Measurements in 75 Ω Coaxial Transmission Lines



Similar documents
Validation of On-Wafer Vector Network Analyzer Systems

Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows

Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements

S-PARAMETER MEASUREMENTS OF MEMS SWITCHES

2. The Vector Network Analyzer

Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters

February 2010 Number Vector Network Analyzer upgrade to SureCAL Power Sensor and RF Components Packages

Category 8 Cable Transmission Measurements Comparative Study between 4-port single wire measurements and 2-port balun measurements

Broadband Push-Pull Power Amplifier Design at Microwave Frequencies

Shielding Effectiveness Test Method. Harbour s LL, SB, and SS Coaxial Cables. Designs for Improved Shielding Effectiveness

Impedance 50 (75 connectors via adapters)

Agilent PN In-fixture Microstrip Device Measurements Using TRL* Calibration. Product Note

Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer. Application Note

spinner Measurement & Calibration equipment for network analyzers

Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers

Copyright 1996 IEEE. Reprinted from IEEE MTT-S International Microwave Symposium 1996

Agilent Measuring Noninsertable Devices

Applying Error Correction to Network Analyzer Measurements. Application Note Table of Contents. Page

ELECTRICAL CHARACTERISATION OF SEMI-RIGID COAXIAL CABLES WITH SMA AND K CONNECTORS. Carmen Diez Rafael García Juan Daniel Gallego.

MEASUREMENT OF CRYOGENIC PERFORMANCE OF 4-8 GHz PAMTECH ISOLATORS S/N Juan Daniel Gallego Isaac López Fernández Carmen Diez González

Agilent Advanced impedance measurement capability of the RF I-V method compared to the network analysis method. Application Note

Utilizing Time Domain (TDR) Test Methods For Maximizing Microwave Board Performance

Experiment 7: Familiarization with the Network Analyzer

ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER

A Study of RF Absorber for Anechoic Chambers Used in the Frequency Range for Power Line Communication System

ACHIEVING accurate broadband on-wafer calibrations

Agilent 8753ET/8753ES Network Analyzers

Politecnico di Torino. Porto Institutional Repository

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

EMC Basics. Speaker : Alain Lafuente. Alain.lafuente@we-online.com

RF and Microwave Accessories. CD-ROM Catalog. Find the right component for your Rohde & Schwarz test & measurement equipment

Connectivity in a Wireless World. Cables Connectors A Special Supplement to

Techniques for precise cable and antenna measurements in the field

A New Concept of PTP Vector Network Analyzer

MEASUREMENT OF COMPLEX PERMITTIVITY OF LIQUIDS USING WAVEGUIDE TECHNIQUES

Calibration Procedure for Measuring S-Parameters in Balun Applications on 150-Ω High-Speed Cables

75 Ω Transmission System

Using S-Parameter and Load Pull Measurements to Validate Transistor Large-Signal Fundamental and Harmonic Tuning Performance

Making a Case for a New Connector

Agilent AN In-Fixture Measurements Using Vector Network Analyzers

Probes and Setup for Measuring Power-Plane Impedances with Vector Network Analyzer

UNDERSTANDING NOISE PARAMETER MEASUREMENTS (AN )

Advanced Calibration Techniques for Vector Network Analyzers

Tuning a Monopole Antenna Using a Network Analyzer

Cable Analysis and Fault Detection using the Bode 100

Measurement & Modeling Device Characterization Solutions

AGILENT E5071C ENA NETWORK ANALYZER

Agilent EEsof EDA.

1000BASE-T1: 3-Port Conversion/Balance Measurement for Cables and Components

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER

Pre-Compliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions

A Comparison of Measurement Uncertainty in Vector Network Analyzers and Time Domain Reflectometers. White Paper

Agilent Test Solutions for Multiport and Balanced Devices

Techniques for Precise Cable and Antenna Measurements in the Field

Measurement of Inductor Q with the MSA Sam Wetterlin 3/31/11. Equation 1 Determining Resonant Q from Inductor Q and Capacitor Q

MICRO-COAX Space Grade Cable Assembly Integration Manual

Introduction to Network Analyzer Measurements

Distance-To-Fault (DTF) Measurement in 75 Ohm Systems with the R&S FSH3-TV

PRINTED LOOP ANTENNA INTEGRATED INTO A COMPACT, OUTDOOR WLAN ACCESS POINT WITH DUAL-POLARIZED RADIATION

Application Guide to RF Coaxial Connectors and Cables

Vector Network Analyzer (VNA) Calibration: The Basics

Agilent E5063A ENA Series Network Analyzer. 100 khz to 4.5/ 8.5/18 GHz

Times Microwave Systems Hermetically Sealed Assemblies

A Reflection-Type Vector Modulator with Balanced Loads

Cable Impedance and Structural Return Loss Measurement Methodologies

Measuring RF Parameters of Networks Bill Leonard NØCU

Design of a Planar Omnidirectional Antenna for Wireless Applications

LOW LOSS CABLE PAG. 1

Understanding Power Splitters

Scalar Network Analysis with the HP 8590 Series Spectrum Analyzers Product Overview

db MISER ULTRA LOW LOSS CABLE ASSEMBLIES Are you a design or test engineer fighting a challenging loss budget?

Characterisation of Schottky Varactor Diodes for Pulse Compression Circuits

R3765/67 CG Network Analyzer

Capacitor Self-Resonance

HP 8970B Option 020. Service Manual Supplement

MAURY. VNA Calibration Kits, Microwave Components & Adapters. Calibrate With Confidence IN THIS CATALOG:

Characterization of Spatial Power Waveguide Amplifiers

Analysis of Quick Lock N Type Connectors

J. Zhang, J.-Z. Gu, B. Cui, andx. W. Sun Shanghai Institute of Microsystem & Information Technology CAS Shanghai , China

Design of an U-slot Folded Shorted Patch Antenna for RF Energy Harvesting

LARGE-SIGNAL NETWORK ANALYZER MEASUREMENTS AND THEIR USE IN DEVICE MODELLING

Coaxial Cable Products Guide. Connectivity for Business-Critical Continuity

VIETNAM NATIONAL UNIVERSITY HOCHIMINH CITY INTERNATIONAL UNIVERSITY SCHOOL OF ELECTRICAL ENGINEERING

Field Calibration Software

Radio Frequency Interference (RFI) Shielding Principles

APN1001: Circuit Models for Plastic Packaged Microwave Diodes

The Application of Vector Network Analyzers in Balanced Transmission Line Signal Integrity Measurements

Cancellation of Load-Regulation in Low Drop-Out Regulators

iva Cable & Antenna Analyzer

DSA800 Series Spectrum Analyzer


Digital Systems Ribbon Cables I CMPE 650. Ribbon Cables A ribbon cable is any cable having multiple conductors bound together in a flat, wide strip.

Data Receivers. Digital Data Receivers

"FP", "FR", "FQ" Series Bandpass Filters

Transcription:

Measurements in 75 Ω Coaxial Transmission Lines Using CMT PLANAR Vector Network Analyzers (VNAs) with 50 Ω Test Ports One application of a vector network analyzer (VNA) is the measurement of 75 Ω coaxial transmission lines. In this article, we will discuss making these measurements using the most common type of VNAs, which have 50 Ω test ports, in conjunction with 50Ω to 75Ω Minimum Loss Pads (MLP), i.e. impedance matching attenuators with insertion loss of 5.7 db. As 50Ω VNAs we suggest using the VNAs manufactured by Copper Mountain Technologies: PLANAR TR1300/1, PLANAR 304/1, and PLANAR 804/1 (as of 2012). The use of an MLP affects accuracy of measurements by changing the calibration error and, depending on the location of the attenuator in the measurement circuit, impacts stability of measurements related to test cable bending. To assess the impact of MLP on the accuracy of the measurements, we will refer to the calibration comparison method [1,2]. We will also compare the additional errors with the typical errors of the VNAs with 75Ω coaxial lines after the most widely used and affordable SOLT calibration (using SHORT, OPEN, LOAD, and THRU standards). To assess the errors dependent on the location of the attenuator in the measurement circuit, it is possible to use the results of practical measurements. Assessing the additional calibration error caused by the MLP, we will use a number of calibration schemes for a measurement system based on PLANAR VNAs and will apply the calibration comparison method. Let us perform calibrations of the measurement system with the MLPs as shown in Figure 1. TRLM calibration is one of the most accurate calibrations in the industry today. It is performed using the following standards: THRU, REFLECT, LINE, and MATCH. This calibration technique is recognized as the benchmark calibration all over the world, and a TRLM calibrated measurement system achieves the effective parameters of directivity much better than 50 db and source match much better than 50 db [3]. 1 2012 Copper Mountain Technologies

Figure 1 SOLT calibration is the most widely used and affordable calibration technique. In our example we use a high quality Agilent 85036B calibration kit, which provides the calibrated measurement system with the following effective parameters: directivity no less than 46 db at frequencies up to 2 GHz, directivity no less than 40 db at frequencies up to 3 GHz, source match no more than 36 db, and reflection tracking no more than 0.03 db. To assess the measurement system from Figure 1 we will compare two successive TRLM calibrations. Results from Figure 2 confirm that the proposed measurement system allows comparing two calibrations with the following effective parameters: directivity at least 73 db, source matches at least 65 db, and reflection tracking being less than 0.003 db. This measurement proves that the quality of the TRLM calibration performed on PLANAR VNAs practically depends on quality of the LINE and SHORT calibration standards.. Comparing TRLM and SOLT calibrations, we can see that the use of an MLP insignificantly impaired the SOLT calibration parameters compared to the parameters claimed by the calibration kit manufacturer. For example, the measured effective directivity of the given measurement system is better than 46 db at the frequency of 2.8 GHz, and thus complies with the specifications of the calibration kit manufacturer, i.e. better than 41.2 db. Therefore, the additional errors introduced into the measurement system due to the use of MLPs are comparable to the range of calibration error when using a commercially available SOLT calibration kit. 2 2012 Copper Mountain Technologies

Figure 2 Correspondingly, using MLP in a measurement system has a low impact on the measurement errors. That allows recommending the use of PLANAR VNAs (PLANAR TR1300/1, PLANAR 304/1, PLANAR 804/1) or other 50Ω VNAs of sufficient dynamic range (over 120 db) and measurement stability with a high quality MLP for performing most of VNA RF measurements in 75Ω transmission lines. It is important to analyze two possible ways of connecting MLP into the measurement circuit (Figure 3). 3 2012 Copper Mountain Technologies

Figure 3 The MLPs can be either connected to the VNA ports, or to the end of test cables. The second variant allows using widely available 50Ω test cables that surpasses 75Ω test cables in all the parameters. Even though it looks like the measuring circuit with a high quality 50Ω test cable has obvious advantages, such circuit is at a strong disadvantage due to the high sensitivity to test cable bending. Figure 4 shows S 11 parameter of the LOAD standard being measured after calibration and the subsequent test cable bending using the cheapest and simplest 75Ω cable and a several times more expensive and stable 50Ω cable. The bending of the 75Ω cable keeps the S 11 value of the LOAD at about 50 db, while the bending of a much higher quality 50Ω cable results in degradation of the S 11 to 30 db. Such S11 change actually results in 4 2012 Copper Mountain Technologies

deterioration of the effective directivity of the measurement system to 30 db, and thus increases the measurement error to an unacceptable value. For example, the measurement of S 11 equal to 16 db will have an error of 2.1 db compared to the possible error of 0.35 db; and the measurement of S 21 equal to 3 db will have an error of 0.16 db compared to the possible error of 0.08 db. Connection of MLP to the VNA test ports and use of the 75Ω test cables allows for minimizing the impact of the measuring circuit on the measurement accuracy. Figure 4 Using CMT PLANAR vector network analyzers with 50Ω test ports, 5.7 db MLP connected to the VNA test ports, and performing SOLT calibration with a high quality calibration kit, we can expect the measurement error to be dependent only on the quality of the calibration kit, keeping impact of the 50Ω to 75Ω adapter to a minimum. References: [1] Dylan Williams, Arkadiusz Lewandowski, Denis LeGolvan, and Ron Ginley Electronic Vector Network Analyzer Verification, IEEE Microwave Magazine, October 2009. [2] D. F. Williams, R. B. Marks, and A. Davidson, Comparison of onwafercalibrations, Automat. RF Tech. Group Conf. Dig., vol. 38, pp.68 81, Dec. 1991. [3]Doug Rytting, Advances in Microwave Error Correction Techniques, Hewlett Packard, RF & Microwave Measurement Symposium and exhibition, 1987. 5 2012 Copper Mountain Technologies