Construction of a Gas Electron Multiplier (GEM) Detector for Medical Imaging

Size: px
Start display at page:

Download "Construction of a Gas Electron Multiplier (GEM) Detector for Medical Imaging"

Transcription

1 Construction of a Gas Electron Multiplier (GEM) Detector for Medical Imaging N. N. Mondal, S. Chattopadhyay, M. R. Dutta Mazumdar, A. K. Dubey and Y. P. Vioygi. Physics Group, Variable Energy Cyclotron Center (VECC) 1/AF, Bidhannagar, Kolkata , India. Abstract: A prototype Gas Electron Multiplier (GEM) detector is under construction for medical imaging purposes. A single thick GEM of size cm 2 is assembled inside a square shaped air-tight box which is made of Perspex glass. In order to ionize gas inside the drift field two types of voltage supplier circuits were fabricated, and array of 2 4 pads of each size 4 8 mm 2 were utilized for collecting avalanche charges. Preliminary testing results show that the circuit which produces high voltage and low current is better than that of low voltage and high current supplier circuit in terms of x-ray signal counting rates. 1. Introduction: Gas Electron Multiplier (GEM) detector system was first introduced by F. Sauli [1] at CERN in A GEM is a composite grid consisting of two metal layers separated by a thin insulator, etched with a regular array of pin hole (diameter of a few hundredth of mm). This sandwiched package is kept inside a non explosive gas (mainly inert gas) or mixture of gases with optimized proportion under normal temperature and pressure in general. A suitable potential difference is applied between the two copperplates, which generate drift field along the holes perpendicular to the plates. When energetic photons or charged particles enter into this drift region gas is ionized and create pair of ions. Fig. 1: A schematic diagram of a standard double GEM detector (not in scale). 2D image reconstruction is

2 possible by using the data X- and Y-grids. Electrons are accelerated by the drift field and ionize more electrons and those avalanched electrons cloud are accumulated on the anode plates, made of copper strips/pads on the printed circuit board (PCB) and electronics. Principle of GEM is similar to that of a photomultiplier tube (PMT). Single, double, triple or multiple GEM detectors are possible to develop in order to improve gain, better time and energy resolutions. A schematic view of a double GEM detector is depicted in Fig. 1. The upper plate of GEM is a beryllium window of a few µ m thicknesses through which photons or charged particles can easily pass without loss of significant energies. This layer is useful for gas confinement also. GEM1 and GEM2 are thick copper foils consists of many holes, each diameter is µ m. Data are collected using the two grids (X and Y) and 2D image reconstruction can be done by image reconstruction algorithms. The advanced features of GEM based detectors are high counting rate, excellent spatial resolution, and good imaging capability, operation in electric/magnetic fields, large sensitive area, flexible geometry and low cost [2]. The remarkable features stated above of multi-gem structure are very attractive for the numerous applications [3]. GEM-based detectors like other micro-pattern devices, offer localization to around 40 microns. However, their unique feature is that two co-ordinates can be recorded on the pickup electrode. Both X- and Y-strips are at ground potential, essential when using high-density readout. Such two-dimensional localization, useful in particle tracking, is necessary for medical imaging. X-rays are efficient tool for diagnostics in medicine and industry and are widely used for biological, chemical and material science researches. In astro-particle physics x-rays provide significant information on the type of x-ray sources and the objects that scatter and affect the polarization of x-rays such as magnetic fields. Several important domains of x-ray imaging are medical diagnostics, industrial x-ray imaging, synchrotron radiation experiments and astro-particle physics. A comparison among the medical scanners with GEM can be found elsewhere [4]. 2. Materials and Methods: Fig. 2 Schematic diagram of a thick GEM foil. The GEM works in the proportional regime as an avalanche preamplifier, which means that the amplitude of the output signal is proportional to the initial number of ion pairs produced by the incident photon. The gap between the drift plane and the GEM top is called the drift region or conversion gap, and the lower space, where amplified electrons are collected by the readout plate (anodes), is named the induction region or collection gap (see Fig. 1.). Upon application of a potential difference between the GEM copper plates, a high electric field is developed inside the holes focusing the field lines between the drift electrode and the readout strips/plates. The field density in the amplifying channel can be varied either by increasing the potential difference between the upper and the lower electrodes. Or by reducing the diameter of the GEM holes. The length of the amplifying

3 channel for a single GEM grid is fixed by the thickness of the insulating foil. So far most standard GEM foils have an insulator thickness of 50 µ m. Along the field directions perpendicular to the axis of the holes the field strength is almost uniform in the center and increases sharply near the edges, particularly close to the copper-kapton interface. This determines how a high voltage could be applied to the GEM, i.e., the point of electrical breakdown (appears as a spark signal). The charge avalanche traverses the GEM hole mostly in the center, but a fraction of it could approach this extremely high field region on the edges of the hole due to diffusion and trigger a discharge. Also electron emission from the GEM electrodes in this area is possible. Fig.3 Voltage supplier circuits: (a) High voltage and low current, input voltage to the Mesh was supplied separately, (b) Low Voltage and high current supply across the GEM and Mesh. Both Mesh and GEM are kept in floating conditions. Two types of voltage supplier circuit are demonstrated in fig. 3. Total impedance of (a) is about five times higher than (b). For various input voltage in the circuit DV G across the GEM foil and corresponding currents are obtained for two different circuits. In Fig. 4, DV G versus current are shown. 4. Results & Discussion Using the 55 Fe RI source (a few µ Ci) we have tried to get signals and counting rates. In Fig.5 signals with and without sources are depicted. The amplitude of the noise level is about 2 4 mv and with source amplitude is about 6 10 mv without amplifier. The voltage divider circuit (fig. 3(a) was used for this testing purpose. Using more than 5 mv discriminator threshold much of the noise and other background is reduced but sparks and cosmic

4 background remain the same. In Fig.6 a counting rate measurement circuit is depicted. Many sparks signals come at a time with highest amplitude are easily countable separately from the original signals. We have found the cosmic ray signals come out with higher amplitude than the x-ray signals. Amplitudes of x-ray signals are also varying and the average amplitude is about 7 mv without further amplification. The signal and the spark counting rates with corresponding DV G are depicted in Fig.7. It is very difficult to get pure x-ray signals out of sparking from this system, because nothing is optimized yet. The amplitude of the spark signals have no limits, therefore no discriminator threshold level can be set in order to get rid off it. Figure 4. Voltage across the single thick GEM and currents in the circuit (fig.3(a)) are shown. Signals of x-rays are obtained only in the sparking area. 3. Testing of a prototype GEM detector: It is quite known from many other experiments that avalanche electron gain increases with increasing DV G and more of it is required to obtain same gain of pure high Z inert gases. Optimized mixture of gases, e.g., Ar:CO 2 (70:30), Xe:Ar(95:5), Ar:CH 4 (90:10) etc. show better gain at lower DV G than pure gases at atmospheric pressure. We have used Ar:CO 2 (70:30) gas mixture to obtain better gain, special, time and energy resolutions [CBM report]. Gain or the detection efficiency depends on many parameters of the detector such as geometry, size, shape and materials of GEM foil, cathode mesh, pads, electronic accessories, gas, mixture of gases, drift, ionization and transfer distances, and the voltage across the Mesh and GEM. For the test of this prototype GEM we used two types of voltage divider circuits (see fig. 3) and obtained rare signals which are depicted in Fig. 5.

5 Fig. 5: Signals of 5.9 kev x-rays and noises are shown. The voltage divider circuit (see fig.3(a) was used to process these signals. 4. Results & Discussion Using the 55 Fe RI source (a few µ Ci) we have tried to get signals and counting rates. In Fig.5 signals with and without sources are depicted. The amplitude of the noise level is about 2 4 mv and with source amplitude is about 6 10 mv without amplifier. The voltage divider circuit (fig. 3(a) was used for this testing purpose. Using more than 5 mv discriminator threshold much of the noise and other background is reduced but sparks and cosmic background remain the same. In Fig.6 a counting rate measurement circuit is depicted. Many sparks signals come at a time with highest amplitude are easily countable separately from the original signals. We have found the cosmic ray signals come out with higher amplitude than the x-ray signals. Amplitudes of x-ray signals are also varying and the average amplitude is about 7 mv without further amplification. The signal and the spark counting rates with corresponding DV G are depicted in Fig.7. It is very difficult to get pure x-ray signals out of sparking from this system, because nothing is optimized yet. The amplitude of the spark signals have no limits, therefore no discriminator threshold level can be set in order to get rid off it. A lot of sparks almost half of the signals come out due to high DV G which is not recommended because of the safety reason of the detector. Sparks generally appear due to dust particles inside the holes or some other cracks or distortion among the copper layers and insulator. The rate of sparks increases due to avalanche diffusion of electrons. We also design another type of circuit [see

6 fig. 3(b)] for low DV G and high current supply. We increase current upto 700 µ A at DV G =1000 V. Fig. 6: Signal processing and counting circuit with Phosphor Digital Oscilloscope, Model No. DPO 7254, Tektronix. A lot of spark signals were seen without any x-ray signal. It seems to us that second type of circuit is not suitable for thick GEM. Our applied GEM foil is initially damaged a few areas and after cleaning we used it for testing purpose. Thickness of the top layer of the detector was 5 mm and drilled with 2 mm hole where 55 Fe source was kept and drift gap was 1 cm. Therefore x-ray exposure area was very limited and a very few ionization was occurred and less number of electrons were detected by the small size pad. Fig. 7. Counting rates are shown with different voltage drops across the single thick GEM foil.

7 Hence it is difficult to conclude that the geometry of the system, circuits or foil itself which are not providing efficient signals. In order to test this detector high intense x-rays and γ-rays with various energies are required to optimize DV G, current, geometry, gain etc of the system. Conclusion Advantages of GEM detector are its flexible shape, size, spatial resolution and above all cost. A prototype GEM detector is constructed and tested roughly. A lot of improvement in design, circuits and electronics are required in order to get pure signals of x-rays, good detection efficiency and measurement of spatial, energy and time resolutions. Preliminary test show that high DV G and low current supply circuit is better than low DV G and high current supply circuit for getting much counting rate of x-rays. Further improvements of the system are going on and the fruitful results are commensurate soon. Acknowledgements: Authors are pleased to acknowledge the sincere help of scientific assistants. One of the authors is pleased to acknowledge the fellowship support from the VECC and grateful to Dr. D. K. Srivastava for encouraging the project. Reference: 1. F. Sauli NIMA 386 (1997) S. Bachmann et.al. NIMA 471 (2001) A. Buzulutskov, NIMA 494(2002) N. N. Mondal et.al., Annuual report 2009, VECC, Kolkata, India.

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (no PiN and pinned Diodes) Peter Fischer P. Fischer, ziti, Uni Heidelberg, Seite 1 Overview Reminder: Classical Photomultiplier

More information

Design and Construction of a First Prototype Muon Tomography System with GEM Detectors for the Detection of Nuclear Contraband

Design and Construction of a First Prototype Muon Tomography System with GEM Detectors for the Detection of Nuclear Contraband Design and Construction of a First Prototype Muon Tomography System with GEM Detectors for the Detection of Nuclear Contraband M. Hohlmann, Member, IEEE, K. Gnanvo, Member, IEEE, L. Grasso, J. B. Locke,

More information

Evaluation of various planar gaseous detectors with CsI photocathodes. for the detection of primary scintillation light from noble gases

Evaluation of various planar gaseous detectors with CsI photocathodes. for the detection of primary scintillation light from noble gases Evaluation of various planar gaseous detectors with CsI photocathodes for the detection of primary scintillation light from noble gases L. Periale 1,2, V. Peskov 3, P. Carlson 3, C. Iacobaeus 4, T. Francke

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

Information about the T9 beam line and experimental facilities

Information about the T9 beam line and experimental facilities Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions

More information

Radiation Strip Thickness Measurement Systems

Radiation Strip Thickness Measurement Systems Radiation Strip Thickness Measurement Systems During the past years we have increased our sales of radiometric Vollmer strip thickness measurement systems, i.e. X-ray or isotope gauges, dramatically. Now,

More information

Vacuum Evaporation Recap

Vacuum Evaporation Recap Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.

More information

Results from first tests of TRD prototypes for CBM. DPG Frühjahrstagung Münster 2011 Pascal Dillenseger Institut für Kernphysik Frankfurt am Main

Results from first tests of TRD prototypes for CBM. DPG Frühjahrstagung Münster 2011 Pascal Dillenseger Institut für Kernphysik Frankfurt am Main Results from first tests of TRD prototypes for CBM DPG Frühjahrstagung Münster 2011 Pascal Dillenseger Institut für Kernphysik Contents Overview of the CBM experiment CBM-TRD General TRD requirements The

More information

Cathode Ray Tube. Introduction. Functional principle

Cathode Ray Tube. Introduction. Functional principle Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the

More information

Expression of Interest in providing design and manufacturing of detectors for the construction of the European X-ray Free Electron Laser Facility

Expression of Interest in providing design and manufacturing of detectors for the construction of the European X-ray Free Electron Laser Facility Wrocław Technology Park 54-424 Wrocław, Expression of Interest in providing design and manufacturing of detectors for the construction of the European X-ray Free Electron Laser Facility Introduction This

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

The photoionization detector (PID) utilizes ultraviolet

The photoionization detector (PID) utilizes ultraviolet Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).

More information

Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge

Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge , Firenze, Italy Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge A. Bulletti, L. Capineri B. Dunn ESTEC Material and Process

More information

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

Measurement of Charge-to-Mass (e/m) Ratio for the Electron Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic

More information

Modern Physics Laboratory e/m with Teltron Deflection Tube

Modern Physics Laboratory e/m with Teltron Deflection Tube Modern Physics Laboratory e/m with Teltron Deflection Tube Josh Diamond & John Cummings Fall 2010 Abstract The deflection of an electron beam by electric and magnetic fields is observed, and the charge

More information

Methods of plasma generation and plasma sources

Methods of plasma generation and plasma sources Methods of plasma generation and plasma sources PlasTEP trainings course and Summer school 2011 Warsaw/Szczecin Indrek Jõgi, University of Tartu Partfinanced by the European Union (European Regional Development

More information

GAMMA-RAY SPECTRA REFERENCES

GAMMA-RAY SPECTRA REFERENCES GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.

More information

Coating Technology: Evaporation Vs Sputtering

Coating Technology: Evaporation Vs Sputtering Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information

More information

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.

More information

2 Absorbing Solar Energy

2 Absorbing Solar Energy 2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could

More information

High Rate Oxide Deposition onto Web by Reactive Sputtering from Rotatable Magnetrons

High Rate Oxide Deposition onto Web by Reactive Sputtering from Rotatable Magnetrons High Rate Oxide Deposition onto Web by Reactive Sputtering from Rotatable Magnetrons D.Monaghan, V. Bellido-Gonzalez, M. Audronis. B. Daniel Gencoa, Physics Rd, Liverpool, L24 9HP, UK. www.gencoa.com,

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

Storage Tubes and Their Basic Principles, pp. 93-96

Storage Tubes and Their Basic Principles, pp. 93-96 Storage Tubes and Their Basic Principles, pp. 93-96 Description (Fig. 26). This device employs a large number of short parallel beams of electrons emitted from elongated flat cathodes which lie side by

More information

ISTITUTO NAZIONALE DI FISICA NUCLEARE

ISTITUTO NAZIONALE DI FISICA NUCLEARE ISTITUTO NAZIONALE DI FISICA NUCLEARE Sezione di Catania INFN/TC-11/02 February 28, 2011 PRELIMINARY TESTS OF A SCINTILLATOR-BASED MINI-STATION FOR EXTENSIVE AIR SHOWERS MEASUREMENTS S.Aiola 1, P. La Rocca

More information

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the

More information

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada

More information

Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission

Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission MAVEN Science Community Workshop December 2, 2012 Particles and Fields Package Solar Energetic Particle Instrument (SEP) Davin Larson and the SEP

More information

PoS(PhotoDet 2012)068

PoS(PhotoDet 2012)068 Characterization of the Hamamatsu R11265 multi-anode photomultiplier tube with single photon signals Luca CADAMURO, Marta CALVI, Andrea GIACHERO,, Matteo MAINO, Clara MATTEUZZI, Gianluigi PESSINA Dipartimento

More information

TOF FUNDAMENTALS TUTORIAL

TOF FUNDAMENTALS TUTORIAL TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] info@rmjordan.com [e-mail] This

More information

A.17. OXIDIZING PROPERTIES (SOLIDS)

A.17. OXIDIZING PROPERTIES (SOLIDS) A.17. OXIDIZING PROPERTIES (SOLIDS) 1. METHOD 1.1. INTRODUCTION It is useful to have preliminary information on any potentially explosive properties of the substance before performing this test. This test

More information

07 - Cherenkov and transition radiation detectors

07 - Cherenkov and transition radiation detectors 07 - Cherenkov and transition radiation detectors Jaroslav Adam Czech Technical University in Prague Version 1.0 Jaroslav Adam (CTU, Prague) DPD_07, Cherenkov and transition radiation Version 1.0 1 / 30

More information

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics

More information

How to measure absolute pressure using piezoresistive sensing elements

How to measure absolute pressure using piezoresistive sensing elements In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article

More information

E/M Experiment: Electrons in a Magnetic Field.

E/M Experiment: Electrons in a Magnetic Field. E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.

More information

Introduction to vacuum gauges. Vacuum Gauges where the Pressure Readings are Independent of the Type of Gas (Mechanical Vacuum Gauges)

Introduction to vacuum gauges. Vacuum Gauges where the Pressure Readings are Independent of the Type of Gas (Mechanical Vacuum Gauges) Introduction to vacuum gauges Vacuum Gauges where the Pressure Readings are Independent of the Type of Gas (Mechanical Vacuum Gauges) BOURDON Vacuum Gauge The inside of a tube which is bent into a circular

More information

Electricity. Investigating spontaneous gas discharge in air as a function of pressure. LD Physics Leaflets P3.9.2.1. 0210-Sel

Electricity. Investigating spontaneous gas discharge in air as a function of pressure. LD Physics Leaflets P3.9.2.1. 0210-Sel Electricity Electrical conduction in gases Gas discharge at reduced pressure LD Physics Leaflets P3.9.2.1 Investigating spontaneous gas discharge in air as a function of pressure Objects of the experiments

More information

Silicon Drift Detector Product Brochure Update 2013

Silicon Drift Detector Product Brochure Update 2013 Silicon Drift Detector Product Brochure Update 2013 Content Classic Silicon Drift Detector High Resolution Silicon Drift Detector Multielement Silicon Detector Extra Large Detector Series July 2013 About

More information

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS Contents Foreword --------------------------------------------------------------------------------------------------------------------

More information

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY I. BIBICU 1, G. NICOLESCU 2, L. CIOLACU 2, L. SERBINA 2 1 National Institute for Materials Physics, Bucharest 77125, Romania, bibicu@infim.ro 2

More information

Introduction to Geiger Counters

Introduction to Geiger Counters Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists

More information

High resolution and two-dimensional focal plane detector for intermediate energy heavy ions

High resolution and two-dimensional focal plane detector for intermediate energy heavy ions INS Rep.-1088 INSTITUTE FOR NUCLEAR STUDY Jan. 1995 UNIVERSITY OF TOKYO T u- T i,an INS 1088 Tanashi, Tokyo 188 jp9503014 Japan High resolution and two-dimensional focal plane detector for intermediate

More information

RX-AM4SF Receiver. Pin-out. Connections

RX-AM4SF Receiver. Pin-out. Connections RX-AM4SF Receiver The super-heterodyne receiver RX-AM4SF can provide a RSSI output indicating the amplitude of the received signal: this output can be used to create a field-strength meter capable to indicate

More information

Graphical displays are generally of two types: vector displays and raster displays. Vector displays

Graphical displays are generally of two types: vector displays and raster displays. Vector displays Display technology Graphical displays are generally of two types: vector displays and raster displays. Vector displays Vector displays generally display lines, specified by their endpoints. Vector display

More information

The OPERA Emulsions. Jan Lenkeit. Hamburg Student Seminar, 12 June 2008. Institut für Experimentalphysik Forschungsgruppe Neutrinophysik

The OPERA Emulsions. Jan Lenkeit. Hamburg Student Seminar, 12 June 2008. Institut für Experimentalphysik Forschungsgruppe Neutrinophysik The OPERA Emulsions Jan Lenkeit Institut für Experimentalphysik Forschungsgruppe Neutrinophysik Hamburg Student Seminar, 12 June 2008 1/43 Outline The OPERA experiment Nuclear emulsions The OPERA emulsions

More information

On the first Townsend coefficient at high electric field

On the first Townsend coefficient at high electric field 1 On the first Townsend coefficient at high electric field Yu.I. Davydov Joint Institute for Nuclear Research, 141980, Dubna, Russia arxiv:physics/0409156v2 [physics.ins-det] 18 Sep 2006 Abstract Based

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

First Measurements with U-probe on the COMPASS Tokamak

First Measurements with U-probe on the COMPASS Tokamak WDS'13 Proceedings of Contributed Papers, Part II, 109 114, 2013. ISBN 978-80-7378-251-1 MATFYZPRESS First Measurements with U-probe on the COMPASS Tokamak K. Kovařík, 1,2 I. Ďuran, 1 J. Stöckel, 1 J.Seidl,

More information

"""""""FGRCTVOGPV"QH"GNGEVTKECN"GPIKPGGTKPI"

FGRCTVOGPVQHGNGEVTKECNGPIKPGGTKPI Introduction to High Voltage Measurement It is essential to measure the voltages and currents accurately ensuring perfect safety to the personnel and equipment in industrial testing and research laboratories.

More information

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION David Bernard Dage Precision Industries Inc. Fremont, California d.bernard@dage-group.com ABSTRACT The x-ray inspection of PCB assembly processes

More information

Sound absorption and acoustic surface impedance

Sound absorption and acoustic surface impedance Sound absorption and acoustic surface impedance CHRISTER HEED SD2165 Stockholm October 2008 Marcus Wallenberg Laboratoriet för Ljud- och Vibrationsforskning Sound absorption and acoustic surface impedance

More information

Archimedes Palimpsest Metadata Standard XRF Extensions DRAFT

Archimedes Palimpsest Metadata Standard XRF Extensions DRAFT Archimedes Palimpsest Metadata Standard XRF Extensions DRAFT [These metadata extensions to the Archimedes Palimpsest Metadata Standard 1.0X are currently under review. Comments may be provided to Bob Morton

More information

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,

More information

Automotive Electronics Council Component Technical Committee

Automotive Electronics Council Component Technical Committee AEC - Q101-005 - REV- ATTACHMENT 5 AEC - Q101-005 Rev- CAPACITIVE DISCHARGE MODEL (CDM) ELECTROSTATIC DISCHARGE (ESD) TEST METHOD - 005 DISCRETE COMPONENT CHARGED DEVICE MODEL (CDM) ELECTROSTATIC DISCHARGE

More information

Solar Photovoltaic (PV) Cells

Solar Photovoltaic (PV) Cells Solar Photovoltaic (PV) Cells A supplement topic to: Mi ti l S Micro-optical Sensors - A MEMS for electric power generation Science of Silicon PV Cells Scientific base for solar PV electric power generation

More information

Single Electron Detection with the Large Volume Spherical Proportional Counter Ilias Savvidis

Single Electron Detection with the Large Volume Spherical Proportional Counter Ilias Savvidis Single Electron Detection with the Large Volume Spherical Proportional Counter Ilias Savvidis Collaboration S. Andriamonje, S. Aune, E. Bougamont, M. Chapelier, P. Colas, J. Derré, E. Ferrer, G. Gerbier,

More information

LUXEON LEDs. Circuit Design and Layout Practices to Minimize Electrical Stress. Introduction. Scope LED PORTFOLIO

LUXEON LEDs. Circuit Design and Layout Practices to Minimize Electrical Stress. Introduction. Scope LED PORTFOLIO LED PORTFOLIO LUXEON LEDs Circuit Design and Layout Practices to Minimize Electrical Stress Introduction LED circuits operating in the real world can be subjected to various abnormal electrical overstress

More information

PREFAULT MONITOR FOR AIR COOLED GENERATORS

PREFAULT MONITOR FOR AIR COOLED GENERATORS PREFAULT MONITOR FOR AIR COOLED GENERATORS Steve Kilmartin Utility Products Specialist Environment One Corporation 2773 Balltown Road Schenectady, NY 129 518 346-6161 George F. Skala Senior Engineer Environment

More information

An option for the SHiP Muon Detector: Scintillator bars with WLS fibers and SiPMs readout

An option for the SHiP Muon Detector: Scintillator bars with WLS fibers and SiPMs readout An option for the SHiP Muon Detector: Scintillator bars with WLS fibers and SiPMs readout M. Anelli, W. Baldini, P. Ciambrone, M. Dallavalle, F. Fabbri, G. Lanfranchi, A. Montanari INFN-LNF, INFN-Ferrara,

More information

Diagnostics. Electric probes. Instituto de Plasmas e Fusão Nuclear Instituto Superior Técnico Lisbon, Portugal http://www.ipfn.ist.utl.

Diagnostics. Electric probes. Instituto de Plasmas e Fusão Nuclear Instituto Superior Técnico Lisbon, Portugal http://www.ipfn.ist.utl. C. Silva Lisboa, Jan. 2014 IST Diagnostics Electric probes Instituto de Plasmas e Fusão Nuclear Instituto Superior Técnico Lisbon, Portugal http://www.ipfn.ist.utl.pt Langmuir probes Simplest diagnostic

More information

A receiver TDC chip set for accurate pulsed time-of-flight laser ranging

A receiver TDC chip set for accurate pulsed time-of-flight laser ranging A receiver TDC chip set for accurate pulsed time-of-flight laser ranging Juha Kostamovaara, Sami Kurtti, Jussi-Pekka Jansson University of Oulu, Department of Electrical Engineering, Electronics Laboratory,

More information

Assembly of LPCC Packages AN-0001

Assembly of LPCC Packages AN-0001 Assembly of LPCC Packages AN-0001 Surface Mount Assembly and Handling of ANADIGICS LPCC Packages 1.0 Overview ANADIGICS power amplifiers are typically packaged in a Leadless Plastic Chip Carrier (LPCC)

More information

AC-DC Converter Application Guidelines

AC-DC Converter Application Guidelines AC-DC Converter Application Guidelines 1. Foreword The following guidelines should be carefully read prior to converter use. Improper use may result in the risk of electric shock, damaging the converter,

More information

Balancing the Electrical and Mechanical Requirements of Flexible Circuits. Mark Finstad, Applications Engineering Manager, Minco

Balancing the Electrical and Mechanical Requirements of Flexible Circuits. Mark Finstad, Applications Engineering Manager, Minco Balancing the Electrical and Mechanical Requirements of Flexible Circuits Mark Finstad, Applications Engineering Manager, Minco Table of Contents Abstract...............................................................................................

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

Application Note AN1

Application Note AN1 TAKING INVENTIVE STEPS IN INFRARED. MINIATURE INFRARED GAS SENSORS GOLD SERIES UK Patent App. No. 799A USA Patent App. No. 9/78,7 World Patents Pending SENSOR OVERVIEW Application Note AN The Dynament

More information

Electron Charge to Mass Ratio Matthew Norton, Chris Bush, Brian Atinaja, Becker Steven. Norton 0

Electron Charge to Mass Ratio Matthew Norton, Chris Bush, Brian Atinaja, Becker Steven. Norton 0 Electron Charge to Mass Ratio Matthew Norton, Chris Bush, Brian Atinaja, Becker Steven Norton 0 Norton 1 Abstract The electron charge to mass ratio was an experiment that was used to calculate the ratio

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

PRELIMINARY. J-Series High PDE and Timing Resolution, TSV Package. High-Density Fill Factor Silicon Photomultipliers. Overview

PRELIMINARY. J-Series High PDE and Timing Resolution, TSV Package. High-Density Fill Factor Silicon Photomultipliers. Overview High-Density Fill Factor Silicon Photomultipliers SensL s J-Series low-light sensors feature an industry-leading low dark count rate and a high PDE that extends much further into the blue part of the spectrum

More information

Cambridge International Examinations Cambridge International General Certificate of Secondary Education

Cambridge International Examinations Cambridge International General Certificate of Secondary Education Cambridge International Examinations Cambridge International General Certificate of Secondary Education *0123456789* PHYSICS 0625/04 Paper 4 Theory (Extended) For Examination from 2016 SPECIMEN PAPER 1

More information

RF data receiver super-reactive ASK modulation, low cost and low consumption ideal for Microchip HCS KEELOQ decoder/encoder family. 0.

RF data receiver super-reactive ASK modulation, low cost and low consumption ideal for Microchip HCS KEELOQ decoder/encoder family. 0. Receiver AC-RX2/CS RF data receiver super-reactive ASK modulation, low cost and low consumption ideal for Microchip HCS KEELOQ decoder/encoder family. Pin-out 38.1 3 Component Side 1 2 3 7 11 13 14 15

More information

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator. PHYS 222 Spring 2012 Final Exam Closed books, notes, etc. No electronic device except a calculator. NAME: (all questions with equal weight) 1. If the distance between two point charges is tripled, the

More information

Fabrication of Complex Circuit Using Electrochemical Micromachining on Printed Circuit Board (PCB)

Fabrication of Complex Circuit Using Electrochemical Micromachining on Printed Circuit Board (PCB) 5 th International & 26 th All India Manufacturing Technology, Design and Research Conference (AIMTDR 2014) December 12 th 14 th, 2014, IIT Guwahati, Assam, India Fabrication of Complex Circuit Using Electrochemical

More information

Ion Beam Sputtering: Practical Applications to Electron Microscopy

Ion Beam Sputtering: Practical Applications to Electron Microscopy Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a

More information

Gain Measurements of a GridPix detector operated in Ar/iC 4 H 10 at different pressures

Gain Measurements of a GridPix detector operated in Ar/iC 4 H 10 at different pressures Gain Measurements of a GridPix detector operated in Ar/iC 4 H 10 at different pressures Analysis of data recorded at the Nikhef Detector R&D-group Eric Drechsler Universiteit van Amsterdam 16. January

More information

CHAPTER 7 SCINTILLATION COUNTING

CHAPTER 7 SCINTILLATION COUNTING CHAPTER 7 SCINTILLATION COUNTING Radiation of various types is widely utilized for non-destructive inspection and testing such as in medical diagnosis, industrial inspection, material analysis and other

More information

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment. Helium-Neon Laser Experiment objectives: assemble and align a 3-mW HeNe laser from readily available optical components, record photographically the transverse mode structure of the laser output beam,

More information

AORC Technical meeting 2014

AORC Technical meeting 2014 http : //www.cigre.org B1-196 AORC Technical meeting 214 Evaluation of Insulating Materials for HVDC Transmission Cable using Space Charge Measurement Y. TANAKA Tokyo City University Japan SUMMARY Some

More information

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change

More information

T(CR)3IC Testbed for Coherent Radio Cherenkov Radiation from Cosmic-Ray Induced Cascades

T(CR)3IC Testbed for Coherent Radio Cherenkov Radiation from Cosmic-Ray Induced Cascades T(CR)3IC Testbed for Coherent Radio Cherenkov Radiation from Cosmic-Ray Induced Cascades R. Milinčić1, P. Gorham1, C. Hebert1, S. Matsuno1, P. Miočinović1, M. Rosen1, D. Saltzberg2, G. Varner1 1 University

More information

How Does One Obtain Spectral +Imaging Data

How Does One Obtain Spectral +Imaging Data How Does One Obtain Spectral +Imaging Data What we observe depends on the instruments that one observes with! In x and γ-ray spectroscopy we have a wide variety of instruments with different properties

More information

A study of breakdown limits in microstrip gas counters with preamplification structures

A study of breakdown limits in microstrip gas counters with preamplification structures Nuclear Instruments and Methods in Physics Research A 416 (1998) 23 31 A study of breakdown limits in microstrip gas counters with preamplification structures P. Fonte, V. Peskov, B.D. Ramsey * LIP/Coimbra

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF) Advanced Physics Laboratory XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF) Bahia Arezki Contents 1. INTRODUCTION... 2 2. FUNDAMENTALS... 2 2.1 X-RAY PRODUCTION... 2 2. 1. 1 Continuous radiation...

More information

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.

More information

Physics 221 Experiment 5: Magnetic Fields

Physics 221 Experiment 5: Magnetic Fields Physics 221 Experiment 5: Magnetic Fields August 25, 2007 ntroduction This experiment will examine the properties of magnetic fields. Magnetic fields can be created in a variety of ways, and are also found

More information

Lab 4: Magnetic Force on Electrons

Lab 4: Magnetic Force on Electrons Lab 4: Magnetic Force on Electrons Introduction: Forces on particles are not limited to gravity and electricity. Magnetic forces also exist. This magnetic force is known as the Lorentz force and it is

More information

Displays. Cathode Ray Tube. Semiconductor Elements. Basic applications. Oscilloscope TV Old monitors. 2009, Associate Professor PhD. T.

Displays. Cathode Ray Tube. Semiconductor Elements. Basic applications. Oscilloscope TV Old monitors. 2009, Associate Professor PhD. T. Displays Semiconductor Elements 1 Cathode Ray Tube Basic applications Oscilloscope TV Old monitors 2 1 Idea of Electrostatic Deflection 3 Inside an Electrostatic Deflection Cathode Ray Tube Gun creates

More information

Reaction Engineering of Polymer Electrolyte Membrane Fuel Cells

Reaction Engineering of Polymer Electrolyte Membrane Fuel Cells Reaction Engineering of Polymer Electrolyte Membrane Fuel Cells A new approach to elucidate the operation and control of Polymer Electrolyte Membrane (PEM) fuel cells is being developed. A global reactor

More information

RIEGL VZ-400 NEW. Laser Scanners. Latest News March 2009

RIEGL VZ-400 NEW. Laser Scanners. Latest News March 2009 Latest News March 2009 NEW RIEGL VZ-400 Laser Scanners The following document details some of the excellent results acquired with the new RIEGL VZ-400 scanners, including: Time-optimised fine-scans The

More information

Appendix A. An Overview of Monte Carlo N-Particle Software

Appendix A. An Overview of Monte Carlo N-Particle Software Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation

More information

The Ultimate British Valve Tester

The Ultimate British Valve Tester TEST EQUIPMENT The Ultimate British Valve Tester AVO VCM 163 by Kurt Schmid, DH3PJ To evaluate the quality of thermionic valves, ambitious valve testers (USA: tube tester) are able to diagnose a couple

More information

Development of High-Speed High-Precision Cooling Plate

Development of High-Speed High-Precision Cooling Plate Hironori Akiba Satoshi Fukuhara Ken-ichi Bandou Hidetoshi Fukuda As the thinning of semiconductor device progresses more remarkably than before, uniformity within silicon wafer comes to be strongly required

More information

Electronics Technology

Electronics Technology Teacher Assessment Blueprint Electronics Technology Test Code: 5907 / Version: 01 Copyright 2011 NOCTI. All Rights Reserved. General Assessment Information Blueprint Contents General Assessment Information

More information

Flex Circuits for the ATLAS Pixel Detector

Flex Circuits for the ATLAS Pixel Detector Flex Circuits for the ATLAS Pixel Detector P. Skubic University of Oklahoma Outline ATLAS pixel detector ATLAS prototype Flex hybrid designs Performance simulations Performance measurements Wire bonding

More information

Construction of an Alpha- Beta and Gamma-Sensitive Radiation Detector on the Basis of a Low-Cost PIN-Diode

Construction of an Alpha- Beta and Gamma-Sensitive Radiation Detector on the Basis of a Low-Cost PIN-Diode Construction of an Alpha- Beta and Gamma-Sensitive Radiation Detector on the Basis of a Low-Cost PIN-Diode Bernd Laquai, 12.6.2012 Encouraged by the observation, that the Am241 alpha source of an old smoke

More information

Diodes and Transistors

Diodes and Transistors Diodes What do we use diodes for? Diodes and Transistors protect circuits by limiting the voltage (clipping and clamping) turn AC into DC (voltage rectifier) voltage multipliers (e.g. double input voltage)

More information

Charged cable event. 1 Goal of the ongoing investigation. 2 Energy sources for the CDE. Content

Charged cable event. 1 Goal of the ongoing investigation. 2 Energy sources for the CDE. Content Charged cable event David Pommerenke, david_pommerenke@hp.com, 916 785 4550 Last update: Feb.23, 2001 Content Goal Energy sources, which may lead to CDE. Complexity of the different discharge modes. Possible

More information

CHAPTER 2 BASIC PRINCIPLES OF PHOTOMULTIPLIER TUBES 1)-5)

CHAPTER 2 BASIC PRINCIPLES OF PHOTOMULTIPLIER TUBES 1)-5) CHAPTER 2 BASIC PRINCIPLES OF PHOTOMULTIPLIER TUBES 1)-5) A photomultiplier tube is a vacuum tube consisting of an input window, a photocathode, focusing electrodes, an electron multiplier and an anode

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Section 10.0: Electrode Erosion

Section 10.0: Electrode Erosion Section 10.0: Electrode Erosion The primary reason for failure of plasma torches usually involves the inability of the electrodes to operate as they were designed, or operation under adverse conditions.

More information