1 Analytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination J. Wang, M. K. Balazs Balazs Analytical Laboratory, Sunnyvale, CA P. Pianetta, K. Baur and S. Brennan Stanford Synchrotron Radiation Laboratory (SSRL), Stanford University, Stanford, CA Abstract Direct measurement of trace elements on wafer surfaces by is quick and nondestructive. SR- has similar features to, but much better detection limits mainly due to the high flux. VPD- and VPD-SR- improve the detection capability to various degrees. However, some elements (e.g. Cu) may have recovery problems at certain concentrations. VPD-ICP-MS is a destructive technique because of dissolving native oxide, but it can analyze most of elements in the periodic table and especially well for low Z elements. By using NIST standards for calibration, quantification can be verified relatively easily. All of these techniques can be complementary to each other and provide comprehensive analyses for the semiconductor industry. Introduction Ultra-clean Si wafer surfaces are one of the most critical factors to the ULSI manufacturing because the uncontrolled contamination during the wafer processing can change the electrical characteristics, resulting in yield loss 1. Metallic impurities at certain concentrations on wafer surfaces can cause serious device degradation such as diminished carrier lifetimes, dielectric breakdown of gate oxides, threshold voltage shift and leakage current of P-N junction. Sources of trace metal contamination include processing chemicals used during cleaning, stripping, photolithography, and deposition; processing equipment (furnaces, reactors, implanters); and wafer handling. Current methods for metal analyses on wafer surfaces include total reflection x-ray fluorescence () 2-7, vapor phase decomposition total reflection x-ray fluorescence (VPD-) 8-11, vapor phase decomposition inductively coupled plasma mass spectrometry (VPD- ICP-MS) 12,13 and synchrotron radiation total reflection x-ray fluorescence (SR-) 14,15. In analysis, x-rays at an incidence angle below the critical angle strike on the highly polished silicon surface and are reflected in a manner similar to the total reflection of visible light from a mirror surface. Under this condition, the penetration 1
2 depth can be restricted to a few nano-meters, leading to a substantially reduced background signal compared to conventional XRF. The excited fluorescence signals of the contaminants are emitted from the wafer surface and detected by an energy dispersive, Si(Li) solid state detector. Using synchrotron radiation (SR) as a primary excitation source can increase the overall sensitivity of. Compared to regular x-ray sources produced by electron bombarding on metal targets, synchrotron radiation is produced as a natural byproduct of circulating or oscillating electrons in a storage ring. It contains all the wavelengths of the electromagnetic spectrum and is 100 or more times powerful than a x-ray generator of the standard instrument. Primary excitation source of synchrotron radiation has several advantages over a conventional x-ray tube: The high incidental flux combined with low divergence results in higher fluorescence intensities and so lower detection limits. Due to its linear polarization, the elastic scattering of the incoming beam can be reduced. The spectral background, which is already reduced by total reflection, is lowered further. The tunability of synchrotron radiation allows improving the sensitivity for special elements by increasing the photon absorption cross-section. In the methods of VPD-ICP-MS or VPD-, a wafer is exposed to saturated HF vapor, which reacts with the surface native or thermal oxide (vapor phase decomposition). The matrix of silicon dioxide is removed by formation of SiF 4 (gas form). A liquid droplet is then pipetted onto the wafer surface and moved over the entire surface area of the wafer for VPD residue collection. The droplet is transferred either automatically (by an instrument) or manually into a sample vial for elemental analysis by ICP-MS. In ICP-MS analysis, solutions are nebulized into an argon plasma. The contaminants dissolved in the solutions are vaporized, atomized, ionized and extracted into a mass spectrometer for analysis. During the past 17 years, ICP-MS has been developed to a mature technique for many applications. One of the advantages of VPD- ICP-MS is that a wide range of elements in the periodic table from Li to U can be analyzed at one time. The disadvantage of the VPD preparation is the tedious procedure of collecting the reaction products on the wafer surface. If the VPD droplet dries on the wafer surface, the resulting residue can be measured by. The sensitivity enhancement of VPD- over direct measurement can be estimated from the ratio of the total wafer surface area (except edge exclusion) to the instrumental sampling area (corresponding to the solid angle of instrument detector). For example, a 200-mm wafer with 5-mm edge exclusion has a total surface area of cm 2. The sampling area is 0.5 cm 2 for an Atomika 8030W, and is cm 2 for the SR- at SSRL. The ratio of sensitivity enhancement by this pre-concentration process will be 1:567 and 1:2250, respectively. Although the VPD process loses the spatial contamination information because contamination from the entire wafer surface are collected, it may be advantageous since it integrates all contaminants into one residual droplet, thereby reducing the numbers of analysis spots on the wafer surface. Figure 1 shows a summary of detection limits of, VPD-, VPD-ICP-MS and SR-. Table 1 lists their advantages and disadvantages. 2
3 Copper (Cu) has been well known for its high diffusivity and solubility in silicon bulk. Copper contamination could cause gate oxide integrity degradation, premature breakdown and P-N junction leakage. Trace amounts of copper could be introduced into silicon wafers during the thermal processing, wet cleaning or other steps of silicon fabrication 16. In addition, new copper interconnection processes introduce greater potential for copper contamination by diffusion, migration from the backside of the wafer and by cross contamination (airborne or wafer handling) to copper sensitive processes. This paper will give a comparison of, SR- and VPD by applying these techniques on the analysis of a Cu droplet residue on a Si wafer surface. In addition, issues associated with the quantification of Cu will be discussed. Advantage Disadvantage Nondestructive VPD- Destructive VPD- ICP- MS Destructive SR- Capable of Mapping Minimum matrix effect DL = ~10 10 atoms/cm 2 Some DL s are lower than VPD-ICP-MS DL = ~ atoms/cm 2 A wide range of elements from Li to U can be analyzed. Very sensitive for Na, Mg, Al, Li, Be, also B, P. DL = ~ atoms/cm 2 Lowest DL for direct measurement Capable of Mapping; Minimum matrix effect; Energy tunable for special applications DL = ~10 8 atoms/cm 2 Cl, Br, S can be measured Cl, Br, S not available Nondestructive Cl, Br, S can be measured Low Z Elements Na, Mg, Al difficult, Li, Be, B not available Low Z Elements Na, Mg, Al difficult, Li, Be, B not available Lose local contamination information Tedious procedure; Drying conditions critical and some problems with quantification Tedious procedure; Lose local contamination information; must consider matrix effects Instrumentation cost very high and not easily available Table 1. Comparison of, VPD-ICP-MS, VPD- and SR-. DL s for VPD-ICP- MS and VPD- are based on 200-mm wafers Experimental Wafer preparation and analyses by VPD- and VPD-ICP-MS were performed in an ultra clean environment in a cleanroom. After depositing a known amount of Cu solution (e.g. 50 µl) on a bare Si wafer surface, the size of the droplet was found to be 4-5 mm in diameter. After drying at room temperature within a clean box with nitrogen purge, the size of the dry spot decreased to less than 1 mm in diameter. The dry spot on the wafer was then analyzed by a instrument (Atomika 8030W). The surface concentration was calculated based on a sampling area of 0.5 cm 2. A standard wafer with a Ni dry spot on the wafer surface provided by the manufacturer was used to calibrate the instrument. Standard procedures were used to ensure the 3
4 1.E+12 VPD-ICP-MS Surface Concentration 1.E+11 1.E+10 1.E+09 VPD- VPD-ICPMS VPD- S- 1.E+08 S- SIA 99 Roadmap 1.E+07 Li B Na Mg Al P S Cl K Ca Ti V Cr Mn Fe Co Ni Cu Zn As Br Sr Ag Sn Sb Ba Ta W Pt Au Pb Th/U Figure 1. Detection limits of, VPD-ICP-MS, VPD- and SR- on 200 mm Si wafer. (DL s for are based on the instrument detection limits from Atomika; DL s for VPD-ICP-MS are based on the routinely reporting detection limits from Balazs Analytical Lab; DL s for VPD- are based on the measurements of 13 dry spots of diluted HF on a clean wafer and their standard deviation multiplied by 3 at Balazs Analytical Lab; DL s of SR- are estimated based on the findings from SSRL) reproducibility for all analyses. Synchrotron radiation SR- experiments were performed at the Stanford Synchrotron Radiation Laboratory (SSRL) at a horizontally focussed wiggler beamline (6-2) using a double multilayer monochromator. The surface concentration was calculated based on a sampling area of cm 2. Experimental conditions can be found in reference 14. Discussion Elements 1. Dry spots of Cu solution on a wafer surface studied by : A single droplet of copper (Cu) solution was dried on a wafer surface. It was found that in the analysis of Cu dry residue by, the results of Cu usually were lower than expected when the surface concentration of Cu was above a certain concentration. For example, a single droplet containing 1 ng Cu deposited on a wafer should have a surface concentration of 1890 E10 atoms/cm 2 (theoretical value) under a sampling area 0.5 cm 2. However, an average of 1612 E10 atoms/cm 2 from five replicates was found, indicating 15% lower than the theoretical result, as shown in figure 2. Similarly, a single droplet of 0.5 ng Cu deposited on the wafer surface should result in a surface concentration of 940 E10 atoms/cm 2. An average of 767 E10 atoms/cm 2 indicates 18% lower than the theoretical results. It is worth noting that at lower surface concentrations, e.g. for a dry spot of 0.05 ng Cu, this phenomenon no longer exists. It needs to be clarified that the 4
5 ng Cu on 0.5 cm 2 sampling area =1890xE10 atoms/cm ng Cu on 0.5 cm 2 sampling area =940xE10 atoms/cm ng Cu on 0.5 cm 2 sampling area =94xE10 atoms/cm Dry spot Cu (ng) Figure 2. Cu result of vs. calculated surface concentration on Cu dry residue. The calculation is based on 0.5 cm 2 sampling area; each bar represents a dry residue analysis on the wafer surface. Average of 1612 E10 atoms/cm 2 for 1 ng Cu dry spot and average of 767E10 atoms/cm2 for 0.5 ng Cu dry spot were found from 5 replicates. analysis of a known amount of Cu (e.g. 1 ng Cu) in a single droplet on the wafer surface is similar to the one resulting from a VPD process. If the spot of 1 ng Cu was equally distributed onto the entire 200-mm wafer surface as a reverse VPD process (with 5-mm edge exclusion), the resulting surface concentration would be 3.3E10 atoms/cm 2. This amount has been found to be a common range of Cu contamination on wafers. There are still some open questions on how Cu diffuses into Si substrate and how the drying process is influenced by the concentration of Cu. In addition, the assumption of formation of various micro-crystals or particulates, different compounds or islands of materials 10,11,17 during the drying process remains debatable. 2. Dry spots of Cu/Ni solution on wafer surface by and ICP-MS: For further investigation, a droplet of Cu/Ni mixed solution was deposited on a wafer surface. The dry spot was analyzed by. Then the dry spot was extracted twice with 0.5 ml solution of H 2 O 2 /HF. The extracted solutions were analyzed by ICP-MS. Figure 3a shows Cu recoveries by and ICP-MS. The Cu results of ICP-MS show consistent recoveries 100% ±10%. By contrast, the Cu recoveries of show ~ 80%. By taking into account that both and ICP-MS have Ni recoveries near 100%, as seen in figure 3b, the possibility of mis-location during measurement can be excluded. It was found that our finding from this work differs from a recent report, in which the results of Cr, Fe, Ni, Cu, and Zn by are all lower than that by ICP-MS 17. One of the possibilities for the discrepancy may attribute to different drying conditions. The mixed Cu/Ni droplet was dried in a chamber 5
6 at room temperature with nitrogen purge for this work; while the mixed elements solution was dried by using an IR heat lamp from that report. Nevertheless, the lower recovery of Cu by could be due to some unknown factors during the drying process. A detailed investigation is on the way Figure 3. Recoveries of (a) Cu (left) and (b) Ni (right) by (dot bars) and ICP-MS (empty bars). Numbers on the x-axis represent results from replicate dry spots. A droplet of solution containing 1 ng of both Ni and Cu (diluted from NIST stock solution) was dried on the wafer surface and analyzed by. The dried spots were then extracted into a solution of H 2 O 2 /HF and analyzed by ICP-MS. The recovery is based on the analytical results vs. known 1 ng of Cu and Ni. 3. Dry spots of Cu solution studied by SR-: To quantify Cu at very low concentration by SR-, a droplet of 0.1 pg Cu in 0.1% HNO 3 (diluted from NIST standard solution) was deposited and dried on the wafer surface. The wafers were transferred to SSRL for SR- analysis within a wafer box sealed with cleanroom tapes. The surface concentration was calculated based on a sampling area of cm 2. Thus theoretical value of 0.1 pg Cu is 7.5 x 10 9 atoms/cm 2. Figure 4a shows that the Cu signal is plotted as a function of position vs. surface concentration as an area on the wafer is scanned through the SR- beam in steps of 0.5 mm. When the detector sampling area 9.00E E E E E E E E E E E E E E E E E E E Wafer Location (mm) Figure 4a (left) and b (right): Cu surface concentration vs. the SR- beam position scanned across the wafer in steps of 0.5 mm. 0.00E Wafer Location (mm) 6
7 covers entire dry spot, a maximum Cu intensity can be seen. When the detector sampling area covers only part of the dry spot, the Cu intensity decreases. Given the potential sources of error when working at these low concentrations, the theoretical value (7.5 x 10 9 atoms/cm 2 ) corresponds well to the measured value of 5.8 x 10 9 atoms/cm 2 in Figure 4a. Another spot shows a result of 8.0 x 10 9 atoms/cm 2 in Figure 4b. This provides an independent verification of the system calibration. It also confirms that the quantification problem is no longer an issue if the surface concentration is at a low level. 4. Residue analysis after Cu removal studied by SR-: A single droplet of 1 ng Cu was deposited and dried on the wafer surface. The surface concentration of the Cu dry spot was measured by. Similar to what was mentioned earlier, the result was found to be lower than the theoretical value. The surface 6.00E E+10 Fe from Marker 4.00E+10 Co from marker 3.00E E E+10 Cu left 4E9 0.00E W afer location (m m) Figure 5. Cu, Ni and Co surface concentrations vs. the SR- beam scanned across the wafer in steps of 0.5 mm. of the wafer was then extracted with H 2 O 2 /HF by scanning the wafer surface using a Padscan instrument (Programmable Automatic Droplet Scanner) for removing Cu residue. The wafer surface was then rinsed with ultra pure water (UPW). At that time, the resulting spot after Cu removal by H 2 O 2 /HF could barely be seen by eyes. It was found to be necessary that two marks by a marker pen be put on the both sides (10 mm apart) of the residue in order to locate the spot. The mark ink contains high levels of Fe and Co can be used as an indicator for estimating the location of the drop when the synchrotron x-ray beam is striking on the wafer. Figure 5 shows the concentrations of Cu, Fe and Co plotted as a function of the wafer position in steps of 0.5 mm. The signals of Fe and Co increase as the x-ray beam 7
8 moves to the marker and excites the ink materials. The maximum surface concentration of Cu was found to be 4E9 atoms/cm 2. It seems that even copper diffuses into Si substrate during the drying process, copper residue can be completely removed after H 2 O 2 /HF extraction because 4E9 is insignificant compared to a theoretical value of 7 E13 atoms/cm 2 (i.e. 1 ng Cu on an area of cm 2 ). Carryover is only 0.006% by calculation in this droplet study. Summary We have made a comparison of, SR-, ICP-MS and VPD techniques to study the Cu concentration when a droplet is dried on a wafer surface. In general, direct measurement of trace elements on wafer surface by is quick and nondestructive. SR- has similar features to, but much better detection limits mainly due to the high flux. Measurement for low Z elements by and SR- are improving but are still in a developing process at the present time. VPD- and VPD-SR- improve the detection capability to various degrees. However, we found that some elements (e.g. Cu) may have recovery problems above a certain concentration. It is worth mentioning that many investigations have been conducted on detection limits on wafer surfaces by, but only a few have assessed recoveries. Therefore many issues concerning the quantification are still unknown. On the other hand, VPD-ICP-MS is a destructive technique because of dissolving the native oxide on wafers, but it can analyze most of the elements in the periodic table and especially well for low Z elements, e.g. B Na, Mg and Al. By using NIST standards for calibration, quantification can be verified relatively easily. All of these techniques can be complementary to each other and provide comprehensive analyses for the semiconductor industry. Acknowledgements JW would like to acknowledge Dr. Mark Camenzind for his comments, suggestions and proofreading. References: 1. Sources of surface contamination affecting electrical characteristics of semiconductors, G. J. Slusser and L. MacDowell,, J. Vac. Sci. Technol. A, Vac. Surf. Films VOL. 5, NO. 4, , Total reflection X-ray fluorecence spectroscopy, R. Klockenkaemper, J. Knoth, A. Prange and H. Schwenke, Anal. Chem., Vol.64, No. 23, 1115A, Dec Analysis of trace metals on silicon surfaces, Streckfuss, N.; Frey, L.; Zielonka, G.; Kroninger, F.; Ryzlewicz, C.; Ryssel, H., Fresenius' J. Anal. Chem. VOL. 343, NO. 9-10, Jul 1992, P Total reflection X-ray spectrometry: method and applications, A. Prange, Spectrochimica Acta. Vol. 44B, No. 5, pp , Basic features of total-reflection X-ray fluorescence analysis on silicon wafers. Berneike, W., Spectrochim. Acta, Part B, VOL. 48B, NO. 2, Feb 1993, P Total Reflection X-Ray Fluorescence Spectrometry, Klockenkaemper, R.1996,
9 7. Calibration procedure for quantitative surface analysis by total-reflection X-ray fluorescence, Torcheux, L.; Degraeve, B.; Mayeux, A.; Delamar, M., SIA, Surf. Interface Anal. VOL. 21, NO. 3, Mar. 1994, P Ultra-trace analysis of metallic contaminations on silicon wafer surfaces by vapour-phase decomposition-total-reflection X-ray fluorescence (VPD/), Neumann, C.; Eichinger, P., Spectrochim. Acta, Part B. VOL. 46B NO. 10, 1991, P Characterization and application of the vapour-phase decomposition technique for tracemetal analysis on silicon oxide surfaces, Hall, L. H.; Sees, J. A.; Schmidt, B. L. Surface and Interface Analysis, VOL. 24, NO. 8, 1996, PP Novel methods of analysis for silicon wafer surface inspection, L. Fabry, S. Pahlke, L. Kotz, P. Wobrauschek and C. Streli, Fresenius J. Anal. Chem (1999) 363: Accurate calibration of using microdroplet samples. Fabry, L.; Pahlke, S.; Kotz, L., Fresenius' J. Anal. Chem. VOL. 354, NO. 3, Feb Measurement of trace metallic contaminants on silicon wafer surfaces in native and dielectric silicon oxide by vapor phase decomposition flow injection inductively coupled plasma mass spectrometry, J. Fucsko, S. S. Tan, and M. K. Balazs, J. of the Electrochemical Society, Vol. 140, No , P Using VPD ICP-MS to monitor trace metals on unpatterned wafer surfaces, G. Settebre, E.Debrah, Micro, June 1998, P Application of synchrotron radiation to analysis of metal contamination on silicon wafer surfaces, P. Pianetta, K. Baur, S. Brennan, D. Werho, and J. Wang, accepted by Thin Solid Films, with synchrotron radiation. Analysis of nickel on silicon-wafer surfaces, Wobrauschek, P., Kregsamer, P.; Ladisich, W.; Streli, C.; Pahlke, S.; Fabry, L.; Garbe, S.; Haller, M.; Knoechel, A.; Radtke, M., Nucl. Instrum. Methods Phys. Res., Sect. A, VOL. A363, NO. 3, 1995, P Effects of copper contamination in silicon on thin film oxide breakdown, D. A. Ramappa and W. B. Henley, J of the Electrochemical Society, 146, (6) , Application of vapor phase decomposition/total reflection x-ray fluorescence in the silicon semiconductor manufacturing environment, G. Buhrer, Spectrochim. Acta, Part B, VOL. 54B, 1999, P
h e l p s y o u C O N T R O L
contamination analysis for compound semiconductors ANALYTICAL SERVICES B u r i e d d e f e c t s, E v a n s A n a l y t i c a l g r o u p h e l p s y o u C O N T R O L C O N T A M I N A T I O N Contamination
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE
Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
Coating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
OLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
Wipe Analysis to Determine Metal Contamination on Critical Surfaces
By Albert Dato, Ph.D., Warren York, Jennifer Jew, Laarni Huerta, Brice Norton, and Michael Coste On-wafer metallic contamination is detrimental to the fabrication and performance of semiconductor devices.
Elemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
Contamination. Cleanroom. Cleanroom for micro and nano fabrication. Particle Contamination and Yield in Semiconductors.
Fe Particles Metallic contaminants Organic contaminants Surface roughness Au Particles SiO 2 or other thin films Contamination Na Cu Photoresist Interconnect Metal N, P Damages: Oxide breakdown, metal
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
CVD SILICON CARBIDE. CVD SILICON CARBIDE s attributes include:
CVD SILICON CARBIDE CVD SILICON CARBIDE is the ideal performance material for design engineers. It outperforms conventional forms of silicon carbide, as well as other ceramics, quartz, and metals in chemical
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser
Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser V.Nassisi #, G.Caretto #, A. Lorusso #, D.Manno %, L.Famà %, G.Buccolieri %, A.Buccolieri %, U.Mastromatteo* # Laboratory of Applied
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING Brian L. Riise and Michael B. Biddle MBA Polymers, Inc., Richmond, CA, USA Michael M. Fisher American Plastics Council, Arlington, VA, USA X-Ray Fluorescence
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers
Key Questions & Exercises Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers 1. The atomic weight of carbon is 12.0107 u, so a mole of carbon has a mass of 12.0107 g. Why doesn t a mole of
Analyses on copper samples from Micans
PDF rendering: DokumentID 1473479, Version 1., Status Godkänt, Sekretessklass Öppen Analyses on copper samples from Micans P. Berastegui, M. Hahlin, M. Ottosson, M. Korvela, Y. Andersson, R. Berger and
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
Conductivity of silicon can be changed several orders of magnitude by introducing impurity atoms in silicon crystal lattice.
CMOS Processing Technology Silicon: a semiconductor with resistance between that of conductor and an insulator. Conductivity of silicon can be changed several orders of magnitude by introducing impurity
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with
Universal Data Acquisition (UDA)
Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte
EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray
Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray Hannu Teisala a, Mikko Tuominen a, Mikko Aromaa b, Jyrki M. Mäkelä b, Milena Stepien c, Jarkko J. Saarinen c, Martti Toivakka c
MILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE
MILESTONE CLEAN CHEMISTRY LINE A Breakthrough in Reduction and Control of the Analytical Blank DUOPUR TRACECLEAN ULTRA-TRACE INSERTS MILESTONE MILESTONE CLEAN CHEMISTRY LINE An innovative and complete
Secondary Ion Mass Spectrometry
Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown,
Lectures about XRF (X-Ray Fluorescence)
1 / 38 Lectures about XRF (X-Ray Fluorescence) Advanced Physics Laboratory Laurea Magistrale in Fisica year 2013 - Camerino 2 / 38 X-ray Fluorescence XRF is an acronym for X-Ray Fluorescence. The XRF technique
Vacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
Damage-free, All-dry Via Etch Resist and Residue Removal Processes
Damage-free, All-dry Via Etch Resist and Residue Removal Processes Nirmal Chaudhary Siemens Components East Fishkill, 1580 Route 52, Bldg. 630-1, Hopewell Junction, NY 12533 Tel: (914)892-9053, Fax: (914)892-9068
3. What would you predict for the intensity and binding energy for the 3p orbital for that of sulfur?
PSI AP Chemistry Periodic Trends MC Review Name Periodic Law and the Quantum Model Use the PES spectrum of Phosphorus below to answer questions 1-3. 1. Which peak corresponds to the 1s orbital? (A) 1.06
X Ray Flourescence (XRF)
X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose
Dry Etching and Reactive Ion Etching (RIE)
Dry Etching and Reactive Ion Etching (RIE) MEMS 5611 Feb 19 th 2013 Shengkui Gao Contents refer slides from UC Berkeley, Georgia Tech., KU, etc. (see reference) 1 Contents Etching and its terminologies
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist
Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist University of Puerto Rico Mayagüez Campus Department of Geology X-ray / Microanalysis Laboratory F - 304
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS
COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS Bérenger Caby (1), Fabio Brigidi (2), Dieter Ingerle (3), Blanka Detlefs
Photolithography. Class: Figure 12.1. Various ways in which dust particles can interfere with photomask patterns.
Photolithography Figure 12.1. Various ways in which dust particles can interfere with photomask patterns. 19/11/2003 Ettore Vittone- Fisica dei Semiconduttori - Lectio XIII 16 Figure 12.2. Particle-size
RoHS Test Methods. Presented by: Bruce Peterson
RoHS Test Methods Presented by: Bruce Peterson Motivation for Presentation The EU RoHS directive takes effect July 1, 2006 Customer confusion about - Scope of directive How much testing needs to be performed
Quality Control on Aerospace Components Using Handheld X-ray Fluorescence (XRF)
International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada Quality Control on Aerospace Components Using Handheld X-ray
Improved Contact Formation for Large Area Solar Cells Using the Alternative Seed Layer (ASL) Process
Improved Contact Formation for Large Area Solar Cells Using the Alternative Seed Layer (ASL) Process Lynne Michaelson, Krystal Munoz, Jonathan C. Wang, Y.A. Xi*, Tom Tyson, Anthony Gallegos Technic Inc.,
The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I
5 The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I Mg Kα excited photoemission. Abstract The oxidation of Cu(100)-Fe surfaces was studied using XPS. Surfaces
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield INTRODUCTION Stoichiometry calculations are about calculating the amounts of substances that react and form in a chemical reaction. The word stoichiometry
Sample Analysis Design Isotope Dilution
Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with
III. Wet and Dry Etching
III. Wet and Dry Etching Method Environment and Equipment Advantage Disadvantage Directionality Wet Chemical Solutions Atmosphere, Bath 1) Low cost, easy to implement 2) High etching rate 3) Good selectivity
Solar Photovoltaic (PV) Cells
Solar Photovoltaic (PV) Cells A supplement topic to: Mi ti l S Micro-optical Sensors - A MEMS for electric power generation Science of Silicon PV Cells Scientific base for solar PV electric power generation
Find a pair of elements in the periodic table with atomic numbers less than 20 that are an exception to the original periodic law.
Example Exercise 6.1 Periodic Law Find the two elements in the fifth row of the periodic table that violate the original periodic law proposed by Mendeleev. Mendeleev proposed that elements be arranged
MCAL Spectrophotometry. Spectrophotometry
MCAL Spectrophotometry Instruments include: Cary 50 UV-vis Spectrophotometer Eclipse Spectrofluorometer HPLC Diode Array and Fluorescence ICP-OES with CCD detection Spectrophotometry The instruments all
Instrumentation. (Figure 2)
X-Ray Fluorescence Lab Report Nydia Esparza Victoria Rangel Physics of XRF XRF is a non destructive analytical technique that is used for elemental and chemical analysis. X-Ray Fluorescence Spectroscopy
for Low power Energy Harvesting Sun to fiber' Solar Devices
Nanostructured Energy Conversion for Low power Energy Harvesting Devices and Beyond for High power Sun to fiber' Solar Devices Michael Oye and Nobuhiko Nobby Kobayashi Advanced Studies Laboratories and
3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
Paper Chromatography: Separation and Identification of Five Metal Cations
Paper Chromatography: Separation and Identification of Five Metal Cations Objectives Known and unknown solutions of the metal ions Ag +, Fe 3+, Co 2+, Cu 2+ and Hg 2+ will be analyzed using paper chromatography.
The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün
N.B. A shorter version of this article was published in the ICOM-CC Ethnographic Conservation Newsletter, Number 30, January 2009, pp. 5-8. The Limitations of Hand-held XRF Analyzers as a Quantitative
Lecture 12. Physical Vapor Deposition: Evaporation and Sputtering Reading: Chapter 12. ECE 6450 - Dr. Alan Doolittle
Lecture 12 Physical Vapor Deposition: Evaporation and Sputtering Reading: Chapter 12 Evaporation and Sputtering (Metalization) Evaporation For all devices, there is a need to go from semiconductor to metal.
Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:
Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T
Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror
Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum
Quantitation of Drugs in Dried Bloodstains. Thomas Meyer Anna Gomenyuk Nadiah Lester
Quantitation of Drugs in Dried Bloodstains Thomas Meyer Anna Gomenyuk Nadiah Lester Objective To find the quantitative value of drugs in dried bloodstains Currently Cloth With Dried Bloodstain Containing
MODERN ATOMIC THEORY AND THE PERIODIC TABLE
CHAPTER 10 MODERN ATOMIC THEORY AND THE PERIODIC TABLE SOLUTIONS TO REVIEW QUESTIONS 1. Wavelength is defined as the distance between consecutive peaks in a wave. It is generally symbolized by the Greek
Quartz Glass. Tubes and Rods
Quartz Glass Tubes and Rods PH 300, GE 214, QI PN GVB Solutions in Glass Schlackstrasse 3 52080 Aachen Germany +49-241/9108588 +49-241/9108589 E- [email protected] www.g-v-b.de Table of contents General Information
Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
2. Write the chemical formula(s) of the product(s) and balance the following spontaneous reactions.
1. Using the Activity Series on the Useful Information pages of the exam write the chemical formula(s) of the product(s) and balance the following reactions. Identify all products phases as either (g)as,
Physical & Chemical Properties. Properties
Physical & Chemical Properties Properties Carbon black can be broadly defined as very fine particulate aggregates of carbon possessing an amorphous quasi-graphitic molecular structure. The most significant
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control
EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system
EDS system Most common X-Ray measurement system in the SEM lab. Major elements (10 wt% or greater) identified in ~10 secs. Minor elements identifiable in ~100 secs. Rapid qualitative and accurate quantitative
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
DI-water technology for photoresist removal. shifts from 200- to 300-mm wafers, the use of ozonated deionized water (DIO 3
Using an ozonated- DI-water technology for photoresist removal Jae-Inh Song, Richard Novak, Ismail Kashkoush, and Pieter Boelen, Akrion As the semiconductor industry shifts from 200- to 300-mm wafers,
QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
Lecture 11. Etching Techniques Reading: Chapter 11. ECE 6450 - Dr. Alan Doolittle
Lecture 11 Etching Techniques Reading: Chapter 11 Etching Techniques Characterized by: 1.) Etch rate (A/minute) 2.) Selectivity: S=etch rate material 1 / etch rate material 2 is said to have a selectivity
Electrochemistry Voltaic Cells
Electrochemistry Voltaic Cells Many chemical reactions can be classified as oxidation-reduction or redox reactions. In these reactions one species loses electrons or is oxidized while another species gains
Chapter 8 - Chemical Equations and Reactions
Chapter 8 - Chemical Equations and Reactions 8-1 Describing Chemical Reactions I. Introduction A. Reactants 1. Original substances entering into a chemical rxn B. Products 1. The resulting substances from
Flow Injection Analysis
Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since
Redox and Electrochemistry
Name: Thursday, May 08, 2008 Redox and Electrochemistry 1. A diagram of a chemical cell and an equation are shown below. When the switch is closed, electrons will flow from 1. the Pb(s) to the Cu(s) 2+
LASER ENGRAVING REFLECTIVE METALS TO CREATE SCANNER READABLE BARCODES Paper P516
LASER ENGRAVING REFLECTIVE METALS TO CREATE SCANNER READABLE BARCODES Paper P516 Paul M Harrison, Jozef Wendland, Matthew Henry Powerlase Ltd, Imperial House, Link 10, Napier Way, Crawley, West Sussex,
Electron Microscopy 3. SEM. Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts
Electron Microscopy 3. SEM Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts 3-1 SEM is easy! Just focus and shoot "Photo"!!! Please comment this picture... Any
Electrochemistry - ANSWERS
Electrochemistry - ANSWERS 1. Using a table of standard electrode potentials, predict if the following reactions will occur spontaneously as written. a) Al 3+ + Ni Ni 2+ + Al Al 3+ + 3e - Al E = -1.68
Spectral Measurement Solutions for Industry and Research
Spectral Measurement Solutions for Industry and Research Hamamatsu Photonics offers a comprehensive range of products for spectroscopic applications, covering the, Visible and Infrared regions for Industrial,
Scanning Electron Microscopy Services for Pharmaceutical Manufacturers
Scanning Electron Microscopy Services for Pharmaceutical Manufacturers Author: Gary Brake, Marketing Manager Date: August 1, 2013 Analytical Testing Laboratory www.atl.semtechsolutions.com Scanning Electron
Spectral distribution from end window X-ray tubes
Copyright (C) JCPDS-International Centre for Diffraction Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg 2 1 FORTEX - E.N.S.A.I.S. 24, bld de la Victoire,
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
47374_04_p25-32.qxd 2/9/07 7:50 AM Page 25. 4 Atoms and Elements
47374_04_p25-32.qxd 2/9/07 7:50 AM Page 25 4 Atoms and Elements 4.1 a. Cu b. Si c. K d. N e. Fe f. Ba g. Pb h. Sr 4.2 a. O b. Li c. S d. Al e. H f. Ne g. Sn h. Au 4.3 a. carbon b. chlorine c. iodine d.
EXPERIMENT 4 The Periodic Table - Atoms and Elements
EXPERIMENT 4 The Periodic Table - Atoms and Elements INTRODUCTION Primary substances, called elements, build all the materials around you. There are more than 109 different elements known today. The elements
Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM
Original Paper Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM Satoshi Hashimoto 1, Tsuguo Sakurada 1, and Minoru Suzuki 2 1 JFE-Techno research corporation, 1-1 Minamiwatarida, Kawasaki,
Gold Refining and Coin Manufacturing at the Royal Canadian Mint
INTERNET Sub title will go here The Application of XRF to Gold Refining and Coin Manufacturing at the Royal Canadian Mint Michael W. Hinds, Ph.D. The Business of the Mint Circulation Coins (Canada & Other
Name period AP chemistry Unit 2 worksheet Practice problems
Name period AP chemistry Unit 2 worksheet Practice problems 1. What are the SI units for a. Wavelength of light b. frequency of light c. speed of light Meter hertz (s -1 ) m s -1 (m/s) 2. T/F (correct
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS
IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS INTRODUCTION Each REACH registrant is required to file its
CLASS TEST GRADE 11. PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change
CLASS TEST GRADE PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change MARKS: 45 TIME: hour INSTRUCTIONS AND INFORMATION. Answer ALL the questions. 2. You may use non-programmable calculators. 3. You may
Ion Beam Sputtering: Practical Applications to Electron Microscopy
Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a
How To Use The Nexion 300
APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of
Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)
Ionizing Radiation, Czech Republic, (Czech Metrology Institute) Calibration or Measurement RADIOACTIVITY 1.0E+00 1.0E+02 Bq cm -2 C-14 1.0E+01 1.0E+02 Bq cm -2 Co-60 1.0E+01 1.0E+02 Bq cm -2 Sr-90 1.0E+01
Unit 3 Study Guide: Electron Configuration & The Periodic Table
Name: Teacher s Name: Class: Block: Date: Unit 3 Study Guide: Electron Configuration & The Periodic Table 1. For each of the following elements, state whether the element is radioactive, synthetic or both.