Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES
|
|
|
- Andrew Chapman
- 10 years ago
- Views:
Transcription
1 Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES
2 AGILENT ENHANCED MATRIX REMOVAL LIPID ARE LIPIDS WEIGHING YOUR SAMPLES DOWN? Interference from lipids is a problem for labs measuring trace residues in fatty foods or complex biological matrices. Lipids can build up in the instrument and column, decreasing lifetime and reducing analyte sensitivity due to ion suppression. The need for MS maintenance increases too, because of lipid deposits on the source. The need for lipid removal is well understood, but current methods often sacrifice analyte recovery, removing some of your target analytes along with the lipids. Now, you don t have to choose between lipid removal and analyte recovery, because innovative Agilent Enhanced Matrix Removal Lipid delivers the most complete matrix removal and analyte recovery of any sample prep product. Enhanced Matrix Removal Lipid, UNLIKE any other type of sample prep, is a unique sorbent that selectively removes lipids in complex matrices and challenging high-fat samples such as avocado, so you can remove lipids without losing your analytes. x Avocado Quechers extract before dspe Avocado Quechers extract with C8/PSA cleanup Avocado Quechers with EMR Lipid cleanup 9% Matrix Removal with EMR Lipid vs. acquisition time (min) Figure. GC/MS full-scan chromatogram overlay of an untreated QuEChERS avocado extract (black), compared to traditional C8/PSA cleanup (red) and an extract treated with Agilent Enhanced Matrix Removal Lipid (blue).
3 STANDARD QuEChERS Extraction/partition (6 steps) Mix and centrifuge Transfer to dspe (sorbents) Mix and centrifuge Evaporate/reconstitute or dilute Filter out precipitate Transfer to A/S vial EMR LIPID QuEChERS Extraction/partition (6 steps) Mix and centrifuge Transfer to dspe (EMR Lipid & HO) Mix and centrifuge Polish* Evaporate/reconstitute or dilute Transfer to A/S vial The innovative sorbent in EMR Lipid replaces C8/PSA in your QuEChERS methodology to significantly reduce matrix effects and improve analyte recoveries. It can be universally applied to the analysis of polar, mid-polar and non-polar target analytes, providing effective matrix removal. *Note: Polish step is omitted for tetracycline analysis. Sample Type Removal (%) Avocado 9 Olive oil 73 Avocado oil 8 Canola oil 7 Beef liver 97 Wet cat food 9 Paprika 9 Cumin 7 Hops 7 Tumeric 47 Black pepper Onion 4 Strawberry Plum Red pepper 6 Spinach 6 Orange Apple 66 3
4 EMR LIPID MINIMIZE LIPIDS. MAXIMIZE RESULTS. Analyte recovery % (n=6) in ppb pesticides fortified avocado Methamidophos EMR Lipid cleanup C8/PSA dspe cleanup Zirconia sorbent cleanup Thiophanate methtyl Bentazone,4-D Acid Dichlorprop Penconazole TPP EPN Captafol EMR Lipid significantly removes matrix contributions, improving recovery across many analytes and minimizing your need for separate workup and methods. Figure 3. Recovery data for multiple pesticide residue analysis in avocado, showing some of the 73 pesticides analyzed. Using Agilent Enhanced Matrix Removal Lipid, 9% of the pesticides fell within the acceptable recovery range of 7 to %. Analytes Recovery % (n=6) in ppb vet drugs fortified in bovine liver samples 3 REC<7% REC in 7 % REC> Higher efficiency matrix removal with EMR Lipid results in higher precision and greater accuracy in results. Figure 4. Number of analytes with % recoveries within the acceptable % recovery range of 7-% for the analysis of vet drugs in Bovine Liver, demonstrating the higher number of acceptable results after Agilent Enhanced Matrix Removal Lipid Cleanup vs zirconia sorbent. EMR Lipid cleanup C8/PSA cleanup Zirconia sorbent cleanup 4
5 Chromatographic benefits of sample preparation using EMR Lipid QuEChERS Efficiently removing lipids from samples without removing analytes is the key to improving chromatographic performance for the best quality data, especially with high-sensitivity MS detectors. An EMR Lipid QuEChERS protocol considerably improves accuracy, reproducibility, and low-level quantitation when using GC/MS and LC/MS (Figure ). BENEFIT BETTER ANALYTE SENSITIVITY IMPROVED S/N RATIO LESS INTERFERENCES FOR ACCURATE INTEGRATION Example EPN in avocado on LC-QQQ Captan in avocado on GC-QQQ Permethrin in avocado on GC-QQQ EMR Lipid cleanup Zirconia sorbent cleanup C8/PSA cleanup x 3.. x 3.. x 3.. +MRM (34. -> 96.) AVBVR POQ4.d min Acqusition Time (min) +MRM (34. -> 96.) AVZ-Sep+ POQ 4.d min MRM (34. -> 96.) AVC8 PR-Q3.d 4.8 min x 4.6 +MRM (. -> 79.)AV MBPI-a-CD C spike ppb-.d 3.63 min x x MRM (. -> 79.) AV Z-Sep+ C spike ppb-.d MRM (83. -> 68.)AV x 4 MBPI-a-CD C spike ppb-.d.788 min Figure. Improvements in chromatography can be seen in the analysis of samples prepared with Agilent Enhanced Matrix Removal Lipid, including better sensitivity due to reduced matrix ion suppression, better signal-to-noise ratio, and fewer interferences providing accurate integration with GC or LC triple quadrupole systems min. +MRM (. -> 79.) Post-S ppb Fatty cspe C-.D *3.74 min MRM (83. -> 68.) AV MZ-Sep+ C spike ppb-.d.89 min. x MRM (83. -> 68.) AV Fatty dspe spike ppb-r.d.833 min. x
6 FASTER SAMPLE PREP STREAMLINES LAB PRODUCTIVITY Without effective sample cleanup, analytical efficiency and quality are quickly compromised as fatty matrices build up in the instrument and column. With EMR Lipid, system performance is maintained even after injections of a fatty-matrix sample like avocado. Reduced maintenance allows for increased sample throughput and enables you to use your assets to their fullest extent, maximizing productivity in your lab TPP peak area Figure 6. Analyte response across avocado sample injections. Triphenyl phosphate (TPP) is a commonly used internal standard. Signal suppression or enhancement can result from insufficient cleanup of samples, which can contribute to poor data quality, errors, and re-running samples. The superior EMR Lipid cleanup. Response drops 4% Zirconia sorbent cleanup. Response drops % C8/PSA cleanup. Response increases 3% matrix removal ability of Agilent Enhanced Matrix Removal Lipid, results in a cleaner source and more consistent MS response over time, for higher quality data, fewer re-runs, and reduced need for time-consuming troubleshooting and source maintenance Injections of avocado extract samples on GC/QQQ Relative abundance for tuning ion (%) PARAMETER EM voltage Rel. abundance MS ion 9 Rel. abundance MS ion METHOD PERFORMANCE CRITERIA >6V needs EM replacement Min 7 Min EM Voltage With EMR Lipid there is less need for system maintenance and calibration due to fewer sample matrix interferences. As a result, fewer samples need to be re-run. This advantage lets you operate at higher throughput, resulting in cost savings and a more efficient lab. Figure 7. GC/MS source conditions during injections of avocado samples, demonstrating significantly reduced frequency of MS source maintenance (cleaning or replacement) when using Agilent Enhanced Matrix Removal Lipid. Injections of avocado samples after QuEChERS EMR Lipid cleanup 6
7 Analytes RSD over injections on GC/QQQ (n = ) Pesticides EMR Lipid Cleanup C8/PSA Cleanup Zirconia sorbent Cleanup Dichlorvos Phenylphenol Ethalfluralin Sulfotep Atrazine Lindane Chlorothalonil Diazinon Chlorpyriphosmethyl Dichlofluanid Aldrin Tolylfluanid Captan Procymidone Bupirimate Endrin Endosulfan sulfate DDT Iprodione Permethrin Parathion ethyl TPP Over a hundred runs of fatty sample analysis, the consistently better, effective clean up of EMR Lipid results in lower % RSDs and greater confidence in results, compared to alternative QuEChERS procedures. Higher quality data with better precision results in less need for data re-verification, justification and fewer costly re-runs. Figure 8. Reproducibility of pesticides analyzed (%RSD) over multiple injections of avocado samples on GC/QQQ. ppb fortified avocado samples were tested every five injections, and avocado matrix blanks were injected in between. RSD > % are considered as unacceptable and marked in red. The benefits of using new and innovative EMR Lipid for QuEChERS are clear Improved productivity: better sensitivity and signal-to-noise from fewer matrix interferences enable faster data processing, and greater sample throughput. Reduced lab costs: cleaner samples using EMR Lipid can offer significantly reduced MS source maintenance, giving you more time to analyze samples rather than spend time on costly troubleshooting, downtime, and instrument repair. Simplified workflows: standardize on an easy-to-use single-sorbent procedure that maximizes analyte recovery from a wide variety of fatty samples. Save time and money by reducing material costs, inventory, training time, and documentation, thereby streamlining your lab productivity. Higher quality results: a cleaner sample profile leads to greater data integrity and confidence in results, faster data processing, and fewer re-runs. Part Number Description 98- Bond Elut QuEChERS EMR Lipid dspe 98- Bond Elut QuEChERS EMR Lipid Final polish 7
8 The best analysis requires the best overall workflow Explore more food testing solutions at Increase analysis throughput with current instruments and minimal investment check how to increase your laboratory productivity here: Liquid chromatography analysis workflow agilent.com/chem/productivitylc Gas chromatography analysis workflow agilent.com/chem/productivitygc Learn more agilent.com/chem/emr-lipid Find a local Agilent customer center USA and Canada [email protected] Europe [email protected] Asia Pacific [email protected] India [email protected] For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., Published in the USA, July 7, 99-6EN
Multiresidue Analysis of Pesticides in Avocado with Agilent Bond Elut EMR Lipid by GC/MS/MS
Multiresidue Analysis of Pesticides in Avocado with Agilent Bond Elut EMR Lipid by GC/MS/MS Application Note Food Testing and Agriculture Authors Limian Zhao and Derick Lucas Agilent Technologies, Inc.
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
Multiresidue Analysis of Pesticides in Avocado with Agilent Bond Elut EMR Lipid by LC/MS/MS
Multiresidue Analysis of Pesticides in Avocado with Agilent Bond Elut EMR Lipid by LC/MS/MS Application Note Food Testing and Agriculture Authors Limian Zhao and Derick Lucas Agilent Technologies, Inc.
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Multi Pesticides Residue analysis in Ayurvedic cough Syrup by GCMS/MS using QuEChERS extraction method
PO-CON679E Multi Pesticides Residue analysis in Ayurvedic cough Syrup by GCMS/MS using QuEChERS extraction method ASMS 206 ThP 47 Durvesh Sawant (), Ankush Bhone (), Dheeraj Handique (), Prashant Hase
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Application Note Food Testing & Agriculture Authors Edgar Naegele and Thomas
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference
Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas
[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS. Simplifed
[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS Simplifed The QuEChERS method (an acronym for Quick, Easy, Cheap, Effective, Rugged, and Safe) followed by dispersive solid-phase extraction
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
A New QuEChERS Dispersive SPE Material and Method for Analysis of Veterinary Drug Residue by LC-MS-MS
A New QuEChERS Dispersive SPE Material and Method for Analysis of Veterinary Drug Residue by LC-MS-MS Olga Shimelis, Michael Ye, and Len Sidisky Supelco, Div. of Sigma-Aldrich, Bellefonte, PA 16823 USA
Automation of Solid Phase Extraction and Column Chromatographic Cleanup
Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup
Application Note 25570306 Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup Keywords Gel Permeation Chromatography GPC AutoPrep 2000 Pesticides RapidVap N2 System
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
AppNote 1/2008. Automated Disposable Pipette Extraction of Pesticides from Fruits and Vegetables KEYWORDS ABSTRACT
AppNote 1/2008 Automated Disposable Pipette Extraction of Pesticides from Fruits and Vegetables William E. Brewer, Hongxia Guan, Stephen L. Morgan The University of South Carolina, Columbia, SC, USA Fred
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
The Chemistry of QuEChERS. Quick, Easy, Cheap, Effective, Rugged and Safe Method for Determining Pesticide Residues
The Chemistry of QuEChERS Quick, Easy, Cheap, Effective, Rugged and Safe Method for Determining Pesticide Residues 1 Fruits and Vegetables 2 Milk and Honey 3 Tobacco 4 Animal Tissues 5 Food Analysis QuEChERS
Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
Pesticide residue analysis: New Trends. Michal Godula, Ph.D. EU Marketing Manager Food Safety & Environmental Thermo Fisher Scientific
Pesticide residue analysis: New Trends Michal Godula, Ph.D. EU Marketing Manager Food Safety & Environmental Thermo Fisher Scientific Why pesticide analysis? Extensively used in agriculture to control
Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS
Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS AGILENT 6000 SERIES LC/MS SOLUTIONS UNRIVALED ACHIEVEMENT AT EVERY LEVEL The Agilent 6000 Series of LC/MS instruments
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.
Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class
Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process
- Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM
DRUG METABLISM Drug discovery & development solutions FR DRUG METABLISM Fast and efficient metabolite identification is critical in today s drug discovery pipeline. The goal is to achieve rapid structural
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS
AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,
LC-MS/MS, the new reference method for mycotoxin analysis
LC-MS/MS, the new reference method for mycotoxin analysis What is necessary for the implementation of the multi-residue mycotoxin method in an analytical laboratory? Azel Swemmer [email protected] AFMA
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
Sample Preparation for Forensic Toxicology and Clinical Research. Part 1: Sample Preparation Options
Sample Preparation for Forensic Toxicology and Clinical Research Part 1: Sample Preparation Options Agenda E-seminar Series Introduction Challenges for Today s Forensic Toxicology Laboratories Agilent
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
A Novel Approach to Low Volume Sample Preparation
A Novel Approach to Low Volume Sample Preparation Tony Edge, Jon Bardsley, Ken Meadows, Luisa Pereira International Symposium on Chromatography Salzburg, Austria September 2014 1 The world leader in serving
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment
Tracer Detection Systems
Tracer Detection Systems WASSON ECE Engineered Solutions, Guaranteed Results. Wasson-ECE Tracer Analysis System The industry standard for low level perfluoronated tracer analysis. Tracer tagging technology
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
aurora ICP-MS Innovation with Integrity The Science of Sensitivity ICP-MS
aurora ICP-MS The Science of Sensitivity Innovation with Integrity ICP-MS Powers the Business of Quantification aurora ICP-MS enhanced value package Bruker s ICP-MS is famous for its innovative and unique
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs
AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide
chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo
AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS
AppNote 1/2012 Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS Oscar G. Cabrices, Fred D. Foster, John R. Stuff, Edward
How To Analyze Pesticide Levels In Lemon Oil
AppNote 9/2006 Target GC-MS Analysis using Accelerated Column Heating and Interactive Deconvolution Software Albert Robbat Tufts University, Chemistry Department, Medford, MA 02155, USA Andreas Hoffmann
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
Signal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
AppNote 9/2013. Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides KEYWORDS ABSTRACT
AppNote 9/2013 Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides Oscar G. Cabrices Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum, MD 21090, USA
Agilent LCMS products and the application strategies in veterinary drug analysis
Agilent LCMS products and the application strategies in veterinary drug analysis Experience A New Level of Confidence Ma Xin 马 鑫 LCMS Application Engineer Internal Sales Training - The new 6495 QQQ LC/MS
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software
Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,
Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report
PRODUCT BULLETIN Cliquid ChemoView 3.0 Software for Routine Screening and Quantitation Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report KEY FEATURES Secure user login that
Innovative vs Traditional
And in the red corner, introducing the challenger NASDAQ April 4 2011 243.47 $ 1.38 $ 0.58 % 0.07 $ 0.02 $ 39.4 % Technology DVD Digital Streamline Instant Play Internet Access Unlimited Selection No Late
AxION edoor. Web-Based, Open-Access Mass Spectrometry Software
AxION edoor Web-Based, Open-Access Mass Spectrometry Software 2 AxION edoor is the open-access software solution for today s fast-paced laboratory. It is designed to operate and optimize the management
S olid Ph ase Extraction (SP E) Method:
Solid Phase Extraction of US EPA 535 for Chloroacetanilide and Acetamide Degradates in Drinking Water Samples Prior to Liquid Chromatography/ Tandem Mass Spectrometry Toni Hofhine, Horizon Technology,
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS
EKSIGENT EKSPERT NANOLC 4 Unmatched flexibility for low flow LC/MS EKSIGENT EKSPERT NANOLC 4 Focused ekspertise In 23, Eksigent changed the way scientists looked at nanoscale chromatography with the introduction
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet
TargetQuan 3 Software Leading the way in regulatory POPs quantification Bullet Bullet Bullet Leading the way in regulatory POPs quantification Analyse Samples Open Sequence Process Sequence Analyse Samples
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS
Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS Thorsten Bernsmann, Peter Fürst Chemical and Veterinary Analytical Institute Münsterland-Emscher-Lippe Joseph-König-Str. 40,
MassHunter for Agilent GC/MS & GC/MS/MS
MassHunter for Agilent GC/MS & GC/MS/MS Next Generation Data Analysis Software Presented by : Terry Harper GC/MS Product Specialist 1 Outline of Topics Topic 1: Introduction to MassHunter Topic 2: Data
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative