Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity of mass spectrometry results in an extremely useful analytical technique. GC/MS has long been the technique of choice for identification and quantification of analytes in complex mixtures. Because of its utility, extreme efforts are often employed in sample preparation to make difficult compounds amenable to GC/MS analysis. This sample preparation often includes liquid-liquid extraction (LLE), solidphase extraction (SPE), and/or derivatization. Liquid chromatography interfaced to mass spectrometry was the natural progression of chromatographic/mass spectral techniques. However, early attempts at interfacing LC to mass spectrometry had very limited capabilities. The introduction of electrospray ionization helped LC/MS/MS develop from a research tool into a routine technique. Further developments in electronics and computers have resulted in the economical benchtop instruments that are present in a majority of analytical labs today. LC/MS/MS is a very complementary technique to GC/MS. Many of the compounds that are difficult to analyze by GC/MS such as amines and semi-volatile compounds are ideal candidates for LC/MS/MS. By switching assays from GC/MS to LC/MS/MS, many advantages can be realized. First of all, GC is most commonly interfaced to single quadrupole mass analyzers. By moving from single MS to tandem MS/MS, selectivity is greatly enhanced, and therefore improved detection and quantitation limits can be achieved. Secondly, two limitations with GC/MS are that the sample had to be in an organic injection solvent and derivatization is often necessary to improve peak shape, ionization, and/or volatility. LC/MS/MS typically does not have these two requirements, and therefore sample preparation can be greatly simplified. Although sample preparation can include solid-phase extraction (SPE) or liquid-liquid extraction (LLE), it can be as simple as direct injection, dilution, or protein precipitation. By minimizing the amount of sample handling required, savings of both time and money can be attained. Additionally, decreasing the number of preparation steps can minimize areas where errors or contamination can be introduced. Several assays were transferred from GC/MS to LC/ MS/MS and the methods and data compared and contrasted. Unless otherwise stated, all data were collected using an AB SCIEX API 32 LC/MS/MS system. Ion 371. (37.7 to 371.7): 213.D\data.ms TIC 15 1 5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 9.5 1 5 Ion 386. (385.7 to 386.7): 213.D\data.ms 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 9.5 3 2 1 Ion 387. (386.7 to 387.7): 213.D\data.ms 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 9.5 -COOH 5 Ion 473. (472.7 to 473.7): 213.D\data.ms 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 9.5 -COOH 3 2 1 Ion 474. (473.7 to 474.7): 213.D\data.ms 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 9.5 Figure 1. Chromatogram of a low QC (1 ng/ml) analyzed using GC/MS. The total run time was 1 minutes. p 1
in blood Cannabis is one of the most commonly used drugs in the world and is illegal in most countries. The active ingredient is 9-, which is the target analyte along with its major metabolite, 11- nor-9-carboxy- 9-tetrahydrocannabinol (-COOH). An LC/MS/MS method for analysis of and -COOH in blood was developed to replace an existing GC/MS method. For GC/MS, sample preparation consisted of solid-phase extraction followed by derivatization and sample concentration. Use of LC/MS/MS allowed omission of the derivatization step. Representative GC/MS and LC/MS/MS data from a low QC (1 ng/ml) are shown in Figures 1 and 2, respectively. Note the reduction in run times: 1 minutes for the GC/MS analysis and 4.5 minutes for the LC/MS/MS analysis. The long GC/MS run time was necessary to bake out the GC column of excess derivatizing reagent. The LLOQ for LC/MS/MS was also better at.5 ng/ml, versus 2 ng/ml observed for GC/MS analysis. By switching the analysis of and its metabolite from GC/MS to LC/MS/MS, sample preparation was simplified and run times shortened, resulting in significant time savings. Furthermore, the LLOQ improved by a factor of 4x. TIC: from Sample 5 (1. std) of 18.wiff (Turbo Spray) 2.5e5 2.e5 1.5e5 1.e5 5.e4. 1.65 2.6 Max. 2.6e5 cps..5 1. 1.5 2. 2.5 3. 3.5 4. 3 6 89 119 147 177 26 236 XIC of +MRM (4 pairs): Period 2, 315.2/193.2 amu from Sample 5 (1. std) of 18.wiff (Turbo Spray) 8.6e4 8.e4 7.e4 6.e4 5.e4 4.e4 3.e4 2.e4 1.e4. 2.62 2.6 2.96 Max. 8.6e4 cps. 2.4 2.5 2.6 2.7 2.8 2.9 3. 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4. 4.1 4.2 4.3 4.4 6 12 18 24 3 36 42 48 53 59 65 71 77 83 89 95 11 17 112 118 124 XIC of -MRM (4 pairs): Period 1, 343.1/299.3 amu from Sample 5 (1. std) of 18.wiff (Turbo Spray) 5824 5 4 3 2 1 TIC -COOH Max. 5824. cps..2.4.6.8 1. 1.2 1.4 1.6 1.8 2. 2.2 13 25 36 48 6 72 83 95 17 119 13 Figure 2. Chromatogram of a low QC (1 ng/ml) analyzed using LC/MS/MS. The total run time was 4.5 minutes, which is less than half the time required for GC/MS analysis. 1.65 2.7 Opiates in urine Opiates are a class of drugs frequently prescribed for pain relief. However, these drugs are often illegally obtained and abused. Because of the high prevalence of illicit use, frequent testing may be required. This method analyzes urine for the presence of morphine, oxymorphone, hydromorphone, codeine, oxycodone, hydrocodone, and 6-monoacetylmorphine (6-MAM). A comparison of the sample preparation protocols for GC/MS and LC/MS/MS is shown in Figure 3. For GC/MS, sample preparation consisted of hydrolysis, derivatization, and an extensive SPE cleanup procedure. The steps highlighted in blue were omitted for LC/MS/MS analysis and sample preparation was reduced to hydrolysis, centrifugation, and sample dilution. Also note that GC/MS required a sample volume of 1 ml that was subsequently concentrated 1x for analysis. For LC/MS/MS, the sample volume was reduced to.25 ml with a dilution factor of approximately 5x, demonstrating the improved selectivity and sensitivity of LC with tandem mass spectral detection. By eliminating SPE clean-up, significant reductions in both time and consumable costs were achieved. savings were not limited to the front-end sample preparation, but also to the instrumental analysis. Total run time was reduced by 4%, dropping from 1 minutes for the GC/MS method to 6 minutes for the LC/MS/MS method. Also, due to the 1 ml urine sample Add 1 µl IS Add1 µl β-glucuronidase (1, units/ml) Add2 ml.1 M sodium acetate Buffer (ph 4.) Vortex and incubate for 12 min at6 C; Cool to room temp. Add175 µl of 5% HCland 125 µl of 1% methoxyamine HCl 25 µl Urine sample Add 75 µl IS Vortex Add21 µl Of NaOH Add5 µl.1 M sodium acetate Buffer (ph 4.) Add2 µl β-glucuronidase (1, units/ml) Vortex and Centrifuge 13-step SPE Process Evaporate to Dryness under N 2 at3 C Reconstitute In 1 µl EtOAc Vortex and incubate for 12 min at6 C; Cool to room temp. Add85 µl of DI water Vortex and centrifuge Inject Inject Figure 3. Sample preparation for analysis of opiates in urine. (Top) GC/MS sample preparation protocol. The steps highlighted in blue could be omitted when LC/MS/MS analysis was used. (Bottom) Sample preparation protocol for LC/MS/MS analysis uses a smaller sample volume and is much simpler than the GC/MS preparation. p 2
XIC of +MRM (21 pairs): 3./199.2 Da from Sample 1 (5ng/ml) of 586-8_pos.wiff (Turbo Spray), Smoothed 24 22 2 18 16 14 12 1 8 6 4 2 Max. 245. cps. 6-MAM 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2. 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3. 3.1 3.2 3.3 3.4 3.5 3.6 Figure 4. Representative chromatogram for opiates in urine. Data shown are from a 5 ng/ml calibrator, which is the LLOQ for this assay. The linear range was 5 1, ng/ml. 3.1 3.2 sensitivity and derivatization requirements, 6-MAM was a separate, dedicated GC/MS method. However, all opiates, including 6-MAM, were included in the LC/MS/MS extraction and analytical method. Figure 4 shows a representative chromatogram at 5 ng/ml, which was the working LLOQ for this method. When results from the LC/MS/MS method were compared with the validated GC/MS method, there was a correlation of 9%. The inter- and intra-day CVs were <1% for both the high and low QCs. The data quality was also greatly improved, as measured by a reduction in the sample re-run rate, which dropped from an average of 5 samples per batch of 3 samples to 1 sample per batch of 3. By transferring the analysis of opiates in urine to an LC/MS/MS method, significant advantages were realized while maintaining and even improving data quality, which greatly reduced the number of samples requiring repeat analysis. Previous use of GC/MS required two runs and long, extensive sample preparation protocols for complete panel analysis. When analyzed using LC/MS/MS, sample size was reduced, sample preparation simplified, run time shortened, and only one run was necessary. These benefits of LC/MS/MS resulted not only in time savings but also significant reductions in consumable and labor costs. Benzodiazepines in urine Figure 5. Total ion chromatogram for GC/MS analysis of ten benzodiazepines. Total run time was 12 minutes per sample. XIC of +MRM (39 pairs): 286.1/121.1 Da from Sample 4 (STD URINE 5 ng/ml) of 28927_Curves.wiff (Turbo Spray), Smoothed 1.6e4 1.5e4 1.4e4 1.3e4 1.2e4 1.1e4 1.e4 9. 8. 7. 6. 5. 4. 3. 2. 1.. 7-Aminoclonazepam 1.97 7-Aminoflunitrazepam Midazolam Flurazepam Oxazepam Max. 1.6e4 cps..42 1.591.76.62.78 2.37 1.33 2.76 3.12 3.32 1.13 2.58 3.54 3.77.29 3.84 4.19 4.49 5.74 5.87 4. 4.5 6..5 1. 1.5 2. 2.5 3. 3.5 5. 5.5 Figure 6. Extracted ion chromatograms for analysis of 14 benzodiazepines using LC/MS/MS. Not all analytes are labeled in the figure. Data shown were obtained from a 5 ng/ml calibrator, and the LLOQ for all analytes was 1 ng/ml or better. Total run time for this analysis, including column equilibration, was about 8 minutes. Clonazepam Flunitrazepam Temazepam Diazepam Alprazolam Benzodiazepines are frequently prescribed for sleep disorders, anxiety, stress, and several other conditions. Because of the high prevalence of these drugs, they are often involved in illicit use and forensic cases. Traditionally, benzodiazepine screening has been performed utilizing immunoassay with GC/MS confirmation and quantification. GC/MS analysis of benzodiazepines requires hydrolysis followed by extensive SPE clean-up and derivatization in order to detect the low concentrations, similar to the sample preparation for opiates. For GC/MS analysis of benzodiazepines, 1 ml of urine was hydrolyzed, cleaned up using a multi-step SPE procedure, derivatized, and reconstituted in.1 ml (a 1x concentration) prior to analysis. Sample preparation for LC/MS/MS analysis started with.5 ml of urine and consisted of hydrolysis, centrifugation, and 1x dilution, similar to the preparation procedure outline in Figure 3. Total analytical run time was again significantly reduced, as illustrated in Figures 5 and 6, which show chromatograms from GC/MS and LC/MS/MS analyses, respectively. When GC/MS analysis is utilized, total analytical run time is about 12 minutes, versus 8 minutes using LC/MS/MS. As with the opiate analysis, transferring the method from GC/MS to LC/MS/MS resulted in several advantages: simplified sample p 3
preparation, including omission of time-consuming SPE and derivatization steps, and shorter analytical run time. 5.e4 Norfentanyl 1.93 Analysis across drug classes One of the most significant advantages of being able to use a simple, relatively generic sample preparation is that it allows analysis of drugs across multiple compound classes in a single method. An example of this multi-class drug analysis is demonstrated in Figure 7, which shows analysis of 9 opiates and 1 benzodiazepines in 6 minutes. Sample preparation was the same as depicted in Figure 3. 4.5e4 4.e4 3.5e4 3.e4 2.5e4 2.e4 1.5e4 1.e4 5. Morphine Oxymorphone Hydromorphone Codeine 6-MAM Hydrocodone 7-Aminoclonazepam Fentanyl Oxazepam Desalkylflurazolam Temazepam Diazepam The ability to analyze across drug classes can be extended to confirmation panels or comprehensive drug screening. Figure 8 shows a representative chromatogram of a confirmation panel for 3 commonly abused illicit and prescription drugs. Even with a high number of analytes, precision and accuracy can be maintained by proper method setup and use of appropriate internal standards (Figure 9). Although Figure 9 presents statistics for morphine, statistics for other compounds were of similar quality. Figure 1 shows a multi-target drug screen for over 7 drugs using an AB SCIEX 32 QTRAP instrument. The method takes advantage of the hybrid triple quadrupole/linear ion trap capabilities of the instrument by using the selective MRM mode to detect the target compounds, and the sensitive full-scan capabilities of the ion trap to obtain fullscan MS/MS spectra of the analytes. These spectra can be searched against a library for compound identification and confirmation. A representative report showing library confirmation for temazepam is presented in Figure 11. Summary Many toxicological analyses are being transferred from GC/MS to LC/MS/MS to take advantage of: reduced sample preparation and handling; improved detection limits; and faster analytical run times. This reduction in sample handling, consumables, and time helps reduce costs and minimize potential for errors or contamination. Furthermore, because the sample preparation protocols are relatively generic, it is easy to analyze drugs across multiple classes, resulting in even greater time savings. LC/MS/MS can be utilized to analyze multiple drug classes in a single assay for quantitation/confirmation, as well as comprehensive drug screening. Because of these advantages, as well as the versatility of hybrid instruments, the use of LC/MS/MS is rapidly increasing in the forensic toxicology field and is quickly becoming the analytical technique of choice...5 1. 1.5 2. 2.5 3. 3.5 4. Figure 7. Combined benzodiazepine and opiate analysis. Nine opiates and ten benzodiazepines were analyzed in a single method. Data shown are from a 1 ng/ml calibrator (1 ng/ml for fentanyl), and total run time was 6 minutes. The LC conditions used for this analysis are different from the conditions used for the benzodiazepine and opiate analyses previously discussed. XIC of +MRM (71 (71 pairs): pairs): 278.2/234.2 amu Expected amu Expected RT: 4.4 ID: RT: EDDP 4.4 ID: 1 from EDDP Sample 1 from 9 (Calib Sample 3: dil 5x) 9 (Calib of Test 3: Pos dil Stock 5x) of A Test smrm_sept... Pos Stock A smrm_sept... 3.e5 2.8e5 2.6e5 2.4e5 2.2e5 2.e5 1.8e5 1.6e5 1.4e5 1.2e5 1.e5 8.e4 6.e4 4.e4 2.e4 Morphine Hydromorphone Benzoylecgonine Methamphetamine MDMA Norfentanyl Max. 3.1e5 Max. cps. 3.1e5 cps.. 1. 1.2 1.4 1.6 1.8 2. 2.2 2.4 2.6 2.8 3. 3.2 3.4 3.6 3.8 4. 4.2 4.4 4.6 4.8 5. 5.2 5.4, min Figure 8. Analysis of a 5 ng/ml calibrator for 3 illicit and prescription drugs in urine. Total run time was 6.5 minutes. Not all drugs are labeled in this figure. Tramadol Cocaine Fentanly EDDP 4.47 Methadone p 4
Figure 9. Data quality for a multi-analyte panel was maintained as supported by the precision, accuracy, and linearity. Results for morphine are displayed, and similar results were obtained for other analytes, though the upper limit of quantification was lower than 5, ng/ml for some compounds. Figure 11. Identification of temazepam from a urine sample analyzed using the 7-compound multi-target LC/MS/MS screening method. The MS/MS spectrum from the unknown sample is on top and the library spectrum is on the bottom. 2.4e6 2.2e6 2.e6 1.8e6 Acknowledgments The author would like to thank Albert Elian, Massachussetts State Police; Timothy D. Dahn, AIT Laboratories; and Sumandeep Rana, Redwood Toxicology for their contribution of data and/or samples. 1.6e6 1.4e6 1.2e6 1.e6 8.e5 6.e5 4.e5 2.e5. 2.9 1. 2. 3. 4. 5. 6. 7. 8. 9. 1. 11. 12. 13. 14. Figure 1. Multi-target LC/MS/MS screening for >7 drugs. MRM mode was combined with dependent full-scan MS/MS spectral acquisition. The MS/MS spectra were searched against a library for compound identification and confirmation. For Research Use Only. Not for use in diagnostic procedures. 21 AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Publication number: 13721-1 p 5