Application Note FTMS-55 Laser/Desorption Ionization FT-ICR Mass Spectrometry as a Tool for Statistical Analysis of Crude Oils
|
|
- Virgil Pitts
- 7 years ago
- Views:
Transcription
1 Application Note FTMS-55 Laser/Desorption Ionization FT-ICR Mass Spectrometry as a Tool for Statistical Analysis of Crude Oils Introduction Crude oil is an extremely complex mixture of hydrocarbons and heteroatom-containing chemicals such as thiophenes, phenols, pyrrols, and pyridines. The composition can vary widely, depending on the origin of the oil and how the petroleum was formed. The maturity of each crude oil has a huge influence on its molecular composition. Each raw petroleum or crude oil has a characteristic composition and properties. Therefore, chemical analysis can be used to fingerprint the source of the petroleum. Highboiling point hydrocarbons and highly polar compounds cannot be characterized by GC/MS analysis. However, these compounds can be analyzed at the molecular level by Fourier Transform Ion Cyclotron Resonance (FT-ICR) mass spectrometry, which is a well-established method in petroleomics [1-3]. On the basis of the detected compounds, the relative abundance of different compound classes can be calculated and compared between different crude oils. The chemical composition of crude oil can also be correlated with the chemical properties of the oil. Such analyses complement characterization using standard crude oil parameters such as API gravity, TAN acidity, sulfur and nitrogen content. Authors Matthias Witt Bruker Daltonik GmbH, Bremen, Germany Kolbjørn Zahlsen, Anders Brunsvik SINTEF Materials and Chemistry, Trondheim, Norway Keywords Instrumentation and Software Petroleomics solarix XR, 12 T FTMS ProfileAnalysis 2.1 LDI MTP II MALDI source Composer 1.0.6
2 It should be noted that the abundance of the detected compound classes depends heavily on the ionization method used. Therefore, crude oils can be compared by FT-ICR mass spectrometry, but this method is only semiquantitative [4]. The types of compound classes that are detected are also influenced by the presence of salts and additives used to provide better ionization efficiencies. In this study, laser desorption/ionization (LDI) was coupled to a Fourier Transform Ion Cyclotron Resonance (FT-ICR) mass spectrometer for the analysis of crude oil [5]. A thin film of crude oil was applied onto a stainless steel target and analyzed by LDI using a UV laser at 355 nm. No matrix was added to the sample for these measurements. Nine different crude oils were analyzed and compared at the molecular level. The relative abundances of compound classes with different numbers of heteroatoms were compared. In some cases, the relative abundances of the compounds classes were quite low. Therefore, high reproducibility of the LDI measurements is a prerequisite for reliable characterization of crude oils. Each crude oil sample was measured several times to calculate the standard deviation of the detected compound classes. Additionally, the results were analyzed using the statistical method principle component analysis (PCA). The results showed that the oils were well-separated in the PCA scoring plot, with low variation between the replicate measurements. The statistical analysis also indicated that extremely high resolution mass spectra are required to distinguish between oils. Therefore, LDI coupled to FT-ICR mass spectrometry is a perfect tool for distinguishing between crude oils using statistical methods such as PCA. In addition, LDI is independent of solvent effects observed in atmospheric pressure ionization methods such as ESI or APPI, and sample preparation is extremely simple. Experimental Sample preparation: Samples were kindly provided by SINTEF, Department of Materials and Chemistry, Trondheim, Norway. The crude oil samples were measured by LDI without the addition of matrix and without any purification. Crude oil (10 mg) was dissolved in 990 μl toluene and 1 μl of this solution was spotted onto a stainless steel target and dried for 15 min at room temperature. The samples were placed in the source chamber for 30 minutes at around 3 mbar before mass analysis by laser/desorption ionization. Mass analysis: Mass spectra were acquired using a Bruker solarix XR Fourier transform ion cyclotron resonance mass spectrometer (Bruker Daltonik GmbH, Bremen, Germany) equipped with a 12 T refrigerated actively-shielded superconducting magnet (Bruker Biospin, Wissembourg, France) and the new dynamically harmonized analyzer cell (ParaCell ). The samples were ionized in positive ion mode using the MALDI ion source (Bruker Daltonik GmbH, Bremen, Germany). The mass range was set to m/z using 4M data points with a transient length of 1.6 s resulting in a resolving power of 450,000 at m/z 400 in magnitude mode. One hundred single scans were averaged for a final mass spectrum. The laser power was lowered to 12% to reduce fragmentation during the ionization process. The laser focus was set to a diameter of 30 μm and 4 laser shots were used for each scan. Five replicate measurements were performed for each sample for meaningful statistical analysis. Spectra were zero-filled to processing size of 8M data points before sine apodization. Mass calibration: The mass spectra were calibrated externally using arginine clusters in electrospray ionization positive ion mode. A linear calibration was applied. A 10 μg ml -1 solution of arginine in 50% methanol was used to generate the clusters. To improve the mass accuracy, the crude oil spectra were recalibrated internally with the N 1 series. The RMS mass error of the recalibration of all acquired spectra was better than 150 ppb using a linear calibration. Molecular formula calculation: The molecular formula calculation was performed in Composer (Sierra Analytics, Modesto, CA, USA) using a maximum formula of C n H h N 3 O 3 S 3. Due to the formation and detection of radical cations and protonated species, the electron configuration was set to odd and even, and a mass tolerance of 500 ppb was used. The relative abundances of the compound classes were calculated with the Composer software. The statistical analysis was performed with the software ProfileAnalysis 2.1 (Bruker Daltonik GmbH, Bremen, Germany). Results and Discussion The mass spectra of three different crude oil samples measured in LDI positive ion mode are shown in Figure 1. The mass distribution of the samples is very similar, with an average mass of around m/z 300. Masses of up to m/z 800 were detected. The average number of carbons in compounds detected by LDI can be estimated by mass distribution. Therefore, LDI-TOF-MS can be used as a general tool to measure the average mass and the mass distribution analysis of crude oil, lubricants, bitumen, and asphaltenes. However, ultrahigh-resolution mass spectrometry is required to analyze these samples at the molecular level. Only ultra-highresolution FT-ICR is capable of resolving all mass peaks in the extremely complex mass spectrum, thereby enabling calculation the correct molecular formula of each peak based on highly accurate mass measurements (Figure 2).
3 Mass spectra of three different crude oil samples Figure 1: Mass spectra of three different crude oil samples measured in LDI positive ion mode. Magnified view of crude oil mass spectra Figure 2: Magnified view of mass spectra of crude oils 2, 3, and 4 at m/z 301. Significant differences are only apparent using ultrahigh resolution.
4 The molecular formulas of mass peaks were calculated with a mass tolerance of 500 ppb. Using a maximum formula of C n H h N 3 O 3 S 3, single-candidate molecular formulas were obtained in the mass range up to m/z 500. Higher molecular formulas were obtained by extending the homologous series obtained in the lower mass region. This calculation was performed using the Composer software. The mass spectra showed significant differences in ion abundance (Figure 2) resulting in different relative abundances of compound classes (Figure 3). Oil 8 and oil 8a are identical oils from different batches. Oil 9a corresponds to a sample of oil 9 that was stored in the source vacuum chamber for one day at 3 mbar, which resulted in the evaporation of volatile compounds. Distinguishing between crude oils on the basis of compound class abundances alone only provides approximate and sometimes even inaccurate results. Therefore, our approach for distinguishing between crude oils was performed at the molecular level using all mass peaks detected by LDI-FT-ICR mass spectrometry. The mass spectra of the different oil samples were first analyzed on the basis of relative abundance of compound classes (Figure 3) for an approximate characterization of the crude oils. For a more precise characterization, the mass spectra of the crude oils were analyzed using the statistical method principle component analysis (PCA). To obtain statistically meaningful results, five replicate measurements were performed on each sample. The results of the PCA calculated using the average mass list show that the oils are well-separated in the PCA scoring plot, with low variation between the replicate measurements (Figure 4). Relative abundances of compound classes in crude oil samples Figure 3: Three-dimensional column diagram showing relative abundances of compound classes detected as protonated species and radical cations by LDI-FT-ICR mass spectrometry. Class of protonated species is marked as [H]. PCA scoring plots of LDI-FT-ICR mass spectra using average Figure 4: PCA scoring plots of LDI-FT-ICR mass spectra of all oil samples using average mass lists with a) logarithmic scaling and b) linear scaling. Oil 8 and oil 8a are identical oils from different batches. Oil 9a corresponds to a sample of oil 9 that was stored in the source vacuum chamber for one day at 3 mbar.
5 Improved separation of the oils in the PCA scoring plot was observed using logarithmic scaling (Figure 4a) instead of linear scaling (Figure 4b). However, statistical analysis indicated that extremely high resolution mass spectra were required for efficient separation of the oils by PCA (Figure 5). Using bucketing with a mass resolution of 2 mda for PCA calculation instead of average mass lists resulted in inefficient separation of crude oils (Figure 5a). No separation of the crude oils was observed with low mass resolution of 20 mda bucketing (Figure 5b). A bucketing mass resolution of 20 mda corresponds to a mass spectrum with resolving power of only 20,000, as typically obtained using a MALDI-TOF instrument. Therefore, the ultra-high resolution provided by FT-ICR MS is required to distinguish between the crude oils using LDI and PCA. The high reproducibility of LDI mass spectra was demonstrated by the standard deviation of the relative abundances of the compound classes (Table 1). Low relative standard deviations of less than 4% were observed for highly abundant compound classes such as hydrocarbons or N 1. Relatively low standard deviations (< 10%) were observed for very low-abundance classes. Therefore, LDI-FT-ICR mass measurements are highly reproducible, a factor that is crucial for the efficient characterization of crude oil using this technique. PCA scoring plots of LDI-FT-ICR mass spectra using mass bucketing Figure 5: PCA scoring plots of LDI-FT-ICR mass spectra of all oil samples using logarithmic scaling and a) 2 mda or b) 20 mda mass bucketing. Oil 8 and oil 8a are identical oils from different batches. Oil 9a corresponds to a sample of oil 9 that was stored in the source vacuum chamber for one day at 3 mbar. Table 1: Relative abundances of compound classes found in five replicate measurements of oil 8. LDI-FT-ICR mass spectra Class Oil 8 rep1 Oil 8 rep2 Oil 8 rep3 Oil 8 rep4 Oil 8 rep5 Std. deviation [%] HC HC [H] N N [H] N N 2 [H] NO NO [H] O O [H] O O 2 [H] OS [H] S
6 Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. Bruker Daltonics , FTMS-55, Conclusion Crude oil samples can be distinguished at the molecular level by LDI ultra-high resolution mass spectrometry. LDI- FT-ICR mass spectrometry is capable of classifying crude oils by principle component analysis (PCA). LDI-FT-ICR measurements provide the reproducibility required for efficient PCA characterization. Crude oils are best separated in PCA scoring plots using average mass lists and logarithmic scaling. References [1] Rodgers, R. P.; Schaub, T. M.; Marshall, A. G. Anal. Chem. 2005, 21A. [2] Marshall, A. G.; Rodgers, R. P. Acc. Chem. Res. 2004, 37, [3] Cho, Y.; Ahmed, A.; Islam, A., Kim, S. Mass Spectrom. Rev. 2014, doi: /mas [4] Witt, M. Comparison of LDI and APPI for correlation studies between physical and chemical properties of crude oil and FT-ICR mass spectra, Poster at Petrophase [5] Cho, Y.; Witt M.; Kim, Y. H.; Kim S. Energy & Fuels 2012, 84, For research use only. Not for use in diagnostic procedures. Bruker Daltonik GmbH Bremen Germany Phone +49 (0) Fax +49 (0) Bruker Daltonics Inc. Billerica, MA USA Phone +1 (978) Fax +1 (978) Fremont, CA USA Phone +1 (510) Fax +1 (510)
Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method
Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Introduction During the last decade, the complexity of samples
More informationamazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS
amazon SL Setting New Standards in Performance, Simplicity and Value Innovation with Integrity Ion Trap Best-In-Class Ion Trap Mass Spectrometer for Routine Analysis The amazon SL entry-level system is
More informationKendrick Mass Defect Spectrum: A Compact Visual Analysis for Ultrahigh-Resolution Broadband Mass Spectra
JSB is an authorised partner of Kendrick Mass Defect Spectrum: A Compact Visual Analysis for Ultrahigh-Resolution Broadband Mass Spectra Christine A. Hughey, Christopher L. Hendrickson, Ryan P. Rodgers,
More information13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
More information# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationApplication Note # MS-14 Fast On-site Identification of Drugs with the mobile GC/MS system E²M
Bruker Daltonics Application Note # MS- Fast n-site Identification of Drugs with the mobile GC/MS system E²M For the detection of drugs a couple of quick tests are available which were used by the police
More informationAppendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
More informationApplication Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software
Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software Abstract Presented here is a case study of a walk-up liquid chromatography
More informationAMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
More informationElectrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
More informationAccurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis
Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** mariadellarcopassaro@gmail.com
More informationWeight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System
Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Carlton G. Slough TA Instruments, 109 Lukens Drive, New Castle DE 19720, USA ABSTRACT The use of
More informationApp. Note # FTMS-47+MT-113 MALDI Imaging with Single Cell Resolution at 10 µm Pixel Size
pp. Note # FTMS-47+MT-113 MLDI Imaging with Single Cell Resolution at 10 µm Pixel Size bstract Key parameters for MLDI imaging are spatial resolution and mass spectral quality. In this application note,
More informationApplication Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008
Bruker Daltonics Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008 In the ABRF-PRG study 2008 [*] the ability to characterize
More informationOverview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
More informationEleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
More informationProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry
ProteinScape Proteomics Data Analysis & Management Innovation with Integrity Mass Spectrometry ProteinScape a Virtual Environment for Successful Proteomics To overcome the growing complexity of proteomics
More informationSimultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
More information[ Care and Use Manual ]
PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample
More informationWaters Core Chromatography Training (2 Days)
2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.
More informationL. Yu et al. Correspondence to: Q. Zhang (dkwzhang@ucdavis.edu)
Supplement of Atmos. Chem. Phys. Discuss., 5, 967 974, 5 http://www.atmos-chem-phys-discuss.net/5/967/5/ doi:.594/acpd-5-967-5-supplement Author(s) 5. CC Attribution. License. Supplement of Molecular transformations
More informationBruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
Bruker ToxScreener TM A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS ToxScreener - Get the Complete Picture Forensic laboratories are frequently required
More informationAppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT
AppNote 6/2003 Analysis of Flavors using a Mass Spectral Based Chemical Sensor Vanessa R. Kinton Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA Kevin L. Goodner USDA, Citrus & Subtropical
More informationGC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
More informationTECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature
ProteoMass Peptide & Protein MALDI-MS Calibration Kit Catalog Number MSCAL1 Store at Room Temperature TECHNICAL BULLETIN Description This kit provides a range of standard peptides and proteins for the
More informationSupplementary information
Supplementary information What is under the hump? Mass Spectrometry based analysis of complex mixtures in processed food Lessons from the characterisation of black tea thearubigins, coffee melanoidines
More informationThermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software
Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,
More informationElectrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)
Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III) Purpose: Verify the synthesis of the synthesis product, Cr(acac) 3. The presence of any urea containing byproducts will also be determined.
More informationCorrelation of the Mass Spectrometric Analysis of Heat-Treated Glutaraldehyde Preparations to Their 235nm / 280 nm UV Absorbance Ratio
an ABC Laboratories white paper Correlation of the Mass Spectrometric Analysis of Heat-Treated Glutaraldehyde Preparations to Their 235nm / 280 nm UV Absorbance Ratio A. Sen, R. Dunphy, L. Rosik Analytical
More informationAnalyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
More information0 10 20 30 40 50 60 70 m/z
Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationMass Spectrometry. Overview
Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2
More informationApplication Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow Abstract There is high demand in clinical research and forensic toxicology for comprehensive,
More informationTechnical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
More informationIncreasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening
Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening Application Note Forensics Author Erin Shonsey Director of Research Alabama Department of Forensic Science Abstract
More informationAiping Lu. Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn
Aiping Lu Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn Proteome and Proteomics PROTEin complement expressed by genome Marc Wilkins Electrophoresis. 1995. 16(7):1090-4. proteomics
More informationMarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis
MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis Overview MarkerView software is a novel program designed for metabolomics applications and biomarker profiling workflows 1. Using
More informationNMR Chemical Shifts of Common Laboratory Solvents as Trace Impurities
7512 J. Org. Chem. 1997, 62, 7512-7515 NMR Chemical Shifts of Common Laboratory Solvents as Trace Impurities Hugo E. Gottlieb,* Vadim Kotlyar, and Abraham Nudelman* Department of Chemistry, Bar-Ilan University,
More informationPosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
More informationLaboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik
Laboration 1 Identifiering av proteiner med Mass Spektrometri Klinisk Kemisk Diagnostik Sven Kjellström 2014 kjellstrom.sven@gmail.com 0702-935060 Laboration 1 Klinisk Kemisk Diagnostik Identifiering av
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationApplication Note # ET-29 Screening for novel natural products from myxobacteria using LC-MS and LC-NMR
Application Note # ET-29 Screening for novel natural products from myxobacteria using LC-MS and LC-NMR Introduction Because known pathogens become increasingly resistant to antibiotics, and at the same
More informationF321 THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important are... in the nucleus of an atom
Atomic Structure F32 TE STRUCTURE OF ATOMS ATOMS Atoms consist of a number of fundamental particles, the most important are... Mass / kg Charge / C Relative mass Relative Charge PROTON NEUTRON ELECTRON
More informationTutorial for Proteomics Data Submission. Katalin F. Medzihradszky Robert J. Chalkley UCSF
Tutorial for Proteomics Data Submission Katalin F. Medzihradszky Robert J. Chalkley UCSF Why Have Guidelines? Large-scale proteomics studies create huge amounts of data. It is impossible/impractical to
More informationGENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
More informationEnhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference
Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas
More informationBridging the analytical gap
Bridging the analytical gap Thermal analysis provides perfect tools for the characterization of all kinds of organic and inorganic solids or liquids. Thermodynamic transitions, thermal stability, decomposition
More informationElement of same atomic number, but different atomic mass o Example: Hydrogen
Atomic mass: p + = protons; e - = electrons; n 0 = neutrons p + + n 0 = atomic mass o For carbon-12, 6p + + 6n 0 = atomic mass of 12.0 o For chlorine-35, 17p + + 18n 0 = atomic mass of 35.0 atomic mass
More informationUltra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
More informationAnalysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
More informationFunctional Data Analysis of MALDI TOF Protein Spectra
Functional Data Analysis of MALDI TOF Protein Spectra Dean Billheimer dean.billheimer@vanderbilt.edu. Department of Biostatistics Vanderbilt University Vanderbilt Ingram Cancer Center FDA for MALDI TOF
More informationUsing Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
More informationGuide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
More informationSimultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
More informationaurora ICP-MS Innovation with Integrity The Science of Sensitivity ICP-MS
aurora ICP-MS The Science of Sensitivity Innovation with Integrity ICP-MS Powers the Business of Quantification aurora ICP-MS enhanced value package Bruker s ICP-MS is famous for its innovative and unique
More informationUHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
More informationMaster course KEMM03 Principles of Mass Spectrometric Protein Characterization. Exam
Exam Master course KEMM03 Principles of Mass Spectrometric Protein Characterization 2010-10-29 kl 08.15-13.00 Use a new paper for answering each question! Write your name on each paper! Aids: Mini calculator,
More informationThe Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
More informationAB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs
AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF
More informationAnalysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
More informationFast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
More informationNoise reduction of fast, repetitive GC/MS measurements using principal component analysis (PCA)
Analytica Chimica Acta 401 (1999) 35 43 Noise reduction of fast, repetitive GC/MS measurements using principal component analysis (PCA) M. Statheropoulos a,, A. Pappa a, P. Karamertzanis a, H.L.C. Meuzelaar
More informationIntelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
More informationMetabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants
Metabolomic Profiling of Accurate Mass LC- Data to Identify Unexpected Environmental Pollutants André Schreiber 1, David Cox 1, Nadia Pace 1, Christopher Borton 2 1 AB SCIEX, Concord, ntario, Canada; 2
More informationLC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
More informationSUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
More informationRAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH Cristiana C. Leandro 1, Peter Hancock 1, Christian Soulier 2, Françoise Aime 2 1 Waters Corporation, Manchester,
More informationUltraviolet Spectroscopy
Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer
More informationExtraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in uman Serum Technical verview Clinical Research Authors Anthony Macherone, Ph.D. Agilent Technologies,
More informationInfrared Spectroscopy 紅 外 線 光 譜 儀
Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed
More informationELECTRONIC SUPPLEMENTARY INFORMATION
ELECTRONIC SUPPLEMENTARY INFORMATION General.. 1 1 HNMR titration fitplots.. 2 1 HNMR titration Job plots... 2 1 HNMR spectra of 1 + TBAacetate. 3 Isothermal Titration Calorimetry.. 3 High Resolution Mass
More informationAlkoxycarbonylation of Ethylene with Cellulose in Ionic Liquids
Supplementary data Alkoxycarbonylation of Ethylene with Cellulose in Ionic Liquids Anna sichow and Stefan Mecking* Chair of Chemical Materials Science, Dept. of Chemistry, University of Konstanz, 78464
More informationDiscovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
More informationIn-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics
More information1. Details of the ultrahigh-resolution ESI FT-ICR MS analysis of WSOC
Supplement of Atmos. Chem. Phys., 15, 5047 5068, 2015 http://www.atmos-chem-phys.net/15/5047/2015/ doi:10.5194/acp-15-5047-2015-supplement Author(s) 2015. CC Attribution 3.0 License. Supplement of Molecular
More informationFast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD
Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai) Co., Ltd. 412 Ying Lun
More informationModule 3: Liquid Fossil Fuel (Petroleum) Lecture 17: Evaluation of crude
1 P age Module 3: Liquid Fossil Fuel (Petroleum) Lecture 17: Evaluation of crude 2 P age Keywords: Evaluation, characterization factor, TBP, ASTM, EFV 3.2 Evaluation of crude The assessment of a crude
More informationDetermination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS
Application Note No. 051 Determination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS by P Jüngel ¹), J Wittenzellner ²) and E Melcher ¹) ¹) Federal Research Centre
More informationProteinChip Energy Absorbing Molecules (EAM)
ProteinChip Energy Absorbing Molecules (EAM) Instruction Manual Catalog #C30-00001, #C30-00002, #C30-00003, #C30-00004 For technical support, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD
More informationPhysical & Chemical Properties. Properties
Physical & Chemical Properties Properties Carbon black can be broadly defined as very fine particulate aggregates of carbon possessing an amorphous quasi-graphitic molecular structure. The most significant
More informationApplication of a New Immobilization H/D Exchange Protocol: A Calmodulin Study
Application of a New Immobilization H/D Exchange Protocol: A Calmodulin Study Jiang Zhao; Mei Zhu; Daryl E. Gilblin; Michael L. Gross Washington University Center for Biomrdical and Bioorganic Mass Spectrometry:
More informationProton Nuclear Magnetic Resonance Spectroscopy
CHEM 334L Organic Chemistry Laboratory Revision 2.0 Proton Nuclear Magnetic Resonance Spectroscopy In this laboratory exercise we will learn how to use the Chemistry Department's Nuclear Magnetic Resonance
More informationSTATE UNIVERSITY OF NEW YORK COLLEGE OF TECHNOLOGY CANTON, NEW YORK COURSE OUTLINE CHEM 150 - COLLEGE CHEMISTRY I
STATE UNIVERSITY OF NEW YORK COLLEGE OF TECHNOLOGY CANTON, NEW YORK COURSE OUTLINE CHEM 150 - COLLEGE CHEMISTRY I PREPARED BY: NICOLE HELDT SCHOOL OF SCIENCE, HEALTH, AND PROFESSIONAL STUDIES SCIENCE DEPARTMENT
More informationHigh Resolution LC-MS Data Output and Analysis
High Resolution LC-MS Data Output and Analysis Your Sample LC/MS Result Sample Sample Efficient Separation Preparation Introduction Gradient (column) Today Data (computer) LC- MS Ions Detection Ions Separation
More informationA Ratiometric NMR ph Sensing Strategy Based on Slow- Proton-Exchange (SPE) Mechanism
Electronic Supplementary Material (ESI) for Chemical Science. This journal is The Royal Society of Chemistry 2015 A Ratiometric NMR ph Sensing Strategy Based on Slow- Proton-Exchange (SPE) Mechanism Loïse
More informationEffects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests AB SCIEX TOF/TOF 5800 System with DynamicExit Algorithm and ProteinPilot Software for Robust Protein
More informationNuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass
Protons, neutrons and electrons Nuclear Structure particle relative charge relative mass proton 1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons and neutrons make up
More informationCH3 Stoichiometry. The violent chemical reaction of bromine and phosphorus. P.76
CH3 Stoichiometry The violent chemical reaction of bromine and phosphorus. P.76 Contents 3.1 Counting by Weighing 3.2 Atomic Masses 3.3 The Mole 3.4 Molar Mass 3.5 Percent Composition of Compounds 3.6
More informationSignal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
More informationA Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas
A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas Paper #46 Heidi C. Hayes Air Toxics, Ltd., 180 Blue Ravine Rd. Ste. B, Folsom, CA 95630 Diane J.
More informationChemical Structure of the polymer showing end groups: CH 3 CH 2 CH H
Polymer Reference Materials-Polystyrene Introduction Reference materials are used for calibration and performance evaluation of instruments used as part of overall quality assurance programs of polymer.
More informationApplication of Structure-Based LC/MS Database Management for Forensic Analysis
Cozette M. Cuppett and Michael P. Balogh Waters Corporation, Milford, MA USA Antony Williams, Vitaly Lashin and Ilya Troisky Advanced Chemistry Development, Toronto, ntario, Canada verview Application
More informationNMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum.
NMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum. Dr. John Jackowski Chair of Science, Head of Chemistry Scotch College Melbourne john.jackowski@scotch.vic.edu.au
More informationSupporting Information
Supporting Information Simple and Rapid Synthesis of Ultrathin Gold Nanowires, Their Self-Assembly and Application in Surface-Enhanced Raman Scattering Huajun Feng, a Yanmei Yang, a Yumeng You, b Gongping
More informationCHE334 Identification of an Unknown Compound By NMR/IR/MS
CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared
More informationElemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
More informationGetting the most from this book...4 About this book...5
Contents Getting the most from this book...4 About this book....5 Content Guidance Topic 1 Atomic structure and the periodic table...8 Topic 2 Bonding and structure...14 Topic 2A Bonding....14 Topic 2B
More information