1 Detection of Multiple Reporter Dyes in Real-time, On-line PCR Analysis with the LightCycler System Gregor Sagner, Cornelia Goldstein, and Rob van Miltenburg Roche Molecular Biochemicals, Penzberg, Germany Introduction The LightCycler System is a fast PCR amplification and analysis system. Central to the function of the LightCycler System are the use of air as a medium for heating and cooling, and the use of capillaries as reaction vessels for efficient temperature exchange between the air and the reagents contained in the capillary. The capillaries are made of borosilicate glass and can therefore also be used as a cuvette. Addition of selected fluorescent dyes to the reaction components allows the PCR to be monitored in real-time and on-line. Fluorescence data are recorded by the optical unit of the instrument, are displayed on a computer screen as PCR proceeds, and are stored for further analysis. A detailed description of the LightCycler System and real-time on-line monitoring of PCR can be found at: The most prominent fluorescent formats used as reporters in the LightCycler System are the doublestranded DNA binding dye SYBR Green I, and the so-called Hybridization Probes format. Whereas signal development in the presence of the SYBR Green I dye is dependent on the formation of double-stranded DNA, regardless of the DNA sequence, the Hybridization Probes format requires 2 specially designed, sequencespecific oligonucleotides. These Hybridization Probes, labeled with two different fluorescent molecules, hybridize next to each other on the target DNA molecule (Figure 1). The first fluorescent dye, the donor dye fluorescein, is excited at 470 nm by the light source of the LightCycler Instrument. Instead of emitting light at 530 nm, the fluorescein can transfer its energy in a nonfluorescent manner to an acceptor dye, for example LightCycler-Red 640 (LC-Red 640) or, alternatively, LightCycler-Red 705 (LC-Red 705), in a process called Fluorescence Resonance Energy Transfer (FRET). The acceptor or reporter dye emits light of a longer wavelength and the intensity of this signal can be correlated with the amount of target DNA molecules. Efficient FRET can only take place when the 2 dyes are in direct local proximity and when the emission spectrum of the donor dye overlaps the absorption spectrum of the reporter dye. In this article, we describe the simultaneous use of two reporter dyes, allowing the measurement of two independent target sequences. Fluorescent dyes for use with the LightCycler System The LightCycler Instrument s optical unit, a microvolume fluorimeter (Figure 2), is capable of measuring fluorescence in three separate channels simultaneously. The choice of band pass filters and dichroic mirrors for the optical unit is optimized to the fluorescent dyes that are used. Channel 1 measures fluorescence (F1) at Figure 1: Hybridization Probes format. 7
2 Figure 2: Microvolume fluorimeter of the LightCycler Instrument 530 nm and is the channel used for SYBR Green I and for fluorescein. Channel 2 (F2; 640 nm) is used to measure signals from LC-Red 640. Channel 3 (F3; 710 nm) is designed for use with LC-Red 705. The optical unit for the LightCycler System was developed with the objective of performing dual color detection. To achieve this objective, fluorescent dyes with optimized spectral separation had to be developed. The spectra of the dyes are shown in Figure 3. The spectral separation of excitation and emission spectra of the two FRET acceptor dyes (LC-Red 640 and LC-Red 705) is optimized to simultaneously: provide minimized spectral crosstalk from each dye into the other detection channels (e.g., a minimal fluorescence signal measured in channel 3 coming from LC-Red 640), provide adequate overlap of the excitation spectrum of the respective acceptor dyes with the fluorescein emission spectrum to allow efficient FRET. In meeting both objectives, a residual crosstalk of emitted light from LC-Red 640 into channel 3 cannot be avoided. Minor crosstalks from LC-Red 705 into channel 2 and from fluorescein into channels 2 and 3 occur as well. When measuring the two dyes simultaneously, correction of this crosstalk must be performed in order to obtain signals that can directly be correlated with the amount of hybridized probes and thus with the amount of the respective target. This color compensation is achieved by correcting the measured raw fluorescence values. The LightCycler-Color Compensation Set (Cat. No ) is used to determine the level of crosstalk by measuring the fluorescence of each individual dye in all detection channels, thereby providing the data for a so-called color compensation file. Factors influencing the relative crosstalk values The emitted fluorescence intensity and spectrum of a fluorescent dye is strongly temperature-dependent. Thus the temperature chosen for data acquisition (fluorescence measurement) during a PCR run influences the crosstalk significantly. This is particularly important for applications where measurements are performed at different temperatures (e.g., during melting curve analysis). During melting curve analysis, continuous measurements of fluorescence are made at slowly increasing temperatures. The measured fluorescence crosstalk is not only dependent on the temperature of data acquisition but also on the specific hardware components of the optical system of each specific instrument. Minor variations in the optical filters, lenses, and dichroic mirrors of each LightCycler Instrument s fluorimeter have an influence Figure 3: Fluorescence excitation and emission spectra of fluorescein, LightCycler-Red 640, and LightCycler-Red BIOCHEMICA No
3 Figure 4a: without color compensation Figure 4: Temperature-dependent color compensation. on crosstalk values. Therefore, it is necessary to generate and store an instrument-specific color compensation file before starting dual color applications on a LightCycler Instrument. It has been shown that the optimized LC reagents and kits used in the LightCycler PCR or RT-PCR systems do not display a relevant influence on relative crosstalk values. Furthermore, it is shown that different oligonucleotide sequences used for Hybridization Probes do not influence the crosstalk. The 5 coupling chemistry of the LC-Red 640-NHS-ester in combination with commonly used C6-Phosphoramidites (ABI, Glen Research), or the coupling chemistry with LC-Red 705-Phosphoramidite, results in a uniform structure of the linkers between the dye and Figure 4b: with color compensation oligonucleotide for each of the two fluorophores, and thus does not influence relative crosstalk values. In contrast, fluorescein can be coupled to the 3 end of an oligonucleotide using different linker chemistries. The most frequently used are thiourea linkers (FITCderived, for example fluorescein-cpg s from Glen Research or ChemGene) and amide-linkers (fluorescein-nhs-ester-derived, like fluorescein-cpg from BioGenex), or 3 -amino-cpg s that require coupling of fluorescein-nhs-ester after oligo synthesis. Thiourealinked fluorescein oligos display different relative crosstalk values compared to amide-linked fluorescein oligos. Therefore, both linker types are not compatible for quantitative crosstalk correction. For the LightCycler-Color Compensation Set, a fluorescein thiourea linker is used. As a result, it is Vial Label Contents and use 1 Blank standard buffer excluding fluorophores to determine the standard buffer fluorescence 2 Fluorescein Calibrator standard oligonucleotide, fluorescein-labeled at the 3' end [0.3 µm] to determine the crosstalk from fluorescein (channel 1 fluorophore), into channels 2 and 3 3 LightCycler-Red 640 Calibrator standard oligonucleotide, LC-Red 640-labeled at the 3' end [1 µm] to determine the crosstalk from LC-Red 640 (channel 2 fluorophore), into channels 1 and 3 4 LightCycler-Red 705 Calibrator standard oligonucleotide, LC-Red 705-labeled at the 3' end [1 µm] to determine the crosstalk from LC-Red 705 (channel 3 fluorophore), into channels 1 and 2 Table 1: Components of the LightCycler-Color Compensation Set 9
4 necessary that user-specific Hybridization Probes are also fluorescein-labeled with a thiourea-linker when performing dual color applications. When ordering Hybridization Probes from licensed oligonucleotide suppliers (e.g., GENSET, page 20), the fluoresceinlabeled Hybridization Probe will automatically be labeled using the thiourea-linker. Generation and use of a color compensation file (ccc file) An instrument-specific color compensation file can be created that subsequently can be used for color compensation of the raw data from different dual color experiments and applications. The color compensation file is generated using the LightCycler-Color Compensation Set. The set consists of three calibrators to determine the relative contribution of each individual fluorophore to the total fluorescence observed in the 3 detection channels (Table 1). This relative contribution does not depend on the FRET process and thus does not depend on the presence of a hybridization target. In addition, the set provides a standard buffer to determine the general influence of non-dye components on the crosstalk. The procedure to set up and save a color compensation file takes approximately 45 minutes, including the LightCycler Instrument calibration run, with only 5 minutes hands-on-time. All 4 components of the set Figure 5a: without color compensation Figure 5b: with color compensation Figure 5: Color compensation of a dual color duplex PCR experiment. Target sequences within the GAPDH-gene and the Cyclophilin A-gene were either amplified simultaneously [1: 10 7 copies each; 2: 10 6 copies each; 3: 10 5 copies each] or as a single target [4:10 9 copies GAPDH; 5: 10 8 copies Cyclophilin A]. GAPDH is detected with a fluorescein/lc-red 640 probe pair and Cyclophilin A with a fluorescein/lc-red 705 probe pair. Figure 5a: raw data from channels 2 and 3. Figure 5b: data after color compensation. 10 BIOCHEMICA No
5 are ready-to-use solutions. Each is pipetted into a single capillary and placed into the LightCycler Instrument. The experimental protocol consists of 4 program steps: heating step, cycling step, temperature gradient step (with continuous data acquisition), and cooling step. For detailed conditions, see the pack insert of the LightCycler-Color Compensation Set or the Light- Cycler Operator s Manual. Applications The color compensation file must be used for any application where two reporter dyes are used simultaneously. This includes: Use of internal controls (e.g., to exclude PCR inhibition). After the run, the data are stored as a color compensation file. Detection of more than one target sequence in a single sample. In subsequent dual color experiments, the color compensation file can either be used to compensate raw fluorescence data on-line during a LightCycler Instrument run, or after the run during data analysis. Detection of complex mutations. See the article Mutation Detection Using Multi-color Detection of the LightCycler System on page 12 of this Biochemica issue. Figure 4 illustrates the temperature-dependent contribution of all fluorophores to the fluorescence in each channel of the LightCycler Instrument. Figure 4a demonstrates the fluorescence raw data that are measured in each channel during the temperature gradient step while generating a color compensation file. Figure 4b shows the same data after color compensation using a previously stored color compensation file. It is obvious that after color compensation, each channel displays specifically the signal from the channel-specific fluorophore, whereas non-specific signals are compensated. Development formats of even more precise quantification using internal standards. Measurement of a known number of copies of a standard sequence, and simultaneous measurement of the copy number of a target sequence in the same capillary, guarantees identical PCR conditions for both amplicons. Appropriate software for the evaluation of the results of these types of experiments is under development. Figure 5 demonstrates the application of color compensation to a duplex PCR experiment using two Hybridization Probe pairs for the detection of amplicons from two different equimolar targets. Target 1 amplicons are detected using a fluorescein LC-Red 640-labeled probe pair and target 2 amplicons are detected using a fluorescein LC-Red 705-labeled probe pair. In addition to the duplex PCR, both targets are also amplified separately in different capillaries. Raw data and color compensated data are shown for channels 2 and 3 for all three capillaries. After color compensation, the contributions of the nonspecific dyes to the fluorescence signals are eliminated, as clearly demonstrated in the uniplex reactions. 11
Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation
Roche Applied Science Technical Note No. LC 19/2004 Purpose of this Note Color The LightCycler System is able to simultaneously detect and analyze more than one color in each capillary. Due to overlap
More informationTechnical Note. Roche Applied Science. No. LC 18/2004. Assay Formats for Use in Real-Time PCR
Roche Applied Science Technical Note No. LC 18/2004 Purpose of this Note Assay Formats for Use in Real-Time PCR The LightCycler Instrument uses several detection channels to monitor the amplification of
More informationEssentials of Real Time PCR. About Sequence Detection Chemistries
Essentials of Real Time PCR About Real-Time PCR Assays Real-time Polymerase Chain Reaction (PCR) is the ability to monitor the progress of the PCR as it occurs (i.e., in real time). Data is therefore collected
More informationCo Extra (GM and non GM supply chains: Their CO EXistence and TRAceability) Outcomes of Co Extra
GM and non GM supply chains: Their CO EXistence and TRAceability Outcomes of Co Extra Comparison of different real time PCR chemistries and their suitability for detection and quantification of genetically
More informationIntroduction To Real Time Quantitative PCR (qpcr)
Introduction To Real Time Quantitative PCR (qpcr) SABiosciences, A QIAGEN Company www.sabiosciences.com The Seminar Topics The advantages of qpcr versus conventional PCR Work flow & applications Factors
More informationRealStar HBV PCR Kit 1.0 11/2012
RealStar HBV PCR Kit 1.0 11/2012 RealStar HBV PCR Kit 1.0 For research use only! (RUO) Product No.: 201003 96 rxns INS-201000-GB-02 Store at -25 C... -15 C November 2012 altona Diagnostics GmbH Mörkenstraße
More informationReal-time PCR: Understanding C t
APPLICATION NOTE Real-Time PCR Real-time PCR: Understanding C t Real-time PCR, also called quantitative PCR or qpcr, can provide a simple and elegant method for determining the amount of a target sequence
More informationLightCycler 480 Real-Time PCR System
Roche Applied Science LightCycler 480 Real-Time PCR System Planned introduction of the LightCycler 480 System: September 2005 Rapid by nature, accurate by design The LightCycler 480 Real-Time PCR System
More informationCreating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II
Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II Due to ROX and Cy5 dye emmision spectra overlap Roche LightCycler 480 can pick up signals from a dye measured by another channel.
More informationReal-Time PCR Vs. Traditional PCR
Real-Time PCR Vs. Traditional PCR Description This tutorial will discuss the evolution of traditional PCR methods towards the use of Real-Time chemistry and instrumentation for accurate quantitation. Objectives
More informationMolecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
More informationQuando si parla di PCR quantitativa si intende:
Quando si parla di PCR quantitativa si intende: A. Una PCR che produce grandi quantità di DNA B. Una PCR che emette quanti di luce C. Una PCR che quantifica il numero di molecole stampo presenti all inizio
More informationHBV Quantitative Real Time PCR Kit
Revision No.: ZJ0002 Issue Date: Aug 7 th, 2008 HBV Quantitative Real Time PCR Kit Cat. No.: HD-0002-01 For Use with LightCycler 1.0/LightCycler2.0/LightCycler480 (Roche) Real Time PCR Systems (Pls ignore
More informationDNA Detection. Chapter 13
DNA Detection Chapter 13 Detecting DNA molecules Once you have your DNA separated by size Now you need to be able to visualize the DNA on the gel somehow Original techniques: Radioactive label, silver
More informationFactors Influencing Multiplex Real-Time PCR
APPLICATION NOTE Multiplex Real-Time PCR Factors Influencing Multiplex Real-Time PCR Introduction Multiplex PCR is the simultaneous amplification of more than one target sequence in a single reaction [1].
More information2.500 Threshold. 2.000 1000e - 001. Threshold. Exponential phase. Cycle Number
application note Real-Time PCR: Understanding C T Real-Time PCR: Understanding C T 4.500 3.500 1000e + 001 4.000 3.000 1000e + 000 3.500 2.500 Threshold 3.000 2.000 1000e - 001 Rn 2500 Rn 1500 Rn 2000
More informationHighly specific and sensitive quantitation
PRODUCT ULLETIN SYR Select Master Mix SYR Select Master Mix Highly specific and sensitive quantitation SYR Select Master Mix offers advanced performance at an affordable price. SYR Select Master Mix is
More informationValidating Microarray Data Using RT 2 Real-Time PCR Products
Validating Microarray Data Using RT 2 Real-Time PCR Products Introduction: Real-time PCR monitors the amount of amplicon as the reaction occurs. Usually, the amount of product is directly related to the
More informationSYBR Green Realtime PCR Master Mix -Plus-
Instruction manual SYBR Green Realtime PCR Master Mix -Plus- 0810 F0925K SYBR Green Realtime PCR Master Mix -Plus- Contents QPK-212T 1mLx1 QPK-212 1mLx5 Store at -20 C, protected from light [1] Introduction
More informationmircute mirna qpcr Detection Kit (SYBR Green)
mircute mirna qpcr Detection Kit (SYBR Green) For detection of mirna using real-time RT-PCR (SYBR Green I) www.tiangen.com QP110302 mircute mirna qpcr Detection Kit (SYBR Green) Kit Contents Cat. no. FP401
More informationReal time and Quantitative (RTAQ) PCR. so I have an outlier and I want to see if it really is changed
Real time and Quantitative (RTAQ) PCR or.. for this audience so I have an outlier and I want to see if it really is changed Nigel Walker, Ph.D. Laboratory of Computational Biology and Risk Analysis, Environmental
More informationStratagene QPCR Mouse Reference Total RNA
Stratagene QPCR Mouse Reference Total RNA Instruction Manual Catalog #750600 Revision C.0 For Research Use Only. Not for use in diagnostic procedures. 750600-12 LIMITED PRODUCT WARRANTY This warranty limits
More informationApplications Guide. Real-Time PCR Applications Guide
Applications Guide Real-Time PCR Applications Guide table of contents Table of Contents 1. Overview of Real-Time PCR 2 1.1 Key Concepts of Real-Time PCR 2 1.1.1 What Is Real-Time PCR? 2 1.1.2 How Real-Time
More informationQPCR Applications using Stratagene s Mx Real-Time PCR Platform
QPCR Applications using Stratagene s Mx Real-Time PCR Platform Dan Schoeffner, Ph.D Field Applications Scientist Dan.Schoeffner@Stratagene.com Tech. Services 800-894-1304 Polymerase Chain Reaction Melt
More informationamplification tech Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye
amplification tech note 6047 Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye Liz Jordan and Richard Kurtz, Gene Expression Division, io-rad
More informationReagent Guide. Applied Biosystems StepOne and StepOnePlus Real-Time PCR Systems
Reagent Guide Applied Biosystems StepOne and StepOnePlus Real-Time PCR Systems Applied Biosystems StepOne and StepOnePlus Real-Time PCR Systems Reagent Guide Copyright 2008, 2010 Applied Biosystems. All
More informationAbsolute Quantification Getting Started Guide
5 cdna Reverse Primer Oligo d(t) or random hexamer Synthesis of 1st cdna strand 3 5 cdna Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Absolute Quantification Getting Started Guide Introduction
More informationSmartFlare RNA Detection Probes: Principles, protocols and troubleshooting
Technical Guide SmartFlare RNA Detection Probes: Principles, protocols and troubleshooting Principles of SmartFlare technology RNA detection traditionally requires transfection, laborious sample prep,
More informationIntroduction to Quantitative PCR
Introduction to Quantitative PCR Methods and Applications Guide Introduction to Quantitative PCR Methods and Applications Guide IN 70200 D US and Canada Orders: 800-227-9770 x3 Technical Service: 800-227-9770
More informationquantitative real-time PCR, grain, simplex DNA extraction: PGS0426 RT-PCR: PGS0494 & PGS0476
BioScience quantitative real-time PCR, grain, simplex DNA extraction: PGS0426 RT-PCR: PGS0494 & PGS0476 This method describes a Real-time semi-quantitative TaqMan PCR procedure for the determination of
More informationReal-time PCR handbook
Real-time PCR handbook Single-tube assays 96- and 384-well plates 384-well TaqMan Array cards OpenArray plates The image on this cover is of an OpenArray plate which is primarily used for mid-density real-time
More informationProtocol. Introduction to TaqMan and SYBR Green Chemistries for Real-Time PCR
Protocol Introduction to TaqMan and SYBR Green Chemistries for Real-Time PCR Copyright 2008, 2010 Applied Biosystems. All rights reserved. Ambion and Applied Biosystems products are for Research Use Only.
More informationCOMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
More informationEco TM 48 Real Time PCR System. Fastest block-based qpcr. Four colour multiplexing. Supports all chemistries. Fast cycling protocols
A Bibby Scientific Company Eco TM 48 Real Time PCR System Fast cycling Fastest block-base HRM functionality Four colour multiplexing MIQE compliant Easy to use software Supports all chemistries Small footprint
More informationBeginner s Guide to Real-Time PCR
Beginner s Guide to Real-Time PCR 02 Real-time PCR basic principles PCR or the Polymerase Chain Reaction has become the cornerstone of modern molecular biology the world over. Real-time PCR is an advanced
More informationDNA Sequence Analysis
DNA Sequence Analysis Two general kinds of analysis Screen for one of a set of known sequences Determine the sequence even if it is novel Screening for a known sequence usually involves an oligonucleotide
More informationMethods and Application Guide. Introduction to Quantitative PCR
Methods and Application Guide Introduction to Quantitative PCR Introduction to Quantitative PCR Methods and Application Guide Stratagene USA and Canada Order: 800-424-5444 x3 Technical Services: 800-894-1304
More informationJBS FUNDAMENT Thermofluor Screen
Cat. No. CS-330 Amount 1 Kit For in vitro use only. Quality guaranteed for 3 months. Store at 4 C. Application Screen for thermal stability of proteins as a function of the FUNDAMENTAL variables ph and
More informationqpcr Technical Guide Detection Methods Primer and Probe Design Instrumentation Applications Guide
qpcr Technical Guide Detection Methods Primer and Probe Design Instrumentation Applications Guide Table of Contents qpcr Technical Guide Introduction... 1 Quantitative PCR: How does it work?... 2 qpcr
More informationThermo Scientific DyNAmo cdna Synthesis Kit for qrt-pcr Technical Manual
Thermo Scientific DyNAmo cdna Synthesis Kit for qrt-pcr Technical Manual F- 470S 20 cdna synthesis reactions (20 µl each) F- 470L 100 cdna synthesis reactions (20 µl each) Table of contents 1. Description...
More informationTaqMan Fast Advanced Master Mix. Protocol
TaqMan Fast Advanced Master Mix Protocol For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use. Information in this document is subject to change without notice. APPLIED
More informationSpeed Matters - Fast ways from template to result
qpcr Symposium 2007 - Weihenstephan Speed Matters - Fast ways from template to result March 28, 2007 Dr. Thorsten Traeger Senior Scientist, Research and Development - 1 - Overview Ạgenda Fast PCR The Challenges
More informationSEQUENCING. From Sample to Sequence-Ready
SEQUENCING From Sample to Sequence-Ready ACCESS ARRAY SYSTEM HIGH-QUALITY LIBRARIES, NOT ONCE, BUT EVERY TIME The highest-quality amplicons more sensitive, accurate, and specific Full support for all major
More informationArtisan Scientific is You~ Source for: Quality New and Certified-Used/Pre:-awned ECJuiflment
Looking for more information? Visit us on the web at http://www.artisan-scientific.com for more information: Price Quotations Drivers Technical Specifications. Manuals and Documentation Artisan Scientific
More informationGene Expression Assays
APPLICATION NOTE TaqMan Gene Expression Assays A mpl i fic ationef ficienc yof TaqMan Gene Expression Assays Assays tested extensively for qpcr efficiency Key factors that affect efficiency Efficiency
More informationPower SYBR Green PCR Master Mix and Power SYBR Green RT-PCR Reagents Kit
USER GUIDE Power SYBR Green PCR Master Mix and Power SYBR Green RT-PCR Reagents Kit Catalog Number 4368577, 4367659, 4367660, 4368706, 4368702, 4368708 (Master Mix) and 4368711 (RT-PCR Reagents Kit) Publication
More informationReal-time quantitative RT -PCR (Taqman)
Real-time quantitative RT -PCR (Taqman) Author: SC, Patti Lab, 3/03 This is performed as a 2-step reaction: 1. cdna synthesis from DNase 1-treated total RNA 2. PCR 1. cdna synthesis (Advantage RT-for-PCR
More informationExperiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
More informationOptimizing and Analyzing Real-Time Assays on the SmartCycler II System
Optimizing and Analyzing Real-Time Assays on the SmartCycler II System Cepheid Technical Support Overview This technical note provides guidelines on transferring, optimizing, and evaluating real-time PCR
More informationRoche Applied Science Technical Note No. LC 13/2001
Roche Applied Science Technical Note No. LC 13/2001 LightCycler Relative Quantification Purpose of this Note The LightCycler provides great flexibility especially to the user interested in quantitative
More informationPitfalls in qpcr. Primer and probe design and synthesis. Mary Span Eurogentec S.A.
Pitfalls in qpcr Primer and probe design and synthesis Mary Span Eurogentec S.A. Steps in qpcr assay Set up experiment statistical relevant # samples/experimental group controls Design and synthesis primers
More informationSYBR Green PCR Master Mix and SYBR Green RT-PCR Reagents Kit
USER GUIDE SYBR Green PCR Master Mix and SYBR Green RT-PCR Reagents Kit Catalog Number 4309155 (Master Mix) and 4306736 (RT-PCR Reagents Kit) Publication Part Number 4310251 Rev. G Revision Date September
More informationQuantiFast Multiplex PCR Handbook
May 2011 QuantiFast Multiplex PCR Handbook QuantiFast Multiplex PCR Kit with master mix containing ROX passive reference dye QuantiFast Multiplex PCR +R Kit with separate tube of ROX passive reference
More informationPATHOGEN DETECTION SYSTEMS BY REAL TIME PCR. Results Interpretation Guide
PATHOGEN DETECTION SYSTEMS BY REAL TIME PCR Results Interpretation Guide Pathogen Detection Systems by Real Time PCR Microbial offers real time PCR based systems for the detection of pathogenic bacteria
More informationReal-time qpcr Assay Design Software www.qpcrdesign.com
Real-time qpcr Assay Design Software www.qpcrdesign.com Your Blueprint For Success Informational Guide 2199 South McDowell Blvd Petaluma, CA 94954-6904 USA 1.800.GENOME.1(436.6631) 1.415.883.8400 1.415.883.8488
More informationQuantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit
Product Bulletin Human Identification Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit The Quantifiler kits produce reliable and reproducible results, helping to
More informationPlexor Systems Instrument Setup and Data Analysis for the Roche LightCycler 2.0 and LightCycler 1.5 Systems using LightCycler Software Version 4.
TECHNICAL MANUAL Plexor Systems Instrument Setup and Data Analysis for the Roche LightCycler 2.0 and LightCycler 1.5 Systems using LightCycler Software Version 4.0 Instruc ons for use of Products A4011,
More informationPlexor Systems Instrument Setup and Data Analysis for the Applied Biosystems 7300 and 7500 Real-Time PCR Systems
Technical Manual Plexor Systems Instrument Setup and Data Analysis for the Applied Biosystems 7300 and 7500 Real-Time PCR Systems INSTRUCTIONS FOR USE OF PRODUCTS A4011, A4021, A4031, A4041, A4051 AND
More informationADVANCES IN REAL TIME PCR: APPLICATION TO CLINICAL LABORATORY DIAGNOSTICS
ADVANCES IN CLINICAL CHEMISTRY, VOL. 40 ADVANCES IN REAL TIME PCR: APPLICATION TO CLINICAL LABORATORY DIAGNOSTICS Bernhard Kaltenboeck and Chengming Wang Department of Pathobiology, College of Veterinary
More informationGetting Started Guide
Primer Express Software Version 3.0 Getting Started Guide Before You Begin Designing Primers and Probes for Quantification Assays Designing Primers and Probes for Allelic Discrimination Assays Ordering
More informationDyNAmo cdna Synthesis Kit for qrt-pcr
DyNAmo cdna Synthesis Kit for qrt-pcr Instruction manual F- 470S Sufficient for 20 cdna synthesis reactions (20 µl each) F- 470L Sufficient for 100 cdna synthesis reactions (20 µl each) Description...
More informationDNA Integrity Number (DIN) For the Assessment of Genomic DNA Samples in Real-Time Quantitative PCR (qpcr) Experiments
DNA Integrity Number () For the Assessment of Genomic DNA Samples in Real-Time Quantitative PCR (qpcr) Experiments Application Note Nucleic Acid Analysis Author Arunkumar Padmanaban Agilent Technologies,
More informationApplication Note. Biotechnology Explorer Crime Scene Investigator PCR Basics. Kit: A Real-Time PCR Extension
Biotechnology Explorer Crime Scene Investigator PCR Basics Kit: Table of Contents Introduction.............................................. 2 Learning Objectives......................................
More informationIntroduction. Preparation of Template DNA
Procedures and Recommendations for DNA Sequencing at the Plant-Microbe Genomics Facility Ohio State University Biological Sciences Building Room 420, 484 W. 12th Ave., Columbus OH 43210 Telephone: 614/247-6204;
More informationIntroduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
More informationChapter 12 Filters for FISH Imaging
Chapter 12 Filters for FISH Imaging Dan Osborn The application of in situ hybridization (ISH) has advanced from short lived, non-specific isotopic methods, to very specific, long lived, multiple color
More informationVLLM0421c Medical Microbiology I, practical sessions. Protocol to topic J10
Topic J10+11: Molecular-biological methods + Clinical virology I (hepatitis A, B & C, HIV) To study: PCR, ELISA, your own notes from serology reactions Task J10/1: DNA isolation of the etiological agent
More informationReal-time monitoring of rolling circle amplification using aggregation-induced emission: applications for biological detection
Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 215 Supplementary Information Real-time monitoring of rolling circle amplification using aggregation-induced
More informationMystiCq microrna cdna Synthesis Mix Catalog Number MIRRT Storage Temperature 20 C
microrna cdna Synthesis Mix Catalog Number MIRRT Storage Temperature 20 C Product Description The microrna cdna Synthesis Mix has been designed to easily convert micrornas into cdna templates for qpcr
More informationqstar mirna qpcr Detection System
qstar mirna qpcr Detection System Table of Contents Table of Contents...1 Package Contents and Storage Conditions...2 For mirna cdna synthesis kit...2 For qstar mirna primer pairs...2 For qstar mirna qpcr
More informationLightCycler DNA Master SYBR Green I
For general laboratory use. Not for use in diagnostic procedures. FOR IN VITRO USE ONLY. LightCycler DNA Master SYBR Green I Version April 2008 Easy-to-use Reaction Mix for PCR using the LightCycler Carousel-Based
More informationQuantitative Real Time PCR Protocol. Stack Lab
Quantitative Real Time PCR Protocol Stack Lab Overview Real-time quantitative polymerase chain reaction (qpcr) differs from regular PCR by including in the reaction fluorescent reporter molecules that
More informationTable of Contents. I. Description... 2. II. Kit Components... 2. III. Storage... 2. IV. 1st Strand cdna Synthesis Reaction... 3
Table of Contents I. Description... 2 II. Kit Components... 2 III. Storage... 2 IV. 1st Strand cdna Synthesis Reaction... 3 V. RT-PCR, Real-time RT-PCR... 4 VI. Application... 5 VII. Preparation of RNA
More informationA closed tube format for amplification and detection of DNA based on energy transfer
2516 2521 Nucleic Acids Research, 1997, Vol. 25, No. 12 1997 Oxford University Press A closed tube format for amplification and detection of DNA based on energy transfer I. A. Nazarenko*, S. K. Bhatnagar
More informationReal-Time PCR UNIT 10.3 OVERVIEW AND PRINCIPLES
UNIT.3 Real-Time PCR Dean Fraga, 1 Tea Meulia, 2 and Steven Fenster 3 1 College of Wooster, Wooster, Ohio 2 Ohio Agricultural Research and Development Center, Wooster, Ohio 3 Ashland University, Ashland,
More informationREAL TIME PCR USING SYBR GREEN
REAL TIME PCR USING SYBR GREEN 1 THE PROBLEM NEED TO QUANTITATE DIFFERENCES IN mrna EXPRESSION SMALL AMOUNTS OF mrna LASER CAPTURE SMALL AMOUNTS OF TISSUE PRIMARY CELLS PRECIOUS REAGENTS 2 THE PROBLEM
More informationFluorescence and Fluorescence Applications
Fluorescence and Fluorescence Applications Contents 1. Introduction... 1 2. Principles of Fluorescence... 1 3. Fluorescence Spectra... 3 4. Fluorescence Resonance Energy Transfer (FRET)... 5 5. Dark Quenchers...
More informationApplication Note. Single Cell PCR Preparation
Application Note Single Cell PCR Preparation From Automated Screening to the Molecular Analysis of Single Cells The AmpliGrid system is a highly sensitive tool for the analysis of single cells. In combination
More informationFRET Basics and Applications an EAMNET teaching module
FRET Basics and Applications an EAMNET teaching module Timo Zimmermann + Stefan Terjung Advanced Light Microscopy Facility European Molecular Biology Laboratory, Heidelberg http://www.embl.de/almf/ http://www.embl.de/eamnet/
More informationRelative Quantification Applied Biosystems 7300/7500 Real Time PCR System
5 3 5 cdna Reverse Primer Oligo d(t) or random hexamer 3 5 cdna Getting Started Guide Relative Quantification Applied Biosystems 7300/7500 Real Time PCR System Introduction and Example RQ Experiment Designing
More informationTroubleshooting Sequencing Data
Troubleshooting Sequencing Data Troubleshooting Sequencing Data No recognizable sequence (see page 7-10) Insufficient Quantitate the DNA. Increase the amount of DNA in the sequencing reactions. See page
More informationOptimizing Performance of the Transcreener ADP Assay for the BioTek Synergy 2 and 4 Multi-Mode Microplate Readers
Optimizing Performance of the Transcreener ADP Assay for the BioTek Synergy 2 and 4 Multi-Mode Microplate Readers Brad Larson 1, Karen Kleman-Leyer 1, Xavier Amouretti 2 1 BellBrook Labs, Madison, WI,
More informationSupporting Information. The G-triplex DNA could function as a new variety of DNA peroxidase
Supporting Information The G-triplex DNA could function as a new variety of DNA peroxidase Shaoru Wang +, Boshi Fu +, Shuang Peng, Xiaoe Zhang, Tian Tian*, Xiang Zhou* College of Chemistry and Molecular
More informationFast SYBR Green Master Mix. Protocol
Fast SYBR Green Master Mix Protocol Copyright 2007, 2010 Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. Information in this document is subject to
More informationUniversidade Estadual de Maringá
Universidade Estadual de Maringá Disciplina: Biologia Molecular Sequenciamento de ácidos nucléicos Profa. Dra. Maria Aparecida Fernandez Maxan e Gilbert - quebra química Berg, Gilbert and Sanger dideoxinucleotideos
More informationProcedures For DNA Sequencing
Procedures For DNA Sequencing Plant-Microbe Genomics Facility (PMGF) Ohio State University 420 Biological Sciences Building 484 W. 12th Ave., Columbus OH 43210 Telephone: 614/247-6204 FAX: 614/292-6337
More informationAutomated DNA Sequencing. Chemistry Guide
Automated DNA Sequencing Chemistry Guide Copyright 2000, Applied Biosystems For Research Use Only. Not for use in diagnostic procedures. ABI PRISM and its design, Applied Biosystems, and MicroAmp are registered
More informationA pretty picture, or a measurement? Retinal Imaging
Big Data Challenges A pretty picture, or a measurement? Organelles Dynamics Cells Retinal Imaging Physiology Pathology Fundus Camera Optical coherence tomography Fluorescence Histology High Content Screening
More informationAPPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
More informationOligonucleotide Stability Study
Oligonucleotide Stability Study IDT is performing a long term study of oligo stability. The study has been ongoing for 2 years and has tested oligo stability under various conditions. An overview of the
More informationFluorChem Q. A Quantitative Western Blot Imaging Solution
FluorChem Q A Quantitative Western Blot Imaging Solution Contents FluorChem Q Innovative Design The FluorChem Q is a total solution for quantitative fluorescent and chemiluminescent Western blot imaging
More informationBacReady TM Multiplex PCR System
BacReady TM Multiplex PCR System Technical Manual No. 0191 Version 10112010 I Description.. 1 II Applications 2 III Key Features.. 2 IV Shipping and Storage. 2 V Simplified Procedures. 2 VI Detailed Experimental
More informationReal-Time PCR UNIT 10.3 OVERVIEW AND PRINCIPLES
UNIT.3 Dean Fraga, 1 Tea Meulia, 2 and Steven Fenster 3 1 College of Wooster, Wooster, Ohio 2 Ohio Agricultural Research and Development Center, Wooster, Ohio 3 Ashland University, Ashland, Ohio OVERVIEW
More informationQuantitative PCR Systems
QUANTITATIVE PCR Quantitative PCR Systems Our QPCR systems meet your current research needs, offer flexibility for future research goals, and match your budget. HIGH-PERFORMANCE FULL-FEATURED FLEXIBLE
More informationEpstein Barr Virus (Human Herpes virus 4) genesig Standard Kit. DNA testing. Everything... Everyone... Everywhere...
TM Primerdesign Ltd TM Primerdesign Ltd Epstein Barr Virus (Human Herpes virus 4) nonglycosylated membrane protein (BNRF1) gene genesig Standard Kit 150 tests DNA testing Everything... Everyone... Everywhere...
More informationHiPer RT-PCR Teaching Kit
HiPer RT-PCR Teaching Kit Product Code: HTBM024 Number of experiments that can be performed: 5 Duration of Experiment: Protocol: 4 hours Agarose Gel Electrophoresis: 45 minutes Storage Instructions: The
More informationMEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:
INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals
More informationKatharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
More informationScott Reierstad. Field Applications Scientist
Real-time PCR Training Scott Reierstad Field Applications Scientist Topics we ll cover Quick introduction to real-time PCR ABI-supported real-time chemistries Instrumentation Quantitation via Standard
More informationIdaho Technology Food Security Systems. System Components. Idaho Technology. Food Security and Pathogen Detection
Incorporated in 1990, has years of experience developing exciting instruments, software at Inc. Part of s Incorporated in 1990, has years of experience developing exciting instruments, software at Inc.
More information