Anisotropy Measurement Using the Agilent Cary Eclipse Fluorescence Spectrophotometer



Similar documents
Molecular Spectroscopy

Agilent Cary 60 UV-Vis

Chemical Basis of Life Module A Anchor 2

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Lecture 1: Basic Concepts on Absorption and Fluorescence

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Copyright by Mark Brandt, Ph.D. 12

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1

h e l p s y o u C O N T R O L

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

CNAS ASSESSMENT COMMITTEE CHEMISTRY (CH) DEGREE PROGRAM CURRICULAR MAPPINGS AND COURSE EXPECTED STUDENT LEARNING OUTCOMES (SLOs)

Fluorescent dyes for use with the

Experiment #5: Qualitative Absorption Spectroscopy

COURSE TITLE COURSE DESCRIPTION

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

STUDENT BACKGROUND READING FOR EXPERIMENT C: COLORIMETRIC GOLD NANOSENSOR

Enzymes: Practice Questions #1

Keystone Review Practice Test Module A Cells and Cell Processes. 1. Which characteristic is shared by all prokaryotes and eukaryotes?

Time out states and transitions

Uses of Derivative Spectroscopy

BIOLOGICAL MEMBRANES: FUNCTIONS, STRUCTURES & TRANSPORT

Milk protein analysis with the Agilent 2100 Bioanalyzer and the Agilent Protein 80 kit

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Ultraviolet Spectroscopy

SPECTROSCOPY. Light interacting with matter as an analytical tool

Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening

Real-time PCR: Understanding C t

Common Defects in Digital Printing. Paul Geldenhuys & Amir Shapira January, 2009

Infrared Spectroscopy 紅 外 線 光 譜 儀

Marmara Üniversitesi Fen-Edebiyat Fakültesi Kimya Bölümü / Biyokimya Anabilim Dalı PURIFICATION AND CHARACTERIZATION OF PROTEINS

UV-Vis spectrophotometers

Which regions of the electromagnetic spectrum do plants use to drive photosynthesis?

Joshua Blose, Ph.D., Biochemistry

Energy & Enzymes. Life requires energy for maintenance of order, growth, and reproduction. The energy living things use is chemical energy.

Biological cell membranes

USE OF REFERENCE MATERIALS IN THE LABORATORY

Fractional Distillation and Gas Chromatography

APPLIED. Measuring Caffeine Concentration. Application Summary. Analytes: caffeine (C 8 H 10 N 4 O 2

Chemical reactions allow living things to grow, develop, reproduce, and adapt.

Sample Liver Enzyme Lab

3 - Atomic Absorption Spectroscopy

Spectroscopic Characterization of Cytochrome C Entrapped in AOT Reverse Micelles. Kristen Grinstead

Using Spectrophotometers to Examine Photosynthetic Rates Under Various Qualities of Light

Chapter 2. The Chemistry of Life Worksheets

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

TBA-FX8 Automated Clinical Chemistry Analyzer with Flexibility to Respond to Demands of Large-Scale Clinical Laboratories

Enzymes: Amylase Activity in Starch-degrading Soil Isolates

FTIR Instrumentation

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

Master's Programs. Applied Chemisty. Subject

5.33 Lecture Notes: Introduction to Spectroscopy

Tissue Culture 1 Cell/ Microplates 2 HTS- 3 Immunology/ HLA 4 Microbiology/ Bacteriology Purpose Beakers 5 Tubes/Multi-

Laboratory 5: Properties of Enzymes

Design of Experiments for Analytical Method Development and Validation

ZETA POTENTIAL ANALYSIS OF NANOPARTICLES

How to Biotinylate with Reproducible Results

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Thermo Scientific NanoDrop Products. Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules

Co Extra (GM and non GM supply chains: Their CO EXistence and TRAceability) Outcomes of Co Extra

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

Biology 3A Laboratory: Enzyme Function

Secondary Ion Mass Spectrometry

Features of the formation of hydrogen bonds in solutions of polysaccharides during their use in various industrial processes. V.Mank a, O.

It has long been a goal to achieve higher spatial resolution in optical imaging and

Separation of Amino Acids by Paper Chromatography

2. Spin Chemistry and the Vector Model

DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED

Basic principles and mechanisms of NSOM; Different scanning modes and systems of NSOM; General applications and advantages of NSOM.

Transmembrane proteins span the bilayer. α-helix transmembrane domain. Multiple transmembrane helices in one polypeptide

Quantitation of Peptides and Amino Acids with a Synergy HT using UV Fluorescence

ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.

Prentice Hall. Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition. High School. High School

Review of Chemical Equilibrium 7.51 September free [A] (µm)

LED Lighting - Error Consideration for Illuminance Measurement

PCV Project: Excitons in Molecular Spectroscopy

Qualitative NIR Analysis for Ingredients in the Baking Industry

PHYSIOLOGY AND MAINTENANCE Vol. II - On The Determination of Enzyme Structure, Function, and Mechanism - Glumoff T.

CHM333 LECTURE 13 14: 2/13 15/12 SPRING 2012 Professor Christine Hrycyna

CHEMISTRY. Faculty. Programs Offered. Bachelor of Science in Chemistry (certified by the American Chemical Society) Careers in Chemistry

VWR SPECTROPHOTOMETERS

Front face fluorescence analysis to monitor food process contaminants

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION

Guidance for Industry

Using the Spectrophotometer

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

The Lipid Bilayer Is a Two-Dimensional Fluid

An Introduction to Algae Measurements Using In Vivo Fluorescence

Keystone Exams: Chemistry Assessment Anchors and Eligible Content. Pennsylvania Department of Education

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

RNA) = 1 1 = 1 EU

VTT TECHNICAL RESEARCH CENTRE OF FINLAND

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Absorption of Drugs. Transport of a drug from the GI tract

Figure 5. Energy of activation with and without an enzyme.

3/5/2014. iclicker Participation Question: A. MgS < AlP < NaCl B. MgS < NaCl < AlP C. NaCl < AlP < MgS D. NaCl < MgS < AlP

BSC Exam I Lectures and Text Pages. The Plasma Membrane Structure and Function. Phospholipids. I. Intro to Biology (2-29) II.

Transcription:

Anisotropy Measurement Using the Agilent Cary Eclipse Fluorescence Spectrophotometer A Versatile Instrument for Accurate Measurements Technical Overview Authors Dr. Fabian Zieschang, Katherine McNamara, and Kevin Grant Agilent Technologies, Inc. Introduction A scientific measurement is only useful if it accurately records the characteristics of the sample. To ensure that each measurement is representative, the careful analyst must usually check that sample concentration or volume does not change during sample analysis, that there is no deviation from the expected relationship between analyte concentration and signal intensity, and that no physical change or photodegradation of the sample takes place. However, none of these complicating factors are of concern for the analyst who uses an Agilent Cary Eclipse Fluorescence spectrophotometer to undertake fluorescence anisotropy measurements. This is because fluorescence anisotropy is a nondestructive yet powerful analytical technique that provides robust and reliable measurements that are free from the influence of external variables. The Cary Eclipse Fluorescence spectrophotometer is the only instrument that can reliably measure samples and not cause photodegradation. Because of the intrinsic analytical robustness of fluorescence anisotropy, it is now routinely used to yield accurate, precise, and repeatable data in several key application areas (Table 1). Fluorescence anisotropy measurements have led to a heightened degree of accuracy, significant scientific advances, and important application developments. The advantage of fluorescence measurements made using the Cary Eclipse is that all of this can be accomplished without damaging your precious samples. Technique summary Fluorescence anisotropy measurements are made using waves of light that are all aligned in a single direction (polarized). When light hits a sample, it interacts with the molecules and can generate a secondary signal that can be measured. If polarized light is used to analyze the sample and molecules in the sample are free to move or rotate, not all of this secondary signal will be polarized. In anisotropy, we accurately measure how highly polarized the emitted light is after it has interacted with the sample. This can be a powerful probe into the characteristics of a sample. It also means that anisotropy measurements can provide detailed information about the size and shape of molecules, how easily they can move, and how they change during chemical processes or through variations of time or temperature.

A polarizer is an optical device that allows light of only one orientation to pass through, while rejecting light of all other orientations. In the Cary Eclipse Fluorescence spectrophotometer, polarizers can be placed in the beam before and after the sample. Agilent Cary WinFLR software is then used to record the results of each polarized measurement, and to swap between different types of polarization measurement (anisotropy, polarization, or sum function). The Cary WinFLR software automatically shows the sample spectra and, if appropriate, the result of the anisotropy calculation. Cary WinFLR software can also be used to automatically guide the user through each anisotropy measurement by providing real-time, step-by-step tuition as the experiment progresses. Table 1. Area Polymer science Biological systems Molecular biology Immunology Academia Materials Some Key Application Areas for Fluorescence Anisotropy Application Understand polymer viscosity Identify molecular orientation Characterize phase transition temperature and type Local microviscosity in liquids and polymers Monitoring FRET between identical molecules (homo-fret) Investigate enzyme reactions Determine molecular alignment Estimate the internal viscosity of membranes Investigate membrane fluidity and permeability Size and denaturation of proteins Real-time study of protein-protein interactions Study of protein crowding and oligomerization Measure protein dynamics Characterize DNA-protein interactions Investigate DNA/RNA hybridization Study DNA/RNA structures Examine myosin reorientation Ligand affinity screening Degradative assays Antigen-antibody interaction Immunoassay studies Receptor-ligand binding Understand fundamental physical properties of species Investigate solvent relaxation effects Chain dynamics Multicomponent determination Molecular aggregation Quantify the size of nanoparticles (quantum dots) and macromolecules Energy transfer in optoelectronic materials Investigate guest-host systems (micelles, porous materials, metal-organic frameworks) Anisotropic properties of metal-ligand complexes Samples in applications highlighted in bold text may be photosensitive, and can be readily and irreparably altered when exposed to light from many fluorescence spectrophotometers. The lamp in the Agilent Cary Eclipse Fluorescence spectrophotometer only flashes to acquire a data point. This key technological advantage prevents photodegradation and ensures that these critical samples can be measured without concern [1]. 2

Interpreting anisotropy data The fluorescence anisotropy of a sample is largely controlled by how easily the constituent molecules can move around and change orientation. If all the light that first interacts with a sample is polarized in a single direction, but the Cary Eclipse Fluorescence spectrophotometer records that the emitted light is less polarized, then we can learn a great deal about that sample. Generally, we know that the more viscous a sample, the more rigidly the molecules are held in position. We also know that larger molecules rotate slower than smaller ones. These general principles form the basis of the anisotropy measurements that are routinely used in countless life science applications, to monitor how molecules behave in different key substances, or as a function of, for example, temperature or time. Example application: a well-characterized molecule dissolved in three different media Molecules in a thermoplastic polymethylmethacrylate (PMMA) block are held so tightly that they are essentially frozen. The molecules have a limited ability to move and so yield a high anisotropy value (Figure 1). In a highly viscous liquid such as ethylene glycol they are less strongly held, and in highly fluid media such as water they are largely free to move. Anisotropy 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 Anisotropy: PMMA Anisotropy: Ethylene glycol Anisotropy: Water Emission: Rhodamine B in PMMA 1,000 900 800 700 600 500 400 300 200 100 Intensity (a.u) Figure 1. 0 0 530 550 570 590 610 630 650 670 690 Wavelength (nm) The results of anisotropy measurements of rhodamine B in water (green line), ethylene glycol (red line), and thermoplastic (PMMA) block (blue line) illustrate that large differences in the viscosity of each medium are reflected by large differences in the calculated anisotropy. In water, which does not inhibit the movement of molecules, the anisotropy value is close to zero, whereas the PMMA block is exceptionally good at holding molecules extremely tightly; so tightly, in fact, that the high anisotropy value calculated for rhodamine B in the PMMA sample approaches the theoretical maximum [2]. The anisotropy of rhodamine B in ethylene glycol sits comfortably between the values calculated for water and the PMMA block. This reflects the higher viscosity of ethylene glycol compared to water but greater fluidity relative to the PMMA block. The purple line shows an emission spectrum for rhodamine B in PMMA as collected with vertically polarized light and with a vertical polarizer in the emission beam using the Agilent Cary Eclipse Fluorescence spectrophotometer. 3

Example application: accurately characterize molecular structures It can be difficult to quantitatively determine the amount of a compound in any experiment that follows a temperature profile. This is because sample signals may be as affected by temperature variations as a genuine change in sample concentration. Fluorescence anisotropy is a reliable technique for removing the complicating effects of the physical environment because it only yields information about the molecular sample of interest. Therefore, the effects of temperature are automatically isolated from variations in sample chemistry (Figure 2). Anisotropy 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 Anisotropy: PMMA Anisotropy: Ethylene glycol Intensity: PMMA 900 880 860 840 820 800 780 Intensity W (a.u) Figure 2. 0 760-10 10 30 50 70 90 Temperature ( C) The results of anisotropy measurements of rhodamine B dissolved in ethylene glycol and a thermoplastic (PMMA) block at different temperatures. The green line shows the effect of temperature on the absolute intensity of the rhodamine B signal as a function of temperature. This signal is the result of a complex interaction between different factors associated with the behavior of molecules as temperature changes and, if interpreted incorrectly, may lead the unwary to believe that the concentration of rhodamine B decreased as temperature increased or, perhaps, that the structure of rhodamine B altered significantly with temperature. However, the blue line shows the results of anisotropy measurements made across an identical temperature range. It shows that the fluorescence intensity does not affect the anisotropy, and that the structure of the PMMA block is not temperature-dependent. The data show that the thermoplastic block is exceptionally good at holding rhodamine B molecules tightly from 10 to 100 C. The same cannot be said for ethylene glycol. The results of the anisotropy experiments (red line) show that the structure of ethylene glycol changes as temperature changes, and that, as temperature increases, this allows the rhodamine B molecules more freedom to move around (the anisotropy decreases). 4

Advantages of the Agilent Cary Eclipse Fluorescence spectrophotometer: accurate anisotropy measurements of photosensitive samples You just have to look at any product that has been exposed to sunlight for a while to know that light can easily degrade many organic and inorganic compounds. This common phenomenon can also be a huge concern in the laboratory where the fluorescence spectrophotometer might be progressively degrading your samples during every measurement. More critical still, many of the species that benefit most from the increased sensitivity offered by fluorescence anisotropy measurements (Table 1) are susceptible to photobleaching. In essence, structural and compositional changes to your sample caused by interaction with the light beam will bring the accuracy of the analyses into question. This is potentially frustrating because the integrity of your data is at risk from the very instrument you use to characterize your samples, and doubly so because it can easily be prevented. With its unique optical configuration, the Cary Eclipse Fluorescence spectrophotometer gives you huge analytical advantages. The source lamp in the Eclipse is only on when it is acquiring a data point (it flashes 80 times every second), and because it is switched off completely between each flash, energy from the lamp that can cause photochemical damage does not interact with the sample. This high-speed process has been proven to eliminate photodegradation [1] and ensures that true and meaningful data can be collected from various critical chemical and biological, light-sensitive samples without compromise to accuracy and reproducibility. Conclusions The examples in this study demonstrate that fluorescence anisotropy measurements are a powerful tool in the collection of important molecular information from a range of sample types. Fluorescence anisotropy allows the molecular environment of materials to be accurately imaged and analyzed quickly and without the interfering influences that plague many similar analytical techniques. We have shown that the Agilent Cary Eclipse Fluorescence spectrophotometer can easily be used to illustrate differences in the molecular properties of three common media. References 1. Gavin, P.; Prescott, M.; Fyfe, D. J.; Comerford, J. J. Minimizing photobleaching of Blue Fluorescent Protein (BFP) using the Agilent Cary Eclipse fluorescence spectrophotometer, Agilent Technologies Application Note, publication number 5990-7791EN (2011). 2. Lakowicz, J. R. Ch. 10: Fluorescence Anisotropy. In Principles of Fluorescence Spectroscopy, 3rd Edn. Springer Science + Business Media, LLC, New York (2006). 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2015 Printed in the USA January 7, 2015 5991-6475EN