Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS
|
|
- Howard Garrett
- 7 years ago
- Views:
Transcription
1 Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS Solfrid Hegstad, PhD Department of Clinical Pharmacology St.Olavs Hospital
2 St.Olavs Hospital
3 Department of Clinical Pharmacology Clinical samples: Urine, serum, blood and hair 2010: Urine: (~65 %) Serum: (~30 %) Blood: 3600 (~4 %) Hair: 110 (~0.1 %) Postmortem samples: 200
4 Contents Screening of drugs in urine Ethyl glucuronide and Ethyl sulfate in urine
5 Background-Screening of drugs in urine urine samples per year Opioid-dependent patients in a medicationassisted rehabilitation program (LAR-Midt) Confirm buprenorphine or methadone treatment Control side-drug use Liquid-liquid extraction and singel LC-MS 80 % are positive, % 10 confirmed (IS14) by GC-MS
6 Aims Rapid, simple, specific and robust methods Reduce the costs per sample Dilution of sample (MSMS) Eliminate the use of toxic organic LLE solvents Small sample volume µl: Less consumptions of IS/std/enzyme Automation: Less manually workload Short runtime (UPLC): Reduced number of instruments
7 Why two separate methods? Concentration level Analytical signal (MS) Chemical properties (glucuronides)
8 Method 1
9 Method 2
10 Sample preparation Method 1 Pipetting robot Tecan Freedom EVO Separation UPLC Acquity (Waters) 50 µl urine, IS, 1: % methanol in water Analytical column BEH C18 2,1X50 Mobile phase Flow 0.1 % NH 3, 2mM NH 3 formate/ methanol 0.6 ml/min Injection volume 5 µl Column temperature 50 ºC Run time 3.0 min
11 Sample preparation Method 2 Pipetting robot Tecan Freedom EVO Separation UPLC Acquity (Waters) 100 µl urine, 1:10 methanol,buffer, Enzymatic hydrolysis Analytical column HSS T3 2,1X100 Mobile phase Flow 0.1 % HCOOH/acetonitrile 0.7 ml/min Injection volume 7.5 µl Column temperature 50 ºC Run time 2.8 min
12 Detection Metode 1 Metode 2 Xevo TQ-S ESI MRM Positive mode Positive mode Negative mode Number of transition 2 per analyte* 2 per analyte Tramadol*; One transition, a metabolite is included to confirm identity Load ahead
13 Challenges To obtain linear calibration curve; detuning, alternative transition or C13 isotope CID fragmentation of amphetamine Stability of BEH C18 column Analytical separation of hydromorphone and morphine
14 BEH C18 Morphine and hydromorphone Hydromorphone HSS T3 Morphine
15 Summary: Screening of drugs in urine Two methods Simple and automatic sample preparation Short run time Validation Implementation of method
16 Ethyl glucuronide and Ethyl sulfate in urine
17 Ethyl glucuronide (EtG) and ethyl sulfate (EtS) in urine Ethanol mainly transformed to acetic acid Formation of EtG/EtS < 0.1 % of alcohol elimination Excreted in urine Markers of ethanol intake A larger surveillance window compared to ethanol (12 h)
18 Detection in urine EtG/EtS Up to 24 h after intake of 1-2 bottles of beer. Up to 48 h after intake of 3-4 bottles of beer Up to 130 h after heavy drinking
19 Sample preparation EtG and EtS in urine Pipetting robot Tecan Freedom EVO Separation UPLC Acquity (Waters) 100 µl urine, 1:20 Water and IS Analytical column HSS T3 2,1X100 Mobile phase Flow 0.1 % HCOOH/methanol 0.6 ml/min Injection volume 5 µl Column temperature 50 ºC
20 Gradient table Time (min) 0.1 % HCOOH in Methanol (%) water (%) Initial
21 Detection Waters Xevo TQ-S Negativ mode ESI MRM Analyte Transition (m/z) Cone (V) CE (ev) Internal standard EtG EtG EtS > EtG-d > > EtS-d > EtS-d >
22
23 Table 1 Calibration range, Correlation Coefficient, LOQ, Within-assay precision, Betweenassay precision and Accuracy Analyte Calibration range, (ng/ml) Correlation Coefficient R 2 n=5 LOQ, (ng/ml) Theoretical Concentration (ng/ml) Within -assay RSD (n=10) % Between -assay RSD (n=10) % Accuracy (n=10) % EtG EtS
24 Table 2 Matrix Effect Analyte Theoretical Concentration (ng/ml) ME (%) RSD (%) ME corrected with internal standard (%) RSD (%) EtG EtS
25 Potential applications Driving license renewal Workplace testing Control of prison inmates Social and child welfare services Forensic toxicology Victims of sexual abuse
26 Victims of sexual abuse 74 samples 2010/2011 Number of samples Ethanol EtG EtS 24 Positive Positive Positive 30 Negative Positive Positive 20 Negative Negative Negative
27 Summary: EtG and EtS in urine Simple and automatic sample preparation Specific and robust Validation Implementation of method
28 Acknowledgement Technical assistance: Lisa Sylten Lisbeth Michelsen Audhild Naavik Method development: Sigurd Hermansson
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
More informationSIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS
SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS Russell Watts 1, Allan Traynor 2, Michelle Wood 1. 1 Waters Corporation, MS Technologies Centre,
More informationJennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
More informationEthyl Glucuronide. Where is the Drug? Detection Windows. Urine. Blood. Absorption and Distribution. Metabolism. Excretion
Where is the Drug? Ethyl Absorption and Distribution A Biomarker for Ethanol Use Metabolism Excretion Anthony G. Costantino, Ph.D. D-ABFT NMS Labs,Willow Grove, PA John J. Treuting, Ph.D. Treuting & Assoc.
More informationA High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
More informationSimultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
More informationA Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System
A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System Kent Johnson Fortes lab, Wilsonville Oregon List of drug of abuse candidates for
More informationSIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
More informationIntroduction. Methods. Sample Processing
PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko
More informationAnalysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
More informationAppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS
AppNote 1/2012 Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS Oscar G. Cabrices, Fred D. Foster, John R. Stuff, Edward
More informationAnalysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
More information# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
More informationComprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine
Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine Application Note Authors Peter JW Stone and Kevin McCann Agilent Technologies, Inc. Santa Clara,
More informationOverview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
More informationMultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
More informationHPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
More informationDEVELOPMENT OF PFCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS
DEVELOPMENT O PCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS Tatsuya Ezaki 1, Nobutake Sato 1, Jun Yonekubo 1, Takeshi Nakano 2 1 Nihon Waters K.K,.1-3-12 Kitashinagawa,
More informationNew drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
More informationDetermination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
More informationMaking the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity
More informationQuantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
More informationA Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples
A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples Mary Lame, Hua Yang, Sherri Naughton, and Erin Chambers Waters Corporation, Milford,
More informationSalman Azimi, 1 Nayan S. Mistry, 2 and Michelle Wood 2. Drug Quality Control Laboratory, Supreme Council of Health, Doha, Qatar 2
UPLC/MS/MS for the Screening, Confirmation, and Quantification of Drugs Illegally Added to Herbal/Dietary Supplements for the Enhancement of Male Sexual Performance Salman Azimi, 1 Nayan S. Mistry, 2 and
More informationTechnical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
More informationCliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
More informationAppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT
AppNote 1/2014 Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine Fred D. Foster, John R. Stuff, Edward A. Pfannkoch GERSTEL, Inc., 701 Digital Dr. Suite J,
More informationOpiates in Urine by SAMHSA GC/MS
application Note Gas Chromatography/ Mass Spectrometry Author Timothy D. Ruppel PerkinElmer, Inc. Shelton, CT 06484 USA Opiates in Urine by SAMHSA GC/MS Introduction The United States Department of Health
More informationAnalytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
More informationSuccessful Sample Preparation Strategies for LC/MS/MS Analysis of Drugs in Complex Biological Matrices for Forensic Toxicology Applications
Thank you for joining us! Our session will begin shortly Successful Sample Preparation Strategies for LC/MS/MS Analysis of Drugs in Complex Biological Matrices for Forensic Toxicology Applications Jonathan
More informationUHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
More informationTechnical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis
Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis 1.0 Purpose - This procedure specifies the required elements for the solid phase extraction of acidic,
More informationUsing Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
More informationOverview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationAppendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
More informationDaniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
More informationSimultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationAppNote 4/2014. Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS.
AppNote 4/2014 Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS. Oscar G. Cabrices, Fred D. Foster Gerstel, Inc., 701 Digital Dr. Suite J,
More informationAccurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
More informationHenry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS
Development of Integrated Informatics Workflows for the Automated Assessment of Comparability for Antibody Drug Conjugates (ADCs) Using LC-UV and LC-UV/MS Henry Shion, Robert Birdsall, Steve Cubbedge,
More informationHair Drug & Alcohol Testing FAQs
HAIR DRUG TESTING More than DNA Hair Drug & Alcohol Testing FAQs Need advice about testing? Contact us using our online SMS service or call Cellmark free on 0800 043 4247 2045 Introduction Cellmark is
More informationUrine Specimen Dilution: Assessment and Policy Recommendations
Urine Specimen Dilution: Assessment and Policy Recommendations Dr. Leo Kadehjian uly 21, 2003 The following comments and recommendations are for informational use only. Consultation with program counsel
More informationHydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns Jonathan P. Danaceau, Kenneth J. Fountain, and Erin E. Chambers Waters
More informationDetermination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS
Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS Thorsten Bernsmann, Peter Fürst Chemical and Veterinary Analytical Institute Münsterland-Emscher-Lippe Joseph-König-Str. 40,
More informationMaking the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
TECHNICAL NOTE Toxicology Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with
More informationHigh sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
More informationConversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna
Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna Paula Hong and Pat McConville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E
More informationTherapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS
Therapeutic Drug Monitoring of Antiretroviral Drugs with PLC-M Ursula Gutteck-Amsler, Katharina M. Rentsch Abstract Prospective and retrospective studies have provided some evidence of the clinical and
More informationCONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
More informationOverview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
More informationExtraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
More informationAnalysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
More informationThe Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
More informationConfirmation by Triple Quadrupole LC/MS/MS for HHS-compliant Workplace Urine Drug Testing
Confirmation by Triple Quadrupole LC/MS/MS for HHS-compliant Workplace Urine Drug Testing A complete applications solution for forensic toxicology laboratories John Hughes Agilent Technologies Pleasanton,
More informationOne Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas 77504
Validation of Analysis of Amphetamines, Opiates, Phencyclidine, Cocaine, and Benzoylecgonine in Oral Fluids by Liquid Chromatography Tandem Mass Spectrometry Subbarao V. Kala*, Steve E. Harris, Tom D.
More informationInternational GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories
More informationFast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.
More informationHow To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
More informationDrug Testing Reference Tables for Drug Courts
Drug Testing Reference Tables for Drug Courts July, 2009 Page 1 TABLE I. Specimen Detection Period Advantages Disadvantages URINE SWEAT (patch) ORAL FLUID (saliva) Provides a profile of both current and
More informationANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of carbohydrates in foodstuff (LC/RID method) Exercise guarantor: Assoc. Prof. Karel Cejpek, Ph.D. CONTENT Required knowledge...
More informationDiscovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
More informationCLEAN-UP PROCESS FOR MASS SPECTRAL STUDY OF AMPHETAMINES IN PUTREFIED BODY MATERIALS
CLEAN-UP PROCESS FOR MASS SPECTRAL STUDY OF AMPHETAMINES IN PUTREFIED BODY MATERIALS Kenji HARA, Seiichi KASHIMURA, Masayuki KASHIWAGI, Tomoko HAMANAKA, Mitsuyoshi KAGEURA Department of Forensic Medicine,
More informationAppropriate utilization of drug tests for pain management patients
Appropriate utilization of drug tests for pain management patients Gwen McMillin, PhD, DABCC (CC, TC) Medical Director, Toxicology, ARUP Laboratories Associate Professor (clinical), University of Utah
More informationThe Use of EtG & EtS Monitoring in Drug Court. By: Paul L. Cary Toxicology Laboratory University of Missouri
The Use of EtG & EtS Monitoring in Drug Court By: Paul L. Cary Toxicology Laboratory University of Missouri Current alcohol testing issues: legal drug - without prescription screening tests specific for
More informationInnovative vs Traditional
And in the red corner, introducing the challenger NASDAQ April 4 2011 243.47 $ 1.38 $ 0.58 % 0.07 $ 0.02 $ 39.4 % Technology DVD Digital Streamline Instant Play Internet Access Unlimited Selection No Late
More informationBruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
Bruker ToxScreener TM A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS ToxScreener - Get the Complete Picture Forensic laboratories are frequently required
More informationAn Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System
An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System Mary E. Lame, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA,
More informationPractical Aspects of Drug Testing in Human Hair
Practical Aspects of Drug Testing in Human Hair Laboratory Perspective James A. Bourland, Ph.D., DABFT FOR INTERNAL USE ONLY. NOT FOR PRINT OR DISTRIBUTION Overview of Collection Process FOR INTERNAL USE
More informationThe Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.
The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen. Mark Crawford Single Presenter Name Abstract With the recent decline in hive bee populations this year the investigation of acaricides
More informationStep-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
More informationSimultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
More informationINTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE
Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND
More informationMS/MS analysis of Polyphenols
Purdue-UAB Botanicals Center for Age-Related Disease MS/MS analysis of Polyphenols Jeevan Prasain Ph.D. Pharmacology & Toxicology UAB Polyphenols H Caffeic acid Phenolic acids and derivatives H EGC (Flavanol)
More informationGENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
More informationUltra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
More informationImproved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS
htslabs.com folks@htslabs.com 800 541.4792 760 757.8080 760 757.9367 Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS Nadine Koenig 2,
More informationFast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
More informationRapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Derek Hillbeck, Thermo Fisher Scientific, Runcorn, UK Application Note 2999 Key
More informationSPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
More informationAMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
More informationMultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
More informationIncorporation of Ethyl Glucuronide into rat hair: influence of hair h pigmentation and ethanol dose
Incorporation of Ethyl Glucuronide into rat hair: influence of hair h pigmentation and ethanol dose Hicham Kharbouche University Center of Legal Medicine, Lausanne-Geneva ICADTS, Oslo 24 August 2010 Reference:
More informationElectrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
More informationAPPLICATIONS MANUAL. Ifosfamide in blood serum... 2. Ingredients in blood serum... 3. Organophosphorus pesticides in tea leaf... 5
1 APPLICATIONS MANUAL CONTENTS Ifosfamide in blood serum... 2 Ingredients in blood serum... 3 Organophosphorus pesticides in tea leaf... 5 Sudan i ii iii iv in chilli sauce... 6 Pah in water... 7 Phenols
More informationApplication Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
More informationReversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
More informationA feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection LGC/R/2011/181 LGC Limited 2011 GC11/0005A A feasibility study of the determination of B-group
More informationApplication Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow Abstract There is high demand in clinical research and forensic toxicology for comprehensive,
More informationEleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
More informationPrevalence of Ethyl Glucuronide and Ethyl Sulphate Among Patients Injured When Driving or at Work
Alcohol and Alcoholism Vol. 50, No. 1, pp. 68 73, 2015 Advance Access Publication 15 October 2014 doi: 10.1093/alcalc/agu070 Prevalence of Ethyl Glucuronide and Ethyl Sulphate Among Patients Injured When
More informationLC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
More informationUnited States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science
CLG-MG/CV2.00 Page 1 of 15 Contents A. INTRODUCTION... 2 B. EQUIPMENT... 3 C. REAGENTS AND SOLUTIONS... 4 D. STANDARDS... 5 E. SAMPLE PREPARATION... 6 F. ANALYTICAL PROCEDURE... 6 G. CONFIRMATION... 8
More informationIntegrating Automated Systems for Regulated Bioanalysis Farmen, R.H., Struwe, P. and Groeschl, M.
Integrating Automated Systems for Regulated Bioanalysis Farmen, R.H., Struwe, P. and Groeschl, M. EBF Meeting Nov 2014 Implementation Challenges Integration of robotic systems into existing bioanalytical
More informationWhat Does the Drug Test Tell Us
What Does the Drug Test Tell Us QUALITATIVE TOXICOLOGY TESTING for CLINICAL MANAGEMENT of the PATIENT Background The qualitative drug test is often referred to as the drug screen - a misnomer Testing performed
More information2013 Annual Toxicology Report
213 Annual Toxicology Report Department of Justice Forensic Science Division Toxicology Department 2679 Palmer Street Missoula, MT 5988 DOJTOX@mt.gov Personnel Phil Kinsey Scott Larson Scott Schlueter
More information