Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Size: px
Start display at page:

Download "Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview."

Transcription

1 Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide the user with the ability to create MS methods that continually alternate between scanning and selected ion monitoring (SIM) acquisitions in what is known as SIM-Scan. The ability to obtain scan information and SIM information almost simultaneously throughout the chromatographic run can be a powerful tool. For example, in pesticide screening the analyst may have a list of target compounds that must be assayed at very low detection limits and is therefore conducted by SIM. To avoid missing compounds of concern that may not be included in the SIM target list, a second analysis (duplicate injection) by scan acquisition is no longer necessary since this scan acquisition can be added to the SIM method. Similarly, in some environmental analyses, certain legislated methods (for example, some U.S. EPA methods) require scan acquisition of target compounds that provides data for comparison to spectral libraries as a form of compound confirmation and simultaneously requires quantitative (scan) acquisitions of these target compounds. Because the scan quantitation is typically based on a subset of the ions in each compound s spectrum (usually three or four), the SIMscan method can provide SIM acquisition of these ions yet also generate full scan data to satisfy the library match requirements. The advantage of this approach is target compound ion ratios provided by the SIM portion of the SIM-scan acquisition will have better statistics (better reproducibility and accuracy) than a scan-only method. This results in improved detection limits, linearity, etc., while the scan portion of the SIM-Scan acquisition can be arranged to provide compound spectra of high quality. This technical overview will provide guidance on how to make these arrangements and cite the tools available in the MSD Productivity Software package G1701DA (D.03 revision) of the 5975B MSD software to assist SIM-Scan method development. SIM-Scan Function In synchronous SIM-Scan, the instrument rapidly and sequentially switches between a scan acquisition and a SIM acquisition. This is best illustrated in the diagram shown in Figure 1. Over the course of the elution of the chromatographic peak, spectra are recorded according to the arranged SIM or scan parameters. The bar beneath the peak indicates an important idea; the time spent in scan or SIM acquisition modes is not the same and is determined by the acquisition parameters in those modes. Success in SIM-Scan is determined by optimizing the time spent in either mode relative to the chromatographic peak width.

2 Figure 1. SIM-Scan Data Chromatographic peak time Scan time SIM time The SIM-Scan situation. Scan data points SIM data points Figure 2 shows the SIM-Scan data file. The upper window (2) contains the scan data and the next lower window the SIM data. Either signal can be loaded independently by the menu item File\Select Signals. In the scan data window, the cursor allows spectra to be selected and displayed in the lowest window. In the SIM data window, the cursor performs integrations. SIM-Scan Considerations There are two existing application notes that are critical to an understanding of the considerations needed for SIM-scan methods: 1. New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods, Agilent Technologies, publication EN Inert Performance Electronics: Considerations for GC/MS Methods in Scan and Selected Ion Monitoring Modes, Agilent Technologies, publication EN It is assumed that these publications have been read and understood in the discussion that follows. If users wish to add SIM to their existing scan method or add scan to their SIM method, the final SIM-Scan method should be arranged to provide Scan data SIM data Figure 2. SIM-Scan data. 2

3 quantitative quality data from the SIM data and qualitative data from the scan signal in the vast majority of cases. The scenario of using both SIM and scan acquisitions for quantitation should immediately be understood as a less optimal approach. This is because the guiding rule of mass spectroscopy is sample all chromatographic data as slowly as possible or sample only as rapidly as necessary. The best quality data is always obtained at the lowest possible speeds because there is more opportunity to obtain good sampling of the peaks. So good qualitative data (higher quality library matches) is obtained at low speeds and the rule of thumb is four or five scan data points over the peak. The rule in SIM is generally eight to 12 scans over the peak. A refinement is worth mentioning. If you are using SIM ion areas to quantitate the peak, then 6 to <12 SIM data points will suffice. However, if quantitation by peak height is used, then eight or more are required. The situation is always checked by examining the lowest concentration standard. The objective of the SIM-scan method then is four scan data points over the peak and eight SIM data points over the peak. The MSD Productivity Software package G1701DA (D.03 revision) has some enhanced tools to assist method development. AutoSIM Method. This will present the panel shown in Figure 3 and described in detail in the previously cited publication ( EN). The key parameters from a SIM-Scan point of view are: a) The minimum time between peak and SIM ion group start. Shorter time intervals are required for target lists with many compounds. A guiding rule is to look at the chromatographic peak width at the peak base for the highest concentration standard and multiply that 1.5 to get a starting parameter. Shorten this time if there are too many compounds as this will divide the chromatogram into more SIM groups. b) The minimum number of scan or SIM cycles on a peak. Since this is a higher concentration standard then the typical analysis and the method will be a SIM-Scan method, choose a higher number like 10 or 12. This parameter will determine the ion dwell times. SIM-Scan Method Development The more detailed description of setting up SIM methods ( EN) applies here as well. In brief: 1. Apply Retention Time Locking (RTL) to your GC-MS method. RTL provides two important characteristics to your compound retention times (RTs) and your method: compound RTs become permanent and universal. Permanent meaning that they can be reproduced by your method after column maintenance or replacement; and universal meaning that compound RTs can be replicated by any similar instrument anywhere within your lab or elsewhere on the planet or beyond for that matter. 2. Under this method acquire a high concentration standard using scan. Use the resulting data file to select SIM ions for each desired target compound and add these to your quantitation database. 3. With the acquired data file loaded in Data Analysis, select under Method\Generate Note: The minimum dwell time used in the automated SIM calculations is 10 ms, while in the Instrument Control the minimum dwell time is 1 ms. Figure 3. AutoSIM-Scan setup. 4. After a panel appears to print reports, another panel will provide three options (Figure 4) for the new method: make no changes, make a SIM method, or make a SIM-Scan method. Select the SIM-Scan method option. 3

4 Figure 4. Method selection parameters. 5. The following panel asks which signal will be used for quantitation in the database. One should always select the SIM signal as the target option. This means the database will use the SIM data file to build calibration tables and generate quantitative data (Figure 5). 6. Lastly, acquire another standard and check the following: In the scan data, see if there are four cycles across the chromatographic peaks. If too many scans are across the peak, reduce the samples; if too few, increase the samples. In the SIM data, similarly check the data sampling over the chromatographic peaks. Use the SIM table Editor (Method\Edit SIM Parameters) to modify or refine the dwell times for the Select SIM ion Figure 5. SIM-Scan compound quantitation database. 4

5 groups (Figure 6). SIM ion dwell times as short as 1 msec are supported by the 5975B MSD (see below in Employing Rapid SIM ) but are accessible in Instrument Control. Frequently Asked Questions Will I lose signal in SIM-Scan? If you examine the extracted ion signals in scan and SIM for a particular ion, you should find the areas and heights comparable if you are not oversampling in either mode. This is because the same ion is being measured in both modes. What can SIM-Scan tell me? When properly configured, a SIM-Scan method will produce high-quality scan data providing good matches to library spectra. The scan peaks may look less smooth or a little rough but this improves the quality and intensity of the peaks over those that have been scanned more rapidly and show more Gaussian shapes. The SIM data will result in good ion ratio reproducibilities that can be confirmed by replicate injections. These are the criteria for a suitable method. Additionally, typical SIM-only acquisitions of prepared samples may suffer from shifting retention times, poor peak shapes, issues with rapid deterioration of signal, etc., due to the properties of the injected samples themselves; however, because SIM-Scan provides a scan acquisition to examine, problems in sample preparation that produce these effects can be rapidly and readily apprehended. Even a cursory examination of the scan data can be helpful in these cases, so the addition of the scan component to a SIM method can provide highly valuable quality control to SIM work that would otherwise be invisible to a SIM-only method. When do methods require Fast Scan and very fast SIM? Over-sampling chromatographic peaks will result in degraded spectral data in both scan and SIM modes but for very fast analysis such as with microbore capillary columns or hydrogen carrier gas, the fastest speed may be required as discussed below. Employing Rapid SIM The 5973 and 5975 series MSDs have always allowed SIM ion dwell times to be adjusted in 1-msec increments; however, the new 5975B MSD allows ion dwell times to be set as short as 1 msec. For fast chromatography, the combination of rapid scanning and very rapid SIM can be necessary. Figure 7 shows relatively narrow peaks of 600 msec (base width) acquired in SIM-Scan. Because scan cycling usually takes longer than SIM cycling, the enhanced flexibility of very rapid dwells, here 2 msec for the 10 ions in the group, allows very tight peaks to be quantitatively studied. The reader should immediately recognize that the scan peak is oversampled and know that two Edit SIM group ion dwell times to increase or reduce scans over peak Edit SIM group start times to prevent peak clipping Note: The minimum dwell time used in the Edit SIM Parameters is 5 msec. In the Instrument Control, the minimum dwell time is 1 msec. Figure 6. SIM Table editor. 5

6 opportunities exist: 1) expand the mass range being scanned with no change in the speed or 2) decrease the speed by a factor of two. To determine whether the SIM method should be similarly slowed, the lowest concentration standard should be injected. If six or more SIM cycles of each peak are obtained, quantitation by area will be good. To use peak heights, the rule of eight or more SIM cycles must be applied. In this way, all MS methods are refined. Enhanced SIM Data Analysis An annoyance of SIM work in the past is shown in Figure 8. In this case, an ion of interest is superimposed upon a rising baseline, here due to column bleed. Figure 9 shows the normalization of the SIM ions at the time of the beginning of the SIM ion group. To produce this, a separate data file is created in the data file directory: dataorig.ms. This file contains the non-normalized or original data while a new data.ms is written and automatically loaded from the data file. This adjustment aids visual evaluations of the SIM data but will not affect the results. This is a feature of the new MSD Productivity Software package G1701DA (D.03 revision). 10 scans of the peak for a mass range >350 m/z 600-msec wide peaks 10 SIM scans for 10 ions in the SIM group Figure 7. Applying rapid SIM and scan. 6

7 Figure 8. SIM ion offset due to baseline effects. Figure 9. SIM ion offset adjustment. 7

8 For More Information For more information on our products and services, visit our Web site at Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc Printed in the USA September 20, EN

Building Agilent GC/MSD Deconvolution Reporting Libraries for Any Application Technical Overview

Building Agilent GC/MSD Deconvolution Reporting Libraries for Any Application Technical Overview Building Agilent GC/MSD Deconvolution Reporting Libraries for Any Application Technical Overview Authors Xiaofei Ping Agilent Technologies (Shanghai) Co Ltd 412 Yinglun Road, Shanghai, 200131 P.R. China

More information

Setting up a Quantitative Analysis MS ChemStation

Setting up a Quantitative Analysis MS ChemStation Setting up a Quantitative Analysis MS ChemStation Getting Ready 1. Use the tutorial section "Quant Reports" on the MSD Reference Collection CD-ROM that came with your ChemStation software. 2. Know what

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

Background Information

Background Information 1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software

More information

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,

More information

Analysis of Liquid Samples on the Agilent GC-MS

Analysis of Liquid Samples on the Agilent GC-MS Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >

More information

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS

More information

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics

More information

Alignment and Preprocessing for Data Analysis

Alignment and Preprocessing for Data Analysis Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise

More information

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

C146-E087F. GCMS-QP2010 Plus. Shimadzu Gas Chromatograph Mass Spectrometer

C146-E087F. GCMS-QP2010 Plus. Shimadzu Gas Chromatograph Mass Spectrometer C146-E087F GCMS-QP2010 Plus Shimadzu Gas Chromatograph Mass Spectrometer The performance you have been waiting for is here. GCMS-QP2010 Plus GCMS-QP2010 Plus Shimadzu Gas Chromatograph Mass Spectrometer

More information

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

How To Use An Ionsonic Microscope

How To Use An Ionsonic Microscope Sample Analysis Design Element2 - Basic Software Concepts Scan Modes Magnetic Scan (BScan): the electric field is kept constant and the magnetic field is varied as a function of time the BScan is suitable

More information

Expectations for GC-MS Lab

Expectations for GC-MS Lab Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will

More information

The Use of Hydrogen Carrier Gas for GC/MS

The Use of Hydrogen Carrier Gas for GC/MS Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors

More information

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,

More information

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative

More information

Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012.

Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012. Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012 Introduction Screen captures in this document were taken from MassHunter B.05.00

More information

MassHunter for Agilent GC/MS & GC/MS/MS

MassHunter for Agilent GC/MS & GC/MS/MS MassHunter for Agilent GC/MS & GC/MS/MS Next Generation Data Analysis Software Presented by : Terry Harper GC/MS Product Specialist 1 Outline of Topics Topic 1: Introduction to MassHunter Topic 2: Data

More information

Tutorial 9: SWATH data analysis in Skyline

Tutorial 9: SWATH data analysis in Skyline Tutorial 9: SWATH data analysis in Skyline In this tutorial we will learn how to perform targeted post-acquisition analysis for protein identification and quantitation using a data-independent dataset

More information

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can

More information

Micromass LCT User s Guide

Micromass LCT User s Guide Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 33 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael

More information

Fractional Distillation and Gas Chromatography

Fractional Distillation and Gas Chromatography Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils

More information

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview

More information

Tuning & Mass Calibration

Tuning & Mass Calibration Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within

More information

ProteinPilot Report for ProteinPilot Software

ProteinPilot Report for ProteinPilot Software ProteinPilot Report for ProteinPilot Software Detailed Analysis of Protein Identification / Quantitation Results Automatically Sean L Seymour, Christie Hunter SCIEX, USA Pow erful mass spectrometers like

More information

Quick Start Guide Chromeleon 7.2

Quick Start Guide Chromeleon 7.2 Chromeleon 7.2 7229.0004 Revision 1.0 July 2013 Table of Contents 1 Introduction... 1 1.1 About this Document... 1 1.2 Other Documentation... 2 2 Using Chromeleon... 3 2.1 Overview... 3 2.2 Starting Chromeleon...

More information

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting

More information

Analysis of Phthalate Esters in Children's Toys Using GC-MS

Analysis of Phthalate Esters in Children's Toys Using GC-MS C146-E152 Analysis of Phthalate Esters in Children's Toys Using GC-MS GC/MS Technical Report No.4 Yuki Sakamoto, Katsuhiro Nakagawa, Haruhiko Miyagawa Abstract As of February 29, the US Consumer Product

More information

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan

More information

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH Cristiana C. Leandro 1, Peter Hancock 1, Christian Soulier 2, Françoise Aime 2 1 Waters Corporation, Manchester,

More information

Uses of Derivative Spectroscopy

Uses of Derivative Spectroscopy Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative

More information

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

SCION SQ Series GC-MS

SCION SQ Series GC-MS SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT AppNote 6/2003 Analysis of Flavors using a Mass Spectral Based Chemical Sensor Vanessa R. Kinton Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA Kevin L. Goodner USDA, Citrus & Subtropical

More information

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas

More information

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening Application Note Forensics Author Erin Shonsey Director of Research Alabama Department of Forensic Science Abstract

More information

23 The Thermal Conductivity Detector

23 The Thermal Conductivity Detector 23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing

More information

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment

More information

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ)

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ) Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) Closing Date and Time: Wednesday, April 13, 2016 at 5:00 p.m. (Pacific Time) Contact: Aaron Galloway Fax: 541-888-3250

More information

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005 CAC/GL 56-2005 Page 1 of 6 GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005 CONFIRMATORY TESTS When analyses are

More information

CHE334 Identification of an Unknown Compound By NMR/IR/MS

CHE334 Identification of an Unknown Compound By NMR/IR/MS CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared

More information

Storage Conditions of Sausages

Storage Conditions of Sausages Storage Conditions of Sausages The allows to quickly and easily control the storage conditions of food and beverages by a fingerprint / pattern analysis taken from the sample s headspace. Sausage samples

More information

Agilent Cary 60 UV-Vis

Agilent Cary 60 UV-Vis Agilent Cary 60 UV-Vis Efficient. Accurate. Flexible. Specifications Introduction The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is designed to meet both current

More information

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas

More information

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone

More information

Analyst 1.6 Software. Getting Started Guide

Analyst 1.6 Software. Getting Started Guide Analyst 1.6 Software Getting Started Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment. This

More information

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory

More information

Analyst 1.6 Software. Software Reference Guide

Analyst 1.6 Software. Software Reference Guide Analyst 1.6 Software Software Reference Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment.

More information

9.2.3.7 Retention Parameters in Column Chromatography

9.2.3.7 Retention Parameters in Column Chromatography 9.2.3.7 Retention Parameters in Column Chromatography Retention parameters may be measured in terms of chart distances or times, as well as mobile phase volumes; e.g., t R ' (time) is analogous to V R

More information

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird

More information

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Guaranteed specifications Design overview Double beam, ratio recording, double out-of-plane Littrow monochromator UV-Vis-NIR spectrophotometer (Agilent Cary

More information

Application of TargetView software in the fragrance industry: Accurate identi cation of allergens with fast GC

Application of TargetView software in the fragrance industry: Accurate identi cation of allergens with fast GC Application Note: ANTV12 Application of TargetView software in the fragrance industry: Accurate identi cation of allergens with fast GC Introduction The need to screen consumer products for speci c organic

More information

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP and TripleTOF 5600 System Technology Targeted peptide quantification is a rapidly growing application

More information

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies INEF Penn State Conference 2013 Albert Robbat, Jr. and Patrick Antle Tufts University, Chemistry

More information

A Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015

A Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015 A Navigation through the Tracefinder Software Structure and Workflow Options Frans Schoutsen Pesticide Symposium Prague 27 April 2015 Kings day in The Netherlands 1 Index Introduction Acquisition, Method

More information

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS

Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS Thorsten Bernsmann, Peter Fürst Chemical and Veterinary Analytical Institute Münsterland-Emscher-Lippe Joseph-König-Str. 40,

More information

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start. Q-TOF User s Booklet Enter your username and password to login. After you login, the data acquisition program will automatically start. ~ 1 ~ This data acquisition window will come up after the program

More information

Mass Frontier 7.0 Quick Start Guide

Mass Frontier 7.0 Quick Start Guide Mass Frontier 7.0 Quick Start Guide The topics in this guide briefly step you through key features of the Mass Frontier application. Editing a Structure Working with Spectral Trees Building a Library Predicting

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische

More information

INTRODUCTION. The principles of which are to:

INTRODUCTION. The principles of which are to: Taking the Pain Out of Chromatographic Peak Integration Shaun Quinn, 1 Peter Sauter, 1 Andreas Brunner, 1 Shawn Anderson, 2 Fraser McLeod 1 1 Dionex Corporation, Germering, Germany; 2 Dionex Corporation,

More information

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Introduction During the last decade, the complexity of samples

More information

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

HRMS in Clinical Research: from Targeted Quantification to Metabolomics A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier

More information

ForensicDB: A Web-accessible Spectral Database

ForensicDB: A Web-accessible Spectral Database ForensicDB: A Web-accessible Spectral Database Table of Contents 1. Introduction to ForensicDB... 1 2. Accessing ForensicDB... 1 3. Setting System Properties... 5 Spectrum Properties... 5 Frame Properties...

More information

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,

More information

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software Abstract Presented here is a case study of a walk-up liquid chromatography

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

The Theory of HPLC. Gradient HPLC

The Theory of HPLC. Gradient HPLC The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims

More information

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

Thermal Conductivity Detector

Thermal Conductivity Detector Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)

More information

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:

More information

INDUSTRIAL AUTOMATION Interactive Graphical SCADA System INSIGHT AND OVERVIEW. IGSS Online Training. Exercise 8: Creating Templates

INDUSTRIAL AUTOMATION Interactive Graphical SCADA System INSIGHT AND OVERVIEW. IGSS Online Training. Exercise 8: Creating Templates INDUSTRIAL AUTOMATION Interactive Graphical SCADA System INSIGHT AND OVERVIEW IGSS Online Training Exercise 8: Creating Templates Exercise: Create Templates and Template Based Objects Purpose Learn how

More information

RF Measurements Using a Modular Digitizer

RF Measurements Using a Modular Digitizer RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.

More information

TotalChrom. Chromatography Data Systems. Streamlining your laboratory workflow

TotalChrom. Chromatography Data Systems. Streamlining your laboratory workflow TotalChrom Chromatography Data Systems Streamlining your laboratory workflow maximize productivity with TotalChrom CDS Acquiring, processing, reporting, reviewing and approving data is a streamlined series

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1 GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

A4, Empower3 Processing Tips and Tricks

A4, Empower3 Processing Tips and Tricks A4, Empower3 Processing Tips and Tricks Rune Buhl Frederiksen, Manager, Waters Educational Services 2012 Waters Corporation 1 Content Basic Chromatography Workflow Processing Workflow Integration Theory

More information

Data Analysis Software

Data Analysis Software Data Analysis Software for VISION, BioCAD 700E, SPRINT, and INTEGRAL Workstations Version 3 Series Software Getting Started Guide DRAFT August 10, 2001 2:47 pm DASgsg_Title.fm Copyright 1998, 2001, Applied

More information