Multislice Computed Tomography: Basic Principles and Clinical Applications

Size: px
Start display at page:

Download "Multislice Computed Tomography: Basic Principles and Clinical Applications"

Transcription

1 New Applications with the Assistance of Innovative Technologies Multislice Computed Tomography: Basic Principles and Clinical Applications A. F. Kopp, K. Klingenbeck-Regn 2, M. Heuschmid, A. Küttner, B. Ohnesorge 2, T. Flohr 2, S. Schaller 2, C. D. Claussen Eberhard-Karls-University Tübingen, Department of Diagnostic Radiology, Tübingen, Germany 2 Siemens AG, Medical Engineering Group, Forchheim, Germany 94 Introduction Since its clinical introduction in 99, volumetric CT scanning using spiral or helical scanners has resulted in a revolution for diagnostic imaging. In addition to new applications for CT, such as CT angiography and the assessment of patients with renal colic, many routine applications such as the detection of lung and liver lesions have substantially improved. Helical CT has improved over the past eight years with faster gantry rotation, more powerful x-ray tubes, and improved interpolation algorithms [, 2]. However, in practice the spiral data sets from monoslice systems suffered from a considerable mismatch between the transverse (in plane) and the longitudinal (axial) spatial resolution. In other words the isotropic 3-dimensional voxel could not be realized apart from some very specialized cases [3]. Similarly, in routine practice a number of limitations still remained which prevented the scanning protocol from being fully adapted to the diagnostic needs [4]. The introduction of subsecond spiral scanning with the SOMATOM Plus 4 at the RSNA 994 was a first step to facilitate routine clinical work with respect to scannable volumes, total scan time and axial resolution [5]. Compared the sec scanners which were standard at that time, the 750 ms rotation time allowed one to scan 33% longer volumes or to correspondingly reduce the total scan time or to correspondingly improve axial resolution. The latest advance has been the recent introduction of multislice CT (MSCT) scanners. At the RSNA 998, this new technology was introduced by several manufacturers representing an obvious quantum leap in clinical performance [6-8]. Currently capable of acquiring four channels of helical data simultaneously, MSCT scanners have achieved the greatest incremental gain in scan speed since the development of helical CT and have profound implications for clinical CT scanning. Fundamental advantages of MSCT include substantially shorter acquisition times, retrospective creation of thinner or thicker sections from the same raw data, and electromedica 68 (2000) no. 2 improved three-dimensional rendering with diminished helical artifacts [9]. For example, the SOMATOM Volume Zoom with a 500 ms rotation time and the simultaneous acquisition of 4 slices offers an 8-fold increase of performance compared to previous s, single-slice scanning. Obviously, such a quantum leap opens up a new area in spiral CT affecting all existing applications and allowing the realization of new clinical applications. The key issue is correspondingly increased volume coverage per unit time at high axial resolution and a correspondingly improved temporal resolution [6]. In general terms, the capabilities of spiral CT can be expanded in various ways: to scan anatomical volumes with standard techniques at significantly reduced scan times; or to scan larger volumes previously not accessible in practical scan times; or to scan anatomical volumes with high axial resolution (narrow collimation) to closely approach the isotropic voxel of high-quality data sets for excellent 3-dimensional postprocessing and diagnosis. Basic Principles For discussion of spiral imaging the following definition of pitch is commonly used: Table travel per rotation P = Collimation of single slice With this definition the present standard range p 2 for single slice systems is extended to p 8. In Fig. various sampling patterns of a 4-slice spiral are shown for representative values of the pitch. In this schematic view the projections are symbolized by single arrows, which for simplicity are drawn in parallel. From those examples we can draw the following conclusions:

2 Rot. Rot. 2 Rot. 3 Rot. 4 Rot. Rot. 2 Rot. 3 Rot. 4 Rot. Rot. 2 Rot. 3 Rot. 4 Pitch Pitch 2 Rot. Rot. 2 Rot. 3 Pitch.5 Pitch 4 Figure Sampling Patterns of a 4-slice spiral scan at different pitch values. At pitch and 2, each z-position is sampled 4 and 2 times respectively. The spacing between samples decreases from d to d/2 when going from pitch to pitch.5, then increases again to d when increasing the pitch to 2. At a pitch of 4, each sample is acquired once and the sampling distance is d (d denotes the slice collimation).. For pitch values smaller than 4 the four slices overlap to a certain degree after one rotation. Pitch 2 is a transparent case with double sampling. Therefore in this regime p 4 the multiple sampling can be used to reduce the tube current for a desired image noise and for a desired patient dose, respectively. 2. However, collecting data from multiple rotations degrades the temporal resolution of the system. Therefore for imaging of moving organs with high image quality, a pitch smaller than 4 should be avoided. 3. The distance between neighboring samples varies with pitch periodically. Consequently, 80 LI and 360 LI spiral interpolators would yield a corresponding nonmonotonic dependence of slice width on pitch and are therefore not useful for practical purposes. 4. The distance between neighboring samples never exceeds the slice collimation up to a pitch of 8. This opens up the possibility to realize a slice width independent of pitch up to a pitch of 8 and to completely eliminate broadening of the slice width. Design Considerations for MSCT It is easy to design a multi-slice scanner for a fixed slice collimation, the challenge is to design the detector in such a way as to meet the clinical requirement of different slice collimations adjustable to the diagnostic needs. There are basically two different approaches, the matrix detector with elements of a fixed size or the adaptive array principle. Both principles will be briefly described and compared. An example of a matrix detector is sketched in Fig. 2. In axial direction the detector is divided into 6 small elements each providing a mm thick slice at the center of rotation [0]. The outer detector rows cannot be used individually. In the example of Fig. 2 the mm slice is smeared over about 6 mm. The only way out is to sum the signals of the outer rows to generate thick slices. However, the unnecessary mechanical cuts and FOV sw: 4.0 mm Figure 2 Fixed Array Detector. The slice width is compared to the smearing of the slice caused by the cone-angle. It is shown that for the example of a 6-row detector, the outermost slice of nominal thickness 4.7 mm 6.6 mm 6 detector rows sw:.0 mm.0 mm is broadened to 6.6 mm by smearing (right half of figure). The problem can be overcome by combining several slices to a wider slice (left half of figure). Then, however, separators between the slices are not needed for the outer slices. electromedica 68 (2000) no z

3 New Applications with the Assistance of Innovative Technologies 96 optical separations between the small elements, correspondingly reduce the geometrical efficiency and therefore the dose efficiency of the system. In summary, the matrix detector is well suited to scan at 4 x mm collimation but not more than 4. Wider collimations 4 x 2 mm, etc. can be realized by signal combination during read out but at the expense of dead zones and a corresponding waste of dose. These arguments lead to the development of the Adaptive Array Detector [9]. The design of the Adaptive Array Detector (AAD) takes into account the cone beam constraints for optimal image quality, optimizes the dose efficiency and in conjunction with an Adapted Axial Interpolator (AAI) provides a flexible selection of slice widths [6,, 2]. The design principle is depicted in Fig. 3. Narrow detector elements are close to the center; the width of the detector rows increases with distance from the center. Unnecessary dead spaces are avoided and with the corresponding prepatient collimator and the proper read-out schemes the following combinations of collimation can be achieved: 2 x 0.5 mm, 4 x mm, 4 x 2.5 mm, 4 x 5 mm, 2 x 8 mm and 2 x 0 mm (Fig. 4). These combinations represent the collimation of the X-ray beam at center: e. g. a 4 x 5 mm collimation means a X-ray beam width at center of 20 mm. Consequently with sequential imaging four 5 mm slices would be generated during one rotation. In spiral imaging the variety of axial sampling patterns as a function of pitch allows both: obtaining slice widths independent of pitch and reconstructing a multiplicity of slice widths from a scan with collimation narrower than s. To achieve this the AAI-scheme provides a set of linear and nonlinear interpolators which are adapted to the desired pitch, collimation and slice width. To give some examples: from a spiral scan with 4 x mm collimation slice widths of,.25,.5, 2.0, 2.5, 3.0, 4.0, 5.0 up to 0 mm can be obtained by adjusting the width and the functional form of the interpolator. On the other hand, a selected slice width s, like s = 5 mm, may be obtained from different collimator settings, like 4 x mm or 4 x 2.5 mm. This is important to remember for clinical applications, as the narrower collimation is preferable for image quality reasons, i. e. the reduction of partial volume effects [9]. Yet another design criterion should be emphasized: spiral scanning at large pitch, e. g. large table velocity, implies a more severe inconsistency between direct projections and the complementary projections, taken half a rotation later with a quarter detector offset. The reason is changing anatomy in axial direction, which is particularly pronounced for large pitch applications, i. e. rapid table movement. Consequently, image quality at high spatial resolution and large pitch is becoming more dependent on the flying spot technology which provides a quasi-instant doubling of the in plane sampling rate. electromedica 68 (2000) no. 2 2 x 0.5 mm 4 x.0 mm 4 x 2.5 mm 2 x 8.0 mm 4 x 5.0 mm z Magnification center to detector by a factor of approximately 2 in mm appr. 40 mm 5 Figure 3 Design of the Adaptive Array Detector. The physical width of the detector is approximately 40 mm. Slice widths at the axis of rotation range from mm for the inner slices to 5 mm for the outer slices. Figure 4 Available Collimations and read-out schemes for the Adaptive Array Detector (AAD). The dotted bar indicates the collimation at the detector. Prepatient collimation is adjusted correspondingly.

4 In conclusion, image quality in multi-slice spiral scanning must be optimized with respect to several factors [3]:. The narrowest collimation, consistent with the coverage of a certain volume and with a certain scan time, to minimize partial volume effects and to optimize image quality. 2. Fastest rotation time to maximize z-coverage and to minimize motion blurring. 3. Pitch greater than 4 to preserve temporal resolution and to minimize motion blurring. 4. The exploitation of flying focal spot technology to avoid artifacts at high spatial resolution. To measure section profiles of multi-spiral scanning a thin gold plate (thickness 50 µm) in air has been aligned orthogonal to the scanner axis and has been scanned in spiral mode. Some of the resulting slice sensitivity profiles (SSP) are shown in Fig. 5 for a 4 x mm collimation, a slice width of 2 mm and different pitch values. Obviously the AAI indeed results in slice widths independent of pitch; but even more important, the functional form of the SSPs is also identical and practically independent of pitch. Slice broadening and long-range tails of the SSP, which prohibited the use of fast table speeds in single slice spiral scanning, can be completely avoided. From this perspective the whole range up to a pitch of 8 can be used for practical purposes in multi-slice scanning [6]. The second new and attractive feature of the multispiral AAI is illustrated in Fig. 6. From a scan with 4 x mm collimation SSPs with different width can be obtained:.25 mm, 2.0 mm and 4.0 mm respectively. Excellent agreement between theory and measurement is observed. This feature is the basis of Combi Scans. This provides considerable flexibility in image reconstruction [4] especially for imaging of the base of the skull and lung (Fig. 7). Anatomical structures that generate spiral artifacts are well known from single slice spirals. The most difficult situations arise from bony structures (high Pitch 3 Measured: -FWHM = 2.05 mm Pitch 5 FWHM = 2. mm Pitch 7 FWHM = 2. mm Figure 5 Slice Sensitivity Profiles, Collimation 4 x mm, slice width 2 mm, different pitch values : calculated : measured z [SW] Measured: -FWHM =.35 mm FWHM = 2.05 mm FWHM = 4.0 mm Figure 6 Slice Sensitivity Profiles, Collimation 4 x mm, pitch 3, different slice widths : calculated : measured electromedica 68 (2000) no. 2 97

5 New Applications with the Assistance of Innovative Technologies a b c d Figure 7 Combi Scan: lung study with 4 x mm collimation, pitch 6. Reconstruction of 5 mm images for soft tissue (a) and standard lung window (b);.25 mm high resolution images for detection of interstitial disease (c). HR-MPR (d) clearly depicts segmental anatomy. 98 electromedica 68 (2000) no. 2

6 contrast) which are strongly inhomogeneous in axial direction. A demanding example is the base of the skull with bony structures abruptly ending in an image plane. Phantom studies have shown that a larger pitch and narrower collimation is much more favorable for suppressing artifacts than a lower pitch and wider collimation for equal z-coverage. The underlying reason is the better elimination of partial volume artifacts. From image quality reasons, only, the strength of multislice spiral scanning is the ability to cover anatomical volumes with narrowest collimation and thereby to minimize partial volume effects. This is only a matter of scanning technique (collimation) and is independent of the selected slice width. For optimization of image quality we derive the rule: the narrowest collimation should be selected which is consistent with volume coverage and scan time. This generally results in large pitch values (e. g. from 4 to 6) which are also helpful to avoid motion blurring [3]. Clinical Applications The advantages of MSCT are important to many applications of CT scanning, including survey exams in oncologic or trauma patients and the characterization of focal lung and liver lesions through the creation of thin sections retrospectively. However, the greatest impact has been on CT angiography, cardiac imaging, virtual endoscopy, and high resolution imaging [5]. CT Angiography A fundamental advantage of a multislice scanner over monoslice systems is its ability to obtain a first circulation study of a rapidly injected contrast bolus with thinner images. Determining circulation time either by a preliminary minibolus or by online bolus tracking software is important in matching the acquisition interval to the first-circulation time of the injected bolus. As a result of the shorter acquisition time, the contrast dose can be significantly reduced. CT angiography of the intracranial vessels benefits from the quick and detailed scanning technology of MSCT. At mm-collimation, the circle of Willis can be scanned within 0 seconds. This permits the entire scan to be completed during the first pass of iodinated contrast material through the arterial system. The use of MSCT with CT angiography demonstrates the spatial relationship of an aneurysm with the feeding vessel, as well as the shape of the aneurysm itself, because the same bolus can be followed continuously throughout its course. For CT angiography of the thoracoabdominal aorta a collimation of 2.5 mm, a pitch of 5-6 at a rotation speed of 0.5 s are used. Volume coverage is from the thoracic inlet to the inguinal region, a 50 to 55 cm area that can be covered in an acquisition interval of approx. 20 s (Fig. 8). Reconstructions with 50% overlap are used, generating images for the threedimensional data set. A complete lower extremity study a Figure 8 MSCTA of the thoracoabdominal aorta. Stanford Type B Dissection. MPRs from images with 3 mm slice-width and mm increment. The spiral scan was acquired with 4 x 2.5 mm collimation, pitch 5, and 0.5 s rotation time. The total scan time for a spiral length of 550 mm was 22 s. Courtesy of Dr. Baum, Erlangen. electromedica 68 (2000) no. 2 99

7 New Applications with the Assistance of Innovative Technologies can be performed from the level of the renal vascular pedicles to the ankles. A 2.5 mm collimation, a pitch of 6 and 50% overlap reconstructions are used (Fig. 9). A first-pass circulation study can be obtained without venous overlay. The total number of images generated is 800 to 000. With multislice CT operating at pitch 6 and collimation of 4 x mm, a CTA of the pulmonary arteries can be acquired in less than 25 s. Image thickness of.25 mm significantly increases detectability of subsegmental emboli in comparison to monoslice spiral CT using 2 to 3 mm image thickness [6]. CTA of the visceral branch vessels can be performed significantly faster and/or with increased spatial resolution. Using a Figure 9 Peripheral Runoffs. Spiral scan with 4 x 2.5 mm collimation, pitch 6, 0.5 s rotation time. The total scan time for a spiral length of 000 mm was 34 s. Left: MIP from images with 3 mm slice width, 2 mm increment. Right: VRT from images with 6 mm slice width, 4 mm increment. a 00 electromedica 68 (2000) no. 2 b Figure 0 CTA (a, b) of visceral branch vessels (collimation 4 x mm, pitch 5, 20 cc contrast). c On the VR images a stenosis in the SMA (arrows) can be readily appreciated which was confirmed at conventional angiography (c).

8 collimation of mm allows for detection of even minor branches (Fig. 0). The ability to cover the entire chest in0 s allows the scanning of children without sedation. This can be used to easily perform CTA of the large thoracic vessels before or after surgery of complex malformations [8]. Cardiac Imaging Electron Beam CT scanning (EBCT) has been established as a non-invasive imaging modality for imaging coronary calcification. Major clinical applications are the detection and quantification of coronary calcium and non-invasive CT angiography (CTA) of the coronary arteries [7]. Current limitations of EBCT imaging include the limited reproducibility of coronary calcium quantification [8], the inability to detect noncalcified atherosclerotic plaques and the limited spatial resolution of 3D visualizations of the coronary arteries. Because of the restriction to axial, non-spiral scanning in ECG-synchronized cardiac investigations, acquisition of 3D volume images by using EBCT can only provide limited z-resolution within a single breath-hold scan. Retrospectively ECG-gated single-slice spiral scanning does not allow for sufficient continuous volume coverage within reasonable scan times. Retrospectively ECG-gated multi-slice spiral scanning, however, has the potential to completely cover the heart volume without gaps within one breath-hold [9]. Mechanical multislice CT systems with simultaneous acquisition of four slices, half-second scanner rotation and 25 ms maximum temporal resolution allows for considerably faster coverage of the heart volume, compared to singleslice scanning. This increased scan speed allows using thinner collimated slice widths and thus to increasing the z-resolution of high-resolution examinations such as CTA of the coronary arteries [20]. ECG-synchronized multi-slice spiral scans are acquired with heart rate dependent table feed ( pitch ) adaptation. Dedicated spiral algorithms provide 25 ms (60 ms as theoretical limit) temporal resolution and are optimized with regard to volume coverage. This allows reconstruction of overlapping images (increment < slicewidth) at arbitrary z-positions and during any given heart phase [2]. This reconstruction technique combines partial scan reconstruction and multi-slice spiral weighting in order to compensate for table movement and to provide a well-defined slice sensitivity profile [9]. For retrospectively ECG-gated reconstruction each image is reconstructed using a multi-slice partial scan data segment with an arbitrary temporal relation to the R-wave of the ECG-trace. Image reconstruction during different heart phases is feasible by shifting the start point of image reconstruction relative to the R-wave. For a given start position, a stack of images at different z-positions covering a small sub-volume of the heart can be reconstructed owing to mutlislice data acquisition Figure Reconstruction with retrospectively ECG-gated 4-slice spiral scanning. t [sec] (0.5 sec/rotation) Slice q = 0 Slice q = Slice q = 2 Slice q = 3 ECG-Signal (~ 70 bpm) Data ranges are selected with certain phase relation to the RR-intervals. 3D volume images are generated from image stacks reconstructed in consecutive heart cycles. Fig. shows an example of how the cardiac volume is successively covered with stacks of axial images (shaded stacks) reconstructed in consecutive heart cycles. All image stacks are reconstructed at identical time-points during the cardiac cycle. At the same time, the 4 detector slices travel along the z-axis relative to the patient table. In each stack, single-slice partial scan data segments are generated with equidistant spacing in the z-direction depending on the selected image reconstruction increment [2]. Continuous volume coverage can only be achieved, when the spiral pitch is adapted to the heart rate in order to avoid gaps between image stacks that are reconstructed using data from different heart cycles. In order to achieve full volume coverage, the image stacks reconstructed in subsequent heart cycles must cover all z-positions. Thus, the pitch, which can be used for image acquisition is limited by the patient s RR-interval time [9]. The cardiac multi-slice data acquisition on the SOMATOM Volume Zoom are performed using 500 ms full rotation time and 4 x mm or 4 x 2.5 mm collimated slice width. Non-contrast enhanced spiral scans for coronary calcium scoring are performed with 3 mm slice-width (SW = 3 mm, SW coll = 2.5 mm) and mm image reconstruction. For CTA of the coronary arteries and for functional heart imaging 2 different scan protocols with 3 mm slice-width and with.25 mm slice-width can be used (Fig. 2). For both protocols, non-ionic contrast material is intravenously injected at a flow rate of 3 ml/s. The delay times between start of contrast injection and scan start for optimal contrast enhancement is determined individually for each patient by injection of a 20 ml test bolus [20]. electromedica 68 (2000) no. 2 0

9 New Applications with the Assistance of Innovative Technologies a a Figure 2 49-year-old male with CHD of the RCA. S/P posterolateral myocardial infarction 4 months ago. Patient underwent balloon angioplasty of subtotal stenosis at the beginning of the descending part of the RCA. Patient came in for follow-up performed with both coronary CT angiography (a) and conventional angiography (b). Both conventional angiography and CT angiography (3D Virtuoso, Siemens) clearly show the patency of the RCA at the level of the ballon angioplasty. There is only minor residual stenosis of approx. 0-20% (arrows). Scan: Siemens SOMATOM Volume Zoom; 40 kv, 300 mas, collimation 4 x mm; slice width.25 mm, increment 0.6 mm. b Figure 3 Noninvasive CT coronary angiography: Volume rendered image (3D Virtuoso, Siemens) depicts three high grade stenoses (arrows) in the LAD and at the origin of the diagonal branches. These findings are confirmed at conventional angiography. b 02 electromedica 68 (2000) no. 2

10 The high spatial resolution, the absence of motion artifacts and the good overall image quality of the clinical MSCT images of the entire heart volume let appear multislice spiral CT as a promising modality for the non-invasive evaluation of coronary calcification. The scan time needed to acquire continuous ECG-gated multislice spiral CT image data is significantly reduced compared to EBCT ( factor 2.5) and single slice CT ( factor 5). Data with 3 mm slice width can be used for volumetric coronary calcium scoring. This alternative scoring method has the potential to improve the reproducibility of repeat calcium scoring compared to the conventional Agatston-score. Phantom studies have shown that non-overlapping sequential scanning is an important contributor to the inter-scan variability of Agatston- and volumetric Ca-scores due to partial volume errors in plaque registration [22]. ECG-gated volume coverage with multi-slice spiral CT and overlapping image reconstruction, however, was found to improve the reliability of coronary calcium quantification especially for small plaques. ECG-gated multislice spiral CT can potentially be of high value for coronary calcium evaluation especially for patients undergoing lipid-lowering statin therapy and for followup evaluations of patients after heart transplantation [23]. In contrast to sequential CT scanning z-resolution of ECG-gated spiral images with 3 mm slice-width can be improved by using overlapping reconstruction with mm slice increment. Moreover, the fast scan speed allows covering the entire heart with.25 mm slices within a single breath-hold (0 cm in s). 3D reconstruction with.25 mm slice-width and sub-millimeter image increment allows generating highresolution visualizations of the coronary arteries, which may be suitable for a highly accurate diagnosis of coronary artery disease (Fig. 3, 4) [24]. Even more important, the first results indicate that MSCTA not only allows non-invasive imaging of coronary plaques but also assessment of plaque composition (i. e., soft, fibrous, calcified) (Fig. 5). Thus, this new technology holds promise to allow for the non-invasive imaging of rupture-prone soft coronary lesions and may have the option to lead to early onset of therapy [25]. High Resolution Imaging Imaging of the temporal bone is a major challenge for clinical CT and a good example of the need for high resolution CT. Imaging of the temporal bone is improved using MSCT because the in-plane resolution can be greatly increased. The temporal bone is a structure of high intrinsic contrast and is routinely scanned with thin collimation. Both the 2 x 0.5 mm and the 4 x mm mode on the SOMATOM Volume Zoom are applicable (pitch 2 or 3.5) and the tube current can be reduced to below 200 mas (Fig. 6). The reconstruction kernels enable maximum spatial resolution up to 24 line pairs/ Figure 4 Virtual angioscopy (bottom right) of circumflex branch of left coronary artery. VRT (left) and MPR (top right) images help navigating the virtual endoscope in the coronary artery (3D Virtuoso, Siemens). 4 5 Figure 5 MSCTA of LAD: Non-calcified soft plaque (arrow) with density of approx. 5 HU. This lipid-core plaque was classified as prone to rupture in intracoronary ultrasound. Scan: collimation 4 x mm, pitch.5, 300 mas, 20 cc contrast. electromedica 68 (2000) no. 2 03

11 New Applications with the Assistance of Innovative Technologies 04 cm. At this resolution, the delicate structures of the middle and inner ear are sharply delineated and even subtle changes can be assessed. Another advantage of using MSCT to image the temporal bone is that the slice thickness is reduced to mm or below, thus minimizing the partial volume effects and further increasing image quality of subtle structures [26]. Impact on Radiological Practice The large number of images generated by MSCT is a major issue for workstation performance, film display, and PACS archiving. The images should be reviewed in a cine paging format at a workstation for diagnostic purposes [5]. Film hard copy will only be used for reference display. Planning for PACS must take into account this exponential growth in image data generated by this new technology. Rapid generation of 3D data sets is important for inclusion in the diagnostic study interpretation and for demonstration to referring clinicians. The hundreds or thousands of thin sections acquired using MSCT are incompatible with traditional viewing practice; thus MSCT will force a rapid transition of radiology from two-dimensional to volumetric imaging. electromedica 68 (2000) no. 2 Figure 6 High resolution image of the temporal bone. Note the excellent definition of the ossicles at a collimation thickness of 0.5 mm (200 mas, 20 kv). Literature [] Crawford, C. R., King, K. F.: Computed tomography scanning with simultaneous patient translation. Med.Phys. 990; 7: [2] Kalender, W. A., Seissler, W., Klotz, E., Vock, P.: Spiral volumetric CT with single-breath-hold technique, continuous transport and continuous scanner rotation. Radiology 990; 76:8-3. [3] Kalender, W. A.: Thin-section threee-dimensional spiral CT: is isotropic imaging possible? Radiology 995; 97: [4] Brink, J. A., Heiken, J. P., Balfe, M., Sagel, S. S., DiCroce, J.: Decreased spatial resolution in vivo due to broadening of sectionsensitivity profile. Radiology 992; 85: [5] Costello, P.: Subsecond scanning makes CT even faster. Diagnostic Imaging 996; 8:76-9. [6] Schaller, S., Flohr, T., Wolf, H., Kalender, W. A.: Evaluation of a Spiral Reconstruction Algorithm for Multirow-CT. Radiology 999; 209 (P):434. [7] Taguchi, K., Aradate, H.: Algorithm for image reconstruction in multi-slice helical CT. Med.Phys. 998; 25: [8] Fox, S. H., Tanenbaum, L. N., Ackelsberg, S., He, H. D., Hsieh, J., Hu, H.: Future directions in CT technology. Neuroimaging Clinics of North America 998; 8: [9] Ohnesorge, B., Flohr, T., Schaller, S. et al: Technische Grundlagen und Anwendungen der Mehrschicht-CT. Radiologe 999; 39: [0] Hu, H., He, H. D., Foley, W. D., Fox, S. H.: Four multidetector-row helical CT: image quality and volume coverage speed. Radiology 2000; 25: [] Schaller, S., Flohr, T., Steffen, P.: A new, efficient Fourier-reconstruction method for approximate image reconstruction in spiral conebeam CT at small cone-angles. Proceedings of the SPIE International Symposium on Medical Imaging 997; 3032: [2] Saito, Y., Suzuki, T.: Evaluation of the Performance of Multi-slice CT System in Non-helical Scanning. Radiology 2000; 209 (P):578. [3] Flohr, T., Schaller, S., Ohnesorge, B., Klingenbeck-Regn, K., Kopp, A. F.: Evaluation of Image Artifacts in Multislice CT. Radiology 999; 23 (P):37. [4] Schöpf, U. J., Becker, C. R., Brüning, R., Huber, A. M., Hong, C.: Multidetector-array spiral CT imaging of focal and diffuse lung disease: thin-collimation data acquisition with reconstruction of contiguous and HRCT sections. Radiology 999; 23 (P):258. [5] Berland, L. L., Smith, J. K.: Multidetector-array CT: once again, technology creates new opportunities. Radiology 998; 209: [6] Leung, A. W., Klein, J. S.: Optimization of spiral CT of the thorax. Radiology 999; 23 (P):73. [7] Achenbach, S., Moshage, W., Ropers, D., Nossen, J., Daniel, W. G.: Value of electron-beam computed tomography for the noninvasive detection of high-grade coronary-artery stenoses and occlusions. N.Engl.J.Med. 998; 339: [8] Flamm, S. D.: Coronary Arterial Calcium Screening: Ready for Prime Time? Radiology 998; 208:57-2. [9] Ohnesorge, B., Flohr, T., Becker, C. et al: Herzbildgebung mit schneller, retrospektiv EKG-synchronisierter Mehrschichtspiral-CT. Radiologe 2000; 40:-7. [20] Kopp, A. F., Ohnesorge, B., Flohr, T. et al: Multidetektor CT des Herzens: Erste klinische Anwendung einer retrospektiv EKGgesteuerten Spirale mit optimierter zeitlicher und örtlicher Auflösung zur Darstellung der Herzkranzgefäße. Fortschr. Röntgenstr. 2000; 72:-7.

12 [2] Ohnesorge, B., Flohr, T., Becker, C., Kopp, A. F., Schoepf, U. J., Baum, U., Knez, A., Klingenbeck-Regn, K., and Reiser, M. F.: Cardiac Imaging with ECG-Gated Multi-Slice Spiral CT Initial Experience. Radiology 2000, in Press. [22] Ohnesorge, B., Flohr, T., Becker, C. R., Kopp, A. F., Knez, A.: Comparison of EBCT and ECG-gated multislice spiral CT: a study of 3D Ca-scoring with phantom and patient data. Radiology 999; 23 (P):402. [23] Becker, C. R., Knez, A., Ohnesorge, B., Flohr, T., Schoepf, U. J., Reiser, M.: Detection and quantification of coronary artery calcifications with prospectively ECG triggered multirow conventional CT and electron beam computed tomography: comparison of different methods for quantification of coronary artery calcifications. Radiology 999; 23 (P):35. [24] Kopp, A. F., Georg, C., Schröder, S., Claussen, C. D.: CT-Angiographie der Herzkranzgefäße bei koronarer 3-Gefäß-Erkrankung. Fortschr.Röntgenstr. 2000; 72:M3-M4. [25] Kopp, A. F., Ohnesorge, B., Flohr, T., Schroeder, S., Claussen, C. D.: Multidetector-row CT for the noninvasive detection of high-grade coronary artery stenoses and occlusions: first results. Radiology 999; 23 (P):435. [26] Klingenbeck-Regn, K., Schaller, S., Flohr, T., Ohnesorge, B., Kopp, A. F., Baum, U.: Subsecond multislice computed tomography: basics and applications. Eur.J.Radiol. 999; 3:0-24. Author s address: Andreas F. Kopp, M.D. Eberhard-Karls-University Tübingen Department of Diagnostic Radiology D Tübingen/Germany Tel: +49-(0) Fax: +49-(0) andreas.kopp@uni-tuebingen.de electromedica 68 (2000) no. 2 05

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999 Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences mmcnittgray@mednet.ucla.edu X-Ray

More information

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad. 1 12/9/2008 Page 1 Master s Program in Medical Physics Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III Chair in Faculty of Medicine Mannheim University of Heidelberg Theodor-Kutzer-Ufer

More information

Purchasing a cardiac CT scanner: What the radiologist needs to know

Purchasing a cardiac CT scanner: What the radiologist needs to know Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage

More information

State-of-the-Art Technology in Cardiac CT

State-of-the-Art Technology in Cardiac CT 1 2 Next Step Evolution or Revolution? State-of-the-Art Technology in Cardiac CT Stefan Ulzheimer, PhD Global Director of Collaborations CT Siemens Medical Solutions Major Innovations in CT Head The 80

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

How To Improve Your Ct Image Quality

How To Improve Your Ct Image Quality Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Imaging of Thoracic Endovascular Stent-Grafts

Imaging of Thoracic Endovascular Stent-Grafts Imaging of Thoracic Endovascular Stent-Grafts Tariq Hameed, M.D. Department of Radiology and Imaging Sciences, Indiana University School of Medicine, Indianapolis, Indiana Disclosures: No relevant financial

More information

Multislice CT: Current Technology and Future Developments

Multislice CT: Current Technology and Future Developments Multislice CT: Current Technology and Future Developments 1 Stefan Ulzheimer and Thomas Flohr Contents 1.1 Introduction 3 1.2 System Design 5 1.2.1 Gantry 6 1.2.2 X-Ray Tube and Generator 6 1.2.3 MDCT

More information

Multi Detector Row CT Systems and Image- Reconstruction Techniques 1

Multi Detector Row CT Systems and Image- Reconstruction Techniques 1 Special Review Radiology Thomas G. Flohr, PhD Stefan Schaller, PhD Karl Stierstorfer, PhD Herbert Bruder, PhD Bernd M. Ohnesorge, PhD U. Joseph Schoepf, MD Published online before print 10.1148/radiol.2353040037

More information

GE Healthcare. Revolution EVO. More than just high tech. Higher purpose.

GE Healthcare. Revolution EVO. More than just high tech. Higher purpose. GE Healthcare Revolution EVO More than just high tech. Higher purpose. Revolution EVO. Designed with purpose. Today s healthcare environment is about creating new solutions to pressing needs. It s about

More information

Multi-slice CT Technology 3

Multi-slice CT Technology 3 41 Multi-slice CT Technology 3 Thomas Flohr and Bernd Ohnesorge Contents 3.1 Evolution from 1 to 64 Slices 41 3.2 Principles of Multi-slice CT System Design 45 3.3 Multi-slice CT Acquisition and Reconstruction

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

Assessment of calcium scoring performance in cardiac computed tomography

Assessment of calcium scoring performance in cardiac computed tomography Eur Radiol (2003) 13:484 497 DOI 10.1007/s00330-002-1746-y CARDIAC Stefan Ulzheimer Willi A. Kalender Assessment of calcium scoring performance in cardiac computed tomography Received: 19 June 2002 Revised:

More information

Clinical Training for Visage 7 Cardiac. Visage 7

Clinical Training for Visage 7 Cardiac. Visage 7 Clinical Training for Visage 7 Cardiac Visage 7 Overview Example Usage 3 Cardiac Workflow Examples 4 Remove Chest Wall 5 Edit Chest Wall Removal 6 Object Display Popup 7 Selecting Optimal Phase 8 Thick

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

Voxar 3D TM. A suite of advanced visualization and analysis software tools

Voxar 3D TM. A suite of advanced visualization and analysis software tools Voxar 3D TM A suite of advanced visualization and analysis software tools The power to deliver advanced visualization throughout the enterprise To effectively manage the rapid growth of large volumetric

More information

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of

More information

Cardiac CT for Calcium Scoring

Cardiac CT for Calcium Scoring Scan for mobile link. Cardiac CT for Calcium Scoring Cardiac computed tomography (CT) for Calcium Scoring uses special x-ray equipment to produce pictures of the coronary arteries to determine if they

More information

Multi-slice Helical CT Scanning of the Chest

Multi-slice Helical CT Scanning of the Chest Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.

More information

AI CPT Codes. x x. 70336 MRI Magnetic resonance (eg, proton) imaging, temporomandibular joint(s)

AI CPT Codes. x x. 70336 MRI Magnetic resonance (eg, proton) imaging, temporomandibular joint(s) Code Category Description Auth Required Medicaid Medicare 0126T IMT Testing Common carotid intima-media thickness (IMT) study for evaluation of atherosclerotic burden or coronary heart disease risk factor

More information

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

Cone Beam Reconstruction Jiang Hsieh, Ph.D. Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition

More information

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80 Revisions Effective January 2012 Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Technologist Instructions Patient must cough several times prior to scan to clear

More information

Computed Tomography, Head Or Brain; Without Contrast Material, Followed By Contrast Material(S) And Further Sections

Computed Tomography, Head Or Brain; Without Contrast Material, Followed By Contrast Material(S) And Further Sections 1199SEIU BENEFIT AND PENSION FUNDS High Tech Diagnostic Radiology and s # 1 70336 Magnetic Resonance (Eg, Proton) Imaging, Temporomandibular Joint(S) 2 70450 Computed Tomography, Head Or Brain; Without

More information

DENTAL Cone beam 3D X-RAY SYSTEM with

DENTAL Cone beam 3D X-RAY SYSTEM with VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic

More information

5 Factors Affecting the Signal-to-Noise Ratio

5 Factors Affecting the Signal-to-Noise Ratio 5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed

More information

Twelve. Figure 12.1: 3D Curved MPR Viewer Window

Twelve. Figure 12.1: 3D Curved MPR Viewer Window Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

Press. Siemens solutions support diagnosis and treatment of cardiovascular diseases

Press. Siemens solutions support diagnosis and treatment of cardiovascular diseases Press Healthcare Erlangen, August 29, 2015 ESC 2015: ExCel London Exhibition and Convention Center, Booth #G700 Siemens solutions support diagnosis and treatment of cardiovascular diseases New cardiovascular

More information

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami

More information

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. 1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Computer-Assisted Coronary CT Angiography Analysis

Computer-Assisted Coronary CT Angiography Analysis Linköping University Medical Dissertations, No. 1237 Computer-Assisted Coronary CT Angiography Analysis From Software Development to Clinical Application Chunliang Wang Division of Radiological Sciences

More information

The effect of pitch in multislice spiral/helical CT

The effect of pitch in multislice spiral/helical CT The effect of pitch in multislice spiral/helical CT Ge Wang a) and Michael W. Vannier Department of Radiology, University of Iowa, Iowa City, Iowa 52242 Received 20 November 1998; accepted for publication

More information

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt.

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt. Medical Image Processing on the GPU Past, Present and Future Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt.edu Outline Motivation why do we need GPUs? Past - how was GPU programming

More information

MYOCARDIAL PERFUSION COMPUTED TOMOGRAPHY PhD course in Medical Imaging. Anne Günther Department of Radiology OUS Rikshospitalet

MYOCARDIAL PERFUSION COMPUTED TOMOGRAPHY PhD course in Medical Imaging. Anne Günther Department of Radiology OUS Rikshospitalet MYOCARDIAL PERFUSION COMPUTED TOMOGRAPHY PhD course in Medical Imaging Anne Günther Department of Radiology OUS Rikshospitalet CORONARY CT ANGIOGRAPHY (CTA) Accurate method in the assessment of possible

More information

Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35

Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35 Development of SCENAIA New Version Software Takayuki Kadomura Toshio Sakamoto Kana Tanaka Takatsugu Ito Yuuko Nishimura Naomi Maekawa MEDIX VOL. 62 P.32 P.35 Development of SCENAIA New Version Software

More information

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro. www.neurologica.com Critical Care Clinic Operating Room Radiology ED Trauma OR Neuro/Spine Pediatric Plastics Thoracic General Interventional Radiology ENT ED Trauma Trauma Stroke Critical Care Neuro Pediatric

More information

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Computed Tomography to Detect Coronary Artery Calcification File Name: computed_tomography_to_detect_coronary_artery_calcification Origination: 3/1994 Last CAP Review 11/2014 Next

More information

Low-dose CT for Pulmonary Embolism

Low-dose CT for Pulmonary Embolism Low-dose CT for Pulmonary Embolism Gautham Gautham P. P. Reddy, Reddy, MD, MD, MPH MPH University University of of Washington Washington Introduction Introduction CT CT accounts accounts for for > 50%

More information

American College of Radiology CT Accreditation Program. Testing Instructions

American College of Radiology CT Accreditation Program. Testing Instructions American College of Radiology CT Accreditation Program Testing Instructions (Revised July 24, 2015) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission

More information

GE Healthcare. Great treasures in small places. BrightSpeed Elite

GE Healthcare. Great treasures in small places. BrightSpeed Elite GE Healthcare Great treasures in small places BrightSpeed Elite The BrightSpeed Elite brings you multi-detector CT capabilities with maximum convenience in an ultra-compact design. Built with LightSpeed

More information

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D. 4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology jiang.steve@mgh.harvard.edu http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Brilliance CT 64-channel confi guration

Brilliance CT 64-channel confi guration Brilliance CT 64-channel confi guration with Essence technology The Brilliance CT 64-channel confi guration is designed to help you conduct the most advanced multislice CT studies possible. These systems

More information

CPT Radiology Codes Requiring Review by AIM Effective 01/01/2016

CPT Radiology Codes Requiring Review by AIM Effective 01/01/2016 CPT Radiology Codes Requiring Review by AIM Effective 01/01/2016 When a service is authorized only one test per group is payable. *Secondary codes or add-on codes do not require preauthorization or separate

More information

Pediatric Hospitals Bring Low-dose CT to the Middle East

Pediatric Hospitals Bring Low-dose CT to the Middle East Pediatric Hospitals ring Low-dose CT to the Middle East For years, radiologists have been cognizant of the importance of limiting pediatric patients exposure to radiation dose. uilding on the LR principle,

More information

CORONARY ARTERY BYPASS GRAFTS, STENTS, AND EXTRACORONARY CARDIAC DZ. Charles White MD

CORONARY ARTERY BYPASS GRAFTS, STENTS, AND EXTRACORONARY CARDIAC DZ. Charles White MD CORONARY ARTERY BYPASS GRAFTS, STENTS, AND EXTRACORONARY CARDIAC DZ Charles White MD Director of Thoracic Imaging Department of Radiology University of Maryland CORONARY ARTERY BYPASS GRAFTS First performed

More information

CT Angiography (CTA) What is CT Angiography?

CT Angiography (CTA) What is CT Angiography? Scan for mobile link. CT Angiography (CTA) Computed tomography angiography (CTA) uses an injection of iodine-rich contrast material and CT scanning to help diagnose and evaluate blood vessel disease or

More information

CTA OF THE EXTRACORONARY HEART

CTA OF THE EXTRACORONARY HEART CTA OF THE EXTRACORONARY HEART Charles White MD Director of Thoracic Imaging Department of Radiology University of Maryland NO DISCLOSURES CWHITE@UMM.EDU CARDIAC CASE DISTRIBUTION Coronary CTA 30% ED chest

More information

Automatic Optimum Phase Selection In Cardiac CT Imaging. Master Thesis Fahmi Noor

Automatic Optimum Phase Selection In Cardiac CT Imaging. Master Thesis Fahmi Noor Automatic Optimum Phase Selection In Cardiac CT Imaging Master Thesis Fahmi Noor Place Siemens Medical Germany CTE Physical Application Siemensstrasse 1, Forchheim Supervisor Dr. Herbert Bruder Referee

More information

ENG4BF3 Medical Image Processing. Image Visualization

ENG4BF3 Medical Image Processing. Image Visualization ENG4BF3 Medical Image Processing Image Visualization Visualization Methods Visualization of medical images is for the determination of the quantitative information about the properties of anatomic tissues

More information

High Quality Image Magnification using Cross-Scale Self-Similarity

High Quality Image Magnification using Cross-Scale Self-Similarity High Quality Image Magnification using Cross-Scale Self-Similarity André Gooßen 1, Arne Ehlers 1, Thomas Pralow 2, Rolf-Rainer Grigat 1 1 Vision Systems, Hamburg University of Technology, D-21079 Hamburg

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Global Business Unit Address. Siemens Medical Solutions USA, Inc. 2501 N. Barrington Road Hoffman Estates, IL 60192-5203. Telephone: +1847 304 7700

Global Business Unit Address. Siemens Medical Solutions USA, Inc. 2501 N. Barrington Road Hoffman Estates, IL 60192-5203. Telephone: +1847 304 7700 Trademarks and service marks used in this material are property of Siemens Medical Solutions USA or Siemens AG. All other company, brand, product and service names may be trademarks or registered trademarks

More information

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction

More information

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA.

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA. DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA. Juan Antonio Martínez Mera Centro Singular de Investigación en Tecnoloxías da Información Universidade de Santiago de Compostela

More information

Sistemas de reducción de dosis en CT para pediatría. Nuevos avances Ángela de Pinto Siemens CT Business Manager angela.pinto@siemens.

Sistemas de reducción de dosis en CT para pediatría. Nuevos avances Ángela de Pinto Siemens CT Business Manager angela.pinto@siemens. Sistemas de reducción de dosis en CT para pediatría. Nuevos avances Ángela de Pinto Siemens CT Business Manager angela.pinto@siemens.com Copyright Siemens AG 2009. All rights reserved. An Innovation Leader

More information

Mx8000 IDT 16 provides dose control methodologies like DoseRight ACS (Automatic Current Selection) and DoseRight DOM (dynamic dose modulation).

Mx8000 IDT 16 provides dose control methodologies like DoseRight ACS (Automatic Current Selection) and DoseRight DOM (dynamic dose modulation). Philips Mx8000 IDT 16 CT Scanner Philips Medical systems Mx8000 IDT 16 delivers 16 simultaneous slices at sub-millimeter collimator, high resolution and isotropic imaging. New large area detectors allow

More information

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography Today s agenda Computed Tomography Presented by: Dina Hernandez, BSRT, RT (R), CT, QM Krista Bush, RT, MBA Leonard Lucey, JD ACR Quality & Safety MIPPA overview and CMS requirements CT accreditation How

More information

Siemens Computed Tomography

Siemens Computed Tomography Siemens Computed Tomography Innovation Leader. The Cosmopolitan Information Topic Waltraud Winter Siemens Copyright Siemens AG 2011. All rights reserved. Page 1 October 2012 Waltraud Winter An Innovation

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

Procedure Codes. RadConsult provides real-time decision support for physicians who order high-cost imaging procedures RADIATION THERAPY

Procedure Codes. RadConsult provides real-time decision support for physicians who order high-cost imaging procedures RADIATION THERAPY Procedure Codes 2011 RadConsult provides real-time decision support for physicians who order high-cost imaging procedures RADIATION THERAPY 2D3D Therapeutic radiology treatment planning; simple 77261 Therapeutic

More information

A d v a n c e d 1 2 8

A d v a n c e d 1 2 8 Advanced 128 SCENARIA Advanced 128 Hitachi Innovation at Work for You. Continuing its commitment to the SCENARIA scalable CT platform, Hitachi introduces cutting-edge 128-slice technology with the SCENARIA

More information

Three Dimensional Ultrasound Imaging

Three Dimensional Ultrasound Imaging Three Dimensional Ultrasound Imaging Hans Torp/ Sevald Berg/Kjell Kristoffersen m/flere Department of circulation and medical imaging NTNU Hans Torp NTNU, Norway Acquisition Reconstruction Filtering Collecting

More information

Ultrasound in Vascular Surgery. Torbjørn Dahl

Ultrasound in Vascular Surgery. Torbjørn Dahl Ultrasound in Vascular Surgery Torbjørn Dahl 1 The field of vascular surgery Veins dilatation and obstruction (varicose veins and valve dysfunction) Arteries dilatation and narrowing (aneurysms and atherosclerosis)

More information

Mammography. What is Mammography?

Mammography. What is Mammography? Scan for mobile link. Mammography Mammography is a specific type of breast imaging that uses low-dose x-rays to detect cancer early before women experience symptoms when it is most treatable. Tell your

More information

Adenosine stress DECT equivalent to SPECT, MRI for myocardium evaluation By Eric Barnes AuntMinnie.com staff writer November 21, 2008

Adenosine stress DECT equivalent to SPECT, MRI for myocardium evaluation By Eric Barnes AuntMinnie.com staff writer November 21, 2008 Adenosine stress DECT equivalent to SPECT, MRI for myocardium evaluation By Eric Barnes AuntMinnie.com staff writer November 21, 2008 A new power combo consisting of adenosine stress and dual-energy CT

More information

Interim policy effective January 1, 2016, through December 31, 2017

Interim policy effective January 1, 2016, through December 31, 2017 IMPORTANT NOTICE Interim policy announced for post-primary structured education requirement Interim policy relaxes exam content outline requirement (November 2, 205) The ARRT announces an interim policy

More information

SUPERCHARGED. CUSTOM-DESIGNED FOR CARDIOLOGY. Cardiology PACS. Cardiology

SUPERCHARGED. CUSTOM-DESIGNED FOR CARDIOLOGY. Cardiology PACS. Cardiology Cardiology PACS CUSTOM-DESIGNED FOR CARDIOLOGY. Now, you can count on powerful PACS performance in your cardiology department. CARESTREAM Vue for Cardiology can consolidate your disparate cardio lab systems

More information

P R E S E N T S Dr. Mufa T. Ghadiali is skilled in all aspects of General Surgery. His General Surgery Services include: General Surgery Advanced Laparoscopic Surgery Surgical Oncology Gastrointestinal

More information

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

Musculoskeletal MRI Technical Considerations

Musculoskeletal MRI Technical Considerations Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T

More information

CT Image Reconstruction. Terry Peters Robarts Research Institute London Canada

CT Image Reconstruction. Terry Peters Robarts Research Institute London Canada CT Image Reconstruction Terry Peters Robarts Research Institute London Canada 1 Standard X-ray Views Standard Radiograph acquires projections of the body, but since structures are overlaid on each other,

More information

GE Healthcare. Optima* CT660

GE Healthcare. Optima* CT660 GE Healthcare Optima* CT660 Your vision of quality care made The Optima CT660 system helps you deliver competent, personalized care that helps fulfill your mission and please your patients. One look at

More information

AW Server 3 for Universal Viewer

AW Server 3 for Universal Viewer GE Healthcare AW Server 3 for Universal Viewer Powering Advanced Applications in GE s Universal Viewer for Centricity PACS and Centricity PACS-IW. In today s productivity-focused Radiology environment,

More information

Computed Tomography (CT) - Chest

Computed Tomography (CT) - Chest Scan for mobile link. Computed Tomography (CT) - Chest Computed tomography (CT) of the chest uses special x-ray equipment to examine abnormalities found in other imaging tests and to help diagnose the

More information

Patient Prep Information

Patient Prep Information Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center

More information

Surgery Support System as a Surgeon s Advanced Hand and Eye

Surgery Support System as a Surgeon s Advanced Hand and Eye Surgery Support System as a Surgeon s Advanced Hand and Eye 8 Surgery Support System as a Surgeon s Advanced Hand and Eye Kazutoshi Kan Michio Oikawa Takashi Azuma Shio Miyamoto OVERVIEW: A surgery support

More information

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative

More information

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry Table of Contents Scan acquisition and user interface basics Dose modulation and reduction tools Multi-Slice Detector Geometry Image Reconstruction and Display Contrast Media Tools Multi-planar formats

More information

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement. DISCLAIMER: This publication is based on sources and information believed to be reliable, but the AAPM, the editors, and the publisher disclaim any warranty or liability based on or relating to the contents

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

FFR CT : Clinical studies

FFR CT : Clinical studies FFR CT : Clinical studies Bjarne Nørgaard Department Cardiology B Aarhus University Hospital Skejby, Denmark Disclosures: Research grants: Edwards and Siemens Coronary CTA: High diagnostic sensitivity

More information

CMS Manual System Pub. 100-03 Medicare National Coverage Determinations

CMS Manual System Pub. 100-03 Medicare National Coverage Determinations CMS Manual System Pub. 100-03 Medicare National Coverage Determinations Department of Health & Human Services (DHHS) Centers for Medicare & Medicaid Services (CMS) Transmittal 21 Date: September 10, 2004

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

MEDICAL IMAGING. 90/30 Remote Controlled RF System

MEDICAL IMAGING. 90/30 Remote Controlled RF System MEDICAL IMAGING 90/30 Remote Controlled RF System The Easy solution for your Remote Controlled room Apollo EZ represents the Easy approach for the radiographic/fluoroscopic room and it comes from the experience

More information

Case Report: Whole-body Oncologic Imaging with syngo TimCT

Case Report: Whole-body Oncologic Imaging with syngo TimCT Case Report: Whole-body Oncologic Imaging with syngo TimCT Eric Hatfield, M.D. 1 ; Agus Priatna, Ph.D. 2 ; John Kotyk, Ph.D. 1 ; Benjamin Tan, M.D. 1 ; Alto Stemmer 3 ; Stephan Kannengiesser, Ph.D. 3 ;

More information

GE Healthcare. Discovery * CT750 HD Great care by design

GE Healthcare. Discovery * CT750 HD Great care by design GE Healthcare Discovery * CT750 HD Great care by design At GE Healthcare, imag lie at the heart of every than 30 years, we ve d that enhance the ways and treat disease. Bett us to render the inner greater

More information

Your Guide to Express Critical Illness Insurance Definitions

Your Guide to Express Critical Illness Insurance Definitions Your Guide to Express Critical Illness Insurance Definitions Your Guide to EXPRESS Critical Illness Insurance Definitions This guide to critical illness definitions will help you understand the illnesses

More information

CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION

CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION CINDY WEILAND AND SANDRA L. KATANICK Continued innovations in noninvasive testing equipment provide skilled sonographers and physicians with the technology

More information

Performance testing for Precision 500D Classical R/F System

Performance testing for Precision 500D Classical R/F System Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)

More information

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan An Brief Introduction to 4D CT Scanning Steve B. Jiang, Ph.D. Dept of Radiation Oncology Univ of California San Diego Outline Problems with free breathing 3D scanning What is 4D CT? How does it work? Acknowledgements

More information

Business Case to replace CT Scanner at Daisy Hill Hospital

Business Case to replace CT Scanner at Daisy Hill Hospital Business Case to replace CT Scanner at Daisy Hill Hospital Version 0.8 Corporate Planning Division Directorate of Performance & Reform 14 th December 2009 TABLE OF CONTENTS 1. EXECUTIVE SUMMARY...4 2.

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

The New Enhanced Multiframe CT and MR DICOM Objects

The New Enhanced Multiframe CT and MR DICOM Objects The New Enhanced Multiframe CT and MR DICOM Objects David Clunie, MB, BS CTO, RadPharm Bangor, PA 18013 Bradley J Erickson, MD PhD Dept Radiology, Mayo Clinic Rochester, MN 55902 (507) 284-8548 Outline

More information

GE Healthcare. Centricity * PACS with Universal Viewer. Universal Viewer. Where it all comes together.

GE Healthcare. Centricity * PACS with Universal Viewer. Universal Viewer. Where it all comes together. GE Healthcare Centricity * PACS with Universal Viewer Universal Viewer. Where it all comes together. Where it all comes together Centricity PACS with Universal Viewer introduces an intuitive imaging application

More information

The IAC Standards and Guidelines for CT Accreditation

The IAC Standards and Guidelines for CT Accreditation The IAC Standards and Guidelines for CT Accreditation Table of Contents All entries in Table of Contents are linked to the corresponding sections. Introduction... 4 Part A: Organization... 5 Section 1A:

More information

Digital Image Processing with DragonflyPACS

Digital Image Processing with DragonflyPACS Digital Image Processing with DragonflyPACS The Smarter PACS The smartest image processing system in veterinary medicine. Outstanding image quality that s compliant with DICOM 3.0 standards. Compatible

More information

University Children s Hospital Basel

University Children s Hospital Basel Cooperation Siemens International Reference Center Pediatric MRI www.siemens.com/skyra University Children s Hospital Basel Siemens International Reference Center Pediatric MRI Answers for life. Partnering

More information