CT artifacts: Causes and reduction techniques
|
|
|
- Phyllis Gardner
- 10 years ago
- Views:
Transcription
1 F Edward Boas & Dominik Fleischmann* Department of Radiology, Stanford University School of Medicine, 300 Pasteur Drive, Stanford, CA 94305, USA *Author for correspondence: Tel.: Artifacts are commonly encountered in clinical computed tomography (CT), and may obscure or simulate pathology. There are many different types of CT artifacts, including noise, beam hardening, scatter, pseudoenhancement, motion, cone beam, helical, ring, and metal artifacts. We review the cause and appearance of each type of artifact, correct some popular misconceptions, and describe modern techniques for artifact reduction. Noise can be reduced using iterative reconstruction or by combining data from multiple scans. This enables lower radiation dose and higher resolution scans. Metal artifacts can also be reduced using iterative reconstruction, resulting in more accurate diagnosis. Dual and multi-energy (photon counting) CT can reduce beam hardening and provide better tissue contrast. Methods for reducing noise and out-of-field artifacts may enable ultra-high resolution limited-field-of-view imaging of tumors and other structures. Keywords: noise, beam hardening, scatter, pseudoenhancement, metal artifact, dose reduction, iterative reconstruction, dual energy CT, micro CT, ring artifact Executive summary Ring artifact Ring artifact is caused by a miscalibrated or defective detector element, which results in rings centered on the center of rotation. This can often be fixed by recalibrating the detector. Noise Poisson noise is due to the statistical error of low photon counts, and results in random thin bright and dark streaks that appear preferentially along the direction of greatest attenuation. This can be reduced using iterative reconstruction, or by combining data from multiple scans. Noise reduction techniques enable diagnostic scans at a much lower radiation dose. With iterative reconstruction, low dose results in decreased resolution, with only a slight increase in noise. Model-based iterative reconstruction (MBIR), for example, attempts to smooth out the noise while preserving edges, resulting in a plastic appearance, where there are small clusters of pixels with similar Hounsfield units. Beam hardening and scatter Beam hardening and scatter both produce dark streaks between two high attenutation objects (such as metal or bone), with surrounding bright streaks. These can be reduced using iterative reconstruction. Dual energy CT reduces beam hardening, but not scatter. Beam hardening and scatter also cause pseudoenhancement of renal cysts. (Author s version) Imaging Med. (2012) 4(2),
2 Metal artifact Metal streak artifacts are caused by multiple mechanisms, including beam hardening, scatter, Poisson noise, motion, and edge effects. The Metal Deletion Technique (MDT) is an iterative technique that reduces artifacts due to all of these mechanisms. In some cases, the improved image quality can change the diagnosis. Out of field artifact Out of field artifacts are due to a suboptimal reconstruction algorithm, and can be fixed using a better algorithm. Images can then be acquired using a field of view that is much smaller than the object being scanned, thus reducing the radiation dose. Higher resolution scanners will likely require iterative reconstruction or limited field of view scans to reduce the radiation dose required to achieve an acceptable level of noise. Introduction In an idealized situation, with high radiation dose and thus high photon counts, monochromatic X-rays, infinite detector resolution, perfect detectors, no motion, and no scatter, computed tomography (CT) images would be a perfect reflection of reality. If any of those conditions are not met, then artifacts will occur. In this article, we illustrate commonly encountered artifacts in clinical CT, how they can obscure or simulate pathology, and how they can be reduced. Ring artifact A miscalibrated or defective detector element creates a bright or dark ring centered on the center of rotation [1]. This can sometimes simulate pathology (Figure 1). Usually, recalibrating the detector is sufficient to fix this artifact, although occasionally the detector itself needs to be replaced. A B C Figure 1. Ring artifact. A. Pelvic CT showing severe ring artifact. B. Head CT with subtle ring artifact simulating a pons lesion (arrow). C. Changing the window / level settings shows the circular reconstruction region, which is centered at the center of rotation. The pons pseudolesion (marked with a small circle) is exactly at the center of the circular reconstruction region, and thus consistent with a ring artifact. Follow-up MRI showed a normal pons. (Author s version) Imaging Med. (2012) 4(2),
3 Noise Poisson noise is due to the statistical error of low photon counts, and results in random thin bright and dark streaks that appear preferentially in the direction of greatest attenuation (Figure 2). With increased noise, high contrast objects such as bone may still be visible, but low contrast soft tissue boundaries may be obscured. For conventional filtered backprojection (FBP) images, the standard deviation in Hounsfield units (HU) due to Poisson noise [2] is proportional to 1/ slice thickness mas. This relationship applies when comparing corresponding regions in two images acquired with a different mas or slice thickness. It also assumes that the underlying tissue has perfectly uniform Hounsfield units. If the underlying tissue is heterogenous, then the standard deviation in Hounsfield units equals, where s 1 is the standard deviation due to the tissue texture, and s 2 is the standard deviation due to Poisson noise. Poisson noise can be decreased by increasing the mas. Modern scanners can perform tube current modulation, selectively increasing the dose when acquiring a projection with high attenuation. They also typically use bowtie filters, which provide a higher dose towards the center of the field of view compared to the periphery. There is a tradeoff between noise and resolution, so noise can also be reduced by increasing the slice thickness, using a softer reconstruction kernel (soft tissue kernel instead of bone kernel), or blurring the image. Noise can also be reduced by moving the arms out of the scanned volume for an abdominal CT. If the arms cannot be moved out of the scanned volume, placing them on top of the abdomen should reduce noise relative to placing them at the sides. Similarly, large breasts should be constrained in the front of the thorax rather than on both sides in thoracic and cardiac CT. This is because the noise increases rapidly as the photon counts approach zero, which means that the maximum attenuation has a bigger effect on the noise than the average attenuation. In filtered backprojection, which is the standard reconstruction method on most scanners, the projection data are filtered to sharpen edges, and the filtered data are then backprojected [1]. This assumes accurate projection data, and ignores the fact that low photon counts result in a large Poisson error. On the other hand, iterative methods [3, 4] use a statistical model of the noise to improve the image on each iteration. A wide range of techniques have been proposed, and all major vendors now offer various implementations of iterative reconstructions algorithms on their systems. The basic concept is to find the most probable image given: the projection data, the relationship between the image and the projection data (which can include Poisson noise, beam hardening, and scatter), and the prior distribution of images (which often assumes that smoother images are more probable). This optimization problem is too difficult to solve analytically, and is thus solved iteratively. With noisy projection data, there is a wide range of different images that are consistent with the measured projection data. The prior distribution of images directs the iterative reconstruction to pick a smoother image out of the range of possible images. Iterative methods require faster computer chips, and have only recently become available for clinical use. One iterative method, Model-based iterative reconstruction (MBIR, General Electric) [5, 6] received U.S. FDA approval in September 2011 [7]. MBIR substantially reduces (Author s version) Imaging Med. (2012) 4(2),
4 image noise and improves image quality, thus allowing scans to be acquired at lower radiation doses (Figure 3) [2]. Furthermore, due to the tradeoff between noise and resolution, these methods will likely also be important for reducing noise in higher resolution images. Compared to conventional FBP, iterative reconstruction has a different relationship between noise and dose, and has a different noise texture. With FBP, as the dose is reduced, both the noise and image quality become worse. On the other hand, with MBIR, noise and image quality are decoupled: as the dose is reduced, the noise increases only slightly, but resolution worsens, and new artifacts may be introduced at very low dose levels [2]. Thus, traditional measures such as the signal-to-noise ratio are not applicable for MBIR and other iterative reconstruction methods. The noise texture depends on the parameters of the MBIR [6]. Specifically, MBIR attempts to generate a smooth image while preserving edges, and has adjustable parameters to control the trade-off between smoothness and edge-preservation. Thus, the noise tends to coalesce into small clusters of pixels with uniform Hounsfield units, resulting in what has been described as a plastic appearance. Noise can also be reduced by combining information from multiple scans, such as multiple contrast phases [8, 9]. This has important implications for whole organ dynamic contrastenhanced ( perfusion ) imaging, where radiation dose is currently one of the limiting factors. A low noise scan is created by averaging scans performed at multiple time points. The temporal resolution is recovered by multiplying the average scan by a per-pixel weighting factor, which is the blurred image at that time point, divided by the blurred average image. A. 60 ma, 120 kvp, slice thickness 5 mm B. 440 ma, 120 kvp, slice thickness 5 mm Figure 2. Effect of ma on Poisson noise. A. Low dose CT image obtained during a CT-guided biopsy shows extensive Poisson noise. These streaks are the same whether or not the abdomen or arms are partially outside the field of view. B. Post-biopsy image obtained at 7.3 times higher dose has times less noise. The images show an enlarged retroperitoneal lymph node (arrow) and infiltration of the right kidney in a patient with Hodgkin s lymphoma. (Author s version) Imaging Med. (2012) 4(2),
5 A. 50 ma, FBP B. 50 ma, MBIR C. 754 ma, FBP Figure 3. Iterative reconstruction reduces noise and improves image quality. A. FBP image obtained at low dose is extremely noisy. B. The same low dose scan reconstructed using Model Based Iterative Reconstruction (MBIR) results in dramatically reduced noise, revealing new soft tissue details. In particular, note the details in the right renal hilum, and the nodular cirrhotic liver. C. The details in the MBIR image are confirmed in a higher dose FBP image. (Figure modified from [4], with permission). Beam hardening and scatter Beam hardening and scatter are different mechanisms that both produce dark streaks between two high attenuation objects, such as metal, bone, iodinated contrast, or barium. They can also produce dark streaks along the long axis of a single high attenuation object (Figure 4 and Figure 7A) [1]. Bright streaks are seen adjacent to the dark streaks. These artifacts are a particular problem in the posterior cranial fossa, and with metal implants. (Metal artifacts are discussed further in the Metal artifact section below.) Beam hardening is seen with polychromatic X-ray sources. As the X-ray passes through the body, low energy X-ray photons are attenuated more easily, and the remaining high energy photons are not attenuated as easily. Thus, beam transmission does not follow the simple exponential decay seen with a monochromatic X-ray. This is a particular problem with high atomic number materials such as bone, iodine, or metal. Compared to low atomic number materials such as water, these high atomic number materials have dramatically increased attenuation at lower energies. (For low energy X-rays, attenuation is primarily due to the photoelectric effect, and is proportional to Z 3 /E 3, where Z is the atomic number, and E is the energy. At high energies, attenuation is primarily due to Compton scatter, and is proportional to 1/E.) Compton scatter causes X-ray photons to change direction (and energy), and thus end up in a different detector [10]. This creates the greatest error when the scattered photon ends up in a detector that otherwise would have very few photons. In particular, if a metal implant blocks all photons, then the corresponding detector element will only detect scattered photons. Scatter also becomes more significant with an increased number of detector rows, because a larger volume of tissue is irradiated. (Author s version) Imaging Med. (2012) 4(2),
6 Thus, for highly attenuated X-ray beams, beam hardening and scatter both cause more photons to be detected than expected, resulting in dark streaks along the lines of greatest attenuation. In addition, the high pass filter used in FBP exaggerates differences between adjacent detector elements, producing bright streaks in other directions (Figure 4). Scanning at a higher kv results in a harder X-ray beam, and thus less beam hardening artifacts. In addition, metal is more transparent to higher energy photons, making it less likely to block all photons, thus reducing scatter artifacts. However, the tradeoff is that there is less tissue contrast at high kv. Modern scanners perform a simple beam hardening correction that assumes an average amount of beam hardening, given the measured attenuation [11]. However, higher atomic number materials such as metal cause a higher than average amount of beam hardening, and will thus not be fully corrected. This can be addressed using iterative reconstruction [12, 13]. The first iteration is reconstructed using uncorrected projection data. Metal and bone are then detected using a Hounsfield unit cutoff, and these are forward projected to determine how much bone and metal are present in each detector measurement. This information is then used to perform a custom beam hardening correction for each detector element. Dual energy CT reduces beam hardening effects by scanning at two different energies. This information can be used to derive virtual monochromatic images, which do not suffer from beam hardening effects. However, the virtual monochromatic images produced by dual energy CT assume that the X-ray absorption spectrum has an idealized shape, without K-edges, which is clearly just an approximation [14]. In addition, dual energy CT does not correct for scatter, which is an important factor in many scans [10, 15], especially if the metal blocks nearly all photons. Figure 4. Simulated scans without (top row) and with (bottom row) beam hardening, showing that dark streaks occur along the lines of greatest attenuation, and bright streaks occur in other directions. Scatter produces artifacts that look similar to this. Also note the subtle decrease in Hounsfield units just beneath the surface of the abdomen, which is caused by beam hardening. This is called cupping artifact, and it is corrected by the simple beam hardening correction built into modern scanners. (Author s version) Imaging Med. (2012) 4(2),
7 Most scanners use an anti-scatter grid in front of the detector to reduce scatter. Scatter can also be estimated (using a scatter kernel, or from measurements made just outside the field of view), and then subtracted from the detector measurements. Finally, the image can be reconstructed iteratively, where the scatter correction is estimated using the image from the previous iteration [16, 17]. However, in cases where metal blocks all photons (and thus all detected photons are due to scatter), soft tissue information for those detector elements is lost, and cannot be retreived using scatter correction. Pseudoenhancement Pseudoenhancement of renal cysts refers to the fact that simple renal cysts have spuriously increased Hounsfield units after administration of intravenous contrast. This is caused by beam hardening and scatter, even though it does not have the streaks that are more classically associated with beam hardening. The same mechanism is responsible for the increased density seen just inside the skull on head CT. Areas that are surrounded by a ring of high density material become brighter due to beam hardening and scatter (last column of Figure 4). One way to understand this phenomenon is by analogy to the third column of Figure 4. Just inside the dark streaks formed by the 3 implants, there is a bright triangle. This is exactly analogous to the apparent high density seen inside a ring of high density. Pseudoenhancement decreases with the distance from enhancing renal tissue. Thus, there is more pseudoenhancement in smaller cysts, and Hounsfield unit measurements should be performed as far away from the enhancing renal tissue as possible. In conventional CT, pseudoenhancement of up to 28 HU is seen [18]. This may be decreased with dual energy CT [19]. However, it is not eliminated, because dual energy CT only gives approximate monoenergetic images, and does not correct for scatter (as discussed above). Motion artifact Motion (patient, cardiac, respiratory, bowel) causes blurring and double images, as well as long range streaks (Figure 5). The streaks occur between high contrast edges and the X-ray tube position when the motion occurs. Faster scanners reduce motion artifact because the patient has less time to move during the acquisition. This can be accomplished with faster gantry rotation or more X-ray sources [4]. More detector rows allows a greater volume to be imaged in a single gantry rotation, thus increasing the distance between step-off artifacts from motion on coronal or sagittal reformats. Rigid body motion artifacts (mainly a problem with head CT, as shown in Figure 5) can be reduced using special reconstruction techniques [20]. Respiratory motion in cone-beam CT with slow gantry rotation can be estimated and corrected, thus reducing artifacts [21]. With a very fast scanner, the heart can be scanned during diastole within a single heartbeat, significantly reducing cardiac motion, thus allowing evaluation of the coronary arteries [22]. (Author s version) Imaging Med. (2012) 4(2),
8 Alternatively, with ECG gating, projection data are acquired over multiple cardiac cycles, and then reconstructed from data acquired during specific phases of the cardiac cycle [4]. This can be used to make 3D movies of a beating heart. With current scanners, evaluation is suboptimal at higher heart rates, and for images obtained during systole [23]. Temporal resolution in cardiac CT can be improved using new techniques that work with limited projection data [24]. Figure 5. Motion causes blurring and double images (left), as well as long range streaks (right). Cone-beam (multidetector row) and windmill (helical) artifacts Helical multidetector row CT has some additional artifacts that are not seen in single detector row step-and-shoot CT. On the other hand, the significantly reduced scan time reduces motion artifact. In helical CT, the table is continuously advanced as the X-ray tube rotates around the patient. As the detector rows pass by the axial plane of interest, the reconstruction oscillates between taking measurements from a single detector row, and interpolating between two detector rows. If there is a high contrast edge between the two detector rows, then the interpolated value may not be accurate. This creates smooth periodic dark and light streaks originating from high contrast edges, which are called windmill artifacts (Figure 7E). These are more prominent on thin slices, and the vanes of the windmill rotate as one scrolls through axial slices. A similar mechanism is responsible for stair-step artifacts (serrations on coronal or sagittal reformats) [25] and zebra artifacts (periodic stripes of more or less noise at the image periphery seen on coronal or sagittal reformats); these are show in in Figure 6. In multidetector row CT, the projection planes (defined by the X-ray source and the detector row) are not exactly parallel to the axial plane (except for the center detector row). In the simplest 2D FBP reconstruction, the projection planes for each detector row are assigned to the closest axial plane based on where they intersect the center of rotation. If there is a high contrast edge in the z direction between the axial plane and the projection plane, this creates streaks, as well as stair-step artifacts (Figure 6). These effects are worse with an increased number of detector rows. These artifacts can be reduced with Adaptive Multiple Plane Reconstruction (AMPR), which uses tilted planes for reconstruction [26]. Cone-beam reconstructions, which reconstruct the entire 3D volume at the same time using the correct multidetector row geometry (Author s version) Imaging Med. (2012) 4(2),
9 [27] also reduce this artifact, but are much slower. Clinical flat panel detector CTs use conebeam reconstruction. Figure 6. A. Zebra artifacts (alternating high and low noise slices, arrows) due to helical interpolation. These are more prominent at the periphery of the field of view. B. Stair-step artifacts (arrows) seen with helical and multidetector row CT. These are also more prominent near the periphery of the field of view. Therefore, it is important to place the object of interest near the center of the field of view. Metal artifact Metal streak artifacts are extremely common: 21% of scans in one series [28]. They are caused by multiple mechanisms, some of which are related to the metal itself, and some of which are related to the metal edges. The metal itself causes beam hardening, scatter effects, and Poisson noise, which are discussed above. Beam hardening and scatter result in dark streaks between metal, with surrounding bright streaks (Figure 7A). The metal edges cause streaks due to undersampling, motion, cone beam, and windmill artifacts [29]. The large discontinuities in detector measurements created by metal edges are amplified by the filter in FBP. In the limit of perfect data with infinite resolution, these edges cancel out away from metal. However, with undersampling, or imperfections in the data (caused by motion, cone beam, or windmill effects), they do not exactly cancel, resulting in thin bright and dark streaks originating from the metal (Figure 7C and E). Metal artifacts are particularly pronounced with high atomic number metals such as iron or platinum, and less pronounced with low atomic number metals such as titanium. In some cases (such as dental fillings on head CTs), patient positioning or gantry tilt can angle the metal outside of the axial slices of interest. (Author s version) Imaging Med. (2012) 4(2),
10 Several techniques have been proposed for metal artifact reduction [28, 30-32]. We developed an iterative method called the Metal Deletion Technique (MDT) [28], which is based on the principle that projection data involving or near metal is less accurate, due to the mechanisms discussed above. MDT starts with raw projection data from the scanner, and then only uses high quality non-metal data to reconstruct the non-metal portions of the image. Metal pixels are deleted from the reconstructed image, and on each iteration, the inaccurate metal data are replaced with forward projected values from the previous iteration. This means that, instead of trying to look through the metal to see soft tissue, we look around the metal. It also means that any features that can only be seen by looking through metal will be lost. In particular, structures within a few millimeters of metal are blurred out. An initial evaluation of MDT showed that it had the best image quality when compared against FBP and two metal artifact reduction methods [28]. In 2 of 11 scans, the improved image quality revealed important new findings. This includes a case of rectal cancer (in a patient with bilateral hip replacements) that was originally missed when reviewing only the images produced by the scanner. Raw projection data from the scanner is stored in a proprietary format, and therefore not always accessible. Fortunately, the raw data can be estimated by forward projecting the reconstructed image. Using this technique, a follow up study of 80 patients showed that MDT improved image quality 73% of the time for small metal implants, and 75% of the time for large metal implants [33]. MDT had better image quality than all three other metal artifact reduction techniques tested. At Stanford Hospital, we have integrated metal artifact reduction into our PACS system. The DICOM send function is used to send scans to a server that automatically reduces artifacts and sends the processed images back to PACS as a new series under the same accession. This procedure works with images from any scanner, and it does not require any manual drawing of regions of interest, or tuning of parameters. We have found this to be particularly useful for radiation oncology [34], interventional radiology [35], orthopedics, and neurosurgery (Figure 7) applications. In some cases, MDT decreases resolution or introduces new artifacts. Thus, MDT images must be reviewed in conjunction with the original images produced by the scanner. Some portions of the image may be more clearly seen on the original image, and other portions are more clearly seen on the MDT images. A review of 102 cases shows the types of metal devices that tend to produce the best results (Table 1). (Author s version) Imaging Med. (2012) 4(2),
11 Beam hardening and scatter A Motion and undersampling C Windmill E B D F Figure 7. MDT reduces many different types of metal artifacts, and can reveal new findings. A. Dark streak between hip replacements is mostly due to beam hardening and scatter. B. The MDT image more clearly shows a fluid collection adjacent to the left hip replacement. C. Sharp thin alternating streaks surrounding an aneurysm coil are mostly due to motion and undersampling. D. MDT image reveals hemorrhage around the coil. E. Smoothly undulating streaks around cholecystectomy clips are due to windmill artifact. F. MDT reduces this artifact. Improved in 75% of cases aneurysm clip (brain) aneurysm coil (brain) dental fillings pacer wire ventricular assist device surgical clip(s) (abdomen) embolization coil(s) (abdomen) bullet(s) / schrapnel / lead shot shoulder replacement unilateral hip replacement bilateral hip replacements knee replacement orthopedic plate(s) femoral neck screw spinal rods Improved in < 75% of cases pedicle screws depth electrodes (brain) cryoablation probes iodinated contrast Table 1. Metal artifact reduction using MDT usually works on smaller implants, but typically results in lower image quality due to decreased resolution for large or long implants (> 5 cm in the axial plane). In general, if the feature of interest can only be seen by looking through metal, then MDT tends to blur it out. Note that MDT works well with femoral neck screws, but not pedicle screws. This is because pedicle screws tend to lie in the axial plane, resulting in loss of resolution, whereas femoral neck screws are angled relative to the axial plane, thus decreasing their length in the axial plane. This table is based on a review of 102 scans. (Author s version) Imaging Med. (2012) 4(2),
12 Out of field artifact Despite popular belief [36, 37], moving an object far outside the field of view does not necessarily create new artifacts. Existing artifacts (such as Poisson noise or metal artifacts) do not change with the field of view. The filter in filtered backprojection is extremely local, meaning that detector measurements far outside the field of view have minimal impact on pixels inside the field of view (Figure 8). Many modern scanners produce bright pixels at the edge of the field of view when the object being scanned extends outside the field of view. This is in fact due to a suboptimal implementation of FBP, and can be fixed with a better reconstruction algorithm (Figure 9). Weight Figure 8. In filtered backprojection, the projection data are filtered to sharpen edges, and the filtered data are then backprojected. The filter (shown above) is extremely local. For example, detector elements ±9 only have a weight of 0.5% relative to detector element 0. This means that detector measurements far outside the field of view have minimal impact on pixels inside the field of view. Detector element number Conclusion Since its introduction in 1972, computed tomography has seen several generations of improvements, including multidetector row helical CT, improved spatial and temporal resolution, dual energy CT, and iterative reconstruction. Many artifacts from the early days of CT are now substantially reduced, but some artifacts remain, and new technologies have introduced new, incompletely characterized artifacts. Remarkable progress has been made in the past few years on iterative techniques for reducing metal artifacts and noise. These techniques not only improve image quality, but also can reduce the radiation dose, improve spatial resolution, and improve diagnosis. However, with iterative reconstruction, noise and image quality are decoupled, which will require new measures of image quality, as well as subjective evaluation. Iterative methods typically have adjustable parameters that control image smoothness, edge preservation, and other features. The effect of these parameters on image quality and noise texture should be studied. Dual energy CT reduces beam hardening, but not scatter. Thus, some dark streaks between high attenuation objects, as well as pseudoenhancement of renal cysts, remain in a dual energy scan. (Author s version) Imaging Med. (2012) 4(2),
13 A B C Reconstructed image Filtered sinogram Sinogram Field of view Figure 9. FBP can reconstruct images acquired using a field of view smaller than the object being scanned. Top row shows the fields of view, second row shows sinograms, third row shows filtered sinograms, and bottom row shows FBP reconstructions. A sinogram is a plot of the projection data (horizontal axis is the tube angle, and vertical axis is the detector number). A. Full field of view. B. Limited field of view, with the sinogram outside the field of view set to zero. This creates a sharp edge, which is amplified by the filter in FBP, creating a bright rim at the edge of the field of reconstruction. This appears to be what many modern CT scanners do. C. Limited field of view, with the sinogram outside the field of view set to the end values in order to prevent discontinuities. This avoids the artifactual bright rim. There is still a small error at the edge of the field of view, which can be reduced using more sophisticated methods [38, 39], or by scanning a slightly larger field of view. (Author s version) Imaging Med. (2012) 4(2),
14 Limited field of view CT (also known as interior CT) enables imaging of a small region of interest inside the body (such as the spine, or tumors) at a lower dose. Future perspective Iterative reconstruction has been studied since the 1970s, but only recently have computer chips become fast enough for their routine clinical use. In commercial scanners, the reconstructions are typically performed using custom chips application specific integrated circuits (ASIC) or field-programmable gate arrays (FPGA). Researchers tend to use the graphics processing unit (GPU) or central processing unit (CPU) in commodity hardware, which is slower than using custom chips, but much cheaper for small numbers of chips, and easier to reprogram [40]. Further improvements in computer power are likely to lead to improved iterative techniques. In particular, more accurate noise and artifact models, as well as cone-beam reconstructions, will require additional calculations. Further advances in CT hardware are also on the horizon. Inverse geometry CT is a new scanner geometry that uses a large array of multiple X-ray sources, and smaller detector array [41, 42], which eliminates cone-beam artifacts and potentially reduces scatter and radiation dose. The highest resolution clinical scanners are flat panel detector (cone beam) scanners with a resolution of 75 µm (Newtom 5G). Scanners with a resolution in the micron range are also known as micro CT scanners. This resolution allows visualization of structures that are not seen on routine clinical CT (Figure 10). However, several issues need to be addressed before this resolution can actually be attained in routine clinical practice. First, high resolution increases noise, which may be acceptable for imaging high contrast structures such as bone, but may obscure soft tissue boundaries. This can be addressed using a higher dose, or by using iterative reconstruction to reduce noise. Second, motion limits resolution, and this can be addressed by motion correction techniques, or with faster tube rotation speed. Laboratory and industrial CT scanners have a resolution as good as 50 nm (Xradia nanoxct). Improved resolution enables visualization of individual cells on pathology specimens [43]. Interestingly, filtered backprojection (but not current iterative techniques) can reconstruct small fields of view using data from tightly collimated beams (Figure 9). This little-known fact could theoretically be used to obtain ultra-high resolution images of specific regions of interest inside the body (spine, tumors, etc) at a lower dose. In addition, it could be used to obtain low dose perfusion images of tumors. Dual energy CT systems scan at two energy levels, which enables beam hardening correction, and produces two Hounsfield unit numbers at each pixel, allowing for greater differentiation of different materials [44]. However, dual energy is not sufficient to capture the full absorption spectrum for example, it does not detect K-edges that are unique to specific materials. In contrast, energy-sensitive photon counting CT [45] measures the full X-ray energy spectrum and thus can be used to detect K-edges, allowing accurate identification of specific materials, such as protein versus hemorrhage [46]. This should also result in improved reduction of beam hardening and scatter artifacts. The main limitation of energy-sensitive photon counting CT is that since each photon must be detected individually, it can currently only be performed at low (Author s version) Imaging Med. (2012) 4(2),
15 dose (20 mas in one study). Iterative methods for noise reduction would be helpful in this application. Although CT is a mature technology, there are many advances still on the horizon. We look forward to seeing what the future brings. Acknowledgements We thank Scott Hsieh, Dan Sze, and Lewis Shin for helpful comments on the manuscript. Mark Riccio from Cornell Imaging provided the micro CT image. Financial disclosure: F.E.B. has a patent pending on the Metal Deletion Technique (MDT) for metal artifact reduction. D.F. has received research support from General Electric Health Care and Siemens Medical Solutions. (Author s version) Imaging Med. (2012) 4(2),
16 Figure 10. Micro CT reveals details of bony trabeculae. A. Micro CT of a dog vertebra at 0.1 mm resolution. The scale bar is 1 cm. Image courtesy of Mark L. Riccio from Cornell Imaging, Cornell University. B. The same scan downsampled to mm resolution, which is a typical resolution for clinical multi-detector row scanners. (Author s version) Imaging Med. (2012) 4(2),
17 References 1. Hsieh J: Computed tomography: Principles, design, artifacts, and recent advances. SPIE, Bellingham, WA. (2003). ** Excellent general reference book on computed tomography. 2. Fleischmann D, Boas FE, Tye GA, Sheahan D, Molvin LZ: Effect of low radiation dose on image noise and subjective image quality for analytic vs iterative image reconstruction in abdominal CT. In: RSNA. Chicago Vandenberghe S, D'Asseler Y, Van de Walle R et al.: Iterative reconstruction algorithms in nuclear medicine. Computerized Medical Imaging and Graphics 25(2), (2001). 4. Fleischmann D, Boas FE: Computed tomography--old ideas and new technology. Eur Radiol 21(3), (2011). 5. Yu Z, Thibault JB, Bouman CA, Sauer KD, Hsieh J: Fast model-based X-ray CT reconstruction using spatially nonhomogeneous ICD optimization. IEEE Trans Image Process 20(1), (2011). 6. Thibault JB, Sauer KD, Bouman CA, Hsieh J: A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34(11), (2007). ** Provides a detailed description of the Model Based Iterative Reconstruction (MBIR) algorithm, which reduces image noise, improves resolution, and reduces helical artifacts Liu X, Primak AN, Krier JD, Yu L, Lerman LO, McCollough CH: Renal perfusion and hemodynamics: Accurate in vivo determination at CT with a 10-fold decrease in radiation dose and HYPR noise reduction. Radiology 253, (2009). 9. Supanich M, Tao Y, Nett B et al.: Radiation dose reduction in time-resolved CT angiography using highly constrained back projection reconstruction. Phys Med Biol 54, (2009). ** Introduces highly constrained backprojection (HYPR) for reducing noise in CT perfusion studies. 10. Joseph PM, Spital RD: The effects of scatter in x-ray computed tomography. Med Phys 9(4), (1982). 11. Herman GT: Correction for beam hardening in computed tomography. Phys Med Biol 24(1), (1979). 12. Joseph PM, Spital RD: A method for correcting bone induced artifacts in computed tomography scanners. J Comput Assist Tomogr 2(1), (1978). 13. Hsieh J, Molthen RC, Dawson CA, Johnson RH: An iterative approach to the beam hardening correction in cone beam CT. Med Phys 27(1), (2000). 14. Alvarez RE, Macovski A: Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol 21(5), (1976). 15. Akbarzadeh A, Ay MR, Ghadiri H, Sarkar S, Zaidi H: Measurement of scattered radiation in a volumetric 64-slice CT scanner using three experimental techniques. Phys Med Biol 55(8), (2010). 16. Ruhrnschopf EP, Klingenbeck K: A general framework and review of scatter correction methods in x-ray cone-beam computerized tomography. Part 1: Scatter compensation approaches. Med Phys 38(7), (2011). (Author s version) Imaging Med. (2012) 4(2),
18 17. Ruhrnschopf EP, Klingenbeck K: A general framework and review of scatter correction methods in cone beam CT. Part 2: Scatter estimation approaches. Med Phys 38(9), (2011). 18. Birnbaum BA, Hindman N, Lee J, Babb JS: Renal cyst pseudoenhancement: influence of multidetector CT reconstruction algorithm and scanner type in phantom model. Radiology 244(3), (2007). 19. Müller J, Vrtiska T, Howe B et al.: The impact of dual energy CT on pseudo enhancement of kidney lesions. Proc. SPIE 7622, 76223I (2010). 20. Yu H, Wang G: Data consistency based rigid motion artifact reduction in fan-beam CT. IEEE Trans Med Imaging 26(2), (2007). 21. Rit S, Wolthaus JW, van Herk M, Sonke JJ: On-the-fly motion-compensated cone-beam CT using an a priori model of the respiratory motion. Med Phys 36(6), (2009). 22. Leschka S, Stolzmann P, Desbiolles L et al.: Diagnostic accuracy of high-pitch dualsource CT for the assessment of coronary stenoses: first experience. Eur Radiol 19(12), (2009). 23. Mahabadi AA, Achenbach S, Burgstahler C et al.: Safety, efficacy, and indications of beta-adrenergic receptor blockade to reduce heart rate prior to coronary CT angiography. Radiology 257(3), (2010). 24. Tang J, Hsieh J, Chen GH: Temporal resolution improvement in cardiac CT using PICCS (TRI-PICCS): performance studies. Med Phys 37(8), (2010). 25. Fleischmann D, Rubin GD, Paik DS et al.: Stair-step artifacts with single versus multiple detector-row helical CT. Radiology 216(1), (2000). 26. Flohr TG, Schaller S, Stierstorfer K, Bruder H, Ohnesorge BM, Schoepf UJ: Multidetector row CT systems and image-reconstruction techniques. Radiology 235(3), (2005). 27. Feldkamp LA, David LC, Kress JW: Practical cone-beam algorithm. J. Opt. Soc. Am. A 1(6), (1984). 28. Boas FE, Fleischmann D: Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 259(3), (2011). ** This paper (by the authors of this review) introduces the Metal Deletion Technique (MDT) for reducing metal artifacts, and shows that it improves image quality and may also affect the diagnosis. 29. De Man B, Nuyts J, Dupont P, Marchal G, Suetens P: Metal streak artifacts in X-ray computed tomography: a simulation study. IEEE Transactions on Nuclear Science 46(3), (1999). 30. Kalender WA, Hebel R, Ebersberger J: Reduction of CT artifacts caused by metallic implants. Radiology 164(2), (1987). 31. Prell D, Kyriakou Y, Kachelrie M, Kalender WA: Reducing metal artifacts in computed tomography caused by hip endoprostheses using a physics-based approach. Invest Radiol 45(11), (2010). 32. Bal M, Spies L: Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering. Med Phys 33(8), (2006). 33. Golden C, Mazin SR, Boas FE et al.: A comparison of four algorithms for metal artifact reduction in CT imaging. Proc. SPIE 7961, 79612Y (2011). 34. Abelson J, Murphy J, Wiegner E et al.: Evaluation of a metal artifact reduction technique in tonsillar cancer delineation. Practical Radiation Oncology 2(1), (2012). (Author s version) Imaging Med. (2012) 4(2),
19 35. Hong R, S. DR, D. LJ, Y. SD: Intravascular ultrasound-guided mesocaval shunt creation in patients with portal or mesenteric venous occlusion. J Vasc Interv Radiol 23(1), (2012). 36. Barrett JF, Keat N: Artifacts in CT: recognition and avoidance. Radiographics 24(6), (2004). 37. Prokop M, Galanski M, Van Der Molen AJ, Schaefer-Prokop C, Telger TC: Spiral and Multislice Computed Tomography of the Body. Thieme, New York. 220 (2003). 38. Hsieh J, Chao E, Thibault J et al.: A novel reconstruction algorithm to extend the CT scan field-of-view. Med Phys 31(9), (2004). 39. Yu H, Wang G: Compressed sensing based interior tomography. Phys Med Biol 54(9), (2009). 40. Yan G, Tian J, Zhu S, Dai Y, Qin C: Fast cone-beam CT image reconstruction using GPU hardware. Journal of X-ray science and technology 16, (2008). 41. Mazin SR, Star-Lack J, Bennett NR, Pelc NJ: Inverse-geometry volumetric CT system with multiple detector arrays for wide field-of-view imaging. Med Phys 34(6), (2007). 42. Schmidt TG, Star-Lack J, Bennett NR et al.: A prototype table-top inverse-geometry volumetric CT system. Med Phys 33(6), (2006). 43. Watz H, Breithecker A, Rau WS, Kriete A: Micro-CT of the human lung: Imaging alveoli and virtual endoscopy of an alveolar duct in a normal lung and in a lung with centrilobular emphysema -- Initial observations. Radiology 236, (2005). * CT of human autopsy lungs at 14 µm resolution shows individual alveoli. 44. Graser A, Johnson TR, Chandarana H, Macari M: Dual energy CT: preliminary observations and potential clinical applications in the abdomen. Eur Radiol 19(1), (2009). 45. Iwanczyk JS, Nygard E, Meirav O et al.: Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging. IEEE Transactions on Nuclear Science 56(3), (2009). ** Shows the first clinical energy-sensitive photon counting CT images, with an energy resolution of 9.8 kev. 46. Boll DT, Patil NA, Paulson EK et al.: Focal cystic high-attenuation lesions: characterization in renal phantom by using photon-counting spectral CT--improved differentiation of lesion composition. Radiology 254(1), (2010). (Author s version) Imaging Med. (2012) 4(2),
How To Improve Your Ct Image Quality
Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo
Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.
1 12/9/2008 Page 1 Master s Program in Medical Physics Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III Chair in Faculty of Medicine Mannheim University of Heidelberg Theodor-Kutzer-Ufer
Purchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage
Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences [email protected] X-Ray
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
Cone Beam Reconstruction Jiang Hsieh, Ph.D.
Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
State-of-the-Art Technology in Cardiac CT
1 2 Next Step Evolution or Revolution? State-of-the-Art Technology in Cardiac CT Stefan Ulzheimer, PhD Global Director of Collaborations CT Siemens Medical Solutions Major Innovations in CT Head The 80
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35
Development of SCENAIA New Version Software Takayuki Kadomura Toshio Sakamoto Kana Tanaka Takatsugu Ito Yuuko Nishimura Naomi Maekawa MEDIX VOL. 62 P.32 P.35 Development of SCENAIA New Version Software
Experimental study of beam hardening artefacts in photon counting breast computed tomography
Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a,
Concepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning Karin M. Andersson,
Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays
J. E. Wilhjelm Ørsted TU Technical University of enmark, Bldg. 348, K-2800 Kongens Lyngby, enmark. X-ray and Computed Tomography Contents History and characterization of X-rays Conventional (projection)
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
DENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
AI CPT Codes. x x. 70336 MRI Magnetic resonance (eg, proton) imaging, temporomandibular joint(s)
Code Category Description Auth Required Medicaid Medicare 0126T IMT Testing Common carotid intima-media thickness (IMT) study for evaluation of atherosclerotic burden or coronary heart disease risk factor
A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology
A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami
P R E S E N T S Dr. Mufa T. Ghadiali is skilled in all aspects of General Surgery. His General Surgery Services include: General Surgery Advanced Laparoscopic Surgery Surgical Oncology Gastrointestinal
Data. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology [email protected] http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
MEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
Page: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
American College of Radiology CT Accreditation Program. Testing Instructions
American College of Radiology CT Accreditation Program Testing Instructions (Revised July 24, 2015) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
Gated Radiotherapy for Lung Cancer
Gated Radiotherapy for Lung Cancer Steve B. Jiang, Ph.D. Depart Of Radiation Oncology University of California San Diego [email protected] radonc.ucsd.edu/research/cart Two Types of Gating Internal gating
CT Image Reconstruction. Terry Peters Robarts Research Institute London Canada
CT Image Reconstruction Terry Peters Robarts Research Institute London Canada 1 Standard X-ray Views Standard Radiograph acquires projections of the body, but since structures are overlaid on each other,
Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose
Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose ipatient is an advanced platform that delivers focused innovations to facilitate patient-centered
Development of High Speed Oblique X-ray CT System for Printed Circuit Board
計 測 自 動 制 御 学 会 産 業 論 文 集 Vol. 6, No. 9, 72/77 (2007) Development of High Speed Oblique X-ray CT System for Printed Circuit Board Atsushi TERAMOTO*, Takayuki MURAKOSHI*, Masatoshi TSUZAKA**, and Hiroshi
Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications
Safety Issue 3.0 T MRI Pro 3.0 T MRI Con Immediate fringe field surrounding magnet A ferromagnetic object inadvertently brought into the scan room will experience a sharp increase in attraction toward
Patient Prep Information
Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
Computed Tomography, Head Or Brain; Without Contrast Material, Followed By Contrast Material(S) And Further Sections
1199SEIU BENEFIT AND PENSION FUNDS High Tech Diagnostic Radiology and s # 1 70336 Magnetic Resonance (Eg, Proton) Imaging, Temporomandibular Joint(S) 2 70450 Computed Tomography, Head Or Brain; Without
An iterative approach to the beam hardening correction in cone beam CT
An iterative approach to the beam hardening correction in cone beam CT Jiang Hsieh a) Applied Science Laboratory, GE Medical Systems, Milwaukee, Wisconsin 53201 Robert C. Molthen Department of Biomedical
Diagnostic Imaging Prior Review Code List 3 rd Quarter 2016
Computerized Tomography (CT) Abdomen 6 Abdomen/Pelvis Combination 101 Service 74150 CT abdomen; w/o 74160 CT abdomen; with 74170 CT abdomen; w/o followed by 74176 Computed tomography, abdomen and pelvis;
Magnetic Resonance Imaging
Magnetic Resonance Imaging North American Spine Society Public Education Series What Is Magnetic Resonance Imaging (MRI)? Magnetic resonance imaging (MRI) is a valuable diagnostic study that has been used
CPT CODE PROCEDURE DESCRIPTION. CT Scans 70450 CT HEAD/BRAIN W/O CONTRAST 70460 CT HEAD/BRAIN W/ CONTRAST 70470 CT HEAD/BRAIN W/O & W/ CONTRAST
CPT CODE PROCEDURE DESCRIPTION CT Scans 70450 CT HEAD/BRAIN W/O CONTRAST 70460 CT HEAD/BRAIN W/ CONTRAST 70470 CT HEAD/BRAIN W/O & W/ CONTRAST 70480 CT ORBIT W/O CONTRAST 70481 CT ORBIT W/ CONTRAST 70482
MRI for Paediatric Surgeons
MRI for Paediatric Surgeons Starship David Perry Paediatric Radiologist Starship Children s Hospital CHILDREN S HEALTH What determines the brightness of a pixel in MRI? i.e. What determines the strength
GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time
Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs
THE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
CPT Radiology Codes Requiring Review by AIM Effective 01/01/2016
CPT Radiology Codes Requiring Review by AIM Effective 01/01/2016 When a service is authorized only one test per group is payable. *Secondary codes or add-on codes do not require preauthorization or separate
International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters
www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in
PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Medical and Industrial Tomography - W.B.Gilboy
MEDICAL AND INDUSTRIAL TOMOGRAPHY Department of Physics, University of Surrey, Guildford, Surrey, U.K. Keywords: Radiography, transmission tomography, emission tomography, microtomography, SPECT (single
X-ray (Radiography) - Chest
Scan for mobile link. X-ray (Radiography) - Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray produces images of
CMS Manual System Pub. 100-03 Medicare National Coverage Determinations
CMS Manual System Pub. 100-03 Medicare National Coverage Determinations Department of Health & Human Services (DHHS) Centers for Medicare & Medicaid Services (CMS) Transmittal 21 Date: September 10, 2004
Procedure Codes. RadConsult provides real-time decision support for physicians who order high-cost imaging procedures RADIATION THERAPY
Procedure Codes 2011 RadConsult provides real-time decision support for physicians who order high-cost imaging procedures RADIATION THERAPY 2D3D Therapeutic radiology treatment planning; simple 77261 Therapeutic
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
The File-Card-Browser View for Breast DCE-MRI Data
The File-Card-Browser View for Breast DCE-MRI Data Sylvia Glaßer 1, Kathrin Scheil 1, Uta Preim 2, Bernhard Preim 1 1 Department of Simulation and Graphics, University of Magdeburg 2 Department of Radiology,
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems GE s Industrial X-ray and CT Forum, Cologne July 7 th 2015 Oliver Brunke GE Sensing & Inspection Technologies
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
DICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
Twelve. Figure 12.1: 3D Curved MPR Viewer Window
Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from
Embedded Systems in Healthcare. Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008
Embedded Systems in Healthcare Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008 About the Speaker Working for Philips Research since 1982 Projects
Two years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions
Two years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions Laurent Desbat TIMC-IMAG, UMR 5525, Grenoble University, France [email protected] Rolf
GE Healthcare. Revolution EVO. More than just high tech. Higher purpose.
GE Healthcare Revolution EVO More than just high tech. Higher purpose. Revolution EVO. Designed with purpose. Today s healthcare environment is about creating new solutions to pressing needs. It s about
Volume visualization I Elvins
Volume visualization I Elvins 1 surface fitting algorithms marching cubes dividing cubes direct volume rendering algorithms ray casting, integration methods voxel projection, projected tetrahedra, splatting
Digital Dental X-ray. Midmark Animal Health Products
Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't
The Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO
The Whys, Hows and Whats of the Noise Power Spectrum Helge Pettersen, Haukeland University Hospital, NO Introduction to the Noise Power Spectrum Before diving into NPS curves, we need Fourier transforms
Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.
www.neurologica.com Critical Care Clinic Operating Room Radiology ED Trauma OR Neuro/Spine Pediatric Plastics Thoracic General Interventional Radiology ENT ED Trauma Trauma Stroke Critical Care Neuro Pediatric
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
CPT * Codes Included in AIM Preauthorization Program for 2013 With Grouper Numbers
CPT * Codes Included in AIM Preauthorization Program for 2013 With Grouper Numbers Computerized Tomography (CT) CPT Description Abdomen 74150 CT abdomen; w/o contrast 6 74160 CT abdomen; with contrast
Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015
INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design
76641 Ultrasound, breast, unilateral, real time with image documentation, including axilla when performed; complete 76642 limited
2015 CPT Code Update The ACR, either alone or in conjunction with other specialty societies, worked on a number of code proposals for the 2015 code cycle. This update provides a listing of code changes
THORACIC DIAGNOSTIC ASSESMENT PROGRAM (DAP) PATIENT INFORMATION FOR:
central east regional cancer program in partnership with cancer care ontario THORACIC DIAGNOSTIC ASSESMENT PROGRAM (DAP) PATIENT INFORMATION FOR: Thoracic dap booklet March2012.indd 1 SCHEDULED TESTS YOUR
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
Characterization of small renal lesions: Problem solving with MRI Gary Israel, MD
Characterization of small renal lesions: Problem solving with MRI Gary Israel, MD With the widespread use of cross-sectional imaging, many renal masses are incidentally found. These need to be accurately
Image-based simulation of the human thorax for cardio-pulmonary applications
Presented at the COMSOL Conference 2009 Boston Image-based simulation of the human thorax for cardio-pulmonary applications F. K. Hermans and R. M. Heethaar, VU University Medical Center, Netherlands R.
Multi Detector Row CT Systems and Image- Reconstruction Techniques 1
Special Review Radiology Thomas G. Flohr, PhD Stefan Schaller, PhD Karl Stierstorfer, PhD Herbert Bruder, PhD Bernd M. Ohnesorge, PhD U. Joseph Schoepf, MD Published online before print 10.1148/radiol.2353040037
The Olympus stereology system. The Computer Assisted Stereological Toolbox
The Olympus stereology system The Computer Assisted Stereological Toolbox CAST is a Computer Assisted Stereological Toolbox for PCs running Microsoft Windows TM. CAST is an interactive, user-friendly,
In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer
In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer Prostate cancer is the second most common cancer in men worldwide, accounting for 15% of all new cancer cases. 1 Great strides have
Computerized Tomography, X-rays, and the Radon Transform
1 Computerized Tomography, X-rays, and the Radon Transform 1.1 Introduction The purpose of this chapter is to give an informal introduction to the subject of tomography. There are very few mathematical
Acknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group
Cone Beam CT guided Radiotherapy Jan-Jakob Sonke Acknowledgements NKI-AVL: Marcel van Herk, Jose Belderbos, Suzanne van Beek, Anja Betgen, Josien de Bois, ~Rianne de Jong, Michel Frenay, Danny Minkema,
College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries
2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,
Computer Aided Engineering (CAE) Techniques Applied To Hip Implant
International Journal Of Computational Engineering Research (ijceronline.com) Vol. 3 Issue. 3 Computer Aided Engineering (CAE) Techniques Applied To Hip Implant 1, M. S. Abo_Elkhair, 2, M. E. Abo-Elnor,
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
Neuro imaging: looking with lasers in the brain
Neuro imaging: looking with lasers in the brain Aim: To image life cells, label free, with cellular resolution in deep tissue Marloes Groot Vrije Universiteit Amsterdam Faculteit Exacte Wetenschappen Natuurkunde
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905
The phantom portion of the American College of Radiology ACR Computed Tomography CT accreditation program: Practical tips, artifact examples, and pitfalls to avoid a Cynthia H. McCollough b) and Michael
Radiation Protection in Radiotherapy
Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France [email protected] Radiation therapy The lecture is oriented to
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
Image Analysis for Volumetric Industrial Inspection and Interaction
Image Analysis for Volumetric Industrial Inspection and Interaction Rasmus Larsen Mads F. Hansen, Hildur Olafsdottir, Thomas H. Mosbech, Lasse F. Laursen DTU Informatics (CT) reconstruction from few projections