1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

Similar documents
ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System

REVEAL LINQ LNQ11. Insertable Cardiac Monitor MRI procedural information. MRI Technical Manual

FDA Guidelines for Magnetic Resonance Equipment Safety

Boston Scientific Spinal Cord Stimulation (SCS) Systems

ES-3305P V2 / ES-3308P V2. Quick Installation Guide / v1.0

IMAGEREADY MR CONDITIONAL DEFIBRILLATION SYSTEM

Numerical Modelling of E-M Occupational Exposures associated with MRI

MRI Case Study MRI Cervical Spine

MRI SAFETY JEOPARDY (Technologist Edition) Questions

Regulatory Compliance Statement

User Guide Microsoft Screen Sharing for Lumia Phones (HD-10)

Subject: Magnetic Resonance Imaging (MRI) Information on Sorin Group Heart Valve Prostheses and Annuloplasty Devices


MR Information Date: March 11, 2014

User Guide Microsoft Screen Sharing for Lumia Phones (HD-10)

Basic Principles of Magnetic Resonance

REGULATORY IMPACT ANALYSIS (RIA) SAFETY, HEALTH AND WELFARE AT WORK. (Electromagnetic Fields) REGULATIONS (S.I. No. of 2016) 5 th January 2015

Millennium Product Inc. Model: Cell Shield / Zorb

Magnetic Resonance Imaging

EXHIBIT C BOSTON SCIENTIFIC PRODUCTS AVAILABLE FOR NIH BRAIN INITIATIVE Page 1 of 9

LifeStent XL Biliary Stent System

2012 Operating Room Design Manual

MRI Safety at 3T versus 1.5T

Introduction. Setting Up Your Remote Installing the Battery. Contents. System Requirements

Patient Information. PORT-A-CATH Implantable Venous Access Systems

Basestation with icom. User Guide

5.8GHz Audio/Video Sender

Magnetic Resonance Imaging

RF EXPOSURE LIMITS AND TESTING REQUIREMENTS

A PATIENT S GUIDE TO ABLATION THERAPY

ARIB TR-T V Specific Absorption Rate (SAR) requirements and regulations in different regions

Introduction. Setting Up Your Remote Installing the Battery. Contents. System Requirements

Invited. Pre-MRI Procedure Screening: Recommendations and Safety Considerations for Biomedical Implants and Devices

Living with Your Pacemaker

User Guide. Cordless Optical Mouse N2953

Working safely around Radiofrequency (RF) Transmitters

THE ANSI/IEEE RF SAFETY STANDARD AND ITS RATIONALE. Om P. Gandhi and Gianluca Lazzi Department of Electrical Engineering Salt Lake City, Utah 84112

Magnetic Resonance Imaging

Corporate Medical Policy

.org. Arthritis of the Hand. Description

University Children s Hospital Basel

Limited Warranty. CAUTION: Federal law restricts this device to sale, distribution and use by or on the order of a physician.

TomTom Hands-Free Car Kit Reference Guide

MRI SAFETY JEOPARDY (NON Technologist Edition) Questions

COUNCIL RECOMMENDATION of 12 July 1999 on the limitation of exposure of the general public to electromagnetic fields (0 Hz to 300 GHz) (1999/519/EC)

Official Journal of the European Communities. (Acts whose publication is not obligatory) COUNCIL

ASUS Tablet. User Guide

Participant Invitation and Information Sheet MRI Test Run

NFC Communication Tray Model: HHX-IT3-Z

Cook Celect and Günther Tulip Vena Cava Filter Patient Guide

Clean, Green and Safe Management System

Neurostimulation: Orthopaedic Institute of Ohio 801 Medical Drive Lima, Ohio

5 Factors Affecting the Signal-to-Noise Ratio

User Guide. Soft-Touch TM Bluetooth Laser Mouse N2953

YOUR MRI EXAM: QUESTIONS AND ANSWERS

Mini-led spotlight with magnetic base

Instructions for Use. Device Description The AMPLATZER Vascular Plug II is a self-expandable nitinol mesh occlusion device (see Figure 1).

Spinal Cord Stimulation (SCS) Therapy: Fact Sheet

GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G USB + WI-FI

USB 2.0 to 10/100Mbps Ethernet Adapter UE User Manual

Wireless Mouse USER GUIDE. for Mac. ONE YEAR LIMITED WARRANTY N2953

Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use

TG-3468 Gigabit PCI Express Network Adapter

Chronic Low Back Pain

THE WORLD S PREMIER. State of the Art OVER-GROUND GAIT & SAFETY SYSTEM. Transforming Rehabilitation

Safeguarding public health The Regulation of Software as a Medical Device

PM1109 Mondo QUICK GUIDE

IEC TC106. Standards for the Assessment of Human Exposure to Electric, Magnetic, and Electromagnetic Fields, 0 to 300 GHz

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager

FFD 2.5" IDE Plus Flash Disk

Magnetic Resonance Imaging Environment Safety in Ontario

zseries 18-Slot Chassis 18-Slot 3U PXI Express Chassis with AC Up to 8 GB/s

GETTING TO KNOW YOUR PRE-PAID TELSTRA WI-FI 4G

Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS

Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands)

UC Davis Imaging Research Center. MRI Online Safety

ZTE WIRELESS HOME PHONE BASE USER GUIDE

MRI Department Goals. Effective Ways to Improve Patient Cooperation and Safety in MRI Exam. Lecture Topics

RADIOFREQUENCY RADIATION, (RFR): (RFR Information - Technology Newsletter, Full Version)

Achat 115MA full-range speaker. user manual

MRI SureScan Pacing System

Spinal cord stimulation

safe, smart, protected

Bluetooth Stereo Headphone. User Guide. Hive

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

TV Ears 5.0 Digital & Dual Digital Quick Start Guide

Laser Treatment Policy

User Guide Nokia Portable Wireless Charging Plate DC-50

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a

Prepublication Requirements

How To Use A U.S. Cell Phone At Home

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.

Intensity-Modulated Radiation Therapy (IMRT)

WLAN600 Wireless IP Phone Administrator s Guide

Procedures/risks: Radiology (CT, DXA, MRI, ultrasound, X-ray)

X-ray (Radiography) - Bone

A Primer on Medical Device Interactions with Magnetic Resonance Imaging Systems

SERVICES BROCHURE info@t-r-s.com.au PO BOX 680, Claremont, 6910, Western Australia

NeuroStar TMS Therapy Patient Guide for Treating Depression

Transcription:

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro Corp. 1800 Bridge Parkway Redwood City, CA 94065, USA Tel: +1.650.251.0005 Fax: +1.650.251.9415 info@nevrocorp.com MDSS GMBH Schiffgraben 41 D-30175 Hannover, Germany Australian Sponsor Emergo Asia Pacific Pty Ltd 201 Sussex Street, Darling Park, Tower II, Level 20 Sydney, NSW 2000 Australia Copyright 2016, Nevro Corp. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system or translated into any language or computer language, in any form or by any means, including, but not limited to, electronic, magnetic, optical, chemical, manual, or otherwise without written permission of Nevro Corp. Registered Trademarks: Nevro, HF10, and Senza are trademarks of Nevro Corp. CE Mark effective on 4 May 2010 Nevro hereby declares that the Senza System is in compliance with the essential requirements and other relevant provisions of the R&TTE Directive (1999/5/EC). IMPORTANT: Do not change or modify any component of the Senza Spinal Cord Stimulation system, unless expressly approved by Nevro Corp. 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 2

Explanation of symbols. Refer to the product for symbols that apply. Symbols Description MR Conditional CE Marking of Conformity Manufacturer 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 3

Contents Introduction... 5 Overview... 5 Definition of Terms... 6 Risks Associated with MRI with Senza System... 6 Contraindications Associated with MRI with Senza System... 7 Conditions for Use of MRI with Senza System... 7 Preparation Prior to MRI Examination... 10 Considerations during the MRI Examination... 11 Considerations after the MRI Examination... 11 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 4

Introduction Nevro s Senza Spinal Cord Stimulation (SCS) system is an MR Conditional device that has been demonstrated to present negligible hazards in a specified MR environment when following specific guidelines as described in this document. MR Conditional This document is a supplement to the Senza system physician and patient manuals and is related only to the use of a transmit / receive radio frequency (RF) head or other transmit/receive RF volume coil (wrist coil, knee / foot coil etc.) of a 1.5T or 3T cylindrical bore MRI system for patients implanted with the Senza system. The implanted components of the Senza system may include Nevro percutaneous leads (Model No: LEAD10x8-xxB), Surgical Lead (LEAD3005-xx(B), LEAD3015-xx(B), LEAD3025-xx(B)), lead extensions (Model No: LEAD2008-xxB), lead anchors (Model No: ACCK5xxx), IPG Port plug (ACCK7000), and the Senza implantable pulse generator (Model Nos: NIPG1000 or NIPG1500). Please note that MR Conditional components of the Senza system do not include The trial stimulator (TSM), patient remote, charger, surgical accessories, programmer wand, and clinician programmer. Do not bring these components into the MR scan room. S8 lead adaptors (Model No.: SADP2008-xxB) and M8 lead adaptors (Model No.: MADP2008-xxB). Surgical lead (LEAD3005-xx(B), LEAD3015-xx(B), LEAD3025-xx(B)) It is important to read this full document prior to conducting or recommending an MRI examination on a patient with the Senza system. These instructions only apply to the Senza system, and do not apply to other products. If you have any questions, please contact Nevro at the address or phone number at the end of this document. These instructions can also be found at Nevro s website (www.nevro.com/ous/mri). This manual is applicable only for patients implanted with the Senza system who are undergoing MR scans. This manual is not applicable for Healthcare professionals implanted with Senza system who are involved in conducting MR scans. Overview Magnetic Resonance Imaging (MRI) is a tool used to diagnose various diseases and conditions. MRI uses powerful static magnetic field, gradient magnetic fields, and RF energy to construct an image of a section of the body. 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 5

Bench top tests have shown that patients implanted with Senza system can be safely exposed to MR environments specified in this guideline. However, MR scans performed outside the guidelines may result in the MRI field interacting with implanted devices, potentially injuring the patient and damaging the implanted device. Due to risks associated with using an MRI with an implanted device, it is important to read, understand, and comply with these instructions to prevent potential harm to the patient and/or damage to the device. Definition of Terms - MR Conditional 1 : An item with demonstrated safety in the MR environment within defined conditions. At a minimum, address the conditions of the static magnetic field, the switched gradient magnetic field and the radiofrequency fields. Additional conditions, including specific configurations of the item, may be required. - Radio frequency (RF) magnetic field: The magnetic field in MRI that is used to flip the magnetic moments. - Specific absorption rate (SAR) 1 : Radiofrequency power absorbed per unit of mass (W/kg). - Tesla (T) 1 : The SI unit of magnetic induction equal to 10 4 gauss (G). - Transmit / Receive RF head coil 1 : A coil used to transmit and receive RF energy that is limited to the head only. - Other Transmit / Receive RF volume coil: RF coil producing a homogeneous RF field over the volume encompassed by the coil. The homogenous RF field is limited to a section of body only (e.g. Knee, Wrist etc.). The coil both transmits and receives RF energy. - Trial Phase: A time during which a person with chronic pain tests SCS therapy to see if and how well it works. During the trial phase, the person will use a Trial Stimulator, which is not implanted in the body. Risks Associated with MRI with Senza System The potential risks of performing MRI on patients with an implanted Senza system include: - Device movement - Excessive heating of or around the implanted device components - Tissue damage - Damage to the device 1 ASTM F 2503-13, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 6

- Uncomfortable sensation - Image artifact Contraindications Associated with MRI with Senza System The contraindications associated with performing MRI on patients with an implanted Senza system include: - Do not use the transmit RF body coil for 1.5T and 3T imaging. - Many head and other transmit / receive RF volume coils (e.g. wrist coil, knee coil etc.) are receive only. Do not use a receive only head, or a receive only RF volume coil (e.g. wrist coil, knee coil etc.) as this can cause significant heating at the lead tip resulting in serious patient injury and / or device damage. - No part of the implanted system (implantable pulse generator (IPG), extensions, leads, lead anchors or IPG port plugs) may be within the transmit / receive RF head coil. - Under no circumstances should the transmit / receive RF volume coil be placed over the implanted Senza system. Because of this restriction, scanning of the area where the Senza system is implanted is not possible. - The trial stimulator (TSM), patient remote, charger, surgical accessories, programmer wand, and clinician programmer are MR Unsafe. These devices may be projectile hazards and should not be allowed into the MRI scan (magnet) room. Conditions for Use of MRI with Senza System 1.5T and 3T MR scan conditions MRI examinations of the head and extremities can be safely conducted in patients with the Senza system if all the instructions in this document are followed. Non-clinical testing has shown the Senza system to be MR conditional when exposed to the MRI environment under the specific conditions listed below: - Conditions for all scans Do not perform an MRI if the patient has a device or device component (lead, extension, etc) from a different manufacturer attached to the Nevro IPG. The risks of performing MRI scan with a Nevro IPG connected to a lead manufactured by a different company have not been evaluated. Use only a transmit / receive RF head coil or other transmit / receive RF volume coil (e.g. Wrist coil, Knee coil etc.). The risk of using other types of RF coils has not been evaluated. 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 7

Use only a 1.5T or 3T cylindrical bore MRI system with a maximum static field spatial gradient of 1110 gauss/cm (11.1 T/m). Please consult the MR manufacturer for the maximum spatial gradient field in the MR scanner. Do not use open-sided MRI systems or systems operating at higher or lower Tesla value (0.5, 1.0 or 4.0T). The risks of using MRI systems operating at other Tesla values have not been evaluated. The trial stimulator (TSM), patient remote, charger, surgical accessories, programmer wand, and clinician programmer are MR Unsafe and should not be allowed into the MRI scan (magnet) room. Stimulation must be turned off. Do not perform MR scan if the patient is undergoing the trial phase. Please refer to the Nevro s patient manual about details on the trial phase. Do not conduct an MRI if the implanted Nevro lead(s) or lead extension(s) are not connected to the IPG. Do not conduct an MRI if the patient has implanted leads / extensions from another manufacturer which are not connected to a IPG. Only use MRI systems which limit slew rate to 200T/m/sec per axis or less. Please consult the MR scanner manufacturer for the maximum possible slew rate on the MR scanner. Limit total scan time to 15 minutes. - Additional conditions for head MR scans Only use transmit / receive RF head coil. No part of the implanted Senza system (implantable pulse generator (IPG), extensions, leads, lead anchors, or IPG port plugs) may be within the transmit / receive RF head coil. The location of implanted Senza system components shall be confirmed prior to the MR scan to ensure compliance with this condition. Use MRI scan parameters that limit the head Specific Absorption Rate (SAR) level lower than 3.2 W/kg. - Additional conditions for extremity MR scans Only use transmit / receive RF volume coils (e.g. Wrist coil, Knee coil etc.). Under no circumstances should the transmit / receive RF volume coil be placed over the implantable Senza system (implantable pulse generator (IPG), extensions, leads, lead anchors, or IPG port plugs). The location of implanted Senza system components shall be confirmed prior to the MR scan to ensure compliance with this condition. 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 8

Conduct wrist MR scans with wrist placed above the head to minimize RF interaction with the implanted Senza device. Risks associated with placing the wrist coil close to the torso during MR scan have not been evaluated. 1.5T and 3T MR scan scenarios Head MR Imaging Figure 1: Head MRI scans are permissible using 1.5T or 3T transmit / receive RF head coil, as long as the implanted Nevro Senza system components are not within the transmit / receive head coil and other aforementioned scan conditions are met. Extremity MR Imaging (a) Figure 2: Extremity scans are permissible using an appropriate transmit / receive RF volume coil, as long as the RF volume coil is not placed over the implanted Nevro Senza system (b) 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 9

components and other aforementioned scan conditions are met. (a) represents a permissible knee scan scenario. (b) represents a permissible wrist MR scan scenario. Although not illustrated, MRI scans of the ankle are also possible using an appropriate transmit / receive RF volume coil. Preparation Prior to MRI Examination - Inform the patients of all the risks associated with undergoing an MRI examination as stated in this document. - A trained professional with the proper knowledge of MRI equipment such as an MRItrained radiologist or MRI physicist must ensure the MRI examination will be conducted according to the information outlined in this document. - Identify if the patient has any other medical device implants. The most restrictive MRI exposure requirements must be used if the patient has multiple medical device implants. Consult with the manufacturers of the devices. - Document the patient s programming parameters. - Perform an impedance check using Nevro s Clinician Programmer. Do not perform an MRI if any impedance is greater than 10 kω. Details on how to conduct impedance check can be found in Nevro s Physician Implant manual. - Turn stimulation off. This can be done using either the programmer, patient remote, or patient charger. Details on how to switch off the simulation can be found in Nevro s Patient manual. - Do not conduct an MRI if the implanted lead(s) or lead extension(s) are not connected to the IPG. - Do not conduct an MRI with a trial stimulator (TSM), patient remote, charger, surgical accessories, programmer wand, and clinician programmer in the MRI scan room. - Any part of the implanted Senza system (IPG, extensions, leads) should not be in the transmit / receive RF head coil. - The transmit / receive RF volume coil should not be placed over any components of the implanted Senza system. Because of this restriction, scanning of the area where the Senza system is implanted is not possible. - If possible, do not sedate the patient so the patient can inform the MRI operator of any problems during the examination. - Instruct the patient to immediately inform the MRI operator if any discomfort, stimulation, shocking, or heating is experienced during the examination. 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 10

Considerations during the MRI Examination - Carefully monitor the patient throughout the MRI examination both visually and audibly. Discontinue the MRI examination immediately if the patient cannot respond to questions or reports any problems. - Once the Nevro IPG has been turned off, the MRI will not turn on the IPG. Considerations after the MRI Examination - Turn the device on and restore the IPG to pre-mri settings. - Confirm that the IPG has been restored to pre-mri settings. 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 11

NEVRO CORP. All questions or concerns about Nevro Corp. products should be forwarded to: Nevro Corp. 1800 Bridge Parkway Redwood City, CA 94065 USA Tel: +1.650.251.0005 Fax: +1.650.251.9415 Email: info@nevrocorp.com 11095 Senza System 1.5T and 3T MRI Guidelines Rev C Page 12