VS-800 Vital signs monitor. Service Manual
|
|
|
- Tracey Burke
- 10 years ago
- Views:
Transcription
1 VS-800 Vital signs monitor Service Manual
2
3 Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaption and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. and are the registered trademarks or trademarks owned by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners. Contents of this manual are subject to changes without prior notice. I
4 FOR YOUR NOTES II
5 Preface Manual Purpose This manual provides detailed information about the assembling, dissembling, testing and troubleshooting of the equipment to support effective troubleshooting and repair. It is not intended to be a comprehensive, in-depth explanation of the product architecture or technical implementation. Observance of the manual is a prerequisite for proper equipment maintenance and prevents equipment damage and personnel injury. This manual is based on the maximum configuration; Therefore, some contents may not apply to your monitor. If you have any question, please contact our Customer Service Department. Intended Audience This manual is for biomedical engineers, authorized technicians or service representatives responsible for troubleshooting, repairing and maintaining the monitors. Revision History This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice Revision number: 3.1 Release time: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Password A password is required to access the service mode. The service password is
6 FOR YOUR NOTES 2
7 Contents 1 Safety Safety Information Danger Warnings Cautions Notes Equipment Symbols Theory of Operation Introduction System Structure Hardware Structure Main Board Power Board Key&displays Board Parameter Boards Recorder Bar Code Scanner Testing and Maintenance Introduction Test Report Recommended Frequency Visual test Power-On Test NIBP Tests SpO 2 Test Temp Test Nurse Call Performance Test Bar Code Scanner Test Electrical Safety Tests Enclosure Leakage Current Test Earth Leakage Current Test Patient Leakage Current Test Patient Auxiliary Leakage Current Test Recorder Check Program upgrade Troubleshooting Introduction Part Replacement
8 4.3 Monitor Status Check Technical Alarm Check Troubleshooting Guide Power On/Off Failures Display Failure LED Digital Display & Indication Lamp Failure Alarm Problems Button Failure Recorder Failures Interface Failures Power Supply Failures Network related problems Software Upgrade Problems Bar Code Scanner Failures NIBP Measurement Failures SpO2 Measurement Failure Temp Measurement Failures Repair and Disassembly Tools Preparation for Disassembly Disassembly Separating the Front and Rear Half of the Monitor Disassembling the Front Housing Assembly Removing the Main Rack Assembly Disassembling NIBP Module Removing the Main Board Removing the Power Board Removing the Battery Assembly Removing the Fan and Speaker Removing the Power Socket Removing the Temp Module Removing the SpO 2 Module Parts Main Unit Main Rack Assembly (For Lead-acid Batteries) Main Rack Assembly (For Lithium Battery) Front housing Assembly Rear Housing Assembly Lithium Battery Charger Assembly Lead-acid Battery Assembly MASIMO SpO 2 Board Assembly Temp Assembly
9 1 Safety 1.1 Safety Information DANGER Indicates an imminent hazard that, if not avoided, will result in death or serious injury. WARNING Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury. CAUTION Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage. NOTE Provides application tips or other useful information to ensure better maintenance operation Danger There are no dangers that refer to the product in general. Specific Danger statements may be given in the respective sections of this manual. 1-1
10 1.1.2 Warnings WARNING Disassembly and repair of this product should be conducted by Mindray authorized personnel only. To avoid explosion hazard, do not use the equipment in the presence of flammable anesthetics, vapors or liquids. Follow the applicable waste control regulations to dispose of the package material and keep it out of children s reach. The equipment must be connected to a properly installed power outlet with protective earth contacts only. If the installation does not provide a protective earth conductor, disconnect the equipment from the power line and operate it on battery power, if possible Cautions CAUTION Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For this reason, make sure that all external devices operated in the vicinity of the equipment comply with the relevant EMC requirements. Mobile phones, X-ray equipment or MRI devices are possible sources of interference as they may emit higher levels of electromagnetic radiation. Before connecting the equipment to the power line, check that the voltage and frequency ratings of the power line are the same as those indicated on the equipment s label or in this manual. Protect the equipment from damage caused by drop, impact, strong vibration or other mechanical force during servicing Notes NOTE Refer to Operation Manual for detailed operation and other information. 1-2
11 1.2 Equipment Symbols Attention (Consulting this manual). Up button Power On/Off Confirm button Alternating current (AC) Down button Battery indicator Pulse Rate (PR) Type CF applied part. Defibrillator-proof Neonate NIBP start/stop button Pediatric Patient Information Adult Main menu Equipotential grounding Set alarms Nurse Call connector Display Tabular Trends/Pleth Wave RS-232 connector NIBP interval Date of manufacture Record Serial number Alarm Silence Network connector 1-3
12 FOR YOUR NOTES 1-4
13 2 Theory of Operation 2.1 Introduction The monitor is intended for monitoring the patient s vital signs including non-invasive blood pressure (NIBP), pulse oxygen saturation (SpO 2 ) and pulse rate (PR) for single adult, paediatric and neonatal patient. It is also designed to monitor the temperature Temp) for single adult and paediatric patient. 2.2 System Structure The whole system consists of mechanical parts, hardware and software, as follows: System Mechanical Hardware Software Main unit Recorder Battery chamber Temp module charmber Barcode scanner support Main board Power board Key&display Parameter boards Recorder board Barcode scanner System software Parameter software Upgrade software Print software Gateway software 2-1
14 The system s PCBAs are connected as shown below: 2-2
15 2.3 Hardware Structure The monitor s hardware structure is shown in the figure below. SpO2 module NIBP module Isolation power board * Communication/p ower supply TEMP module Communication/p ower supply LED & KEY Ethernet Key&display board LCD Main board RS232 Power supply NurseCall Speaker Recorder Fan Power board Battery AC VS-800 * Mindray SpO2 module does not require the isolation power board. The core of the system is the main board which provides power supply for all parameter modules. The parameter modules directly communicate with the main board, and the measurements and status of all modules are processed by the main board and then displayed on the key&display board. The main board also controls communication with the speaker, the recorder and all the external connectors. 2-3
16 2.3.1 Main Board The main board provides resources and supports for the overall system. It controls LCD, LED, keyboard, speaker and recorder. It also enables to communicate with parameter modules and connectors Principle Diagram Watchdog Ethernet PHY RTL8201 RTC SPI E 2 PROM Nurse Call Speaker Audio process circuit CPU Serial port 0 RS232 IC RS232 Serial port 1: NIBP Serial port 2: SpO2 SDRAM Flash FPGA Serial port 3: Temp 1.5V Serial port 4: Recorder Linear power 3.3V 5.0V 12V LCD LED KEY Function Description The main board communicates with all the parameter modules and the recorder through FPGA extended serial port. The main board supplies information through FPGA for the key&display board to display; it detects the keys and implements user interface. The main board controls the alarm indicator via FPGA and the signal wire is adapted from key&display board. The main board controls the speaker to give audible alarm, key tones and Pitch Tone. The main board provides nurse call connector, network connector and R232 connector. The real-time clock is implemented by the RTC chip. The RT clock is powered according to the priority of AC mains, battery, and button cell on the main board. This ensures the normal working status of the clock in the event that AC mains and batteries are not available. SDRAM is responsible for storing data temporarily and running programs. Flash serves as the system memory and trend data memory whereas EEPROM services as the device configuration memory. 2-4
17 2.3.2 Power Board The power board converts the input power (AC mains or battery) to different working voltages for other boards; it also has the function of charging battery Principle Diagram AC input EMI filter Rectifer & filter Flyback converter Rectifier & filter 16.8V DC/DC converter OVP&OCP 12V output DC/DC converter 5V output PWM controller Current detection Charging circuit DC/DC converter Battery 3.3V output Coupler feedback & isolation circuit OVP&OPP Switch circuit OVP&OCP Functions After the AC input reaches the power board, it flows through the EMI filter circuit and rectifier and filter circuit. Then it is converted to a DC voltage, which is further converted to the 16.8V DC voltage by the Flyback converter. The 16.8V DC voltage is the main input of DC/AC converter and charging circuit; it is used to charge the lithium battery or lead-acid battery, and also converted to 12V, 5V and 3.3V DC voltage by corresponding circuit conversion. In case the AC input is unavailable, the battery can supply power for the DC/DC converters to get 12V, 5V and 3.3V DC voltages, which are under the control of device poweron/poweroff signal. The 16.8V DC output is protected against over-voltage and over-power, the 12V and 3.3V DC outputs are protected against over-voltage, short-circuit and over current, the 5V DC output is protected against over-voltage and short-circuit. Two power boards are developed for the lead-acid battery and lithium battery. Each board demonstrates the battery types it supports via the main board connector, facilitating the main unit to process information such as battery capacities. The power board charges the battery merely under AC input. The monitor runs on AC mains if both AC mains and battery are available. 2-5
18 2.3.3 Key&displays Board The key&displays board provides the user s interface. The LCD module, 7-segment digital display, LED indication lamp and keys are integrated on the board Principle Diagram LCD signal FSTN LCD module LED CPLD LED&KEY 3.3V disable Start-up circuit ADV 3.3V Key matrix Switch detection VSB Functions The LCD module has the function of adjusting the contrast and brightness in 10 levels, displaying menus, trend data and Pleth waveform. The 7-segment digit displays are used to display systolic pressure, diastolic pressure, mean pressure/cuff pressure, SpO 2, PR, Temp and so forth. The AC indicator is driven by the ADV output from the power board whilst the working status indicator (built in the Power On/Off key) is driven by 3.3V voltage. The battery indicator is jointly controlled by the flash control signal, ADV signal and /BC signal. The keypad consists of the power ON/OFF key and the other 11 functional keys. 2-6
19 2.3.4 Parameter Boards SpO2 Module The principle diagram of SpO2 module is shown below: WATCHDOG DRIVE DAC LED DRIVE CIRCUITRY /RST WDI CPU DRAM FLASH Isolated Serial In/Output Port 4KV Isolated Power Supply +3.3V +5V -5V +12V HOST SENSOR A/D CONVERTER GAIN CONTROL CIRCUITRY OFFSET DAC PROGRAMMABLE GAIN CIRCUITRY OFFSET AMPLIFIER The SpO2 sensor collects the red and infrared light signal transmitting through the finger or toe which is pulsing and process the collected signal to get the measured result. The drive circuit of the LED and the gain of the amplifying circuit should be controlled according to the perfusions and transmittances of the tested objects. 2-7
20 NIBP Module The principle diagram of NIBP module is shown below: CUFF PRESSURE SENSOR FOR PROTECT PRESSURE SIGNAL AMPLIFIED CIRCUITS FOR PROTECT OVER PRESSURE PROTECT CIRCUIT WATCHDO G CPU ASYNCHRONOUS SERIAL COMMUNICATION PRESSURE SENSOR PRESSURE SIGNAL AMPLIFIED CIRCUITS PRESSURE SIGNAL A/D CONVERTER Windpipe MOTOR CONTROL FEEDBACK SIGNAL Valve and Air Pump PUMP AND VALVES CONTROLED CIRCUITS FLASH DRAM The NIBP is measured based on the pulse vibration principle. Inflate the cuff which is on the forearm till the cuff pressure blocks the arterial blood, and then deflate the cuff according to a specified algorithm. While the cuff pressure is decreasing, the arterial blood has pulses, which are sensed by the pressure transducer in the cuff. Consequently, the pressure transducer connected with the windpipe of the cuff, generates a pulsation signal, which is then processed by the NIBP module to get the NIBP value. NIBP values are based on the oscillometric method of noninvasive blood pressure measurement and correspond to comparisons with auscultative values (applied to adults and pediatrics elder than 3 years) and intra-arterial values (applied to neonates and pediatrics younger than 3 years) within ANSI/AAMI SP10 Standards for accuracy (a mean difference of ± 5 mmhg, and a standard deviation of < 8 mmhg). 2-8
21 Temp Module The principle diagram of Temp module is shown below: Normally, the sensor used for measuring temperature is a thermistor. The resistance of a given thermistor is nonlinearly relative to the temperature. Thus, the resistance of a thermistor can be conversed into temperature. By applying given field current to the thermistor, its resistance can be easily obtained by measuring the voltage on the thermistor Recorder The recorder receives data from the main board and then sends them to the thermal printhead for printing. The recorder has a hardkey(starting/stopping recordings) and a green LED on its front Bar Code Scanner The bar code scanner enables to read the one-dimensional and two-dimensional bar codes, simplifying patient admit. The bar code scanner communicates with the monitor and gets power supply to the power source via the serial ports. The scanner s serial port is defined as follows: PIN Definition 2 Barcode_RX 3 Barcode_TX 5 GND 9 VCC(5VDC) 2-9
22 The monitor s serial port is defined as follows: PIN Definition 2 VS-800_TX 3 VS-800_RX 5 GND 9 VCC(5VDC) Basic settings of the bar code scanner are listed in the table below: Bar code scanner Factory default Host Parameters Baud Rate Data Bits 8 8 Stop Bits 1 1 Calibration bit 0 0 Handshaking None None User Parameters Beeper Tone Medium Medium Beeper Volume Medium High Trigger Mode Level Auto Aim Parameter Scanning Disable Enable Data Format Prefix Value 7013 <CR><LF> 7013 <CR><LF> Suffix 1 Value 7013 <CR><LF> 7013 <CR><LF> Suffix 2 Value Scan Data Transmission Format <PREFIX><DATA><SUFFIX 1><SUFFIX 2> Data only You shall program the bar code scanner to perform various functions. To change parameter settings, scan the following bar codes in sequence. 1. Set Default Parameter 2-10
23 2. Beeper Volume 3. Trigger Mode 4. Scan Data Transmission Format 5. Disable Parameter Scanning 2-11
24 FOR YOUR NOTES 2-12
25 3 Testing and Maintenance 3.1 Introduction To ensure the monitor always functions normally, qualified service personnel should perform regular inspection, maintenance and test. This chapter provides a checklist of the testing procedures for the monitor with recommended test equipment and frequency. The service personnel should perform the testing and maintenance procedures as required and use appropriate test equipment. The testing procedures provided in this chapter are intended to verify that the monitor meets the performance specifications. If the monitor or a module fails to perform as specified in any test, repairs or replacement must be done to correct the problem. If the problem persists, contact our Customer Service Department. CAUTION All tests should be performed by qualified service personnel only. Care should be taken to change the settings in the [MAINTAIN] menu to avoid loss of data. Service personnel should acquaint themselves with the test tools and make sure that test tools and cables are applicable Test Report After completing the tests, service personnel are required to record test results in this table and report them to Mindray Customer Service Department. Test Equipment Name Model/PN Expiry Date Test Record No. Test Item Test Site Test Results 1 2 Conclusion Pass/Fail: Tested by: Date: 3-1
26 3.1.2 Recommended Frequency Check/Maintenance Item Visual test Power on test Frequency When first installed or after reinstalled. 1. When first installed or after reinstalled. 2. Following any maintenance or the replacement of any main unit parts. NIBP tests IBP tests Accuracy test Leakage test Calibration Performance test Pressure calibration 1. If the user suspects that the measurement is incorrect. 2. Following any repairs or replacement of respective module. 3. At least once every two years. SpO 2 test / Temp test / Analog output test / Bar code scanner test / If the user suspects that analog output or bar code scanner is abnormal. Electrical safety tests Enclosure leakage current test Earth leakage current test Patient leakage current test Patient auxiliary current test 1. Following any repair or replacement of the power module. 2. At least once every two years. Recorder check / Following any repair or replacement of the recorder. 3.2 Visual test Inspect the equipment for obvious signs of damage. The test is passed if the equipment has no obvious signs of damage. Follow these guidelines when inspecting the equipment: Carefully inspect the case, the display screen and the buttons for physical damage. Inspect all external connections for loose connectors, bent pins or frayed cables. Inspect all connectors on the equipment for loose connectors or bent pins. Make sure that the labels on the equipment are clearly legible. 3-2
27 3.3 Power-On Test This test is to verify that the monitor can power up correctly. The test is passed if the monitor starts up by following this procedure: 1. Insert the lead-acid batteries or lithium battery in the battery compartment and connect the monitor to the AC mains; the AC mains LED and battery LED light. 2. Press the Power On/Off button to turn on the monitor. The operating status LED which is built in the Power On/Off button lights up. 3. The system gives a beep indicating the self test related to alarm sounds is passed and all the LEDs on the front panel light up. At the same time, the technical alarm lamp turns yellow, and then turns off, indicating that the self test related to alarm lamps is passed. 4. The start-up screen disappears; the monitor enters the main screen and start-up is finished NIBP Tests NIBP Accuracy test Tools required: T-shape connector Appropriating tubing Balloon pump Metal Vessel with volume 500 ± 25 ml Reference manometer with accuracy 1 mmhg Follow this procedure to perform the test: 1. Connect the equipment as shown above. Monitor Manometer NIBP cuff connector Tubing Balloon pump Metal vessel 3-3
28 2. Before inflating the metal vessel, the reading of the manometer should be 0. If not, disconnect the airway and reconnect it until the readings is Press the MENU key; select [MAINTAIN >>] [NIBP CALIBRATE] to start NIBP calibration. Compare the manometer values with the displayed values. The difference should be no greater than ±3mmHg (±0.4kPa). 4. Raise the pressure in the metal vessel to 50 mmhg with the balloon pump. Repeat step3. 5. Raise the pressure in the metal vessel to 200 mmhg with the balloon pump. Repeat step 3. Contact our service personnel if the difference between any manometer value and the displayed value is greater than ±3 mmhg. NOTE You can use an NIBP simulator to replace the balloon pump and the reference manometer to perform the test. NIBP Leakage Test Tools required: NIBP cuff for adult patient Approprating tubing Cylinder Follow this procedure to perform the test: 1. Set [PATIENT TYPE] to [ADU]. 2. Connect the NIBP cuff with the NIBP connector on the monitor. 3. Apply the cuff to the cylinder as shown below. Monitor Cylinder NIBP cuff connector Air tubing Cuff 4. Press the MENU key; select [MAINTAIN >>] [NIBP LEAK TEST]. 5. The cuff automatically deflates in 20s, which means NIBP leakage test is completed. 6. If the system does not leak, no message will be displayed in the technical alarm area/prompt area. If it leaks, the message [PNEUMATIC LEAK] will be displayed. In this case, check for loose connection and perform the test again. Please contact our service personnel if the NIBP test fails. 3-4
29 3.3.2 SpO 2 Test Tool Required: SpO2 simulator. For the monitor equipped with Mindray SpO2 module, BIO-TEK Index-2 SpO2 simulator is recommended. For the monitor equipped with Nellcor SpO2 module, SRC-MAX SpO2 simulator is recommended. For the monitor equipped with Masimo SpO2 module, BIO-TEK Index-2 SpO2 simulator is recommended SpO 2 Test under Normal Condition 1. Connect the SpO2 simulator to the SpO2 sensor. 2. Select the model and the manufacturer of the SpO2 module under test, and then configure the SpO2 simulator as follows: SpO2 96%; PR 80 bpm. 3. The displayed SpO2 and PR values should be within the ranges listed below SpO 2 Test in Motion Mode 1. Connect the SpO2 simulator to the SpO2 sensor. 2. Select the model and the manufacturer of the SpO2 module under test; take measurement in the motion mode preset by the SpO 2 simulator. 3. The displayed SpO2 and PR values should be within the ranges listed below. Manufacturer SpO 2 sensor SpO 2 PR (bpm) Mindray 512B, 512 D, 512E, 512G, 512F, 512H, 518B, ES /envitec, MAX-A, MAX-P, MAX-I, DS-100A, OXI-P/I, 520A, 520P, 520I, Envite, Envite, Envite 520N, MAX-N, OXI-A/N, Envite 96%±2% 96%±3% 80±3 3-5
30 Manufacturer SpO 2 sensor SpO 2 PR (bpm) 96%±3% LNCS-NeoPt-L, LNCS Neo-L (without motion) 96%±3% 80±3 bpm Masimo LNCS Inf-L, LNCS-Pdt, LNCS-Adtx,LNCS DC-I,LNCS DC-I Pt (with motion) 96%±2% (without motion) 96%±3% (with motion) (without motion) 80±5 bpm (with motion) MAX-A, MAX-P, MAX-I, 96%±2% Nellcor DS-100A, OXI-A/N (Adult), OXI-P/I, MAX-N 96%±3% 80±3 bpm OXI-A/N (Neonate) 96%±4% NOTE The SpO 2 simulator can only be used to verify that the pulse oximeter operates properly. It cannot be used to verify the accuracy of the pulse oximeter or the SpO 2 sensor. To verify the accuracy, clinical tests are required. Please contact our service personnel if the SpO 2 test fails Temp Test Tool required: thermostatic oil tank, HART 7102 recommended. 1. Set the temperature of the oil tank to 37 and conduct the test after the temperature stabilizes. 2. Set the [TEMP TYPE] to [MONITOR]. Remove the Temp probe from the probe sheath and then put it into the oil tank. 3. Wait till the Temp value displayed on the monitor stabilizes. Verify that the displayed value is 37±0.2. Please contact our service personnel if the Temp test fails. 3-6
31 3.4 Nurse Call Performance Test Tools required: Multimeter 1. Connect the nurse call cable to the analog output connector. 2. Trigger a psychological alarm or a technical alarm. 3. Select [NURSE CALL>>] in the [MAINTAIN] menu. In the [NURSE CALL] menu, select [ALM LEV] and [ALM TYPE]; set [CONTACT TYPE] to [NORMAL OPEN]. 4. In the [NURSE CALL >>] menu, set [SIGNAL TYPE] to [PULSE]. Trigger an alarm and measure the contact output with the multimeter. The output should be square waves with an interval of 1s, 5. In the [NURSE CALL >>] menu, set [SIGNAL TYPE] to [CONTINUUM]. Trigger an alarm and measure the contact output with the multimeter. The output should be continuous high level. Please contact our service personnel if the nurse call test fails. 3.5 Bar Code Scanner Test Tools required: None 1. Aim the bar code scanner at the target bar code. Adjust the field of view to capture the bar code. 2. Hold the trigger until the bar code scanner gives a beep, indicating the bar code is successfully decoded. Meanwhile, the indication lamp turns to green and scanned characters are displayed on the monitor. Please contact our service personnel if the bar code scanner fails to work normally. 3-7
32 3.6 Electrical Safety Tests WARNING Electrical safety tests are a proven means of verifying the electrical safety of the monitor. They are intended for determining potential electrical hazards. Failure to find out these hazards timely may cause personnel injury. Commercialy available test equipment such as safety analyzer can be used for electrical safety tests. Be sure that the test equipment can be safely and reliably used with the monitor before use. The service personnel should acquaint themselves with the use of the test eqipment. Electrical safety tests should meet the requiremnets of the latest editions of EN and UL These electrical safety tests do not supercede local requiremnets. All devices using the AC mains and connected to medical equipment within patient environments must meet the requiremnets of the IEC medical electrical systems standard and should be put under electrical safety tests at the frequency recommended for the monitor. Electrical safety tests are intended to timely check the potential electrical hazards to the patient, operator or service personnel arising from the equipment. Electrical safety test should be performed under normal ambient conditions of temperature, humidity and atmospheric pressure. The electrical safety test plan described hereunder takes 601 safety analyzer as an example. Different safety analyzers may be used in different areas. Be sure to choose applicable test plan. 3-8
33 Connection of the equipments is shown below: Tools required: Safety analyzer Isolation transformer Enclosure Leakage Current Test 1. Connect the 601 safety analyzer to an AC power supply (264V, 60 Hz). 2. Connect SUM terminal of the applied part connection apparatus to RA input terminal of 601 safety analyzer, another terminal to the applied part of EUT. 3. Connect the EUT to the 601 analyzer s auxiliary output connector by using a power cord. 4. Attach on end of the red lead to the red input terminal of the analyzer, and the other end to tinsel over the enclosure of the EUT. 5. Power on the 601 safety analyzer and then press the 5-Enclosure leakage button on the analyzer s panel to enter the enclosure leakage test screen. 6. Under normal condition, the enclosure leakage current should be no greater than 100µA. Under single fault condition, it should be no greater than 300µA. 3-9
34 3.6.2 Earth Leakage Current Test 1. Connect the 601 safety analyzer to an AC power supply (264V, 60 Hz). 2. Connect the SUM terminal of the applied part connection apparatus to RA input terminal of 601 safety analyzer, another terminal to the applied part of EUT. 3. Connect the EUT to the 601 analyzer s auxiliary output connector by using a power cord. 4. Power on the 601 safety analyzer and then press the 4-Earth leakage button on the analyzer s panel to enter the earth leakage test screen. 5. Under normal condition, the earth leakage current should be no greater than 300µA. Under single fault condition, it should be no greater than 1000 µa Patient Leakage Current Test 1. Connect the 60 safety analyzer to an AC source (264V, 60Hz). 2. Connect the SUM terminal of the applied part connection apparatus to RA input terminal of 601 safety analyzer, another terminal to the applied part of EUT. 3. Connect the EUT to the 601 analyzer s auxiliary output connector by using a power cord. 4. Power on the 601 safety analyzer and then press the 6-Patient leakage on the 601 analyzer s panel. 5. Repeatedly press the APPLIED PART button to measure AC and DC leakage alternatively. DC leakage reading is followed by DC. 6. Under normal status, the patient leakage current should be no greater than 10 µa. Under single fault condition, it should be no greater than 50µA Patient Auxiliary Leakage Current Test 1. Connect the 601 safety analyzer to an AC source (264V, 60Hz). 2. Connect the equipment under test (EUT) to the analyzer s auxiliary output connector by using a power cord. 3. Connect the sensors of the applied part to the applied part connection apparatus, whose RA-P terminal is connected to 601 safety analyzer s RA terminal and SUM terminal to 601 safety analyzer s LA terminal. RA terminal is switched on. 3-10
35 4. Power on the 601 safety analyzer and then press the 8-Patient Auxiliary Current Test button on the analyzer s panel to enter the patient auxiliary current test screen. 5. Repeatedly press the APPLIED PART button to measure AC and DC leakage alternatively. DC leakage reading is followed by DC. 6. Under normal status, the patient auxiliary leakage current should be no greater than 10 µa. Under single fault condition, it should be no greater than 50µA. Please contact our service personnel if the electrical safety test fails. 3.7 Recorder Check 1. Print SpO 2 Pleth waveforms. The recorder should print correctly and printout should be clear. 2. Set the recorder to some problems such as out of paper, paper jam and etc, the monitor should give corresponding prompt messages. After the problem is removed, the recorder should be able to work correctly. 3. Set the recorder to print trend data. Check that the recorder can give printouts accordingly. Please contact our service personnel if the recorder test fails. 3.8 Program upgrade You can upgrade the monitor by downloading the upgrade software through network. At the same tine, you can view the upgrade status log. The upgrade software can directly run on a portable PC or desktop PC. Through the network or by connecting the monitor to a PC via a network cable, you can upgrade the following programs: Bootstrap program System program Bar code recognition configuration Multiligual library General configurations (including passwords, company logo) System functional configuration FPGA program Parameter module programs: SpO2 module (Mindray), NIBP module and Temp module. Please refer to help and instructions for program upgrade for details. 3-11
36 CAUTION Disconnect the monitor from the patient and make sure the important date are saved before upgrade. Do not shut down or power off the equipment when upgrading the bootstrap program. Otherwsie, it may cause the equipemnt to break down. Programs upgrade should be performed by qualified service personnel only. NOTE After upgrading the boot program, re-upgrade the system program and other programs to ensure compatibility. Make sure the version of the upgrade package is your desired one. If you want to obtain the latest upgrade package, contact Mindray Customer Service Department. 3-12
37 4 Troubleshooting 4.1 Introduction In this chapter, monitor problems are listed along with possible causes and recommended corrective actions. Refer to the tables to check the monitor, identify and eliminate the troubles. The troubles we list here are frequently arisen difficulties and the actions were recommended can correct most problems, but not all of them. For more information on troubleshooting, please contact our Customer Service Department. 4.2 Part Replacement Printed circuit boards (PCBs), major parts and components in the monitor are replaceable. Once you isolate a PCB you suspect defective, follow the instruction in 5 Repair and Disassembly to replace the PCB with a known good one and check that the trouble disappear or the monitor passes all performance tests. If the trouble remains, exchange the replacement PCB with the original suspicious PCB and continue troubleshooting as directed in this chapter. Defective PCB can be sent to us for repair. To obtain information on replacement parts or order them, please refer to 6 Parts. 4.3 Monitor Status Check Some troubleshooting tasks may require you to identify the hardware version and status of your monitor. 1. To view system software version and FPGA version, select [MENU] [MAINTAIN>>] [VERSION]. 2. To view the monitor s total runtime, select [MENU] [MAINTAIN>>] [VERSION] [TOTAL RUN TIME>>]. 3. To view modules software version, select [MENU] [MAINTAIN>>] [VERSION] [MODULE VERSION>>]. 4-1
38 4.4 Technical Alarm Check Check whether technical alarm message is displayed on the monitor before troubleshooting. If an alarm message is presented, eliminate the technical alarm first and troubleshoot in accordance with the following instructions. For further information on technical alarm message, possible cause and corrective action, please refer to the monitor s Operation Manual. 4.5 Troubleshooting Guide Power On/Off Failures Symptoms Possible Cause Corrective Actions The monitor fails to start. AC LED or battery LED does not light AC mains not connected or battery too low. Power supply protection Cables defective or poorly connected Power switch & LED board defective Power module defective Main board Defective Check whether AC mains is properly connected or battery capacity is sufficient Refer to 4.5.8Power Supply Failures. 1. Check whether the cables from power switch & LED board to button board, button board to main board, and power module to main board are correctly connected. 2. Check whether cables and connectors are damaged. Replace the power switch & LED board Replace the power module Replace the main board 4-2
39 4.5.2 Display Failure Symptoms Possible Cause Corrective Actions The display is black or blank, but the monitor still works correctly. Images overlapped or distorted Cables defective or poorly connected Backlight defective LCD defective FPGA error Cables defective or poorly connected 1. Check whether cables from the display to the main board are correctly connected. 2. Check whether cables and connectors are damaged. Replace LCD. Replace LCD. Update or upgrade FPGA. 1. Check whether the cable from the display to the main board is correctly connected. 2. Check whether the cables and connectors are damaged LED Digital Display & Indication Lamp Failure Symptoms Possible Cause Corrective Actions LED digital display or indication lamp display abnormally. Signal wires damaged LED digital display or indication lamp damaged 1. Check whether LED signal wires are correctly connected 2. Check whether cables and connectors are damaged Replace LED digital display or indication lamp Alarm Problems Symptoms Possible Cause Corrective Actions The alarm lamp is not light or extinguished but alarm sound is issued Cables defective or poorly connected Alarm LED board failure Button board failure Main board failure 1. Check whether cables from alarm LED board to button board and button board to main board are properly connected. 2. Check that connecting cables and connectors are not damaged. Replace the alarm LED board. Replace the button board. Replace the main board. 4-3
40 Symptoms Possible Cause Corrective Actions No alarm sound is issued but alarm lamp lights properly Audio alarm disabled Cable defective or poorly connected Audio program error identified in the system program Speaker failure Main board failure Select [MENU] [MAINTAIN>>] [USER MAINTAIN >>] enter the required password. In the pop-up menu, set the [MIN ALARM VOL] to a value other than zero. 1. Check whether the cable between speaker and main board is properly connected. 2. Check whether the connecting cables and connectors are damaged. Upgrade system programs Replace the speaker Replace the main board Button Failure Symptoms Possible Cause Corrective Actions Buttons do not work Cables defective or poorly connected Button board failure 1. Check whether the cable between button board and main board is properly connected 2. Check whether connecting cables and connectors are damaged Replace button board Recorder Failures Symptoms Possible Cause Corrective Action No printout Recorder module disabled Paper reversely installed Cable defective or poorly connected Recorder power supply failure Recorder failure 1. Check whether the recorder status LED is on 2. If yes, please reset the recorder. Otherwise, check for other possible causes. Re-install the paper roll 1. Check whether the cable between recorder and main board is appropriately connected 2. Check whether connecting cables and connectors are damaged Check if the power module outputs 5V DC and 12V DC correctly. Replace the recorder 4-4
41 Symptoms Possible Cause Corrective Action Poor print quality or paper not feeding properly Paper roll not properly installed Dirty thermal print head Recorder failure Stop the recorder and re-install the paper roll 1. Check the thermal print head and the paper roller for foreign matter. 2. Clean the thermal print head with an appropriate detergent Replace the recorder Interface Failures Symptoms Possible Cause Corrective Action No nurse call signals are issued Bar code scanner unable to be used Incorrect function settings Main board failure Bar code scanner failure Serial port failure Select [MENU] [MAINTAIN>>] [NURSE CALL >>] select desired nurse call output. Replace the main board Please refer to Bar Code Scanner Failures. Replace the main board Power Supply Failures Symptoms Possible Cause Corrective Action Fluctuating battery voltage Battery fails to be fully charged Battery failure Cable defective or poorly connected Power board failure Battery failure Cable defective or poorly connected Power board failure Replace battery 1. Check whether the cable between battery interface board and power module is correctly connected. 2. Check whether cables and connectors are damaged Replace the power board Replace battery 1. Check whether the cable between battery interface board and the power module is properly connected 2. Check whether the connecting cables and connectors are damaged Replace the power board 4-5
42 Symptoms Possible Cause Corrective Action Battery unable to be recharged No +3.3 V output No +5.0 V output No +12 V output Battery damaged Cable defective or poorly connected Power board failure 1. Power supply protection 2. Power board failure Replace battery and recharge the replacement battery. If the replacement battery can be recharged, the original one fails. 1. Check whether the cable between battery interface board and the power module is properly connected 2. Check whether the connecting cables and connectors are damaged Replace the power board 1. Turn off the monitor then restart it. 2. If the problem remains, disconnect the AC mains for 5 s and reconnect it, then restart the monitor. 3. If the problem still remains, replace power board. NOTE When the power module has a failure, it may cause problems to other components, e.g. the monitor suddenly shuts downs during the start-up, as the power module may have a power supply protection. In this case, troubleshoot the power module as per the procedure described in the table above. Parameter module, recorder and bar code scanner are powered by the power module. In the event that a componet malfunctions, check if the operating voltage is correct. Refer to 2 Theory of Operation for the operating voltage and measurement point of each component. 4-6
43 4.5.9 Network related problems Symptoms Possible Cause Corrective Actions The monitor cannot be connected to the CMS. Frequent dropouts and network disconnections No connection to LAN Improper LAN cable connection Incorrect IP address configuration 1. Confirm whether the cables and connectors are in good condition and whether the network is correctly connected. 2. Check whether the hub or switch facilities is correctly configured Check LAN cable connection. LAN cable shall not be longer than 50 m. Check for IP address conflict. Reconfigure IP address Software Upgrade Problems Symptoms Possible Cause Corrective Actions Bootstrap upgrade fails Program upgrade fails Power failure or unintended power off during bootstrap upgrade Incorrect network connection Wrong upgrade package has been downloaded Incorrect IP address configuration Return the main board to the factory for repair 1. Check that network connector on the monitor is used. 2. Make sure that the hub or the switcher runs normally. Check whether net cables are of the right type and have been connected correctly. Upgrade package shall be.mup files. Select desired package according to the program you want to upgrade. Configure a fixed IP address in range C as specified for the monitor. We recommend not to upgrade a program when the monitor is connected to a network with multiples PCs. 4-7
44 Bar Code Scanner Failures Symptoms Possible Cause Corrective Actions Bar code scanner does not work Bar code scanner cannot decode the target bar codes Scanned characters incorrectly displayed on the monitor Bar code scanner not powered on Incorrect interface cables used Loose Interface cables Bar code scanner not programmed based on bar code categories Unreadable bar code Scanner and bar code at improper distance and angle The monitor s bar code recognition configuration does not meet the format of the target bar code. Scanner serial port output incorrectly set Select [MAINTAIN >>] [USER MAINTAIN >>] enter required user password; set [BARCODE POWER] to [ON] in the pop-up menu to switch on the bar code scanner. Replace interface cables Make sure that interface cables are fixed Make sure to program the bar code scanner so that it can decode the bar codes of specified types. Make sure the bar code is flawless and check if bar codes of the same type can be decoded. Aim the scanner at the bar code and adjust the distance between the scanner and the bar code. Download correct bar code recognition configuration for the monitor. Configure the scanner s serial port output according to the host serial port settings NIBP Measurement Failures Symptoms Possible Cause Corrective Actions Unable to start NIBP measurement Cuff inflation/deflation failure Cables detective Main board failure Pump or valve damaged Air tubing occluded Check whether cables between NIBP module and the main board are in good condition. Replace the main board. Replace NIBP module. Replace the air tubing. 4-8
45 SpO2 Measurement Failure Symptoms Possible Cause Corrective Actions Unable to measure SpO2 SpO2 sensor failure Cables defective SpO2 module failure Main board failure Replace the SpO2 sensor. Check whether the cable between SpO2 module and the main board is in good condition. Replace SpO2 module. Replace the main board Temp Measurement Failures Symptoms Possible Cause Corrective Actions Unable to measure patient temperature. Temp probe damaged Cables detective Main board failure Replace the Temp probe. Check whether the cable between Temp module and the main board is in good condition. Replace the main board 4-9
46 FOR YOUR NOTES 4-10
47 5 Repair and Disassembly 5.1 Tools During disassembly and replacement, the following tools may be required: Philips screwdrivers Wire cutters Sharp nose pliers M6 spanner 5.2 Preparation for Disassembly Before disassembling the monitor, stop monitoring the patients, turn off the monitor and disconnect all the accessories and peripheral devices. Caution Before disassembling the monitor, be sure to eliminate the static charges first. When disassembling the parts labeled with static-sensitive symbols, make sure you are wearing electrostatic discharge protection such as antistatic wristband or gloves to avoid damaging the equipment. Properly connect the cables or wires when reassembling the monitor to avoid short circuit. When assembling the monitor, be sure to select proper screws. If an unfit screw is tightened by force, the monitor may be damaged and the screw or the part may fall off during use to cause unpredictable damage or human injury. Be sure to follow the correct sequence to disassemle the monitor. Otherwise, the monitor may be damaged permanently. Be sure to disconnect all the cables before disassembling any parts. Be sure not to damage any cables or connectors. Place the screws and parts removed from the monitor properly, avoiding avoid falling off, tarnishing or losing the parts. 5-1
48 5.3 Disassembly Separating the Front and Rear Half of the Monitor 1. If the monitor is configured with Temp module, Unscrew the two M3 6 crosshead screws with a screwdriver as shown in the figure below. Disconnect the wires from the PCBA as shown in the figure below. Unscrew the four M3 20 crosshead screws with a screwdriver(102) as shown in the figure below. 5-2
49 2. If the device is not equipped with a Temp module, Unscrew the four M3 20 crosshead screws directly with a screwdriver(102) as shown in the figure below. NOTE When separating the front and the rear part, use screwdriver(102) instead of screwdriver (107) to unscrew the four M3 20 crosshead screws. 3. Release the clips on the front and rear housings; cut the cable tie of the LCD signal wire with a wire cutter; disconnect the cables and air tubing connecting the front and the rear housings; and then separate the front and the rear housings. Clip 5-3
50 NOTE Exercise care when separating the front and the rear housings. Be sure not to damage the cables and connectors Disassembling the Front Housing Assembly 1. Unscrew the six M3 6 crosshead screws with a screwdriver and remove the button board. 2. Unscrew the four M3 8 crosshead screws with a screwdriver; disconnect the flexible cables on the button board; and then remove the LCD. 5-4
51 3. Unscrew the three M3 6 crosshead screws with a screwdriver and take out the SpO 2 rack. 4. Unscrew the M3 6 screw with a screwdriver and take out the SpO 2 module. Mindray SpO 2 module Masimo SpO 2 module Nellcor SpO 2 module 5. Unscrew the plastic nut and take out the SpO 2 signal cable. 5-5
52 5.3.3 Removing the Main Rack Assembly 1. Unscrew the two M3 6 crosshead screws with a screwdriver and take out the recorder by unclenching the clips. 2. Remove the cables from the recorder and unscrew the M3 6 crosshead screw with a screw driver. 3. Unscrew the five M3 6 crosshead screws with a screwdriver and take out the main rack assembly. 5-6
53 4. Provided that the device is configured with a Temp module, unscrew the three PT3 8 tapping screws with a screwdriver and then remove the Temp module Disassembling NIBP Module 1. Cut the cable tie with a wire cutter 2. Unscrew the three M3 6 crosshead screws with a screwdriver and take out the NIBP rack. 5-7
54 3. Unscrew the four M3 6 crosshead screws with a screwdriver and take out the NIBP module Removing the Main Board Disconnect all the cables connecting the main board and then take out the main board by unscrewing the four M3 6 crosshead screws with a screwdriver Removing the Power Board Pull out the cables that connect the power board and remove the power board by unscrewing the three M3 6 crosshead screws with a screwdriver. 5-8
55 5.3.7 Removing the Battery Assembly 1. If the monitor is equipped with the lead-acid battery assembly, unscrew the four M3 8 crosshead screws with a screwdriver. 2. If the monitor is equipped with the lithium battery assembly, unscrew the three M3 6 crosshead screws with a screwdriver Removing the Fan and Speaker Cut the cable ties on the fan, speaker and the main rack; remove the fan cushion; unscrew the three M3 15 crosshead screws with a screwdriver and remove the fan. The speaker can be removed from the main rack directly. 5-9
56 5.3.9 Removing the Power Socket 1. Remove the nut securing the grounding post with a spanner and take out the grounding terminal of the power socket. 2. Unscrew the two M3 8 crosshead screws with a screwdriver and remove the power socket Removing the Temp Module 1. Remove the Temp probe assembly first. 5-10
57 2. Unscrew the two M3 6 cross head screws with a screwdriver and take out the cover plate. 3. Unscrew the four M3 6 crosshead screws with a screwdriver and remove the Temp module main board. 5-11
58 Removing the SpO 2 Module 1. Mindray SpO 2 module needs not be removed. 2. For Masimo SpO 2 module, remove the isolation power board by unscrewing the three M3 nuts with sharp nose pliers; unscrew the three M3 6 crosshead screws with a screwdriver; and then remove the SpO 2 board. 3. For Nellcor SpO 2 module, unscrew the M3 18 crosshead screw with a screwdriver; then remove the isolation power board and the SpO 2 board. 5-12
59 6 Parts 6.1 Main Unit Exploded View Parts List SN P/N Description Qty Front housing assembly Main rack assembly (for lead-acid batteries) Rear housing assembly (with Temp module) 1 4 M Screw M M Crosshead screw M
60 6.2 Main Rack Assembly (For Lead-acid Batteries) Exploded View 6-2
61 Parts List SN P/N Description Qty Main rack (for lead-acid batteries) Lead-acid battery assembly 1 3 M Crosshead screw M Speaker and cables Speaker spacer Power board insulating plate 1 7 M Crosshead screw M Power board for lead-acid battery 1 9 M Bolt M M M3 nut with spring washer Spring strip Fan and cables Fan rack 1 14 M Crosshead screw M Fan spacer Grounding terminal 1 17 M Crosshead screw M Power board AC input cable External connector overlay Battery bar setscrew Battery baffle Torsion spring NIBP rack D D NIBP unit 1 25 M Flat washer Main board Main board insulating plate Key pad insulating plate NIBP rack overlay 1 30 M Flat washer 1 31 M Spring washer 1 32 M Stainless steel nut 1 6-3
62 6.3 Main Rack Assembly (For Lithium Battery) Exploded View 6-4
63 Parts List SN P/N Description Qty Main rack (for lithium battery) Lithium battery assembly 1 3 M Crosshead screw M Speaker and cables Speaker spacer Power board insulating sheet 1 7 M M3 nut with spring washer Power board for lithium battery 1 9 M Bolt M Spring strip Power board Fan rack 1 13 M Crosshead screw M Fan spacer Grounding terminal 1 16 M Crosshead screw M Power board AC input cable External connector overlay Battery baffle setscrew Battery baffle Torsion spring NIBP rack D D NIBP unit 1 24 M Flat washer Main board Main board insulating plate Key pad insulating plate NIBP rack overlay 1 29 M Flat washer 1 30 M Spring washer 1 31 M Stainless steel nut 1 6-5
64 6.4 Front housing Assembly Exploded View Parts List SN P/N Description Qty Front housing Waterproof seal Key pad Key pad fixer 1 5 M Flat washer Dust proof strip LED shade Key pad SpO2 module kit (Masimo) 1 10 M Crosshead screw M M3 nut with spring washer Grounding washer LCD spacer LCD screen Dust proof strip M Tapping screw PT Power button 1 6-6
65 SN P/N Description Qty Battery door link Battery door 1 20 M Tapping screw PT M Flat washer NIBP connector for main unit Front housing overlay (with Temp module) Alarm LED cover Rear Housing Assembly Exploded View Parts List SN P/N Description Qty Rear housing Label (Chinese) 1 3 M Screw M Temp module Foot cushion Speaker spacer Recorder support 1 8 M Tapping screw PT M Washer 1 6-7
66 SN P/N Description Qty 10 M Tapping screw PT M Flat washer Handle cover 2 13 M Crosshead screw M TR6C TR60-C recorder Lithium Battery Charger Assembly Exploded View Parts List SN P/N Description Qty Lithium battery mounting rack Lithium battery receptacle board 1 3 M Bolt M Spring 1 5 M Screw M M Stainless steel screw M M Insulating spacer ф
67 6.7 Lead-acid Battery Assembly Exploded View Parts List SN P/N Description Qty Lead-acid battery mounting rack Spring Battery charging flexible support Battery contact 2 5 M S--- Switch 1 6 M Screw M Spring 1 8 M M3 nut with spring washer 1 9 M Tapping screw M M Flat washer 2 6-9
68 6.8 MASIMO SpO 2 Board Assembly Exploded View Parts List SN P/N Description Qty 1 DA8K insulating plate for SpO2 isolation power board 1 2 M Insulation washer 3 3 M Stainless steel nut, GB6170-M Masimo M7 isolation power board Masimo M7 board(masimo kit) 1 6 M Spring washer 3 7 M Bolt M SpO2 mounting rack 1 9 M Screw M
69 6.9 Temp Assembly Exploded View Parts List SN P/N Description Qty Temp module housing M Tapping screw PT Temp module housing M Screw M Temp module waterproof pad 1 6 M Crosshead screw M M09A Temp main board assembly Temp probe cover pack support Temp probe well
70 P/N: (3.1)
BeneView T1. Patient Monitor. Service Manual
BeneView T1 Patient Monitor Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this
Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF
Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator
imec12/imec10/imec8 Patient Monitor Service Manual
imec12imec10imec8 Patient Monitor Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to
How To Use A Power Supply Unit (Upu)
BRAVER UPS (Uninterruptible Power System) User s Manual Safety CAUTION! This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user replaceable parts.
Installation and Operation Back-UPS 1250, 1300, 1500
Installation and Operation Back-UPS 1250, 1300, 1500 Inventory bu001a Safety and General Information This unit is intended for indoor use only. Do not operate this unit in direct sunlight, in contact with
Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS
+ Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS Inventory Safety and General Information bu001c This unit is intended for indoor use only. Do not operate this unit in direct sunlight,
Laser Barcode Scanner
Laser Barcode Scanner User s Manual FCC Compliance This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
User s Manual AURORA 1.2K/2.2K
User s Manual AURORA 1.2K/2.2K Uninterruptible Power System Safety CAUTION This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user serviceable
StorTrends 3400 Hardware Guide for Onsite Support
StorTrends 3400 Hardware Guide for Onsite Support MAN-3400-SS 11/21/2012 Copyright 1985-2012 American Megatrends, Inc. All rights reserved. American Megatrends, Inc. 5555 Oakbrook Parkway, Building 200
AC-115 Compact Networked Single Door Controller. Installation and User Manual
AC-115 Compact Networked Single Controller Installation and User Manual December 2007 Table of Contents Table of Contents 1. Introduction...5 1.1 Key Features... 6 1.2 Technical Specifications... 7 2.
Infinity C Reference Guide
1 2! Infinity C Reference Guide Table of Contents Components... 1 Hardware Setup... 5 SmartDrive Configuration... 12 Startup... 15 Login... 16 System Configuration... 19 DICOM Setup... 20 Monitoring Status...
Failure code manual. content
Failure code manual content 一 wall split AC series 2 二 floor standing AC series. 4 三 portable AC series.. 5 四 dehumidifer 6 五 DC inverter single split series...7 六 DC inverter multi-split series 10 1 一
USER S MANUAL. MaxPower 400-600 UPS. Uninterruptible Power System 28-2MAXPO0018
USER S MANUAL MaxPower 400-600 UPS Uninterruptible Power System 28-2MAXPO0018 IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS This manual contains important instructions for models MaxPower 400 and
TMS TANK MANAGEMENT SYSTEM
TMS TANK MANAGEMENT SYSTEM Page 1 of 9 Operating Instructions GENERAL The Tank Management System is a bespoke design to control, monitor and accommodate efficient storage and dispensing of TMS. FUNCTIONS
SERVICE INSTRUCTION R410A. WALL MOUNTEDtype INVERTER SPLIT TYPE ROOM AIR CONDITIONER. Models Indoor unit Outdoor unit
SERVICE INSTRUCTION SPLIT TYPE ROOM AIR CONDITIONER WALL MOUNTEDtype INVERTER Models Indoor unit Outdoor unit ASYG07LECA ASYG09LECA ASYG12LECA ASYG14LECA AOYG07LEC AOYG09LEC AOYG12LEC AOYG14LEC R410A CONTENTS
LS1024B / LS2024B/ LS3024B. Solar Charge Controller USER MANUAL
EPSOLAR LS1024B / LS2024B/ LS3024B Solar Charge Controller USER MANUAL Thank you very much for selecting our product! This manual offers important information and suggestions with respect to installation,
MCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
Back-UPS Pro 1300/1500 Installation and Operation
Back-UPS Pro 1300/1500 Installation and Operation Inventory Safety Do not install the Back-UPS in direct sunlight, in excessive heat, humidity, or in contact with fluids. Connect the battery bu059a bu058a
Lenovo IdeaCentre Q180 Series
Machine type: 10087/3110 Lenovo IdeaCentre Q180 Series User Guide Version 1.0 2011.09 31500665 Important Safety Information Before using this manual, it is important that you read and understand all of
Service manual. Website: www.andico.com.au CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL.
Website: www.andico.com.au Service manual CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL. - ONLY FOR AUTHORISED SERVICE PERSONNEL. MODELS: MPK1-09CR-QB8 MPK1-12ER-QB6
Online UPS. PowerWalker VFI 1000R/1U
Online UPS PowerWalker VFI 1000R/1U Manual Uninterruptible Power Supply System Table of Contents 1. Important Safety Warning... 1 1-1. Transportation... 1 1-2. Preparation... 1 1-3. Installation... 1 1-4.
How to read this guide
How to read this guide The following shows the symbols used in this Quick start guide with descriptions and examples. Symbol Description Example P oint Reference Caution [ ] This symbol explains information
HP UPS R1500 Generation 3
HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2. Instruction Manual
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2 Instruction Manual Firmware version: 2.0.1 Issue: 0.6 Version of the GPRS transmitters configurator: 1.3.6.3 Date of issue: 07.03.2012 TABLE OF CONTENTS
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W
MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries
Line Interactive UPS
EN Line Interactive UPS PowerMust 636 LCD (650VA), Line Int., Schuko PowerMust 848 LCD (850VA), Line Int., Schuko PowerMust 636 LCD (650VA), Line Int., IEC PowerMust 848 LCD (850VA), Line Int., IEC PowerMust
RIGOL. Quick Guide. DS1000CA Series Oscilloscope. Aug. 2011. RIGOL Technologies, Inc.
Quick Guide DS1000CA Series Oscilloscope Aug. 2011 Technologies, Inc. Guaranty and Declaration Copyright 2011 Technologies, Inc. All Rights Reserved. Trademark Information is a registered trademark of
Keep it Simple Timing
Keep it Simple Timing Support... 1 Introduction... 2 Turn On and Go... 3 Start Clock for Orienteering... 3 Pre Start Clock for Orienteering... 3 Real Time / Finish Clock... 3 Timer Clock... 4 Configuring
Heat Surge Model X5C Fire Place Insert Service Manual Applies to all units w/30000208 circuit board
Heat Surge Model X5C Fire Place Insert Service Manual Applies to all units w/30000208 circuit board 2012 HS M4417A BR16597R-1 HEAT SURGE 8000 FREEDOM AVE, N. CANTON, OH 44720 330-244-8161 WWW.HEATSURGE.COM
M O D U L E - 7 E Model CS-Caliprompter Operator s Manual
O P E R A T I O N S A N D P R O C E D U R E S F O R C S - C A L I P R O M P T E R A N D A C C E S S O R I E S Model CS-Caliprompter Visual screen display allows easy step-by-step operation On-site programming
MicroTech II McQuay Maverick II Rooftop Unit Controller BACnet Communication Module (MS/TP)
Installation and Maintenance Manual IM 852 Group: Controls Part Number: IM 852 Date: June 2007 Supercedes: New MicroTech II McQuay Maverick II Rooftop Unit Controller BACnet Communication Module (MS/TP)
Square D Clipsal DIN-Rail Four-Channel Auxiliary Input Unit
Square D Clipsal DIN-Rail Four-Channel Auxiliary Input Unit SLCLE5504AUX for Use with Wired C-Bus Networks Instruction Bulletin Retain for future use. Square D Clipsal DIN-Rail Four-Channel Auxiliary Input
Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011
Instruction Manual 2in1 LAN Tester & Multimeter Model: LA-1011 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.
BeneView T8. Patient Monitor. Service Manual
BeneView T8 Patient Monitor Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this
D. Infrared scanning reports with pictures printed in a final report with any deficiencies and actions taken to rectify.
SECTION 263354 STATIC UPS TESTING AND COMMISSIONING PART 1 - GENERAL 1.1 SUMMARY A. Section Includes 1. System start-up services. 2. Battery charging. 3. Load bank testing. 4. Battery load testing. 5.
Ground Resistance Clamp On Tester
USER MANUAL Ground Resistance Clamp On Tester MODEL 382357 Introduction Congratulations on your purchase of Extech s 382357 Ground Resistance Tester. This Clamp on device allows the user to measure ground
Match. GE Digital Energy. Uninterruptible Power Supply 500-1500 VA. Technology for the Digital World. Match UPS. GE Digital Energy.
Match Uninterruptible Power Supply 500-1500 VA Manufactured by: General Electric Company Telephone +41 (0)91 / 850 51 51 CH 6595 Riazzino (Locarno) Fax +41 (0)91 / 850 51 44 Switzerland Website www.gedigitalenergy.com
PRONET WEB SERVER. INSTRUCTIONS FOR INSTALLATION AND USE English. Version 01/15 Ident no. 50950129
PRONET WEB SERVER INSTRUCTIONS FOR INSTALLATION AND USE English EN Version 01/15 Ident no. 50950129 Table of Contents 1. About these instructions 3 2. Important information for your safety 4 2.1. Intended
Laser Barcode Scanner User s Manual
Laser Barcode Scanner User s Manual FCC Compliance This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
Current Monitoring Kit QUICK START GUIDE
Current Monitoring Kit QUICK START GUIDE Pico Technology EL040 Current Monitoring Kit Quick Start Guide CONTENTS 1 Introduction... 1 2 Kit contents... 3 3 Installation... 3 3.1 Connecting the system together...
Line Interactive UPS. PowerWalker VI 500T/HID PowerWalker VI 750T/HID PowerWalker VI 1000T/HID PowerWalker VI 1500T/HID.
Line Interactive UPS PowerWalker VI 500T/HID PowerWalker VI 750T/HID PowerWalker VI 1000T/HID PowerWalker VI 1500T/HID Quick Start Guide CONTT IMPORTANT SAFETY INSTRUCTIONS... 1 1. Introduction... 4 2.
T0118 T2118 T3118. Instruction Manual
Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity
VFS24/32HDIP. Public Display IP Monitor User Manual
VFS24/32HDIP Public Display IP Monitor User Manual 2 Contents Before You Begin...4 Side Panel Control buttons...6 Connections...7 OSD Function...7 LCD monitor Mounting Guide...9 Getting started... 10 Power
Current Monitoring Kit
Current Monitoring Kit QUICK START GUIDE DO090-6 CONTENTS Issues: 1) 2.10.02 WP A4 format 2) 2.10.02 Added safety warning 3) 17.3.06 Word A5 format. S1: Removed relay modules. S2: Added MI010. S4.1: added
Uninterruptible Power Supply ERA LED 1.5 ERA LED 2.0 ERA LED 2.6. User s manual
Uninterruptible Power Supply ERA LED 1.5 ERA LED 2.0 ERA LED 2.6 User s manual Index Safety Warnings... 2 1 Introduction... 3 2 General Characteristics... 4 3 Receipt and site selection... 4 4 EXTERNAL
PITE 3921 Battery Monitoring System User Manual P-140606-V1.0
User Manual P-140606-V1.0 4/F, Bldg A, Chiwan Industrial Park, Shaodi Rd., Chiwan, Shekou Area, Shenzhen, China TEL: +86-755-2680 5759 FAX: +86-755-2688 0310 www.pitetech.com 1 Catalog 1. INTRODUCTION...
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
FCC COMPLIANCE STATEMENT FOR AMERICAN USERS
FCC COMPLIANCE STATEMENT FOR AMERICAN USERS This equipment has been tested and found to comply with the limits for a CLASS A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
430 Power/Electronics Replacement
Replacing the main board WARNING Before proceeding, turn off the main power switch and unplug the power cord. Caution Make sure you are properly grounded with an ESD strap before continuing. The main printed
Wash Select II to WashPay Integration Field Retrofit Installation Guide
Wash Select II to WashPay Integration Field Retrofit Installation Guide Wash Select II Software Version 6.00 and above POS4000 Software Version 8.25 Unitec www.startwithunitec.com WASH SELECT II TO WASHPAY
Quick Installation. A Series of Intelligent Bar Code Reader with NeuroFuzzy Decoding. Quick Installation
Quick Installation A Series of Intelligent Bar Code Reader with NeuroFuzzy Decoding This chapter intends to get your new FuzzyScan scanner working with your existing system within minutes. General instructions
QUICK START GUIDE Bluetooth Cordless Hand Scanner (CHS)
QUICK START GUIDE Bluetooth Cordless Hand Scanner (CHS) 1D Imager Models CHS 7Ci, 7Di, 7DiRx LED Trigger button Power button (also for ios Keyboard Pop-up) Model shown: CHS 7Di This document pertains to
BCST-20 Barcode Scanner. Instruction Manual. www.inateck.com
BCST-20 Barcode Scanner Instruction Manual www.inateck.com IMPORTANT NOTICE Safety Precaution * DO NOT disassemble the scanner, or place foreign matter into the scanner causing a short circuit or circuit
LEN s.r.l. Via S. Andrea di Rovereto 33 c.s. 16043 CHIAVARI (GE) Tel. +39 0185 318444 - Fax +39 0185 472835 mailto: [email protected] url: http//www.len.
MA511 General Index 1 INTRODUCTION... 3 1.1 HARDWARE FEATURES:... 4 2 INTERFACE... 5 2.1 KEYBOARD... 6 2.2 POWER ON... 7 2.3 POWER OFF... 7 2.4 DETECTOR CONNECTION... 7 2.5 DETECTOR SUBSTITUTION...7 3
AutoRanging Digital MultiMeter
Owner's Manual AutoRanging Digital MultiMeter Model No. 82139 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance
UM-X Field display for continous level sensors
Technical Documentation Field display for continous level sensors 10/2007 Edition: 1 Item No.: 207120 FAFNIR GmbH Bahrenfelder Str. 19 D-22765 Hamburg Telephone: +49 (0)40-39 82 07-0 Fax: +49 (0)40-3 90
ADM1TE 5/30A DIN rail single phase two wire energy meter
ADMTE 5/30A DIN rail single phase two wire energy meter. Safety instruction.2 Foreword.3 Performance criteria.4 Specifications.5 Basic errors.6 Description.7 Dimensions.8 Installation.9 Operating.0 Troubleshooting.
PK5500 v1.1 Installation Instructions
PK5500 v1.1 Installation Instructions 1 2 3 4 5 6 7 8 9 * 0 # WARNING: Please refer to the System Installation Manual for information on limitations regarding product use and function and information on
Wireless Laser Barcode Scanner ils 6300BU. User s Manual
Wireless Laser Barcode Scanner ils 6300BU User s Manual FCC Compliance This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
HP ProLiant DL380 G6 Carrier-Grade Server Read Before Install
HP ProLiant DL380 G6 Carrier-Grade Server Read Before Install Carrier-Grade Instructions HP Part Number: AM275-9001A Published: July 2012 Edition: 3 Copyright 2009, 2012 Hewlett-Packard Development Company,
Getting Started. rp5800, rp5700 and rp3000 Models
Getting Started rp5800, rp5700 and rp3000 Models Copyright 2011 Hewlett-Packard Development Company, L.P. The information contained herein is subject to change without notice. Microsoft, Windows, and Windows
CARESCAPE V100 Vital Signs Monitor
GE Healthcare CARESCAPE V100 Vital Signs Monitor Service Manual Silence Systolic HIGH MAP/Cuff Inflate/Stop Alarms Diastolic LOW HIGH ADULT NEONATE AUTO CYCLE Cycle LOW HISTORY History To clear hold 2
CFMRI Physiological Monitoring System. Operator s Manual
CFMRI Physiological Monitoring System Operator s Manual This manual provides a quick guidance to using the physiological monitoring system during an MRI session. It is not intended as a substitute for
Multi-Protocol decoder 76 200 with Load regulation
Multi-Protocol decoder 76 2 with Load regulation For locomotives with universal motors on digital layouts operating in the DCC and Motorola data format. Features 76 2 Load regulated multi-protocol decoder
Taurus Super-S3 LCM. Dual-Bay RAID Storage Enclosure for two 3.5-inch Serial ATA Hard Drives. User Manual March 31, 2014 v1.2 www.akitio.
Dual-Bay RAID Storage Enclosure for two 3.5-inch Serial ATA Hard Drives User Manual March 31, 2014 v1.2 www.akitio.com EN Table of Contents Table of Contents 1 Introduction... 1 1.1 Technical Specifications...
EKOS Cart. Instructions for Use
EKOS Cart Instructions for Use EKOS Corporation 11911 North Creek Parkway South Bothell, WA 98011 USA (425) 415-3100 (tel) (425) 415-3102 (fax) [email protected] (e-mail) - 1-4913-002 REV E Intended Use
RISH CON - Hz. Salient Features : Application : Product Features: FREQUENCY TRANSDUCER
Application : The RISH CON - Hz transducer is used for frequency measurement. The output signal is proportional to measured frequency and is either load independent DC Current or load independent DC Voltage.
3.5 EXTERNAL NETWORK HDD. User s Manual
3.5 EXTERNAL NETWORK HDD User s Manual Table of Content Before You Use Key Features H/W Installation Illustration of Product LED Definition NETWORK HDD Assembly Setup the Network HDD Home Disk Utility
OPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
5-port / 8-port 10/100BaseTX Industrial Ethernet Switch User Manual
5-port / 8-port 10/100BaseTX Industrial Ethernet Switch User Manual Content Overview... 1 Introduction... 1 Features... 3 Packing List... 4 Safety Precaution... 4 Hardware Description... 5 Front Panel...
Unpacking the Product. Rack Installation. Then, use the screws provided with the equipment rack to mount the firewall in the rack.
About This Guide This guide contains step-by-step instructions for setting up the D-Link DFL-260E/860E Firewall. Please note that the model you have purchased may appear slightly different from those shown
Chapter 4: Disassembly and Power
Chapter 4: Disassembly and Power Complete CompTIA A+ Guide to PCs, 6e How to prevent static electricity, RFI, and EMI from harming or interfering with a computer The tools needed to work on computers How
How To Power A Power Control On An Ip40 (Ipl) With A Power Supply (Iplug) With An Ip20 Controller (Iphones) With Power Control (Power Control) With No Antenna) With The Ip20 (Power)
MODEL NUMBER: ISC910-1-0-GB-XX ISC911-5-0-GB-XX IXP20 CONTROLLER SPECIFICATIONS Working Environment Plastic Housing... Power ImproX IXP20 Controller INSTALLATION MANUAL Designed to work in an indoor (dry)
User s manual FLIR VP50/VP52 Non-contact AC voltage detector
User s manual FLIR VP50/VP52 Non-contact AC voltage detector User s manual FLIR VP50/VP52 #T559851; r. AD/ 9134/9134; en-us Table of contents 1 Disclaimers... 1 1.1 Copyright... 1 1.2 Quality assurance...
Manual for Fire Suppression & Methane Detection System
Manual for Fire Suppression & Methane Detection System Fogmaker North America Post address: 150 Gordon Dr Exton, PA 19341 Delivery address: 150 Gordon Dr Exton, PA 19341 Tel: 610-265-3610 Fax: 610-265-8327
User s Manual Network Management Card
User s Manual Network Management Card RMCARD202 Intelligent Network Management Card allows a UPS system to be managed, monitored, and configured Version 1.0 E-K01-SNMP005-0 TABLE OF CONTENTS Introduction
Mercury Helios 2 ASSEMBLY MANUAL & USER GUIDE
Mercury Helios 2 ASSEMBLY MANUAL & USER GUIDE TABLE OF CONTENTS INTRODUCTION...1 1.1 MINIMUM SYSTEM REQUIREMENTS 1.1.1 Apple Mac Requirements 1.1.2 PC Requirements 1.1.3 Supported PCIe Cards 1.2 PACKAGE
BW TECHNOLOGIES GasAlertMicroClip Soft Tools Instruction Sheet
Introduction BW TECHNOLOGIES GasAlertMicroClip Soft Tools Instruction Sheet The GasAlertMicroClip Soft Tools application is required to configure the GasAlertMicroClip detector ( the detector ). Software/Hardware
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
User Manual for CH-PFC76810
AA Portable Power Corp www.batteryspace.com, Email: [email protected] User Manual for CH-PFC76810 1. Overview The CH-PFC76810 charger is suitable for charging lithium ion battery packs such as those
Kramer Electronics, Ltd. Site-CTRL and Web Access Online User Guide (Documentation Revision 2)
Kramer Electronics, Ltd. Site-CTRL and Web Access Online User Guide (Documentation Revision 2) Software Version 1.9. 2902 To check that you have the latest version, go to the DOWNLOADS section of our Web
PCS300 Universal IP Reporting Module V1.0
PCS300 Universal IP Reporting Module V1.0 Reference and Installation Manual Patents: One or more of the following US patents may apply: 7046142, 6215399, 6111256, 6104319, 5920259, 5886632, 5721542, 5287111,
USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR
USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting
IP DSLAM IDL-2402. Quick Installation Guide
IP DSLAM IDL-2402 Quick Installation Guide Table of Contents Package Contents... 3 Overview... 4 Setup the IDL series IP DSLAM... 5 Safety Instruction... 5 Hardware Installation... 6 WEB Configuration...
Memcom Emergency Telephone
Memcom Emergency Telephone Installation Guide Ref No. 450 900 (GB) Version 2 + + Simple wiring for quick installation + + Integrated LCD display shows you what you have programmed + + All code based programming
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
Model 5511 Filler Controller User s Manual Version 1.1 October 2011
Thompson Scale Company WEIGHING SYSTEMS & PACKAGING MACHINERY CONTROLS 2758 Bingle Road Houston, Texas 77055 Phone: 713/932-9071 Fax: 713/932-9379 www.thompsonscale.com Model 5511 Filler Controller User
Cisco TelePresence VCR MSE 8220
Cisco TelePresence VCR MSE 8220 Getting started 61-0008-05 Contents General information... 3 About the Cisco TelePresence VCR MSE 8220... 3 Port and LED location... 3 LED behavior... 4 Installing the VCR
T Sentry 4 Multi-Point Digital Alarm Instruction Manual
T Sentry 4 Multi-Point Digital Alarm Instruction Manual Introduction The T Sentry4 (TS4) is a microprocessor-based temperature monitoring and alarm device with user programmable high-low alarm setpoints
Si10-417_C. Pocket Manual. Service Diagnosis SPLIT & MULTI
Pocket Manual Service Diagnosis SPLIT & MULTI Service Diagnosis SPLIT & MULTI 1. Troubleshooting with LED...5 1.1 Indoor Unit... 5 1.2 Outdoor Unit... 10 2. Troubleshooting by Symptoms...11 2.1 Air conditioner
Ericsson W25. Mobile Networks. Quick Installation Guide
Ericsson W25 Fixed Wireless Terminal for WCDMA/HSDPA Mobile Networks Quick Installation Guide Table of Contents INTRODUCTION... 3 Package Contents... 3 Requirements... 3 INSTALLATION... 4 Unit Preparation...
Wireless Link. Introduction. Setup, Configuration and Connection ENGLISH
ENGLISH Introduction This document is an addendum to the HeartStart MRx Instructions for Use. It explains how to use the features of the HeartStart MRx s Wireless Link Option. The Wireless Link Option
Manual Ranging MultiMeter
Owner s Manual Manual Ranging MultiMeter Model 82345 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product.! Safety! Operation! Maintenance!
CR9971 2 Band Auto Set Dual Alarm Clock Radio ROBERTS. Sound for Generations. Please read this manual before use
ROBERTS Sound for Generations CR9971 2 Band Auto Set Dual Alarm Clock Radio Please read this manual before use Contents Contents... 1 Controls...2-5 Switching on... 6 Setting the time manually... 6 Setting
Back-UPS 650 VA 120 V with AVR (BX650CI-LM)
Back-UPS 650 VA 120 V with AVR (BX650CI-LM) Overview Do not install the unit in direct sunlight, in areas of excessive heat or humidity, or in contact with fluids ON/OFF button Battery connector Circuit
1 General information
1 General information 1.1 About this manual This manual describes the installation, function and operation of an integrated solar control system, which is suitable for split pressurized solar heating system.
USER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING. Non-Programmable Single Stage Heat/Cool Thermostat
Builder MODEL 1010 Series Non-Programmable Single Stage Heat/Cool Thermostat USER MANUAL Compatible with low voltage single stage gas, oil or electric heating or cooling systems, including single stage
14. Troubleshooting Guide
14. Guide 14.1 Refrigeration Cycle System In order to diagnose malfunctions, ensure the air conditioner is free from electrical problems before inspecting the refrigeration cycle. Such problems include
EGCP-3 LS 8406-113 Revision L. Explanation of Software Changes Made for the EGCP-3 LS Control. Application Note 51294
EGCP-3 LS 8406-113 Revision L Explanation of Software Changes Made for the EGCP-3 LS Control Application Note 51294 WARNING DANGER OF DEATH OR PERSONAL INJURY WARNING FOLLOW INSTRUCTIONS Read this entire
Network Interface Panel User s Guide NPD4747-01 EN
User s Guide NPD4747-01 EN Contents Contents Copyright and Trademarks Safety Instructions Important Safety Instructions... 4 Cautions and Notes... 4 Getting to Know Your Network Interface Panel Package