phase loss due to column bleeding. Reduce the column temperature. Back to main Results related problem list. Retention times increasing Increasing carrier leakage. Check the septum and column connections. Carrier gas supply running out. Replace the bottle. 35
36 Back to main Results related problem list. Low reproducibility of retention times Drifting or unstable Monitor the column pneumatic controller. pressure or flow. Check and replace the controller if necessary. Poor injection technique. Start the run at consistent time after injection. Sample size is too large. Reduce the injected amount and/or volume. Unstable column temperature. Check the main oven door and cooling flap. Monitor the column temperature. Back to main Results related problem list. Retention times are inconsistent GC column is in condition. poor Condition or change the column. Insufficient equilibration time set on GC. Increase equilibration time. Poor injection. Repeat with better injection technique. Oven temperature programmed to rise too quickly. Reduce oven temperature ramp rate. Air is leaking into the system at the injector seal or the carrier gas manifold. Trace and repair the leak. 36
37 Back to main Results related problem list. FID Sensitivity problems The sensitivity of the Flame Ionization Detector depends on the carrier and detector gases flow rates, and on the detector temperature. The gas flow rates must be set correctly for proper FID operation. Click here for the gases normally used with FID. The carrier gas flow range depends on the type of gas used and on the type and diameter of the capillary or packed column installed. Note: Make-up gas is not required when a packed column is used. Click here for the recommended range of detector gas flow rates tolerated by the FID. To gain optimum analytical performance from the FID, the hydrogen flow rate should be experimentally optimized, keeping the carrier and air flows constant, to obtain the maximum signal intensity for the components of interest. The FID sensitivity will be reduced as hydrogen flow rate is above or below the optimal value. The flow rate of the air is less critical than the hydrogen one. An excessive amount of air will destabilize the flame, causing noise and eventual flameout. Note: Generally the air flow rate must be set to about 10 times the hydrogen flow rate to keep the flame lit. A low flow rate of air would reduce the detector sensitivity. Click here for typical FID operating conditions. For high sensitivity applications, it is essential that you exclude all traces of organic contamination from the chromatographic system and/or from the gas lines of the detector. Such contamination would cause ghost peaks in the chromatogram or, more commonly, an unstable baseline. WARNING! It is the customer s responsibility to ensure compliance with all local safety regulations concerning gas supplies. Hydrogen is a potentially dangerous gas. Mixed with air it may give rise to an explosive mixture. The use of hydrogen requires the operator s extreme caution due to the risks involved. For further details concerning hydrogen, please refer to Using Hydrogen on page xi of the Maintenance and Troubleshooting Manual. Back to FID troubleshooting problem list. FID gases Capillary Columns Carrier gas Detector Gas Make-up Gas Packed Columns Helium Helium Hydrogen + Nitrogen Nitrogen Nitrogen Air (recommended) Hydrogen Argon Helium FID detector gas flow rates Gas Flow rate Hydroge n ml/min 37
38 FID operating conditions Air ml/min Make-up gas ml/min Gas Capillary Packed Columns Columns Carrier H ydrogen Air 2 ml/min 35 ml/min 350 ml/min 40 ml/min 40 ml/min 500 ml/min Make-up gas 30 ml/min Not used (Nitrogen) FID Flame Ignition problems You can ignite the flame as soon as the detector temperature has reached 150 C. After the flame appears to have ignited, check for water vapor condensed on a mirror or on the polished end of a wrench directly over the FID exit. You should observe steam condensing on the cold surface. If not, the flame is not ignited. WARNING! Do not lean over the FID to see the flame, it is invisible. The table below shows the possible causes of flame ignition problems and the relevant remedies. Incorrect fuel gas flows. Make sure that all gas flows are correct. Defective igniter. Check the igniter element. Refer to the Replace the Ignition Assembly Operating Sequence in Chapter 9 of the Maintenance and Troubleshooting Manual for instructions. Broken or cracked flame jet. Replace jet. Refer to the Replace the Jet Operating Sequence in Chapter 9 of the Maintenance and Troubleshooting Manual for instructions. Blocked jet tip. Check for a blocked jet by measuring the hydrogen flow with a 38
39 flowmeter. Refer to the Measure the Gas Flows with a Bubble Meter Operating Sequence in Chapter 8 of the Maintenance and Troubleshooting Manual for instructions. If required, remove and clean the jet following the Clean the Jet Operating Sequence in Chapter 9 of the manual. Faulty electronics. Contact your service representative. Back to FID troubleshooting problem list. FID Contamination problems All organic compounds that pass through the FID are burnt in the flame. Some compounds can produce detector contamination, resulting in a high noise baseline. If this symptom is found, clean the collecting electrode. Refer to the Clean the Collecting Electrode Operating Sequence in Chapter 9 of the Maintenance and Troubleshooting Manual for instructions. Back to FID troubleshooting problem list. ECD Contamination problems The ECD, if properly used, has a good resistance against contamination thanks to its unique internal geometry. However, some critical operating conditions may give rise, in the long run, to contamination of the collecting electrode. This contamination is highlighted by an excessive increase of the base frequency and a baseline drift when the reference current or the pulse amplitude is changed. The collecting electrode, in these cases, can be easily removed and cleaned without disturbing the radioactive source. Refer to the Clean the Collecting Electrode Operating Sequence in Chapter 10 of the Maintenance and Troubleshooting Manual for instructions. If contamination of the whole cell is suspected, heat the ECD at the maximum operating temperature with carrier and make-up gases flowing through the detector (thermal cleaning). The decontamination process can be followed by monitoring the base frequency. Initially the frequency value tends to increase to very high values, and then it decreases to acceptable ones. If irreversible contamination of the cell is suspected, please contact your local ThermoFinnigan Technical Service office for maintenance or replacement of the source, whichever is necessary. Strictly follow the existing regulations for the transport of radioactive material. Detector contamination could be indicated by the following effects in the chromatogram: reduced signal to noise ratio high-noise baseline 39
40 baseline drift with changing pulse voltage negative dips after peaks Back to ECD troubleshooting problem list. Warning: Radioactivity WARNING! The Electron Capture Detector (ECD) contains a 63Ni beta-emitting radioactive source of 370 MBq (10 mci). The 63Ni radioisotope, electrically deposited as metal on a nickel foil, is inserted in a cylindrical holder made of 6 mm thickness stainless steel. The holder is fixed to the detector body, also made of stainless steel, in order to be totally inaccessible from the outside. The radioisotope is not released by its support at temperatures lower than 450 C. This temperature can never be reached by the detector, whose maximum operating temperature is 400 C. A safety device prevents overheating. In normal operating conditions, no dispersion of solid or gaseous radioactive material is involved. This eliminates any risk of direct or secondary radiation. For no reason should the detector be opened or handled by the user. Any maintenance or service operations involving even partial disassembling of the detector must be performed ONLY by CE Instruments factories (Milano and Austin) or by a laboratory specifically licensed to handle radioactive material. ECD Sensitivity problems The effect of a temperature change on the ECD response generally depends on the compounds analyzed. The ECD behavior can be affected by a change of the total flow of the gases through the cell. The resulting response is also related to the nature and purity of the gases. Click here for the gases normally used on the ECD. When using helium or hydrogen as a carrier gas with capillary or wide bore columns, the detector should be fed with nitrogen or argon/methane through the make-up gas line. The carrier gas flow rate must be set according to the type of the gas used and to the diameter and the stationary phase of the capillary (or packed) column installed. Note: When packed columns are used, the make-up gas (as carrier gas type) is used only if strictly required. WARNING! It is the customer s responsibility to ensure compliance with all local safety regulations concerning gas supplies. Hydrogen is a potentially dangerous gas. Mixed with air it may give rise to an explosive mixture. The use of hydrogen requires the operator s extreme caution due to the risks involved. For further details concerning hydrogen, please refer to Using Hydrogen on page xi of the Maintenance and Troubleshooting Manual. Back to ECD troubleshooting problem list. ECD gases Capillary Columns Helium Carrier Gas Packed Columns Nitrogen Make-up Gas Nitrogen (recommended) Nitrogen Argon-Methane (5- Argon-Methane (5-10%) 10%) 40
41 Hydrogen High base frequency (ECD) Impure gas supply. Use high purity grade gases and filters to trap moisture and oxygen. No (or insufficient) flow of make-up gas. Increase make-up gas flow rate. Excess column bleeding. Condition the column. Leaks on carrier and/or make-up gas lines. Run a leak test. Contaminated collecting electrode. Clean the collecting electrode, as described in the Clean the Collecting Electrode Operating Sequence in Chapter 10 of the Maintenance and Troubleshooting Manual, or replace it. Chemically contaminated radioactive source. Refer to ECD Contamination. Pulse width not Set Pulse width to 1µs correctly set. for Nitrogen and 0.1 µs for Ar/CH4. Back to ECD troubleshooting problem list. Negative dips after peaks (ECD) Contaminated collecting electrode. Clean the collecting electrode, as described in the Clean the 41
42 Collecting Electrode Operating Sequence in Chapter 10 of the Maintenance and Troubleshooting Manual, or replace it. Contaminated radioactive source. Refer to ECD Contamination. Back to ECD troubleshooting problem list. Baseline drift with changing pulse voltage (ECD) Contaminated collecting electrode. Clean the collecting electrode, as described in the Clean the Collecting Electrode Operating Sequence in Chapter 10 of the Maintenance and Troubleshooting Manual. If the problem persists, replace the collecting electrode. Back to ECD troubleshooting problem list. NPD Sensitivity problems The main cause of changes in sensitivity is related to the Thermionic source. Its gradual depletion results in a drop of sensitivity, that can be compensated by increasing the source current. Take into account that increasing current results in a shorter source life. Loss of sensitivity can also be related to source contamination, due to high boiling sample compounds not completely eliminated. Shift of detector temperature can reduce sensitivity. Click here for the gases normally used for NPD. Consistent flows are necessary to maintain a constant and stable sensitivity. Nitrogen is preferred over helium as make-up gas due to its lower thermal conductivity. Using nitrogen, the source requires a lower heating current. Note: When a packed column is used, make-up gas generally is not necessary. WARNING! It is the customer s responsibility to ensure compliance with all local safety regulations concerning gas supplies. Hydrogen is a potentially dangerous gas. Mixed with air it may give rise to an explosive mixture. The use of hydrogen requires the operator s extreme caution due to the risks involved. For further details 42
43 concerning hydrogen, please refer to Using Hydrogen on page xi of the Maintenance and Troubleshooting Manual. Click here for typical flow rates for detector gases. Back to NPD troubleshooting problem list. NPD gases Capillary Columns Helium Hydrogen (only with Carrier gas Detector Gas Make-up Gas Packed Co lumns Helium Hydrogen + Nitrogen Nitrog en Nitrogen Air Helium DPFC) NPD operating conditions Gas Hydrogen Air Make- up gas Flow rate 2-4 ml/min ml/min ml/min No NPD detector response Source heating Increase the heating current too low. current. No hydrogen flow. Turn the hydrogen flow on and set it to a correct value. No air flow. Turn the air flow on and set it to a correct value. Source turned off. Turn the source on. Source faulty. Replace the source. Back to NPD troubleshooting problem list. NPD detector response lower than expected 43
44 Lower source temperature. Check the base body temperature and source heating current. Air contamination in hydrogen line. Turn the source off. Increase the hydrogen pressure for minutes to purge the hydrogen line. Check the line tightness. Back to NPD troubleshooting problem list. High background level (NPD) Heating current too high. Set the correct operating parameter. Hydrogen flow too high. Set the correct operating parameter. Air and/or Make-up gas flow too low. Set the correct operating parameter. Excessive column bleed. Condition the column. Back to NPD troubleshooting problem list. NPD shows FID-like response for solvent and other carbon based compounds Hydrogen flow too high. Reduce hydrogen flow to a lower value. Back to NPD troubleshooting problem list. 44
45 Solvent quenching effect (NPD) The "Solvent quenching effect" is observed when there is a large negative baseline upset at solvent elution with no recovery to the original baseline. Heating current too low. Slightly increase the source heating current. Back to NPD troubleshooting problem list. Unstable baseline (NPD) Background current level too high. Reduce the source heating current value. Hydrogen flow fluctuation. Check pressure regulators on hydrogen line. Back to NPD troubleshooting problem list. Low carbon rejection (NPD) Hydrogen flow too high. Reduce hydrogen flow to the correct operative conditions. Back to NPD troubleshooting problem list. FPD Sensitivity problems The FPD sensitivity is affected by the temperature. The FPD response decreases with an increase in the detector temperature. In addition, the detector response can be severely reduced if a non-sulphur or nonphosphorus compound is partially or fully eluted with the analyte of interest (quenching). Keep clean the interferential filter and the heat shields as described in the maintenance sequences, to avoid reducing sensitivity. Replace the heat shields if the deposits due to contamination cannot be eliminated. A right choice of the flow rates of hydrogen and air (fuel gases) is of primary importance in determining FPD sensitivity. Flow rates will affect also selectivity and, to a great extent, peak shapes. 45
46 Optimum carrier gas flow rate depends on the type of gas used and on the diameter of the column installed. Click here for the gases normally used with FPD. Click here for the recommended values of flow rate for FPD. The optimum air flow rate should be found experimentally, setting a proper value for the hydrogen and performing some analyses on a standard mixture at different air flow rates. WARNING! It is the customer s responsibility to ensure compliance with all local safety regulations concerning gas supplies. Hydrogen is a potentially dangerous gas. Mixed with air it may give rise to an explosive mixture. The use of hydrogen requires the operator s extreme caution due to the risks involved. For further details concerning hydrogen, please refer to Using Hydrogen on page xi of the Maintenance and Troubleshooting Manual. Back to FPD troubleshooting problem list. FPD gases Capillary Columns Carrier gas Fuel Gas Make-up Gas Packed Columns Helium Helium H ydrogen + Nitrogen Nitrogen Air not required H ydrogen Argon FPD gas flow rates Gas Capillary Columns Packed Columns C arrier 1-3 ml/min ml/min Hydrogen ml/min ml/min Air ml/min ml/min Generally The optimum air flow rate should be found experimentally, setting a pr oper value for the hydrogen and performing some analyses on a standard mixture at different air flow rates. High standing current and noise (FPD) Detector conditioning Increase detector not completed. conditioning time. Air flow rate set too high. Reduce air flow rate. Detector not light- tight. Tighten the chimney cap and the knurled nut that fixes the photomultiplier tube. lnterferential filter Check the 46
47 missing. interferential filter has been correctly installed. Back to FPD troubleshooting problem list. No standing current (FPD) Signal or power cables incorrectly connected. Check connections. Faulty photomultiplier tube. Contact your service representative. Back to FPD troubleshooting problem list. Unstable and excessively noisy baseline (FPD) Faulty Contact your service photomultiplier tube. representative. Back to FPD troubleshooting problem list. FPD temperature does not reach the set point Power cable not Check cable properly connected. connection. Faulty heater. Contact your service representative. Back to FPD troubleshooting problem list. Low sensitivity (FPD) 47
48 Low photomultiplier excitation voltage. Set High Voltage Mode to Yes in the FPD control table. Inadequate hydrogen flow rate. Set the hydrogen flow rate to a proper value. Refer to Recommended gas flow rates for FPD. Inadequate air flow rate. Set the air flow rate to a proper value. Refer to Recommended gas flow rates for FPD and relevant note. Dirty interferential filter. Clean the interferential filter. Refer to Clean the Interferential Filter Operating Sequence in Chapter 12 of the Maintenance and Troubleshooting Manual. The mirror of the combustion chamber is dirty. Clean the mirror. Refer to Clean the Mirror Metal Plug Operating Sequence in Chapter 12 of the Maintenance and Troubleshooting Manual. The flame-side heat shield is dirty. Clean the flame-side heat shield. Refer to Clean the Flame-side Heat Shield Operating Sequence in Chapter 12 of the Maintenance and Troubleshooting Manual. Reduced transparency of the heat shields. Replace the heat shields. Refer to Replace the Heat Shields Operating Sequence in Chapter 12 of the Maintenance and Troubleshooting Manual. 48
49 Back to FPD troubleshooting problem list. Low sensitivity and water droplets generate between the heat shields (FPD) Leak at the flameside heat shield. Replace the heat shields. Refer to Replace the Heat Shields Operating Sequence in Chapter 12 of the Maintenance and Troubleshooting Manual. If the symptom does not disappear, contact your service representative. Back to FPD troubleshooting problem list. UV lamps for PID Lamp Type Application 8.4 ev Determination of amines and polycyclic aromatic compounds. 9.6 ev Determination of lowboiling aromatic compounds (BTEX analyses) ev General applications. Filled with krypton, emits at 10.0 and ev ev Determination of aldehydes and ketones. 49
50 PID Sensitivity problems The most common cause of loss of sensitivity is condensation of the effluent on the lamp window, resulting in a progressive increase of absorption of the UV. For long life, the window must be cleaned periodically according to the operating sequence described in the Clean the Lamp Window Operating Sequence in Chapter 13 of the Maintenance and Troubleshooting Manual. A Lamp failure message will be displayed when the lamp does not work. One of the possible causes is the breakdown of the lamp. In this case, replace the lamp following the instructions given in the in the Replace the UV Lamp Operation Sequence in Chapter 13 of the Maintenance and Troubleshooting Manual. See also: Low sensitivity (PID) Back to PID troubleshooting problem list. PID Contamination problems An abnormally high baseline may indicate contamination of the detector. In this case, turning off the lamp will not make the symptom disappear. This means that the detector cell has been contaminated. Try to solve the problem by baking the cell for two hours at 400 C. If the symptom does not disappear, contact your service representative. If you suspect contaminants are carried into the detector cell by the gas lines, try reducing the flow rate in each line, one at a time, to identify the contamination source. To avoid detector contamination install gas filters between the them periodically. gas cylinders and the detector and replace Note: To face more effectively possible chromatographic problems, you should carry out an analysis using the test mixture and evaluate the results. Back to PID troubleshooting problem list. UV lamp does not light immediately (PID) The lamp has not been used for a long time. Set High current mode? to Y until the lamp lights, then reset to N. Back to PID troubleshooting problem list. UV lamp does not light at all (PID) Message: Lamp failure Lamp not installed. Install a suitable UV lamp. Refer to the Replace the UV Lamp Operating 50
51 Sequence in Chapter 13 of the Maintenance and Troubleshooting Manual. Lamp not correctly installed in the lamp holder. Reinstall the UV lamp. Refer to the Replace the UV Lamp Operating Sequence in Chapter 13 of the Maintenance and Troubleshooting Manual. Lamp holder not correctly fixed to the lamp housing. Tighten the two knurled screws that fix the lamp holder to the housing. Lamp power cable disconnected. Check the connection. Polarization cable disconnected. Check the connection. Safety interlock circuit is faulty. Contact your service representative. UV lamp is exhausted. Replace UV lamp. Refer to the Replace the UV Lamp Operating Sequence in Chapter 13 of the Maintenance and Troubleshooting Manual. Back to PID troubleshooting problem list. No standing current (PID) Power and signal cables not properly connected. Check the connections of the cables. 51
52 Back to PID troubleshooting problem list. Tailing of solvent peak (PID) Insufficient make-up Increase the flow rate. gas flow rate. Back to PID troubleshooting problem list. Tailing of sample peaks (PID) Adsorption effects caused by degradation products inside the cell. Remove deposits by baking the cell for two hours at 400 C. If the symptom does not disappear, contact your service representative. Contamination of the cell. Remove contaminants by baking the cell for two hours at 400 C. If the symptom does not disappear, contact your service representative. Inadequate gas flow rate settings. Set gas flow rates to proper values. Leaks on the gas lines. Perform a leak test. Back to PID troubleshooting problem list. Low sensitivity (PID) Lamp window fogged by UVabsorbing deposits. Clean the lamp window. Refer to the Clean the Lamp Window Operating Sequence in Chapter 13 of the Maintenance and Troubleshooting 52
53 Manual. Flow rate of the make-up gas is set too high. Reduce the flow rate of the make-up gas down to10 ml/min or less. UV lamp is not adequate to the compounds to be analyzed. Replace with an appropriate one. Back to PID troubleshooting problem list. TCD Sensitivity problems Sensitivity is related to the detector temperature (increasing temperature reduces sensitivity) and to the flow rate of the carrier, reference and make-up gas. A gain in sensitivity can be obtained by increasing the gap between the temperature of the block and that of the filaments, or increasing the filaments voltage. Sensitivity is strictly related to the state of the filaments and to their operating conditions. A significant reduction of the detector sensitivity may be caused by the contamination of the filaments due to degradation of high molecular weight compounds inside the cell or to contaminated gases. Low temperatures of the detector block may cause high boiling compounds to condensate on the reducing sensitivity. See also Low Sensitivity (TCD). Back to TCD troubleshooting problem list. Negative Peaks (TCD) filaments, Negative peaks are normally generated by the sample components that have a thermal conductivity higher than carrier gas. For instance, using nitrogen or argon as carrier gas, negative peaks are obtained for helium, hydrogen or methane. To revert the polarity of the detector, refer to the Trace GC Operating Manual for instructions. Back to TCD troubleshooting problem list. Baseline Fluctuation (TCD) Unstable regulation of the flow rate of the gases. Check the controllers of the carrier, reference and make-up gases work well. Leaks on the gas lines. Leak test the connections of the carrier, reference and 53
54 make-up gas. Inlet pressure of gases set too low. Set the pressure of the carrier, reference and make-up gas to a proper value. Refer to the Trace GC Operating Manual for instructions. Trans temp set too high. Set Trans temp to a proper value (about 10 to 20 C below the Block temp value). Faulty temperature regulation. Contact your service representative. Back to TCD troubleshooting problem list. Baseline Drift (TCD) A small baseline drift normally occurs during a temperature program and does not indicate a problem. This effect is due to the decrease of the carrier gas flow rate as temperature increases. A baseline that suddenly goes out of scale and a rapid growth of the signal could indicate that filaments are likely to be burnt and have to be replaced. Unstable regulation of the flow rate of the gases. Check the controllers of the carrier, reference and make-up gases work well. Leaks on the gas lines. Leak test the connections of the carrier, reference and make-up gas. Column conditioning not correctly performed. Recondition the column according to the manufacturer s instructions. Septum bleeding. Check the operating temperature of the septum is adequate. Replace the septum if necessary. 54
55 Faulty temperature control. Contact your service representative. Back to TCD troubleshooting problem list. Low Sensitivity (TCD) Leaks on the gas lines. Leak test the connections of the carrier, reference and make-up gas. Leak due to septum wearing. Replace the septum. Operating conditions of the detector not properly set. Optimize working parameters of the detector according to the actual operating mode (Constant Voltage, Constant Temperature). Refer to the Trace GC Operating Manual for instructions. Thermal conductivity of the carrier gas is too close to that of the compound to be analyzed. Use a different carrier gas. Refer to the Trace GC Operating Manual for instructions. Contaminated filaments. Remove contamination by baking the filaments for one hour at a temperature higher than the boiling point of the most highboiling compound. Perform baking twice if necessary. If the symptom does not disappear, contact your service representative. 55
56 Oxidized filaments. Contact your service representative. Back to TCD troubleshooting problem list. TCD does not work Message: Filament power Off Lack of carrier, make-up or reference gas or pressure too low. Check the feed of the carrier, make-up and reference gas. Filament burnt. Contact your service representative. Back to TCD troubleshooting problem list. High Baseline (MS) Contaminated MS source. Clean the Source. Air is leaking into the GC/MS interface. Trace and repair the leak. Back to MS troubleshooting problem list. Peak Tailing (MS) MS Source or GC Interface is not hot enough. Increase the Source or GC Interface temperature. Back to MS troubleshooting problem list. 56
57 Low Sensitivity (MS) MS ion source is dirty. Clean the source. MS ion source temperature is not optimized. Re-tune the MS or adjust the source temperature. MS is not tuned properly. Tune the MS. Incorrect column alignment in the MS source. Re-align the column in the MS source. Detector voltage set too low (Trace MS) Increase the Detector Voltage in the Trace MS Tune window. Back to MS troubleshooting problem list. Poor Reproducibility of Results (MS) MS ion source is dirty or badly assembled. Disassemble, clean and reassemble the ion source. Old or damaged MS source filament. Replace the source filament. Poor MS calibration. Re-calibrate the MS. Poor MS tuning. Re-tune the MS. Loose connections in the MS ion source. Check the source connections. Back to MS troubleshooting problem list. 57
58 PTV Sensitivity problems Low sensitivity is commonly caused by incorrect operating conditions. It might also be due to a dirty liner or to a liner wrong sized with regards to the volume of the sample injected. Lack of sensitivity for selected compounds can also be related to discrimination phenomena induced by inadequate injection conditions. See PTV Discrimination problems and the related topics. Back to PTV troubleshooting problem list. PTV Discrimination problems Possible discrimination of the heavy or volatile fraction of the sample may be caused by operating conditions set incorrectly with regard to the chosen injection mode. Discrimination could also be caused by: degradation of the thermally labile compounds due to the cathalytic effect of bad deactivated quartz wool inside the liner; by-products originated in previous injections that have been not completely eliminated from the liner during the cleaning phase or the cleaning operating sequence. PTV Injector Troubleshooting provides a list of possible symptoms relating to discrimination when using a PTV Injector. Select the one that best matches your problem, and then investigate the probable causes and the suggested remedies. Back to PTV troubleshooting problem list. Discrimination of heavy compounds in splitless mode (PTV) Transfer temperature too low. Set Transfer temp and cleaning (without backflush) temperature to a higher value. Transfer time too low. Set transfer time or cleaning time to a value close to the whole GC run time. Splitless time too short. Set Splitless time to a higher value. Initial temperature too high with regard to the solvent boiling point. Set Inject temp to a value closer to the solvent boiling point. Flooding of the liner Set Inject temp to a 58
59 due to low injection temperature. higher value. If the symptom does not disappear, insert quartz wool in the liner. Refer to Adjust a Quartz Wool Packing Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Flooding of the liner due to injected volume too large. Reduce the amount of sample to be injected. Replace the liner with one of adequate diameter. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Flooding of the liner due to insufficient size. Replace the liner with one of larger diameter. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Dirty liner. Clean or replace the liner. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Liner not suitable for the actual sample. Replace the liner with an adequate one. A lower diameter liner would improve the sample transfer. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance 59
60 and Troubleshooting Manual. Quartz wool causing excessive retention of high molecular compounds. Replace the packed liner with a new one filled will quartz wool. Replace the packed liner with an empty one. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Back to PTV troubleshooting problem list. Discrimination of volatile compounds in splitless mode (PTV) Initial temperature too high. Set Inject temp to a lower value. Liner diameter too small. Replace the liner with one of larger diameter. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Volatiles are eliminated through the purge line. Set Const purge flow? to N, then set Stop purge for to a value corresponding to Splitless time. Back to PTV troubleshooting problem list. 60
61 Discrimination of volatile compounds in solvent split mode (PTV) Initial temperature too high. Set Inject temp to a lower value. Injection and evaporation time are set as to exceed the time required for the solvent elimination. Set Inject time and Evaporation time to adequate values. Split flow set too high. Set Split flow to a lower value. Lack of an adequate quartz wool packing inside the liner. Insert quartz wool in the liner. Back to PTV troubleshooting problem list. Discrimination in split mode (PTV) Initial temperature too low. Set Inject temp to a higher value. Injected volume too large. Reduce the amount of sample to be injected. Replace the liner with the 2 mm ID version. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Lack of an adequate quartz wool packing inside the liner. Replace the packed liner with a new one filled will quartz wool. Fix the quartz wool packing inside the 61
62 liner. Refer to Adjust a Quartz Wool Packing Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Back to PTV troubleshooting problem list. Excessive broadening of solvent peak (PTV) Note: Volatile compounds may also be obscured and a lack of sensitivity observed. Injection pressure set too high. Set Inject pressure to a lower value. Back to PTV troubleshooting problem list. Sample degradation (PTV) Dirty liner. Clean or replace the liner. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Transfer temperature set too high. Set Transfer temp to a value adequate to the nature of the sample compounds. Liner diameter too large. Replace the liner with one of smaller diameter to improve transfer efficiency. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting 62
63 Manual. Cathalytic and thermal degradation of sensitive compounds caused by the quartz wool inside the liner. Use a new packed liner. If the symptom does not disappear, use an empty liner. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Back to PTV troubleshooting problem list. PTV Large Volume - Poor reproducibility of results Note: Excessive broadening of the solvent peak and broadening and irregular shape of sample peaks may also be observed. Injection speed set too high. Reduce injection speed. Injected volume too large. Reduce the amount of sample to be injected. Initial temperature set too low. Set Inject temp to a higher value. Injection pressure set too high. Set Inject pressure to a lower value. Lack of an adequate quartz wool packing inside the liner. Fix the quartz wool packing inside the liner. Refer to Adjust a Quartz Wool Packing Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Replace the liner with another filled with quartz wool. Refer to 63
64 Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Back to PTV troubleshooting problem list. PTV Large Volume - Discrimination of volatile compounds Note: Solvent peak may be significantly reduced. Injection speed set too low. Set Inject speed to a higher value. Initial temperature set too high. Set Inject temp to a lower value. Split flow set too high. Lower the split flow. Lack of an adequate quartz wool packing inside the liner. Fix the quartz wool packing inside the liner. Refer to Adjust a Quartz Wool Packing Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Replace the liner with another filled with quartz wool. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Back to PTV troubleshooting problem list. 64
65 PTV Large Volume - Sample degradation Cathalytic degradation of sensitive compounds caused by the quartz wool. Use an empty deactivated liner reducing the volume injected. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Cathalytic degradation of sensitive compounds caused by byproducts accumulated on the quartz wool. Replace the packed liner with a new one. Refer to Clean or Replace the Liner Operating Sequence in Chapter 6 of the Maintenance and Troubleshooting Manual. Back to PTV troubleshooting problem list. Analyses results are not reproducible (GSV) Loading time is too short (loop not completely filled by sample). Increase the loading time. Injecting time is too short (sample not completely transferred to the column). Increase the injection time. Leaks on the valve or gas line connections. Check all valve and gas line connections for leaks. The rotor of the valve is worn and causes gas leaks. Replace the valve body. 65
66 The diaphragm of the valve is worn and causes gas leaks. Replace the diaphragm. Refer to Replace the Diaphragm on an Automatic Valve in Chapter 7 of the Maintenance and Troubleshooting Manual. The frit of the filter is dirty. Clean or replace the frit. Refer to Clean or Replace the Filter on an Automatic Valve in Chapter 7 of the Maintenance and Troubleshooting Manual. Back to GSV troubleshooting problem list. No signal at all (GSV) The frit of the filter is obstructed. Replace the frit. Refer to Clean or Replace the Filter on an Automatic Valve in Chapter 7 of the Maintenance and Troubleshooting Manual. Back to GSV troubleshooting problem list. GSV does not work Servo air pressure absent or too low. Check the servo air source. Faulty electrovalve on the servo air line. Contact your service representative. The diaphragm is Replace the 66
67 damaged. diaphragm. Refer to Replace the Diaphragm on an Automatic Valve in Chapter 7 of the Maintenance and Troubleshooting Manual. Back to GSV troubleshooting problem list. Checking for leaks - Capillary Columns, DPFC, DGFC 1. With the capillary column installed, cool down the injector and the detector base body to room temperature. 2. Only for FID, NPD, FPD: remove the detector. Do not remove ECD detector. 3. Only for FPD: remove the jet. Use typewriter correction fluid to mark the position of the column entering the detector base body. Lower the column end down to the top of the detector base body, then tighten the M4 fixing nut to restore system tightness. 4. Finger tighten the relevant blind cap (fitted with a silicon seal) on the detector base body, to seal the column flow path. 5. Only for ECD: remove the chimney and seal the line with the relevant blind 6MB nut. 6. Turn off any H2, air and make-up gas flows. Do not close the detector s gases supplies. 7. Turn off the split and septum purge vents (if any). 8. Set the inlet pressure to kpa. 9. Set the flow of the make-up gas to the maximum acceptable value to speed the pressurization of the line. 10. Wait until the system is equilibrated. Air, Make-up and H2values need not necessarily be equal. 11. Turn off the inlet pressure and the make-up gas. The shown values should not change more than 5% in 10 minutes. If the values drop down immediately, one or more leaks are present. In this case, carry on with step Check the accessible, critical connections (column to injector, column to detector, split and purge valves, septum cap) with a handheld electronic leak detector to find possible leaks. 13. If no leak is detectable in this way, contact your service representative. Back to Leak-checking. Back to top level troubleshooting list. Checking for leaks - Capillary Columns, DPFC, Non-DGFC 1. With the capillary column installed, cool down the injector and the detector base body to room temperature. 2. Only for FID, NPD, FPD: remove the detector. Do not remove ECD detector. 67
68 3. Only for FPD: remove the jet. Use typewriter correction fluid to mark the position of the column entering the detector base body. Lower the column end down to the top of the detector base body, then tighten the M4 fixing nut to restore system tightness. 4. Finger tighten the relevant blind cap (fitted with a silicon seal) on the detector base body, to seal the column flow path. 5. Only for ECD: remove the chimney and seal the line with the blind 6MB nut. 6. Turn off the H2 and air gas flows. 7. Turn off the split and septum purge vents (if any). 8. Set the inlet pressure to kpa. 9. Open the make-up gas line to speed the pressurization of the detector side of the system. Use an external pressure gauge (e.g. the gauge installed on the bottle) to monitor pressure. 10. Wait until the system is equilibrated. 11. Turn off the inlet pressure and the make-up gas. The shown values should not change more than 5% in 10 minutes. If the values drop down immediately, one or more leaks are present. In this case, carry on with step Check the accessible, critical connections (column to injector, column to detector, split and purge valves, septum cap) with a handheld electronic leak detector to find possible leaks. 13. If no leak is detectable in this way, contact your service representative. Back to Leak-checking. Back to top level troubleshooting list. Checking for leaks - Capillary Columns, Non- DPFC, Non-DGFC 1. With the capillary column installed, cool down the injector and the detector base body to room temperature. 2. Only for FID, NPD, FPD: remove the detector. Do not remove ECD detector. 3. Only for FPD: remove the jet. Use typewriter correction fluid to mark the position of the column entering the detector base body. Lower the column end down to the top of the detector base body, then tighten the M4 fixing nut to restore system tightness. 4. Finger tighten the relevant blind cap (fitted with a silicon seal) on the detector base body, to seal the column flow path. 5. Only for ECD: remove the chimney and seal the line with the blind 6MB nut. 6. Turn off the H2 and air gas flows. 7. Turn off the split and septum purge vents (if any). 8. Open the ON/OFF knob that controls the carrier gas, then adjust the inlet pressure to kpa using a screwdriver. You can monitor the pressure value on the gauge located close to the regulator. 9. Open the make-up gas line to speed the pressurization of the detector side of the system. Use an external pressure gauge (e.g. the gauge installed on the bottle) to monitor pressure. 10. Wait until the system is equilibrated. 11. Turn off the inlet pressure and the make-up gas. The shown values should not change more than 5% in 10 minutes. If the values drop down immediately, one or more leaks are present. In this case, carry on with step Check the accessible, critical connections (column to injector, column to detector, split and purge valves, septum cap) with a handheld electronic leak detector to find possible leaks. 68
69 13. If no leak is detectable in this way, contact your service representative. Back to Leak-checking. Back to top level troubleshooting list. Checking for leaks - Packed Columns, DPFC, DGFC 1. With the packed column installed, cool down the injector and the detector base body to room temperature. 2. Only for FID, NPD, FPD: remove the detector. Do not remove ECD detector. 3. Only for FPD: remove the jet. 4. Finger tighten the relevant blind cap (fitted with a silicon seal) on the detector base body, to seal the column flow path. 5. Only for ECD: remove the chimney and seal the line with the blind 6MB nut. 6. Turn off any H2, air and make-up gas flows. Do not close the detectors gases supplies. 7. Turn off the split and septum purge vents (if any). 8. Set the inlet pressure to kpa. 9. Wait until the system is equilibrated (Air, Make-up and H2 values need not necessarily be equal). 10. Turn off the inlet pressure. Pressure value should not change more than 5% in 10 minutes. If the value drops down immediately, one or more leaks are present. In this case, carry on with step Check the accessible, critical connections (column to injector, column to detector, split and purge valves, septum cap) with a handheld electronic leak detector to find possible leaks. 12. If no leak is detectable in this way, contact your service representative. Back to Leak-checking. Back to top level troubleshooting list. Checking for leaks - Packed Columns, DPFC, Non- DGFC 1. With the capillary column installed, cool down the injector and the detector base body to room temperature. 2. Only for FID, NPD, FPD: remove the detector. Do not remove ECD detector. 3. Only for FPD: remove the jet. 4. Finger tighten the relevant blind cap (fitted with a silicon seal) on the detector base body, to seal the column flow path. 5. Only for ECD: remove the chimney and seal the line with the blind 6MB nut. 6. Turn off any H2, air and make-up gas flows. 7. Turn off the split and septum purge vents (if any). 8. Set the inlet pressure to kpa. 9. Wait until the system is equilibrated. 10. Turn off the inlet pressure. The shown value should not change more than 5% in 10 minutes. If the value drops down immediately, one or more leaks are present. In this case, carry on with step
70 11. Check the accessible, critical connections (column to injector, column to detector, split and purge valves, septum cap) with a handheld electronic leak detector to find possible leaks. 12. If no leak is detectable in this way, contact your service representative. Back to Leak-checking. Back to top level troubleshooting list. Checking for leaks - Packed Columns, Non-DPFC, Non-DGFC 1. With the capillary column installed, cool down the injector and the detector base body to room temperature. 2. Only for FID, NPD, FPD: remove the detector. Do not remove ECD detector. 3. Only for FPD: remove the jet. 4. Finger tighten the relevant blind cap (fitted with a silicon seal) on the detector base body, to seal the column flow path. 5. Only for ECD: remove the chimney and seal the line with the blind 6MB nut. 6. Turn off any H2, air and make-up gas flows. 7. Turn off the split and septum purge vents (if any). 8. Open the ON/OFF knob that controls the carrier gas, then adjust the inlet pressure to kpa using a screwdriver. You can monitor the pressure value on the gauge located close to the regulator. 9. Wait until the system is equilibrated. 10. Turn off the inlet pressure. The shown value should not change more than 5% in 10 minutes. If the value drops down immediately, one or more leaks are present. In this case, carry on with step Check the accessible, critical connections (column to injector, column to detector, split and purge valves, septum cap) with a handheld electronic leak detector to find possible leaks. 12. If no leak is detectable in this way, contact your service representative. Back to Leak-checking. Back to top level troubleshooting list. 70
71 Multi GC system 71
72 Index A Air leaks Analyses results are not reproducible (GSV) Automatic Column Evaluation Automatic leak check B Baseline - high standing current Baseline abnormally high (PID) Baseline Drift (TCD) Baseline drift with changing pulse voltage (ECD) Baseline drifting Baseline excessively noisy (FPD) Baseline falling Baseline falling away slowly after a high initial value Baseline Fluctuation (TCD) Baseline high (MS) Baseline high frequency noise Baseline irregular shape...23, 24 dip after solvent peak square waves S-shaped Baseline related problems...11, 20 with specific system components Baseline rising under temperature program control Baseline spiking Baseline unstable (FPD) Baseline upset at solvent elution (NPD) Broad Ghost Peaks Broadening of solvent peak (PTV LV) Broadening of solvent peak (PTV) Broadening peaks C Chair-shaped peaks Checking for leaks...17, 67, 68, 69, 70 Capillary Columns DPFC DGFC Non-DGFC Non-DPFC Non-DGFC Packed Columns DPFC DGFC Non-DGFC
73 Non-DPFC Non-DGFC Clipping peaks at bottom Clipping peaks at top Column Evaluation Contaminant peaks D Decreasing retention times Degradation of sample (PTV) Detector gas flow rates (FID) Discrete high-intensity contaminant peaks Discrimination...58, 60, 61 heavy compounds in splitless mode (PTV) in split mode (PTV) volatile compounds in solvent split mode (PTV) volatile compounds in splitless mode (PTV) Discrimination of volatile compounds (PTV LV) Discrimination problems (PTV) Double Peaks DPFC-equipped systems leak-checking Drifting Baseline E ECD Contamination problems ECD gases ECD Sensitivity problems ECD troubleshooting Electron Capture Detector troubleshooting Excessive broadening of solvent peak (PTV) Excessive slope on left side of peaks Excessive slope on right side of sample peaks Excessive slope on right side of solvent peak F Falling Baseline FID Contamination problems FID detector gas flow rates FID Flame Ignition problems FID gases FID operating conditions FID Sensitivity problems FID troubleshooting Flame Ignition problems Flame Ionization Detector troubleshooting Flame Photometric Detector troubleshooting Flow rates (FID detector gas) FPD gas flow rates FPD gases FPD sensitivity low...47, 49 and water droplets generate between the heat shields FPD Sensitivity problems FPD standing current...46, 47 high
74 none FPD temperature does not reach the set point FPD troubleshooting Fronting G Gas leaks Gas Sampling Valve does not work Gas Sampling Valve troubleshooting GC Troubleshooting - Top Level Ghost Peaks broad GSV does not work GSV troubleshooting H High background level (NPD) High base frequency (ECD) High Baseline (MS) High baseline (PID) High standing current and noise (FPD) How to use this Troubleshooting Help system I Identifying leaks Inconsistent retention times Increasing retention times Irregular Peaks Irregular shape of sample peaks (PTV LV) L Lack of sensitivity (PTV) Lamps (UV) for PID Leak checking Leak-checking...19, 67, 68, 69, 70 Capillary Columns DPFC DGFC Non-DGFC Non-DPFC Non-DGFC lines from the Gas Source to the GC Packed Columns DPFC DGFC Non-DGFC Non-DPFC Non-DGFC Leak-checking DPFC-equipped systems Leak-checking non-dpfc systems Leaks identifying and removing Low carbon rejection (NPD) Low reproducibility...33, 36 of peaks area
75 of retention times Low sensitivity (FPD)...47, 49 and water droplets generate between the heat shields Low Sensitivity (MS) Low sensitivity (PID) Low Sensitivity (TCD) Low sensitivity and water droplets generate between the heat shields (FPD) M Mass Spectrometer troubleshooting MS sensitivity low MS troubleshooting Multi GC system leak-checking N Negative dips after peaks (ECD) Negative Peaks Negative Peaks (TCD) Nitrogen Phosphorous Detector troubleshooting No Peaks...29, 30 after solvent peak at all No signal at all (GSV) No standing current (FPD) No standing current (PID) Non-DPFC systems leak-checking NPD detector response lower than expected none NPD gases NPD operating conditions NPD Sensitivity problems NPD shows FID-like response for solvent and other carbon based compounds NPD troubleshooting O Operating conditions (FID) P Peak Fronting Peak related problems...12, 25 with specific system components Peak tailing sample solvent Peak Tailing (MS) Peaks broadening Peaks clipping at bottom Peaks clipping at top Performing an Automatic Column Evaluation Performing an Automatic Leak Check Photoionization Detector troubleshooting PID Contamination problems
76 PID sensitivity low PID Sensitivity problems PID standing current none PID troubleshooting Poor Reproducibility of Results (MS) Poor reproducibility of results (PTV LV) Poor sensitivity with increased retention time with normal retention time PTV Discrimination problems PTV Injector troubleshooting PTV Large Volume - Discrimination of volatile compounds PTV Large Volume - Poor reproducibility of results PTV Large Volume - Sample degradation PTV Sensitivity problems R Radioactivity warning Removing leaks Reproducibility of peaks area low Reproducibility poor (MS) Results not reproducible (GSV) Results related problems...12, 33 with specific system components Retention times low reproducibility Retention times are inconsistent Retention times decreasing Retention times increasing Rising Baseline S Sample degradation (PTV LV) Sample degradation (PTV) Sample Peak Tailing Sample peak tailing (PID) Sensitivity low (MS) Sensitivity low (PID) Sensitivity low (TCD) Sensitivity poor - increased retention time Sensitivity poor - normal retention time Sensitivity poor (PTV) Sensitivty low (FPD)...47, 49 and water droplets generate between the heat shields Solvent Peak Tailing Solvent peak tailing (PID) Solvent quenching effect (NPD) Spiking (baseline) Standing current (FPD)...46, 47 high none Standing current (PID) none
77 T Tailing of sample peaks (PID) Tailing of solvent peak (PID) TCD does not work TCD sensitivity low TCD Sensitivity problems TCD troubleshooting Thermal Conductivity Detector troubleshooting U Unresolved Peaks Unstable and excessively noisy baseline (FPD) Unstable baseline (NPD) UV lamp does not light at all (PID) UV lamp does not light immediately (PID) UV lamps for PID V Volatile compounds obscured (PTV) W Warning Radioactivity
23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
The Electron Capture Detector (ECD)
Theory of operation The ECD is based on the phenomenon that electronegative species can react with thermal electrons present to form negatively charged ions. The loss of such electrons is related to the
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
Thermal Conductivity Detector
Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
Gas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
Gas Chromatograph GC Consumables. p/n: 220-94898-02
Gas Chromatograph GC Consumables p/n: 220-94898-02 Diag. # 1) Septum Nut 2) Needle Guide 3) Septum 4) Injection Port A Assy 5) O-ring for Glass Insert 6) Glass Insert 7) Gold Gasket 8) Capillary Adaptor
The Use of Hydrogen Carrier Gas for GC/MS
Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors
Clarus 580 Gas Chromatograph
S P E C I F I C A T I O N S Gas Chromatography Clarus 580 Gas Chromatograph The PerkinElmer Clarus 580 Gas Chromatograph (GC) is a fully automated gas chromatograph. The system offers the capabilities
THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles
THERMAL DESORPTION Introduction and Principles Thermal Desorption Thermal desorption is a simple extension of the technique of gas chromatography (GC) It involves the use of heat and a flow of inert gas
Split / Splitless Injection for Capillary GC
Split / Splitless Injection for Capillary GC Speakers John V Hinshaw CHROMacademy GC Dean GC Connections editor Moderator Tony Taylor CHROMacademy Trainer Technical Director, Crawford Scientific Dave Walsh
GC-2010 Gas Chromatograph INSTRUCTION MANUAL
221-40407F Gas Chromatograph INSTRUCTION MANUAL Read this instruction manual thoroughly before using this instrument. Save the manual for future reference. Copyright Shimadzu Corporation 2000. All rights
C184-E014D GC-2014. Shimadzu Gas Chromatograph System
C184-E014D GC-2014 Shimadzu Gas Chromatograph System Big Performance & Small Space GC-2014 GC-2014 Shimadzu Gas Chromatograph System High Performance and Expandability Merged at a higher level High Performance
Analysis of Liquid Samples on the Agilent GC-MS
Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >
INSTALLATION INSTRUCTIONS Fan Coil Replacement Coil Kit EBX & EBXX
Fan Coil Replacement Coil Kit EBX & EBXX These instructions must be read and understood completely before attempting installation. These instructions covers the installation of replacement coil kit into
Trouble Shooting. Pump
Trouble Shooting Pump Trouble Possible Cause Remedy Oil leaking in the area of water pump crankshaft Worn crankshaft seal, bad bearing, grooved shaft, or failure of retainer o-ring. Excessive play on crankshaft
Operating Manual Volume 3. Detectors. Agilent 6890 Series Gas Chromatograph
Operating Manual Volume 3. Detectors Agilent 6890 Series Gas Chromatograph Agilent Technologies 2000 All rights reserved. Reproduction, adaptation, or translation without permission is prohibited, except
CARING FOR YOUR WATER HEATER
http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html Water Heater Inspections CARING FOR YOUR WATER HEATER Venting System (Direct Vent Only) The venting system should be inspected
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Hydrogen as an alternative to Helium for gas chromatography.
Application Note Gas Chromatography. Application Note for Specialty Gases. as an alternative to for gas chromatography. Author: Peter Adam Linde AG, Linde Gas Division Seitnerstrasse 70, 82049 Pullach,
Gas Oven Repair Guide
- /6 - Gas Oven Repair Guide [ FX70*, FX50*] Ver. Aug-0 Ignition Failure Cooktop ignition Oven ignition Heating defect Oven Cooktop Abnormal Flame Button, Motors Lamp Others Smell, Smoke, Noise No Display,
Chapter 5 - Aircraft Welding
Chapter 5 - Aircraft Welding Chapter 5 Section A Study Aid Questions Fill in the Blanks 1. There are 3 types of welding:, and, welding. 2. The oxy acetylene flame, with a temperature of Fahrenheit is produced
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in
San josé OWNER S MANUAL
San josé OWNER S MANUAL Assembling & operating manual San josé 30 mbar - PORTABLE GAS BARBECUE 1. 2. 3. Improper installation, adjustment, alteration, service or maintenance can injury or property damage.
CDS TROUBLESHOOTING SECTION I. VACUUM. 1.0. Weak vacuum at wand. Gauge reads normal (10hg to 14hg)
CDS TROUBLESHOOTING SECTION I. VACUUM 1.0. Weak vacuum at wand. Gauge reads normal (10hg to 14hg) 1.1. Clogged hoses or wand tube. Disconnect hoses and carefully check for an obstruction. 1.2. Excessive
Mass Spectrometry. Fundamental GC-MS. GC Considerations
Mass Spectrometry Fundamental GC-MS GC Considerations i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
Evaluate, Clean, and Tune Guidance
Evaluate, Clean, and Tune Guidance The Evaluate, Clean and Tune (ECT) process serves three essential purposes in the Weatherization Assistance Program (WAP). The first is to evaluate the existing system
Micro Cryo-Trap Model 971
Micro Cryo-Trap Model 971 for Liquid CO2 Cooling Scientific Instrument Services, Inc. 1027 Old York Road Ringoes, NJ 08551 Phone (908) 788-5550 Part #951001M revised 10/ /98 Notice The information in this
Cryogenic Vacuum Pumps
Oxford Instruments Austin Cryogenic Vacuum Pumps Decontamination Procedure Only use helium gas with at least 99.999% purity for decontamination procedure described in this document. Decontamination procedures
Heating, Ventilation, Air Conditioning and Refrigeration (HVACR)
Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) I. Demonstrate safety skills in typical HVACR work situations to NATE Core Installer Knowledge Areas for Technician Excellence for Safety
EnAlgae SOP: Biogas analysis
Table of content 1. PURPOSE... 2 2. PRINCIPLE... 2 3. REQUIREMENTS... 3 3.1. EQUIPMENT AND MATERIALS... 3 3.2. REAGENTS... 3 3.3. PREPARATION OF MEASUREMENT... 3 4. PROCEDURE... 4 4.1. SAMPLE MEASUREMENT...
Electrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
TIG INVERTER INSTRUCTION MANUAL
TIG INVERTER INSTRUCTION MANUAL Contents Warning General Description Block Diagram Main Parameters Circuit Diagram Installation and Operation Caution Maintenance Spare Parts List Troubleshooting 3 4 4
NITROUS TRANSFER PUMP INSTRUCTIONS
NITROUS TRANSFER PUMP INSTRUCTIONS SAFETY TIPS Never directly inhale nitrous oxide. When inhaled in large quantities, nitrous oxide can cause respiratory ailments or in extreme cases, death by suffocation.
Fractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
http://waterheatertimer.org/how-to-troubleshoot-gas-water-heater.html
http://waterheatertimer.org/how-to-troubleshoot-gas-water-heater.html TECHNICAL SERVICE DEPARTMENT Effective October 2007, we transitioned to the White Rodgers (Intelli-Vent TM )Thermostat Control for
Application Requirement
Hazardous Gas Monitors Sensor Selection Overview Safety Monitoring (LEL) Toxic Limit Detection (PEL) Leak Detection Personal Safety Application Requirement Exposure Assessment (TWA) Ambient Air Quality
Compressed Gas Cylinder Program
Department of Environmental Health & Safety Procedures Compressed Gas Cylinder Program March 2010 Compressed Gas Cylinder Program Page 1 of 8 Table of Contents I. Introduction II. III. IV. Policy Scope
Fundamentals of Mass Flow Control
Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of
Threaded Inline Air Heaters
Threaded Inline Air Heaters FOR SAFETY & LONG HEATER LIFE, CAREFULLY READ THIS MANUAL BEFORE USE. Safety SHOCK HAZARD! Only qualified individuals should install this heater and related controls. Follow
SL280UHV SERIES GAS FURNACE WARNING
2010 Lennox Industries Inc. Dallas, Texas, USA 506677 01 11/2010 Supersedes 506409 01 SL280UHV SERIES GAS FURNACE Litho U.S.A. FIRE OR EXPLOSION HAZARD. Failure to follow safety warnings exactly could
Chapter 7 Hydraulic System Troubleshooting
Chapter 7 Hydraulic System Troubleshooting General The following troubleshooting information is provided as a general guide to identify, locate and correct problems that may be experienced with the hydraulic
User Guidelines & Standard Operating Procedure for the Varian 3800 GC
User Guidelines & Standard Operating Procedure for the Varian 3800 GC GC/MS Standard Operating Procedure ii TABLE OF CONTENTS DISCLAIMER...iii ACKNOWLEDGEMENTS... iv 1. INTRODUCTION...1 1.1 Purpose of
orlando OWNER S MANUAL
orlando OWNER S MANUAL 2 Assembling & operating manual Orlando 30 mbar - PORTABLE GAS BARBECUE 1. 2. 3. Improper installation, adjustment, alteration, service or maintenance can injury or property damage.
Please keep this manual for f uture reference.
GAS WATER/AGUA Congratulations! You've just purchased a new Marey Portable tankless water heater and will soon begin to enjoy the benefits of going tankless. Take the time to thoroughly read and understand
12 Volt 30 Amp Digital Solar Charge Controller
12 Volt 30 Amp Digital Solar Charge Controller User s Manual WARNING Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions
Thermo Top - Troubleshooting Tree
Thermo Top - Troubleshooting Tree 07-15-2002 CAUTION Troubleshooting requires comprehensive knowledge about the structure and theory of operation of the Thermo Top heater. Troubleshooting and repairs may
Hydraulic Troubleshooting PRESENTED BY
Hydraulic Troubleshooting PRESENTED BY NORMAN KRONOWITZ Introduction Welcome to the CMA/Flodyne/Hydradyne s Hydraulic Troubleshooting presentation. We will introduce many aspects of troubleshooting hydraulic
E - THEORY/OPERATION
E - THEORY/OPERATION 1995 Volvo 850 1995 ENGINE PERFORMANCE Volvo - Theory & Operation 850 INTRODUCTION This article covers basic description and operation of engine performance-related systems and components.
Chromatograph Maintenance and Troubleshooting Class # 5050
Chromatograph Maintenance and Troubleshooting Class # 5050 Shane Hale Business Development Manager Emerson Process Management Gas Chromatographs Division 5650 Brittmoore Drive Houston, TX 77041 USA Introduction
Your safety and the safety of others are very important.
NATURAL GAS TO PROPANE CONVERSION KIT 090 INSTALLATION INSTRUCTIONS FOR ALTITUDES 0 -,00 FT. ONLY PROPANE CONVERSION KIT SAFETY... INSTALLATION REQUIREMENTS... Tools and Parts... LP Gas Requirements...
CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT
CRYOCOOLER CONTAMINATION STUDY S.W.K. Yuan, D.T.Kuo, and A.S. Loc EI Technologies, Cryocooler Group Sylmar, CA 91342 ASTRACT Eliminating or reduction of contamination is essential for the success of cryocoolers
Toroidal Conductivity Sensor
Instruction Sheet PN 51A-/rev.C June 2012 Toroidal Conductivity Sensor For additional information, please visit our website at www.emersonprocess.com/rosemountanalytical.com SPECIFICATIONS Wetted Materials:
AUTOMATIC CHANGEOVER MANIFOLD Series 915
AUTOMATIC CHANGEOVER MANIFOLD Series 915 FEATURES Fully automatic, simple, hassle-free operation No variance in delivery pressure Designed for ultra high purity gas service Built-in audio and visual alarm
Tuning & Mass Calibration
Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within
SELECTION, APPLICATION AND MAINTENANCE
DIESEL PROTECTION SYSTEMS Automatic Diesel Engine Shut Down System for Safe Area Applications SELECTION, APPLICATION AND MAINTENANCE Series 300 Series 310 SYSTEM DESCRIPTION Suitable for attended engine
Operating Instructions & Maintenance Guidelines for Precision Vaporizers
Operating Instructions & Maintenance Guidelines for Precision Vaporizers This page intentionally blank. Examination and Preparation for Use We want you to be completely satisfied with your vaporizer. Please
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE
ENGINE COOLING SYSTEM
ENGINE COOLING SYSTEM 1988 Toyota Celica 1987-88 TOYOTA Engine Cooling Systems Celica DESCRIPTION The basic liquid cooling system consists of a radiator, water pump, thermostat, cooling fan, pressure cap,
Model 43AP Pneumatic Controller, Style B
Instruction MI 011-476 January 1980 Model 43AP Pneumatic Controller, Style B Installation and Operation Model 43AP Controller continuously detects the difference between a process measurement and its set
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
Hydraulic Trouble Shooting
Hydraulic Trouble Shooting Hydraulic systems can be very simple, such as a hand pump pumping up a small hydraulic jack, or very complex, with several pumps, complex valving, accumulators, and many cylinders
ASK THE EXPERT: Burner Troubleshooting Information & Maintenance
ASK THE EXPERT: Burner Troubleshooting Information & Maintenance The burner is the heart of your BBQ. It is subject to a number of conditions that can cause damage, and lead to potential safety issues.
A.Y. McDonald Mfg. Co. Troubleshooting Submersible and Jet Pumps
A.Y. McDonald Mfg. Co. Troubleshooting Submersible and Jet Pumps Troubleshooting Submersible Pumps Fuse overload or circuit breaker trips when motor is started 1. Incorrect line voltage. Check the line
FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases
FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY RESULT IN INJURY TO YOU OR DAMAGE TO
Why and How we Use Capacity Control
Why and How we Use Capacity Control On refrigeration and air conditioning applications where the load may vary over a wide range, due to lighting, occupancy, product loading, ambient weather variations,
Compressor Service & Maintenance Manual
Compressor Service & Maintenance Manual C Series COMPRESSOR (D)C1103 (D)C1203 (D)C2106 (D)C2206 (D)C3210 Copyright 2006 DCI. All Rights Reserved. 92311, Rev. C, 08/13 1 C1000 Series Service & Maintenance
Oxy-Fuel Gas Welding. Given a functional oxy-fuel gas unit, instruction and demonstration of use, each student will be able to:
I. Competencies Oxy-Fuel Gas Welding Given a functional oxy-fuel gas unit, instruction and demonstration of use, each student will be able to: A. Identify the major parts of the oxy-fuel gas unit. B. Pass
Geothermal Alliance of Illinois. TXVs Theory and Fundamentals John Haug Senior Application Engineer Emerson Climate Technologies - Flow Controls
Geothermal Alliance of Illinois TXVs Theory and Fundamentals John Haug Senior Application Engineer Emerson Climate Technologies - Flow Controls Thermal Expansion Valve Topics Anatomy Operation Terms &
SCION SQ Series GC-MS
SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative
COMPRESSOR REPLACEMENT PROCEDURE
R-22 and R-410A COMPRESSOR REPLACEMENT PROCEDURE Bard Manufacturing Company, Inc. Bryan, Ohio 43506 Since 1914...Moving ahead just as planned Manual No.: 2100-003F Supersedes 2100-003E File: Volume I,
Munchkin Heat Exchanger Maintenance
Munchkin Heat Exchanger Maintenance By: Mike Granahan *Refer to the appropriate boiler manual for your specific model number. Adjustment procedures will vary. Boiler shown below is an HTP Munchkin 80M
Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ)
Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) Closing Date and Time: Wednesday, April 13, 2016 at 5:00 p.m. (Pacific Time) Contact: Aaron Galloway Fax: 541-888-3250
ORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS BEFORE USE GENERAL SAFETY INSTRUCTIONS: ALWAYS OPERATE THE UNIT IN AN UPRIGHT POSITION AND PLACE
TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here. Operating Instructions
TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here Operating Instructions Table of Contents 1 INTRODUCTION Safety 2 Sample Preparation 3 2 BACKGROUND Background Information 4 Resources
36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve
Operator: Save these instructions for future use! FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL COULD CAUSE PERSONAL INJURY AND/OR PROPERTY DAMAGE. DESCRIPTION
580-6001 INSTALLATION AND OPERATION INSTRUCTIONS NOT FOR OXYGEN OR OTHER OXIDIZING GASES
ADI 3205-A Certified ISO 9001 580-6001 Micron Filter Assembly INSTALLATION AND OPERATION INSTRUCTIONS NOT FOR OXYGEN OR OTHER OXIDIZING GASES Before Installing or Operating, Read and Comply with These
Diesel: Troubleshooting
Diesel: Troubleshooting Probable Cause Engine not starting Hard to start engine Runs rough at lower RPM Lack of power Diesel knock / pinking Black White Blue Low compression X X X Low fuel pressure X X
ON-VEHICLE INSPECTION
Sight Glass I11244 ONVEHICLE INSPECTION 1. INSPECT REFRIGERANT VOLUME Observe the sight glass on the liquid tube. AC3 AC22F05 Test conditions: Running engine at 1,500 rpm Blower speed control switch: HI
Micro GC Fusion Pre-Installation Checklist
Site Preparation Proper site preparation prior to installing is very important. Contact INFICON (www.inficon.com) with any questions regarding site preparation. NOTE: Checkboxes ( ) identify hardware that
EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board.
EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board. This Troubleshooting Guide covers the electric coolant valve and control
Expectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
G670 Intermittent Pilot Ignition Controls
Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin Issue Date 0300 Intermittent Pilot Ignition Controls Figure 1: Intermittent Pilot Ignition Control
6.7L Diesel Fuel System Contamination Diagnosis and Service Procedure Job Aid (Revised March, 2012)
6.7L Diesel Fuel System Contamination Diagnosis and Service Procedure Job Aid (Revised March, 2012) Fuel contamination on 6.7L diesel engines can damage fuel system components including the High Pressure
USER S, MAINTENANCE and SERVICE INFORMATION MANUAL
CONTENTS SAFETY INFORMATION................ 2 FOR YOUR SAFETY....................... 2 SYSTEM OPERATION.................. 2 THERMOSTATS........................... 2 INTERMITTENT IGNITION DEVICE...........
ARCO Electric Products Installation and Maintenance Manual Low Voltage Automatic Power Factor Correction Capacitor Systems 2013
ARCO Electric Products Installation and Maintenance Manual Low Voltage Automatic Power Factor Correction Capacitor Systems 2013 READ CAREFULLY These instructions are intended to cover good practices in
Theory and Instrumentation of GC Introduction
Theory and Instrumentation of GC Introduction i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
Volkswagen Jetta, Golf, GTI 1999, 2000 2.8 Liter VR6 2V Engine Mechanical, Engine Code(s): AFP 17 Engine-Lubrication (Page GR-17)
17 Engine-Lubrication (Page GR-17) Lubrication system components, removing and installing Oil filter housing, disassembling and assembling Oil pan, removing and installing Oil pressure and oil pressure
Energy Efficient Operations and Maintenance Strategies for Boilers
Energy Efficient Operations and Maintenance Strategies for Boilers Large, complex and widely used, industrial boilers are major consumers of fuel. Proper operations and maintenance (O&M) procedures must
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
Recharge Procedure for Telemark TVP-2000 and TVP-3500
Recharge Procedure for Telemark TVP-2000 and TVP-3500 Required items: 1.) Refrigeration service manifold. 2.) Charge recovery system. 3.) Dry nitrogen gas. If a high pressure cylinder is used, a pressure
INSTALLATION INSTRUCTIONS & HOME OWNERS MANUAL EEMAX ELECTRIC TANKLESS WATER HEATERS IMPORTANT SAFETY INFORMATION
INSTALLATION INSTRUCTIONS & HOME OWNERS MANUAL EEMAX ELECTRIC TANKLESS WATER HEATERS IMPORTANT SAFETY INFORMATION When installing or using any high voltage electrical appliance, basic safety precautions
Operator Quick Guide EC SENSOR
Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
System Saver 318 Air Compressor for Mack E-Tech and ASET Engines
Maintenance Manual 31 System Saver 318 Air Compressor for Mack E-Tech and ASET Engines Revised 08-05 NON-THROUGH DRIVE THROUGH DRIVE Service Notes About This Manual This manual provides service and repair
