MRI QA Technologist s Tests



Similar documents
MRI QA Technologist s Tests

Breast MRI Quality Control

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

SITE IMAGING MANUAL ACRIN 6698

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

How To Improve Your Ct Image Quality

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

Phantom Test Guidance

Update on ACR Digital Mammography QC Manual

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

32-Channel Head Coil Imaging at 3T

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Quantitative Imaging In Clinical Trials Using PET/CT: Update

INTRODUCTION. A. Purpose

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Quality Control and Maintenance Programs

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

5 Factors Affecting the Signal-to-Noise Ratio

Prepublication Requirements

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Standardized MRI Protocol for Brain Tumor Clinical Trials. Benjamin M. Ellingson, Ph.D. Assistant Professor of Radiology at UCLA

Update on MQSA and ACR Mammography Accreditation

User Manual Microsoft Dynamics AX Add-on LabAX Label Printing

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Magnetic Resonance Imaging

Quality Control of Full Field Digital Mammography Units

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Current Industry Neuroimaging Experience in Clinical Trials Jerome Barakos, M.D.

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Performance testing for Precision 500D Classical R/F System

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Purchasing a cardiac CT scanner: What the radiologist needs to know

Assessing Radiation Dose: How to Do It Right

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

Atomic Force Microscope

Nuclear PACS: Problems, Solutions

Advanced MRI methods in diagnostics of spinal cord pathology

MGH Adult Diffusion Data Scanning Protocols

Education. October, 1985 GE Medical Systems Milwaukee, WI Signa Operators Training Program

Digital Mammography Update: Design and Characteristics of Current Systems

Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9

Super-Resolution Reconstruction in MRI: Better Images Faster?

Imaging Informatics Ann Scherzinger, PhD, CIIP C-TRIC Lecture Series, May 18, 2011

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

MQSA Quality Control Manual for Monochrome Displays for Mammography

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

The Dedicated MRI Factbook. Esaote MRI: A world of difference

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

WHERE IN THE WORLD JILL LIPOTI?

Physics testing of image detectors

Window Glass Design 5 According to ASTM E 1300

Sample Table. Columns. Column 1 Column 2 Column 3 Row 1 Cell 1 Cell 2 Cell 3 Row 2 Cell 4 Cell 5 Cell 6 Row 3 Cell 7 Cell 8 Cell 9.

Microsoft Excel 2010 Charts and Graphs

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute

Kap 8 Image quality, signal, contrast and noise

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Elfring Fonts, Inc. PCL MICR Fonts

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

High Spatial Resolution EPI Using an Odd Number of Interleaves

Automatic Wheel Profile Measuring Device for Tramway Vehicles

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint

Create Charts in Excel

The multitude of symptoms following a whiplash injury has given rise to much discussion because of the lack of objective radiological findings.

Three Dimensional Ultrasound Imaging

MRI DATA PROCESSING. Compiled by: Nicolas F. Lori and Carlos Ferreira. Introduction

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

MRI SEQUENCES. 1 Gradient Echo Sequence

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Musculoskeletal MRI Technical Considerations

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

Basic Specifications. Electrical. EPSON Perfection Color EPSON MatrixCCD TM line sensor. device Effective pixels

Annual MQSA Inspection Questions NMQAAC April 19, 2004

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Shorter Scanning Times Through Efficient Software. Annette Tuffs. Article from the customer magazine Medical Solutions, December 2010

QAV-PET: A Free Software for Quantitative Analysis and Visualization of PET Images

Technical handbook Panel Anchoring System

Failure Modes and Effects Analysis (FMEA)

Education, Training, and Certification of Medical Physicist in Korea

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

DIRECTIONS FOR SETTING UP LABELS FOR MARCO S INSERT STOCK IN WORD PERFECT, MS WORD AND ACCESS

Scatter Plots with Error Bars

Unwarping Echo Planar Images Using CMTK 1 Release 1.2

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

Specifying Laser Scanning Services. A Quantapoint White Paper

Merging Labels, Letters, and Envelopes Word 2013

Transcription:

AAPM 55th Annual Meeting MRI QA Technologist s Tests Guang Jia, PhD Associate Professor Department of Physics Louisiana State University

Background Outline QC importance Technologist s role QC test tips and problems Table OK? CF and TX Gain Phantom distance HR holes LCD spokes Additional QC tests Questions

Background Dr. Moriel S. NessAiver s experience 174 yearly performance tests (98 magnets over 3.5 years) 18 (10.3%) without deficiencies 19 (21.3%) with minor deficiencies, not affecting image quality 137 (78.7%) with deficiencies, directly affecting image quality 144 phased array coils (19.2%) with significant problems 22 systems (12.6%) with homogeneity problems 10-20% scanners: excessive RF noise, excessive ghosting, poor gradient calibration, poor hard copy (film), and soft copy performance 1 vendor s TSE with slice thickness 18-23% thicker than specified 1 vendor s TSE with slice thickness 20-25% thinner than specified Sobol WT, NessAiver MS, Orton CG, Med Phys 35(8)3419

Daily/Weekly QA Program

Technologist s Role

Record Book Examples Data form for weekly QC Template from ACR MRI QC Manual Examples

Record Book Examples QC visual checklist Template from ACR MRI QC Manual Examples

Table OK? Table OK? A test about the accuracy of the laser localizer Page 35 of ACR Manual 2004 version Frequently ignored by technologists Most technologists misunderstand that it is a check of whether the table move smoothly Laser Light

Table OK? on Siemens Scanner Procedure Open the sagittal localizer image User PixelLens to check the top edge of the vertical bars The location should be within ± 2mm < H(ead)2 or F(oot)2

Table OK? on GE Scanner Procedure Open the sagittal localizer image User Crosshair to check the top edge of the vertical bars The location should be within ± 2mm < S2 or I2

Table OK? Offline Measurement Philips EBW Open the sagittal localizer image User Crosshair to check the top edge of the vertical bars

Table OK? Offline Measurement Philips EBW Open the sagittal localizer image User Crosshair to check the top edge of the vertical bars

Magnetic Field Drift Definition Center frequency run down over time Drift reasons Windings are not perfect superconductors Eddy current interactions Drift rate < 0.01 ppm per hour < 107 Hz per week for 1.5T < 214 Hz per week for 3T Enshrined drift level 3 ppm per day (AAPM report 34, 1992) 1 ppm per day (AAPM report 100, 2010): 447 Hz/wk for 1.5T and 894 Hz/wk for 3T 1.5 ppm per day (ACR MRI QC guide, 2004) Corrective action Ask medical physicist or MRI scientist to re-check Ask service to monitor the units to ensure 1. B 0 field within RF transmit/receive frequency range 2. Draft rate does not increase

Central Frequency on GE scanner: 1 GE Signa Horizon 1.5T Place the phantom Click Auto Prescan before run the localizer sequence Record TG (transmit gain) and AX (central frequency)

Central Frequency on GE scanner: 2 GE Signa Horizon 1.5T Text page Series page Record central frequency and transmit gain

Central Frequency on Siemens scanner Procedure Before run a sequence, e.g. ACR T1 axial Options Adjustments Record central frequency and transmit gain

Distance Measurement: WW and WC Setting window and level/center To have the edges at the half-maximum value of the signal intensity Skipped by some Technologists First set window to 0 or 1 Adjust level/center about ½ white and ½ black, record level value width = recorded level value level/center = ½ recorded level value W = 606 C = 303 W = 1 C = 256 W = 256 C = 128 W = 768 C = 384 W = 1 C = 500 W = 500 C = 250

Distance Measurement: Sagittal Sagittal Image Avoid going through the black bar < 148 ± 2 mm

Distance Measurement: Problem Sagittal Loc Length 145.3 mm (Stoneridge) 146.4 mm (Morehouse) 147.0 mm (Siemens SSCBC)

Distance Measurement: Axial Axial slice #5 One vertical One horizontal < 190 ± 2 mm Examples:

Distance Measurement: Problem Vertical distance 194 mm (Siemens SSCBC 3T MRI) Out of spec Axial Curved border Axial Out of spec

Spokes GE Signa Horizon 1.5T Assign image 8 or 9? # of spokes on the image? Image 8 Image 9 Image 10 Image 11

Resolution GE Signa Horizon 1.5T (1.0) Rows 1 through 4 of the UL array are resolved (1.0) no columns or column 4 of the LR array could be resolved? ACR MRI Accreditation Program Phantom Test Guidance

Resolution: ZIP Zero filled interpolation (ZIP) ZIP512: zero fill up to 512x256 Better image resolution Cost 1. Slower reconstruction times 2. Longer to transfer 3. Takes up more disk space FT Unfiltered Image 256x256 k-space data 256x256 Courtesy of Dr. Moriel S. NessAiver

Resolution: ZIP Zero filled interpolation (ZIP) ZIP512: zero fill up to 512x256 Better image resolution Cost 1. Slower reconstruction times 2. Longer to transfer 3. Takes up more disk space FT Unfiltered Image 512x512 k-space data 512x512 Courtesy of Dr. Moriel S. NessAiver

Resolution: ZIP No ZIP UL: 1.0 LR: 1.1 Not passing ACR criteria with ZIP512 UL: 0.9 LR: 0.9 Passing ACR criteria

Common Problems Problems 1. Central frequency: 4 digits? (should be 9 at 3T) 2. Phantom distance: 3 digits? (should be 4) 3. LCD spokes: Always 10 spokes? (because the tech used slice #11, should use specified slice # 8) Slice 11 Slice 8 1 2 3

Additional Tests Advanced clinical MRI procedures MR spectroscopy Diffusion imaging Blood oxygen level dependent contrast (BOLD) imaging Angiographic and blood perfusion methods 1. DCE-MRI

Diffusion

DCE-MRI Purpose of the DCE-MRI Phantom Tests: Provide means for qualitative and quantitative evaluation, including T1 map error estimation, temporal resolution and image quality. Evaluate compliance with image acquisition protocols. ACRIN CQIE Body Phantom

20% 20% 20% 20% Question 1: Allowed weekly CF drift from ACR Manual? 20% 1. 0.1 ppm 2. 1.0 ppm 3. 1.5 ppm 4. 2.0 ppm 5. 2.5 ppm 10

20% 20% 20% 20% Question 2: Transaxial image measurements on slice #? 20% 1. 1 2. 3 3. 5 4. 11 5. Any slice 10

20% 20% 20% 20% Question 3: Table OK is for checking? 20% 1. Bed position and lights 2. Laser light alignment 3. Patient monitors 4. Horizontal bed movement 5. Vertical motion smoothness 10

References 1. Sobol WT, NessAiver MS, Orton CG, Point/counterpoint. The physics components of the ACR MRI Accreditation Program are overly tedious and beyond what is needed to ensure good patient care.med Phys 35(8)3419 2. ACR 2004 Magnetic Resonance Imaging Qaulity Control Manual 3. Ihalainen TM, Lönnroth NT, Peltonen JI, Uusi-Simola JK, Timonen MH, Kuusela LJ, Savolainen SE, Sipilä OE. MRI quality assurance using the ACR phantom in a multi-unit imaging center. Acta Oncol. 2011 Aug;50(6):966-72 4. Chen CC, Wan YL, Wai YY, Liu HL. Quality assurance of clinical MRI scanners using ACR MRI phantom: preliminary results. J Digit Imaging. 2004 Dec;17(4):279-84. 5. NessAiver, Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period, MRI QA RSNA 2011. 6. AAPM Report No. 110, Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities, 2010 7. AAPM Report No. 34, Acceptance Testing of Magnetic Resonance Imaging Systems (Reprinted from Medical Physics, Vol. 19, Issue 1)

Acknowledgements Department of Radiology Rick Layman Brian Raterman Xiangyu Yang Georgeta Mihai Jun Zhang