Raman vs. FTIR Process Analyzers for In-Line Blending Use



Similar documents
FTIR Instrumentation

Introduction to Fourier Transform Infrared Spectrometry

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Industrial Process Monitoring Requires Rugged AOTF Tools

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

Components for Infrared Spectroscopy. Dispersive IR Spectroscopy

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Raman Spectroscopy Basics

Raman Spectroscopy. 1. Introduction. 2. More on Raman Scattering. " scattered. " incident

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

13C NMR Spectroscopy

Signal to Noise Instrumental Excel Assignment

Back to Basics Fundamentals of Polymer Analysis

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

Trans Fats. What is a trans fat? Trans fatty acids, or trans fats as they are known, are certain

Fibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques

Experiment #5: Qualitative Absorption Spectroscopy

ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.

Raman spectroscopy Lecture

UV-Vis spectrophotometers

Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Measurement & Analytics Measurement made easy

Time out states and transitions

Uses of Derivative Spectroscopy

IR READY TOGO. Spectrum Two FT-IR Spectroscopy

Mozzarella Process Analysis Get more out of your production with High Resolution in-line analysis. ProFoss. Dedicated Analytical Solutions

Validation of a FTIR Spectroscopy Method for Measuring and Monitoring Fuel Dilution in Lubricating Oils

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH

Modern Classical Optics

Realization of a UV fisheye hyperspectral camera

High Voltage Power Supplies for Analytical Instrumentation

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

Effective Process Planning and Scheduling

UVC LEDs for. Environmental Monitoring

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

MB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC

Principle of Thermal Imaging

Background Information

Infrared Spectroscopy 紅 外 線 光 譜 儀

A More Efficient Way to De-shelve 137 Ba +

Spectrum 10 Spectroscopy Software

Spectral Measurement Solutions for Industry and Research

Spectroscopy and Regions of the Spectrum

Chemometric Analysis for Spectroscopy

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

SGS: Das Scanning Grating Spektrometer Ein kleines, günstiges Spektrometermodul auf Basis eines dispersiven Mikrosystems

Qualitative NIR Analysis for Ingredients in the Baking Industry

Oil Analysis. It is important that the sampling be done in the following manner: 1.

MB3600-HP12 / Measurement & Analytics. Laboratory FT-NIR Heavy Oils Analyzer Crude oil, fuel oil, heavy oil upgrader unit feeds

Calibration of a High Dynamic Range, Low Light Level Visible Source

3 - Atomic Absorption Spectroscopy

Ultraviolet Spectroscopy

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Using the Spectrophotometer

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

White Paper. Chemical dosing and blending - the role of precision actuators

Multivariate Chemometric and Statistic Software Role in Process Analytical Technology

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

EDXRF of Used Automotive Catalytic Converters

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION

The Measurement of Sensitivity in Fluorescence Spectroscopy

Laboratory 5: Properties of Enzymes

A NEW LOOK AT RISLEY PRISMS. By Craig Schwarze Senior Systems Engineer OPTRA Inc.

APPLIED. Measuring Caffeine Concentration. Application Summary. Analytes: caffeine (C 8 H 10 N 4 O 2

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Department of Aerospace Engineering Indian Institute of Science Bangalore

Flour Milling Process Analysis Get more out of your production with High Resolution in-line analysis. ProFoss. Dedicated Analytical Solutions

Appendix 5 Overview of requirements in English

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

Functional Data Analysis of MALDI TOF Protein Spectra

Shimadzu Simulated Distillation Gas Chromatograph System C184-E030

Function Guide for the Fourier Transformation Package SPIRE-UOL-DOC

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER , CAS NUMBER ] IRON ORE PELLETS

Measurement & Analytics Measurement made easy. MB3600-CH10 FT-NIR oleochemicals analyzer Pre-calibrated for Iodine Value (IV) and %Trans Fat

DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

A whitepaper on: Invisible and Fluorescing Bar Code Printing and Reading

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1

An Introduction to the MTG-IRS Mission

Advanced Solutions for Gas Monitoring

NIRCal Software data sheet

m/z

Agilent Cary 60 UV-Vis

SSO Transmission Grating Spectrograph (TGS) User s Guide

Refinery Planning & Scheduling - Plan the Act. Act the Plan.

1 st day Basic Training Course

INTRUSION PREVENTION AND EXPERT SYSTEMS

OLIVÉR BÁNHIDI 1. Introduction

Chemistry Instrumental Analysis Lecture 1. Chem 4631

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

VECTORAL IMAGING THE NEW DIRECTION IN AUTOMATED OPTICAL INSPECTION

MB3000-PH FT-IR spectrometer for the pharmaceutical industry Designed for QA/QC, formulation, research and development laboratory applications

PROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR)

Application Note Noise Frequently Asked Questions

Infrared Spectroscopy: Theory

Synthetic Sensing: Proximity / Distance Sensors

The photoionization detector (PID) utilizes ultraviolet

Supporting Information

Transcription:

Raman vs. FTIR Process Analyzers for In-Line Blending Use Ara Barsamian Refinery Automation Institute, LLC Jabarsa@refautom.com: Tel: +1-973-644-2270 This paper was provoked by recent interest in using Raman and other spectroscopic methods for certifying in-line blends, particularly after ASTM sponsored a working group in D02 [1] to work on practices (e.g. using generalized calibrations for direct match technique) that would be acceptable to FTC (for certifying octanes) and other US government agencies. Bottom line: spectroscopic methods work, but they are highly dependent on the quality of modeling and design of spectrometers 1. Background Turbulent economic times are pressing refiners to increase profits to simply stay in business. We feel acutely this pressure in gasoline and diesel blending since we make an easy target for news media and politicians readily pointing fingers at us to escape public wrath for their incompetence and bias against the oil industry. Well, how do we increase profits? By lowering production costs (something alien to US Congress). We do this magic by implementing well-established and proven techniques, such as in-line blend property control (1965) to insure we meet specs as close as possible to legal limits, and equally important, that we can certify in-line blends on the fly using on-line analyzers (1965 first in-line blend release based on knock engines!). In US, where we typically produce about 100 to 150kbpd of various grades of gasoline ( a big sea of gasoline worth a couple of billions a year to the refinery), stopping the blending process because of analyzer failure is usually not acceptable because you might miss a shipment (waiting tanker demurrage, missing your pipeline parcel injection time slot), and you never hear the end of it from planning and economics, traders, instant experts, your boss signing your paycheck, aso The first generation of on-line analyzers were electro-mechanical contraptions, like CFR knock engines (1929), Reid Vapor Pressure analyzers, D86 Distillation, etc., one for each property to certify for releasing the blend to sales. As one can imagine, the costs piled up, not just the initial acquisition cost, but also maintenance costs, spares, training and each technology used different design, parts, method of operation and maintenance, a mixture of mechanical, electrical and electronic parts enough to drive the blending people crazy. In the 1970 s refiners were looking desperately for more reliable and cheaper technology for on line analysis, and this how the spectroscopic methods were born with Ashland s pioneering effort with Infra-Tane TM octane analyzer using NIR spectroscopy, and then we had a booming field with every oil major developing its own proprietary method.

2. Spectroscopic Analysis Methods Today we use three different spectroscopic techniques for on-line gasoline analysis Near Infra Red (NIR), Mid-IR Raman Nuclear Magnetic Resonance (NMR); also known as Magnetic Resonance Analyzer (MRA) Of these three methods, we will look at the first two because they turned out to be easier to implement and use. Nevertheless, all are INFERENTIAL type of measurements relying on models to predict properties based on spectral analysis. The proliferation of optical spectroscopic methods for in-line blend property measurements, e.g. IR, NIR, Mid-IR, and Raman type of spectroscopic techniques produced various confusing claims that method A is better than method B, thus it would be useful to briefly review their basic functionality and usefulness for blending. All these methods are available in direct spectra (so-called dispersive) or Fourier Transformed (FT) spectra measurement technique, IR or FTIR. Issues to consider are reliability and robustness of property predictions, ease of developing and implementing property prediction models, and field equipment reliability and cost. 2.1 Basics of Raman and FT-IR Spectroscopy There are two classes of spectroscopic measurement: DISPERSIVE (Fig. 1) and Fourier- Transformed (FT) which apply to both IR/NIR and Raman (Fig 2). A dispersive method simple involves passing a light (reflected or through a sample) through a spectrograph and then a multi-channel detector to determine the spectra. In the FT method, we combine the direct light and the reflected light through a Michelson interferometer; the resulting interferogram is FT-transformed to get the spectra. Fig. 1 Dispersive Raman Spectrometer [2] Fig. 2 FT-type Spectrometer [2]

2.1.1 Raman Basics Raman method illuminates a sample with a monochromatic light from a laser; the illuminated sample material gives off light of a different color (Raman shifted light, which means a different wavelength or color, see Fig. 3). The Raman effect light has specific spectral peaks which indicate specific composition, and the peak intensity indicates the concentration. It is a linear process, which means that models to predict properties of interest are simpler to build and more robust than IR methods (MIR, NIR, FT-IR). On the other hand, the sample also fluoresces (like a chest X-ray phosphor screen) and also the primary illuminating beam punches thru the sample. This requires careful filtering out of these spurious lights. The actual Raman shift light is very weak, which means that special measures have to be taken to intensify it so it can have enough strength (S/N ratio) to be measured. The Raman spectrum is illustrated in Fig. 6. Fig. 3 FT Raman Spectrometer [3] Fig. 4 FTIR Spectrometer [3] 2.1.2 FT-IR Basics The FTIR method illuminates a sample with a multispectral light and measures the absorbance attenuation of the beam of light at various wavelengths as it passes through the sample (Fig. 4). The challenge here is that some of the compositional spectral lines overlap, so they are partially SUPERIMPOSED (and look smeared ), thus one cannot accurately separate individual component spectral lines and peak intensities, and this makes the prediction model and the type of chemometric tool used much more complicated.

2.1.3 Property Prediction Modeling Basics and Practices All spectroscopic methods require a mathematical model that ties the spectra to a property, e.g. octane. The mathematical tools to build these property prediction models use Chemometrics which combines math with statistics, e.g. PLS (partial least squares), PCA (principal component analysis), etc. In theory, the Raman spectral lines being linear with composition concentration would require fewer samples and make it easier, faster, and cheaper to develop, but that is not always the case. The samples needed to build a robust and reasonably accurate model are specified in the ASTM E1655 practice Modeling Bible [5], and there are no short-cuts. Fig. 5 FTIR Spectrum [4] Fig. 6 Raman Spectrum [4] Established industrial practice is to use a Lab analyzer to build property prediction models which are then transferred to on-line analyzer(s) on the in-line blender header; this process is referred to as Calibration Transfer. For example, a successful calibration transfer produces model prediction differences between a Lab analyzer and a field on-line analyzer much closer than ASTM repeatability of the test method, e.g. octanes should agree within 1 standard deviation, or about 0.1 ON. To be able to do this is very demanding of the precision of optical system design, manufacturing tolerances, and thermal stability of the spectrometers, which is not always the case.

Mechanically, a model is built using fuel sample spectra measured by the Lab analyzer which is then correlated to actual properties measured using accepted conventional analysis methods, e.g. ASTM D2885 for octanes. 2.1.4. Enhanced Capability Through Fourier Transform (FT) Techniques It is important to understand that Fourier Transform (FT) techniques apply to both IR, NIR, Mid- IR, and Raman type of spectroscopic instruments. Applying FT techniques improves the sharpness of the spectra and increases the signal to noise (S/N) ratio allowing to see very weak spectral lines (thus lower concentrations). The FT technique is a bit more complicated, the spectrometer and/or interferometer are a bit more complex than a simple dispersive grating and thus potentially more expensive. What is Fourier Transform (FT)? A FT-type instrument works in two steps: 1. first collects an Interferogram (combination of direct polychromatic reference light beam with the transmitted light beam spectrum through the sample using a Michelson interferometer (like in High School elementary physics class) 2. second, it uses Fast Fourier Transform Algorithm mathematics to convert an interferogram (which looks like a shapeless blob) into a frequency-domain spectra identifying hydrocarbon components (like a hologram). The resulting spectrum has a much higher resolution (better discerning closely spaced spectral lines, and thus composition) and a higher signal to noise ratio (higher sensitivity detecting low concentrations). The benefits of using FT type instruments justify the more parts, and higher cost when compared with the ordinary dispersive type instruments like straight NIR or Raman. 3. Raman Vs. FT(IR, mid-ir, NIR) Before briefly comparing Raman with IR/FTIR technology, one needs to put things in perspective: there are over 20,000 NIR/FTIR instruments in the field vs. less than 2000 Raman instruments. Most of the instruments are used in pharmaceutical and foodstuff industry, not refining. Theoretical advantages do not always translate into actual experience; in refining the experience was mixed, so it is good practice to thoroughly investigate technologies before using them! Raman Theoretical Advantages Property prediction model is easier to make, and more robust than NIR

o Chemical and molecular information is rich with high specificity (well defined spectral lines) compared with NIR o Raman spectral bands are usually sharper and interference weaker than NIR, making data analysis easier No interference from water No sample conditioning required Probe is Temperature insensitive if swings are less than 10 deg C/hr Disadvantages Raman spectrograph and Raman laser more complicated than FTIR Spectrometer optical path requires tight temperature control (a must) Calibration transfer between instruments is very sensitive to spectrometer temperature control Laser aging impacts accuracy Smaller number of Raman instruments in the field than FTIR Small number of companies selling Raman for blending applications and support Life-Cycle-Cost and initial price of equipment could be higher, although the cost of modeling could theoretically be lower 4. Conclusion What is the best spectroscopic technique for in-line blend property measurement? Remember to put things in perspective: That we have to deal with ASTM precision of measurement, don t expect zero errors, but use the ASTM reproducibility (or repeatability for in-line measurements) as a guideline Property prediction model development takes know-how (like ASTM E1655) [5], not brute-force feeding of each blend result in the prediction model The goodness of a spectroscopic method is defined against ASTM precision numbers, and ASTM performance measurement practices (like ASTM D6122) [6], not claims in a magazine article If you use wrong methodology for building the model, it will never work right Modeling requires interaction between planning, Lab, blending, and vendor chemometrician Quality of spectroscope design is crucial for accuracy and usable calibration transfer Expensive does not mean guaranteed success Vendor proven experience in fuel blending application and high quality equipment and design is key to success Always buy with process performance guarantees demonstrated over a minimum of 6 months (no fiddling with the model); the extra x% you pay will repay in a matter of weeks and happy end users

There are no short cuts or magic. The best way to decide what is best for a specific application is to rent one for about six to nine months to gain experience by using it preferably, in parallel with conventional in-line blender measurements. During the site testing, spend 2 to 3 months to build and check models; the next 6 months is to check robustness of models (a model should be insensitive to crude changes, to seasonal grade variation, and minor process upsets, e.g. reformer octane). If the finished model requires any adjustments in 6 months (assuming no major process changes or significant new components), you have a poor model most likely because of not following the steps in the ASTM E1655 practice If the results during this post-finished model site evaluation are positive, buy it. 5. Acknowledgement I would like to thank my colleagues Norm Weeks, David Seiver and Daniel C. Mertens who kindly reviewed the paper and provided valuable feedback. References: [1] Mertens, D., ASTM D02-WK38018 Draft: Standard Test Method for Spectroscopic Determination of Properties of Spark Ignition Fuels Using Direct Match Comparison Technique, (June 2012) [2] McCreery, R.L., Raman Spectroscopy for Chemical Analysis, 3 rd ed., Wiley, NY(2000) [3] Baeten, V., Dardenne, P., Spectroscopy: Developments in Instrumentation and Analysis, Grasas y Aceites, p.55, V53, 1, 2002 [4] Smith, L., Process Instruments Inc. Application Brochure, p.4 (2013) [5] ASTM E-1655 Standard Practices for Infrared Multivariate Quantitative Analysis, (2005) [6] ASTM D-6122 Practice for Validation of Performance of Multivariate Online, and Laboratory Infrared Spectrophotometer Based Analyzer Systems Analysis, (2013)