Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis



Similar documents
Digital Breast Tomosynthesis QC Requirements

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Quality Control of Full Field Digital Mammography Units

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Update on ACR Digital Mammography QC Manual

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Annual MQSA Inspection Questions NMQAAC April 19, 2004

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

Digital Mammography Update: Design and Characteristics of Current Systems

American College of Radiology Mammography Accreditation Program

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

MQSA and ACR Digital Breast Tomosynthesis Mammography Accreditation

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Performance testing for Precision 500D Classical R/F System

How To Compare Mammography To A 3D System

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Hologic Selenia Dimensions C-View Software Module. October 24, 2012

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

2014 Meeting of Mid-Atlantic States Radiation Control Programs March 25, 2014

Aspire FCRm/CRm Fujifilm CR for Mammography

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Update on MQSA and ACR Mammography Accreditation

Quality Control and Maintenance Programs

MQSA Quality Control Manual for Monochrome Displays for Mammography

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

INTRODUCTION. A. Purpose

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Physics testing of image detectors

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Digital Mammography Quality Control

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

Breast MRI Quality Control

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

Prepublication Requirements

Measuring Size and Distance in Digital Mammography

Chapter 15. Creating the Digital Image. Objectives. Key Terms

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Supports screening, diagnostic, and multimodality workflows

THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS

The Exposure Index and Its Standardization

Concepts for High-Resolution Low-Dose CT of the Breast

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Mammography. What is Mammography?

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Recommendations for a technical quality control program for diagnostic X-ray equipment

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

The Mammography Quality Standards Act Final Regulations Document #4; Final Guidance for Industry and FDA

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

Fundamentals of Cone-Beam CT Imaging

The Possibilities Tomosynthesis Brings to Lung Cancer Screening

MQSA, FDA Regulations, and Inspections

Page: 1 of 6 Page: 1 of 6

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Elements of required physical infrastructures: space, schielding, and patient flow..

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Needle crystal detector technology in mammography further dose reduction and clinical image quality with

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

PLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT TECHNICAL MANUAL /11 LAT LAT SIN DEX. kv kv. mas. mas CTL

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

MIN-R 2000 Plus Film

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit

American College of Radiology CT Accreditation Program. Testing Instructions

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Purchasing a cardiac CT scanner: What the radiologist needs to know

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Assessing Radiation Dose: How to Do It Right

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

Mammomat Inspiration Mammomat Inspiration Prime

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

Transcription:

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Digital Breast Tomosynthesis was approved by FDA, finally!!! On February 11, 2011, FDA approved Hologic to market Selenia Dimensions Digital Breast Tomosynthesis (DBT) system.

Radiation Tasks for Risks DBT Associated Implementation with CT Task Train staff for DBT Evaluate shielding requirement Install the DBT system Perform Mammography Equipment Evaluation (MEE) Obtain the certificate to operate Performer Hologic or 3 rd party MP Hologic MP Clinical Manager

Staff Qualification for DBT FDA considers the DBT as a new modality Each MP, tech, or radiologist must be qualified for FFDM Each MP, tech, or radiologist must have additional 8 hours training in DBT Hologic installation engineer can provide 5 hours hands-on training to MP. Additional 3 hours training can be acquired online from Hologic.

Facility Certification for DBT 2D portion of the system must be accredited by an approved accreditation body There is no approved accreditation body for DBT Apply to FDA for DBT certification extension

Shielding Requirement for DBT Factors to consider: Selenia Dimensions operates up to 49 kvp W/Rh W/Ag W/Al Mean kvp (MGH) 29.7 30.9 32.1 Workload may be different NCRP 147 workload, scatter data and α, β, γ may not be applicable

Get Ready for MEE Before starting MEE Know tests to be performed in 2D or 3D mode Know performance criteria for each test Collect required testing tools Get familiar with system and user interface

Selenia Dimensions MEE Tests

Selenia Dimensions QC Tests There is no QC standard for all DBT systems. Follow Equipment Vendor s QC manual The Hologic QC manual MAN-01965 covers: -Selenia Dimensions 2D FFDM system -Selenia Dimensions DBT system

Selenia Dimensions QC Tests (MP) Quality Control Test Mammographic Unit Assembly Evaluation Collimation Assessment Artifact Evaluation kvp Accuracy and Reproducibility Beam Quality Assessment HVL Evaluation of System Resolution Automatic Exposure Control (AEC) Function Performance Entrance Exposure, AEC Reproducibility, and Dose Radiation Output Rate Phantom Image Quality Evaluation Signal-To-Noise and Contrast-To-Noise Measurements Diagnostic Review Workstation Quality Control Detector Ghosting (Troubleshooting Use Only) 2D Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes 3D Yes Yes Yes Yes Yes Yes Yes

Selenia Dimensions QC Tests (MP) Quality Control Test Mammographic Unit Assembly Evaluation Collimation Assessment Artifact Evaluation kvp Accuracy and Reproducibility Beam Quality Assessment HVL Evaluation of System Resolution Automatic Exposure Control (AEC) Function Performance Entrance Exposure, AEC Reproducibility, and Dose Radiation Output Rate Phantom Image Quality Evaluation Signal-To-Noise and Contrast-To-Noise Measurements Diagnostic Review Workstation Quality Control Detector Ghosting (Troubleshooting Use Only) Special Tools? Special film Special sensor More Al sheets 2-15 lp/mm

Selenia Dimensions: Image Acquisition Modes 3 Combo: Tomo + Conv under the same 1 compression 2 Conventional Only Tomosynthesis Only

Selenia Dimensions: Specifications Conventional 2D Imaging Tomosynthesis 3D Imaging a-se detector, 24 29 cm area 70 μm pixel size Rh and Ag filters HTC grid in contact mode; No grid in magnification mode a-se detector, 24 29 cm area 140 μm pixel size Al filter No anti-scatter grid Moving tube, 15 sweep Moving detector 15 projections 3-4 seconds acquisition Reconstruction - ~100 μm pixel size - 1 mm slice spacing

You are qualified for screen/film mammography, but not for FFDM. How many hours training do you need before you can perform acceptance test for DBT? 1. 16 hours for FFDM 2. 8 hours for FFDM 3. 8 hours for DBT 4. 8 hours for FFDM + 8 hours for DBT 5. 16 hours for DBT 10

Answer: 4. 8 hours for FFDM + 8 hours for DBT Explanation: 8 hours training is needed for each new modality. Hologic Selenia Dimensions with DBT option is a 2-modality system (FFDM+DBT) Reference: www.fda.gov/radiation-emittingproducts/mammographyqualitystandardsactandprogram /FacilityCertificationandInspection/ucm243765.htm

Your hospital just bought a Hologic Dimensions with DBT option. How do you get MQSA facility certificate for DBT? 1. Apply accreditation for both 2D and DBT from ACR, then apply to FDA for the certificate 2. Apply to FDA for the certificate for both 2D and DBT directly, bypass accreditation body 3. Apply for accreditation for 2D from ACR or other accreditation body, then apply to FDA for certification extension for DBT 4. Apply accreditation for DBT from State of Texas, then apply for the certificate from FDA 5. Apply accreditation for DBT from State of Iowa, then apply for the certificate from FDA 10

Answer: 3 Apply for accreditation for 2D from ACR or other accreditation body, then apply for certification extension for DBT from FDA Explanation: Normal FFDM accreditation process should be followed for 2D portion. No accreditation body for DBT at the moment. Apply for certification extension for DBT directly from FDA Reference: www.fda.gov/radiation-emittingproducts/mammographyqualitystandardsactandprogram /FacilityCertificationandInspection/ucm243765.htm

Available target/filter combinations and the maximum kvp for Hologic Dimensions are: 1. Rh/Rh, W/Ag, W/Al, and 49 kvp 2. Mo/Mo, Mo/Rh, Rh/Rh, and 35 kvp 3. Rh/Rh, Rh/Ag, Rh/Al, and 39 kvp 4. W/Rh, W/Ag, W/Al, and 39 kvp 5. W/Rh, W/Ag, W/Al, and 49 kvp 10

Answer: 5. W/Rh, W/Ag, W/Al, and 49 kvp Explanation: Available target/filter combo in 2D mode are W/Rh and W/Ag Available target/filter combo in DBT mode is W/Al kvp range is between 20 and 49 Reference: Selenia Dimensions User Manual MAN-01964, page 81

Selenia Dimensions: User Interface

Selenia Dimensions: User Interface

Selenia Dimensions: User Interface

QC Test: Artifact Evaluation Flat field images to be acquired with 4 cm acrylic 2D contact mode (LFS) with Rh and Ag filters 2D mag mode (SFS) with Rh and Ag filters 3D DBT (LFS) with Al filter. In 3D, central projection image is evaluated. Artifacts due to reconstruction are not evaluated. Image Review Condition: WL = EI ; WW= 500 Full resolution Actual Pixel button

QC Test: kvp Accuracy Target is W and kvp range is 20-49 May need new sensor or kvp meter Use copper plate or lead sheet to protect detector

QC Test: HVL Use copper plate or lead sheet to protect the detector Use Zero-Degree Tomo in 3D mode Target and filters are unique and beam is harder More al sheets may be needed. Performance Criteria: HVL > kvp/100 + 0.03 (mm Al)

QC Test: System Resolution Image with line pair pattern One 2D acquisition One 3D acquisition Performance Criteria: 2D: > 7 lp/mm at 45 3D: > 3 lp/mm at 45 You may need a new line pair phantom (2-15lp/mm) Note: Z-resolution is not evaluated.

QC Test: AEC AEC modes Auto-Filter - filter, kvp, mas automatically determined Auto-kV - filter manually selected - kvp and mas automatically determined Auto-Time - filter and kvp manual selected - mas automatically determined

QC Test: AEC Pay attention to: AEC sensor position (must be at 2) Compression thickness (must be precise) Dose table or CNR correction table used

QC Test: AEC Performance 2D

QC Test: AEC Performance 3D

QC Test: Dose Performance criteria: - AGD < 3.0 mgy (2D) - AGD < 3.0 mgy (2D+ 3D)

QC Test: Radiation Output Use copper or lead sheet to protect the detector Make exposure at 28 kvp with Rh filter, Max mas ma @ 28 kvp is 160 Performance criteria: Tube output rate >= 230 mr/s

QC Test: Phantom Image Quality Use ACR Phantom View provided by the system Acquire 2D and 3D phantom images using AEC

Phantom Image Quality Evaluation 2D Actual Pixel button Click the Actual Pixel button to bring the image to full resolution Score the phantom following 1999 ACR Mammography Quality Control Manual 2D Phantom Passing Score: - 5 fibers, 4 specs, 4 masses - Due to phantom variation, a score of (4.5, 4.0, 3.5) is acceptable if SNR and high contrast resolution tests pass

Phantom Image Quality Evaluation 3D Click the Actual Pixel button to bring the image to full resolution Scroll to the reconstruction slice in which the phantom elements are in focus Score the phantom following 1999 ACR Mammography Quality Control Manual 3D Phantom Passing Score: 4 fibers, 3 specs, 3 masses Actual Pixel button

For Hologic Selenia Dimensions, which of following phantom scores is acceptable? 1. 4 fibers, 3 speck groups and 3 masses for DBT; 5 fibers, 4 speck groups and 4 masses for 2D 2. 4.5 fibers, 4 speck groups and 3.5 masses for DBT; 4.5 fibers, 4 speck groups and 3.5 masses for 2D 3. 5 fibers, 4 speck groups and 4 masses for DBT; 5 fibers, 4 speck groups and 4 masses for 2D 4. 4 fibers, 4 speck groups and 4 masses for DBT; 4 fibers, 4 speck groups and 4 masses for 2D 5. 4 fibers, 3 speck groups and 3 masses for DBT; 4 fibers, 3 speck groups and 3 masses for 2D 10

Answer: 1. 4 fibers, 3 speck groups and 3 masses for DBT 5 fibers, 4 speck groups and 4 masses for 2D Explanation: Specified in Selenia Dimensions QC manual Reference: Selenia Dimensions QC Manual MAN-01965, page 40

For Hologic Dimensions, the total mean glandular dose to ACR mammographic Accreditation phantom in the combo mode (2D+DBT) must not exceed 1. 6 mgy 2. 3 mgy 3. 4.5 mgy 4. 2.5 mgy 5. 3.5 mgy 10

Answer: 2 3 mgy Explanation: Specified in Selenia Dimensions QC manual Reference: Selenia Dimensions QC Manual MAN-01965, page 34

For Hologic Selenia Dimensions, the HVL must satisfy 1. HVL < kvp/100 + 0.30 mm Al 2. HVL > kvp/100 + 0.03 mm Al 3. kvp/100 + 0.03 mm Al < HVL < kvp/100 + 0.30 mm Al 4. kvp/100 + 0.03 mm Al < HVL < kvp/100 + 0.22 mm Al 5. kvp/100 + 0.03 mm Al < HVL < kvp/100 + 0.19 mm Al 10

Answer: 2. HVL > kvp/100 + 0.03 mm Al Explanation: Specified in Selenia Dimensions QC manual Reference: Selenia Dimensions QC Manual MAN-01965, page 23

QC Test: Collimation Assessment 1. Deviation between X-ray field and light field Measured on the breast holder in 2D Only evaluated for 24 cm x 29 cm collimation

QC Test: Collimation Assessment 2. Deviations between X-ray field and detector 2D: 24 x 29 Center, 18x24 Left, 18x24 Center, 18 x 24 Right 3D: 18 x 24 Center, Zero-Degree Tomo view 18x24 Center 18x24 Left 18x24 Right

QC Test: Collimation Assessment 3. Alignment of paddle and the detector 2D: 18x24 Center and 24x29 Center

QC Test: Collimation Assessment Comments The size displayed = the size on the image plane / ERMF Default ERMF = 1.073

Other Optional Tests Ghosting Artifacts for reconstructed images Z-resolution

Thank You!