1 Gas Chromatography-Mass Spectrometry

2 Gas Chromatographic-Mass Spectrometry Is Olive Oil the healthiest oil? Introduction Gas chromatographic-mass spectrometry (GC-MS) is a very powerful and ubiquitous analytical technique. It is often the analytical method of choice in toxicology, forensics, food science, and environmental research. In essence, this hybrid instrument replaces the traditional thermal-conductivity detector (TCD) or flame-ionization chromatographic detector (FID) with a very sensitive and information-rich mass spectrometer (MS). Not only can a GC-MS separate the volatile components of complex mixtures, but it can also record a mass spectrum of each component. This hybrid instrument provides two separate dimensions of information about the components in the sample, GC retention times and electron ionization (EI) mass spectra. GC retention time is related to specific chemical properties of the molecules in question (e.g. volatility, polarity, presence of specific functional groups) while molecular weight (derived from the mass spectrum) is indicative of atomic composition. Chromatographic techniques separate mixtures of species based on their interactions with a stationary phase and a mobile phase. In the A315 gas chromatography experiment, volatile and semi-volatile species are vaporized into a flow of helium (mobile phase) and blown through a fused silica open tubular capillary (stationary phase). The capillary is usually derivatized to control its polarity (ranging from hydrophobic dimethylsiloxane to hydrophilic polyethylene glycol or any of a wide array of specialty coatings). The gas phase molecules interact with the surface and exist in dynamic equilibrium between being adsorbed onto the inner wall of the column and dissolved in the carrier gas. Gas chromatographs most often use temperature to control this equilibrium (higher temperatures shift the equilibrium more towards the mobile phase while lower temperatures shift it to the stationary phase). This is analogous to changing the polarity of the solvent mixture in thin-layer chromatography (TLC). Various components of mixtures will have different equilibria with the stationary phase and will consequently move through the column at different rates. The temperature can be left constant for maximum resolution or varied to minimize analysis time. A detector at the end of the column records the amount of time required for a given compound to elute off of the column (cf. RF values in TLC). With the proper conditions and column chemistries, very similar compounds can be separated by GC (e.g. isomers of dichlorobenzene, cis- and trans- fatty acids, racemic mixtures, etc.). The GC-MS places an electron ionization quadrupole mass spectrometer at the end of the column. The mass spectrum provides two valuable pieces of information; the retention time and the EI mass spectrum. Molecules eluting from the column are directed into the source of the mass spectrometer and ionized by a beam of 70 ev electrons from a hot filament (M + e - M + + 2e - ). It is important to point out that EI generates odd-electron radical cations. The resulting positive ions are then analyzed by a quadrupole mass filter. Ideally, the molecular weight of the compound can be directly inferred from its mass spectrum. The mass spectrum indicates the mass to charge ratio of the ions, not the molecular weight of the neutral species (amu/e or Da/z

3 where e is the charge on an electron, Da is Daltons (1 Da = 1 amu) and z is the number of positive charges). However, the vast majority of ions generated by EI are singly charged and the charging is accomplished by removing an electron (an electron weighs amu), thus the masses of the ions in an EI mass spectrum are nearly identical to the neutral species from which they are derived. Because EI is a rather energetic process, most organic compounds fragment in the MS ion source, therefore the EI mass spectrum also indicates the masses of these fragments. While this may seem to be a weakness at first (spectra are more complex and sensitivity for the intact species will suffer), a careful analysis of a fragmentation pattern can be used to determine the connectivity of the atoms in the original molecule. In addition, the GC-MS system in the 315 laboratory contains a computerized library that can match mass spectral fingerprints with known spectra to aid in the identification of unknowns. Figure 1 contains the mass spectrum of 9 - tetrahydrocannabinol (THC, the hallucinogenic agent in marijuana) from the NIST 02 mass spectrum library. The NIST 02 database contains the 70 ev EI mass spectra of over 150,000 compounds OH 29 O (mainlib) Dronabinol Figure 1: 70 ev mass spectrum of delta-9 tetrahydrocannabinol The Quadrupole Mass Filter and Spectral Skew The mass analyzer of the Agilent 5973 inert mass-selective detector is a quadrupole mass filter. As the name implies, this device allows only a narrow range of ions (having a specified mass-to-charge ratio, m/z) to pass through at any given time. Thus, the quadrupole is directly analogous to the grating monochromator used in optical spectroscopy techniques. Consequently, quadrupole mass spectrometers are scanning instruments. To generate a complete mass spectrum, they pass ions of a particular m/z for a brief period of time, record the ion current, move on to the next m/z value, and repeat the entire process over the specified range. Depending on the range of m/z ratios scanned, the electronics driving the instrument, and the quality of data required, a quadrupole mass analyzer may require sec to construct a single mass spectrum (i.e., spectra are created at a rate of Hz). When a quadrupole mass analyzer is used to monitor the output of a chromatographic column (as in GC-MS), a phenomenon referred to as spectral skew results. As compounds elute from the column, they are physically separated into bands which arrive at the detector in

4 sequence. As viewed from a fixed point near the end of the column, the concentration of any given compound will appear to rapidly rise and fall upon the arrival of its band, creating the familiar chromatographic peak. Since a quadrupole mass analyzer requires a finite amount of time to monitor each m/z ratio, the intensities of different ions will be recorded at different points within a single chromatographic peak. Depending on which direction the quadrupole is scanning and where the scan lies within the chromatographic peak, the intensities of either larger or smaller m/z ratios are artificially inflated relative to other ions in the mass spectrum, causing a skewed spectrum to be recorded. The critical question one must ask to assess spectral skew is, How does the scan rate compare with the width of a typical chromatographic peak? If the scan time is not significantly less than the width of a peak, spectral skew will be problematic. One way to reduce the effect of spectral skew is simply to increase the scanning speed of the mass analyzer. For a quadrupole mass filter, this is most often achieved by narrowing the range of m/z ratios scanned. As the time required to obtain a single scan becomes smaller, the width of the chromatographic peak traversed during the scan becomes smaller, and the degree of skew is lessened. If mass spectra can be recorded instantaneously (which can be approached by some time-of-flight [TOF] mass spectrometers), spectral skew is eliminated entirely. SCAN and Select-Ion Monitoring (SIM) The 5973N GC MS quadrupole mass analyzer can be operated in two modes: SCAN and select-ion monitoring (SIM). In the SCAN mode, the quadrupole continuously and repeatedly ramps the monitored m/z ratio from a preset lower limit to a preset upper limit, generating a series of complete mass spectra. At the conclusion of each individual scan, the intensities of all the m/z ratios within the scan are summed, giving a total ion current. A chromatogram is then constructed by plotting the series of total ion current versus retention time. This plot is called the total ion chromatogram, or TIC. Analyzing a GC MS chromatogram obtained in the SCAN mode therefore consists of selecting the portion of the TIC that corresponds to a given peak and extracting the mass spectra from that time period. In the select-ion monitoring (SIM) mode, the quadrupole remains fixed on a small set of m/z ratios, effectively allowing only those predetermined masses to pass through to the detector. An analysis in the SIM mode is useful when one is looking for small quantities of known compounds under circumstances in which they cannot be separated from other compounds chromatographically (i.e., in very complicated mixtures). In addition, the SIM detection scheme often yields substantially lower detection limits than the SCAN mode (more time is spent monitoring the m/z of interest) and also minimizes spectral skew. Automated Data Collection The 6890 GC is fitted with a 100-position autosampler which allows the GC-MS to record data without user intervention. Analytical chemists use autosamplers to increase the number of samples run by an instrument over the course of a day (injections can be made after the technicians have gone home) and to increase productivity (analysts can perform other tasks while the autosampler is working). Furthermore, they free the analyst to focus on data interpretation rather than data collection. Another advantage of robotic sample handling is that it is ensures that all injections are made in a highly reproducible manner. Human users may inject 1.5 ul or 0.95 µl instead of 1.0 µl or push the plunger with a different

5 speed for each injection. These small variations can lower the precision of the data. The robotic injector always injects samples the same way every time. Another useful feature is that autosamplers can perform many different analyses within a sample set. For example, one can direct the instrument to record data using a SCAN mode method followed by a second analysis in SIM mode method. As long as a hardware change (e.g. switching column chemistries) is not required, the autosampler can perform an analysis. You will take full advantage of these capabilities. The instrument will continue to inject samples after the assigned laboratory period is over. The data can be analyzed at any time on any computer that has Agilent s ChemStation 1701DA software. Analysis of Mass-Spectral Data - Important terms and ideas: 1. Molecular ion peak (M+) or parent ion peak: the peak that results from the loss of an electron from the molecule. Since the electron is of negligible mass, the M+ peak gives you the mass of the original molecule. Note that you may not be able to find a molecular ion peak in mass spectra of molecules that fragment easily (e.g., aliphatic hydrocarbons). Furthermore, the M + peak is considered the mass peak containing atoms of the most abundant isotope for each element (i.e., 12 C, 1 H, 16 O, 35 Cl, 79 Br, etc.) and not the less abundant isotopes (i.e., 13 C, 37 Cl, 81 Br, etc.). This is often one of the highest m/z ratio ions in an EI mass spectrum. The official name for this peak is the monoisotopic peak. The second peak in an isotope cluster is called the M+1 peak; the third peak in an isotopic envelope is called the M+2 peak, and so on. Be aware that co-eluting compounds, background contamination in the ion source, and portions of the column stationary phase bleeding into the mass spectrometer can create high m/z ions. 2. Base Peak: the largest peak (highest abundance) in the spectrum. The abundance of each mass peak in the spectrum is representative of the ion s stability in the source, thus the base peak is almost always the most stable ion. Occasionally, the base peak is the same as the molecular ion peak, but this not always the case. Since aliphatic compounds tend to fragment extensively, the base peak is usually a fragment of the original compound. 3. Recall that the atomic weights reported on the periodic table are weighted averages for all isotopes. The mass spectrometer can distinguish between the isotopes. Thus, use the mass of the predominant isotope for each atom. For example, the mass of 12 C (the most abundant isotope of carbon with a nominal mass of 12) is (not ) Da (Daltons, 1 Da = 1 atomic mass unit), the mass of 1 H is and the mass of 16 O is (the resolution of the 5973 inert is such that you can round off to the nearest 0.1 Dalton). These differences may seem slight, but they add up quickly in a large molecule (e.g. for octane, C 8 H 18 has a monoisotopic mass of , not ( = * 18)). The molecular weight calculated using mass values from the periodic

6 table is called the average mass. This is the molecular weight written on the sides of reagent bottles and used for stoichiometric calculations. 4. The intensity of a particular ion is proportional to its concentration in the source at any given time. Thus, a doubling in the intensity of a particular ion implies that the concentration of the molecular species responsible for that ion has also doubled. This is analogous to Beer s Law used in absorption spectroscopy and enables the analyst to use the same quantification methodologies used for GC-FID or GC-TCD experiments for GC-MS. You will calculate response ratios for various fatty acids to allow you to compare their abundances in commercial cooking oils. Steps to identifying a compound based on its mass spectrum: 1) Preliminary survey a. How many carbon atoms are present? b. Are there any N, Cl, Br, or S atoms? c. Is the molecule aromatic or aliphatic? d. What is the molecular weight of the molecule? 2) What functional groups are present? 3) The molecular formula can be determined by piecing the functional groups together. Hydrogen and oxygen can be filled in as needed. 1a. How many carbon atoms are present? Recall that most carbon is in the form of 12 C ( amu), but 1.1% of carbon is 13 C ( amu). Start with the M + peak. This is the monoisotopic mass of the molecule. In carbon containing molecules there will be a less abundant M+1 peak due to the presence of 13 C (~1 Dalton heavier). The rule of thumb that follows does begin to break down for molecules larger than 500 Da (as the chance of a molecule having two or more 13 C s becomes significant), but very few compounds with molecular weights greater than 500 are volatile enough for gas chromatography. Although the plotted mass spectrum is useful for obtaining an overview of the main spectral features, the tabulated list of mass-to-charge ratios (m/z) versus abundance values is essential for an accurate determination of the number of carbon atoms. Below is an example of how the number of carbons in a given molecule can be calculated: Peak: m/z: Abundance: Abundance as a % of M+ M /73.0*100 = 100 M /73.0*100 = 4.5

7 Number of carbon atoms = relative abundance of M+1 peak / 1.1 Using the numbers from above: # of carbons = 4.5/1.1 = 4 carbons. 1b. Are there any N, Cl, Br or S atoms? According to the nitrogen rule, a molecule with an odd M + must contain an odd number of nitrogen atoms and an even M + implies an even number of nitrogen atoms (0 counts as an even number in this case). For Cl, S, and Br, look for at the M+2 peak. These elements, like carbon, have heavy isotopes that can be detected by the mass spectrometer. However the heavy isotopes of these atoms are two Daltons heavier than the most abundant isotope. Lack of an M+2 peak is a clear indication that the molecule does not contain these atoms. If there is a significant M+2 peak present, then you need to look a little closer at the ratio. If (M+/M+2 = 3), then a single chlorine is present; if (M+/M+2 = 19), then one sulfur is present; if (M+/M+2 = 1), then one bromine is present. These ratios are based on the relative abundances (just as in the case of carbon) of each heavy isotope. While these ratios are useful, there is one note of warning that ought to be made about using them to identify which atom is present. If there are multiple Cl, Br or S atoms present in the molecule, then a distribution of peaks in a series (M+, M+2, M+4, M+6, etc.) will result. In these cases, the intensities of these peaks will follow a statistical distribution that is more complicated than the simple ratios above imply. However, in this case there are other ways to determine which atom is present. Look for losses of 35 (for chlorine) or 79 (for bromine) since halogens usually fragment from molecules rather easily. 1c. Is the molecule aromatic or aliphatic? Since aromatic compounds tend to be quite stable, they show less fragmentation under electron impact ionization. Thus, you will see more intense peaks at higher masses with a strong M + peak. If the compound is primarily a substituted benzene ring, then a strong m/z = 77 is certain to show up in the spectrum due to the C 6 H + 5 fragment ion. An aliphatic compound will show extensive fragmentation and exhibit a very weak M + peak. Straight chain aliphatic compounds should show a series of peaks at m/z = n, where n is the number of CH 2 groups in the chain and m/z = 15 corresponds to the terminal methyl group. 1d. What is the molecular weight? By now it should be clear that the molecular weight of the molecule corresponds to the m/z of its M + peak. 2. What functional groups are present?

8 Frequently the functional groups on the molecule can be determined by searching for particular patterns. Here are some examples: a) Alcohols: M = n (formation of R-CH=OH +, where R is composed of n CH 2 groups) b) Aliphatics: M = n c) Benzene: M = 78 (subtract one mass unit for each substitution on the ring) d) Ketones: M = n (formation of R-CH 2 -C O + ) More examples can be found in textbooks (e.g. Interpretation of Mass Spectra, 4 th edition by F. W. McLafferty and F. Tureček). Sometimes identification is not this straightforward. It is not uncommon for a hydrogen atom or a methyl group to migrate to form a more stable ion. Generally, fragmentation will occur where functional groups connect, such as places where branching occurs or where a side chain is attached to an aromatic ring. 3. Piecing it all together At this point you ought to have some idea of what your molecule looked like before fragmentation. You should know the number of carbon atoms and whether or not any N, Cl, Br or S atoms are present. Any mass left over can be accounted for by adding oxygen (M = 16) and hydrogen (M = 1) until you manage to assemble a coherent structure for the molecule. Figure 2 is an example. Figure 2: Mass spectrum of unknown 136 Da compound The first thing to notice is that the base peak is M = 94. The M+ peak is M = 136.

9 Based on the data the molecular weight of the compound is 136. Computing the number of carbons yields: 2.5/25*100 = 10 = M+1 relative to M+ peak Therefore, M+1/1.1 = 10/1.1 = carbons There is no significant M+2 peak, so there likely aren t any chlorine, bromine or sulfur atoms in the molecule. Since M+ is even, there is either an even number of nitrogen atoms or none at all. Notice that there is an intense m/z = 77 peak, which corresponds to a mono-substituted benzene. This leaves three other carbons to figure out. Molecular weight = from the benzene ring -36 (from 3 12 C atoms) 23 mass units left over There must be an even number of nitrogen atoms present, however since two nitrogen atoms (m/z = 14) would give a mass of 28, the possibility of nitrogen can be eliminated. The remaining mass can be accounted for by filling in with oxygen and hydrogen. Obviously the three remaining carbons are not going to carry 23 hydrogen atoms, however if we use one oxygen atom: 23 (mass unaccounted for) 16 (oxygen) = 7 mass units left over. It is reasonable to imagine seven hydrogen atoms on three carbon atoms. With all of the mass accounted for the next step is to propose a likely structure. There are several possible combinations, but only a few actually make sense. Perhaps the molecule is an alcohol. Recall that alcohols show a n mass pattern. There are no m/z = 31, m/z = 45, or m/z -= 59 peaks in the spectrum, so it is probably not an alcohol. How about a ketone? There is an m/z = 43 peak present. However if the molecule were a ketone then it could not accommodate seven hydrogen atoms. What about an ether? That seems to work. What if the structure were: (C 6 H 5 )-CH 2 -O-CH 2 -CH 3?

10 As mentioned above, molecules generally fragment between functionalities. If this were the actual structure, we would probably see an intense fragment with M = 91, resulting from the bond breakage between the leftmost CH 2 and the oxygen atom. However there is no fragment at this mass in the spectrum. Perhaps the molecule is an isomer: (C 6 H 5 )-O-CH 2 -CH 2 -CH 3 If the molecule fragmented such that the oxygen was attached to the aromatic ring, but the rest of the chain broke off, then these fragments would be left: (C 6 H 5 )-O and + CH 2 -CH 2 -CH 3. Generally a hydrogen atom will shift to stabilize an oxyanion, resulting in this fragment: (C 6 H 5 )-OH + (m/z = 94) Referring back to the mass spectrum, m/z = 94 is the most prominent peak, making this structure the most likely candidate. Thus the final proposed structure is (C 6 H 5 )-O-CH 2 -CH 2 -CH 3. Clearly, the interpretation of mass spectra takes a lot of finesse. If you were to do this more frequently, common fragmentation patterns would become much more obvious to you. Fortunately modern chemists can utilize the on-line spectral libraries of GC-MS systems to narrow down the likely choices. The Mass Spectrometry Facility has purchased a copy of the NIST 02 library which contains reference spectra for over 150,000 compounds. Most database searching programs arrive at their suggestions by comparing the analyte mass spectrum to all of those in the library. However it is essential to understand how to properly interpret mass spectra in order to properly assess the validity of a computer-generated database matches. 4. Incomplete Mass Spectra Although mass spectra databases are very powerful tools for identifying the components of a complex mixture, they are far from perfect. No database is complete, and there will be times (especially when synthesizing novel compounds) that the molecule of interest simply will not be in the database. The NIST 02 library contains over 170,000 entries (150,000 compounds, there is some redundancy in this database), but there are millions of potential organic compounds. Another challenge arises when a component of a mixture is present at very low levels. Often the mass spectra observed will be incomplete. That is, only the most intense fragment ions in the mass spectrum will show up because the other peaks will be too weak to be considered by the software. One solution to this would be to inject a more concentrated sample, thus increasing the

11 concentration of the component in question. This is not always an option however as analysts are often sample limited in the real world. Furthermore, injection of a more concentrated sample may foul the column or EI source of the GC-MS significantly shortening the lifetime of those components. Remember, the concentration of the major components in a mixture will increase as well. Figure 3 contains the mass spectra of dodecane, 1-dodecene, and 1-dodecyne and an unknown, low abundance compound with 14 carbons. Look at the three 12-carbon molecule spectra and notice the differences in the masses of the low mass (<100) fragment ions. By comparing the low mass portion of the unknown spectrum to the other three, the identity should be clear. Check with your AI during the lab period for the correct answer. A similar analysis can be to identify compounds with shared structure features. In this experiment, one of the components of olive oil will likely be insufficiently abundant for a proper database match. You will have to compare its mass spectrum to the mass spectra from olive oil s other components to determine its identity.

12 Figure 3: Mass spectra of dodecane, 1-dodecene, 1-dodecyne and an unknown 14-carbon hydrocarbon. Note: the bottom 25% of the 14-carbon mass spectrum has been deleted to simulate the mass spectrum from a low abundance compound. Response Factors and Quantitation As stated above, the response observed for a given compound is proportional to its concentration in the gas stream. Unfortunately, not all compounds are ionizable or detectable with similar efficiencies. This problem is also encountered in optical spectroscopy. Recall that Beer s law contains a proportionality constant (the molar absorptivity, ε) that relates how well any compound absorbs the wavelength in question. If ε is not readily known, a line relating peak area and concentration of a compound can be created by analyzing a handful of standards with known concentrations of the compound in question. This method can become cumbersome as the number of analytes in a given sample increases. An alternative to generating a calibration curve for each analyte is computing response factors. A response factor is a ratio between the observed response (peak area) for a given concentration of a molecule with the response from the same amount of a closely related standard (e.g. a stable isotope labeled version or a structural isomer). According to method 8270C from the United States Environmental Protection Agency, the response factor is computed as follows: areaanalyte conc RF = area conc std std analyte The response factor is used to normalize the peak areas of the other analytes using this equation: In this experiment, you will normalize all response factors to that of palmitic acid methyl ester. The corrected areas will then be used to compute the compositions of the oils. Fatty Acid Analysis area area corrected = RF analyte analyte The calorie, vitamin, mineral, protein, carbohydrate, and fat content of nearly every food item sold in the United States must be clearly shown on the package to allow consumers to make informed nutrition choices. These values are determined using a wide array of analytical chemistry techniques including atomic absorption spectroscopy, bomb calorimetry, gas chromatography, gravimetric analysis, liquid chromatography, and ultraviolet-visible

13 spectroscopy. This particular experiment will use GC-MS to determine the fatty acid profiles of a handful of commercially available cooking oils. For most foods, a chemical method is required to isolate the nutrient of interest from the food prior to analysis (typically homogenization followed by refluxing solvent extraction (aka Soxhlet extraction), supercritical fluid extraction, solid phase extraction, enhanced microwave extraction, etc.). The extracted nutrients are often then further modified to make them more amenable to a particular analysis method (e.g. a chromophore may be added, the nutrient may be converted to an easily quantified small molecule, the nutrient may be esterified, etc.). Fats exist as triglycerides, three fatty acids attached to a glycerol molecule with ester bonds. Fatty acids are long chain carboxylic acids that are isolated from naturally occurring fats. They can be saturated (fully hydrogenated, i.e. all carbon-carbon bonds are single bonds), monounsaturated (one double bond between carbon bonds in the chain), and poly-unsaturated (more than one double bond in the carbon chain). All double bonds in natural fatty acids are cis isomers. The degree of saturation determines many important properties of the fat including flavor, resistance to oxidation, melting point, and smoke point. Olive oil, which contains more than 70% unsaturated fatty acids melts at -6 C and has a smoke point of about 210 C. Refined coconut oil (8% unsaturated fatty acids) melts at 25 C and does not smoke until 232 C. Food processors often partially hydrogenate vegetable oils to raise their melting points, lengthen their shelf lives, or change the texture of the finished product. The high temperatures and pressures of the hydrogenation process can cause isomerization in unsaturated fatty acids, creating trans fatty acids. Table 1 contains a list of common fatty acids and their sources. There is an overwhelming body of evidence concerning the health effects of the various types of fatty acids, and as a result food labels must now list the amount of saturated, unsaturated, and trans fatty acids in each serving. These effects are summarized in an American Heart Association publication ( Saturated fatty acids have been shown to raise low density lipoproteins (LDL or bad cholesterol). Cis isomers of mono- and poly-unsaturated fatty acids have been shown to lower blood cholesterol levels. Trans isomers of unsaturated fatty not only raise LDL but lower heart protecting high density lipoproteins (HDLs or good cholesterol); some researchers consider them worse than saturated fats. The federal government has mandated that the saturated, unsaturated, and transfat content of all foods be included on the label. It also stipulated that foods with less than 0.5 g of trans fatty acids per serving could be labeled as having 0 g trans fat.

14 Table 1: Names, properties, and sources of common fatty acids. Common Name Carbon atoms double bonds IUPAC name Sources Butyric acid 4 0 butanoic acid Butterfat Caproic Acid 6 0 hexanoic acid Butterfat Caprylic Acid 8 0 octanoic acid coconut oil Capric Acid 10 0 decanoic acid coconut oil Lauric Acid 12 0 dodecanoic acid coconut oil Myristic Acid 14 0 tetradecanoic acid palm kernel oil Palmitic Acid 16 0 hexadecanoic acid palm oil Palmitoleic Acid hexadecenoic acid animal fats Stearic Acid 18 0 octadecanoic acid animal fats Oleic Acid 18 1 cis 9-octadecenoic acid olive oil Elaidic Acid 18 1 trans 9-octadecenoic acid Hydrogenated oils Ricinoleic acid hydroxy-9-octadecenoic acid castor oil Vaccenic Acid octadecenoic acid Butterfat Linoleic Acid ,12-octadecadienoic acid grape seed oil Alpha-Linolenic Acid ,12,15-octadecatrienoic acid Flaxseed (linseed) oil Gamma-Linolenic Acid ,9,12-octadecatrienoic acid borage oil Arachidic Acid 20 0 eicosanoic acid peanut oil, fish oil Gadoleic Acid eicosenoic acid fish oil Arachidonic Acid ,8,11,14-eicosatetraenoic acid liver fats EPA ,8,11,14,17- eicosapentaenoic acid fish oil Behenic acid 22 0 docosanoic acid rapeseed oil Erucic acid docosenoic acid rapeseed oil DHA ,7,10,13,16,19- docosahexaenoic acid fish oil Lignoceric acid 24 0 tetracosanoic acid Table adapted from small amounts in most fats In this experiment, we will saponify the oils to release the fatty acids. Saponification, or treating the fat with a strong base, hydrolyzes the triglyceride, creating one molecule of glycerol and three fatty acid molecules. Since the free fatty acids are not very volatile, they will be converted into methyl esters by reacting them with boron triflouride in methanol. The resulting fatty acid methyl esters (FAMEs) are then injected into the GC-MS. Mass spectra will be used to determine the identities of the FAMEs; GC retention times will be used to discriminate between various FAME isomers.

15 Gas Chromatography-Mass Spectrometry Experiment Materials Glassware and other equipment: 5 8 test tubes ml beaker 1 hot plate 1 10 ml graduated cylinder 2 25 ml graduated cylinders ml Erlenmeyer flasks with stoppers 100 ul variable pipette 1 ml pipette 7 autosampler vials Reagents: 75 µl of olive oil 75 mg of 0 g trans fat shortening 75 µl/mg of 3 different fats of your choosing (not olive or 0 g trans fat shortening) 18 ml of 0.5 M NaOH in CH 3 OH 24 ml of 14% BF 3 in CH 3 OH 120 ml of saturated NaCl in H 2 O 150 ml of hexane 100 g of anhydrous Na 2 SO 4 1 ml of standard FAME mix in hexane 1 ml of 25 ppm oleic acid (cis 18:1) methyl ester in hexane 1 ml of 25 ppm elaidic acid (trans 18:1) methyl ester in hexane Procedure (adapted from J. F. Robinson and J. Neyer-Hilvert in J. of Chem. Educ. 74: (1997)) Part 1: Saponification of the Oils 1) Place a half-filled 250 ml beaker of water onto a hot plate. Bring this to a slow boil. 2) Get small samples of olive oil, 0 g trans fat shortening, and three other fats for analysis. NOTE: The solids should be scooped out with a spatula and put on weighing paper. You will need ~75 mg of the solid greases. Be sure to place the glob of grease at the BOTTOM of the test tube so it can react with the NaOH. Also, you will be unable to quantitatively transfer the solid greases; this is acceptable as all measurements in this experiment are relative to the total amount in the test tube. You will need 75 µl of each of the liquid fats. 3) Be sure to CLEARLY LABEL all test tubes, flasks and autosampler vials as all solutions in this experiment are clear. 4) Pour 3 ml of the NaOH solution into five 8 test tubes. 5) Add 75 µl/mg of each fat into a clean test tube and place it in a boiling water bath. 6) Boil them until the solutions are homogenous. 7) Remove the test tubes from the bath and let them cool for 3-5 minutes.

16 Part 2: Esterifying the free fatty acids 8) Add a boiling chip and 5 ml of 14% BF 3 /MeOH solution to each tube. 9) Boil the solution for 5 minutes in the water bath. 10) Remove the test tubes from the bath and let them cool. Part 3: Extracting the FAMEs from the reaction mixture 11) Pour 20 ml of saturated NaCl and 25 ml of hexane into five 125 ml flasks. 12) Pour each cooled reaction mixture into a separate 125 ml flask containing the NaCl solution and hexane. 13) Seal the flask and shake vigorously for 2 minutes. You want to make sure the 2 layers are mixed very well. Make sure to vent the flask occasionally to release the pressure. 14) Allow the layers to settle for ~1 minute. 15) Decant the hexane layer (the upper layer) into another 125 ml flask containing several grams of sodium sulfate. 16) Add 25 ml of hexane to the salt/methanol solution and extract any remaining FAMEs from the salt water and methanol. 17) Combine the 2 nd hexane extract with the first extract. Part 4: Diluting the FAME extract for GC-MS analysis 18) Add 20 µl of each FAME extract into separate autosampler vials containing 1 ml of hexane and mix well. 19) Place 1 ml of each standard solution into separate autosampler vials. 20) Take all the autosampler vials and some clean hexane up to the mass spectrometry lab (A411) for analysis. Part 5: Recording Data with the GC-MS 21) Setting up the instrument for the experiments: If the software is not already open, log on to the computer using your network ID and password and start the ChemStation software by opening Instrument 1 (located on the desktop as well as in the Start menu). Select Method >> Load and 315_FAME.m (it is in the directory C:\MSDCHEM\1\315\Methods ). This step MUST be completed first. 22) Loading the autosampler: Place the autosampler vials containing the three standard solutions into positions 1, 2, and 3 of the autosampler tray; put the five vegetable oil FAME extracts into positions 4-8; put an autosampler vial containing clean hexane into

17 position 9. Make sure the vial labeled A on the turret is full (nearly to the top) with clean hexane. 23) Creating the autosampler sequence table: This step creates the table that determines types of analyses to perform on each sample and specifies the order in which they are to be performed. The table is generated as follows: a. Click the Open Sequence: icon (it has a folder and three autosampler vials in it) and select the sequence BLANK.S from the C:\MSDCHEM\1\315\Sequences directory. b. Click on the Edit Sequence icon (it has a pencil and three autosampler vials). Make sure the data directory is C:\MSDCHEM\1\315\Data\ and the method directory is set to C \MSDCHEM\1\315\Methods. c. Select sample type Sample for all entries. The other sample types are only used with automated data analyses. Enter 1 for the vial. Put a descriptive sample name in the Sample column (e.g. standard FAME mix or corn oil extract ). Select the method 315_FAME after clicking inside the box under Method/Keyword. Enter a file name for the run. It is suggested that you use your initials and a three digit number (e.g. ABC_001). Enter a descriptive comment (similar to the Sample column) in the Comment/Keyword String. Leave the other columns (Sample Amt, Multiplier, Level, etc.) blank. When you are finished, your table row should look something like this: d. Make a two more entries so the contents of the autosampler vial 1 are analyzed three times (for computing averages and standard deviations). Make sure to increment the number in the file name. Create similar triplicate entries for each vial. You should analyze the blank only once with 315_FAME. e. Your completed sample table should have 25 entries and should look similar to the table below.

18 24) Validate the Autosampler Sequence Table: This step ensures that no errors were made in typing the names of the methods and that there is enough room on the hard drive to record your data. Save your completed table by clicking the Save Sequence icon (it has a red disk and 3 vials on it). Save the sequence in the directory C:\MSDChem\1\315\Sequences. Give it a file name involving your initials so it will be easy to find again, if necessary. Click the Simulate Sequence icon (it has a red check mark and 3 vials on it). Make sure the Full Method button is selected in the Method Sections To Run box. Then click Run Sequence. Correct any errors that arise and resave your sequence. Make sure to view and print the verification report so you can check your notebook entries. Close the sequence verification report after you have printed it. 25) Start the Instrument: Click on the Run Sequence icon (it has a running figure and 3 autosampler vials on it). Make sure the Full Method button is selected in the Method Sections To Run box. Then click Run Sequence. The instrument will perform the 26 analyses specified in the sequence table.

19 At the end of each day Dispose of any unneeded hexane solutions in the ORGANIC WASTE container. Dispose of the BF 3 /methanol/salt water solution in the AQUEOUS WASTE container. If you need to keep a hexane solution overnight, cap it and seal the top with a small strip of Parafilm. Make sure any flasks you keep are labeled and stored upright in the refrigerator or freezer rather than a drawer. Always clean your glassware, the balance area, and your bench space before you leave. Part 6: Analyzing the GC-MS Data (Note: this section may be performed either in the Mass Spectrometry Facility, A454. See the Appendix entitled How to use ChemStation in the Library for C315 Students for instructions on remote access to the GC-MS computer. It also contains a brief introduction to ChemStation.) 26) Load the software and method: Open Enhanced Data Analysis by double clicking on its icon. Load the 315_FAME.M method. 27) Analyze the Data from the FAME mixture: a. Open the data from the standard FAME mixture. b. Click the integrate button on the software (it looks like a chromatogram with 2 peaks and an arrow pointing between them). If you are unsure what a particular button does, mouse over it for a few seconds, and a brief description of that button s function will appear. c. Print the chromatogram by clicking the printer icon. You will need this for your report. d. Select Integration Results from the Chromatogram menu. Copy the table by clicking the Copy button and paste the data into a spreadsheet for later processing. e. Zoom in on each peak and extract its mass spectrum by dragging across the peak with the right mouse button. f. Make sure the NIST 02 library is selected and search the mass spectra against the NIST 02 library. Use the results from the searches to add names to the table in your lab report. g. Using the peak areas from the table and the equations in the handout, determine a response factor for each fatty acid. Use palmitic acid methyl ester as the basis of your factors. Ask your AI for the Certificate of Analysis for the FAME standard. 28) Determine the elution order for cis and trans 9-octadecenoic acid methyl esters: Open the data file from the oleic acid methyl ester (cis 9-octadecenoic acid) solution. Note the retention time. Do the same for the elaidic acid methyl ester (trans 9-octadecenoic acid).

20 29) Determine the response factor for elaidic acid methyl ester: Using a procedure similar to that in step 26, compute the response factor for elaidic acid methyl ester. 30) Analyze the data from the oil samples: a. Open the data file from the olive oil sample and integrate the chromatogram. b. Using both mass spectra and retention times, identify the components of each oil. You will need to tabulate these data for your report. c. Using the response factors calculated in steps 26 and 27, compute the fatty acid composition for each oil. Add these results to your table. Be sure to list all fatty acids more than 0.5% abundant. d. Repeat steps 29a-29c for the remaining oil samples. e. For any fatty acid for which a response factor was not computed, assume a response factor of 1.00 f. For the olive oil sample, determine the identity of the low intensity feature that elutes just before palmitic acid methyl ester.

21 Part 7: Lab Report 1. Attach at the end of your report one representative chromatogram for each standard sample analyzed (standard FAME mixture chromatogram, oleic acid chromatogram, and elaidic acid chromatogram). DO NOT print each chromatogram it is not necessary and wastes paper. 2. Create a table summarizing the chromatographic conditions used for the analyses. Include the type of column used, the dimensions of the column, the injection volumes, split ratios, and the temperature program. Column data are posted on the instrument and the temperature program can be viewed from the Oven icon in the Acquisition window. Discuss how the method should be modified if the fatty acids had longer hydrocarbon chains. 3. Create a table summarizing the retention times for the FAMEs in the standard mixture. Include the averages and standard deviations for the retention times for each FAME in the standard. In separate columns in this table, include the name, concentration (from the certificate of analysis), peak area (average and standard deviation), and response factor for each standard FAME. 4. Include in your report the mass spectra for the oleic acid and elaidic acid FAMEs. Describe how you can determine which of the two isomers is present when only one is found in an unknown sample. 5. For each of the unknown oils, create a table that lists each of the fatty acid components with a relative abundance of 0.5% or higher. This table should include name, retention time, peak area, corrected peak area, and relative concentration (average and standard deviation). Find a reliable reference and report the amount of each fatty acid known to be present in each oil. Compare the relative concentration you measured to this known amount. Comment on how well these numbers agree with each other. 6. Based on the data collected for your unknowns, which oil would you report to be the healthiest? 7. In the chromatogram for olive oil, you should find a small peak with a retention time of approximately 7.9 minutes. Include a mass spectrum of this compound in your report and any evidence that you can use to help identify this compound. Explain the quality of the mass spectrum and the impact of this quality on the data interpretation. 8. Based on your data table for trans-fat free Crisco, compute the number of grams of trans fat in one serving (one tablespoon, which is 14 g for liquid oils and 12 g for solid oils/greases). Is this product truly trans-fat free? Comment on your findings.

22 9. Other than the balance and the glassware used, what are the major sources of error in GC-MS? How can these errors be minimized or eliminated? Does the autosampler address any of these issues? If so, how? Gas Chromatography Mass Spectrometry Report Grade Sheet Section Points Qualifier 25 Introduction 10 Experimental 10 Results and Discussion FAME Analysis (a) Chromatographic conditions Table of chromatographic conditions 5 Conditions for longer chain fatty acid 5 (b) Standard results Chromatograms 5 Table of results 5 Response factors 5 Oleic & elaidic acid MS 5 (c) Unknown oils Table of results 10 Comparison to known values 10 Healthiest oil 5 Olive oil peak 5 Trans-fat free Crisco 5 (d) Sources of error 5 Conclusions 5 Literature references 5

23 Appendix: How to use ChemStation in the Library for C315 Students: ChemStation is installed on PC # 5 in the library. This PC is located next to the emergency exit at the far end of the library (near where the journal, Tetrahedron Letters is shelved). Login to PC #5 using your IU username and password. When the PC boots up, double click on the Instrument #1 Data Analysis icon. It should be on the desktop. If not, it can be found on the start menu in Programs Dept. Sponsored Chemistry MSD Chem Instrument #1 Data Analysis You will need to map C315 directory on the GC-MS computer in the mass spectrometry lab as follows: Click on the start bar and select Run Put \\bl-chem-msfgcmf into the Open dialog box. Right click on the C315_Data_Methods icon. Select Map Network Drive Select a drive letter for the connection (e.g. W:). DO NOT SELECT Reconnect at logon! Click on Finish (see below) Your data will now be on the drive with the letter you selected. For example, if you chose W:, your data will be found in the directory W:\Data. A copy of the NIST 02 library is also found in the C315 directory. Be sure to go to the Spectrum pull-down menu and choose Select Library

24 Click on the Browse icon and go to the C315 directory (e.g. W:) and select NIST02. This will allow you to do your data searches. You will need to check with library personnel to see where printouts from PC 5 appear.

25 Quick Guide to Data Analysis on ChemStation: The left mouse button is used to zoom in on peaks. This feature works in either the chromatogram or mass spectrum windows. You draw a box around the area you want to zoom in on. The Y axis does NOT autoscale in the chromatogram window. Whatever is in the box is what fills the chromatogram window after the zoom. I suggest you start your zoom boxes from the bottom of the chromatogram. This way you can be sure that you do not cut off the tops of peaks. Double left-clicking will undo a previous zoom. If you want to redraw and entire chromatogram, click on the Draw Chromatogram icon (the icon looks like a little teal chromatogram with no arrows, letters, or other stuff in it). The GC method (containing temperature program, split ratio, mass spec conditions, etc.) can be found by selecting List Header from the File menu. The instrument parameters are found in the acqmeth.txt window. The Copy Window item under the Tools menu can be used for copying chromatograms and/or spectra into other programs. Window 1 is mass spectrum, Window 2 is the chromatogram. You can direct the computer to integrate the chromatogram by clicking on the integrate icon. It looks like a little blue chromatogram with a single, downward pointing arrow on it. Alternatively you can select Integrate from the Chromatogram menu. The results of the integration are viewed by selecting the Integration Results icon from the Chromatogram menu. The right mouse button is used to select the portion of the chromatogram from which you wish to extract mass spectra. I suggest you draw the extraction boxes INSIDE chromatographic peaks to avoid extracting too much noise. Double right-clicking on a mass spectrum searches it against the selected library. You should only search SCAN mode data and use the NIST02 library. The Tabulate option under the Spectrum menu generates a table of masses and intensities from the displayed mass spectrum. This table can then be copied and pasted into Excel for further analysis. ChemStation does have a rather complete set of online help documents which can be accessed from the Help menu. Direct any further questions to Jonathan Karty, jkarty@indiana.edu.

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

Analyzing Small Molecules by EI and GC-MS. July 2014

Analyzing Small Molecules by EI and GC-MS. July 2014 Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,

More information

Background Information

Background Information 1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software

More information

13C NMR Spectroscopy

13C NMR Spectroscopy 13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number

More information

Analysis of Liquid Samples on the Agilent GC-MS

Analysis of Liquid Samples on the Agilent GC-MS Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >

More information

Expectations for GC-MS Lab

Expectations for GC-MS Lab Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will

More information

Mass Spectrometry. Overview

Mass Spectrometry. Overview Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2

More information

Lab 2 Biochemistry. Learning Objectives. Introduction. Lipid Structure and Role in Food. The lab has the following learning objectives.

Lab 2 Biochemistry. Learning Objectives. Introduction. Lipid Structure and Role in Food. The lab has the following learning objectives. 1 Lab 2 Biochemistry Learning Objectives The lab has the following learning objectives. Investigate the role of double bonding in fatty acids, through models. Developing a calibration curve for a Benedict

More information

Organic Chemistry Lab Experiment 4 Preparation and Properties of Soap

Organic Chemistry Lab Experiment 4 Preparation and Properties of Soap Organic Chemistry Lab Experiment 4 Preparation and Properties of Soap Introduction A soap is the sodium or potassium salt of a long-chain fatty acid. The fatty acid usually contains 12 to 18 carbon atoms.

More information

Fractional Distillation and Gas Chromatography

Fractional Distillation and Gas Chromatography Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils

More information

Chapter 5 Classification of Organic Compounds by Solubility

Chapter 5 Classification of Organic Compounds by Solubility Chapter 5 Classification of Organic Compounds by Solubility Deductions based upon interpretation of simple solubility tests can be extremely useful in organic structure determination. Both solubility and

More information

Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012.

Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012. Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012 Introduction Screen captures in this document were taken from MassHunter B.05.00

More information

Separation of Amino Acids by Paper Chromatography

Separation of Amino Acids by Paper Chromatography Separation of Amino Acids by Paper Chromatography Chromatography is a common technique for separating chemical substances. The prefix chroma, which suggests color, comes from the fact that some of the

More information

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080 Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien allis@stanford.edu 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD

More information

Colorimetric Determination of Iron in Vitamin Tablets

Colorimetric Determination of Iron in Vitamin Tablets Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered

More information

Micromass LCT User s Guide

Micromass LCT User s Guide Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project

More information

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated

More information

CHE334 Identification of an Unknown Compound By NMR/IR/MS

CHE334 Identification of an Unknown Compound By NMR/IR/MS CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared

More information

Suggested solutions for Chapter 3

Suggested solutions for Chapter 3 s for Chapter PRBLEM Assuming that the molecular ion is the base peak (00% abundance) what peaks would appear in the mass spectrum of each of these molecules: (a) C5Br (b) C60 (c) C64Br In cases (a) and

More information

Setting up a Quantitative Analysis MS ChemStation

Setting up a Quantitative Analysis MS ChemStation Setting up a Quantitative Analysis MS ChemStation Getting Ready 1. Use the tutorial section "Quant Reports" on the MSD Reference Collection CD-ROM that came with your ChemStation software. 2. Know what

More information

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start. Q-TOF User s Booklet Enter your username and password to login. After you login, the data acquisition program will automatically start. ~ 1 ~ This data acquisition window will come up after the program

More information

Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass

Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons, neutrons and electrons Nuclear Structure particle relative charge relative mass proton 1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons and neutrons make up

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis Main points of the chapter 1. Hydrogen Nuclear Magnetic Resonance a. Splitting or coupling (what s next to what) b. Chemical shifts (what type is it) c. Integration

More information

SCION SQ Series GC-MS

SCION SQ Series GC-MS SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative

More information

Making Biodiesel from Virgin Vegetable Oil: Teacher Manual

Making Biodiesel from Virgin Vegetable Oil: Teacher Manual Making Biodiesel from Virgin Vegetable Oil: Teacher Manual Learning Goals: Students will understand how to produce biodiesel from virgin vegetable oil. Students will understand the effect of an exothermic

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND #3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric

More information

Reactions of Fats and Fatty Acids

Reactions of Fats and Fatty Acids Reactions of Fats and Fatty Acids Outline Fats and Oils Fatty Acid Biosynthesis Biodiesel Homework We hear quite a lot about the place of fats and oils in human nutrition. Foods high in fat are at the

More information

0 10 20 30 40 50 60 70 m/z

0 10 20 30 40 50 60 70 m/z Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions

More information

Separation by Solvent Extraction

Separation by Solvent Extraction Experiment 3 Separation by Solvent Extraction Objectives To separate a mixture consisting of a carboxylic acid and a neutral compound by using solvent extraction techniques. Introduction Frequently, organic

More information

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS PRINCIPLE This method confirms the identity of volatile compounds in biological

More information

Chapter 1: Moles and equations. Learning outcomes. you should be able to:

Chapter 1: Moles and equations. Learning outcomes. you should be able to: Chapter 1: Moles and equations 1 Learning outcomes you should be able to: define and use the terms: relative atomic mass, isotopic mass and formula mass based on the 12 C scale perform calculations, including

More information

Honors Chemistry: Unit 6 Test Stoichiometry PRACTICE TEST ANSWER KEY Page 1. A chemical equation. (C-4.4)

Honors Chemistry: Unit 6 Test Stoichiometry PRACTICE TEST ANSWER KEY Page 1. A chemical equation. (C-4.4) Honors Chemistry: Unit 6 Test Stoichiometry PRACTICE TEST ANSWER KEY Page 1 1. 2. 3. 4. 5. 6. Question What is a symbolic representation of a chemical reaction? What 3 things (values) is a mole of a chemical

More information

Name Lab #3: Solubility of Organic Compounds Objectives: Introduction: soluble insoluble partially soluble miscible immiscible

Name  Lab #3: Solubility of Organic Compounds Objectives: Introduction: soluble insoluble partially soluble miscible immiscible Lab #3: Solubility of rganic Compounds bjectives: - Understanding the relative solubility of organic compounds in various solvents. - Exploration of the effect of polar groups on a nonpolar hydrocarbon

More information

Trans Fatty Acids Labeling

Trans Fatty Acids Labeling Medallion Laboratories ANALYTICAL PRGRESS Trans Fatty Acids Labeling Byjonathan W. DeVries Ph.D. and Aruna Prakash, TRANS FAT LABELING Effective January 1, 2006, trans Fat Labeling on Food Products is

More information

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C

More information

Synthesis of Isopentyl Acetate

Synthesis of Isopentyl Acetate Experiment 8 Synthesis of Isopentyl Acetate Objectives To prepare isopentyl acetate from isopentyl alcohol and acetic acid by the Fischer esterification reaction. Introduction Esters are derivatives of

More information

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,

More information

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview

More information

Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance

Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance 1 Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance Read pp 142-155, 161-162, Chapter 10 and pp 163-173, Chapter 11, in LTOC. View the videos: 4.2 Extraction (Macroscale);

More information

methods Electrospray mass spectrometry of human hair wax esters Mark Fitzgerald and Robert C. Murphy 1

methods Electrospray mass spectrometry of human hair wax esters Mark Fitzgerald and Robert C. Murphy 1 Electrospray mass spectrometry of human hair wax esters methods Mark Fitzgerald and Robert C. Murphy 1 Department of Pharmacology, University of Colorado at Denver and Health Sciences Center, Aurora, CO

More information

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies INEF Penn State Conference 2013 Albert Robbat, Jr. and Patrick Antle Tufts University, Chemistry

More information

How To Use Gc-Ms

How To Use Gc-Ms The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,

More information

For example: (Example is from page 50 of the Thinkbook)

For example: (Example is from page 50 of the Thinkbook) SOLVING COMBINED SPECTROSCOPY PROBLEMS: Lecture Supplement: page 50-53 in Thinkbook CFQ s and PP s: page 216 241 in Thinkbook Introduction: The structure of an unknown molecule can be determined using

More information

Symmetric Stretch: allows molecule to move through space

Symmetric Stretch: allows molecule to move through space BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed

More information

Properties of Alcohols and Phenols Experiment #3

Properties of Alcohols and Phenols Experiment #3 Properties of Alcohols and Phenols Experiment #3 Objectives: To observe the solubility of alcohols relative to their chemical structure, to perform chemical tests to distinguish primary, secondary and

More information

Organic Chemistry Calculations

Organic Chemistry Calculations Organic Chemistry Calculations There are three basic units for measurement in the organic laboratory mass, volume, and number, measured in moles. Most of the other types of measurements are combinations

More information

EnAlgae SOP: Biogas analysis

EnAlgae SOP: Biogas analysis Table of content 1. PURPOSE... 2 2. PRINCIPLE... 2 3. REQUIREMENTS... 3 3.1. EQUIPMENT AND MATERIALS... 3 3.2. REAGENTS... 3 3.3. PREPARATION OF MEASUREMENT... 3 4. PROCEDURE... 4 4.1. SAMPLE MEASUREMENT...

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,

More information

experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.

experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration. 81 experiment5 LECTURE AND LAB SKILLS EMPHASIZED Synthesizing an organic substance. Understanding and applying the concept of limiting reagents. Determining percent yield. Learning how to perform a vacuum

More information

Calculation of Molar Masses. Molar Mass. Solutions. Solutions

Calculation of Molar Masses. Molar Mass. Solutions. Solutions Molar Mass Molar mass = Mass in grams of one mole of any element, numerically equal to its atomic weight Molar mass of molecules can be determined from the chemical formula and molar masses of elements

More information

PREPARATION AND PROPERTIES OF A SOAP

PREPARATION AND PROPERTIES OF A SOAP (adapted from Blackburn et al., Laboratory Manual to Accompany World of Chemistry, 2 nd ed., (1996) Saunders College Publishing: Fort Worth) Purpose: To prepare a sample of soap and to examine its properties.

More information

EXPERIMENT 12: Empirical Formula of a Compound

EXPERIMENT 12: Empirical Formula of a Compound EXPERIMENT 12: Empirical Formula of a Compound INTRODUCTION Chemical formulas indicate the composition of compounds. A formula that gives only the simplest ratio of the relative number of atoms in a compound

More information

Element of same atomic number, but different atomic mass o Example: Hydrogen

Element of same atomic number, but different atomic mass o Example: Hydrogen Atomic mass: p + = protons; e - = electrons; n 0 = neutrons p + + n 0 = atomic mass o For carbon-12, 6p + + 6n 0 = atomic mass of 12.0 o For chlorine-35, 17p + + 18n 0 = atomic mass of 35.0 atomic mass

More information

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration GOAL AND OVERVIEW Antacids are bases that react stoichiometrically with acid. The number of moles of acid that

More information

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93)

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93) INTRODUCTION HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93) Headache, sore muscles, arthritis pain... How do you spell relief?

More information

CHEM 2423 Recrystallization of Benzoic Acid EXPERIMENT 4 - Purification - Recrystallization of Benzoic acid

CHEM 2423 Recrystallization of Benzoic Acid EXPERIMENT 4 - Purification - Recrystallization of Benzoic acid EXPERIMENT 4 - Purification - Recrystallization of Benzoic acid Purpose: a) To purify samples of organic compounds that are solids at room temperature b) To dissociate the impure sample in the minimum

More information

CHEMISTRY II FINAL EXAM REVIEW

CHEMISTRY II FINAL EXAM REVIEW Name Period CHEMISTRY II FINAL EXAM REVIEW Final Exam: approximately 75 multiple choice questions Ch 12: Stoichiometry Ch 5 & 6: Electron Configurations & Periodic Properties Ch 7 & 8: Bonding Ch 14: Gas

More information

electron does not become part of the compound; one electron goes in but two electrons come out.

electron does not become part of the compound; one electron goes in but two electrons come out. Characterization Techniques for Organic Compounds. When we run a reaction in the laboratory or when we isolate a compound from nature, one of our first tasks is to identify the compound that we have obtained.

More information

Formulas, Equations and Moles

Formulas, Equations and Moles Chapter 3 Formulas, Equations and Moles Interpreting Chemical Equations You can interpret a balanced chemical equation in many ways. On a microscopic level, two molecules of H 2 react with one molecule

More information

Austin Peay State University Department of Chemistry CHEM 1021 TESTING FOR ORGANIC FUNCTIONAL GROUPS

Austin Peay State University Department of Chemistry CHEM 1021 TESTING FOR ORGANIC FUNCTIONAL GROUPS TESTING FOR ORGANIC FUNCTIONAL GROUPS Caution: Chromic acid is hazardous as are many of the organic substances in today s experiment. Treat all unknowns with extreme care. Many organic substances are flammable.

More information

SUCRALOSE. White to off-white, practically odourless crystalline powder

SUCRALOSE. White to off-white, practically odourless crystalline powder SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA

More information

Organic Chemistry Tenth Edition

Organic Chemistry Tenth Edition Organic Chemistry Tenth Edition T. W. Graham Solomons Craig B. Fryhle Welcome to CHM 22 Organic Chemisty II Chapters 2 (IR), 9, 3-20. Chapter 2 and Chapter 9 Spectroscopy (interaction of molecule with

More information

CH243: Lab 4 Synthesis of Artificial Flavorings by Fischer Esterification

CH243: Lab 4 Synthesis of Artificial Flavorings by Fischer Esterification H243: Lab 4 Synthesis of Artificial Flavorings by Fischer Esterification PURPSE: To prepare esters by reaction of carboxylic acids and alcohols. To modify a known procedure to prepare an unknown. DISUSSIN:

More information

Performing Calculatons

Performing Calculatons Performing Calculatons There are three basic units for measurement in the organic laboratory mass, volume, and number, measured in moles. Most of the other types of measurements are combinations of them,

More information

CHM220 Addition lab. Experiment: Reactions of alkanes, alkenes, and cycloalkenes*

CHM220 Addition lab. Experiment: Reactions of alkanes, alkenes, and cycloalkenes* CM220 Addition lab Experiment: Reactions of alkanes, alkenes, and cycloalkenes* Purpose: To investigate the physical properties, solubility, and density of some hydrocarbon. To compare the chemical reactivity

More information

GOED VOLUNTARY MONOGRAPH (v. 4)

GOED VOLUNTARY MONOGRAPH (v. 4) GOED VOLUNTARY MONOGRAPH (v. 4) Omega-3 EPA Omega-3 DHA Omega-3 EPA & DHA DEFINITION Omega-3 fatty acids, EPA and DHA, consist of the all cis forms of 5, 8, 11, 14, 17-eicosapentaenoic acid and 4, 7, 10,

More information

H H H O. Pre-Lab Exercises Lab 6: Organic Chemistry. Lab 6: Organic Chemistry Chemistry 100. 1. Define the following: a.

H H H O. Pre-Lab Exercises Lab 6: Organic Chemistry. Lab 6: Organic Chemistry Chemistry 100. 1. Define the following: a. Lab 6: Organic hemistry hemistry 100 1. Define the following: a. ydrocarbon Pre-Lab Exercises Lab 6: Organic hemistry Name Date Section b. Saturated hydrocarbon c. Unsaturated hydrocarbon 2. The formula

More information

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS 1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years

More information

Tuning & Mass Calibration

Tuning & Mass Calibration Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within

More information

Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis

Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis 1.0 Purpose - This procedure specifies the required elements for the solid phase extraction of acidic,

More information

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas

More information

Experiment 5: Column Chromatography

Experiment 5: Column Chromatography Experiment 5: Column Chromatography Separation of Ferrocene & Acetylferrocene by Column Chromatography Reading: Mohrig, Hammond & Schatz Ch. 18 pgs 235-253 watch the technique video on the course website!

More information

Quick Start Guide Chromeleon 7.2

Quick Start Guide Chromeleon 7.2 Chromeleon 7.2 7229.0004 Revision 1.0 July 2013 Table of Contents 1 Introduction... 1 1.1 About this Document... 1 1.2 Other Documentation... 2 2 Using Chromeleon... 3 2.1 Overview... 3 2.2 Starting Chromeleon...

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

CHEM 121. Chapter 19, Name: Date:

CHEM 121. Chapter 19, Name: Date: CHEM 121. Chapter 19, Name: Date: 1. A lipid is any substance of biochemical origin that is A) soluble in water but insoluble in nonpolar solvents B) insoluble in both water and nonpolar solvents C) insoluble

More information

Laboratory 28: Properties of Lipids

Laboratory 28: Properties of Lipids Introduction Lipids are naturally occuring substances that are arbitrarily grouped together on the basis of their insolubility in water (a polar solvent) and solubility in nonpolar solvents. Lipids include

More information

Experiment 8 Preparation of Cyclohexanone by Hypochlorite Oxidation

Experiment 8 Preparation of Cyclohexanone by Hypochlorite Oxidation Experiment 8 Preparation of Cyclohexanone by ypochlorite xidation In this experiment we will prepare cyclohexanone from cyclohexanol using hypochlorite oxidation. We will use common household bleach that

More information

Solving Spectroscopy Problems

Solving Spectroscopy Problems Solving Spectroscopy Problems The following is a detailed summary on how to solve spectroscopy problems, key terms are highlighted in bold and the definitions are from the illustrated glossary on Dr. Hardinger

More information

Paper Chromatography: Separation and Identification of Five Metal Cations

Paper Chromatography: Separation and Identification of Five Metal Cations Paper Chromatography: Separation and Identification of Five Metal Cations Objectives Known and unknown solutions of the metal ions Ag +, Fe 3+, Co 2+, Cu 2+ and Hg 2+ will be analyzed using paper chromatography.

More information

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas

More information

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10

More information

SALTS of FATTY ACIDS

SALTS of FATTY ACIDS SYNONYMS INS No. 470 SALTS of FATTY ACIDS Prepared at the 33rd JECFA (1988), published in FNP 38 (1988) and in FNP 52 (1992). Metals and arsenic specifications revised at the 55th JECFA (2000). An ADI

More information

Survival Organic Chemistry Part I: Molecular Models

Survival Organic Chemistry Part I: Molecular Models Survival Organic Chemistry Part I: Molecular Models The goal in this laboratory experience is to get you so you can easily and quickly move between empirical formulas, molecular formulas, condensed formulas,

More information

ISOLATION OF CAFFEINE FROM TEA

ISOLATION OF CAFFEINE FROM TEA ISLATIN F CAFFEINE FRM TEA Introduction In this experiment, caffeine is isolated from tealeaves. The chief problem with the isolation is that caffeine does not exist alone in the tealeaves, but other natural

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Determination of Specific Nutrients in Various Foods. Abstract. Humans need to consume food compounds such as carbohydrates, proteins, fats,

Determination of Specific Nutrients in Various Foods. Abstract. Humans need to consume food compounds such as carbohydrates, proteins, fats, Determination of Specific Nutrients in Various Foods Abstract Humans need to consume food compounds such as carbohydrates, proteins, fats, and vitamins to meet their energy requirements. In this lab, reagents

More information

Chapter 6. Solution, Acids and Bases

Chapter 6. Solution, Acids and Bases Chapter 6 Solution, Acids and Bases Mixtures Two or more substances Heterogeneous- different from place to place Types of heterogeneous mixtures Suspensions- Large particles that eventually settle out

More information

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT AppNote 6/2003 Analysis of Flavors using a Mass Spectral Based Chemical Sensor Vanessa R. Kinton Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA Kevin L. Goodner USDA, Citrus & Subtropical

More information

Part One: Mass and Moles of Substance. Molecular Mass = sum of the Atomic Masses in a molecule

Part One: Mass and Moles of Substance. Molecular Mass = sum of the Atomic Masses in a molecule CHAPTER THREE: CALCULATIONS WITH CHEMICAL FORMULAS AND EQUATIONS Part One: Mass and Moles of Substance A. Molecular Mass and Formula Mass. (Section 3.1) 1. Just as we can talk about mass of one atom of

More information

Infrared Spectroscopy 紅 外 線 光 譜 儀

Infrared Spectroscopy 紅 外 線 光 譜 儀 Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed

More information

Distillation of Alcohol

Distillation of Alcohol CHEM 121L General Chemistry Laboratory Revision 1.6 Distillation of Alcohol To learn about the separation of substances. To learn about the separation technique of distillation. To learn how to characterize

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

A Beer s Law Experiment

A Beer s Law Experiment A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining

More information