Triple Stage Raman spectrograph/spectrometer Raman system with scanning microscopy attachment: QTY: One

Size: px
Start display at page:

Download "Triple Stage Raman spectrograph/spectrometer Raman system with scanning microscopy attachment: QTY: One"

Transcription

1 Specifications: Triple Stage Raman spectrograph/spectrometer Raman system with scanning microscopy attachment: QTY: One A. Triple Stage Raman spectrograph/spectrometer: 1. Spectral range : UV_Vis_NIR : nm optimized for UV-visible 2. Various Modes of operation : i) double additive + third : Triple additive ii) double subtractive + third : Spectrograph iii) single Stage : Spectrograph Easy software operated switching among all modes of operations. 3. Spectral resolution: i) Bandwidth per pixel (26 µm) at 500 nm with 1800 l/mm Triple (subtractive) and Single stage spectrograph : 0.75 cm -1 CCD coverage : ~ 1024 X bandwidth per pixel ii) Triple stage spectrometer (additive) : 0.2 cm -1 at 10 µm slit width at 500 nm with 1800 l/mm. 4. Reproducibility : ± one pixel or better 5. Low frequency : Triple stage spectrograph : up to 5 cm -1 With Triple stage additive : 5 cm Focal length of first two monochromators : 500 mm Third monochromator (spectrograph stage) : 600 mm 7. Gratings : holographic gratings 2400 l/mm : Quantity : three 1800 l/mm : Quantity : three 600 l/mm : Quantity : one 900 l/mm : Quantity : two With grating case for protection or mounted on grating turret/s. *The grating details to be mounted on turrets will be given at the time of P.O. Reflectivity curves for gratings and mirrors should be provided 8. Slits: Entrance, exit and intermediate slits. Entrance and exit slits with height limiter : 0-2 mm or better with resolution of 5 µm

2 Motorized intermediate slits: 0-12 mm or better with resolution 15 µm 9. Sample chamber : Macro and Micro with option to view the entrance slit image i) Macro sample chamber : Scattering geometry for Macro: 90 0 and (backscattering) Appropriate Focusing / collection optics with F/ 7.5 or better coupling with the monochromator Entrance slit viewing arrangement for focusing adjustment ii) Micro sample chamber: Microscope detail specifications are given in section B 10. Optics for single stage spectrograph operation: Interference Filters with 1 nm FWHM, 1 diameter for 363.8, 488 and nm Edge Filter with Holder for 488 and nm with transition Width 110 cm -1, O. D. > 6 UV Edge Filter with Holder for nm with transition Width 305 cm -1, O. D. > Optics for Polarization measurements: Half wave plate/s for UV, visible range. Scrambler/s for UV, visible range Analyzer/s for UV, visible range with extinction ratio of ³ 10 4

3 12. Detectors : 1. Single channel : Peltier cooled PMT : Qty : one each: PMT 1 and PMT 2 PMT 1 specifications: Head ON Dark current: 30 cps at C Gain: ³ 10 5 Wavelength range : nm Dynamic range : typical ³ cps QE: typical value ³ 15% at 250nm and ³ 20% at 500nm PMT 2 specifications: Head ON Dark current: 1 cps at C or 100 cps at Room temperature (25 0 C). Gain: 10 5 Wavelength range: nm Dynamic range: typical cps QE: typical value 15% at 500 nm Note: The PMTs should be inclusive of common peltier cooled housing and required electronics for cooling and photon counting data acquisition. 2. Multichannel : Peltier cooled CCD (inclusive of required housing and electronics for cooling) Open Electrode front illuminated. Pixel size : 26 µm x 26 µm or less* * Resolution mentioned in point 3 should be scaled accordingly.

4 Dimension : 1024 x 256 pixels with 26 µm pixel or equivalent for smaller pixel size. Spectral range : nm cooling temperature : C System Read-out noise : typical 6 e - rms or 100-kHz digitization / 4 e - 20-kHz digitization. Dark noise : minimum cooling temperature Quantum Efficiency at 25 0 C : 40% over nm and 20% over nm and nm Dynamic range : 16 bit v Certification for QE curve for PMT and CCD should be provided. v Two exit ports on third monochromator with appropriate opto-mechanical arrangement to switch from one to other port (CCD and PMT detector) should be provided. v The Raman system should include motorized ND filter wheel/assembly to allow variation of the intensity of the excitation wavelengths by at least 2 orders of magnitude for the entire range of operation with one beam stopper.

5 B. Microscope attachment 1. Working distance range up to 130 mm 2. Motorized X-Y stage: XY stage for Raman mapping/imaging : 75 mm x 50 mm or larger With step resolution: 0.1 µm 3. UV-visible confocal microscopy : Z stage motion: step size: 0.1 µm Aperture (variable): µm Revolving nosepiece : 6 objectives 4. Objectives: UV-visible range: Plano Achromatic Wavelength range: nm SWD (short working distance) o 20 X, NA 0.40 or better, WD 1.50 mm o 50X, NA 0.75 or better, WD 0.30 mm o 100 X, NA 0.90 or better, WD 0.20 mm LWD: (long working distance) o 20 X, NA 0.4 or better, WD mm o 50 X, NA 0.5 or better, WD mm Wavelength range: nm SWD: 40 X, NA 0.50 or better, WD 1.0 mm LWD: 15 X, NA 0.30 or better, WD 8.5 mm

6 5. Arrangement for One for viewing the sample with CCD camera** Second for coupling to the Raman spectrograph/meter ** CCD camera with frame grabber for viewing and taking optical image of the sample should be provided. 6. White light Illumination with halogen lamp (12V,100W) C. Software, hardware and power supply: 1. Complete Windows based software control for data acquisition in all modes. All the necessary manuals (both soft and hard copy) of the software should be supplied. 2. All the power supplies, interfaces, data acquisition electronics and control units required for Raman measurements in all the modes specified should be inclusive in the offer. 3. Raman spectroscopy software should include multiple spectral window joining with CCD background subtraction to increase the spectral window of measurement in multichannel detection mode with high spectral resolution. 4. PC and Monitor : Latest Windows operating system based-software pre-installed and ready to run, LCD display 23 or larger with up to date PC 5. To be operated at : 230 V ± 10% at 50 Hz ( Power supply ) D. Other facilities: 1. Hg / Ne lamp with power supply for calibration of a spectrometer should be provided. 2. Sokes /Anti Stokes accessory to allow low frequency measurements of both Stokes and Anti-Stokes data. E. Optional: 1. Spares: please also quote for important spares required for smooth operation of the system for 5 years. 2. Motor driven Rotational mounting stage for rotating the sample by 360 0

7 3. Thermometric cell attachment for microscope: lowest temperature : 80 K Highest temperature : 800 K Temperature stability : ± 1 0 K or better Complete with electronic interface, controls and mechanical components required for programmed operation using data acquisition software. 4. Half wave plates one each for 488 nm and nm wavelength. 5. Semi-apochromatic objectives for visible range: 10 X, NA 0.25 or better, WD 15 mm 100 X, NA 0.7 or better, WD 3.0 mm 6. Filters: i) Interference Filters with 1 nm FWHM, 1 diameter for and nm ii) Stokes and AntiStokes Edge Filters with Holder for and nm with transition width 250 cm -1 with O. D. > 6 iii) AntiStokes Edge Filter with Holder for 488 and 514 nm with transition width 100 cm -1 with O.D. > 6 7. Power meter for microscope, Range : ~ 100 µw mw, spectral range nm F. Warranty : All components of the system should have at least one year warranty from the date of installation at RRCAT site. G. Maintenance contract : Please also quote separately for our consideration an yearly maintenance contract (AMC) renewable up to 3 years after the warranty period, clearly indicating scope of AMC. H. Installation and training: The system should be satisfactorily installed at the RRCAT site. Pre-requirements for the installation should be communicated at the time of confirmation of P.O. The training for operation of the Raman system at the RRCAT site for two days should be provided after completion of the installation of the system. I. Acceptance Criteria: The item shall be accepted on the basis of successful operation and demonstration of all the quoted parameters at RRCAT site, by the service Engineer of the supplier or it s representative.

8 Before dispatch following spectra recorded on Instrument (Raman system) should be provided i) Hg/Ne lines to show claimed resolution/s in various modes of operation. ii) Observation of 9.5 cm -1 peak of L-Cystine. iii) Low frequency data showing confined acoustic phonon/s in the range cm -1 of commercially available Schott filter OG515/OG530 All the above measurements to be reproduced at the RRCAT site at the time of installation. The Raman mapping may be separately demonstrated on a suitable sample. Notes: 1. Operating manuals containing technical details and some details of the instruments should be provided. 2. The supplier shall confirm support for maintenance and after sales service. i.e. The manufacturers should have authorized sales / service representatives in India, if the principal is not located in India. 3. The suppliers should include, A) A list of customers to whom similar systems are supplied in India with their complete contact details. B) Compliance confirmation separately for each specification/parameters mentioned in the indent. 4. Quotations from the original manufacturers and their authorized representatives will only be considered. The authorization certificate should be provided along with the quotation and the original should be reproduced as and when required for verification.

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Guaranteed specifications Design overview Double beam, ratio recording, double out-of-plane Littrow monochromator UV-Vis-NIR spectrophotometer (Agilent Cary

More information

Advanced Research Raman System Raman Spectroscopy Systems

Advanced Research Raman System Raman Spectroscopy Systems T600 Advanced Research Raman System Raman Spectroscopy Systems T600 Advanced Research Raman System T600 Triple stage Raman Spectrometer: The only solution for unprecedented stability and performance! Robust

More information

Raman Spectroscopy Basics

Raman Spectroscopy Basics Raman Spectroscopy Basics Introduction Raman spectroscopy is a spectroscopic technique based on inelastic scattering of monochromatic light, usually from a laser source. Inelastic scattering means that

More information

Raman spectroscopy Lecture

Raman spectroscopy Lecture Raman spectroscopy Lecture Licentiate course in measurement science and technology Spring 2008 10.04.2008 Antti Kivioja Contents - Introduction - What is Raman spectroscopy? - The theory of Raman spectroscopy

More information

S2000 Spectrometer Data Sheet

S2000 Spectrometer Data Sheet Description The Ocean Optics OEM S2000 Spectrometer includes the linear CCD-array optical bench, plus the circuits necessary for spectrometer operation. The result is a compact, flexible system with no

More information

APPLICATION NOTE. Basler racer Migration Guide. Mechanics. www.baslerweb.com. Flexible Mount Concept. Housing

APPLICATION NOTE. Basler racer Migration Guide. Mechanics. www.baslerweb.com. Flexible Mount Concept. Housing 62 62 APPLICATION NOTE www.baslerweb.com Basler racer Migration Guide This paper describes what to consider when replacing the Basler L100 Camera Link or the Basler runner Gigabit Ethernet (GigE) line

More information

instruments Analytical Instruments for Science

instruments Analytical Instruments for Science instruments Analytical Instruments for Science instruments Contents PAGE NO. Introduction 4 T60 UV-Vis Spectrophotometer 6 T70 UV-Vis Spectrophotometer 10 T80 UV-Vis Spectrophotometer 14 T90+ UV-Vis Spectrophotometer

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

VWR SPECTROPHOTOMETERS

VWR SPECTROPHOTOMETERS VWR SPECTROPHOTOMETERS Reliable Accurate Easy to Use New_Spectrophotometry_v2.indd 2 Your first choice for spectrophotometry 5/9/2014 2:22:18 PM Quality Guaranteed VWR SPECTROPHOTOMETERS Single Beam Units

More information

Agilent Cary 60 UV-Vis

Agilent Cary 60 UV-Vis Agilent Cary 60 UV-Vis Efficient. Accurate. Flexible. Specifications Introduction The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is designed to meet both current

More information

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory FIELD APPLICATION REPORT UV/Vis/NIR Spectroscopy Author: Frank Padera Shelton, CT Contributor: Chris Lynch Shelton, CT Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR

More information

Fig.1. The DAWN spacecraft

Fig.1. The DAWN spacecraft Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics

More information

Development of the Extreme Ultraviolet Spectrometer: EXCEED

Development of the Extreme Ultraviolet Spectrometer: EXCEED Development of the Extreme Ultraviolet Spectrometer: EXCEED Go Murakami*, Kazuo Yoshioka, Atsushi Yamazaki, Tomoki Kimura Institute of Space and Astronautical Science, Japan Aerospace Exploration Agency

More information

Spectroscopy Using the Tracker Video Analysis Program

Spectroscopy Using the Tracker Video Analysis Program Spectroscopy Using the Tracker Video Analysis Program Douglas Brown Cabrillo College Aptos CA 95003 dobrown@cabrillo.edu Spectroscopy has important applications in many fields and deserves more attention

More information

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION G. Thuillier1, D. Bolsee2 1 LATMOS-CNRS, France 2 Institut d Aéronomie Spatiale

More information

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this

More information

Radiant Dyes Laser Accessories GmbH

Radiant Dyes Laser Accessories GmbH New NarrowScan New Resonator Design Improved Sine Drive Unit Autotracking Frequency doubling, tripling and mixing Wavelength Separation Unit Frequency Stabilization Temperature Stabilization Wavelength

More information

Introduction to reflective aberration corrected holographic diffraction gratings

Introduction to reflective aberration corrected holographic diffraction gratings Introduction to reflective aberration corrected holographic diffraction gratings By Steve Slutter, Wu Jiang, and Olivier Nicolle The reflective diffraction grating is the heart of most spectroscopy systems

More information

Fully Automated Imaging Spectrometer User Manual

Fully Automated Imaging Spectrometer User Manual Fully Automated Imaging Spectrometer User Manual Part Number 81092 Revision 2 Copyright February, 06 HORIBA Jobin Yvon Inc., Optical Spectroscopy Division. All rights reserved. Portions of the software

More information

Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red

Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red Changing the economics of space Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red J. Fernandez-Saldivar 1, F. Culfaz 1,

More information

Realization of a UV fisheye hyperspectral camera

Realization of a UV fisheye hyperspectral camera Realization of a UV fisheye hyperspectral camera Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM Outline Purpose of the instrument Required specs Hyperspectral technique Optical

More information

Spectral Measurement Solutions for Industry and Research

Spectral Measurement Solutions for Industry and Research Spectral Measurement Solutions for Industry and Research Hamamatsu Photonics offers a comprehensive range of products for spectroscopic applications, covering the, Visible and Infrared regions for Industrial,

More information

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters

More information

UV-NIR LASER BEAM PROFILER

UV-NIR LASER BEAM PROFILER CinCam CMOS - Technical Data - CMOS-1201 CMOS-1202 CMOS-1203 CMOS-1204 Standard Series Standard Series Standard Series Standard Series SENSOR DATA Format: 1/2 1/1.8 1/1.8 1/2.5 Active area (without cover

More information

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

WAVELENGTH OF LIGHT - DIFFRACTION GRATING PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant

More information

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy Microscopy from Carl Zeiss AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy New Dimensions in Performance AxioCam MR from Carl Zeiss Both

More information

Copyright 1999 2010 by Mark Brandt, Ph.D. 12

Copyright 1999 2010 by Mark Brandt, Ph.D. 12 Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits

More information

MCAL Spectrophotometry. Spectrophotometry

MCAL Spectrophotometry. Spectrophotometry MCAL Spectrophotometry Instruments include: Cary 50 UV-vis Spectrophotometer Eclipse Spectrofluorometer HPLC Diode Array and Fluorescence ICP-OES with CCD detection Spectrophotometry The instruments all

More information

LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes

LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes LBS-300 Beam Sampler for C-mount Cameras P/N SP90183, SP90184, SP90185 and SP90186 YAG Focal Spot Analysis Adapter P/N SP90187, SP90188, SP90189, SP90190, SP90191 User Notes Ophir-Spiricon Inc. 60 West

More information

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough

More information

DIODE PUMPED CRYSTALASER

DIODE PUMPED CRYSTALASER DIODE PUMPED CRYSTALASER Ultra-compact CW & Pulsed Turnkey Systems UV Visible to IR High Reliability High Stability High Efficiency TEMoo & SLM Low Noise Low Cost ULTRA-COMPACT DIODE-PUMPED CRYSTAL LASER

More information

UV-VISIBLE SPECTROPHOTOMETER

UV-VISIBLE SPECTROPHOTOMETER instruments UV-VISIBLE SPECTROPHOTOMETER PRODUCT BROCHURE Analytical Instruments for Science Contents PAGE NO. Introduction 4 T60 UV-Vis Spectrophotometer 6 T70 UV-Vis Spectrophotometer 10 T80 UV-Vis Spectrophotometer

More information

AxioCam HR The Camera that Challenges your Microscope

AxioCam HR The Camera that Challenges your Microscope Microscopy from Carl Zeiss AxioCam HR The Camera that Challenges your Microscope Documentation at the edge of the visible The Camera for Maximum Success: AxioCam HR Low light fluorescence, live cell imaging,

More information

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical

More information

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Some limitations (disadvantages) of NSOM A trade-off

More information

Quasi-Continuous Wave (CW) UV Laser Xcyte Series

Quasi-Continuous Wave (CW) UV Laser Xcyte Series COMMERCIAL LASERS Quasi-Continuous Wave (CW) UV Laser Xcyte Series Key Features 355 nm outputs available Quasi-CW UV output Field-proven Direct-Coupled Pump (DCP ) TEM00 mode quality Light-regulated output

More information

Raman Spectroscopy. 1. Introduction. 2. More on Raman Scattering. " scattered. " incident

Raman Spectroscopy. 1. Introduction. 2. More on Raman Scattering.  scattered.  incident February 15, 2006 Advanced Physics Laboratory Raman Spectroscopy 1. Introduction When light is scattered from a molecule or crystal, most photons are elastically scattered. The scattered photons have the

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

A down-under undergraduate optics and photonics laboratory

A down-under undergraduate optics and photonics laboratory A down-under undergraduate optics and photonics laboratory Barry Perczuk and Michael Gal School of Physics, The University of New South Wales, Sydney, NSW 2052, Australia ABSTRACT Our senior undergraduate

More information

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale Nano-Spectroscopy Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale Since its introduction in the early 80 s, Scanning Probe Microscopy (SPM) has quickly made nanoscale imaging an affordable

More information

ISO9001 ISO14001. Environmentally-friendly product All optical components in CX41 microscope use lead-free eco-glass.

ISO9001 ISO14001. Environmentally-friendly product All optical components in CX41 microscope use lead-free eco-glass. Specifications Item Optical system Illumination CX41 UIS2 (Universal Infinity-corrected) optical system Built-in transmitted Koehler illuminator 6V30W halogen bulb 100-120V/220-240Vg 0.85/0.45A 50/60Hz

More information

MRC High Resolution. MR-compatible digital HD video camera. User manual

MRC High Resolution. MR-compatible digital HD video camera. User manual MRC High Resolution MR-compatible digital HD video camera User manual page 1 of 12 Contents 1. Intended use...2 2. System components...3 3. Video camera and lens...4 4. Interface...4 5. Installation...5

More information

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements. T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting

More information

ING LA PALMA TECHNICAL NOTE No. 130. Investigation of Low Fringing Detectors on the ISIS Spectrograph.

ING LA PALMA TECHNICAL NOTE No. 130. Investigation of Low Fringing Detectors on the ISIS Spectrograph. ING LA PALMA TECHNICAL NOTE No. 130 Investigation of Low Fringing Detectors on the ISIS Spectrograph. Simon Tulloch (ING) June 2005 Investigation of Low Fringing Detectors on the ISIS Spectrograph. 1.

More information

hsfc pro 12 bit ultra speed intensified imaging

hsfc pro 12 bit ultra speed intensified imaging hsfc pro 12 bit ultra speed intensified imaging four MCP-image intensifier camera modules ultra fast shutter down to 3 ns in single mode excellent sensitivity of the system allows single photon detection

More information

SFG spectrometers. Sum Frequency Generation Vibrational Spectroscopy. More than 20 years experience

SFG spectrometers. Sum Frequency Generation Vibrational Spectroscopy. More than 20 years experience Sum Frequency Generation Vibrational Spectroscopy More than 20 years experience Sum Frequency Generation Vibrational Spectroscopy Sum Frequency Generation Vibrational Spectroscopy (SFG-VS) is powerful

More information

Meteor Spectroscopy First Steps

Meteor Spectroscopy First Steps Meteor Spectroscopy First Steps Martin Dubs AGG, FMA Content Introduction Hardware Software Spectrum extraction Summary Martin Dubs, Meteor Spectroscopy, Nov. 2014 page 2 Meteor spectroscopy Additional

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn

More information

SGS: Das Scanning Grating Spektrometer Ein kleines, günstiges Spektrometermodul auf Basis eines dispersiven Mikrosystems

SGS: Das Scanning Grating Spektrometer Ein kleines, günstiges Spektrometermodul auf Basis eines dispersiven Mikrosystems the leading supplier of Micro Scanning Devices Your Micro Optical Solution IPHT-Workshop 11.-12.03.08 We move the light for you! SGS: Das Scanning Grating Spektrometer Ein kleines, günstiges Spektrometermodul

More information

Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP

Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP Now Powered by LightField FEATURES BENEFITS Back Illuminated CCD (248B) For highest sensitivity Front illuminated CCD (248F) Affordable technology for moderate light level applications Ultra low noise

More information

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,

More information

It has long been a goal to achieve higher spatial resolution in optical imaging and

It has long been a goal to achieve higher spatial resolution in optical imaging and Nano-optical Imaging using Scattering Scanning Near-field Optical Microscopy Fehmi Yasin, Advisor: Dr. Markus Raschke, Post-doc: Dr. Gregory Andreev, Graduate Student: Benjamin Pollard Department of Physics,

More information

Fraunhofer Diffraction

Fraunhofer Diffraction Physics 334 Spring 1 Purpose Fraunhofer Diffraction The experiment will test the theory of Fraunhofer diffraction at a single slit by comparing a careful measurement of the angular dependence of intensity

More information

Recording the Instrument Response Function of a Multiphoton FLIM System

Recording the Instrument Response Function of a Multiphoton FLIM System Recording the Instrument Response Function of a Multiphoton FLIM System Abstract. FLIM data analysis in presence of SHG signals or extremely fast decay components requires the correct instrument response

More information

The Measurement of Sensitivity in Fluorescence Spectroscopy

The Measurement of Sensitivity in Fluorescence Spectroscopy The Measurement of Sensitivity in Fluorescence Spectroscopy Among instrumental techniques, fluorescence spectroscopy is recognized as one of the more sensitive. In fluorescence, the intensity of the emission

More information

Limiting factors in fiber optic transmissions

Limiting factors in fiber optic transmissions Limiting factors in fiber optic transmissions Sergiusz Patela, Dr Sc Room I/48, Th. 13:00-16:20, Fri. 9:20-10:50 sergiusz.patela@pwr.wroc.pl eportal.pwr.wroc.pl Copying and processing permitted for noncommercial

More information

Nederland België / Belgique

Nederland België / Belgique Cronus Brochure 2 Introduction Cronus: The first spectrometer and colorimeter in one The Cronus is world s first spectrocolorimeter combining a high en VIS spectrometer with a high speed XYZ colorimeter.

More information

scmos discover new ways of seeing andor.com Features and Benefits Scientific CMOS - Fast, sensitive, compact and light. Specifications Summary

scmos discover new ways of seeing andor.com Features and Benefits Scientific CMOS - Fast, sensitive, compact and light. Specifications Summary Ultra Sensitive Imaging Low Light Imaging Features and Benefits Compact and light Ideal for integration into space restrictive set-ups 1.2 e - read noise Lower detection limit than any CCD 5.5 megapixel

More information

Alpy guiding User Guide. Olivier Thizy (olivier.thizy@shelyak.com) François Cochard (francois.cochard@shelyak.com)

Alpy guiding User Guide. Olivier Thizy (olivier.thizy@shelyak.com) François Cochard (francois.cochard@shelyak.com) Alpy guiding User Guide Olivier Thizy (olivier.thizy@shelyak.com) François Cochard (francois.cochard@shelyak.com) DC0017A : april 2013 Alpy guiding module User Guide Olivier Thizy (olivier.thizy@shelyak.com)

More information

A Guide to Acousto-Optic Modulators

A Guide to Acousto-Optic Modulators A Guide to Acousto-Optic Modulators D. J. McCarron December 7, 2007 1 Introduction Acousto-optic modulators (AOMs) are useful devices which allow the frequency, intensity and direction of a laser beam

More information

Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.

Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away. Revision problem Chapter 18 problem 37 page 612 Suppose you point a pinhole camera at a 15m tall tree that is 75m away. 1 Optical Instruments Thin lens equation Refractive power Cameras The human eye Combining

More information

Time out states and transitions

Time out states and transitions Time out states and transitions Spectroscopy transitions between energy states of a molecule excited by absorption or emission of a photon hn = DE = E i - E f Energy levels due to interactions between

More information

UV-Vis spectrophotometers

UV-Vis spectrophotometers UV-Vis spectrophotometers an Advanced Digital Technology EXPERIENCE SECOMAM is proud to provide the new fourthgeneration UVIKON ultraviolet-visible spectrophotometers.with more than five decades of experience

More information

How To Understand Light And Color

How To Understand Light And Color PRACTICE EXAM IV P202 SPRING 2004 1. In two separate double slit experiments, an interference pattern is observed on a screen. In the first experiment, violet light (λ = 754 nm) is used and a second-order

More information

73 Series Spectrophotometers

73 Series Spectrophotometers 73 Series Spectrophotometers The 73 series introduces four new spectrophotometers to the extensive Jenway range. These spectrophotometers use icon driven software and have an improved navigation system

More information

Color holographic 3D display unit with aperture field division

Color holographic 3D display unit with aperture field division Color holographic 3D display unit with aperture field division Weronika Zaperty, Tomasz Kozacki, Malgorzata Kujawinska, Grzegorz Finke Photonics Engineering Division, Faculty of Mechatronics Warsaw University

More information

Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012

Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012 Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012 Figure 1.1 illustrates the parts of an upright compound microscope and indicates the terminology that I use in these notes. Figure 1.1. Parts of a Compound

More information

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES ICP OPTICAL ATOMIC EMISSION SPECTROSCOPY Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES HORIBA Scientific Longjumeau, France Keywords: metallurgy, non-ferrous, halogens 1 Introduction

More information

Application Report: Running µshape TM on a VF-20 Interferometer

Application Report: Running µshape TM on a VF-20 Interferometer : Running µshape TM on a VF-20 Interferometer General This report describes how a fiber interferometer from Arden Photonics Ltd was used together with the µshape TM Generic software package. The VF-20

More information

Advances in scmos Camera Technology Benefit Bio Research

Advances in scmos Camera Technology Benefit Bio Research Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical

More information

Spherical VLS grating RIXS spectrometers: From basics to the hv 2 concept

Spherical VLS grating RIXS spectrometers: From basics to the hv 2 concept Spherical VLS grating RIXS spectrometers: From basics to the hv 2 concept Vladimir N. Strocov (ADRESS beamline, Swiss Light Source) Outline: 1. Optical design of spherical VLS grating spectrometers: -

More information

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information 1 Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information Instructions for the Operation of the Cary 300 Bio UV-Visible Spectrophotometer See the Thermo OMNIC Help reference on page 49. Ultraviolet-Visible

More information

LabRAM HR. Research Raman Made Easy! Raman Spectroscopy Systems. Spectroscopy Suite. Powered by:

LabRAM HR. Research Raman Made Easy! Raman Spectroscopy Systems. Spectroscopy Suite. Powered by: LabRAM HR Research Raman Made Easy! Raman Spectroscopy Systems Powered by: Spectroscopy Suite Cutting-Edge Applications with the LabRAM HR Deeply involved in Raman spectroscopy for decades, HORIBA Scientific

More information

Rodenstock Photo Optics

Rodenstock Photo Optics Rogonar Rogonar-S Rodagon Apo-Rodagon N Rodagon-WA Apo-Rodagon-D Accessories: Modular-Focus Lenses for Enlarging, CCD Photos and Video To reproduce analog photographs as pictures on paper requires two

More information

HP 70950B OPTICAL SPECTRUM ANALYZER

HP 70950B OPTICAL SPECTRUM ANALYZER HP 71450B, 71451B, and 71452B Optical Spectrum Analyzers Technical Specifications Spectral Measurements from 600 to 1700 nm HP 70950B OPTICAL SPECTRUM ANALYZER OPTICAL INPUT The HP 71450B, 71451B, and

More information

Simplicity. Reliability. Performance

Simplicity. Reliability. Performance Simplicity Reliability Performance ICP Without Compromise At Teledyne Leeman Labs atomic spectroscopy is our business, our only business. We are deeply committed to providing you with technically superior

More information

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Application Note: 52020 Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Timothy O. Deschaines, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words Array Automation

More information

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution Product Information Version 1.0 ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution ZEISS Axiocam 506 color Sensor Model Sensor Pixel

More information

GRID AND PRISM SPECTROMETERS

GRID AND PRISM SPECTROMETERS FYSA230/2 GRID AND PRISM SPECTROMETERS 1. Introduction Electromagnetic radiation (e.g. visible light) experiences reflection, refraction, interference and diffraction phenomena when entering and passing

More information

SENSORS. Miniature Sensors - S3Z. Advanced line of miniature Asian style of photoelectric sensors. 40-300 mm background suppression

SENSORS. Miniature Sensors - S3Z. Advanced line of miniature Asian style of photoelectric sensors. 40-300 mm background suppression Advanced line of miniature Asian style of photoelectric sensors 4-3 background suppression.7 m proximity, 15 with narrow beam 4 m polarized retroreflective 3 m LASER through beam Standard 3-wire output

More information

WHITE PAPER. Are More Pixels Better? www.basler-ipcam.com. Resolution Does it Really Matter?

WHITE PAPER. Are More Pixels Better? www.basler-ipcam.com. Resolution Does it Really Matter? WHITE PAPER www.basler-ipcam.com Are More Pixels Better? The most frequently asked question when buying a new digital security camera is, What resolution does the camera provide? The resolution is indeed

More information

Connect to a remote PC via Ethernet with a monitor, keyboard, and mouse. This network monitor offers both long distance and wireless transmission.

Connect to a remote PC via Ethernet with a monitor, keyboard, and mouse. This network monitor offers both long distance and wireless transmission. 17" (43 cm) LCD Monitor Connect to a remote PC via Ethernet with a monitor, keyboard, and mouse. This network monitor offers both long distance and wireless transmission. Long Distance Transmission Easily

More information

Optical laser beam scanner lens relay system

Optical laser beam scanner lens relay system 1. Introduction Optical laser beam scanner lens relay system Laser beam scanning is used most often by far in confocal microscopes. There are many ways by which a laser beam can be scanned across the back

More information

ID Objective Requirements Description of Test Date & Examiner 15 Verify that the optics?? OMC

ID Objective Requirements Description of Test Date & Examiner 15 Verify that the optics?? OMC NAOMI OMC/NCU Acceptance Tests at the University of Durham ATC Document number AOW/GEN/RAH/15.0/06/00 OMC/NCU acceptance tests DRAFT (Version date: 2 nd June 2000) wht-naomi-44 The ID numbers are those

More information

GIANT FREQUENCY SHIFT OF INTRAMOLECULAR VIBRATION BAND IN THE RAMAN SPECTRA OF WATER ON THE SILVER SURFACE. M.E. Kompan

GIANT FREQUENCY SHIFT OF INTRAMOLECULAR VIBRATION BAND IN THE RAMAN SPECTRA OF WATER ON THE SILVER SURFACE. M.E. Kompan GIANT FREQUENCY SHIFT OF INTRAMOLECULAR VIBRATION BAND IN THE RAMAN SPECTRA OF WATER ON THE SILVER SURFACE M.E. Kompan Ioffe Institute, Saint-Peterburg, Russia kompan@mail.ioffe.ru The giant frequency

More information

MB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC

MB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC Measurement & Analytics Measurement made easy MB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC MB3600 FT-NIR spectrometer for your industry

More information

D/N VANDAL RES. VARI-FOCAL IR DOME CAMERA

D/N VANDAL RES. VARI-FOCAL IR DOME CAMERA D/N VANDAL RES. VARI-FOCAL IR DOME CAMERA PIH-0746XW/0746XS INSTRUCTION MANUAL IMPORTANT SAFEGUARDS CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN CAUTION TO REDUCE THE RISK OF ELECTRIC SHOCK, DO NOT REMOVE

More information

Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998)

Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998) Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998) Geoff Andersen and R. J. Knize Laser and Optics Research Center

More information

Theremino System Theremino Spectrometer Technology

Theremino System Theremino Spectrometer Technology Theremino System Theremino Spectrometer Technology theremino System - Theremino Spectrometer Technology - August 15, 2014 - Page 1 Operation principles By placing a digital camera with a diffraction grating

More information

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps 37 500:1 > 70 % pco. low noise high resolution high speed high dynamic range

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps 37 500:1 > 70 % pco. low noise high resolution high speed high dynamic range edge 4.2 LT scientific CMOS camera high resolution 2048 x 2048 pixel low noise 0.8 electrons USB 3.0 small form factor high dynamic range 37 500:1 high speed 40 fps high quantum efficiency > 70 % edge

More information

Active noise control in practice: transformer station

Active noise control in practice: transformer station Active noise control in practice: transformer station Edwin Buikema 1 ; Fokke D. van der Ploeg 2 ; Jan H. Granneman 3 1, 2, 3 Peutz bv, Netherlands ABSTRACT Based on literature and extensive measurements

More information

Incoherent beam combining using stimulated Brillouin scattering in multimode fibers

Incoherent beam combining using stimulated Brillouin scattering in multimode fibers Incoherent beam combining using stimulated Brillouin scattering in multimode fibers Timothy H. Russell and Won B. Roh Air Force Institute of Technology, Wright-Patterson AFB, Ohio 45433 timothy.russell@afit.edu;

More information

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F) Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F) Please contact Dr. Amanda Young for training requests and assistance: 979-862-6845, amandayoung@tamu.edu Hardware Our spectrophotometer is made up of

More information

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented User's Guide Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented Introduction Congratulations on your purchase of Extech's Laser Photo/Contact Tachometer with Non- Contact IR Thermometer,

More information

COOKBOOK. for. Aristarchos Transient Spectrometer (ATS)

COOKBOOK. for. Aristarchos Transient Spectrometer (ATS) NATIONAL OBSERVATORY OF ATHENS Institute for Astronomy, Astrophysics, Space Applications and Remote Sensing HELMOS OBSERVATORY COOKBOOK for Aristarchos Transient Spectrometer (ATS) P. Boumis, J. Meaburn,

More information

D/N VARI-FOCAL IR CAMERA

D/N VARI-FOCAL IR CAMERA D/N VARI-FOCAL IR CAMERA PIH-0364X/0368X/0384X/0388X W/S INSTRUCTION MANUAL IMPORTANT SAFEGUARDS 1. Read Instructions All the safety and operating instructions should be read before the unit is operated.

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope. MICROSCOPY I. OBJECTIVES To demonstrate skill in the proper utilization of a light microscope. To demonstrate skill in the use of ocular and stage micrometers for measurements of cell size. To recognize

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information