Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan



Similar documents
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

How To Test For Contamination In Large Volume Water

Introduction. Methods. Sample Processing

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Gas Chromatography Liner Selection Guide

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

A Novel Approach to Low Volume Sample Preparation

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Oasis HLB Cartridges and 96-Well Plates

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

Overview. Purpose. Methods. Results

A New QuEChERS Dispersive SPE Material and Method for Analysis of Veterinary Drug Residue by LC-MS-MS

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Determination of Tetracycline Antibiotics in Milk Using a Simple Strong Cation-Exchange SPE Cleanup Procedure and LC-MS/MS Analysis

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

METHOD OF ANALYSIS. N methyl 2-pyrrolidone

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

A Complete Solution for Method Linearity in HPLC and UHPLC

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

A Determination of Veterinary Drug Residues in Milk Using Polymeric SPE and UHPLC-MS/MS Analysis

A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System

ApplicationNOTE Introduction

High-Throughput 3-D Chromatography Through Ion Exchange SPE

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

LC-MS/MS, the new reference method for mycotoxin analysis

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

Development and Validation of Methods for Chemotherapy Drugs on the New Prelude SPLC LC-MS/MS System

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Innovative vs Traditional

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

CORESTA RECOMMENDED METHOD N 72

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

New drugs of abuse? A case study on Phenazepam & Methoxetamine

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Rapid LC/MS/MS Method Development for Drug Discovery

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France

S olid Ph ase Extraction (SP E) Method:

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System

High Sensitivity MS Determination of Carryover in a New Autosampler Design

ORGANIC SAMPLE PREPARATION

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

LC-MS/MS for Chromatographers

Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES

Supplementary Materials and Methods (Metabolomics analysis)

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

Thermo Scientific Mass Spectrometric Immunoassay (MSIA)

How To Develop A Recipe Card

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS

Optimizing the Post-implementation Monitoring of a LC-MS/MS Method

Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Transcription:

Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract A liquid chromatography tandem mass spectrometry method for the analysis of dextromethorphan and dextrorphan from rat plasma has been developed. A limit of quantification of 0.1 ng/ml is readily achieved. When using Thermo Scientific SOLA CX cartridges or plates, sample preparation is fast and reproducible giving excellent precision, low matrix suppression, and good recovery. The analysis was carried out on a Thermo Scientific Accucore C18 2.1 mm HPLC column for a fast separation with a cycle time of 2 minutes and good peak shape. Introduction (Figure 1) is used as a cough suppressant and can be found in many over-the-counter cough and cold remedies. is quickly metabolized to dextrorphan (Figure 2) in the liver. This process can be used to differentiate fast and slow metabolizers, which can be useful information when starting metabolism drug research. The extraction from plasma using SOLA solid phase extraction (SPE) products and LC analysis of dextromethorphan and dextrorphan are demonstrated in this application. SOLA SPE products introduce next-generation, innovative technological advancements, which give unparalleled performance characteristics compared to conventional SPE, phospholipid, and protein precipitation products. Figure 1. These include: Higher levels of reproducibility Higher levels of extract cleanliness Reduced solvent requirements Increased sensitivity SOLA SPE plates or cartridges have significant advantages when analyzing compounds in complex matrices, particularly in high-throughput bioanalytical and clinical research laboratories where reduced failure rates, higher analysis speed, and lower solvent requirements are critical. SOLA products superior performance gives higher confidence in analytical results and lowers cost without compromising ease-of-use or requiring complex method development. Figure 2.

2 Accucore HPLC columns use Core Enhanced Technology TM to facilitate fast and highly efficient separations. The 2.6 µm diameter particles are not totally porous, but instead have a solid core and a porous outer layer. The optimized phase bonding creates a series of high coverage, robust phases. The carbon loading of Accucore C18 columns provides high retention of non-polar analytes via a predominantly hydrophobic interaction mechanism. The tightly controlled 2.6 µm diameter of Accucore particles results in much lower backpressures than typically seen with sub-2 µm materials. Experimental Details Sample Handling Fisher Scientific TM LC-MS grade water W/011217 Fisher Scientific LC-MS grade methanol M/62/17 Fisher Scientific LC-MS grade acetonitrile A/0626/17 Fisher Scientific Analytical grade formic acid F/1900/PB08 Fisher Scientific HPLC grade ammonia solution A/3295/PB05 Thermo Scientific TM National TM Mass Spec Target DP Certified 2 ml clear vial with ID patch, MSCERT00-34W blue DP cap with bonded PTFE/silicone septum Sample Handling Equipment Thermo Scientific 96 well plate vacuum manifold 103-351 Thermo Scientific TM UltraVap TM CLS-229070 Sample Pretreatment spiking solutions of dextromethorphan and dextrorphan were prepared in acetonitrile. A working internal standard solution (dextromethorphan-d 3 ) was prepared in acetonitrile. Spiking solutions for each level of calibrant were made in acetonitrile. 1 μl of blank plasma or sample was taken and added to a clean vial. For standards and quality control (QC) samples, 10 μl of standard spiking solution was added, with 10 μl of acetonitrile added for blanks. For standards and QCs, 10 μl of working internal standard solution was added. For blanks, 10 μl of acetonitrile was added. All samples were vortexed for 30 seconds and 0 μl of 0.1% formic acid in water was added. All samples were vortexed for 30 seconds and centrifuged at 14,000 rpm for 5 min. Sample Preparation Compound(s):, dextromethorphan-d 3, and dextrorphan Matrix: Plasma Plate type: Thermo Scientific SOLA CX 309-002 Conditioning stage: Apply 0 μl of methanol, then 0 μl 0.1% formic acid in water to the SPE plate Application stage: Washing stage: Elution stage: Additional stage: Apply all supernatant to the SPE plate at a flow rate of 0.5 ml/min Apply 0 μl of methanol / water (: v/v) to the SPE plate Apply 4 2 μl 5% ammonia in methanol to the SPE plate and dry well Dry down under nitrogen and reconstitute in 0 μl acetonitrile / water (: v/v). Mix well.

Separation Conditions Instrumentation: Thermo Scientific TM Dionex TM UltiMate TM 3000 RSLC HPLC System Column: Accucore C18 2.6 μm, 2.1 mm 17126-052130 Mobile phase A: Mobile phase B: Water + 0.1% formic acid Acetonitrile + 0.1% formic acid Gradient: Time (min) %B Flow rate: 0 5 1 95 1.01 5 2 5 1.4 ml/min Column temperature: C Pressure: Injection details: 3 Bar 2 μl 3 MS Conditions Instrumentation: Thermo Scientific TM TSQ Vantage TM MS Ionization conditions: HESI Polarity: Positive Spray voltage (V): 00 Vaporizer temperature ( C): 4 Sheath gas pressure (Arb): Aux gas pressure (Arb): Capillary temp ( C): 300 Collision pressure (m Torr): 1.5 Q1 (FWHM): 0.7 Q3 (FWHM): 0.7 Compound transition details are provided in Table 1. Compound -d 3 Parent (m/z) 272.2 258.2 275.2 Products (m/z) 215.2 157.1 215.1 Collision energy (V) 22 37 22 S Lens (Arb) 76 84 77 Table 1. Compound transition details Data Processing Software: Thermo Scientific TM LC QUAN TM 2.6

4 Results Chromatography The Accucore C18 HPLC column gave excellent peak shape. The chromatography of the QCM at 2.5 ng/ml is shown in Figure 3. Relative Abundance 90 70 30 10 0 90 70 2 30 10 0 90 70 30 2 0 1 1 1 1 10 0 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5 Time (min) Figure 3: Representative chromatogram of dextromethorphan and dextrorphan SRM, extracted from plasma at 2.5 ng/ml 0 10 30 70 90 110 Linearity s of dextromethorphan and dextrorphan extracted from spiked rat plasma gave a linear calibration curve over the dynamic range of 0.1 to ng/ml with an r 2 of 0.998 and 0.999, respectively (Figures 4 and 5 and Table 2). Area Ratio 2 2 5 0 0 1 4 1 0 1 3 Area Ratio 1 Area Ratio 300 2 0 1 0 10 30 70 90 110 Figure 4: linearity over the dynamic range 0.1 to ng/ml 0 10 30 70 90 110 Figure 5: linearity over the dynamic range 0.1 to ng/ml 5 0 4

Accuracy Eight standards were run over the linear range 0.1 to ng/ml. The accuracy of the back calculated values was 14% in all cases. 5 Specified Concentration (ng/ml) Calculated Concentration (ng/ml) % Diff Calculated Concentration (ng/ml) % Diff S1 0.1 0.101 1 0.090-10 S2 0.2 0.4 2 0.228 14 S3 0.5 0.4-4 0.3 1 S4 1 1.10 10 0.904-10 S5 5 1.73-5 5.06 1 S6 10 9.86-1 10.3 3 S7 47.2-6.4 1 S8 103 3 99.2-1 Table 2: Accuracy data for eight extracted dextromethorphan and dextrorphan standards over the linear range 0.1 to ng/ml Precision QC samples were run in replicates of six at concentrations of 0.25, 3, and 25 ng/ml. The precision of the QC level was 8% CV in all cases (Table 3). Concentration (ng/ml) Average Calculated Concentration (n=6) % CV Average Calculated Concentration (n=6) % CV QCL 0.25 0.236 7 0.236 8 QCM 3 2.67 3 3.08 7 QCH 25 24.5 8 27.4 8 Table 3: Average precision data for six replicate QCs at three levels for dextromethorphan and dextrorphan Recovery Overspikes (post extracted fortified blanks) were run in triplicate at concentrations of 0.25, 3, and 25 ng/ml and used to calculate the percentage recovery level for dextromethorphan and dextrorphan (Table 4). The recovery of detromethorphan was 95.4% and that of dextrorphan was 101.6%. The recovery proved to be consistent throughout the batch for all compounds. Response Ratio % Recovery Average % Recovery Average QCL response ratio 0.596 Average QCL overspike response ratio 0.594 Average QCM response ratio 5.45 Average QCM overspike response ratio 5.71 Average QCH response ratio 49.8 Average QCH overspike response ratio 54.2 Average QCL response ratio 1.18 Average QCL overspike response ratio 1.15 Average QCM response ratio 15.1 Average QCM overspike response ratio 14.8 Average QCH response ratio 134 Average QCH overspike response ratio 134 98.8 95.6 91.9 102.9 102.0 99.8 95.4 101.6 Table 4: Recovery data for dextromethorphan, dextrorphan and dextromethorphan-d 3

Matrix Suppression Overspikes and solution standards were run in triplicate at concentrations of 0.25, 3, and 25 ng/ml and used to calculate the percentage matrix suppression for dextromethorphan and dextrorphan (Table 5). There was very little matrix suppression: -0.5% and 7.4% for dextromethorphan and dextrorphan, respectively. The matrix effects were consistently low throughout the batch for all compounds. Response Ratio Average QCL overspike response ratio 0.594 Average QCL solution response ratio 0.5 Average QCM overspike response ratio 5.71 Average QCM solution response ratio 5.72 Average QCH overspike response ratio 54.2 Average QCH solution response ratio 56.5 Average QCL overspike response ratio 1.15 Average QCL solution response ratio 1.34 Average QCM overspike response ratio 14.8 Average QCM solution response ratio 15.1 Average QCH overspike response ratio 134 Average QCH solution response ratio 142 % Matrix Suppression -5.9 0.2 4.2 14.6 2.3 5.4 Average % Matrix Suppression -0.5 7.4 Application Note 685 Table 5: Matrix suppression data for dextromethorphan, dextrorphan and dextromethorphan-d 3 Conclusion SOLA CX SPE plates and Accucore C18 HPLC columns allow for simple extraction and rapid quantification of dextromethorphan and dextrorphan from plasma. Accucore C18 HPLC columns give a fast runtime for dextromethorphan and dextrorphan of 2 min. Accucore C18 HPLC columns give excellent peak shape for dextromethorphan and dextrorphan. An LLOQ of 0.1 ng/ml was achieved. Extraction recovery was 95.4% and 101.6% for dextromethorphan and dextrorphan, respectively. SOLA CX SPE plates achieved excellent precision on the extraction of dextromethorphan and dextrorphan from plasma samples at low concentrations with % CV (n=6) 8% SOLA CX SPE plates achieved low matrix recovery. This gave excellent matrix suppression effects at 7% and 12% for dextromethorphan and dextrorphan, respectively. For Research Use Only. thermoscientific.com/chromatography 13 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. USA and Canada +1 0 332 3331 France +33 (0)1 92 48 34 Germany +49 (0) 2423 9431 or 21 United Kingdom +44 (0)1928 534110 Japan +81 3 5826 1615 AN685_E 01/13S China +86 21 68654588 +86 10 84193588 +86 83145199 0 810 5118 India +91 22 6742 9494 +91 27 1766 2352 Australia 1 300 735 292 (free call domestic) New Zealand 00 933 966 (free call domestic) All Other Enquiries +44 (0) 1928 534 0 Technical Support North America +1 0 332 3331 Outside North America +44 (0) 1928 534 4