Computed Radiography: Acceptance Testing and Quality Control



Similar documents
INTRODUCTION. A. Purpose

Quality Control and Maintenance Programs

Overview of Digital Detector Technology

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

Page: 1 of 6 Page: 1 of 6

Physics testing of image detectors

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Quality Control of Full Field Digital Mammography Units

The Exposure Index and Its Standardization

Digital Mammography Update: Design and Characteristics of Current Systems

Performance evaluation of computed radiography systems

Digital Radiography Image Quality: Image Acquisition

Understanding Digital Modalities: System Integration and Use

Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

Performance testing for Precision 500D Classical R/F System

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

A PACS-Aware DICOM Image Object

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Best Practices in Digital Radiography

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Interfacing Fuji CR in a DICOM Network Environment

Digital Breast Tomosynthesis QC Requirements

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

Digital radiography conquers the veterinary world

Update on ACR Digital Mammography QC Manual

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Learning Objectives: Quality Assurance (QA) is. Quality Assurance Procedures for Digital Radiography. Alternate definition of Quality Assurance (QA)

Digital radiography has become the norm in U.S.

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

General requirements. 1. Administrative Controls.

Aspire FCRm/CRm Fujifilm CR for Mammography

Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Chapter 15. Creating the Digital Image. Objectives. Key Terms

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

Using Automatic Exposure Control in Digital Radiography

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

Prepublication Requirements

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Checklist for the Qualification of Digital Detector Array Systems

Module 13 : Measurements on Fiber Optic Systems

Diagnostic Exposure Tracking in the Medical Record

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

CR30-ORACLE CR. with Office Viewer. A Revelation in Computed Radiography. Innovations in Digital Imaging. Solution for Quantum Medical Imaging

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

Lecture 14. Point Spread Function (PSF)

Required RS Training Info

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

REGIUS History. Touch REGIUS110. REGIUS170 ImagePilot REGIUS150 REGIUS330 CS-1

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Medical imaging monitors specification guidelines

A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose

AAPM REPORT NO. 15 PERFORMANCE EVALUATION AND QUALITY ASSURANCE IN DIGITAL SUBTRACTION ANGIOGRAPHY

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

DICOM Correction Item

MQSA Quality Control Manual for Monochrome Displays for Mammography

Recommendations for a technical quality control program for diagnostic X-ray equipment

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

RIS. Complete -Complete exam on RIS *** [! - 9 ] ***

Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

X-ray Imaging Systems

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Digital Fluoroscopic Imaging: Acquisition, Processing & Display

FTIR Instrumentation

Transcription:

Computed Radiography: Acceptance Testing and Quality Control J. Anthony Seibert, Ph.D. University of California Davis Department of Radiology Sacramento, California Introduction CR is the primary means to capture D images in a PACS environment Acceptance testing validates performance Quality control verifies optimal operation Considerations: Knowledge of CR attributes and operation Understanding the tests Determining appropriate results Presentation Outline Overview of CR How does it work? What are the issues? Acceptance test procedures What tests? Why? How? Quality control When? What? How often? Computed Radiography (CR)...is the generic term applied to an imaging system comprised of: Photostimulable Storage Phosphor to acquire the x-ray x projection image CR Reader to extract the electronic latent image Digital electronics to convert the signals to digital form. X-ray Exposure Patient CR Image Acquisition unexposed 5. Computed Radiograph. Acquisition Transmitted x-raysx through patient Digital Pixel Matrix. Display Digital to Analog Conversion X-ray system exposed. Image Reader Phosphor plate 3. Image Scaling. Image Record Charge collection device Analog to Digital Conversion X-ray converter x-rays electrons Digital processing 3. Archiving

CR Reader CR - QC Workstation DICOM Film laser printer PACS Soft-copy review CR Networking PACS and DICOM Digital Imaging COmmunications in Medicine Provides standard for modality interfaces, storage/retrieval, and print Modality Worklist Input (from RIS via HL-7) Technologist QC Workstation Image manipulation and processing Processed / Unprocessed images DICOM image output PSL 3. ev CR: How does it work? Photostimulated Luminescence phonon f 6 5d f 7 t Eu F/F + Eu 3+ / Eu + Incident x-rays t tunneling e - e t recombination Laser stimulation. ev Conduction band 8.3 ev Energy Band BaFBr Valence band PSLC complexes (F centers) are created in numbers proportional to incident x-ray x intensity Relative intensity..5. 8.5 Stimulation and Emission Spectra Stimulation 7.75 Diode 68 nm 6 BaFBr: Eu + Optical Barrier 5.5 Emission 3 λ (nm) Energy (ev) 3 Exposed Imaging Plate Photostimulated Luminescence Incident Laser Beam Light Scattering Light guide Photostimulated Luminescence Laser Light Spread "Effective" readout diameter PMT PSL Signal Protective Layer Phosphor Layer Base Support Laser Source Reference detector Polygonal Mirror Laser beam: Scan direction CR: Latent Image Readout f-θ lens Plate translation: Sub-scan direction Cylindrical mirror Light channeling guide PMT Output Signal ADC x= 79 y= To 333 image processor z= 5

Phosphor Plate Cycle Scan Direction Laser beam deflection Sub-scan Direction Plate translation Typical CR resolution: 35 x 3 cm --.5 lp/mm ( µm) x 3 cm -- 3.3 lp/mm (5 µm) 8 x cm -- 5. lp/mm ( µm) Screen/film resolution: 7- lp/mm (8 µm - 5 µm) reuse x-ray exposure laser beam scan light erasure PSP Base support plate exposure: create latent image plate readout: extract latent image plate erasure: remove residuals Computed Radiography Exposure Latitude: Dynamic Range Acquisition, Display and Archive are separate functions Variable speed detector to speed Wide dynamic range. to mr Image processing is a crucial requirement Signal output Film : CR : Log relative exposure Raw image Processing the Image Inherent subject contrast displayed Contrast inverted (to screen-film) PSL signal amplitude log amplified Image pre-processing processing Find pertinent image information (histogram analysis) Scale data to an appropriate range Contrast enhancement Anatomy specific grayscale manipulation Spatial frequency enhancement 3

Finding the Image Location Histogram analysis Image recognition phase Collimation (Agfa) EDR, automatic mode (Fuji) Segmentation (Kodak) Finding collimation borders and edges Frequency distribution of pixel values within a defined area in the image Shape is anatomy specific Sets minimum and maximum useful pixel values Histogram: frequency distribution of pixel values in an image Histogram Distribution Frequency Value 3 Frequency 6 8 6 3 5 Value Frequency Histogram Distribution Collimated area Anatomy Useful signal Direct x-ray area Pixel value The shape is dependent on radiographic study, positioning and technique Frequency of Digital Number.mR Histogram Distribution S.mR S K mr S Latitude (L) mr 3 Q 5 Q mr Sensitivity (S) Digital value

Relative PSL Data conversion Exposure into digital number - Histogram - Exposure input min max 3 5 3 Raw Digital Output Grayscale transformation Input to output digital number, number, Output digital number 8 6 6, Input digital number. Find the. Scale to 3. Create film signal range look-alike Frequency 8 6 Histogram: pediatric image to 9368 to 833 6 8 Digital value Useful image range for anatomy Relative PSL Data conversion for overexposure - Exposure into digital number Reduce overall gain Pre-processed raw image Scaled and inverted: unprocessed image Exposure input overexposure - min 3 max 5 3 Raw Digital Output (scaled and log amplified) Screen-Film Computed Radiography Underexposed Overexposed Underexposed Overexposed 5

Data conversion for wide latitude Exposure into digital number Change gradient (auto mode) Screen-Film 8 kvp, 8 mas 8 kvp, 6 mas CR 8 kvp, 8 mas Relative PSL - Exposure input - 3 low kvp (wide range) min max 5 3 Raw Digital Output (scaled and log amplified) speed screen - film L=, wide latitude Contrast Enhancement Look-up up-table transformation Optimize image contrast via non-linear transformation curves Unprocessed images display linear subject contrast Gradation processing (Fuji) Tone scaling (Kodak) MUSICA (Agfa) Output digital number, 8 6 M L E A GT Gradient Type Fuji System Example LUTs 6 8, Input digital number Spatial Frequency Processing Edge Enhancement Solid: Edge Difference: Original Enhanced: original MTF response Dash: Difference Original low pass filtered - + filtered Original Response Sum Original Blurred Raw Unprocessed Contrast Enhanced low low low high high Spatial frequency Difference Edge enhanced 6

Fuji CR Parameter Settings Anatomy Anatomical region GA LUT shape parameters GT GC GS Frequency enhancement parameters General chest (LAT). B.6 -.. R. General chest (PA).6 D.6 -.5. R. Port Chest GRID.8 F.8 -.5. T. Port Chest NO GRID. D.6 -.5. R.5 Peds chest NICU/PICU. D.6 -. 3. R.5 Finger.9 O.6.3 5. T.5 Wrist.8 O.6. 5. T.5 Forearm.8 O.6.3 5. T.5 Plaster cast (arm).8 O.6. 5. T.5 Elbow*.8 O.6. 7. T. Upper Ribs*.8 O.6. 5. R. Pelvis*.9 O.6. 6. T. Pelvis portable.9 O.6.. T.5 Tib/Fib.9 N.6.5 5. F.5 Foot.8 O.6.3 5. T.5 Foot*. N.6 -.5 7. T.5 Os Calcis.8 O.6. 5. F. Foot cast.8 O.6.5 5. F.5 C-spine. F.6.5 5. P.5 T-spine.8 F.8 -.5. T. Swimmers. J.9.3 5. T.5 Lumbar spine. N.9. 5. T. Breast specimen.5 D.6.35 9. P. RN RT RE Acceptance Test / QC considerations Image acquisition Electro-Mechanical readout Image processing PACS / RIS interfaces Image handling Recommended Acceptance Tests Recommended Acceptance Tests Physical Inspection Inventory Inventory PACS Interfaces Imaging Plate Uniformity and Dark Noise Signal Response: Linearity and Slope Signal Response: Exposure calibration and beam quality Laser Beam Function High Contrast Resolution Noise / Low-Contrast Response Distortion Erasure Thoroughness Artifact Analysis: Hardware/Software Positioning and collimation robustness Imaging Plate Throughput Acceptance test tools required Exposure meter/dosimeter Spatial resolution phantom Low contrast phantom Vendor QC phantom (periodic tests) SMPTE test pattern Anthropomorphic phantom Documentation log / spreadsheet / instructions CR: Spatial Resolution Phosphor plate sizes: impact on resolution 35x3 (x7). mm pixels x3 (x) 8x (8x). mm pixels. mm pixels 7

High Contrast (Spatial) Resolution MTF Curves 35 x 3 cm 8 x cm 35 x 3 cm MTF Sampled MTF: Standard CR K x K matrix 35 x 3 cm.8.6.. Screen-film 6 8 Spatial Frequency (lp/mm) Pre-sampled MTF Scan Subscan Hi res CR Standard CR X-ray Absorption Efficiency Low Contrast Response: Leeds TO-6 Photon absorption fraction.8.6.. BaFBr, 5 mg/cm² BaFBr, mg/cm² 6 8 Energy (kev) Gd O S, mg/cm 3.5 mr 7 kvp.5 mr Uniformity 98 58 537 8 9 53 mas mas 97 55 5 87 8

Radiation Dose for CR Variable Speed Detector Optimal dose (UC Davis) Adult chest image: S=-3 Neonates/pediatrics: S=-6 Extremities: S= 75- Lower detection efficiency, luminance and readout noise Anti-scatter grids needed Sensitivity number, S Estimate of the incident exposure on the IP Comparable to screen-film speed Amplification required to map median value of histogram to 5 ( to 3 grayscale) Dependent on histogram shape and examination selected Date: 7//98 Location: UCDMC, ACC, 3 Physicist: Anthony Seibert, Identification: CR unit 3 Medical Ph.D. System UC Davis Medical Center CR Reader and Screens Signal Response: Calibration and Beam Quality Note: Use mas values to provide an approximate exposure of of mr to to the IP. Menu = TEST Exposure Conditions x7 SubMenu = Ave. Focal spot Time delay SID (cm) SMD (cm) IP Type: ST IP SN: L =,, EDR = semi. mm ~ min 3 kvp Dependency kvp Filtration mas mr-meter mr-ip S S (mr) OD NA 6 Al/.5 Cu 5..6.9..59. NA 8 Al/.5 Cu.5.6.9 8. 98.7.38 NA 5 Al/.5 Cu.3..98 5. 3..5 NA Maximum Difference:.33.7 NA Filtration Dependency kvp Filtration mas mr-m eter mr-ip S S (mr) OD NA 8 none.5.96.83 87. 5.79. NA 8 Al/.5 Cu.5.6.9 8. 98.7.38 NA 8 Al/.5Cu 6..99.85. 5.85. NA Maximum Difference: 56.8. NA Response.. 8. 6... S (mr). 5 5 7 7 9 9 3 kvp Response 8. 6.... 8. 6.. S (mr).. none Al/.5 Cu Cu Al/.5Cu Filtration Guidelines for QC based on Exposure Sensitivity number > <. mr 6.3-.. mr 3-6.-.3.3 mr 5-3.3-.. mr 75-5.3-.7 mr 5-7.-.7 mr <5 >. mr Indication Underexposed: repeat Underexposed: QC exception Underexposed: QC review Acceptable range Overexposed: QC review Overexposed: QC exception Overexposed: repeat Sensitivity number, S Imaging plate (HR vs. ST) differences Examination specific histogram shapes S number varies with examination type EDR mode effects Automatic (determines S and S values on histogram) Semi-automatic (average value within ROI) Fixed (system acts like screen-film detector) X-ray beam spectrum effects S number varies with beam hardness (calibration required) What S value is appropriate? Determined by examination UCDMC targets Adult exams (CXR, abdomen, etc) 5 3 Extremities (ST plates) 75 5 Pediatrics 3 6 CR s variable speed should be used to advantage Anatomical information can be lost with too high or too low exposure 9

Adult portable chest calculated exposures #exams 6 5 3 Low First half, 99, 57 exams 5 38.3% 53.9% 7.8% 3 Incident Exposure System speed (S #) Q Q <5 High Target exposure range Adult portable chest calculated exposures #exams 6 5 3 Low Second half, 99, 66 exams 5 3.% 73.5% 3.% 3 System speed (S #) Incident Exposure Q3 Q <5 High Target exposure range Number of examinations 8 6 8 6 April - 7, 996 Adult Portable Chest > > > > > 6-699666-6996-699699699 5-59555-595-595959 -9-9-999 35-37353535-3735-373737 3-333-333-333 5-7 5-755-777 - -- 5-7 5-755-777 - -- 5-7 5-755-777 Sensitivity number Exposure Creep Grid technique without a grid Wrong exam 5% Motion 6% Reprinting 9% Underexposure % Other % Repeated Examinations with CR Overexposure % Positioning 6% Total # repeats = 3 from Willis, RSNA 996 Variable Speed Detector Radiation Dose for CR Optimal dose for typical adult chest image is X higher than speed screen/film Lower absorption efficiency Quantum and electronic noise Readout inefficiencies of latent image Anti-scatter grids necessary for most procedures AEC adjustment procedures A -speed equivalent exposure is desirable Empirically determine AEC setting(s) with simple uniform phantoms UCDMC technique: use fixed mode (S=) and sensitivity test menu; adjust AEC response according to changes in film optical density Verify settings with semi-auto mode Verify patient exposure S number; recheck often

Date: 7//98 Location: UCDMC, ACC, 3 Medical Physicist: Anthony Seibert, Ph.D. System Identification: CR unit 3 UC Davis Medical Center CR Reader and Screens Inspection Results Summary Acceptable. Physical Inspection - Inventory Yes. Imaging Plate Uniformity and Dark Noise Yes 3. Signal Response: Linearity and Slope Yes. Signal Response: Calibration and Beam Quality Yes 5. Laser Beam Function Yes 6. High-Contrast Resolution Yes 7. Noise/Low-Contrast Response Yes 8. Distortion Yes 9. Erasure Thoroughness Yes*. Anti-Aliasing Yes. Positioning and Collimation Errors Yes. Throughput Yes Problem areas Film-based performance measurements Digital image analysis tools and evaluation methods not readily available Low contrast resolution measurements Lack of a standardized QC phantom Comments: Quality Control Three levels of system performance quality control. Routine: Technologist level - no radiation measurements. Full inspection: Physicist level - radiation measurements and non-invasive adjustments 3. System adjustment: Vendor service level - hardware and software maintenance Daily (technologist) Periodic Quality Control Inspect CR system and status. Interfaces: PACS broker, ID terminal, QC workstation Erase image receptors (if status unknown). Periodic Quality Control Weekly / Biweekly (technologist) Calibrate review workstation monitors (SMPTE). Acquire QC phantom test images. Verify performance. Check filters / vents and clean as necessary. Periodic Quality Control Quarterly (Technologist) Inspect cassettes. Clean with recommended agents. Review image retake rate and exposure trends. Update QC log. Review out-of of-tolerance issues. Clean screens with recommended agents.

Periodic Quality Control Annually (Physicist) Perform linearity / sensitivity / uniformity tests Inspect / evaluate image quality Re-establish establish baseline values (Acceptance Tests) What is needed? Computer friendly phantoms Objective quantitative analysis methods System performance tracking and database logs Exposure monitoring tools and database tracking Review retakes, exposures, service records. Fuji Agfa Line pair phantoms (contrast transfer tests) Lumisys Diagonal bar (laser jitter test) line/cm grid (visual aliasing) Home built Fiducial Markers (distance accuracy) Lead attenuator (dynamic range) Notches (geometric accuracy tests) Resolution Bar Pattern (qualitative) Single exposure, qualitative and quantitative Step wedge (signal, signal to noise and linearity response tests) Open area (scan uniformity test) Copper step wedge (dynamic range, linearity, SNR) Edge for Presampled MTF Additional Information / Help AAPM Task Group # document: Email: jaseibert@ucdavis ucdavis.edu Dr. Ehsan Samei spreadsheets http://deckard deckard.mc.duke..mc.duke.edu/~samei/downloads Vendor efforts for QC phantom development and analysis Summary CR is the mainstay for direct digital acquisition of projection radiographs CR acceptance testing and QC are essential for optimal operation A TEAM approach is necessary Technologists, Radiologists, Physicists, Clinical Engineering, Information System Group