Extraction Heights for STIS Echelle Spectra



Similar documents
SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION

COOKBOOK. for. Aristarchos Transient Spectrometer (ATS)

FTIR Instrumentation

ING LA PALMA TECHNICAL NOTE No Investigation of Low Fringing Detectors on the ISIS Spectrograph.

SSO Transmission Grating Spectrograph (TGS) User s Guide

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

Spectral Cube Builder for IRS Scubi

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Measuring the Doppler Shift of a Kepler Star with a Planet

Uses of Derivative Spectroscopy

EDXRF of Used Automotive Catalytic Converters

Spectroscopy Using the Tracker Video Analysis Program

Reduced data products in the ESO Phase 3 archive (Status: 15 May 2015)

Spectral Line II. G ij (t) are calibrated as in chapter 5. To calibrated B ij (ν), observe a bright source that is known to be spectrally flat

Introduction to reflective aberration corrected holographic diffraction gratings

Introduction to Fourier Transform Infrared Spectrometry

Signal to Noise Instrumental Excel Assignment

WFC3 Image Calibration and Reduction Software

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Passive Remote Sensing of Clouds from Airborne Platforms

Lecture 14. Point Spread Function (PSF)

MY FIRST STEPS IN SLIT SPECTROSCOPY


Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Spectrophotometry of Ap Stars

Detecting and measuring faint point sources with a CCD

Shimadzu UV-VIS User s Guide

Function Guide for the Fourier Transformation Package SPIRE-UOL-DOC

Alignment and Preprocessing for Data Analysis

An Airborne A-Band Spectrometer for Remote Sensing Of Aerosol and Cloud Optical Properties

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Technical Bulletin on the Evaluation of the Kinetics Tuned Absorber/Diffuser Panel by Peter D'Antonio RPG Diffusor Systems, Inc Upper Marlboro, MD

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

How To Process Data From A Casu.Com Computer System

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July Reprinted from Photonics Spectra, Laurin Publishing

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Nonlinear Iterative Partial Least Squares Method

Galaxy Morphological Classification

Analytical Test Method Validation Report Template

Status of the Free Electron Laser

Introduction to X-Ray Powder Diffraction Data Analysis

Implementing and Using the EMVA1288 Standard

LANDSAT 8 Level 1 Product Performance

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

1 Genzyme Corp., Framingham, MA, 2 Positive Probability Ltd, Isleham, U.K.

Labs in Bologna & Potenza Menzel. Lab 3 Interrogating AIRS Data and Exploring Spectral Properties of Clouds and Moisture

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Measuring Line Edge Roughness: Fluctuations in Uncertainty

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

QUALITY CONTROL OF THE IUE FINAL ARCHIVE

A high accuracy, small field of view star guider with application to SNAP.

Quick Start Guide Chromeleon 7.2

Overview. What is EMR? Electromagnetic Radiation (EMR) LA502 Special Studies Remote Sensing

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

Learning Objectives:

Experiment 5. Lasers and laser mode structure

RightMark Audio Analyzer 6.0. User s Guide

Quick look tool for long-slit spectroscopy

MiSeq: Imaging and Base Calling

PDA Software Getting Started Guide

Module 13 : Measurements on Fiber Optic Systems

The Measurement of Sensitivity in Fluorescence Spectroscopy

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

Empower PDA Software. Getting Started Guide. 34 Maple Street Milford, MA , Revision A

Quick Start Instructions

Determining the Resolution of Scanned Document Images

Meteor Spectroscopy First Steps

2 Absorbing Solar Energy

How To Use An Ionsonic Microscope

Fig.1. The DAWN spacecraft

The Concept(s) of Mosaic Image Processing. by Fabian Neyer

Introduction to CCDs and CCD Data Calibration

Tutorial 9: SWATH data analysis in Skyline

Asteroid Compositions: Spectra S. K. Croft

COMPENSATION MIT Flow Cytometry Core Facility

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

RESULTS FROM A SIMPLE INFRARED CLOUD DETECTOR

FLUX / GOT-It Finite Element Analysis of electromagnetic devices Maccon GmbH

Convolution. 1D Formula: 2D Formula: Example on the web:

Making OSNR Measurements In a Modulated DWDM Signal Environment

Pesticide Analysis by Mass Spectrometry

Trace Gas Exchange Measurements with Standard Infrared Analyzers

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND

Transcription:

Instrument Science Report STIS 98-09 Extraction Heights for STIS Echelle Spectra Claus Leitherer and Ralph Bohlin March 1998 ABSTRACT The optimum extraction height (h) for STIS echelle spectra of 7 pixels maximizes the signal-to-noise for all modes and encompasses more than 98% of the target flux. Reducing h from 11 to 7 pixels reduces the error in the background correction by 40% at Lyman-α. This 7 pixel extraction height perpendicular to the dispersion should be the default for echelle extractions in the pipeline. 1. Introduction The trade-offs among different considerations for optimum extraction heights for firstorder STIS gratings are discussed in ISR STIS 97-13 (Leitherer & Bohlin 1997). The preferred slit heights for first-order modes are a compromise between spectral purity, signalto-noise, and photometric precision. Two additional issues must be addressed when extracting STIS echelle spectra. First, individual orders have a finite spatial separation, imposing a mode and order dependent upper limit on the maximum extraction height. Second, the E140M and E140H modes suffer from significant scattered light, whose contribution to the extracted spectrum increases with increasing extraction height. In this ISR we determine the optimum echelle extraction height and recommend the implementation of this height in the calibration pipeline. 2. Observations and Data Reduction The data which define the optimum extraction height are summarized in Table 1. The observations in this table were taken as part of several SMOV and Cycle 7 calibration programs. The target was a spectrophotometric standard star in all cases. The available data are sufficient to characterize all four STIS echelle modes (E140M, E140H, E230M, E230H) over the entire wavelength region of interest. 1

Table 1. Observation details Rootname Mode Aperture Wavelength Detector Target Prop ID O3ZX02X5Q E140M 0.2X0.2 1425 FUV-MAMA BD+28D4211 7096 O3ZX02XAQ E230M 0.2X0.2 1978 NUV-MAMA BD+28D4211 7096 O3ZX02XEQ E230M 0.2X0.2 2707 NUV-MAMA BD+28D4211 7096 O3ZX02XGQ E230M 0.2X0.2 2707 NUV-MAMA BD+28D4211 7096 O3ZX10DAQ E140H 0.2X0.2 1416 FUV-MAMA BD+75D325 7096 O3ZX10DEQ E230H 0.2X0.2 2513 NUV-MAMA BD+75D325 7096 O45930010 E140H 0.2X0.2 1416 FUV-MAMA BD+28D4211 7673 O45930020 E230H 0.2X0.2 2263 NUV-MAMA BD+28D4211 7673 O45930030 E140M 0.2X0.2 1425 FUV-MAMA BD+28D4211 7673 O45930040 E230M 0.2X0.2 1978 NUV-MAMA BD+28D4211 7673 O45930050 E230M 0.2X0.2 2707 NUV-MAMA BD+28D4211 7673 O47S01020 E140H 0.2X0.09 1307 FUV-MAMA BD+33D2642 7104 O47S01030 E140H 0.2X0.09 1307 FUV-MAMA BD+33D2642 7104 O4DD05020 E230H 0.2X0.2 2013 NUV-MAMA BD+75D325 7657 O4DD05030 E230H 0.2X0.2 1763 NUV-MAMA BD+75D325 7657 O4DD05040 E230H 0.2X0.2 2263 NUV-MAMA BD+75D325 7657 O4DD05050 E230H 0.2X0.2 2762 NUV-MAMA BD+75D325 7657 O4DD05060 E230H 0.2X0.2 3012 NUV-MAMA BD+75D325 7657 O4DD05070 E140H 0.2X0.2 1234 FUV-MAMA BD+75D325 7657 O4DD05080 E140H 0.2X0.2 1598 FUV-MAMA BD+75D325 7657 After sequentially performing the standard reduction steps to the raw data up to the flatfield corrections, the two-dimensional images are geometrically corrected. The onedimensional spectra are extracted and wavelength calibrated from 1024 x 1024 images using extraction heights of 3, 5, 7, 9, 11, 15, and 21 pixels. The corresponding background data are the inter-order count rates from the two-dimensional images after filtering and fitting with 3rd-order polynomials for each order. The background spectra are subtracted from the gross science spectra to produce the net spectra for each observation, order, and extraction height. There is no optimal extraction algorithm that could satisfy every observer s needs. Even for a point source, the optimal height of an extraction aperture depends on the PSF width, the target flux relative to the background, and the trade-off between flux conservation, maximum signal-to-noise, and spectral purity. As in the case for the first-order 2

gratings, a simple, robust extraction algorithm with few parameters is adequate for the scientific analysis of most point source data. 3. Spectral Purity of STIS Echelle Observations As with most other echelle spectrographs, STIS echelle data are affected by spectral order overlap at the shortest wavelengths where adjacent spectral orders are not well separated due to the degraded PSF in the UV. Some fraction of the spectral signal from the next higher (or lower) order is included in the background, introducing spectral impurity. In addition, the echelle data suffer from significant scattered light, especially at the shortest wavelengths. Figure 1 highlights the spectral purity issue. Plotted in this figure is the net count rate versus wavelength of BD+28D4211 observed with E140M at central wavelength 1425 Å. The wavelength region around Lyman-α is extracted from m = 36. Lyman-α is a black line in BD+28D4211 and should be at the zero level, independent of the extraction height. The actual net spectrum is overcorrected at Lyman-α, with the overcorrection increasing with larger extraction height. Figure 2 compares the gross, net, and background count rates of BD+28D4211 for a fixed extraction height of 7 pixels. Figure 2a is the region around Lyman-α. The fitted background spectrum amounts to about 25% of the peak rate of the gross spectrum and is higher than the count rate in the Lyman-α core. The net count rate becomes negative in the Lyman-α line core. The principal source for the overcorrection is scattered light, which becomes even more important for larger extraction height (dotted line for h = 11 in Figure 2a). Lyman-α bottoms out at 0.5 counts/s/p for h = 11, while the h = 7 extraction improves by 40% to 0.3 counts/s/p. The effects of scattered light are far less important for E230H. In Figure 2b we show the gross, net, and background count rates of BD+75D325 observed in the E230H mode for an extraction height of 7 pixels. The spectral region is around the Mg II doublet. The structure of Figure 2b is the same as 2a. The background shows no wavelength dependence and is small in comparison with the gross spectrum. The black Mg II lines are properly at 0 level after the background correction. A small extraction height minimizes the uncertainty in the background correction. In the future, post-processing software will become available to correct for the oversubtraction of the background due to scattered light. 4. Photometric Precision Although not the first priority for most echelle observers, photometric precision is a consideration for choosing the optimum extraction height. Figure 3 shows the net count rates of BD+28D4211 for extraction heights of 3, 5, 7, 9, 11, 15, and 21 pixels. The rates 3

are normalized by dividing by the values measured through the 11 pixels extraction aperture. Three modes and wavelength/order combinations are selected for inclusion in panels a), b), and c) to illustrate the typical behavior. Heights larger than 5 pixels result in throughputs of at least 90%, except at the extreme wavelengths for each order. The 21-pixel height extracts close to 100% of the source flux. However, the spectra extracted with h = 3 and 5 pixels have significant quasi-periodic noise. The results in Figure 3 argue in favor of an extraction height of at least 5 pixels. 5. Signal-to-noise versus Extraction Height for Continuum Sources The signal-to-noise initially increases with extraction height, as more target flux is extracted. After a critical height is reached, the signal-to-noise decreases, as more and more noise from the background is added. In Figure 4 we show the signal-to-noise versus extraction height for representative gratings and wavelengths. In each figure, the results for all orders of a particular grating and central wavelength combination are plotted after averaging over the central 200 pixels in the dispersion direction. The signal-to-noise values are normalized to the peak value in each order. As expected, a turn-over of the maximum signal-to-noise is seen. The location of the turn-over depends on the target signal relative to the background. In all studied cases, most of the background is due to scattered light, which is stronger at shorter than at longer wavelengths. Therefore, the drop-off in the signal-to-noise with h is more pronounced for the E140 modes and for the shortest wavelengths of each mode. The behavior of the curves in Figure 4 is independent of the target flux, as long as the dark noise is negligible. In the absence of dark noise, the background is predominantly due to scattered light, which is approximately a constant fraction of the source flux at each wavelength. The simulations demonstrate that extraction heights between 5 and 11 pixels retain at least 98% of the maximum signal-to-noise at all wavelengths for all modes. 6. Recommendations. The optimum extraction height for STIS echelle spectra of point sources depends on the amount of scattered light, the desired photometric precision, and the signal-to-noise. There is no height that would optimize all constraints simultaneously. Since our goal is to define a height that would fit most needs, and at the same time has as few variables as possible, our selection is guided by the requirement to maximize photometric precision and spectral purity. A height of 7 pixels for all echelle modes is acceptable as the default for point source extractions. An extraction height of 7 pixels is very close to the extraction height which maximizes the signal-to-noise for all modes and encompasses more than 4

98% of the target flux. The STScI pipeline will use h = 7 pixels for echelle extractions. GOs will be able to extract spectra for different heights with the x1d STSDAS task. 5

7. Figures Figure 1: Net count rate versus wavelength of BD+28D4211 observed with E140M at central wavelength 1425 Å (m = 122). The seven spectra plotted are extracted with heights of 3, 5, 7, 9, 11, 15, and 21 pixels. Larger extraction heights have larger count rates at the peak (1211 Å) and more negative rates in the Lyman-α absorption line. 6

Figure 2: a) Gross, net, and background count rate of BD+28D4211 around Lyman-α. The net count rate (spectrum reaching about 3 counts/s at 1211 Å) is obtained after subtracting the background (smooth curve peaking at about 1.3 counts/s) from the gross spectrum (top spectrum) for an extraction height of 7 pixels. For comparison, an extraction height of 11 pixels (dotted curve) gives an even stronger overcorrection of the Lyman-α absorption. b) Gross, net, and background count rate of BD+75D325 around the Mg II doublet. As in a), the net, gross, and background are shown for an extraction height of 7 pixels. At this wavelength the background is small; and the gross and net differ by a few percent only. The Mg II lines in the net spectrum are black, as expected. a) BD+28D4211 b) BD+75D325 7

Figure 3: Net count rates of BD+28D4211 for extraction heights of 3, 5, 7, 9, 11, 15, and 21 pixels. The rates are normalized by dividing by the values measured with the 11 pixels extraction height. a) Results for m = 107 of E140M around 1380 Å. b) Results for m = 104 of E230M around 1960 Å. c) Results for m = 346 of E230H around 2230 Å. a) m=107, E140M, 1380Å b) m=104, E230M, 1960Å c) m=346, E230H, 2230Å 8

Figure 4: Signal-to-noise versus extraction height for representative gratings and wavelengths. Each figure contains the results for all orders of a particular grating and central wavelength combination. All signal-to-noise values are normalized to the peak value found for each order. a) Results for BD+28D4211 and E140M at 1425 Å. b) BD+28D4211 and E230M at 1978 Å. c) BD+28D4211 and E230M at 2707 Å. d) BD+75D325 and E140H at 1234 Å. e) BD+75D325 and E140H at 1416 Å. f) BD+75D325 and E140H at 1598 Å. g) BD+75D325 and E230H at 2013 Å. h) BD+75D325 and E230H at 2762 Å. i) BD+75D325 and E230H at 3012 Å. a) b) c) d) e) f) g) h) i) 9