EEG COHERENCE AND PHASE DELAYS: COMPARISONS BETWEEN SINGLE REFERENCE, AVERAGE REFERENCE AND CURRENT SOURCE DENSITY



Similar documents
Jitter Measurements in Serial Data Signals

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA

Spectrum Level and Band Level

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

7. DYNAMIC LIGHT SCATTERING 7.1 First order temporal autocorrelation function.

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

Signal to Noise Instrumental Excel Assignment

Analytical Test Method Validation Report Template

Welcome to SKIL 3 - the most comprehensive quantitative EEG analysis software available

AN-007 APPLICATION NOTE MEASURING MAXIMUM SUBWOOFER OUTPUT ACCORDING ANSI/CEA-2010 STANDARD INTRODUCTION CEA-2010 (ANSI) TEST PROCEDURE

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

AP1 Oscillations. 1. Which of the following statements about a spring-block oscillator in simple harmonic motion about its equilibrium point is false?

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

Frequency Response of Filters

AP Physics 1 and 2 Lab Investigations

SR2000 FREQUENCY MONITOR

Lecture Notes Module 1

Audio Engineering Society. Convention Paper. Presented at the 129th Convention 2010 November 4 7 San Francisco, CA, USA

MICROPHONE SPECIFICATIONS EXPLAINED

Probability and Random Variables. Generation of random variables (r.v.)

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

AC : MEASUREMENT OF OP-AMP PARAMETERS USING VEC- TOR SIGNAL ANALYZERS IN UNDERGRADUATE LINEAR CIRCUITS LABORATORY

Optimum Resolution Bandwidth for Spectral Analysis of Stationary Random Vibration Data

Fairfield Public Schools

Taking the Mystery out of the Infamous Formula, "SNR = 6.02N dB," and Why You Should Care. by Walt Kester

L9: Cepstral analysis

This document is downloaded from DR-NTU, Nanyang Technological University Library, Singapore.

Appendix A: Science Practices for AP Physics 1 and 2

In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data.

Determination of source parameters from seismic spectra

SENSITIVITY ANALYSIS AND INFERENCE. Lecture 12

TECHNICAL SPECIFICATIONS, VALIDATION, AND RESEARCH USE CONTENTS:

Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Chapter 8 - Power Density Spectrum

MATRIX TECHNICAL NOTES

Simple Regression Theory II 2010 Samuel L. Baker

Positive Feedback and Oscillators

Lecture 1-10: Spectrograms

Classic EEG (ERPs)/ Advanced EEG. Quentin Noirhomme

Measuring Line Edge Roughness: Fluctuations in Uncertainty

Advanced Signal Processing and Digital Noise Reduction

Trigonometric functions and sound

Uses of Derivative Spectroscopy

PeakVue Analysis for Antifriction Bearing Fault Detection

SIGNAL PROCESSING & SIMULATION NEWSLETTER

= δx x + δy y. df ds = dx. ds y + xdy ds. Now multiply by ds to get the form of the equation in terms of differentials: df = y dx + x dy.

Overview of Methodology. Human Electrophysiology. Computing and Displaying Difference Waves. Plotting The Averaged ERP

CBS RECORDS PROFESSIONAL SERIES CBS RECORDS CD-1 STANDARD TEST DISC

PHASOR DIAGRAMS HANDS-ON RELAY SCHOOL WSU PULLMAN, WA. RON ALEXANDER - BPA

Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment

Noise reduction of fast, repetitive GC/MS measurements using principal component analysis (PCA)

Fig. 1 :Block diagram symbol of the operational amplifier. Characteristics ideal op-amp real op-amp

Convention Paper Presented at the 118th Convention 2005 May Barcelona, Spain

Experiment #1, Analyze Data using Excel, Calculator and Graphs.

Data Analysis Methods: Net Station 4.1 By Peter Molfese

Appendix D Digital Modulation and GMSK

Supporting Information

Lecture 14. Point Spread Function (PSF)

B3. Short Time Fourier Transform (STFT)

Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide)

CHI-SQUARE: TESTING FOR GOODNESS OF FIT

The accurate calibration of all detectors is crucial for the subsequent data

MUSICAL INSTRUMENT FAMILY CLASSIFICATION

Trigonometry Review with the Unit Circle: All the trig. you ll ever need to know in Calculus

Georgia Department of Education Kathy Cox, State Superintendent of Schools 7/19/2005 All Rights Reserved 1

PRACTICAL GUIDE TO DATA SMOOTHING AND FILTERING

Step Response of RC Circuits

Short-time FFT, Multi-taper analysis & Filtering in SPM12

S-Parameters and Related Quantities Sam Wetterlin 10/20/09

Dynamic sound source for simulating the Lombard effect in room acoustic modeling software

Precision Diode Rectifiers

Nonlinear Signal Analysis: Time-Frequency Perspectives

Predictive Indicators for Effective Trading Strategies By John Ehlers

Author: Dr. Society of Electrophysio. Reference: Electrodes. should include: electrode shape size use. direction.

Definition 8.1 Two inequalities are equivalent if they have the same solution set. Add or Subtract the same value on both sides of the inequality.

7. Beats. sin( + λ) + sin( λ) = 2 cos(λ) sin( )

The Calculation of G rms

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz

Real-time PCR: Understanding C t

The continuous and discrete Fourier transforms

*UDYLW\)LHOG7XWRULDO

Statistical Confidence Calculations

Statistics, Probability and Noise

E190Q Lecture 5 Autonomous Robot Navigation

Estimation of Loudness by Zwicker's Method

Algebra. Exponents. Absolute Value. Simplify each of the following as much as possible. 2x y x + y y. xxx 3. x x x xx x. 1. Evaluate 5 and 123

Generation and Detection of NMR Signals

AP1 Waves. (A) frequency (B) wavelength (C) speed (D) intensity. Answer: (A) and (D) frequency and intensity.

The two dimensional heat equation

Time Series Analysis: Introduction to Signal Processing Concepts. Liam Kilmartin Discipline of Electrical & Electronic Engineering, NUI, Galway

application note Directional Microphone Applications Introduction Directional Hearing Aids

DVB-T. The echo performance of. receivers. Theory of echo tolerance. Ranulph Poole BBC Research and Development

Using Optech LMS to Calibrate Survey Data Without Ground Control Points

Algebra Academic Content Standards Grade Eight and Grade Nine Ohio. Grade Eight. Number, Number Sense and Operations Standard

EE362L, Power Electronics Triac Light Dimmer

Transcription:

Version 1, June 13, 2004 Rough Draft form We apologize while we prepare the manuscript for publication but the data are valid and the conclusions are fundamental EEG COHERENCE AND PHASE DELAYS: COMPARISONS BETWEEN SINGLE REFERENCE, AVERAGE REFERENCE AND CURRENT SOURCE DENSITY Robert W. Thatcher, Ph.D., 1,2, Carl J. Biver, Ph.D. 1 and Duane North, M.S. 1 NeuroImaging Laboratory, Bay Pines VA Medical Center 1 and Department of Neurology, University of South Florida College of Medicine Tampa, Florida 2 Send Reprint Requests To: Robert W. Thatcher, Ph.D. NeuroImaging Lab, VA Medical Center, Bldg. 23, Room 117 Bay Pines, Fl 33744 (727) 391-0890

Thatcher et al, 2 Abstract Mathematical evaluations of coherence and phase differences were conducted using different signal-to-noise ratios based on mixtures of sine waves and random noise. Comparisons were conducted between linked ears reference, single or a common reference, average reference and the Laplacian reference. The results of the analyses demonstrate that linked ears and common references provide valid and systematic relations between the magnitude of coherence and signal-to-noise ratio. In contrast, an average reference and Laplacian reference fail to systematically vary as a function of the signal-to-noise ratio. The reasons for the failure of the average reference and Laplacian is because both methods involve mixing signals from each channel into all other channels whereas a common reference and linked ears do not add signals to other channels. 1.0 Introduction EEG coherence is a measure of the degree of association or coupling of frequency spectra between two different time series while EEG phase delays are a measure of the temporal lead or lag of spectra. Mathematically coherence is defined as the normalized cross-power spectrum and phase delay as the phase angle and it is computed between two simultaneously recorded EEG signals from different scalp locations per frequency band. Coherence is often interpreted as a measure of coupling and as a measure of the functional association between two brain regions (1 Sklar et al, 1972; 2 Shaw et al, 1976; 3 1978; 4 Nunez, 1981; Nunez et al, 1997 5). Coherence is a sensitive measure that can reveal subtle aspects of the network dynamics of the brain which complement the data obtained by power spectral analyses. For example, coherence has been applied to studies of cognition (6 Beaumont et al, 1978; 7 Gevins et al, 1987; 8,9 Thatcher et al, 1983; 1987; 10 Petsche et al, 1993; 11 Rappelsberger and Petsche, 1988), brain maturation (Thatcher et al, 1987 13, Gasser et al, 1988 14; van Baal et al, 1999 15) spatial tasks (Rappelsberger and Petsche 11), agenesis of the corpus callosum (Nielsen et al, 1993 16; Koeda et al, 1995 17; and various clinical populations (Hughes and John, 1999; John et al, 1988 20, Thatcher et al, 1998 ). EEG phase delays are often used to compute directed coherence which is a measure of the directional flow of information between two EEG electrode sites (21, 22). Similar to coherence, EEG phase delays have been used in the analysis of cognition (21, 22) and a variety of clinical conditions (Thatcher et al, 1991 ).

Thatcher et al, 3 EEG coherence and EEG phase delays are statistical estimates and are dependent on the number of degrees of freedom used to smooth or average spectra as well as the reference used to derive the raw digital data. Thus, while there is only one general mathematical equation for the computation of coherence, nonetheless, differences in the accuracy and sensitivity of the computation of coherence and phase delays depends on the amount of averaging across frequency bands or across records, e.g., replications ( ). Fein et al (1988 26) showed that interhemispheric coherence was inflated when a common reference with a strong signal was used such as Cz. The authors also computed coherence using reconstructed reference-free signals using the source derivation method of Hjorth (1975 27) with ambiguous results. As pointed out by Rappelsberger (1989 28) source derivation is not reference free because it measures the difference between the potential at one electrode with respect to the weighted average potential surrounding that electrode. Another commonly used EEG reference is the average reference in which the average of the potentials from all electrodes is subtracted for the electrical potentials at each electrode. Rappelsberger (1989 28) used EEG simulations to evaluate EEG coherence recorded with a single reference, the average reference and source derivation and concluded that EEG coherence is invalid using either the average reference or source derivation, e.g., Tremendous distortions of the theoretical assumptions of common average reference recording and by source derivation are found in the coherence maps pg. 66. This same conclusion was arrived at by A. Korzeniewska, M. Mañczak, M. Kamiñski, K. Blinowska, S. Kasicki Determination of information flow direction between brain structures by a modified Directed Transfer Function method (ddtf) Journal of Neuroscience Methods 125, 195-207, 2003 The purpose of this paper is to explore and extend the findings of Rappelsberger, 1989 (28) by comparing EEG coherence as well as EEG phase delays using a single reference, an average reference and a source derivation. Simulations of EEG will involve the use of a systematic mixture of sine waves mixed with noise and recorded from the 19 channels of the international 10/20 system. 2.0 Methods 2.1 EEG Simulations

Thatcher et al, 4 The EEG was simulated by mixing a known sine wave (1 to 10 uv) at a sample rate of 128 Hz at a given scalp location with varying amplitudes of random gaussian noise (0 to 100 uv) and a constant phase delay of 30 degrees.. The sine waves are defined as: Eq.1: x = standard sine wave equation and amplitude, frequency and phase parameters Where xx is the parameter that controls amplitude, xx controls frequency and xx controls phase shifts. Coherence was evaluated by comparing a single channel with a pure10 uv sine wave (Fp1) with respect to the remaining 18 channels of the 10/20 electrode system in which varying amounts and this same sine wave was mixed with varying amplitudes of gaussian random noise. Signal-to-Noise (S/N) was defined as the amplitude of the pure sine wave divided by the amplitude of noise. Figure 1 is an example of the simulated EEG sine waves in which Fp1 = 10 uv and Fp2 = 10 uv with a 30 degree phase shift and no noise. Figure 2 is an example of the simulated EEG sine waves in which Fp1 = 10 uv and Fp2 = 10 uv with a 30 degree phase shift and 10 uv of white noise added to Fp2.

Thatcher et al, 5 Figure 3 is an example of the simulated EEG in which Fp1 = 10 uv, Noise = 0; Fp2 = 10 uv, Noise = 0, F3 = 10 uv, Noise = 1 uv; F4 = 10 uv, Noise = 2 uv; C3 = 10 uv, Noise = 3 uv; C4 = 10 uv, Noise = 4 uv..... to Pz = 10 uv, Noise = 18 uv. The range of S/N was from 10 to.01 involving

Thatcher et al, 6 200 1 uv noise increments. Figure 4 is the FFT of the signals in figure 3.

Thatcher et al, 7 2.2 EEG Reference Computations The single reference was defined as A i X i, where A i = an ideal reference and X i is a single EEG channel for each time point i. Average reference is defined as A i X i, where A. Where N = i X i N 19 electrode channels. In other words, the average electrical potential at each time point is subtracted from each electrode at each time point. The surface Laplacian or current source density (CSD) was computed using the spherical harmonic Fourier expansion of the EEG scalp potentials to estimate the CSD directed at right angles to the surface of the scalp in the vicinity of each scalp location (Pascual- Marqui, Gonzalez-Andino, Valdes-Sosa, & Biscay-Lirio, 1988 29 ). The CSD is the second spatial derivative or Laplacian of the scalp electrical potentials which is independent of the linked ear reference itself. The Laplacian is free of a single reference but it is not reference free because it is dependent upon the electrical potential gradients surrounding each electrode. 2.3 Computation of Coherence The sample length of the simulated EEG samples = 1 minute, sample rate = 128 and the FFT was computed using 2 second overlapping windows with a 75% overlap as described by Kaiser and Sterman

Thatcher et al, 8 (1999). This provided a total of 117 windows that were subjected to cross-spectral analyses. Coherence is mathematically defined two channels X and Y as: Eq. 2 - Coherence (f) = Cross Spectrum( f ) XY ( Autospectrum( f )( X ))( Autospectrum( f )( Y )) 2 However, this standard mathematical definition of coherence hides some of the essential statistical nature and structure of coherence. The algebraic notation of equation 3 helps illustrate the fundamental statistics of coherence: Eq. 3 - Coherence (f) = and: ( N ( a( u( y) b( v( y))) N 2 ( a( b( ) 2 ( N N ( a( v( y) b( u( y))) u( y) 2 2 v( y) ) 2 a( = cosine coefficient for the frequency (f) for channel x b( = sine coefficient for the frequency (f) for channel x u(y) = cosine coefficient for the frequency (f) for channel y v(y) = sine coefficient for the frequency (f) for channel y Where N and the summation sine represents averaging over frequencies in the raw spectrogram or averaging replications of a given frequency or both. The numerator and denominator of coherence always refers to smoothed or averaged values, and, when there are N replications or frequencies then each coherence value has 2N degrees of freedom. Note that if spectrum estimates were used which were not smoothed or averaged over frequencies nor over replications, then coherence = 1 (Bendat and Piersol, 1980; Benignus, 1968; Otnes and Enochson, 1972). In order to compute coherence, averaged cospectrum and quaspectrum smoothed values with degrees of freedom > 2 and error bias = 1/N must be used (xx. In the EEG simulations in this study spectra were averaged across 0.5 Hz frequency bands from 4 to 7 Hz, N = 7 and across replications N = 117 and the degree of freedom = 124 x 2 = 248. The error bias = 1/248 =.00403 2.4 Computation of Phase Delays

Thatcher et al, 9 Coherence and phase angle are linked by the fact that the average temporal consistency of the phase angle between two EEG time series is directly proportional to coherence. For example, when coherence is computed with a reasonable number of degrees of freedom and approaches unity, then the phase angle between the two time-series becomes meaningful because the confidence interval of phase is a function of the magnitude of the coherence and the degrees of freedom. If the phase angle is random between two time series then coherence = 0. Another way to view the relationship between phase consistency and coherence is to consider that if Coherence = 1, then once the phase angle relation is known the variance in one channel can be completely accounted for by the other. The phase relation is also critical in understanding which time-series lags or leads the other or, in other words the direction and magnitude of the delay. The phase angle is defined as: Eq. 4: Phase angle (f) = Arctan ( Smoothedquadspectrum( f )) ( Smoothed cos pectrum( f )) Where the quadspectrum and cospectrum are defined using the same algebraic notation as in equation xx: Eq. 5 : Cospectrum ( f ) ( a( u( y) b( v( y)) N Eq. 6: Quadspectr um( f ) ( a( v( y) b( u( y)) N As mentioned previously, the confidence internal for the estimation of phase delay or phase angle is directly related to the magnitude of coherence. This means that when coherence is too low, e.g., < 0.2, then the estimate of the phase angle may not be reliable. 3.0 Results 3.1 EEG Coherence and Phase using a single reference

Thatcher et al, 10 Figure 2 are the results of the computation of EEG coherence and EEG phase delays using the single reference EEG simulation. The y-axis in figure 2 (top) is coherence and the x-axis is the signalto-noise ratio (S/N). The y-axis in figure 2 (bottom) is phase delay (degrees) and the x-axis is the same signal-to-noise ratio (S/N) as in figure xa. It can be seen in Figure 2 that coherence decreases as a linear function of signal-to-noise ratio. It can also be seen in Figure 2 that an approximate average the 30 degree phase delay is present as signal-to-noise increases, however, the variance of the phase delay increases as a function of S/N, especially at coherence values < 0.2. Figure 3 shows the relationship between the standard deviation of coherence and phase delays as a function of signal-to-noise. These results show that EEG coherence standard deviations are relatively stable as S/N increases, however, phase delay standard deviations dramatically increase when coherence < 0.2.

Thatcher et al, 11 3.2 - EEG Coherence and Phase using the average reference Figure 4 are the results of the computation of EEG coherence and EEG phase delays using the average reference EEG simulation. The y-axis in figure 4 (top) is coherence and the x-axis is the signalto-noise ratio (S/N). The y-axis in figure 4 (bottom) is phase delay (degrees) and the x-axis is the same signal-to-noise ratio (S/N) as in figure 2. It can be seen in Figure 4 that coherence is extremely variable and does not decrease as a linear function of signal-to-noise ratio. It can also be seen in Figure 4 that EEG phase delays never approximate 30 degrees and are extremely variable at all levels of noise.

Thatcher et al, 12 3.3 - EEG Coherence and Phase using the source derivation Figure 5 are the results of the computation of EEG coherence and EEG phase delays using the average reference EEG simulation. The y-axis in figure 5 (top) is coherence and the x-axis is the signalto-noise ratio (S/N). The y-axis in figure 5 (bottom) is phase delay (degrees) and the x-axis is the same signal-to-noise ratio (S/N) as in figure 2. It can be seen in Figure 5 that coherence is extremely variable and does not decrease as a linear function of signal-to-noise ratio. It can also be seen in Figure 5 that EEG phase delays never approximate 30 degrees and are extremely variable at all levels of noise.

Thatcher et al, 13 4.0 Discussion The choice of reference in the computation of EEG coherence is important in quantitative EEG. Studies by Beaumont and Rugg (1979) and French and Beaumont (1984) emphasize the fact that any signal contained in a single reference will be shared by all of the scalp EEG electrodes, depending on the location of the single reference also called a common reference. Coherence, however, is amplitude independent thus the absolute magnitude of the EEG signal in a single or common reference is irrelevant, except where the magnitude of the signal is very large, such as a Cz reference (Fein et al, 19x. The results of this study are consistent with those by Rappelsberger, 1989 who emphasized the value and validity of using a single reference and linked ears in estimating the magnitude of shared or coupled activity between two scalp electrodes. The use of re-montage methods such as the average reference and Laplacian source derivation are useful in helping to determine the location of the sources of EEG of different amplitudes at different locations. However, the results of this study which again confirm the findings of Rappelsberger, 1989 show that coherence is invalid when using either an average reference or a source derivation which is also arrived at in a later study by A. Korzeniewska, M.

Thatcher et al, 14 Mañczak, M. Kamiñski, K. Blinowska, S. Kasicki Determination of information flow direction between brain structures by a modified Directed Transfer Function method (ddtf) Journal of Neuroscience Methods 125, 195-207, 2003 It is easy to understand why coherence is invalid when using an average reference since the summation of signals from all channels is subtracted or added to the electrical potentials recorded at each electrode. In the present study, a comparison of Figure 2 and figure 4 shows the results of the average reference where noise and signal from each channel is incorporated into all of the channels by being subtracted from the electrical potential recorded from each channel. Thus, signals and noise are mixed and added to the recordings from each channel making coherence and phase delays invalid. A similar situation prevails with source derivation since spatially weighted signals and noise from other channels are averaged and subtracted from the electrical potential recorded from each electrode site. Coherence when using the average reference or source derivation is especially sensitive to the presence of artifact or noise since the artifact will be mixed with and added to all channels. The results of this study also show that measures of phase delay are invalid when using average reference or source derivations. The same arguments for the average reference can be applied to source derivation as they pertain to EEG phase delays. An important finding in the present study concerns the relationship between coherence and phase delays using a single reference or common reference in which the variance of the phase angle dramatically increases at low levels of coherence. We have designated a threshold value of coherence < 0.2 below which EEG phase is highly variable and misleading. Although mean phase delay remains at 30 degrees even at high levels of noise (see figures 2 & 3), nonetheless, the extreme variability invalidates EEG phase delays at low levels of coherence (e.g., < 0.2). A similar finding was reported by xxx and xxx. XXX suggested a coherence threshold < 0.4, however, this value was a result of using lower degrees of freedom than in the present study for the computation of coherence. It is important to use as high a number of degrees of freedom as possible since, in this study, coherence values were valid even when the signal-to-noise = 0.01. It is also important to recognize that estimates of EEG phase delays are extremely variable for low values of coherence, e.g., < 0.2 when using a single or common reference. 5.0 References

Thatcher et al, 15