THE EFFICIENCY DEPENDENCE ON THE ANALYZED SAMPLE CHARACTERISTICS IN GAMMA-RAY SPECTROMETRIC ANALYSIS

Size: px
Start display at page:

Download "THE EFFICIENCY DEPENDENCE ON THE ANALYZED SAMPLE CHARACTERISTICS IN GAMMA-RAY SPECTROMETRIC ANALYSIS"

Transcription

1 THE EFFICIENCY DEPENDENCE ON THE ANALYZED SAMPLE CHARACTERISTICS IN GAMMA-RAY SPECTROMETRIC ANALYSIS LAURENTIU DONE, LIVIU CORNEL TUGULAN, FELICIA DRAGOLICI, CARMELA ALEXANDRU, MARIA SAHAGIA National Institute for Physics and Nuclear Engineering Horia Hulubei 30, Reactorului str., RO , P.O. Box MG-6, Magurele (Ilfov) Romania, EU Received February 10, 2014 This paper proposes a comparative study of a HP-Ge detector efficiencies for various geometrical configurations of the calibration samples with properties close to the environmental ones. The detection efficiency for the studied analysis geometries was calculated using the LabSOCS software (Laboratory Sourceless Calibration Software). This program calculates the detector efficiency versus incident photons energy, for the operator s set analysis geometry, by the Monte Carlo simulation. The detection efficiency variation was studied as function of the analyzed sample shape, height and mass. Key words: gamma-ray spectrometry, environmental samples, detection efficiency. 1. INTRODUCTION In the last couple of decades the communities have become more and more aware of the importance of preserving the natural environment and the deployment of all the human activities with a decreased pollution degree. The monitoring and supervision of all kinds of parameters, which can give us an insight into the Earth health, has become increasingly a more important worldwide activity. As a result, the scientists are trying, on the one hand to discover and devise new methods to measure the environment pollution level, and on the other hand to increase the precision and accuracy of these measurements. The gamma-ray spectrometric analysis of environmental samples (soil, sediment, water, vegetation, air) gives us informations on both natural radionuclides concentration and the time evolution of these concentrations, but mostly we can warn of the occurrence in the environment of anthropogenic radionuclides, due to the human activities. Rom. Journ. Phys., Vol. 59, Nos. 9 10, P , Bucharest, 2014

2 2 The efficiency of gamma-ray spectrometric analysis 1013 The activity calculation of a gamma-ray emitter radionuclide in a sample, by the gamma spectrometric quantitative analysis, is based on the recording of the net area expression A n (E) of the Full-Energy Peak (FEP) of the photon specific energy E, in a gamma-ray spectrometric installation, and its relation with the activity, expressed by equation (1): A ( E ) = Λ I t ε ε f (1) n γ where: Λ is the radionuclide activity from the sample, I γ (Yield) is the emission probability of the photons with energy E, t is the effective time of the spectrum acquisition (live time), ε int is the detector intrinsic efficiency at energy E, ε geo is the detection geometrical efficiency at energy E, f is the coincidence summing correction factor, especially important in the case of the cascade emission of the gamma ray radionuclides. For a sample with a Λ activity and a given detector intrinsic efficiency, a better statistical analysis is obtained if the peak net area of E energy is larger. From equation (1) is noticed that the net peak area is directly proportional to two parameters on which the operator can act: a) the spectrum acquisition time, t; b) the geometrical detection efficiency, ε geo. The gamma-ray spectrum aquisition time can be increased only in the reasonable limits, depending on the needed number of samples to be analyzed in a given period. The second solution to increase the peak net area is to perform the analysis in a larger efficiency geometry. Note: Further, for a given detector, the calculations refer to the total detection efficiency, ε T = ε int ε geo, where the variation is due to the geometrical efficiency, ε geo. This paper especially refers to the environmental samples analysis, having relatively low activity; therefore, in this study the approached geometries are those in which the sample is in contact with the detector (close measurement geometry). O. Sima in the paper [1] and G. Gilmore in the paper [2] mention the importance of the coincidence summing effects in the case of the closest detectorsource geometry and the necessity of applying corrections in activity calculation, with the specific factors of this phenomenon. int geo 2. METHODS AND EQUIPMENT The efficiency was calculated for a hyperpure germanium detector, REGe type (Reverse-Electrode Germanium), GR4020 model, with thin carbonate window (0.6 mm), manufactured by CANBERRA, with the following characteristics: kev energy range; detector diameter: 76 mm, Ge cristal diameter: 61 mm; HPGe cristal length: 65 mm; resolution: kev at 122 kev and

3 1014 Laurentiu Done et al kev at 1332 kev; peak/compton ratio: 58.9; relative efficiency (measured by the manufacturer): 47.5%. The detector is mounted in a CANBERRA manufactured lead shield, model 747, and was ISOXCALL characterized by the manufacturer. The multichannel analyzer is of DSA 1000 type. The acquisition and processing of gamma spectra software is GENIE Canberra company provided us the LabSOCS software, based on the ISOXCALL detector characterization and the MCNP code (Monte Carlo N-Particle), developed by Los Alamos National Laboratory. This software is customized for the mentioned detector and uses the Monte Carlo simulation to calculate the efficiency and the coincidence summing corrections, for various geometries set by the operator or supplied by the manufacturer. 3. RESULTS AND DISCUSSIONS 3.1. THE DETECTION EFFICIENCY DEPENDENCE ON THE SAMPLES GEOMETRICAL SHAPE (i) A hypotethical sample with the following characteristics was analyzed, using the LabSOCS software: density ρ = 1.4 g/cm 3, corresponding approximately to the density of the dry sediment, or the soil samples, with particles size less than 1 mm; sample mass m = 1042 g, which would be required to fill a Marinelli beaker, type 130G to the overall height of 111 mm, which means that the samples height on the flat detector surface would be 36 mm. The detector efficiency was calculated for the following sample s geometries: Marinelli beaker 130G type, D mean inside = 122 mm; plexiglas cylinder with D inside = 122 mm, the bottom thickness of 2.09 mm and the wall thickness of 1 mm, plexiglas density 1.2 g/cm 3 (characteristics for the Sarpagan type box); cube with the side l inside = 92 mm and wall thickness of 2.09 mm; sphere with R inside = 57 mm and wall thickness of 2.09 mm; cylinder with D = 70 mm (diameter of a Sarpagan beaker) and bottom thickness 2.09 mm. The detection efficiency variation versus photon s energy, for these analysed geometries, was graphically represented in Figure 1. For this study energy interest range ( kev), the following remarks are to be done: a) The detection efficiency for the sample analyzed in the Marinelli beaker is the highest; b) In the case of the sample analyzed in cylindrical geometry with the diameter D = 122 mm, the detection efficiency is greater than in the vessel with the diameter D = 70 mm; c) The detection efficiency is greater for cylinder with D = 122 mm than for a cube with the same volume (with side size l = 92 mm);

4 4 The efficiency of gamma-ray spectrometric analysis 1015 d) If one notes with h c the distance between the sample gravity center and the flat top surface of the cylindrical detector, it is observed that the detection efficiency increases as the distance h c is lower. The sample described in the Marinelli beaker, due to its shape (the beaker "envelops" a certain volume of the detector) has the gravity center inside the detector at 7.1 mm below its flat side. In this case, the virtual sample concentrated in the volumic sample s gravity center, are inside the detector, and consequently h c = 7.1 mm. Fig. 1 The detection efficiency dependence on the energy, for samples in different geometries. The detection efficiency calculated in Marinelli and in various diameters of cylindrical geometries, as well as the h c distance for each geometry, corresponding to four significant energies, in the range kev, as presented in Fig. 1, are given in Table 1. Table 1 Calculated efficienciy values for various geometries, for the analysed sample Geometry h c [mm] Eff. at 60 kev Eff. at 120 kev Eff. at 600keV Eff. at 1400 kev Marinelli E E E E-02 D=130 mm E E E E-03 D=122 mm E E E E-03 D=100 mm E E E E-03 D=70 mm E E E E-03

5 1016 Laurentiu Done et al. 5 If one considers the detection efficiency graphical representation of the volume sample above, depending on the h c distance between the detector and the sample s centroid from Figure 2, one may fit an efficiency dependence on the h c represented by the following semiempirical equation, deduced by supposing a relation of the type (2). a1 + a 2 hc ε ( h c ) = a + a h + a h 3 4 c 5 where a 1 a 5 are real coefficients, with h c variation between 10 mm and +200 mm. For example, the detection efficiency at 60 kev, depends on the distance h c by a semiempirical relation of the form: h c ε ( h c ) = (3) h h c 2 c c (2) Fig. 2 The detection efficiency at four energy values as function of the distance h c. (ii) The second hypothetical sample. Let s suppose that, for practical reasons, we can not take a soil sample with a mass of 1042 g; for a soil sample, in the Marinelli 130G type beaker, we need a sample mass greater than 500 g, to reach the detector front. Note: especially for a beryllium or carbonate detector window (for the low energy photons detection), is important that a large part of the sample be located on the detector plane front. In this case we are forced to analyze the sample in a lower geometry efficiency, suppose in a cylindrical vessel. Naturally, the question of choice the optimum diameter of the beaker in which we prepare the sample arises, so that we have the maximum detection efficiency. Using the same software, we calculated that the detection efficiency of the detector above, considering a sample with mass m = 500 g and a density of 1.4 g/cm 3, for cylindrical geometry with the following

6 6 The efficiency of gamma-ray spectrometric analysis 1017 diameters: 70, 80, 100, 120, 140, 160, 180 mm. Figure 3 represents the detection efficiency variation depending on the sample diameter, for a few gamma-ray energies. Fig. 3 The detection efficiency as function of the sample cylindrical geometry diameter. From this graphical representation it results that the detection efficiency is maximum for the cylindrical geometry with a diameter of mm THE DEPENDENCE OF THE DETECTION EFFICIENCY ON THE SAMPLE HEIGHT We consider the gamma-ray spectrometric analysis of a sample with mass of 200 g, in Sarpagan type box with the following dimensions: plexiglass material, density ρ = 1.2 g/cm 3 ; diameter, D = 70 mm; the bottom beaker thickness, G = 2.09 mm. The initial height of the sample is h 1 = 50 mm and density ρ 1 = 1.04 g/cm 3. Suppose that we can compress the sample by 5 mm steps up to h 10 = 5 mm, density of sample taking the values in the Table 2. Table 2 Densitiy values of the sample compressed at various heights Sample height in Sarpagan vessel [mm] Sample density [g/cm 3 ] The calculated values of efficiency at 5 significant energies within the interval 45 kev 1400 kev as function of the sample height is presented in Figure 4.

7 1018 Laurentiu Done et al. 7 Fig. 4 The detection efficiency at 45, 60, 120, 600 and 1400 kev, depending on the sample height in a Sarpagan beaker. From the graphical representation of Figure 4 we can see that the detection efficiency is better with decreasing the sample height. In fact we have a competitive process, which partially compensates: as more we subtract the sample height, as more the radioactive matrix is closer to the detector (h c decreases), but at the same time, increasing the sample density, the photons attenuation in the radioactive matrix, selfattenuation, increases. The graphical representation demonstrates that, within the studied limits, carrying out the whole radioactive mass near the detector (the sample height loss) leads to higher efficiency growth than the attenuation by self-absorption, which would lead to the efficiency decrease. This means that we can analyze a soil or a sediment sample in Sarpagan beaker (cylindrical) with better efficiency, if by some method we compress the sample into a cylindrical mold. The same exercise for a sample in Marinelli-type vessel leads to similar results regarding the increase of the detection efficiency with the sample height decreasing. However, compressing a sample in the form of a Marinelli beaker is difficult to achieve in practice PEAK NET AREA VARIATION WITH THE ANALYZED SAMPLE MASS As a result of the environmental sample preparation according to the specific procedures, they are volumic ones, homogeneous in terms of the contained radionuclides activity distribution. The question is how much of the prepared sample is subjected to analysis so that the accuracy of the analysis to be as good as possible. This is translated into obtaining a gamma spectrum of the net areas peaks as large as possible. From equation (1) it results that a net peak area increases

8 8 The efficiency of gamma-ray spectrometric analysis 1019 proportionally to the sample activity. The activity of a homogeneous sample is an additional magnitude in relation to the mass. If the test sample is not subject to the radionuclides concentration by chemical methods, the net area of a peak is proportional to the mass of the sample. It results that, for the analysis accuracy increase, the sample mass must be as large as posible. On the other hand, with the sample mass increase, and implicit its volume, the detection efficiency is changed. Further on, we want to see how the detection efficiency is changed depending on the sample mass Sample in Sarpagan beaker Let us suppose that there are two series of ten samples analyzed, a series with density of 1.0 g/cm 3 and another with density of 1.4 g/cm 3, in Sarpagan type box with the following dimensions: the inner box diameter, 70 mm; sidewall thickness, 1 mm; bottom box thickness, 2.09 mm; plexiglass density, 1.2 g/cm 3 ; detector sample distance, 0 mm. Samples with densities of ρ 1 = 1.0 g/cm 3 and ρ 2 = 1.4 g/cm 3 with masses within the interval 25 g 250 g, varying in steps of 25 g. For each set of 10 samples, the detection efficiency for the kev incident photon energy range was calculated; Figures 5 present the detection efficiency, depending on the incident photons energy, for the samples with mass m = 25, 125 and 250 g. Fig. 5 The detection efficiency versus energy, for samples of m = 25 g, 125 g and 250 g, with density of 1.0 g/cm 3, in Sarpagan beaker. From the Figures 5 we find that the detection efficiency decreases with sample mass increase. The same efficiency evolution with the sample mass is also observed for the samples of 1.4 g/cm 3 density. It results that in the relation (1), with the mass increase, the net area on the one hand increases with the sample mass, and on the other hand it decreases due to the detection efficiency decrease. With these

9 1020 Laurentiu Done et al. 9 two competitional trends, the next obvious question arises: how appropriate is to increase the mass of an analyzed sample to obtain a higher net peak area, so a better statistical analysis. In equation (1), if we neglect the coincidence summation phenomena, the net area of the E peak energy, for the sample mass m 1 = 25 g is given by the relation: n ( ) ( ) ( ) A m =Λ m I t ε ε m (4) 1 1 γ int geo 1 where: Λ(m 1 ) is the sample activity of mass m 1 and ε geo (m 1 ) is the detection geometrical efficiency for the sample m 1. If all the samples of the same density were obtained from a homogeneous radioactive matrix, the ratio between the activity of any two samples is equal to the ratio of these samples masses and the equation (4) written for a some m mass, becomes: ( m) ( m ) m ε geo An ( m) = An ( m1 ) (5) m ε 1 geo 1 If in the relation (5) we consider that A n (m 1 ), m 1 and ε geo (m 1 ) are constant quantities characterizing the sample of mass m 1 = 25 g, then the net area A n (m) of an uncertain mass sample depends only on the following product: ( ) ( ) P m = m ε m (6) For the two samples series with density ρ = 1 g/cm 3 and ρ = 1.4 g/cm 3, the product variations P(m), for a few energy lines is graphically represented in Figures 6 and 7. geo Fig. 6 The P(m) function for energy peaks E = 45, 60, 120, 600 and 1400 kev, sample density ρ = 1 g/cm 3 - Sarpagan beaker.

10 10 The efficiency of gamma-ray spectrometric analysis 1021 Fig. 7 P(m) function for energy peaks E = 45, 60, 120, 600 and 1400 kev, sample density ρ = 1.4 g/cm 3 - Sarpagan beaker Sample in Marinelli beaker 130G type Let us suppose that the same two samples, with densities of 1.0 and 1.4 g/cm 3 are analyzed in Marinelli geometry, 130G type, in contact with the detector. The samples mass (Table 3) was established so that it meets the following conditions: a) the m 1 samples masses, 535 g and respectively 750 g, were chosen so as to fill the beaker to a reasonable analysis height (15 mm on the detector front, mean overall height of 90 mm in Marinelli beaker); b) from one sample to the next, the mass increased so that the sample height rised steadily with 5 mm; c) the beaker size does not allow a total sample height greater than 140 mm. Table 3 The sample masses calculated for various heights of the samples No. Sample height [mm] m (ρ = 1.0 g/cm 3 ) [g] m (ρ = 1.4 g/cm 3 ) [g] The detection efficiency was calculated for the kev energy range, for all samples. For example, in Figures 8, a) and b), the detection efficiency is graphically represented depending on the photon s energy, for samples no. 1, 5 and

11 1022 Laurentiu Done et al of each density group. It was observed a decrease of the detection efficiency when the sample mass increases, similar to the samples contained in Sarpagan beaker. a) b) Fig. 8 The detector efficiency depending on the energy for: a) m = 535, 761 and 1048 g, the sample density of 1.0 g/cm3; b) m = 750, 1066 and 1468 g, the sample density of 1.4 g/cm3, in the Marinelli beaker. In Figures 9 and 10, the product P(m) variation described by the equation (6), which characterizes the net peak area evolution depending on the sample mass, for the samples with a density of 1 g/cm3 and 1.4 g/cm3 respectively, analyzed in Marinelli geometry, is graphically presented. In both cases, the Sarpagan and Marinelli - 130G beakers, is observed that with increasing the sample mass, within the limits imposed by the container volume, the net peak area increases, too. It follows that for the samples with a density of 1.0 g/cm3 and 1.4 g/cm3 we get a bigger net peak area, so a better statistical analysis, as that the sample mass is higher. Fig. 9 The P(m) function for energy peaks E = 45, 60, 120, 600 and 1400 kev, sample density ρ = 1.0 g/cm3 Marinelli beaker.

12 12 The efficiency of gamma-ray spectrometric analysis 1023 Fig. 10 The P(m) function for energy peaks E = 45, 60, 120, 600 and 1400 kev, sample density ρ = 1.4 g/cm 3 Marinelli beaker. The graphical representations show that the net peak area increasing has the tendency of saturation, the trend being increasing with the sample density and the peaks energy decrease. This trend shows, that an increase in sample mass over a certain value to get a better statistics is becoming useless for the purpose. Example: for the sample with density of 1.4 g/cm 3 in Marinelli beaker, the increase of the sample mass over aprox. 990 g, leads to an insignificant increase of the net peak area of 45 kev THE EFFICIENCY UNCERTAINTY CALCULATED BY THE LABSOCS SOFTWARE The technical documentation for validation and verification of the software [3] contains the comparison data between the efficiency calculated with LabSOCS and the efficiency calculated with volumic standard sources (detector with 42% relative efficiency, sources with 1.15 g/cm 3 density, in Marinelli beaker of 1 liter and in cylindrical geometry about 125 ml, placed on the detector). The conclusion of the validation tests for the geometries used in the laboratory, according to [3] is: the standard deviation average is 7.1% for kev, 5.5% for kev and 4.3% for E>400 kev. It is noticed that at low energies, i.e kev, Am-241, the efficiency calculated by LabSOCS is closer to the efficiency value determined with standard sources in cylindrical geometry, the difference being of 9% against 20%, for Marinelli geometry. At high energies, over 800 kev, i.e. Co-60, Y-88, the tendency is reverse. The final conclusion is that the deviations between the two manners of calculating the efficiencies are larger than two standard deviations (k=2); it means that some other supplementary corrections, such as coincidence summation etc., sould be taken into account.

13 1024 Laurentiu Done et al. 13 The same type of tests were performed by J.P.Steward and D. Groff, in order to verify the LabSOCS calculated efficiency using volumic standard cylindrical sources, as described in paper [4]. The results are similar with those obtained in the technical validation documentation. 4. CONCLUSIONS From the efficiency calculations performed with the LabSOCS software, by Monte Carlo simulation, for the volumic samples analyzed in contact with the detector, the following conclusions can be drawn: i) the detection efficiency depends on the sample geometrical shape, observing that for the constant mass samples analyzed in Marinelli, cylindrical, cubical and spherical geometries, the detection efficiency is maximum in the case of the Marinelli geometry; ii) the detection efficiency increases with the decreasing of the distance between the sample gravity center and the detector plane surface; iii) the detection efficiency for the samples analyzed in the cylindrical geometry depends on the sample diameter, an optimum sample diameter interval is characteristic for each type of the detector; iv) for the Sarpagan and the Marinelli analysis geometries type, the analysis statistics improves (the peak net area increases) with the sample mass increasing, in the beaker dimensions limits. REFERENCES 1. O. Sima, Efficiency Calculation of Gamma Detectors by Monte Carlo Methods, Enciclopedia of Analytical Chemistry, JohnWilley & Sons, Ltd. (2012). 2. G. Gilmore, Practical Gamma-ray Spectrometry, 2 nd Edition, John Wiley & Sons, Ltd (2008). 3. CANBERRA Industries, Inc.: ISOCS/LabSOCS, Validation & Verification Manual (2002). 4. John P. Stewart (Sequoyah Nuclear Power Plant, Tennessee Valley Authority), David Groff (Canberra Industries, Meriden, CT., USA): LabSOCS vs. SOURCE-BASED GAMMA-RAY DETECTOR EFFICIENCY, COMPARISONS FOR NUCLEAR POWER PLANT GEOMETRIES, 48th Annual Radiobioassay & Radiochemical Measurements Conference, Knoxville, Tennessee (November 2002). 5. A. Luca, B. Neacsu, A. Antohe, M. Sahagia, Calibration of high and low resolution gamma-ray spectrometers, Romanian Reports in Physics, 64, (2012). 6. R. Suvaila, O. Sima, Complex Gamma Ray spectra analysis, Romanian Reports in Physics, 63, (2011). 7. E. Robu, C. Giovani, Gamma-ray self-attenuation corrections in environmental samples, Romanian Reports in Physics, 61, (2009).

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,

More information

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector

More information

ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE

ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE Gevaldo Lisboa de Almeida, Carlos Feu Alvim Agência Brasileiro-Argentina de

More information

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation

More information

Appendix A. An Overview of Monte Carlo N-Particle Software

Appendix A. An Overview of Monte Carlo N-Particle Software Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation

More information

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY I. BIBICU 1, G. NICOLESCU 2, L. CIOLACU 2, L. SERBINA 2 1 National Institute for Materials Physics, Bucharest 77125, Romania, bibicu@infim.ro 2

More information

Application Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System

Application Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System Application Note Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System Based on the German KTA 1503.1 and respective international regulations,

More information

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Ronald M. Keyser, Timothy R. Twomey, Sam Hitch ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831

More information

Chapter Test B. Chapter: Measurements and Calculations

Chapter Test B. Chapter: Measurements and Calculations Assessment Chapter Test B Chapter: Measurements and Calculations PART I In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1.

More information

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner

More information

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy An Innovative Method for Dead Time Correction in Nuclear Spectroscopy Upp, Daniel L.; Keyser, Ronald M.; Gedcke, Dale A.; Twomey, Timothy R.; and Bingham, Russell D. PerkinElmer Instruments, Inc. ORTEC,

More information

WM2012 Conference, February 26 -March 1, 2012, Phoenix, Arizona, USA

WM2012 Conference, February 26 -March 1, 2012, Phoenix, Arizona, USA Upgrade and Certification of the Los Alamos National Laboratory SHENC 2011-12270 ABSTRACT S.B. Stanfield 1, M. Villani 1, P. T. Barton 1, C. Gerlock 1, D. Nakazawa 1, R. C. Baumann 1, R. Mowry 1, J.P Harvill

More information

Tutorial 4.6 Gamma Spectrum Analysis

Tutorial 4.6 Gamma Spectrum Analysis Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning

More information

Chapter 3 Student Reading

Chapter 3 Student Reading Chapter 3 Student Reading If you hold a solid piece of lead or iron in your hand, it feels heavy for its size. If you hold the same size piece of balsa wood or plastic, it feels light for its size. The

More information

Evaluation of 55-gallon Lead-Lined Calibration for the Gamma Modality of the High Efficiency Neutron Counter

Evaluation of 55-gallon Lead-Lined Calibration for the Gamma Modality of the High Efficiency Neutron Counter Evaluation of 55-gallon Lead-Lined Calibration for the Gamma Modality of the High Efficiency Neutron Counter S.B. Stanfield 1, J.R. Wachter 1, D.L. Cramer 1, S. L. Chambers 1, R.C. Baumann 1, J.P Harvill

More information

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA ABSTRACT Comparison of Activity Determination of Radium 226 in FUSRAP Soil using Various Energy Lines - 12299 Brian Tucker*, Jough Donakowski**, David Hays*** *Shaw Environmental & Infrastructure, Stoughton,

More information

A methodological approach for sediment core dating

A methodological approach for sediment core dating Radioprotection - Colloques, volume 37, Cl (2002) Cl-1009 A methodological approach for sediment core dating R. Ocone, M. Belli, L. Schiozzi 1 and A. Fazio 2 ANPA, Agenzia Nazionale perla Protezione dell'ambiente,

More information

Efficiency calibration of the Ge(Li) detector of the BIPM for SIR-type ampoules

Efficiency calibration of the Ge(Li) detector of the BIPM for SIR-type ampoules Efficiency calibration of the Ge(Li) detector of the BIPM for SIR-type ampoules C. Michotte Bureau international des poids et mesures, Pavillon de Breteuil, Sèvres. Abstract The absolute full-energy peak

More information

A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials

A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials Illicit Trafficking, Sub/cross-national threats, Poster presentation A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials Ronald M. Keyser, Timothy R. Twomey,

More information

INVESTIGATION OF FALLING BALL VISCOMETRY AND ITS ACCURACY GROUP R1 Evelyn Chou, Julia Glaser, Bella Goyal, Sherri Wykosky

INVESTIGATION OF FALLING BALL VISCOMETRY AND ITS ACCURACY GROUP R1 Evelyn Chou, Julia Glaser, Bella Goyal, Sherri Wykosky INVESTIGATION OF FALLING BALL VISCOMETRY AND ITS ACCURACY GROUP R1 Evelyn Chou, Julia Glaser, Bella Goyal, Sherri Wykosky ABSTRACT: A falling ball viscometer and its associated equations were studied in

More information

Chapter 1: Chemistry: Measurements and Methods

Chapter 1: Chemistry: Measurements and Methods Chapter 1: Chemistry: Measurements and Methods 1.1 The Discovery Process o Chemistry - The study of matter o Matter - Anything that has mass and occupies space, the stuff that things are made of. This

More information

arxiv:1401.3092v1 [physics.ins-det] 14 Jan 2014

arxiv:1401.3092v1 [physics.ins-det] 14 Jan 2014 Monte Carlo Simulation of Indoor External Exposure due to Gamma-emitting Radionuclides in Building Materials arxiv:1401.3092v1 [physics.ins-det] 14 Jan 2014 DENG Jun 1) CAO Lei SU Xu 2) Key Laboratory

More information

Density Determinations and Various Methods to Measure

Density Determinations and Various Methods to Measure Density Determinations and Various Methods to Measure Volume GOAL AND OVERVIEW This lab provides an introduction to the concept and applications of density measurements. The densities of brass and aluminum

More information

Introduction to Geiger Counters

Introduction to Geiger Counters Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists

More information

S.3.6. BULK DENSITY AND TAPPED DENSITY OF POWDERS. Final text for addition to The International Pharmacopoeia

S.3.6. BULK DENSITY AND TAPPED DENSITY OF POWDERS. Final text for addition to The International Pharmacopoeia March 2012 S.3.6. BULK DENSITY AND TAPPED DENSITY OF POWDERS Final text for addition to The International Pharmacopoeia This monograph was adopted at the Forty-sixth WHO Expert Committee on Specifications

More information

MIDLAND ISD ADVANCED PLACEMENT CURRICULUM STANDARDS AP ENVIRONMENTAL SCIENCE

MIDLAND ISD ADVANCED PLACEMENT CURRICULUM STANDARDS AP ENVIRONMENTAL SCIENCE Science Practices Standard SP.1: Scientific Questions and Predictions Asking scientific questions that can be tested empirically and structuring these questions in the form of testable predictions SP.1.1

More information

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino

More information

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM Force measurement Forces VECTORIAL ISSUES In classical mechanics, a force is defined as "an action capable of modifying the quantity of movement of a material point". Therefore, a force has the attributes

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

Laue lens for Nuclear Medicine

Laue lens for Nuclear Medicine Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of

More information

Measurement of Germanium Detector Efficiency

Measurement of Germanium Detector Efficiency Measurement of Germanium Detector Efficiency Marcus H. Wiggs 2009 Notre Dame Physics REU Advisor: Dr. Philippe Collon Mentors: Matthew Bowers, Daniel Robertson, Chris Schmitt ABSTRACT: A possible discrepancy

More information

Proton Nuclear Magnetic Resonance Spectroscopy

Proton Nuclear Magnetic Resonance Spectroscopy Proton Nuclear Magnetic Resonance Spectroscopy Introduction: The NMR Spectrum serves as a great resource in determining the structure of an organic compound by revealing the hydrogen and carbon skeleton.

More information

Application Note. ipix A Gamma imager to support various applications

Application Note. ipix A Gamma imager to support various applications Application Note ipix A Gamma imager to support various applications Introduction ipix is a unique gamma imager that quickly locates low level radioactive sources from a distance and estimates the dose

More information

EDXRF of Used Automotive Catalytic Converters

EDXRF of Used Automotive Catalytic Converters EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,

More information

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry John Keightley NSUF 18 May 2010 Acknowledgement: Several discussions with Stefaan Pommé, IRMM, EC-JRC, Geel Some diagrams

More information

CFD SIMULATION OF SDHW STORAGE TANK WITH AND WITHOUT HEATER

CFD SIMULATION OF SDHW STORAGE TANK WITH AND WITHOUT HEATER International Journal of Advancements in Research & Technology, Volume 1, Issue2, July-2012 1 CFD SIMULATION OF SDHW STORAGE TANK WITH AND WITHOUT HEATER ABSTRACT (1) Mr. Mainak Bhaumik M.E. (Thermal Engg.)

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues

Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues Journal of mathematics and computer science 9 (2014), 41-45 Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues S. A. Mahdipour 1, H. R. Abdi Roknabadi

More information

PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE

PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE Test Procedure for PARTICLE SIZE ANALYSIS OF SOILS TxDOT Designation: Tex-110-E Effective Date: August 1999 1. SCOPE 1.1 This method covers the quantitative determination of the distribution of particle

More information

ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY. Donald L. Horrocks

ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY. Donald L. Horrocks ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY Donald L. Horrocks Nuclear Systems Operations Beckman Instruments, Inc. SmithKlirie Beckman Irvine,

More information

Introduction to the Monte Carlo method

Introduction to the Monte Carlo method Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw

More information

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics

More information

XI / PHYSICS FLUIDS IN MOTION 11/PA

XI / PHYSICS FLUIDS IN MOTION 11/PA Viscosity It is the property of a liquid due to which it flows in the form of layers and each layer opposes the motion of its adjacent layer. Cause of viscosity Consider two neighboring liquid layers A

More information

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY 243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution

More information

Algebra 1 2008. Academic Content Standards Grade Eight and Grade Nine Ohio. Grade Eight. Number, Number Sense and Operations Standard

Algebra 1 2008. Academic Content Standards Grade Eight and Grade Nine Ohio. Grade Eight. Number, Number Sense and Operations Standard Academic Content Standards Grade Eight and Grade Nine Ohio Algebra 1 2008 Grade Eight STANDARDS Number, Number Sense and Operations Standard Number and Number Systems 1. Use scientific notation to express

More information

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness

More information

GAMMA-RAY SPECTRA REFERENCES

GAMMA-RAY SPECTRA REFERENCES GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.

More information

Injection moulding and modelling on a micro scale

Injection moulding and modelling on a micro scale Injection moulding and modelling on a micro scale Technology Update Injection moulding and welding of plastics 11 November 2014 Research Projects (National / European) Micro/Nano/Multimaterial Manufacturing

More information

Solving Simultaneous Equations and Matrices

Solving Simultaneous Equations and Matrices Solving Simultaneous Equations and Matrices The following represents a systematic investigation for the steps used to solve two simultaneous linear equations in two unknowns. The motivation for considering

More information

Chapter 1 Units, Physical Quantities, and Vectors

Chapter 1 Units, Physical Quantities, and Vectors Chapter 1 Units, Physical Quantities, and Vectors 1 The Nature of Physics Physics is an experimental science. Physicists make observations of physical phenomena. They try to find patterns and principles

More information

p atmospheric Statics : Pressure Hydrostatic Pressure: linear change in pressure with depth Measure depth, h, from free surface Pressure Head p gh

p atmospheric Statics : Pressure Hydrostatic Pressure: linear change in pressure with depth Measure depth, h, from free surface Pressure Head p gh IVE1400: n Introduction to Fluid Mechanics Statics : Pressure : Statics r P Sleigh: P..Sleigh@leeds.ac.uk r J Noakes:.J.Noakes@leeds.ac.uk January 008 Module web site: www.efm.leeds.ac.uk/ive/fluidslevel1

More information

Math Matters: Dissecting Hydrometer Calculations

Math Matters: Dissecting Hydrometer Calculations Math Matters: Dissecting Hydrometer Calculations By Jonathan Sirianni, Laboratory Assessor In the 16th century Galileo and the Tower of Pisa made famous the fact that two objects, no matter their mass,

More information

Vector Treasure Hunt Teacher s Guide

Vector Treasure Hunt Teacher s Guide Vector Treasure Hunt Teacher s Guide 1.0 Summary Vector Treasure Hunt is the first activity to be done after the Pre-Test. This activity should take approximately 30 minutes. 2.0 Learning Goals Driving

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

Physics 181- Summer 2011 - Experiment #8 1 Experiment #8, Measurement of Density and Archimedes' Principle

Physics 181- Summer 2011 - Experiment #8 1 Experiment #8, Measurement of Density and Archimedes' Principle Physics 181- Summer 2011 - Experiment #8 1 Experiment #8, Measurement of Density and Archimedes' Principle 1 Purpose 1. To determine the density of a fluid, such as water, by measurement of its mass when

More information

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.

More information

For convenience, we may consider an atom in two parts: the nucleus and the electrons.

For convenience, we may consider an atom in two parts: the nucleus and the electrons. Atomic structure A. Introduction: In 1808, an English scientist called John Dalton proposed an atomic theory based on experimental findings. (1) Elements are made of extremely small particles called atoms.

More information

THE LABORATORY NOTEBOOK

THE LABORATORY NOTEBOOK THE LABORATORY NOTEBOOK In scientific work keeping a permanent record of all raw data, observations, calculations, et cetera obtained during an experiment is important. Therefore, a student must become

More information

Clearance Laboratory. Capability and measurement sensitivity DD-R-15(EN)

Clearance Laboratory. Capability and measurement sensitivity DD-R-15(EN) DD-R-15(EN) Clearance Laboratory Capability and measurement sensitivity Per Hedemann Jensen, Bente Lauridsen, João Silva, Jens Søgaard-Hansen, Lisbeth Warming Danish Decommissioning, Roskilde September

More information

Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software

Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software ORTEC AN55 APPLICATION NOTE Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software 1. Introduction Quality Assurance (QA) measurements are made by analysis laboratories

More information

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün N.B. A shorter version of this article was published in the ICOM-CC Ethnographic Conservation Newsletter, Number 30, January 2009, pp. 5-8. The Limitations of Hand-held XRF Analyzers as a Quantitative

More information

MECHANICAL PRINCIPLES HNC/D MOMENTS OF AREA. Define and calculate 1st. moments of areas. Define and calculate 2nd moments of areas.

MECHANICAL PRINCIPLES HNC/D MOMENTS OF AREA. Define and calculate 1st. moments of areas. Define and calculate 2nd moments of areas. MECHANICAL PRINCIPLES HNC/D MOMENTS OF AREA The concepts of first and second moments of area fundamental to several areas of engineering including solid mechanics and fluid mechanics. Students who are

More information

Computer Animation of Extensive Air Showers Interacting with the Milagro Water Cherenkov Detector

Computer Animation of Extensive Air Showers Interacting with the Milagro Water Cherenkov Detector Computer Animation of Extensive Air Showers Interacting with the Milagro Water Cherenkov Detector Miguel F. Morales Department of Physics, University of California, Santa Cruz, CA 95064, USA We employ

More information

MONTE CARLO SIMULATION OF THE FULL MULTIPLICITY DISTRIBUTIONS MEASURED WITH A PASSIVE COUNTER

MONTE CARLO SIMULATION OF THE FULL MULTIPLICITY DISTRIBUTIONS MEASURED WITH A PASSIVE COUNTER International Conference on Mathematics, Computational Methods & Reactor Physics (M&C 9) Saratoga Springs, New York, May 3-7, 9, on CD-ROM, American Nuclear Society, LaGrange Park, IL (9) MONTE CARLO SIMULATION

More information

Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to:

Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to: I. OBJECTIVE OF THE EXPERIMENT. Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to: 1) Viscosity of gas (cf. "Viscosity of gas" experiment)

More information

ENRICHMENT AND URANIUM MASS FROM NMIS FOR HEU METAL

ENRICHMENT AND URANIUM MASS FROM NMIS FOR HEU METAL ABSTRACT ENRICHMENT AND URANIUM MASS FROM NMIS FOR HEU METAL J. K. Mattingly, T. E. Valentine, J. T. Mihalczo, L. G. Chiang, R. B. Perez Oa Ridge National Laboratory Oa Ridge, Tennessee 37831-6004 Recently,

More information

Secondary Ion Mass Spectrometry

Secondary Ion Mass Spectrometry Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown,

More information

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in

More information

NOTE: FOR PROJECTS REQUIRING CONTRACTOR MIX DESIGN, THE DESIGN PROCEDURES ARE SPECIFIED IN THE SPECIAL PROVISIONS OF THE CONTRACT.

NOTE: FOR PROJECTS REQUIRING CONTRACTOR MIX DESIGN, THE DESIGN PROCEDURES ARE SPECIFIED IN THE SPECIAL PROVISIONS OF THE CONTRACT. September 1, 2003 CONCRETE MANUAL 5-694.300 MIX DESIGN 5-694.300 NOTE: FOR PROJECTS REQUIRING CONTRACTOR MIX DESIGN, THE DESIGN PROCEDURES ARE SPECIFIED IN THE SPECIAL PROVISIONS OF THE CONTRACT. 5-694.301

More information

GRADATION OF AGGREGATE FOR CONCRETE BLOCK

GRADATION OF AGGREGATE FOR CONCRETE BLOCK GRADATION OF AGGREGATE FOR CONCRETE BLOCK Although numerous papers have been written concerning the proper gradation for concrete mixes, they have generally dealt with plastic mixes, and very little published

More information

When the fluid velocity is zero, called the hydrostatic condition, the pressure variation is due only to the weight of the fluid.

When the fluid velocity is zero, called the hydrostatic condition, the pressure variation is due only to the weight of the fluid. Fluid Statics When the fluid velocity is zero, called the hydrostatic condition, the pressure variation is due only to the weight of the fluid. Consider a small wedge of fluid at rest of size Δx, Δz, Δs

More information

An Overview of the Finite Element Analysis

An Overview of the Finite Element Analysis CHAPTER 1 An Overview of the Finite Element Analysis 1.1 Introduction Finite element analysis (FEA) involves solution of engineering problems using computers. Engineering structures that have complex geometry

More information

Lecture L2 - Degrees of Freedom and Constraints, Rectilinear Motion

Lecture L2 - Degrees of Freedom and Constraints, Rectilinear Motion S. Widnall 6.07 Dynamics Fall 009 Version.0 Lecture L - Degrees of Freedom and Constraints, Rectilinear Motion Degrees of Freedom Degrees of freedom refers to the number of independent spatial coordinates

More information

Stack Contents. Pressure Vessels: 1. A Vertical Cut Plane. Pressure Filled Cylinder

Stack Contents. Pressure Vessels: 1. A Vertical Cut Plane. Pressure Filled Cylinder Pressure Vessels: 1 Stack Contents Longitudinal Stress in Cylinders Hoop Stress in Cylinders Hoop Stress in Spheres Vanishingly Small Element Radial Stress End Conditions 1 2 Pressure Filled Cylinder A

More information

1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST 2003. Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors

1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST 2003. Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors 1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST 2003 Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors Carolyn E. Lehner, Student Member, IEEE, Zhong He, Senior

More information

Sample preparation for X-ray fluorescence analysis

Sample preparation for X-ray fluorescence analysis Technical articles Sample preparation for X-ray fluorescence analysis III. Pressed and loose powder methods Gakuto Takahashi* 1. Introduction There are two main sample preparation techniques for measurement

More information

5-Axis Test-Piece Influence of Machining Position

5-Axis Test-Piece Influence of Machining Position 5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,

More information

Figure 2.31. CPT Equipment

Figure 2.31. CPT Equipment Soil tests (1) In-situ test In order to sound the strength of the soils in Las Colinas Mountain, portable cone penetration tests (Japan Geotechnical Society, 1995) were performed at three points C1-C3

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

2 Absorbing Solar Energy

2 Absorbing Solar Energy 2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could

More information

Measuring Size and Distance in Digital Mammography

Measuring Size and Distance in Digital Mammography On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval

More information

Area & Volume. 1. Surface Area to Volume Ratio

Area & Volume. 1. Surface Area to Volume Ratio 1 1. Surface Area to Volume Ratio Area & Volume For most cells, passage of all materials gases, food molecules, water, waste products, etc. in and out of the cell must occur through the plasma membrane.

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

Instructors Guide: Atoms and Their Isotopes

Instructors Guide: Atoms and Their Isotopes Instructors Guide: Atoms and Their Isotopes Standards Connections Connections to NSTA Standards for Science Teacher Preparation C.3.a.1 Fundamental structures of atoms and molecules. C.3.b.27 Applications

More information

Experiment 1: Measurement and Density

Experiment 1: Measurement and Density Experiment 1: Measurement and Density Chemistry 140 Learning Objectives Become familiar with laboratory equipment and glassware Begin to see the link between measurement and chemical knowledge Begin to

More information

The volume of a penny will be calculated from its mass and density.

The volume of a penny will be calculated from its mass and density. Measurement and Density In science a key concern is the quantities involved in chemical processes. These amounts can be directly measured or calculated from other measurements. A measurement consists of

More information

Basic Nuclear Concepts

Basic Nuclear Concepts Section 7: In this section, we present a basic description of atomic nuclei, the stored energy contained within them, their occurrence and stability Basic Nuclear Concepts EARLY DISCOVERIES [see also Section

More information

Big Ideas in Mathematics

Big Ideas in Mathematics Big Ideas in Mathematics which are important to all mathematics learning. (Adapted from the NCTM Curriculum Focal Points, 2006) The Mathematics Big Ideas are organized using the PA Mathematics Standards

More information

Main properties of atoms and nucleus

Main properties of atoms and nucleus Main properties of atoms and nucleus. Atom Structure.... Structure of Nuclei... 3. Definition of Isotopes... 4. Energy Characteristics of Nuclei... 5. Laws of Radioactive Nuclei Transformation... 3. Atom

More information

Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document 74-461

Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document 74-461 Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/0/202 Adapted from National Bureau of Standards Document 74-46 The purpose of calibrating glassware is to determine the volume of

More information

The Basic Principles of Sieve Analysis

The Basic Principles of Sieve Analysis The Basic Principles of Sieve Analysis Introduction Many natural and manufactured materials occur in a disperse form, which means that they consist of differently shaped and sized particles. The particle

More information

Module 7 (Lecture 24 to 28) RETAINING WALLS

Module 7 (Lecture 24 to 28) RETAINING WALLS Module 7 (Lecture 24 to 28) RETAINING WALLS Topics 24.1 INTRODUCTION 24.2 GRAVITY AND CANTILEVER WALLS 24.3 PROPORTIONING RETAINING WALLS 24.4 APPLICATION OF LATERAL EARTH PRESSURE THEORIES TO DESIGN 24.5

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

CHAPTER: 6 FLOW OF WATER THROUGH SOILS

CHAPTER: 6 FLOW OF WATER THROUGH SOILS CHAPTER: 6 FLOW OF WATER THROUGH SOILS CONTENTS: Introduction, hydraulic head and water flow, Darcy s equation, laboratory determination of coefficient of permeability, field determination of coefficient

More information

Acceleration levels of dropped objects

Acceleration levels of dropped objects Acceleration levels of dropped objects cmyk Acceleration levels of dropped objects Introduction his paper is intended to provide an overview of drop shock testing, which is defined as the acceleration

More information

CONSTANT HEAD AND FALLING HEAD PERMEABILITY TEST

CONSTANT HEAD AND FALLING HEAD PERMEABILITY TEST CONSTANT HEAD AND FALLING HEAD PERMEABILITY TEST 1 Permeability is a measure of the ease in which water can flow through a soil volume. It is one of the most important geotechnical parameters. However,

More information

Measurement and Calibration

Measurement and Calibration Adapted from: H. A. Neidig and J. N. Spencer Modular Laboratory Program in Chemistry Thompson Learning;, University of Pittsburgh Chemistry 0110 Laboratory Manual, 1998. Purpose To gain an understanding

More information