The amount of dispersed suspended solids in natural
|
|
|
- Percival Barnett
- 10 years ago
- Views:
Transcription
1 Application Notes / Useful & Timely Information from Hach Hydromet Turbidity Measurement: A Simple, Effective Indicator of Water Quality Change INSTRUMENTATION AND BEST PRACTICE TECHNIQUES ARE KEY TO EFFECTIVELY MONITOR TURBIDITY IN NATURAL WATER BODIES by Mike Sadar solids in lakes, rivers, and streams. High levels of suspended sediment often create negative effects on aquatic life. For example, suspended sediment can interfere with photosynthesis by blocking light from reaching submerged aquatic plants. This not only directly damages vegetation but also results in reduced levels of dissolved oxygen because of the reduced rates of photosynthesis. Moreover, waters with high levels of suspended solids absorb more sunlight and can therefore cause an increase in water temperature that can cause dissolved oxygen levels to drop even further. Low dissolved oxygen then stresses aerobic aquatic organisms and could ultimately lead to fish kills. The amount of dispersed suspended solids in natural water bodies is an important indicator of water quality. These solids that often include silt, clay, algae, organic matter, and other minute particles, obstruct the transmittance of light through the water and impart a qualitative characteristic known as turbidity. Turbidity is often closely correlated to climatological or surface water conditions and changes in turbidity are therefore indicators of changes in environmental conditions. Turbidity data can be used as a surrogate measurement because it is strongly correlated with sediment, nutrients and bacteria, and can be measured in-stream on a continuous basis. Turbidity can be interpreted as a measure of the relative clarity of water. It is not a direct measure of suspended particles in water but, instead, a measure of the scattering and attenuation effects these particles have on light. The higher the intensity of the scattered or attenuated light, the higher the value of turbidity. Human activities such as logging, mining, road building, and commercial construction can often lead to chronic levels of suspended Turbidity is a Valuable Surrogate Section 303(d) of the federal Clean Water Act requires States to establish total maximum daily loads (TMDLs) of various pollutants to help meet water quality goals and to attain designated beneficial uses for each water body. 1 The ability to continuously measure water constituents that are most commonly associated with impairments is often limited by technical and financial constraints. Turbidity, however, can be used effectively as a surrogate measurement because it can be measured in-stream on a continuous basis and it is strongly correlated with sediment, nutrients, and bacteria concentrations. The United States Geological Survey (USGS) describes a surrogate as an environmental measurement than can be reliably correlated with an in-stream characteristic, such as concentration or particle-size distribution of fluvial sediment. Surrogate data are typically easier, less expensive, and (or) safer to collect than the target variable, and may enable reliable estimates of uncertainty associated with the measurement. 2 Turbidity can be used as a surrogate measurement for many environmental influencers, such as for: Monitoring the impact of humans on natural water bodies. Monitoring pathogens in water, such as E.coli in stormwater runoff from cattle pastures. Monitoring sediments to track erosion and landscape change.
2 Monitoring natural streams below mining and dredging operations. Measuring total phosphorous in water is very difficult, but an increase in phosphate or phosphorous typically correlates to an increase in turbidity levels. Many parameters influence overall water quality, yet turbidity is especially noteworthy because it is a simple and undeniable indicator of water quality. This allows it to serve as a surrogate for other factors or conditions. For example, high turbidity can mean higher concentrations of bacteria, nutrients, pesticides, or metals. So, a sudden change in turbidity may indicate the presence of a new pollution source (biological, organic or inorganic) in natural waters. Turbidity Measurement Technologies It s important to remember that turbidity is not a measure of the quantity of suspended solids in a sample but, instead, an aggregate measure of the combined scattering effect of the water sample s suspended particles on an incident light source. All turbidity measurements detect the amount of light either transmitted through or scattered by the particles in a sample of water. Over the past several decades, instrument technology has advanced dramatically and many turbidity measurement techniques have resulted. These new approaches evolved to address interferences or inconveniences associated with earlier turbidity measurement techniques. Different technological approaches (often using different light sources and detector designs) have been used to compensate for or minimize measurement errors attributed to color, bubbles, stray light, absorption, and path length. Common Turbidimeter Design Criteria Common turbidimeters involve a light source and one or more detectors with a specific orientation to one another. The vast number of turbidity monitoring technologies can be categorized by three design criteria -- the type of incident light source that is used, the detection angle for the scattered light, and the number of scattered or attenuated light detectors used. Incident light sources: There are three types of light sources used commonly in turbidimeters: incandescent, LED, and laser. INCANDESCENT LIGHT SOURCES are typically polychromatic with tungsten filaments with a color temperature of 2200 to 3000º Kelvin and emit relatively short wavelengths. These shorter wavelengths will be more effectively scattered by smaller particles. Those methods that are typically compliant to USEPA Method or Standard Methods 2130B will utilize an incandescent light source. LED (LIGHT EMITTING DIODE) LIGHT sources are lower energy emitting light sources when compared to tungsten filament lamps. The most common wavelengths used in turbidity measurements emit nm light that is typically not absorbed by visible color in the sample, thus eliminating a common error source in most turbidity measurements. The International Standardization Organization (ISO) method 7027 requires the use of a light source in this range. LASER-BASED LIGHT SOURCES. A small portion of turbidity measurement techniques will have laser-based incident light sources that emit energy at a discrete wavelength. Laser-based light sources are very sensitive to small changes in turbidity and are often used to monitor filtration performance for clean waters, such industrial processes requiring ultrapure water. Detection Angle: The detection angle is the angle formed between the centerline of the incident light beam and the centerline of the detector s receive angle. The vertex of the angle is located in the center of the sample view volume and is represented by a red dot depicted in the following figures. The detection angle can have a significant impact on the detection of particles from a size perspective and on the turbidity range of the instrument. Also, the number of the detectors and their relative angle to the incident light beam can help reduce the impact of interferences such as color and subtle changes in the instrument components, such as in the degradation of a light source or fouling of a detector. A 90-DEGREE DETECTION ANGLE is often referred to as the nephelometric detection angle and is the most common detection angle because of its sensitivity to a broad range of particle sizes (Figure 1). A slight variation of this approach is to utilize a design that does not come into contact with the sample with its measurement optics. This technique measures light scatter below the surface of the water and is commonly referred to as a surface scatter technology. Figure 1.The 90-degree detection angle is the default technology for most regulatory applications. It is very susceptible to color interference and is best used at low turbidity levels. THE ATTENUATED DETECTION ANGLE is geometrically oriented at an angle that is 180-degrees relative to the incident light beam. This detection angle measures the attenuation of the incident light beam due to both light scatter and absorption. This angle has the greatest susceptibility to absorbance and color interferences. Figure 2.The attenuated detection angle is very susceptible to color as a negative interference and color absorption interference. Measurement sensitivity is highly wavelength dependent. 2
3 Figure 3.The backscatter angle measures the light that is scattered in the direction of the incident light beam. It is used for high turbidity samples, but is a poor application for monitoring clean waters. DUAL LIGHT SOURCE DETECTOR. This unique ratio approach uses a combination of light sources that are geometrically oriented at 90-degree angles to each other. The detectors are also oriented at 90-degrees to each other and at 90 and 180-degrees to each of the light sources. In one phase of measurement, a detector will be the nephelometric (90-degree) detector and the other detector will be at 180-degrees to the light source that is powered. In the second phase of the measurement, the second light source will be powered and the detector positions from phase one are reversed. A software algorithm is then used to generate the turbidity value from different measurement phases. The combination of the two phases provides a turbidity measurement that is corrected for color absorption, fouling of the optics, and any optical changes that can occur. THE BACKSCATTER DETECTION ANGLE has a detector that is geometrically centered at an angle of between 0 and 45 degrees relative to the directional centerline of the incident light beam. This angle will be sensitive to light scatter that is reflected back in the direction of the incident light source, which is characteristic with extremely high turbidity samples. This is not an appropriate technique for low level turbidity monitoring because it has inherent poor sensitivity at these levels. Number of Detectors: Some turbidity measurement technology involves the use of signal ratios of two or more detectors to determine the turbidity value. One detector is at 90-degrees from the incident source and the other detector(s) can be at any of several different angles. MULTIPLE DETECTION ANGLES. This ratioing approach will utilize one primary detector, which is typically oriented at a 90-degree angle relative to the incident light beam, and it is often referred to as the primary nephelometric detector. Other detectors will be at various angles including an attenuated, backscatter, and forward scatter angles. A software algorithm is often used to produce the turbidity measurement from the combination of detectors. These detectors can help compensate for color interference and in optical changes such as light source degradation. Figure 4. An In-Situ probe turbidimeter with a ratio design. The In-situ designates the measurement to be made in the process (the river, lake, sample transport pipe) itself, Two detectors extend the range of measurement to low turbidity levels and reduce color interference. Figure 5. A dual-light source, dual-detector design. Two LED light two detectors combine to perform the turbidity measurement. The measurement consists of two phases. This design eliminates interferences due to color, fouling of optical components, and is very stable over time. Variability Based on Measurement Method It is normal for different turbidity measurement technologies to deliver different results on the same sample. This is because natural suspended solids - whether algae, silt, organic material, etc. - have wildly different and practically unpredictable optical characteristics, such as the way they absorb and reflect light. From technology to technology, differences in the combination of incident light source, detection angle, and number of detectors, together with the natural variation in sample optical characteristics, lead to different measurement results. Because of this potential for variability, it is important to also provide information on the type of technology used to collect a given set of data. Nephelometers, or nephelometric turbidimeters, measure the light scattered at an angle of 90 by one detector from the incident light beam generated by an incandescent light bulb. Readings are reported in Nephelometric Turbidity Units, or NTUs. NTU has been the traditional reporting unit for turbidity and is still recognized by some as the universal unit of measure, regardless of the technology used. However, because of the potential to generate data with a high degree of variability when different technologies are used, the American Society for the Testing of Materials (ASTM) has revised its turbidity measurement methods to incorporate a unit reporting protocol that provides traceability to the type of technology used. ASTM D was the first turbidity method to incorporate these changes. For examples, when a 90-degree detection angle is used, the letter N for nephelometric should be used and will be either the first or second letter of the reporting unit. When an attenuated detection angle is utilized, the reporting unit will contain an A for 3
4 attenuated and will usually be the second letter of the reporting unit. (See Table 1) The push by ASTM and other regulatory bodies to have units indicate the type of technology used is a relatively recent development to address reporting discrepancies between different technologies. Stakeholders who are aware of these conventions can now avoid errors such as directly comparing readings measured by a nephelomentric instrument reported in NTU with instruments applying absorption or transmission optical designs. This is merely an example of one comparison that can lead to errors but, essentially, any combination of misaligned technologies will lead to the same problem. Table 1. Instrument Design Nephelometric non-ratio turbidimeters Ratio White Light turbidimeters Nephelometric, near-ir turbidimeters, non-ratiometric Nephelometric near-ir turbidimeters, ratio metric Surface Scatter turbidimeters Formazin Back Scatter Unit Backscatter Unit Formazin attenuation Unit Light attenuation Unit Nephelometric Turbidity Multibeam Unit Reporting Unit (NTU) (NTRU) (FNU) (FNRU) (SSU) (FBU) (BU) (FAU) (AU) (NTMU) Additional Sources of Variation Perhaps the most significant practical consideration in turbidity measurement is the difference in measured values among different instruments that have been calibrated with a natural sample as a standard. Differences in the spectral characteristics of the light source/detector combination are the most important reason for different instruments giving different values for the same sample. This is why it is important to use formazin or at least some other laboratory-made substance that has repeatable spectral characteristics as a primary standard. An instrument s capability to measure a wide turbidity range is dependent on three critical design components light source, detector angle, and sensitivity. Another variation source is the depth at which the sensor is placed in a water system. For example, if the sensor is placed in a 10-feet-deep river that is flowing at high velocity, the lightest solids will be near the top of the water column and the heavier sand will be at the bottom. The instrument readings will be different depending on where the monitoring instrument is submerged. Ideally, the sensor should be placed where the water is as homogenous as possible, such as over a waterfall or flume. Best Practices Best practices are essential for reliable data collection and use. When deciding on which technologies to use for a specific application, it is important to adhere to the following guidelines: Select the proper monitoring site. The site should produce a homogenous sample and be readily accessible. Monitoring sites are often put in incredibly remote areas, and if the site goes down for some reason, it may be months before personnel can attend it. The site must be able to be physically accessed during all seasons, or the site should have a wireless system of data transmittal that confirms both successful data acquisition and data quality. Know the environmental limits of the instrument for the specific sample location. It s critical to know the environmental limits of the sensor. For example, if it s to be placed in a stream that will freeze solid in wintertime, it s unwise to put in an instrument not specified for that type of condition. Conversely, some instruments will fail readily in a creek bed where it may reach 120 degrees F and water only runs through it once a month. It s important to pay close attention to the manufacturer s specifications especially operating temperature, sample temperature, and storage temperature. Select the proper instrument based on light source, detection angle, and sensitivity. Modern instruments are required to measure very high and low turbidity levels over an extreme range of sample particulate sizes and composition. An instrument s capability to measure a wide turbidity range is dependent on the instrument s design, with the three critical design components being light source, detector angle (or optical geometry), and path length. Differences in these three components affect an instrument s turbidimetric measurement. Normally, the longer the path length, the better the detection limit or the instruments ability to see very small changes of turbidity at very low turbidity levels. With a very long pathlength, however, the sacrifice is a wide measurement range. Thus, if one wishes to measure to 10,000 units, a technology will likely have to be selected that does not have a good sensitivity at the very low levels of turbidity (typically below 1 unit). Calibrate the instrument according to the manufacturer recommendations. The process of calibrating and verifying calibration of turbidimeters at ultra-low turbidity levels is very sensitive to both user technique and the surrounding environment. As measured turbidity levels drop below 1.0 NTU, the interferences caused by bubbles and particulate contamination (usually only slightly problematic at higher levels), and ambient light can result in a false-positive reading and invalid verification results. To obtain the most accurate calibration for this linear range, Hydrolab turbidimeters use a 20.0 NTU formazin standard. Formazin is the primary standard for turbidity measurements. This concentration is used because the standard is easy to prepare 4
5 accurately from a concentrated stock formazin standard, plus the standard remains stable long enough to maintain its accuracy for calibration. In addition, the standard concentration is in the middle of the linear nephelometric range, and contamination and bubble errors have less effect on the calibration accuracy at 20 NTU than they would have on a lower calibration standard. Formazin is the only true primary standard for turbidity, but an acceptable and approved alternative includes Hach StablCal. It is the same formazin polymer that has been stabilized with a special matrix. Another marketed standard is a commercially manufactured liquid suspension of Styrene divinylbenzene and referred to as polymer beads (SDVB). The optical characteristics of formazin and polymer beads are very different. Polymer bead standards are rated based a specific technology type and this varies between different technologies. For example, if a user chooses to use SDVB beads in non-nephelometric technologies, the same concentration of beads may deliver a different value on a different technology. With SDVB materials, the majority of their standards are instrument specific, designed for a specific make and model of turbidimeter and cannot be used anywhere else. The merits of a calibration standard for turbidity is that it be interference free. Formazin based standards are the only standards that meet this criteria and this allows them to be the universal calibration for ALL technologies. Accurately tracing the turbidity measurement of an instrument design technology and adhering to best practices will help effectively ensure the qualification and quantification of turbidity measurements in natural water bodies Verify the calibration using same or similar technology. The key to proper verification is to verify calibration with the same or similar technology. Often, users have an instrument with one technology and will verify it using a totally different measurement technology. For example, an instrument is placed in a stream and it reads 20 units. Then, another sample is measured with another instrument and it reads 40 units. If it is not known that the technologies are different, it will often be assumed that one of the instruments is wrong. If a comparative calibration or comparative verification is performed and read on a different instrument, it is necessary to use a technology that best matches. At the very least, match the most important factor, which would be the light source, and the second most important would be the detection angle. Also, because most turbidity measurement technologies are characterized based on formazin, it is the best choice as a calibration standard for both technologies. Maintain the instrument. Maintenance frequency should be based on data value (hourly, daily, weekly, monthly) and monitoring goals. If the monitoring goals are being regulated, for example, those instruments will require much more stringent maintenance and monitoring. Accurately understanding what the needs are of a given site, and how that impacts decisions, oftentimes will dictate maintenance scheduling. Maintenance scheduling can be far more stringent than what the manufacturer suggests, but maintenance activities should never be deferred beyond the manufacturer s recommendations. Conclusion Whether profiling ponds, lakes, or reservoirs, or grab sampling in small streams or rivers, monitoring turbidity provides useful information about the relative clarity of water. Turbidity measurement results can, in turn, serve as a valuable surrogate indicator of biological, organic or inorganic pollution. However, it is now general knowledge that different turbidity measuring technologies can deliver very different results. These differences are related to the type of technology used and how this technology is impacted by the different absorptive and attenuation characteristics that are exhibited by the particles in a given sample. The ability of water resource professionals to accurately trace the turbidity measurement to an instrument design technology and adhering to best practices and proper measurement techniques will help effectively ensure the qualification and quantification of turbidity measurements in natural water bodies. Michael Sadar is Principal Research Scientist, Hach Company, Loveland, Colorado. 1. USEPA. 2003a. Guidance for 2004 Assessment, Listing, and Reporting Requirements Pursuant to Sections 303(d) and 305(b) of the Clean Water Act; TMDL Ziegler, Andrew C. Issues Related To Use Of Turbidity Measurements As A Surrogate For Suspended Sediment Proceedings Of The Federal Interagency Workshop On Turbidity And Other Sediment Surrogates, April 30-May 2, 2002, Reno, Nevada. 3. ASTM D a Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode. 5
Ginger Paige and Nancy Mesner University of Wyoming Utah State University
Effective Water Quality BMP Monitoring Tools Ginger Paige and Nancy Mesner University of Wyoming Utah State University Overview BMP Monitoring Guidance Document for Stream Systems Lessons learned CEAP
LIMNOLOGY, WATER QUALITY
LIMNOLOGY, WATER QUALITY PA RANI ET E R S, AN D c 0 IV D IT I 0 N S AND ECOREGIONS Water Quality Parameters Nutrients are important parameters because phosphorous and nitrogen are major nutrients required
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Part B Integrated Monitoring Design for Comprehensive Assessment and Identification of Impaired Waters Contents
Part B Integrated Monitoring Design for Comprehensive Assessment and Identification of Impaired Waters Contents Chapter 10. Selecting Metrics or Indicators of WQS Attainment... 10-2 Chapter 11. Monitoring
An Introduction to Algae Measurements Using In Vivo Fluorescence
An Introduction to Algae Measurements Using In Vivo Fluorescence Submersible fluorescence sensors enable real-time field estimates of phytoplankton that can be directly correlated to standard laboratory
TurbiDirect TurbiCheck TurbiCheck WL
Turbidity 128 TurbiDirect TurbiCheck TurbiCheck WL www.lovibond.com General Catalogue February 2013 Turbidity measurement The term "turbidity is used to describe the cloudy or milky appearance of liquid
Pamela Birak, Jordan Lake State Park, Chatham County, NC
Pamela Birak, Jordan Lake State Park, Chatham County, NC 3 Lakes, Reservoirs, and Ponds Forty-six states, Puerto Rico, and the District of Columbia (collectively referred to as states in the rest of this
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this
Christine E. Hatch University of Nevada, Reno
Christine E. Hatch University of Nevada, Reno Roadmap What is DTS? How Does it Work? What Can DTS Measure? Applications What is Distributed Temperature Sensing (DTS)? Temperature measurement using only
Total Suspended Solids Total Dissolved Solids Hardness
Total Suspended Solids (TSS) are solids in water that can be trapped by a filter. TSS can include a wide variety of material, such as silt, decaying plant and animal matter, industrial wastes, and sewage.
UVC LEDs for. Environmental Monitoring
UVC LEDs for Environmental Monitoring Environmental monitoring relies extensively on molecular spectroscopy for tracking air quality, water and wastewater quality, and detecting hazardous substances. Advances
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
Electromagnetic Radiation (EMR) and Remote Sensing
Electromagnetic Radiation (EMR) and Remote Sensing 1 Atmosphere Anything missing in between? Electromagnetic Radiation (EMR) is radiated by atomic particles at the source (the Sun), propagates through
Welcome to the Understanding Dissolved Oxygen learning module. This section provides information on the following topics:
Introduction Welcome to the learning module. This section provides information on the following topics: How dissolved oxygen is defined and measured in numbers Why dissolved oxygen is important Natural
WASTEWATER TREATMENT OBJECTIVES
WASTEWATER TREATMENT OBJECTIVES The student will do the following: 1. Define wastewater and list components of wastewater. 2. Describe the function of a wastewater treatment plant. 3. Create a wastewater
Lesson Plan: How Do We Clean Polluted Water?
Lesson Plan: How Do We Clean Polluted Water? Oil Spill Cleanup / Phosphate Cleanup / Groundwater Contamination / Water Treatment Simulation Estimated Time: 2-4 days State Standards taught and addressed
HOW-TO GUIDE FOR STORMWATER MONITORING
White Paper January 2014 by Kevin Simpson HOW-TO GUIDE FOR STORMWATER MONITORING The Dos, Don ts and Must Haves! Stormwater monitoring doesn t have to be a headache for your organization. With proper site,
Lesson 4: What Makes Water Healthy?
Lesson 4: What Makes Water Healthy? Activity: Students make observations and measurements of several water samples. This activity helps students think about different ways to determine water quality. Grade
A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA
A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - [email protected]
Clean Water Services. Ecosystems Services Case Study: Tualatin River, Washington
Viewed broadly, the concept of ecosystem services describes the many resources and services provided by nature. Typically, traditional planning and development practices do not adequately represent the
MARYLAND DEPARTMENT OF THE ENVIRONMENT 1800 Washington Boulevard Baltimore Maryland 21230 (410) 537-3000 1-800-633-6101 http://www.mde.state.md.
MARYLAND DEPARTMENT OF THE ENVIRONMENT 1800 Washington Boulevard Baltimore Maryland 21230 (410) 537-3000 1-800-633-6101 http://www.mde.state.md.us FACT SHEET General Discharge Permit No. 12-SI NPDES Permit
Application Note: Absorbance
Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light
Curt Kerns, M.S., R.P.Bio., C.F.S. WetlandsPacific Corp. 250-722-7117 www.wetlandspacific.com
Curt Kerns, M.S., R.P.Bio., C.F.S. WetlandsPacific Corp. 250-722-7117 www.wetlandspacific.com Presentation Historical perspectives Present perspectives How wetlands function Examples of natural wetlands
Improving Steam Methane Reformer Performance with the ZoloSCAN-SMR
Improving Steam Methane Reformer Performance with the SCAN-SMR The global demand for hydrogen continues to increase as heavier crudes are processed and stricter governmental mandates require reductions
ICC Recommendations for Color Measurement
White Paper #3 Level: Introductory Date: Dec 2004 ICC Recommendations for Color Measurement Introduction In order to prepare a useful device profile based on the International Color Consortium s Specification,
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property
Handbook for Developing Watershed Plans to Restore and Protect Our Waters
This document is one chapter from the EPA Handbook for Developing Watershed Plans to Restore and Protect Our Waters, published in March 2008. The reference number is EPA 841-B-08-002. You can find the
Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry
Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry Models UVS-1, MoniSpec-UV & MoniSpec-UVd Model UVS-1 general information
Overview. What is EMR? Electromagnetic Radiation (EMR) LA502 Special Studies Remote Sensing
LA502 Special Studies Remote Sensing Electromagnetic Radiation (EMR) Dr. Ragab Khalil Department of Landscape Architecture Faculty of Environmental Design King AbdulAziz University Room 103 Overview What
Quantification of Reactive Phosphorus in Lake Mendota Sediments
Quantification of Reactive Phosphorus in Lake Mendota Sediments By: Jeremy, Advisor Dr. Mike Penn Most of the phosphorus (P) that enters lakes from external inputs is attached to soil or sediment. This
Phosphorus. Phosphorus Lake Whatcom Cooperative Management. www.ecy.wa.gov/programs/wq/nonpoint/phosphorus/phosphorusban.html
Phosphorus Phosphorus Brochure Lake Whatcom Cooperative Management Reducing Phosphorus Website Washington State Department of Ecology www.ecy.wa.gov/programs/wq/nonpoint/phosphorus/phosphorusban.html Nutrients
Active Treatment Systems
Categories EC Erosion Control SE Sediment Control TC Tracking Control WE Wind Erosion Control Non-Stormwater NS Management Control Waste Management and WM Materials Pollution Control Legend: Primary Category
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this document
Application Note Noise Frequently Asked Questions
: What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random
Explanation of Reflection Features in Optical Fiber as Sometimes Observed in OTDR Measurement Traces
Explanation of Reflection Features in Optical Fiber as Sometimes Observed in OTDR Measurement Traces WP1281 Issued: November 2015 Supersedes: 2012 Author: Dr. Russell Ellis ISO 9001 Registered Background
Pervious Pavers. By: Rich Lahren. Hebron Brick & Block Supply
Pervious Pavers By: Rich Lahren Hebron Brick & Block Supply Stormwater Management and Control Issues Past emphasis was on flood control Today s emphasis is also on pollution More impermeable areas are
Pond Ecosystem Field Study MOLS
This multi-week lab involves field studies comparing ecosystem-level ecology between 2 freshwater ponds in Marshfield Outdoor Learning Sanctuary. We will be investigating a correlation between weather
Total Dissolved Solids:
Total Dissolved Solids: Going From Tedious to TDS Michael Bechtold, Technical Specialist David Smith, Technical Director Total Dissolved Solids (TDS) What is it and why does it matter? How is it currently
Optical Design Tools for Backlight Displays
Optical Design Tools for Backlight Displays Introduction Backlights are used for compact, portable, electronic devices with flat panel Liquid Crystal Displays (LCDs) that require illumination from behind.
ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL
1 of 8 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.
Steven R. McManus Environmental Product Support Teledyne Isco
Steven R. McManus Environmental Product Support Teledyne Isco Flow Measurement Applications and Technologies. Methods of flow measurement Technologies of flow measurement Flow Rate Applications Channel
Blackbody Radiation References INTRODUCTION
Blackbody Radiation References 1) R.A. Serway, R.J. Beichner: Physics for Scientists and Engineers with Modern Physics, 5 th Edition, Vol. 2, Ch.40, Saunders College Publishing (A Division of Harcourt
Experts Review of Aerobic Treatment Unit Operation and Maintenance. Bruce Lesikar Texas AgriLife Extension Service
Experts Review of Aerobic Treatment Unit Operation and Maintenance Bruce Lesikar Texas AgriLife Extension Service Overview Overview of Aerobic Treatment Units Installing for accessibility to system components
Inventory of Performance Monitoring Tools for Subsurface Monitoring of Radionuclide Contamination
Inventory of Performance Monitoring Tools for Subsurface Monitoring of Radionuclide Contamination H. Keith Moo-Young, Professor, Villanova University Ronald Wilhelm, Senior Scientist, U.S. EPA, Office
Various Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
SOLAR WATER PURIFICATION WITH THE HELP OF CSP TECHNOLOGY
Sci. Revs. Chem. Commun.: 3(2), 2013, 128-132 ISSN 2277-2669 SOLAR WATER PURIFICATION WITH THE HELP OF CSP TECHNOLOGY JINESH S. MACHALE *, PRACHI D. THAKUR, PIYUSH S. LALWANI and GAYATRI M. APTE Department
Lake Monitoring in Wisconsin using Satellite Remote Sensing
Lake Monitoring in Wisconsin using Satellite Remote Sensing D. Gurlin and S. Greb Wisconsin Department of Natural Resources 2015 Wisconsin Lakes Partnership Convention April 23 25, 2105 Holiday Inn Convention
Which of the following can be determined based on this model? The atmosphere is the only reservoir on Earth that can store carbon in any form. A.
Earth s Cycles 1. Models are often used to explain scientific knowledge or experimental results. A model of the carbon cycle is shown below. Which of the following can be determined based on this model?
Cabling & Test Considerations for 10 Gigabit Ethernet LAN
Introduction Current communication data rates in local networks range from 10/100 megabits per second (Mbps) in Ethernet to 1 gigabit per second (Gbps) in fiber distributed data interface (FDDI) and Gigabit
Back to Basics Fundamentals of Polymer Analysis
Back to Basics Fundamentals of Polymer Analysis Using Infrared & Raman Spectroscopy Molecular Spectroscopy in the Polymer Manufacturing Process Process NIR NIR Production Receiving Shipping QC R&D Routine
High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering
High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering Power spectrum development with heterodyne technology advances biotechnology and nanotechnology measurements. M. N. Trainer
CREATING A GREEN FUTURE THROUGH WATER REMEDIATION
CREATING A GREEN FUTURE THROUGH WATER REMEDIATION Cornerstone Principles for Conducting Business Active Treatment Systems, Inc. goes to great lengths to instill in every employee the importance of ethical
GUIDELINES FOR LEACHATE CONTROL
GUIDELINES FOR LEACHATE CONTROL The term leachate refers to liquids that migrate from the waste carrying dissolved or suspended contaminants. Leachate results from precipitation entering the landfill and
How To Plan A Buffer Zone
Backyard Buffers Protecting Habitat and Water Quality What is a buffer? A buffer (also called a riparian buffer area or zone) is the strip of natural vegetation along the bank of a stream, lake or other
Measuring of optical output and attenuation
Measuring of optical output and attenuation THEORY Measuring of optical output is the fundamental part of measuring in optoelectronics. The importance of an optical power meter can be compared to an ammeter
Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC
Common Problems with Online Water Quality Analyzers Greg Macy Director AquatiPro LLC Objectives Diagnosing Colorimetric Analyzers Identifying Common Problems with Sodium Analyzers Troubleshooting ph Analyzers
DEPARTMENT OF ENVIRONMENTAL PROTECTION BUREAU OF WATER QUALITY PROTECTION. Policy for Permitting Surface Water Diversions
DEPARTMENT OF ENVIRONMENTAL PROTECTION BUREAU OF WATER QUALITY PROTECTION DOCUMENT NUMBER: 362-2000-003 TITLE: Policy for Permitting Surface Water Diversions EFFECTIVE DATE: March 1, 1998 AUTHORITY POLICY:
How do you treat water based on water quality from different water sources?
How do you treat water based on water quality from different water sources? Why? Authors: Wendy Lane and Kim Sciarrone Seattle Public Schools; Seattle, WA Water from different sources will contain different
Light Control and Efficacy using Light Guides and Diffusers
Light Control and Efficacy using Light Guides and Diffusers LEDs 2012 Michael Georgalis, LC Marketing Manager, Fusion Optix October 11, 2012 Agenda Introduction What Is Light Control? Improves Application
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July 2001. Reprinted from Photonics Spectra, Laurin Publishing
Reprint (R22) Avoiding Errors in UV Radiation Measurements By Thomas C. Larason July 2001 Reprinted from Photonics Spectra, Laurin Publishing Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1
ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners
ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners Tony Harrison and Bob Latimer ISO 21501- A Standard Methodology to Optical Particle Counter
Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory
FIELD APPLICATION REPORT UV/Vis/NIR Spectroscopy Author: Frank Padera Shelton, CT Contributor: Chris Lynch Shelton, CT Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
Watershed-Wide Modeling for TMDL/MS4 Permit Compliance. Greg Wilson, Barr Engineering
Watershed-Wide Modeling for TMDL/MS4 Permit Compliance Greg Wilson, Barr Engineering Acknowledge City of Golden Valley Eric Eckman Heather Hegi Laura Jester and BCWMC Technical Advisory Committee members
Town of New Castle Utility Department Introduction
Town of New Castle Utility Department Introduction Town of New Castle Utility Department Mission Statement Our commitment is to ensure that our customers receive high quality water and wastewater treatment
Trace Gas Exchange Measurements with Standard Infrared Analyzers
Practical Environmental Measurement Methods Trace Gas Exchange Measurements with Standard Infrared Analyzers Last change of document: February 23, 2007 Supervisor: Charles Robert Room no: S 4381 ph: 4352
Chapter 2 Stormwater Pollution Prevention Plan (SWPPP) for Park Operations
SWPPP for Park Operations 2 Chapter 2 Stormwater Pollution Prevention Plan (SWPPP) for Park Operations Bordered by Lake Washington & Lake Sammamish, the City of Bellevue has more than 60 miles of streams,
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
Infrared Thermometry. Introduction, History, and Applications. Optical Pyrometry. Jason Mershon, Advanced Energy Industries, Inc
Infrared Thermometry Introduction, History, and Applications Jason Mershon, Advanced Energy Industries, Inc In manufacturing environments, measuring the temperature of an object without contact has proven
4. Environmental Impacts Assessment and Remediation Targets
4. Environmental Impacts Assessment and Remediation Targets 4.1 Environmental Impacts Significant additional development in the Alder Creek watershed is not anticipated at this time; however, there are
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Detection of (unusual) changes in wastewater composition using UV/VIS spectroscopy
Detection of (unusual) changes in wastewater composition using UV/VIS spectroscopy G. Langergraber 1 *, N. Fleischmann 2, F. Hofstaedter 3, A. Weingartner 3 and W. Lettl 3 1 Department for Sanitary Engineering
Department of Environmental Engineering
Department of Environmental Engineering Master of Engineering Program in Environmental Engineering (International Program) M.Eng. (Environmental Engineering) Plan A Option 1: (1) Major courses: minimum
About Coffee and Refractometers 2008-2010 Voice Systems Technology, Inc. (VST)
About Coffee and Refractometers 2008-200 Voice Systems Technology, Inc. (VST) www.mojotogo.us.0 Coffee and Refractive Index Refractive Index measurements have been used for process control in the food
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
Principle of Thermal Imaging
Section 8 All materials, which are above 0 degrees Kelvin (-273 degrees C), emit infrared energy. The infrared energy emitted from the measured object is converted into an electrical signal by the imaging
Scheduling Maintenance for Infiltration Basins and Trenches
Visual Inspection for Infiltration Practices Visual inspection is a rapid assessment procedure for qualitatively evaluating the functionality of a stormwater best management practice (BMP). Visual inspections
1-4 kg/m3. Long in-line calibration cycles of the gamma density systems may improve measurement accuracy, but this is often not practical subsea.
Generating Greater Accuracy and Robustness from Subsea Multiphase Meters By Finn Erik Berge, Emerson Process Management Subsea multiphase meters have faced a growing number of challenges linked to the
IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS
IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS 1998 A.P. Malcolmson Malvern, Inc. 10 Southville Road Southborough, MA D. J. Holve Malvern/Insitec, Inc. 2110 Omega
Monitoring Water Quality For Today s Demanding World
Monitoring Water Quality For Today s Demanding World Visualize Analyze Optimize EMERSON PROCESS MANAGEMENT BRINGING WATER QUALITY MONITORING TO A WHOLE NEW LEVEL WHAT MATTERS MOST: MAKE IT SAFE Like never
Particle Monitoring Requirements in Pharmaceutical Cleanrooms
Particle Monitoring Requirements in Pharmaceutical Cleanrooms All drugs must be manufactured in accordance with the current Good Manufacturing Practice (cgmp) regulations. Pharmaceutical manufacturers
Atomic Force Microscopy Observation and Characterization of a CD Stamper, Lycopodium Spores, and Step-Height Standard Diffraction Grating
Atomic Force Microscopy Observation and Characterization of a CD Stamper, Lycopodium Spores, and Step-Height Standard Diffraction Grating Michael McMearty and Frit Miot Special Thanks to Brendan Cross
Don t spit in the well - you may need to drink from it! -- Unknown, Russia.
Don t spit in the well - you may need to drink from it! -- Unknown, Russia. WATER POLLUTION Water is a unique substance, because it can naturally renew and cleanse itself, by allowing pollutants to settle
Sand and Silt Removal from Salmonid Streams
Sand and Silt Removal from Salmonid Streams Stream bank erosion Poor land use practices Sources of Sand and Silt Impacts of Sand and Silt Interstitial spaces infilled Little or no flow through the streambed
Alignment Laser System.
- O T 6 0 0 0 Alignment Laser System. The OT-6000. Multi -Target,Two Dimensional Alignment. Introducing the most powerful way to measure alignment at distances up to 300 feet. The OT-6000 Alignment Laser
Water Water Treatment Plant Tour
Water Water Treatment Plant Tour Don Rankin, Superintendent Topeka, KS 66606 785-368-3111 Tel 785-368-3825 FAX For a complete list of contacts with phone numbers, FAX numbers and email addresses, click
Treasure Hunt. Lecture 2 How does Light Interact with the Environment? EMR Principles and Properties. EMR and Remote Sensing
Lecture 2 How does Light Interact with the Environment? Treasure Hunt Find and scan all 11 QR codes Choose one to watch / read in detail Post the key points as a reaction to http://www.scoop.it/t/env202-502-w2
Molecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
RESULTS FROM A SIMPLE INFRARED CLOUD DETECTOR
RESULTS FROM A SIMPLE INFRARED CLOUD DETECTOR A. Maghrabi 1 and R. Clay 2 1 Institute of Astronomical and Geophysical Research, King Abdulaziz City For Science and Technology, P.O. Box 6086 Riyadh 11442,