Improving digital image quality for larger patient sizes without compromises
|
|
|
- Myrtle Ward
- 9 years ago
- Views:
Transcription
1 Improving digital image quality for larger patient sizes without compromises Christiaan Kok, Technology Physicist Grids Smit Röntgen, Best, The Netherlands fiber grids: 25 lp/cm with variable height versus standard 36R12 grid SPR = 6 SPR = 20 36R12; SPR = 6 36R12; SPR = 20 SNR % % h (mm) In recent years the average patient has increased in weight. It is well-known that an X-ray image of an overweighed or obese patient has a lower image quality than an X-ray image of a standard patient. Due to the fact that the X-rays have to transverse more body mass, the radiologist has to choose a higher kv setting to obtain an image whilst maintaining a reasonable dose and tube load. At higher kv the penetrating power of X- rays increases, but the contrast between different parts of the body decreases. Also the amount of scatter increases very rapidly with increasing body mass, which will further decrease contrast. For digital imaging systems, the DQE (detective quantum effi ciency) is the most important image quality determining factor. The DQE is related to the Signal to Noise Ratio (SNR). If scatter is present in an image, the DQE is reduced. An improved anti-scatter grid is introduced with better scatter reduction. This new grid improves the imaging system s DQE for high scatter conditions. Therefore the image quality of X-ray images of overweighed patients is signifi cantly improved. The performance for standard patients is not compromised. White paper, February 2008
2
3 Introduction to X-ray imaging systems Radiography is the general imaging of body parts by X- ray radiation. A schematic of the imaging setup is given in Figure 1. produced. In many applications a thin metal sheet (spectral fi lter) is put in the X-ray beam to preferentially attenuate the low-energy X-rays which are incapable of reaching the X-ray detector but may harm the patient. A B The symbol B denotes the next important component of an imaging system: the collimator. A collimator is used to limit the irradiated area on the patient. Since X-ray radiation is detrimental for humans it is necessary to limit the X-ray exposure to only the part of the patient which is of interest. The letter C denotes the anti-scatter grid, or grid, which will be dealt with in much detail in a later section. C D Figure 1 Schematic of imaging setup (meaning of symbols denoted in main text) A short description of the main components, denoted by the letters A-D in Figure 1, is now given. In Figure 1, the symbol A denotes the X-ray source, also called the tube. In the tube a beam of electrons is accelerated to high potential ( kv, denoted as kvp setting) and impinges on a metal disc, the anode. Under this violent bombardment with electrons the metal will emit X-ray radiation, although much of the energy of the electrons is dissipated as heat. The amount of heat which the anode can withstand limits both the current of the electrons and the total amount of X-ray which can be generated. The amount of X-rays is related to the notion of mas which is the product of the electron current in milliamperes and the time the electron beam is running in seconds. The physics of the X-ray generation process dictates that a broad X-ray spectrum is obtained, meaning that X-rays of many different energies are Finally, the letter D denotes the detector, which in most modern systems is a digital fl at detector. Most fl at detectors have a scintillator material on top of the detector plate which converts the X-rays into visible light. The visible light is detected by the detector plate, which is subdivided into a square pattern of pixels. These pixels each give an output signal proportional to the measured intensity and from these signals a digital image is built [1]. The many peripherals, which are connected to modern imaging systems, are not discussed here, since they have minor relevance for the concepts discussed in this document. A more thorough description of an X-ray system can be found in [2]. White paper, February
4 Scatter and noise in X-ray imaging Scatter is defi ned as X-ray radiation not emanating from the X-ray source. Most scatter is produced in the patient by a number of physical processes which all produce secondary X-ray radiation propagating in random directions and with energies lower or equal to the primary radiation. Detailed descriptions of the physical processes involved are described in the literature. To demonstrate the effect of scatter on image quality a simplifi ed model, called a phantom, is depicted in Figure 2. The symbols are familiar from Figure 1, only the grid (C) has been removed for simplicity and the patient is replaced by a phantom (F) comprising a rectangular slab of material with a sphere shaped feature in dark blue inside. The plastic PMMA is often used as phantom material since its X-ray properties are similar to soft body tissue and the solid material can be easily shaped and handled. For the objects often real bone material or artifi cial mixtures of metals and plastics are used. In this case the sphere attenuates the X-rays stronger than the surrounding PMMA. Attenuation of X-rays in materials in governed by the law of Lambert-Beer, which states that the X-ray intensity decays exponentially in the path length (x) through the material. l = l 0 exp(-µx) A B The scaling coeffi cient, µ, is called the linear attenuation coeffi cient, which describes the strength of attenuation. The µ coeffi cient is depending both on the material as well as the energy of the X-ray radiation. X-ray opaque materials have high µ values and X-ray transparent materials have (very) low µ values. F For the setup of Figure 2, one can easily simulate images by calculating the attenuation along all possible X-ray trajectories. Such an ideal simulated image of the phantom of Figure 2 is given in Figure 3. D Figure 2 Ideal imaging of a phantom Figure 3 Ideal simulated image 4 White paper, February 2008
5 These simulated images should only be used for demonstration purposes since some important image quality limiting features are not incorporated, such as the focal spot size and the infl uence of the scintillator layer. In an ideal image, which is noise- and scatter-free, the image contrast can be digitally enhanced to any given value. Thus, both low-contrast as well as high-contrast objects can be visualized equally well. Unfortunately, all real images are infl uenced by scatter. A schematic of a scatter image setup is given in Figure 4. Some X-rays are defl ected away from their original direction by interactions with the phantom material and end up at arbitrary locations on the detector (scattered rays are denoted in red in Figure 4). Figure 5 Ideal scatter profi le A B For typical radiography conditions, the amount of scatter dominates the primary (=ideal) radiation by a factor varying from 2 to even above 20. Let us assume a common value for the Scatter to Primary Ratio (SPR) of about 6. Adding the images in Figure 3 and 5 scaled by the appropriate values, gives Figure 6. F D Figure 4 Imaging system with scatter It is well-known that slab-shaped objects imaged with a wide beam give a circular shaped scatter profi le as shown in Figure 5. Figure 6 Noise-free image of feature and scatter White paper, February
6 Scatter and noise in X-ray imaging One immediately sees that the almost perfect contrast of Figure 3 is deteriorated into a weak contrast in Figure 6 and that the contrast is not constant around the peri phery of the sphere. A digital image processing tech nique called har mo ni zation can be employed to re move for a large part of the addi tive scatter contri bution. This technique is used frequently. However, scatter is not the only deteriorating effect which must be taken into account. Another important effect is noise. Quantum physics dictates that radiation is transported in quantized energy packets called photons. Due to the limited allowed irradiation of the patient (dose limitation), a limited number of X-ray photons is used to make an image. Therefore, a limited number of such photons reach each detector pixel. These quantized processes are governed by Poisson statistics. This means that the number of photons per pixel varies from pixel to pixel even in homogeneous regions. The amplitude of this variation is of the order of the square root of the average photon count. This variation is called noise. Figure 7 Scatter-free image with noise Noise is often visible as a speckle pattern in the image. The effect is demonstrated in Figures 7, 8 and 9 for the scatter-free image, the scatter image and the total image. In noisy images, the contrast cannot be arbitrarily increased by simple digital means. An increase in contrast would also lead to an increase in noise and the image would not improve. The effect of noise only decreases if a higher dose is applied, since the signal scales linearly in the dose while the noise scales with a square root in the dose. However, the dose is limited by radiation safety regulations and heat constraints of the tube. Figure 8 Scatter profi le with noise An important point, which must be stressed here, is that harmonization of Figure 9 by digital signal processing can for a large part remove the scatter intensity distribution but it cannot remove the associated noise. Therefore it is always better to remove scatter before it can reach the detector. The removal of the noise associated with scatter (by an anti-scatter grid) will lead to an improvement in the DQE of the system. The precise details on the infl uence of scatter on DQE and SNR can be found in the literature [3]. Figure 9 Total image including both scatter and noise 6 White paper, February 2008
7 Scatter reduction Several methods exist to prevent the scattered radiation to reach the detector at least to a large part. One method is to increase the distance between the patient and the detector (air gap method) such that most of the highly defl ected scatter misses the detector. Several disadvantages of this method include the decreased fi eldof-view and the decreased effi ciency of the X-ray source which increases exposure time and heat load of the X-ray source. A method which has already a long history in X-ray imaging is the use of a so-called antiscatter grid or grid. Its intended use in X-ray imaging is depicted in Figure 10. A B the label B the X-ray transparent interspacers (light grey). The lamellae are stacked in a slightly tilted way such that they all point towards a line (perpendicular to the paper) on which the X-ray focus lies. The focusing of the grid in Figure 11 is exaggerated. Typical dimensions for radiography grids are lamella thickness (d) of 0.04 mm, interspacer thickness (D) of 0.2 mm, height (h) of 3 mm, lateral dimensions 450 x 350 mm and a focus distance (f) of 1000 mm. Most high end grids use lead as lamella material and fi ber as interspacer material. Aluminum is also used as interspacer material. Aluminum has the disadvantage of an increased absorption of primary radiation compared to fi ber and is therefore more and more replaced with fi b e r, es p e c i a l l y for dose critical applications. A typical grid consists of more than 1500 carefully stacked lamella. Often the grid is enclosed in a cover for robustness and moisture protection. D C F Other often used grid-defi ning parameters can be derived from the geometrical parameters defi ned above. An important grid parameter is the line rate (N), which is defi ned as N=1/(d+D). The other important parameter is the grid ratio (r), which is defi ned as r=h/d. Figure 10 Function of antiscatter grid in X-ray imaging The grid (C in Figure 10) blocks preferentially the X-rays which do not emanate directly from the source (red arrows in Figure 10). Several grid types exist, but the most commonly used one is the linear focused grid, which consists of many thin layers of X-ray opaque material interspaced with X-ray transparent layers. A schematic cross section is shown in Figure 11. The label A denotes the X-ray opaque lamellae (in dark grey) and White paper, February
8 Scatter reduction Focus f d D d B h A Cross section D Top view Figure 11 Schematic of focused linear grid Figure 12 Grid function with scattered radiation Figure 12 depicts the grid function in more detail. Most of the primary radiation (black arrows) passes unchanged through the grid in between the lead lamellae, although some of the primary radiation is absorbed by the interspacer material. However, the scattered X-rays not emanating directly from the source (red arrows) have a high probability of being absorbed by one or more lead lamellae. The probability for a primary X-ray to pass through the grid unchanged is denoted by Tp, the primary transmission coeffi cient. The probability of a scattered X-ray to pass through the grid unchanged is denoted by Ts, the scatter transmission coeffi cient. A proper grid has a high primary transmission coeffi cient and a low scatter transmission coeffi cient. A fi ber-interspaced grid has a higher transmission of primary radiation than an aluminum-interspaced grid. The primary and scatter transmission coeffi cients are functions of imaging conditions and grid parameters. To be able to objectively measure these coeffi cients, international measurement conditions have been standardized [4]. A disadvantage of using a grid is that a stationary grid casts shadows of the lead lamellae onto the image, which gives rise to artifacts. A way to circumvent this disadvantage was already invented in the 1920 s by Bucky and Potter. In their invention, the grid is moved perpendicular to the lamella orientation with a spring system. If the movement during an exposure is over a suffi ciently large number of lamellae, the grid artifacts completely disappear. A system in which the grid is moved during exposure is called a bucky system. A broad overview of topics relating to grid design is given in [6]. 8 White paper, February 2008
9 Image quality and grid parameters To be able to choose the best grid for a given application, an objective criterion for the image quality has to be determined. Many reports in the literature have been written on this subject, but many of those stem from the analog fi lm era [6]. In analog systems a minimum exposure level of the fi lm is important. However this is not so relevant for digital systems, since the signal can be amplifi ed by digital means. For modern digital systems the most important image quality parameter is the Signal-to-Noise Ratio (SNR). Another concept, which is closely related to the SNR, is the DQE of the imaging system. The exact connection between SNR and DQE is explained in [3]. For the discussion here it is important to know that under certain conditions the improvement in DQE is proportional to the improvement in SNR squared. We will use the concept of SNR in the following since it can be more intuitively explained and it can be calculated directly from images. The SNR describes the ratio between the measured contrast (signal) and the noise level. In Section 2 we discussed the fact that the noise level in an image is proportional to the square root of the image intensity (scatter plus primary). The contrast is proportional to the primary transmission. Therefore, we obtain: Tp 2 SNR 2 sf Tp + (1 sf) Ts where sf denotes the scatter fraction which is related to the Scatter to Primary Ratio (SPR) by SPR sf = 1 + SPR Much effort has been put in devising models for Ts as a function of imaging conditions and grid parameters, either by fi tting measurement data, geometrical and analytical methods or by computer simulations [5]. General trends can be easily derived. We will state a number of such trends in the following. The coeffi cient Tp can be calculated with high precision for any set of grid parameters since the X-ray trajectories of the primary photons are known. The coeffi cient Tp increases when the relative amount of interspacer with respect to the lead lamellae increases and decreases with increasing height of the grid, due to absorption in the interspacer material. The coeffi cient Ts is much more diffi cult to calculate accurately since their point of origin is randomly distributed in the phantom (or patient). A model for Ts which is derived from experiments has been devised at Smit-Röntgen. This model has been validated for a wide range of grids at X-ray conditions which are widely used [4]. We substitute the derived functions Tp and Ts into the equation stated above for SNR 2 to obtain an equation for SNR 2 as a function of grid parameters under standard conditions. To investigate the infl uence of the height on grid performance, the SNR 2 versus the grid height for two different amounts of scatter is plotted in Figure 13 for a grid with 25 line pairs per centimeter and 69 micron thick lead lamellae. The curves show that grids operate optimally at a specifi c height. This optimum height will shift to higher values if the amount of scatter becomes larger. A current state-of-the-art fi ber interspaced grid (36 lp/cm; ratio 12, shorthanded to 36R12) has a height White paper, February
10 Image quality and grid parameters SNR of 3 mm (plus symbols in Figure 13). This grid was designed to be optimal for average-weight patients, with SPR values up to 6. fiber grids: 25 lp/cm with variable height versus standard 36R12 grid SPR = 6 SPR = 20 36R12; SPR = 6 36R12; SPR = h (mm) Figure 13 SNR versus grid height % +8% fiber interspaced grid versus Al interspaced grid (SPR = 6) +4% fiber Al A trend, observed in hospitals, is that an increasingly lar ger fraction of the patient population is obese. For obese patients a large number of examinations will have SPR values easily exceeding 10, sometimes even SPR values above 20 are reached for very large patients. For such examinations, a grid with an increased height will lead to improved image quality. For instance, increasing the height from 3mm to 5mm at an SPR value of 20 leads to an increase in SNR 2 from 5 to 6.5, which is an improve ment of more than 25%. It can also be seen that increasing the grid height to 5 mm for normal patients (SPR 6), the image quality will not be hampered and might even be improved a little. A further increase in height of the grid to values above 5mm would lead to a reduced performance at lower scatter conditions and a reduced fl exibility in application due to the increase in ratio. ploys aluminum interspacer material. The comparison of the performance of such an aluminum grid with a fi ber interspaced grid is depicted in Figure 14 for a normal patient size. From this Figure the great benefi t of the fi b e r i n t e r s p ac er material, especially at increased heights, is obvious. To obtain the same image quality with the aluminum interspaced grid as with the fi ber interspaced grid, the dose has to be increased signifi cantly. Therefore, a 5mm high aluminum interspaced grid would have an intolerable impact on dose for normal patients. The conclusion is that increasing the height of an aluminum interspaced grid would not yield the same benefi ts as with the fi ber interspaced grid. SNR % h (mm) Figure 14 Performance of fi ber grid versus aluminum grid for normal sized patients 10 White paper, February 2008
11 Obesity 1986 In recent years, especially in the USA, obesity has become a major problem. The CDC estimated in 2004 that approximately 130 million people in the USA were classifi ed as overweight, about 60 million were classifi ed as obese, and 9 million were classifi ed as severely obese (Figure 15). No data < 10% 10% - 14% 15% - 19% 20% - 24% 25% - 29% > 30% A measure which is commonly used to measure overweight, is the so-called Body-Mass Index (BMI), which is the mass of a person (in kg) divided by the square of the height of the person (in meter). According to the defi nitions used by the National Institute of Health, a BMI of over 25 is denoted as overweight, a BMI of over 30 is denoted as obese [7] No data < 10% 10% - 14% 15% - 19% 20% - 24% 25% - 29% > 30% 2006 Not only the obesity itself poses a health risk for the patient, X-ray imaging, especially of the waist region, is severely hampered by the excess body tissue. To be able to penetrate the increased patient thickness, normally the kv settings of the X-ray tube are increased and higher energy X-rays are generated. These higher energy X-rays penetrate better through the patient. However, the contrast between the different body tissues decreases for these higher energy X-rays leading to a decrease in contrast in the image and lower image quality. Furthermore, the increased path length of the X- rays in the patient lead to a large increase in the amount of scatter, i.e., the SPR increases rapidly. This increase in scatter amount leads to a further decrease in image quality. Unfortunately, little can be done on the contrast decrease, but the increase in scatter can be successfully battled with an improved anti-scatter grid, leading to an improvement in the system s DQE. No data < 10% 10% - 14% 15% - 19% 20% - 24% 25% - 29% > 30% Figure 15 Obesity in the USA White paper, February
12 A new 5 mm high grid With the insights of the previous sections, a new 5mm high fi ber interspaced grid has been constructed with 25 line pairs per cm and a ratio of 15 (25R15). The grid s form factor has been designed such that it will fi t into a normal digital bucky system. The ratio of the new grid is very close to the current ( standard ) grid. Therefore, a one to one swap of the current grid with the new grid can be made with no or only minor modifi cations to the bucky. This swap has no negative effect on image quality for normal patients, but it will improve image quality for the obese patients. deviation of background signal (noise level). Dividing both numbers gives an indication of the Signal to Noise Ratio. These measurements were performed with approximately 1cm of water thickness on top of a pile of PMMA plates with thicknesses of 20 cm, 30 cm and 40 cm. The measured differences in intensity values and standard deviations are tabulated in Table 1. Several experiments have been performed with this grid to compare its performance with the performance of a standard 3 mm high 36 lines/cm ratio 12 (36R12) grid. Water Low contrast measurements PMMA A common and easy quantifi cation procedure is the low contrast measurement where a small object of low contrast is placed on top of a thick and large homogeneous object (Figure 16). From the obtained image, one can measure the average difference in signal strength between the low contrast object and the background. This difference is compared to the standard Figure 16 Low contrast measurements; circles are Regions of Interest 12 White paper, February 2008
13 thickness grid height kvp mas Signal difference Improvement SNR CDRAD measurements 20 cm 36R12 3 mm cm 25R15 5 mm cm 36R12 3 mm cm 25R15 5 mm cm 36R12 3 mm cm 25R15 5 mm % + 28 % + 68 % A different method to quantify grid performance involves the CDRAD phantom [8]. The CDRAD phantom consists of a thin PMMA plate with holes of various sizes and depths, such that both the Contrast (different hole depths) and the Detail (different hole sizes) performance can be measured. Table 1 Low contrast measurements One can see clearly from the rightmost column in Table 1, that the signal to noise ratio is increased by employing the new grid at constant kvp and mas settings compared to the current 36R12 grid. Especially in the case of 40 cm PMMA, the SNR value increases tremendously. This measured increase in SNR performance is even higher than we expect from the theoretical treatment of section 4. The DQE improvement is the square of the SNR improvement [3] and therefore almost a factor of three is gained. To generate a realistic amount of scatter, 30 cm of PMMA was added above the CDRAD phantom. The leftmost image in Figure 17 is made with a standard 36R12 (3 mm high) grid; the rightmost image in Figure 17 is made with the new 25R15 (5 mm high) grid, under identical conditions (90 kv, 200 mas, SPR 10). It is immediately obvious that the rightmost image suffers much less from the scatter background and more details can be discerned, especially in the center of the image. The software provided by the phantom manufacturer is able to quantify this improvement in image quality into a single number (IQinv, [8] ). A higher number denotes a better image. For the 36R12 grid the software determines IQinv to be 3.67, for the 25R15 grid an IQinv of 4.19 is found. Standard grid Figure 17 CDRAD phantom images New grid White paper, February
14 A new 5 mm high grid Standard grid New grid Figure 18 Body phantom images at equal dose settings Body phantom measurements To investigate the effect the new grid would have on a clinical image, body phantom images were made. The body phantom is an artifi cial chest of approximately 20 cm thickness, fi lled with bones, tissue and air. The body phantom was thickened with the addition of 15 cm of PMMA. The maximum total thickness was 35 cm. The image with the standard (36R12) grid was imaged at 80 kv because at a higher kvp setting the image contrast becomes intolerably compromised. When the new 25R15 grid is used, due to the reduced infl uence of scatter and improved contrast, a higher kvp setting of 90 kv can be chosen at a lower mas setting to have equal dose to the patient and as can be seen from the right image in Figure 18, the image quality is improved, especially in the lower abdominal region. Clinical testing The new 25R15 grid is currently under clinical testing at a leading medical institute. A detailed report will be published in a leading medical journal. 14 White paper, February 2008
15 Summary and conclusions In this document an overview has been given on all the major components of a general radiography X-ray system. The infl uence of scatter and noise on image quality has been discussed and the infl uence of an antiscatter grid is shown. A simple model has been derived for the image quality as a function of grid parameters. From this model it is shown that at high scatter conditions a higher grid is benefi cial. Calculations show also the large benefit of the fiber interspaced grid with respect to the aluminum interspaced grid, especially for higher grids. A grid with increased height of 5 mm has been made and its performance under several experimental conditions has been tested and compared to the performance of a normal grid. The conclusion from these experiments was that an improved image is obtained by using a grid with increased height for patients which have above average weight. The improvement stems from the increased effi ciency of removing scatter at equal dose. The improvement in SNR values was shown to be 69 % in some cases. An increase in SNR value leads to an increase in the system s DQE up to a factor of three. For normal sized patients it is shown that the image quality is not compromised. The grid is designed such that it fi ts into a normal digital bucky system. References [1] Flat detectors and their clinical applications, Spahn, Eur. Radiol. 15 (2005) [2] Handbook of medical imaging Volume 1: Physics and psychophysics, Beutel et al, SPIE Press, Bellingham WA (2000). [3] Generalizing the MTF and DQE to include x-ray scatter and focal spot unsharpness: Application to a new microangiographic system, Kyprianou et al, Med. Phys. 32 (2005) [4] IEC 60627, IEC, Geneva (2001) [5] A detailed experimental and theoretical comparison of the angular and energy dependencies of grid transmission, Bernstein et al, Med. Phys. 10 (1983) [6] Analysis of secondary screening with special reference to grids for abdominal radiography, Strid, Acta Radiologica Supplement 351, Stockholm 1976 [7] ning. htm [8] White paper, February
16 All rights are reserved. Reproduction in whole or in part is prohibited without the prior written consent of the copyright holder. Dunlee reserves the right to make changes in specifi cation and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. Dunlee is part of Philips Healthcare. Printed in Germany: /740 * FEB 2008 Smit Röntgen grids are exclusively available through Dunlee. Contact worldwide Dunlee Postbox MD Eindhoven The Netherlands Tel: Fax:
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
Measurement of Charge-to-Mass (e/m) Ratio for the Electron
Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Coating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Theremino System Theremino Spectrometer Technology
Theremino System Theremino Spectrometer Technology theremino System - Theremino Spectrometer Technology - August 15, 2014 - Page 1 Operation principles By placing a digital camera with a diffraction grating
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis
Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis David Bernard & Steve Ainsworth Dage Precision Industries Abstract Non-destructive testing during the manufacture of
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries
2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,
Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography
Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
A Guide to Acousto-Optic Modulators
A Guide to Acousto-Optic Modulators D. J. McCarron December 7, 2007 1 Introduction Acousto-optic modulators (AOMs) are useful devices which allow the frequency, intensity and direction of a laser beam
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
Geometric Optics Converging Lenses and Mirrors Physics Lab IV
Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION
X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION David Bernard Dage Precision Industries Inc. Fremont, California [email protected] ABSTRACT The x-ray inspection of PCB assembly processes
Measuring Line Edge Roughness: Fluctuations in Uncertainty
Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:
PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics
Production of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
Signal to Noise Instrumental Excel Assignment
Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a
Experiment 3 Lenses and Images
Experiment 3 Lenses and Images Who shall teach thee, unless it be thine own eyes? Euripides (480?-406? BC) OBJECTIVES To examine the nature and location of images formed by es. THEORY Lenses are frequently
Reflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)
X-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
Engineering & Design: Geometric Dimensioning
Section Contents NADCA No. Format Page 1 Introduction -2 S E C T I O N 2 What is GD&T? -2 3 Why Should GD&T be Used? -3 4 Datum Reference Frame -4 4.1 Primary, Secondary, Tertiary Features & Datums -4
Radiography: 2D and 3D Metrics of Performance Towards Quality Index
AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
Fiber Optics: Fiber Basics
Photonics Technical Note # 21 Fiber Optics Fiber Optics: Fiber Basics Optical fibers are circular dielectric wave-guides that can transport optical energy and information. They have a central core surrounded
Principle of Thermal Imaging
Section 8 All materials, which are above 0 degrees Kelvin (-273 degrees C), emit infrared energy. The infrared energy emitted from the measured object is converted into an electrical signal by the imaging
Physics 441/2: Transmission Electron Microscope
Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This
Antenna Deployment Technical Brief
ProCurve Networking Antenna Deployment Technical Brief Introduction... 2 Antenna types... 2 Omni directional antennas... 2 Directional antennas... 2 Diversity antennas... 3 High gain directional antennas...
Handbook on the Ultrasonic Examination. Austenitic Welds
Handbook on the Ultrasonic Examination Austenitic Welds The International Institute of Welding Edition Handbook On the Ultrasonic Examination of Austenitic Welds Compiled by COMMISSION V Testing, Measurement,
Light in the Greenhouse: How Much is Enough?
Light in the Greenhouse: How Much is Enough? by: James W. Brown http://www.cropking.com/articlelghe Most of us know that green plants need light for photosynthesis, growth, and development. As important
4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:
4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing
Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
Lesson 26: Reflection & Mirror Diagrams
Lesson 26: Reflection & Mirror Diagrams The Law of Reflection There is nothing really mysterious about reflection, but some people try to make it more difficult than it really is. All EMR will reflect
Single Mode Fiber Lasers
Single Mode Fiber Lasers for Industrial and Scientific Applications T h e P o w e r t o T r a n s f o r m T M IPG s Single Mode Fiber Lasers Advantages IPG's YLR-SM Series represents a break-through generation
X-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
An Overview of the Finite Element Analysis
CHAPTER 1 An Overview of the Finite Element Analysis 1.1 Introduction Finite element analysis (FEA) involves solution of engineering problems using computers. Engineering structures that have complex geometry
Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.
T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting
Using the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)
GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding
OmniBSI TM Technology Backgrounder. Embargoed News: June 22, 2009. OmniVision Technologies, Inc.
OmniBSI TM Technology Backgrounder Embargoed News: June 22, 2009 OmniVision Technologies, Inc. At the heart of any digital camera lies the image sensor. The image sensor is an integrated circuit, like
Lab 4: Magnetic Force on Electrons
Lab 4: Magnetic Force on Electrons Introduction: Forces on particles are not limited to gravity and electricity. Magnetic forces also exist. This magnetic force is known as the Lorentz force and it is
Algebra 2 Chapter 1 Vocabulary. identity - A statement that equates two equivalent expressions.
Chapter 1 Vocabulary identity - A statement that equates two equivalent expressions. verbal model- A word equation that represents a real-life problem. algebraic expression - An expression with variables.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
Field-Effect (FET) transistors
Field-Effect (FET) transistors References: Hayes & Horowitz (pp 142-162 and 244-266), Rizzoni (chapters 8 & 9) In a field-effect transistor (FET), the width of a conducting channel in a semiconductor and,
Determination of source parameters from seismic spectra
Topic Determination of source parameters from seismic spectra Authors Michael Baumbach, and Peter Bormann (formerly GeoForschungsZentrum Potsdam, Telegrafenberg, D-14473 Potsdam, Germany); E-mail: [email protected]
Radioactivity III: Measurement of Half Life.
PHY 192 Half Life 1 Radioactivity III: Measurement of Half Life. Introduction This experiment will once again use the apparatus of the first experiment, this time to measure radiation intensity as a function
TOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] [email protected] [e-mail] This
Name Date Class ELECTRONS IN ATOMS. Standard Curriculum Core content Extension topics
13 ELECTRONS IN ATOMS Conceptual Curriculum Concrete concepts More abstract concepts or math/problem-solving Standard Curriculum Core content Extension topics Honors Curriculum Core honors content Options
where h = 6.62 10-34 J s
Electromagnetic Spectrum: Refer to Figure 12.1 Molecular Spectroscopy: Absorption of electromagnetic radiation: The absorptions and emissions of electromagnetic radiation are related molecular-level phenomena
Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications
Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
WHITE PAPER. Source Modeling for Illumination Design. Zemax A Radiant Zemax Company
Source Modeling for Illumination Design Source Modeling for Illumination Design Authored by: Ronald F. Rykowski and C. Benjamin Wooley Abstract As computation speeds have increased dramatically over the
ELECTRON SPIN RESONANCE Last Revised: July 2007
QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron
Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
Lecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
Sound absorption and acoustic surface impedance
Sound absorption and acoustic surface impedance CHRISTER HEED SD2165 Stockholm October 2008 Marcus Wallenberg Laboratoriet för Ljud- och Vibrationsforskning Sound absorption and acoustic surface impedance
Multiple Optimization Using the JMP Statistical Software Kodak Research Conference May 9, 2005
Multiple Optimization Using the JMP Statistical Software Kodak Research Conference May 9, 2005 Philip J. Ramsey, Ph.D., Mia L. Stephens, MS, Marie Gaudard, Ph.D. North Haven Group, http://www.northhavengroup.com/
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control
INTERFERENCE OF SOUND WAVES
1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.
Radiation Strip Thickness Measurement Systems
Radiation Strip Thickness Measurement Systems During the past years we have increased our sales of radiometric Vollmer strip thickness measurement systems, i.e. X-ray or isotope gauges, dramatically. Now,
The Basics of FEA Procedure
CHAPTER 2 The Basics of FEA Procedure 2.1 Introduction This chapter discusses the spring element, especially for the purpose of introducing various concepts involved in use of the FEA technique. A spring
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION
Reduction of Electron Contamination Using a Filter for 6MV Photon Beam
J. Korean Soc Ther Radiol Oncol : Vol. 15, No. 2, June, 1997 Reduction of Electron Contamination Using a Filter for 6MV Photon Beam Choul Soo Lee, MPH., Myung Jin Yoo, Ph.D. and Ha Yong Yum, M.D. Department
MD5-26 Stacking Blocks Pages 115 116
MD5-26 Stacking Blocks Pages 115 116 STANDARDS 5.MD.C.4 Goals Students will find the number of cubes in a rectangular stack and develop the formula length width height for the number of cubes in a stack.
Parabolic Trough Solar Collector Analysis
Parabolic Trough Solar Collector Analysis Derek Lipp AME 4043 Capstone Design Project 28 February 2008 Abstract The purpose of this trade study is to examine a parabolic trough and determine the ideal
Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
Scanners and How to Use Them
Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.
Jitter Measurements in Serial Data Signals
Jitter Measurements in Serial Data Signals Michael Schnecker, Product Manager LeCroy Corporation Introduction The increasing speed of serial data transmission systems places greater importance on measuring
Dithering in Analog-to-digital Conversion
Application Note 1. Introduction 2. What is Dither High-speed ADCs today offer higher dynamic performances and every effort is made to push these state-of-the art performances through design improvements
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
THE INTEGRATED CAD/CAM SOLUTION FOR THE WOOD INDUSTRY THE INTEGRATED CAD/CAM/ERP SOLUTION
THE INTEGRATED CAD/CAM SOLUTION FOR THE WOOD INDUSTRY THE INTEGRATED CAD/CAM/ERP SOLUTION ONE CAD/CAM SOLUTION AND ONE PROFESSION: WOOD TopSolid Wood is both a very powerful CAD/CAM solution, and a dedicated
The Sonometer The Resonant String and Timbre Change after plucking
The Sonometer The Resonant String and Timbre Change after plucking EQUIPMENT Pasco sonometers (pick up 5 from teaching lab) and 5 kits to go with them BK Precision function generators and Tenma oscilloscopes
Chemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium
Chemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium Introduction The word equilibrium suggests balance or stability. The fact that a chemical reaction
2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
University of California at Santa Cruz Electrical Engineering Department EE-145L: Properties of Materials Laboratory
University of California at Santa Cruz Electrical Engineering Department EE-145L: Properties of Materials Laboratory Lab 8: Optical Absorption Spring 2002 Yan Zhang and Ali Shakouri, 05/22/2002 (Based
INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer
INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer X-rays were discovered by W.C. Roentgen because of their ability to cause fluorescence. X-ray photons cannot be seen. The image produced
EMC STANDARDS STANDARDS AND STANDARD MAKING BODIES. International. International Electrotechnical Commission (IEC) http://www.iec.
EMC STANDARDS The EMC standards that a particular electronic product must meet depend on the product application (commercial or military) and the country in which the product is to be used. These EMC regulatory