Enhanced Diode Array Detector Sensitivity and Automated Peak Purity Control

Similar documents
Fundamentals of modern UV-visible spectroscopy. Presentation Materials

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Thermo Scientific Dionex UltiMate 3000 Variable Wavelength Detectors

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Uses of Derivative Spectroscopy

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Agilent 1100/1200 HPLC ChemStation Operation

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Reversed phase C-18 chromatography: From analytical scale to preparative scale

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis

Using the Spectrophotometer

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

instruments Analytical Instruments for Science

Alignment and Preprocessing for Data Analysis

MCAL Spectrophotometry. Spectrophotometry

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

The Theory of HPLC. Gradient HPLC

2 Spectrophotometry and the Analysis of Riboflavin

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Turning HTE Agilent LC/MS Data to Knowledge in 1hr

Guidance for Industry

INTRODUCTION. The principles of which are to:

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ChemStation E-Seminars

PDA Software Getting Started Guide

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Appendix 5 Overview of requirements in English

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

Chapter 28: High-Performance Liquid Chromatography (HPLC)

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

HPLC Column Comparison Screening Study for Reversed Phase Columns

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

A Complete Solution for Method Linearity in HPLC and UHPLC

Agilent Cary 60 UV-Vis

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps

Background Information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

SPECTROSCOPY. Light interacting with matter as an analytical tool

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Determination of Caffeine in Beverages HPLC-1

Chemical analysis service, Turner s Green Technology Group

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

ChromQuest 5.0 Chromatography Data System

Pesticide Analysis by Mass Spectrometry

2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY

ß-CYCLODEXTRIN SYNONYMS

Quick Start Guide Chromeleon 7.2

Waters Core Chromatography Training (2 Days)

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

MassHunter for Agilent GC/MS & GC/MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Experiment #5: Qualitative Absorption Spectroscopy

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Introduction to Fourier Transform Infrared Spectrometry

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Shimadzu UV-VIS User s Guide

Smartline HPLC by design

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

Expectations for GC-MS Lab

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

Analysis of Phthalate Esters in Children's Toys Using GC-MS

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Research for leachables of nuclear grade cation exchange resin

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION

Analytical Test Method Validation Report Template

Spectroscopy Using the Tracker Video Analysis Program

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Empower PDA Software. Getting Started Guide. 34 Maple Street Milford, MA , Revision A

AMD Analysis & Technology AG

Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )

23 The Thermal Conductivity Detector

Transcription:

Enhanced Diode Array Detector Sensitivity and Automated Peak Purity Control Technical Note Introduction The most widely used detection technique for HPLC analysis is UV absorption. Over the decades, single and multi-wavelength detectors have been optimized for use as highly sensitive devices for quantitative analysis at specific wavelengths, however one drawback of these detectors is that they only provide limited information for compound identification. The introduction of the diode array detector helped overcome this problem by providing access to quantitative as well as qualitative spectral information for a wide wavelength range. Agilent Technologie introduced the first diode array detector for liquid chromatography in 1982. Since that time significant improvements in the design of both the hardware and software have been made, resulting in the introduction of the Agilent 11 Series diode array detector (DAD). A schematic view of the optical design of this detector is given in figure 1. In order to obtain optimum lamp energy in the UV and visible ranges, a tungsten lamp is positioned in line with a deuterium lamp. By using a software-programmable slit combined with the 124 element diode array, the optimum between wavelength resolution and sensitivity is easily achieved. This Technical Note demonstrates the outstanding performance of the Agilent 11 Series DAD as a very sensitive UV-Visible detector and the capabilities of the instrument for peak-purity analysis and peak confirmation of trace level. Agilent Technologies Innovating the HP Way

Results and discussion High sensitivity UV-detection The most important criterion for performance measurement in UV detection is a high and reproducible signal-to-noise ratio for the compound of interest. In addition, detector drift and wander over the whole wavelength range have to be monitored. Baseline stability results typical for the Agilent 11 Series DAD are shown in Figure 2, where baseline noise in the UV range is below 1 1-5 AU. The lower part of the figure illustrates the decrease in baseline noise that is achieved by having a tungsten lamp to cover the visible wavelength range. This allows for high sensitivity measurement when combined with the 13 µl standard flow cell (1 mm cell path length.) Tungsten lamp Deuterium lamp Lens Cell Holmium oxide filter 19 nm Programmable slit Figure 1 The optical design of the Agilent 11 Series diode array detector 95 nm 124 element diode array Grating.15.1.5 -.5 -.1 -.15 -.2 Noise: 1 x 1-5 254 nm Conditions Column Mobile phase Flow rate Slit Hypersil BDS, 3 µm, 125 3 mm Water/Acetonitrile 2/8 isocratic.2 ml/min 1 nm 1 2 3 4 5 6 7 8 9 Without tungsten lamp.6.4.2 -.2 7 nm D2 + tungsten lamp 2 4 6 8 Figure 2 Baseline stability with the Agilent 11 Series diode array detector The upper part of this figure illustrates the typical noise in the UV range. The lower figure shows the difference in noise for operation with and without the tungsten lamp in the visible range.

In order to optimize signal/noise each application requires a thorough optimization of DAD parameters at trace level. The graphical user interface of the Agilent 11 Series DAD gives easy access for bandwidth, slit width, response time and reference wavelength. Figure 3 shows the results of analysis of anthracene at the 1 pg level at 251 nm with a signal/noise ratio of 23 after optimization. Multiwavelength analysis For selective quantification of known compounds multiwavelength detection is better. This requires that the substances show absorption at substantially separated wavelengths. Since diode array detectors can collect data over the whole range of wavelengths, they are ideally suited for use as multiwavelength detectors. Figure 4 shows an example from food analysis. The green appearance of lemonade is created by adding a yellow and blue color. To ensure compliance with regulatory requirements the colors are identified by simultaneous monitoring at non-interfering wavelengths and quantified according to the appropriate directives. 1.7 1.6 Anthracene Signal :.23 mau Baseline noise : -5 1 x 1 AU Conditions Sample Column Mobile phase Flow rate Wavelength Anthracene standard, 1 pg Hypersil ODS, 3 µm, 1 2.1 mm Water/Acetonitrile 3/7 isocratic.4 ml/min 251 nm 1.5 1.4 2 3 Figure 3 UV-signal sensitivity in environmental trace analysis Chinolin yellow Patent blue The peak for patent blue reflects a concentration of 7.45 ng dye 12 1 8 6 4 465 nm/3 nm Conditions Column Hypersil BDS, 3 µm, 125 3 mm 4 ºC Mobile phase A:.1 Disodiumhydrogenphosphate + Tetrabutylammonium dihydrogenphosphate in water, ph 4.2, B: Acetonitrile Flow rate Gradient.8 ml/min 15%B to 4%B in 1 min, to 9%B in 19 min 2 6 nm/4 nm 2 4 6 8 1 12 14 Figure 4 Analysis of colors from Woodruff lemonade by multi-wavelength detection

Extended wavelength detection A variety of chemicals and natural products are available for use as colors and dyes to address current interests in fashioned products. In order to prevent health damage arising from this practice an increasing range of colors are regulated with national directives. The Agilent 11 Series DAD is well suited for monitoring in the visible wavelength range. The addition of a tungsten lamp leads to superior baseline stability (figure 2) and more complete spectra (figure 5). Peak identification at trace level Retention time and absorption at a single wavelength is not sufficient to prove the identity of compounds in environmental analysis. The complexity of real life samples are such that additional information is needed to confirm peak identity. With a diode array detector complete spectral information is available from any time and wavelength during the run. One common practice is to compare the obtained spectra from a peak with a library or to check for peak-purity by comparison of spectra taken throughout a single peak. With the Agilent 11 Series DAD both techniques can be applied automatically as part of a method and therefore run routinely for a sequence of samples, for example overnight. Spectra can either be collected throughout the analytical run or peak controlled to save memory space. Library search Figure 6 shows an example for ultimate spectral sensitivity. A comparison of sample results with library spectra confirms the peak identity. At the limit of detection for pyrene, at pg level, the spectrum taken during the run matches with the library spectrum recorded at high concentration. With the availability of pure standard compounds, customized libraries can be established by the individual user for a particular application. The Agilent ChemStation software calculates the correlation between the library and the experimental spectra and reports match factors. A match factor of 1 (correlation factor = 1) describes identical spectra. Tartrazine Amaranth Patent blue Brilliant blue DAD with limited wavelength range Agilent 11 Series DAD 3 4 5 6 7 8 9 Figure 5 The Agilent 11 Series DAD gives more complete spectra.12 mau FS.17 mau FS *Pyrene 24 26 28 3 32 34 36 nm Wavelength [nm] # Match Entry Time Name 1 914.677 5 2.85 Pyrene 2 572.4 8 3.27 Methylpyrene -5 Peak height = 3 x 1 AU -5 Noise = 1 x 1 AU 1 2 3 4 5 Figure 6 Spectral identification by library search at trace level concentrations

Peak-purity analysis Normalized spectra from a pure peak will not differ whether they are recorded at the apex, at the beginning or at the end of a peak. Impurities with different spectra and comparable spectral absorption can be monitored as soon as some chromatographic resolution of the compounds is achieved. The Agilent ChemStation provides several tools for the calculation and judgement of peak-purity. For notification purposes a threshold level can be set to inform the user about a certain level of spectral impurity. The similarity curve displays the match factors for all spectra in a peak, providing more detailed information and higher sensitivity. In addition, a threshold curve can be calculated to show the effect of noise on the spectral similarity. This gives the range for which spectral impurities are within the noise limit. Both curves can be combined into one graph the peak-purity curve. The following examples demonstrate how the Agilent 11 Series DAD is able to identify the presence of even minor impurities within a chromatographic peak. These are different with respect to chromatographic resolution. The absorption bands of the two pairs show such overlay that there is no wavelength for sensitive quantification of the minor compound away from the major ingredient (figures 8 and 9). Example 1: Phenacetin and salicylic acid In the case of phenacetin and salicylic acid, chromatographic separation is nearly achieved. The R =.1 12 1 8 6 4 2 1 6 2 7 3 4 identification of phenacetin as the impurity is done by analyzing peak-controlled spectra (upslope, apex and downslope.) Acquisition of peak-controlled spectra considerably reduces the requirements for data storage. With this mode, the threshold alarm is displayed if one match factor exceeds the threshold value. R =.9 1 Paracetamol 2 Hydrochlorothiazide 3 Accetylsalicylic acid 4 Phenacetin 5 Carbamazepine 6 Caffeine 7 Salicylic acid 1 2 3 4 5 6 7 8 9 Figure 7 Chromatograms of two analgesics mix 254 nm Conditions Column Mobile phase Flow rate 5 Hypersil BDS, 3 µm, 125 3 mm Water with sulfuric acid to ph 3/Acentonitrile 6/4 isocratic.25 ml/min Mix 1 Mix 2 Two mixtures with common ingredients to analgesic formulations were analyzed under the same conditions. Concentrations of the analytes varied. The overlaid chromatograms in figure 7 highlight the limited resolution of phenacetin and salicylic acid (R =.9) and caffeine and paracetamol (R =.1). 1 8 6 4 2 Phenacetin 32nm Salicylic acid 24 26 28 3 32 34 36 38 Wavelength [nm] Figure 8 UV spectra for phenacetin and salicylic acid

Figure 1 shows the results for different levels of phenacetin as the impurity. It can be seen that phenacetin was detected as the spectral impurity down to the.5 % level. The peak spectra window gives the overlay of spectra taken. The peak-purity window offers a view as used in multiwavelength detection. 8 6 4 Paracetamol Caffeine 254 nm 272 nm Example 2: Caffeine and paracetamol Figure 9 shows that caffeine and paracetamol should be monitored at 254 nm and 272 nm. However, at both wavelengths there is a UV absorption of the coeluting compound in the chromatogram (figure 7). In this case, analysis of peak-controlled spectra is only sufficient to detect an impurity level of 1 % paracetamol. However with the full set of spectra available, calculation of the similarity and threshold curves can be done. Both are displayed as separate curves with the ratio of similarity and threshold values (figure 11). Thus the similarity of spectra is elucidated taking into account the influence of noise, especially at the start and end of a peak. With the example of caffeine and paracetamol, as little of.1 % of paracetamol could be detected as spectral impurity. 2 2 22 24 26 28 3 32 34 Wavelength [nm] Figure 9 UV spectra for caffeine and paracetamol 5 % Phenacetin.5 % Phenacetin.1 % Phenacetin Figure 1 Peak-purity analysis with peak-controlled spectra, phenacetin as the impurity, coeluting with salicylic acid

Summary The Agilent 11 Series DAD gives excellent sensitivity for basic single and multiwavelength detection as well as for spectral identification of compounds. Even at trace levels the software on the Agilent ChemStation allows fast and accurate comparison of spectra with customer contributed libraries. Library searches as well as peak-purity analysis can easily be executed as part of a method. Additionally, the algorithms implemented in the peak-purity analysis can detect impurities down to the.1 % level. 1 % 1 % 1 %.5 %.1 % Similarity* Threshold* High Low } Purity Figure 11 Peak-purity analysis for caffeine with different concentrations of paracetamol. From full spectral information detection of impurities at the.1 % level is possible *) inverse scale

For the latest information and services visit our world wide web site: http://www.agilent.com/chem Copyright 1997 Agilent Technologies All Rights Reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed under the copyright laws. Publication Number 5965-59E Agilent Technologies Innovating the HP Way