UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research



Similar documents
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System

Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

ApplicationNOTE Introduction

Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

How To Test For Contamination In Large Volume Water

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Introduction. Methods. Sample Processing

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

WATERS QUANTITATIVE ANALYSIS solutions

Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Salman Azimi, 1 Nayan S. Mistry, 2 and Michelle Wood 2. Drug Quality Control Laboratory, Supreme Council of Health, Doha, Qatar 2

Oasis HLB Cartridges and 96-Well Plates

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

A Novel Approach to Low Volume Sample Preparation

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

AMD Analysis & Technology AG

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH

Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )

DEVELOPMENT OF PFCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit

CORESTA RECOMMENDED METHOD N 72

Patients with COPD were recruited according to the inclusion criteria; aged 40-75, current or

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Overview. Purpose. Methods. Results

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

UPLC/MSMS in the analysis of physiological steroids. May Anders Feldthus Waters Corporation 1

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Appendix 5 Overview of requirements in English

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

LC-MS/MS for Chromatographers

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine

High-Throughput 3-D Chromatography Through Ion Exchange SPE

United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

ProteinPilot Report for ProteinPilot Software

New drugs of abuse? A case study on Phenazepam & Methoxetamine

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

Transcription:

UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal of interferences Analytical sensitivity enables the quantification of low physiological levels of aldosterone Facilitates high sample throughput utilizing automation of the multiwell plate format INTRODUCTION is a mineralocorticoid steroid hormone that plays a central role in the regulation of blood pressure. Traditionally, aldosterone has been analysed by radioimmunoassays. However, these methods involve the use of hazardous radioisotopes. In addition, they can suffer from a lack of specificity due to the cross reactivity of structurally similar steroid hormones and metabolites, which may result in greater imprecision and inaccuracy. To minimize specificity issues, radioimmunoassay methods employ time-consuming manual extraction protocols. UPLC-MS/MS combined with automation of the sample preparation with sample tracking capabilities provides an alternative means of aldosterone analysis for clinical research. An integrated workflow solution enables selective and analytically sensitive characterisation of aldosterone with a reduction in sample handling time. In this application note we describe a clinical research method utilizing an Oasis MAX µelution Plate for the extraction of aldosterone from plasma, which has been automated on the Tecan Freedom Evo 1/4 Liquid Handler. Chromatographic separation was performed on an ACQUITY UPLC I-Class System using a CORTECS UPLC C 18 Column followed by detection on a Xevo TQ-S Tandem Quadrupole Mass Spectrometer (Figure 1). WATERS SOLUTIONS ACQUITY UPLC I-Class System Oasis MAX µelution Plate CORTECS UPLC C 18 Column Xevo TQ-S Mass Spectrometer MassLynx Software KEY WORDS, SPE, UPLC-MS/MS Figure 1. The Waters ACQUITY UPLC I-Class System and Xevo TQ-S Mass Spectrometer. 1

EXPERIMENTAL UPLC conditions System: Needle: 3 µl ACQUITY UPLC I-Class Column: CORTECS UPLC C 18, 2.1 x 1 mm, 1.6 µm (Waters p/n 186795) Mobile phase A: Mobile phase B: Needle wash solvent: Purge solvent: Water Methanol Column temp.: 45 C Injection volume: 25 µl Flow rate: 8 methanol (aq) 4 methanol (aq).4 ml/min Gradient: See Table 1 Sample preparation certified reference solution and its stable labeled internal standard ( 4 H 2 ) were purchased from Cerilliant (Round rock, TX) and IsoSciences (King of Prussia, PA), respectively. Calibrators were prepared in a surrogate matrix of MSG4 stripped human serum purchased from Golden West Biologicals (Temecula, CA). The calibration range for aldosterone was 42-4161 pmol/l (15-15 pg/ml). QC materials were prepared using pooled plasma purchased from SeraLab (Haywards Heath, UK) at 99, 5 and 2 pmol/l (36, 18 and 72 pg/ml). To convert SI units (pmol/l) to conventional mass units (pg/ml) divide by 2.774. Sample extraction Extraction was performed using a Tecan Freedom Evo 1/4 Liquid Handler. To 2 µl of sample; 25 µl of 4 pmol/l internal standard, 2 µl.1 M zinc sulphate in 5 methanol and, 45 µl of.5 (v/v) phosphoric acid were added. The samples were mixed between each reagent addition. The samples were centrifuged for 5 minutes at 1g. Run time: MS conditions System: Resolution: Acquistion mode: Polarity: Capillary: 4. minutes Xevo TQ-S MS1 (.7 FWHM) MS2 (.5 FWHM) Multiple reaction monitoring (MRM) (see Table 2 for details) ESI negative 2.4 kv The Oasis MAX µelution Plate (Waters p/n 1861829) was conditioned and equilibrated with 2 µl methanol and water, respectively. An aliquot of each of the pretreated samples (625 µl) was loaded into individual wells of the plate and slowly pulled through at low vacuum. Consecutive washes with 2 µl of.5 (v/v) phosphoric acid, 2 µl.1 ammonia in 1 methanol, and 2 µl water were performed. was eluted using 5 µl of 7 methanol (aq), followed by 4 µl water. Method conditions On instrument setup ensure the mobile phase lines are purged for a minimum of 1 minutes to remove additives, which could negatively impact sensitivity for aldosterone. Source temp.: 15 C Desolvation temp.: 6 C Dwell time:.1 seconds Inter-scan delay:.1 seconds Inter-channel delay:.2 seconds Data management MassLynx v4.1 with TargetLynx Application Manager (min) Flow Rate (ml/min) A B Curve Initial.4 6 4 Initial 1.4 6 4 6 2.4 4 6 6 2.3.4 5 95 11 2.8.4 6 4 11 Table 1. Gradient table for the separation of aldosterone. Operating backpressure at the initial conditions is 12 psi. UPLC-MS/MS Analysis of in Plasma for Clinical Research 2

Analyte Precursor ion (m/z) Product ion (m/z) Cone voltage Collision energy (Quan) 359.2 189.2 55 18 (Qual) 359.2 297.2 55 16-4 H 2 363.2 19.2 55 18 Table 2. MRM parameters for both aldosterone quantifier and qualifier and its internal standard, aldosterone- 4 H 2. RESULTS No interferences were observed at the retention time of aldosterone when nine structurally related compounds were examined (18-hydroxycorticosterone, cortisol, cortisone, 11-deoxycorticosterone, corticosterone, 11-deoxycortisol, 21-deoxycortisol, prednisolone, and prednisone). In Figure 2, the chromatographic selectivity of the column is demonstrated through the baseline resolution of aldosterone and its metabolic precursor, 18-hydroxycorticosterone. Separation of 18-hydroxycorticosterone is necessary due to the detection of its isotopic transition in the aldosterone- 4 H 2 internal standard MRM trace, when present in the sample at 6 nmol/l. In addition, separation of prednisone was required due to ion suppression of aldosterone at prednisone concentrations of 1 µmol/l. This baseline resolution allows for the independent quantification of aldosterone. Analytical sensitivity investigations reveal that the analytical sensitivity of this method would allow precise quantification (<2 RSD) at 42 pmol/l for aldosterone. Signal:noise (S/N) of the 42 pmol/l calibration standard was >1:1 on ten separate occasions. Total precision was determined by extracting and quantifying three replicates of tri-level QC material on two occasions per day over five separate days (n = 3). Repeatability was assessed by analyzing three replicates at each QC level. The results of these experiments are shown in Table 3, where total precision and repeatability at the low (99 pmol/l), mid (5 pmol/l), and high (2 pmol/l) concentrations is 9.8 RSD for aldosterone. 1 1 1 1.6 1.8 2. 2.2 2.4 2.6 2.8 3. 363.2>19.2 4 H 2 1.6 1.8 2. 2.2 2.4 2.6 2.8 3. 18-Hydroxycorticosterone 361.2>189.2 1.6 1.8 2. 2.2 2.4 2.6 2.8 3. Figure 2. UPLC separation of aldosterone, aldosterone -4 H 2, and 18-hydroxycorticosterone using a CORTECS UPLC C 18 Column. Total QC precision (RSD) QC repeatability (RSD) Compound Low Mid High Low Mid High 9.8 7.1 4.8 8.2 7. 3.9 Table 3. Total precision and repeatability for the analysis of aldosterone. UPLC-MS/MS Analysis of in Plasma for Clinical Research 3

The method was shown to be linear over the range of 42 4161 pmol/l when different ratios of high and low concentration pools of aldosterone were combined and analysed. In addition, calibration lines in spiked serum were linear with coefficient of determinations (r 2 ) >.996 on 1 separate occasions. No system carryover was observed from high concentration samples at 14 pmol/l into subsequent blank injections. A 1:2 dilution was successfully performed on an over-range sample with a concentration of 766 pmol/l, providing a mean accuracy of 99 with an RSD of 4.4. Matrix effects were evaluated as the peak area of extracted post spiked aldosterone plasma samples taken as a percentage of extraction solvent samples spiked to equivalent concentrations. Mean (range) matrix effects were 1.1 (1.3 1.2) with RSD of 5.9 for aldosterone. Calculations using analyte:internal standard response ratio indicated compensation for signal enhancement by the internal standard, providing a mean (range) net matrix effect of 1. (.9 1.1) with RSD of 7.7 for aldosterone. Samples were selected (n = 59) for comparison against an independently developed LC-MS/MS method for aldosterone. A selection of samples from this comparison having quantified values <1 pmol/l are shown in Figure 3. These chromatograms demonstrate the selectivity of the extraction protocol and the sensitivity of the analytical method. 1 7 pmol/l S/N : PtP = 2 359.2 > 189.2 (Quantifier) 1.6 1.8 2. 2.2 2.4 2.6 2.8 1 95 pmol/l S/N : PtP = 96 1 1 1 1.6 1.8 2. 2.2 2.4 2.6 2.8 68 pmol/l S/N : PtP = 44 1.6 1.8 2. 2.2 2.4 2.6 2.8 72 pmol/l S/N : PtP = 27 1.6 1.8 2. 2.2 2.4 2.6 2.8 77 pmol/l S/N : PtP = 29 1.6 1.8 2. 2.2 2.4 2.6 2.8 Figure 3. Five representative chromatograms from plasma samples having quantified aldosterone values <1 pmol/l. The signal-tonoise ratio was calculated on the raw data using peak to peak at ± 2 SD. UPLC-MS/MS Analysis of in Plasma for Clinical Research 4

Comparison data were processed using Analyse-it Software v2.3. The comparison between the two independent LC-MS/MS methods yielded a Deming regression of y = 1.7x - 22.94 (Figure 4A), which showed no significant proportional or constant bias (p>.5). Altman-Bland agreement between the methods demonstrates a mean negative bias of -4.9 (95 CI -1.3 to.5) (Figure 4B) for the UPLC-MS/MS method described herein. 35 A 1 8 B Bias (-4.9) 3 6 Waters UPLC-MS/MS (pmol/l) 25 2 15 1 Difference 4 2-2 -4-6 95 CI 95 Limits of agreement (-45.8 to 36.) 5 95 CI bands Deming fit (-22.94 + 1.7x) 5 1 15 2 25 3 35-8 -1 1 2 3 95 CI LC/MS/MS (pmol/l) Mean of (pmol/l) Figure 4. A) Deming regression comparing the Waters UPLC-MS/MS method to another LC-MS/MS method for aldosterone analysis, B) Altman-Bland agreement showing the difference between the Waters UPLC-MS/MS method and another LC-MS/MS method for aldosterone analysis. CONCLUSIONS An analytically sensitive and selective clinical research method has been developed for the analysis of plasma aldosterone. Using only 2 µl sample volume, this analytical method provides sufficient analytical sensitivity to analyze low physiological levels of aldosterone (42 pmol/l). Selective chromatography and SPE using anion exchange to remove interferences provides clean chromatograms, which enables accurate and precise quantification of aldosterone across the concentration range. Automation of the analytical method in combination with sample tracking capabilities improves laboratory workflow and reduces sample handling, which alleviates the potential for operator error. Acknowledgement Professor Brian Keevil and his colleagues at the Department of Clinical Biochemistry, University Hospital of South Manchester, Wythenshawe, UK, are thanked for the provision of anonymized plasma samples for the analysis. For Research Use Only. Not for use in diagnostic procedures. Waters, The Science of What s Possible, ACQUITY UPLC, Oasis, CORTECS, UPLC, Xevo, and MassLynx are registered trademarks of Waters Corporation. TargetLynx is a trademark of Waters Corporation. All other trademarks are the property of their respective owners. 215 Waters Corporation. Produced in the U.S.A. January 215 725262EN AG-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 2 F: 1 58 872 199 www.waters.com