AUTOMATIC NETWORK ANALYZER PROCEDURES FOR 5045 KLYSTRON CAVITIES * J.G. Judkins



Similar documents
ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER

Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows

Agilent PN RF Component Measurements: Amplifier Measurements Using the Agilent 8753 Network Analyzer. Product Note

Agilent Measuring Noninsertable Devices

2. The Vector Network Analyzer

Performing Amplifier Measurements with the Vector Network Analyzer ZVB

1. The Slotted Line. ECE 584 Microwave Engineering Laboratory Experiments. Introduction:

Agilent 8720 Family Microwave Vector Network Analyzers

Power Amplifier Gain Compression Measurements

Applying Error Correction to Network Analyzer Measurements. Application Note Table of Contents. Page

Impedance 50 (75 connectors via adapters)

Making Spectrum Measurements with Rohde & Schwarz Network Analyzers

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

Shielding Effectiveness Test Method. Harbour s LL, SB, and SS Coaxial Cables. Designs for Improved Shielding Effectiveness

One Port Network Analyzer

Experiments Using the HP8714 RF Network Analyzer

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term

Spectrum Analyzers And Network Analyzers. The Whats, Whys and Hows...

Network Analysis. Specifying calibration standards for the HP 8510 network analyzer Product Note A

HPM-5 Performance Testing

Improving Network Analyzer Measurements of Frequency-translating Devices Application Note

"FP", "FR", "FQ" Series Bandpass Filters

Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer

Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements

Agilent AN In-Fixture Measurements Using Vector Network Analyzers

MATRIX TECHNICAL NOTES

14.5GHZ 2.2KW CW GENERATOR. GKP 22KP 14.5GHz WR62 3x400V

RF Network Analyzer Basics

HEWLETT PACKARD. HP 8510C Network Analyzer 45 MHz to 110 GHz. Unmatched excellence in microwave network analysis

Paul Wade, W1GHZ. 161 Center Rd Shirley, MA Figure 1 WR-75 waveguide to coax transition for 10 GHz. 1 Notes appear on page 16.

Tuning a Monopole Antenna Using a Network Analyzer

Revision 1.10 April 7, 2015 Method of Implementation (MOI) for 100BASE-TX Ethernet Cable Tests Using Keysight E5071C ENA Option TDR

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

Investigations on Critical Higher Order Modes in CEBAF Upgrade Cavities

Utilizing Time Domain (TDR) Test Methods For Maximizing Microwave Board Performance

KLYSTRON TUNING PROCEDURES:

MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER

Agilent AN Improving Throughput in Network Analyzer Applications. Application Note

Experiment 7: Familiarization with the Network Analyzer

UNDERSTANDING NOISE PARAMETER MEASUREMENTS (AN )

Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz

Cable Analysis and Fault Detection using the Bode 100

Agilent FieldFox RF Vector Network Analyzer N9923A Quick Reference Guide

Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters

AMERICAN NATIONAL STANDARD

Techniques for Precise Cable and Antenna Measurements in the Field

Site Master Cable and Antenna Analyzer with Spectrum Analyzer

Release Notes. R&S ZVH4/ZVH8 Cable and Antenna Analyzer

iva Cable & Antenna Analyzer

Network Analyzer Basics- EE142 Fall 07

S-PARAMETER MEASUREMENTS OF MEMS SWITCHES

Techniques for precise cable and antenna measurements in the field

spinner Measurement & Calibration equipment for network analyzers

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

Z Analyze Software. User Manual Version April Frontier Technical Services 1

W-band vector network analyzer based on an audio lock-in amplifier * Abstract

Agilent 10 Hints for Making Better Network Analyzer Measurements. Application Note B

Probes and Setup for Measuring Power-Plane Impedances with Vector Network Analyzer

MEASUREMENT SET-UP FOR TRAPS

R&S ZNC Vector Network Analyzer Specifications

EE 186 LAB 2 FALL Network Analyzer Fundamentals and Two Tone Linearity

Successfully negotiating the PCI EXPRESS 2.0 Super Highway Towards Full Compliance

Agilent E5100A Network Analyzer

Transmitter Interface Program

Fast and Accurate Test of Mobile Phone Boards

Software Utility VNA Frequency Converter Leveling Tool Getting Started

Network Analyzer Operation

Copyright 1996 IEEE. Reprinted from IEEE MTT-S International Microwave Symposium 1996

Solving Difficult Cable Measurements

FREQUENCY RESPONSE ANALYZERS

C.-K. Ng. Stanford Linear Accelerator Center. and. T. Weiland. University oftechnology. FB18, Schlossgartenstr. 8. D64289, Darmstadt, Germany.

Product Name Hexa-Band Cellular SMD Antenna GSM / CDMA / DCS / PCS / WCDMA /UMTS /HSDPA / GPRS / EDGE 800 MHz to 2200 MHz

Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers

TECHNOTES. TD1001: The best RF protector for applications not requiring dc on the coax.

Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer. Application Note

Capacitor Self-Resonance

Agilent E5063A ENA Series Network Analyzer. 100 khz to 4.5/ 8.5/18 GHz

ELECTRICAL ENGINEERING DEPARTMENT. California Polytechnic State University SIGNAL TRANSMISSION LABORATORY EE 353

Measurement & Modeling Device Characterization Solutions

Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform

Agilent Network Analysis Applying the 8510 TRL Calibration for Non-Coaxial Measurements

EVAL-UFDC-1/UFDC-1M-16

khz to 2.7 GHz Spectrum Analyzer

Orthomode Transducer at 43GHz

A Guide to Calibrating Your Spectrum Analyzer

R&S ZNBT8 Vector Network Analyzer Specifications

Part 2 Designing Combline Filters with Free Software Paul Wade W1GHZ 2014

R&S ZCxxx Millimeter-Wave Converters Specifications

Sentinel I28 and Blackbelt Mass Flow Calibration and Verification

APPLICATION NOTES POWER DIVIDERS. Things to consider

R&S ZNC Vector Network Analyzer Solid performance on a future-oriented platform

The Design & Test of Broadband Launches up to 50 GHz on Thin & Thick Substrates

Scalar Network Analysis with the HP 8590 Series Spectrum Analyzers Product Overview

Agilent Specifying Calibration Standards for the Agilent 8510 Network Analyzer

OMNIYIG. .5 TO 2 GHz, 1 TO 4 GHz, 6 to 18 GHz Thin Film YIG-TUNED OSCILLATORS OUTLINE DIMESIONS MOUNTING SURFACE GND +15V + TUNE - HTR HTR

MAURY. VNA Calibration Kits, Microwave Components & Adapters. Calibrate With Confidence IN THIS CATALOG:

Field Calibration Software

Introduction. Description of Measurement Techniques

Measuring RF Parameters of Networks Bill Leonard NØCU

Transcription:

Introduction SLAC-TN-91-10 July 1991 AUTOMATIC NETWORK ANALYZER PROCEDURES FOR 5045 KLYSTRON CAVITIES * J.G. Judkins Stanford Linear Accelerator Center Stanford University, Stanford, CA 94309 This Note describes the results of using Automatic Network Analyzers in measuring SLAC 5045 klystron cavities. Two different analyzers were compared; the HP8753 and HP8510. Both analyzers have frequency synthesizer accuracy and stability to perform the measurement without the need for a frequency counter. The klystron has six cavities which can be put into three categories; input, gain and output. The input and output cavities require an external Q measurement (Qe) to determine coupling (p) and center frequency (fo). The gain cavities require a resonant frequency measurement only. Overview Both analyzers were suitable for measuring input and gain cavities. The HP8510 has an advanced calibration technique that enables it to measure high reflections accurately. The HP8753 does not have this technique which makes it unsuitable for output cavity measurements. This will be discussed in more detail later. The input and output cavities are greatly overcoupled (p >> 1) which results in an Sll measurement near unity around the periphery of the Smith chart. The analyzers Smith chart display lends itself to determine Qe easily for such high reflections. The gain cavities require input/output coupling probes to determine resonance. The probes are inserted side-by-side into the drift tube and the analyzer is set for transmission measurement (S21). Th e coupling between the probes is approximately -25dB. This is used as the baseline to determne the resonant frequency peaks. The resonant frequency is determined by loosely coupling the probe assembly to the appropriate cavity. Over-coupling at this point will detune the cavity. Therefore, the probes are inserted until the resonance peak is barely discernible on the LOG magnitude display of the analyzer. Most of the gain cavities have a frequency specification of f2mhz. Moving the probes in and out; you can see the effect of overcoupling on frequency determination. The output cavity requires a waveguide calibration. Two calibration techniques were tried and only one proved accurate. The Thru-Reflect-Line (TRL) method proved accurate and consistent with slotted line measurements. The TRL cal is available on the HP8510 and not the HP8753. The Cal method for the HPS753 (TN) * Work supported by the Department of Energy, contract DE-AC03-76SFOO51.5.

uses a short, quarter-wave short, and a load. The load is the weak part of this method. Although it works well for moderately high reflections, it is not accurate for very high or small reflections in the order of 0.98 < Sll < 0.02. The quality of the load determines the accuracy of the directivity component of the calibration. A waveguide load with a return loss of 50 db would be necessary to accomplish the measurements within spec. This high quality load is attainable but fragile and difficult to maintain. The waveguide load used in the comparison tests has a return loss of 36 db (1.02:1 VSWR). Th is was not sufficient to measure the output cavity reflection magnitude of 0.99 within spec. The TRL calibration standards are lower cost, easy to maintain and more accurate. The short, quarter-wave short, load technique was tried on both the HP 8510 and HP 8753 with the same undesirable results. Summary In comparison to the slotted line, the network analyzer takes more time to calibrate but measures much faster. For consistent results the analyzer should be calibrated every four hours. Several cavities could be measured with one calibration faster than the slotted line. With the addition of a PC, documentation can be down-loaded automatically. The data stored on disc is readily available for statistical analysis. The measurement procedures and configurations can be guided by software. This speeds up the setup process, improves repeatability and consistency of test results. In summary, the HP 8510 will provide all the measurements necessary for 5045, PEP and X-band klystrons. The HP 8753 is limited to 5045 input/gain cavities and PEP tube cavities. Although the PEP tube cavities were not part of this analysis, the methods in general can be applied. The PEP tube input and output cavities have Type-N adapters which makes them suitable for either network analyzer.

INPUT CAVITY MEASUREMENT The input cavity has a Type-N coaxial input to the coupling loop. This connector will be treated as part of the cavity. A Type-N, Sll-1 port cal will work on either the HP 8753 or HP 8510 network analyzers. The HP 8510 requires a sliding load to characterize the directivity; whereas the HP 8753 uses a fixed load. The HP 8753 has a high precision fixed load specified to 3 GHz which covers its entire frequency range. The fixed load for the HP 8510 is specified to 2 GHz. It may be worthwhile to acquire the higher precision Type-N load to negate the need for a sliding load below 3 GHz. NOTE : The input cavity bandwidth is narrower than the output. It should not be necessary to use Frequency Subset to determine the detune short position. If a fine dot cannot be achieved then the Frequency Subset method described in the output cavity section needs to be performed. SETUP 5045 INPUT CAVITY TEST PROCEDURE FOR HP 8510 Press PRESET STIMULUS START FREQ: enter 2.84 GHz STOP FREQ: enter 2.88 GHz NUMBER OF POINTS: select 401 CAL If Cal Kit Type-N B.l is not already in CAL 1 or CAL 2; then load it from DISC or TAPE.

CALIBRATION ELECTRICALDELAY MEASURE : CAL Select TYPE-N Select Sll B.l 1 PORT Perform short, open, load calibratioin as indicated using Type-N Cal Kit. After calibrating, connect network analyzer Port 1 to the cavity under test and insert shorting bar. FORMAT Select INVERTED SMITH CHART Under RESPONSE MENU select ELECTRICAL DELAY. With the shorting bar in place, adjust ELECTRICAL DELAY either by key entry or knob to get a fine dot at the detuned short position (approximately 1.3 nsec). Remove the cavity shorting bar and note fo at the point where the arc crosses the real axis of the Smith Chart. Invoke markers 1,2 and 3 to measure fo and Qe. The Qe bandwidth points are determined where the Sll reflection arc intersects the inverted Smith Chart susceptance arc of jp = 1. See Figure 1. 5045 GAIN CAVITY MEASUREMENTS FOR HP 8510 The gain cavities require a center frequency measurement only. Since there are no coupling mechanisms into these cavities, an input and output probe assembly is inserted into the drift tube. It is important to loosely couple to the cavity so as not to pull the frequency resulting in an erroneous measurement. The probe assembly is inserted just far enough to discern the resonant frequency peak from the baseline. Network analyzer calibration is not required for this measurement. The free space coupling between the probes is approximately -25 db. This is used as the baseline. SETUP Press PRESET STIMULUS

MEASURE START FREQ.; enter 2.8 GHz STOP FREQ.; enter 3.0 GHz NUMBER OF POINTS; select 801 PARAMETER FORMAT SELECT S21 Select LOG MAG display Connect network analyzer Port 1 and Port 2 to the dual probe assembly. Insert the probe assembly into the drift tube until a small resonant peak is noticed on the LOG MAG display. The frequency of resonance can be read by placing a marker on top of the peak. Notice that inserting the probe further will cause the peak to shift frequency. Therefore, it is important to couple as loosely as possible and still be able to discern the resonant peak. It is possible that false resonances can occur in the drift tube. The probe depth is read off the scale on the probe carriage. This will insure that the cavity is excited and not a coaxial mode in the drift tube. See Figures 2, 3 and 4.

5045 OUTPUT CAVITY MEASUREMENTS FOR HP 8510 Output cavity measurements can be made with or without the dual window section attached. The cavity is greatly overocupled (p >> l), which causes a very high reflection (Sll N 0.99). An Sll measurement is used to determine the characteristics of the cavity. The high reflection requires an accurate calibration to minimize errors in connections and characterization of adapters. SETUP Press PRESET STIMULUS START FREQ; enter 2.75 GHz STOP FREQ; enter 2.95 GHz NUMBER OF POINTS; select 801 SWEEP TIME > 200 msec Response Menu: Press MORE, select WAVEGUIDE DELAY; Enter 2.078 GHz for CUTOFF CAL If Cal Kit WR-284 from disc or tape. Press MORE; FREQ. is not already in Cal 1 or Cal 2; then load it select SET 20; enter 1 ohm. The need for TRIM SWEEP routine is that RAMP SWEEP can be as much as 3 MHz off frequency accuracy. TRIM SWEEP corrects for this difference and allows RAMP SWEEP to approach STEP SWEEP accuracy.

ADJUSTTRIMSWEEP PARAMETERMENU : CALIBRATION ELECTRICALDELAY Connect Ports 1 and 2 with or without FORMAT SELECT S21 select PHASE SCALE STIMULUS DISPLAY 1 degree per devision select STEP sweep Wait for display to update with one full sweep (- select DATA-MEMORY select MATH(/). STIMULUS select RAMP sweep. Display should be a flat line. coax to waveguide adapters. 40 seconds). CAL press MORE; select TRIM SWEEP and adjust knob for flattest trace. Select WR-284 calibration Select TRL 2-Port Perform TRL calibration as indicated. There is no need to perform ISOLATION Cal. To omit, select ISOLATION; then select OMIT ISO- LATION. After calibration select Sll, INVERTED SMITH CHART. The FREQUENCY SUBSET is used to reduce sweep width in order to determine electrical delay. It is necessary to use FREQUENCY SUB- SET and TRIM SWEEP method to achieve the accuracy required to determine fo and Qe. The output cavity Specs for fo are 2853 MHz; +3MHz, -8MHz. If we allow the network analyzer to have 10% error in determining fo, then the window of error would be about 1 MHz or f0.5mhz. This tanslates into a tolerance of 1.2 psec for adjusting electrical delay. The reduced sweep width provides a finer dot to achieve the detuned short position accuracy. After determining electrical delay, the full sweep will be turned back on to perform the cavity measurements. 7

CAL Press MORE; Select MODIFY CAL SET; then Select FREQUENCY SUBSET. Select CENTER FREQ; enter 2.853 GHz. Select SPAN; enter 20 MHz. Select CREATE & SAVE, and store in an unused CAL SET. Attach the coax to WR-284 adapter to the cavity waveguide. Place the shorting rod in the cavity. RESPONSEMENU : Select ELECTRICAL DELAY. Adjust DELAY to achieve a fine dot at the detuned short position. This is approximately 6.2 nsec with the window structure attached. After correcting for electrical delay, Select the original-full sweep-cal set to perform the cavity measurements. The electrical delay will be automatically carried over to the wider bandwidth Cal. Measure: Remove the cavity short and note fo as the point where the arc crosses the real axis. The Qe bandwidth points are determined where the Sll arc intersects the susceptance arc of jp = 1. See Figure 5. The TRIM SWEEP adjustment can be checked by switching between wide and narrow Cal sets and noting the fo for each. The fo should be within 1 MHz.

CH2 Sll/M : U FS A 617.63 MS 009.24 ps.5140 CF MHz P /,, \ / \. \ I 2 / \ \ \ \ \ \ \ I -_.------. r- - \ 4 Cor Del Avg a t START 2 650.000 000 MHz 3 STOP 2 870.000 000 MHZ 7-s 1 6979Al Fig. 1

7-91 6979A2 Fig. 2, i

Fig. 3 6979133

t markq 3 I I I I I I I I I J I I I I I I I I 7-9 1 6979A4 Fig. 4

Sll Y REF 1.0 Units 200.0 mun i t*/ -7.792 ms 1.9966 S / I / 0\ L a \\\ N-H\ \ / h. \ -2--- START 2.750000000 GHz STOP 2.950000000 GHZ 7-9 1 6979A5 Fig. 5