Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Size: px
Start display at page:

Download "Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference"

Transcription

1 Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas chromatography (GC) in terms of providing both qualitative and quantitative information - are well known. However, confident identification and accurate measurement of trace toxic and odorous species at the lowest levels of interest may be compromised by chromatographic anomalies such as; column bleed, extended solvent tails, air /water interference and unresolved sample matrix components. To address this issue, a new software algorithm has been developed for reprocessing stored GCMS data. It uses an innovative mathematical approach to distinguish and eliminate background interference from the mass ions contributing to real chromatographic peaks, even at the lowest levels. This should be of tangible benefit to any GCMS work involving detection and measurement of trace target compounds in complex uncharacterised samples environmental applications, homeland defence, forensic studies, food tests, materials emissions tests, flavour/fragrance profiling, odor monitoring, etc.. The performance and potential advantages of the new software to trace GCMS applications generally have been evaluated and are reported here. The examples presented include applications requiring trace enrichment carried out using thermal desorption (TD)-GCMS and high speed analyses carried out using fast- GC with TOFMS. Introduction Confident quantitative and qualitative analysis of trace compounds using GCMS can be compromised by chromatographic anomalies such as; column bleed, extended solvent tails, air /water interference and unresolved sample matrix components. A new GCMS data reprocessing algorithm has been developed which uses an innovative mathematical approach to distinguish and eliminate background interference from real chromatographic peaks, even at the lowest levels. It compensates dynamically for chromatographic background interference as it changes throughout a run without compromising any of the peak-related information. This dynamic baseline compensation (DBC) algorithm can be applied to total ion (TIC) or extracted ion (EIC) data and produces a separate DBC-reprocessed data file in addition to the original GCMS data file. An example of the software in operation is shown below (Figure 1.) Figure 1 shows the TD-GCMS analysis of a landfill gas standard carried out using automated thermal desorption with a thick-film, high-bleed capillary column

2 and quadrupole MS in full scan mode. The TIC trace is shown both before and after DBC-reprocessing. The dramatic impact on the chromatographic baseline is immediately apparent, but what the TIC data doesn t show is the equally dramatic impact on spectral purity and signal to noise. Figures 2 and 3 Figure 1: TD-GCMS analysis of a landfill-gas standard using a thick film capillary column with high bleed. Original TIC data file shown (black), DBC-reprocessed data file (blue). Apparatus: ULTRA-UNITY TD system (Markes International) with GCMS (Agilent Technologies),) and ClearView GCMS reprocessing software Original GCMS data S/N ~ 3:1 Figure 2: Automatic analysis of the original GCMS data file does not correctly identify the trace component at minutes because high level background ions interfere with the mass ion fragmentation (MIF) pattern ClearView reprocessed data S/N ~ 30:1 Figure 3: Automatic analysis of the reprocessed data file shows greatly enhanced spectral purity allowing automatic

3 On the face of it, this new automated approach to eliminating background interference could offer significant benefits to GCMS analyses both in terms of data quality and laboratory throughput. The purpose of this study was to evaluate the performance of this new algorithm with a range of challenging GCMS applications, in order to test its advantages and limitations under different conditions and assess how widely it could be applied. Examination of potential sensitivity enhancement As shown in the above example, selective removal of GCMS background ion contributions reduces the level of noise observed in full scan (TIC) data. The observed increase in signal to noise (S/N) ratio for the thiophene peak was from ~3:1 (RMS) in the original GCMS data file (figure 2) to ~30:1 in the reprocessed data file (figure 3). This represents a ten fold increase in sensitivity. Given the observed significant reduction in background interference, the new software should also facilitate calibration and quantification based upon a sum of characteristic ions rather than just relying on the intensity from a single extracted ion.. For reasons of selectivity, response factors for individual compounds are typically calculated based on a single target ion within the mass ion fragmentation (MIF) pattern of interest. The mass ion selected is typically only one of many generated during the ionisation process and may not even be the most significant ion in the fragmentation pattern (spectrum). This means that, in many cases, only a small proportion of the MIF is actually being used for quantitative analysis. If the risk of interference can be removed to negligible levels, the choice of mass ion selection for quantification could be expanded to encompass potentially the entire MIF profile, a range of ions, or a summation of specific mass ions. Examining the MIF pattern for the peak at minutes in the original data file (figure 2), it is clear that, due to the high background at this point in the TIC, quantification of thiophene would require single extracted ion analysis using parent mass ion 84. The number and range of interfering mass ions precludes use of a mass range for quantification in this case. In contrast to this, it is clear from the reprocessed data file (figure 3) that masses ranging from around 30 to 84 amu could be used for thiophene quantification after dynamic baseline compensation. In this actual example, the extracted ion profile for mass 84 has a peak area of , whereas using mass range for quantification gives an area of This represents nearly 3-fold increase in actual signal / peak area. Chromatographic needles in TIC haystacks The performance of the new GCMS reprocessing algorithm was evaluated for even lower level compounds using an example of trace contaminants in pure air. In this case 500 ml of clean outdoor air was sampled and analysed using on-line thermal desorption with GC-TOF MS. Original and DBC reprocessed data are

4 shown below (Figure 4) with a close-up of the area of the TIC around 12.6 minutes shown in Figure 5. Figure 4: 500 ml outdoor air sampled on-line. Original GCMS data (black). DBC data (blue) Original MS data MIF from original data Tiny peak at mins Spectrum after ClearView Reprocessed data Styrene Figure 5: Close-up of original and reprocessed data around 12.6 minutes. Insets: High mass ion interference in the original data file meant no identification was possible. Much lower mass ion interference in the reprocessed data allowed automatic library search identification of styrene. The high level of background interference in the original GCMS data file made it impossible to identify the trace component at 12.6 minutes without careful manual background subtraction. However, the enhanced spectral purity observed

5 for this peak in the DBC reprocessed data file allowed this component to be automatically identified as styrene without any manual intervention. Discussion of potential improvements in quantitative accuracy and repeatability In order to evaluate the performance of the new software with respect to quantitative accuracy and repeatability, it was presented with a particularly challenging GCMS application High speed analysis of diesel fuel (5 minute total run time) containing trace levels (3 pg on column) of octafluoronaphthalene (OFN). A Time-of-Flight (TOF) mass spectrometer (BenchTOF-dx from ALMSCO International.) was used in this case. Figure 6 shows the original GCMS data file overlaid with the reprocessed data file. Once again the impact of DBCreprocessing in terms of flattening baselines (this time by removing the hump of unresolved sample components) can clearly be seen. The process of flattening the baseline alone facilitates more accurate and repeatable automatic integration of resolved components. Figure 6: TIC analysis of trace levels (3 pg on column) of octafluoronapthalene (OFN) in diesel using high speed chromatography and Time-of- Flight (TOF) MS. 5 minute run time. Original GCMS data (black) shown overlaid with reprocessed data (blue) In order to assess any potential negative impact of DBC on the mass ions contributing to chromatographically resolved peaks, both the data files for this analysis (original and reprocessed) were first examined in extracted ion mode. While the trace OFN signal is buried within the diesel matrix in the TIC trace (figure 6) and cannot be detected, its presence and retention time can be confirmed by extracting the key mass 272 ion (Figure 7). Detailed close comparison of the OFN extracted ion signal from the original data file and the DBC-reprocessed file shows that the peak area data is identical. This indicates

6 that no mass data contributing to the chromatographic peak is lost during the DBC process and that full quantitative consistency with the original data is been maintained. Extracted Ion 272 Identical OFN Data Original GCMS data ClearView reprocessed data Figure 7: Diesel/OFN data in extracted ion mode (The match between the OFN 272 signals (original and reprocessed data) is so perfect that are shown inverted for clarity.) Such quantitative consistency cannot be claimed for conventional static background subtraction methods - as used by many commercial GCMS software systems to reduce background contribution to a TIC profile. The limitations of static background subtraction for ultra-trace level measurement are illustrated below in Figures 8 and 9. Original GCMS data DBC reprocessed data Figure 8: Overlaid extracted ion (mass 273) C 13 chromatograms (original and DBC reprocessed) from the high speed GC- BenchTOF analysis of diesel spiked with 3 pg OFN. Data offset by 5% along the x axis for clarity. Figure 8 show the extracted mass ion 273 chromatograms from the original and DBCreprocessed data files for the C 13 isomer of OFN. OFN molecules containing one C 13 atom in place of a C 12 atom are only ~11% of the abundance OFN molecules containing all C 12 so levels shown in this data are in the order of 300 femtograms. Despite working at these super-trace levels, close examination of the original and reprocessed data once again shows an identical response i.e. DBC reprocessing has not compromised the mass ion data contributing to the OFN peak area at minutes. However, later in the chromatogram (from about 3.6 minutes) the positive impact of DBC in terms of removing the contribution of background 273 mass ion present in the unresolved diesel matrix can be seen.

7 If, alternatively, a conventional manual, static mass ion fragmentation (MIF) background subtraction is performed at 4.2 minutes and applied to the original GCMS data file then the broad hump maximising at this time can be reduced to something similar to that seen in the DBC profile (Figure 9). However, the C 13 OFN peak at minutes completely disappears under these conditions highlighting the limitations of static subtraction. No OFN visible (red plot) after static subtraction of original data 3 pg OFN (C 13 ) post CV reprocessing Figure 9: Overlaid extracted ion (mass 273) C 13 chromatograms (DBC reprocessed (blue) and static background subtraction (red)) from the high-speed GC- TOF analysis of diesel spiked with 3 pg OFN. As above, data offset by 5% along the x axis for clarity. An investigation of the impact of DBC reprocessing on GCMS quantification is illustrated below (Figures 10 and 11). Figure 10 shows the DBC reprocessed data from three consecutive high speed GC-TOF analyses of OFN-spiked diesel injected manually. A region of interest is highlighted in the upper plot, and is shown expanded to full-scale below. Visual inspection of the overlaid data in close-up shows the excellent run-to-run reproducibility obtained using DBC. This degree of repeatability, makes it possible to reliably implement sophisticated, time-dependent integration events within a GCMS data analysis method for application to all the runs in a data set For example, note the integration event inserted within the TIC (see blue oval) for applying a free-hand baseline between 3.1 and 3.17 minutes such that peak areas within this time window can be summed. The area reproducibility obtained was 4% RSD remarkable for summed peaks and manual injection.

8 Figure 10: Three overlaid repeat high speed GC-BenchTOF MS analyses of OFN-spiked diesel, all reprocessed using DBC. Full run (top), expanded area (bottom). Enhanced repeatability and baseline assignment can be seen. Figure 11: Overlay of the original GCMS data from the series of 3 diesel/ofn analyses as shown in figure 10 but without DBC reprocessing. The decrease in performance with respect to repeatability and chromatographic resolution are obvious. If the same data analysis method (i.e. incorporating the integration event for assigning a baseline and summing peaks between 3.1 and 3.17 minutes) was applied to the original GCMS data obtained from this series of three runs (Figure 11) it had no effect In other words, no baseline was constructed on any of the 3 original GCMS data files. When a second attempt was made to implement a timed integration event using one of the original GCMS data files, a baseline was drawn for this one file. However, when this data analysis method was subsequently applied to the other two original GCMS data files in the set, y-axis displacement (differences) between the runs meant it could not be implemented and no baseline was drawn. There are cases, such as that shown for diesel above where the eliminated background ion contribution comprises an unresolved hump of material from the sample itself This is relatively common in GCMS for example during the analysis of hydrocarbon fuels (as shown) and in TD-GCMS analysis of chemical emissions from materials. In all such cases both the original and reprocessed data files would be required for a complete analysis:

9 The original data file would allow the total unresolved area to be determined e.g. for simulated distillation-type applications in the petroleum industry or TVOC measurements in materials emissions testing The reprocessed data file would allow enhanced quantitative and qualitative analysis of the resolved components. Impact on quantitative accuracy and repeatability when common ions are present? As an ultimate challenge to the quantitative performance of the new algorithm, the landfill gas standard data (Figure 1) was used to evaluate how well the software performed when trying to discriminate background ion contributions from chromatographic data when the peak and background contained one or more common ions. As an example, the peak at minutes (butanethiol) was examined. The spectrum of butanethiol includes a small contribution from mass ion 55 which was also present in the background. Figure 12 shows a spectrum obtained from the reprocessed data at the apex of the peak (which was confirmed as butanethiol by library search) and a spectrum from the background immediately after the peak. Both show a contribution from mass ion 55. * * Figure 12: Original and reprocessed data files for the landfill gas standards, highlighting butanethiol at ~18.1 minutes, shown together with the peak apex spectrum from the reprocessed data and a spectrum of the background signal immediately after the peak in the original GCMS data To evaluate whether the new software had correctly compensated for the mass ion 55 contribution due to the background and eliminated it from the reprocessed

10 data without impacting the contribution of mass ion 55 to the peak, the extracted ion 55 data (original and reprocessed) were overlaid with appropriate offset and alignment. (Figure 13). As mass 55 is a minor component of the butanethiol spectrum, any loss during reprocessing would be immediately apparent. However, the degree of correlation was remarkable, indicating that the new DBC algorithm appeared to compensate accurately for the background, without compromising chromatographic data, even when common ions were present. Figure 13: Extracted ion data for the original and reprocessed data. The overlaid trace was obtained by offsetting and aligning downwards the extracted ion response in the original data to match that in the reprocessed data Increasing the scanning mass range and reducing air/water interference Another key issue for trace TD-GCMS work can be air and water interference - Monitoring key ultra-volatile atmospheric pollutants such as freons and C 2 hydrocarbons requires use of super-strong, water retentive sorbents such as carbon molecular sieves which can retain significant masses of water and traces of carbon dioxide during environmental air monitoring studies. The impact of these interferences can usually be reduced by judicious use of dry purging both before and during thermal desorption-gcms analysis. However, traces may remain and cause baseline anomalies, particularly in the first part of the total ion chromatogram. In extreme cases, these perturbations can become significant and may compromise quantitative analysis when scanning below mass 45. However, when studying light compounds it is not always possible / advisable to exclude mass ions below 45. Low mass ions may comprise a major part of the spectrum of the light compounds of interest and their exclusion may compromise the match quality of some key target analytes. The ability of the new software to selectively eliminate this more unusual background interference was tested with an extreme example of air monitoring using strong, non-hydrophobic sorbent tubes, with insufficient dry purging and scanning from mass 15 (figure 14). The beneficial impact of dynamic baseline compensation is immediately apparent.

11 Figure 14: TD-GCMS of humid air sample scanned from mass 15. Original GCMS data file (black). DBC reprocessed data (blue) Summary This paper has illustrated the potential of the new dynamic baseline compensation algorithm to enhance the quality of GCMS data for trace components which is a critical issue for many applications. Improvements in spectral purity have been shown to facilitate more automated / higher quality identifications of trace components while, at the same time, the flatter baselines and reduced interference of the reprocessed data have been demonstrated to offer more sensitivity and more repeatable quantitative analysis without any loss of peak area data. It has also been shown to facilitate scanning from low masses and to apply to both conventional and high speed GCMS data. Over the range of examples tested it seems that the new software offers significant potential for improving the quality of GCMS data in terms of repeatability, sensitivity, ease-of-integration and spectral purity. By reducing the need for time-consuming manual re-integration of complex total ion chromatograms by skilled technicians, it also holds out the possibility of significant reduction in mean time / cost-per-analysis for GCMS applications requiring the detection and measurement of trace target analytes in uncharacterized and complex real-world sample matrices.

Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps

Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps Application Note: ANTV10 Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps Abstract In order to regulate product safety and quality

More information

Alignment and Preprocessing for Data Analysis

Alignment and Preprocessing for Data Analysis Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview

More information

Background Information

Background Information 1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 33 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

WHITE PAPER CONTINUOUS AIR MONITORING USING GRIFFIN 460 MOBILE GC/MS. HEADINGS: ALL CAPS, 12 PT TREBUCHET FONT, ICx BLUE

WHITE PAPER CONTINUOUS AIR MONITORING USING GRIFFIN 460 MOBILE GC/MS. HEADINGS: ALL CAPS, 12 PT TREBUCHET FONT, ICx BLUE WHITE PAPER CONTINUOUS AIR MONITORING USING GRIFFIN 460 MOBILE GC/MS HEADINGS: ALL CAPS, 12 PT TREBUCHET FONT, ICx BLUE Body text is 11 pt Cynthia Georgia Liu, with 1.5 Ph.D., line spacing. Paul Dreier

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

How To Test For Contamination In Large Volume Water

How To Test For Contamination In Large Volume Water Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed

More information

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,

More information

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting

More information

Expectations for GC-MS Lab

Expectations for GC-MS Lab Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will

More information

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies INEF Penn State Conference 2013 Albert Robbat, Jr. and Patrick Antle Tufts University, Chemistry

More information

Analysis of Liquid Samples on the Agilent GC-MS

Analysis of Liquid Samples on the Agilent GC-MS Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >

More information

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1 GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,

More information

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT AppNote 6/2003 Analysis of Flavors using a Mass Spectral Based Chemical Sensor Vanessa R. Kinton Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA Kevin L. Goodner USDA, Citrus & Subtropical

More information

How To Use Gc-Ms

How To Use Gc-Ms The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter

More information

Setting up a Quantitative Analysis MS ChemStation

Setting up a Quantitative Analysis MS ChemStation Setting up a Quantitative Analysis MS ChemStation Getting Ready 1. Use the tutorial section "Quant Reports" on the MSD Reference Collection CD-ROM that came with your ChemStation software. 2. Know what

More information

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D. Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class

More information

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative

More information

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS

More information

Series 2. Continuous monitoring of trace chemical vapours

Series 2. Continuous monitoring of trace chemical vapours Series 2 TT24-7 Continuous monitoring of trace chemical vapours Series 2 TT24-7 TT24-7 Series 2 Continuous monitoring of trace chemical vapours The TT24-7 Series 2 is an advanced thermal desorption (TD)

More information

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

SCION SQ Series GC-MS

SCION SQ Series GC-MS SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

MassHunter for Agilent GC/MS & GC/MS/MS

MassHunter for Agilent GC/MS & GC/MS/MS MassHunter for Agilent GC/MS & GC/MS/MS Next Generation Data Analysis Software Presented by : Terry Harper GC/MS Product Specialist 1 Outline of Topics Topic 1: Introduction to MassHunter Topic 2: Data

More information

Waters Core Chromatography Training (2 Days)

Waters Core Chromatography Training (2 Days) 2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.

More information

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can

More information

Application of TargetView software in the fragrance industry: Accurate identi cation of allergens with fast GC

Application of TargetView software in the fragrance industry: Accurate identi cation of allergens with fast GC Application Note: ANTV12 Application of TargetView software in the fragrance industry: Accurate identi cation of allergens with fast GC Introduction The need to screen consumer products for speci c organic

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES

Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES AGILENT ENHANCED MATRIX REMOVAL LIPID ARE LIPIDS WEIGHING YOUR SAMPLES DOWN? Interference from lipids is a problem for labs measuring

More information

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers

More information

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment

More information

A Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015

A Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015 A Navigation through the Tracefinder Software Structure and Workflow Options Frans Schoutsen Pesticide Symposium Prague 27 April 2015 Kings day in The Netherlands 1 Index Introduction Acquisition, Method

More information

Indoor Air Comfort Test Report

Indoor Air Comfort Test Report Knauf Insulation SPRL Rue de Maestricht, 95 B-4600 Vise Belgium Indoor Air Comfort Test Report 1 Sample Information Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing

More information

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Application of Structure-Based LC/MS Database Management for Forensic Analysis Cozette M. Cuppett and Michael P. Balogh Waters Corporation, Milford, MA USA Antony Williams, Vitaly Lashin and Ilya Troisky Advanced Chemistry Development, Toronto, ntario, Canada verview Application

More information

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform Remco Swart Pittcon March 8-12, 2015 1 The world leader in serving science Outline Thermo Scientific Vanquish UHPLC

More information

Analysis of Phthalate Esters in Children's Toys Using GC-MS

Analysis of Phthalate Esters in Children's Toys Using GC-MS C146-E152 Analysis of Phthalate Esters in Children's Toys Using GC-MS GC/MS Technical Report No.4 Yuki Sakamoto, Katsuhiro Nakagawa, Haruhiko Miyagawa Abstract As of February 29, the US Consumer Product

More information

Quick Start Guide Chromeleon 7.2

Quick Start Guide Chromeleon 7.2 Chromeleon 7.2 7229.0004 Revision 1.0 July 2013 Table of Contents 1 Introduction... 1 1.1 About this Document... 1 1.2 Other Documentation... 2 2 Using Chromeleon... 3 2.1 Overview... 3 2.2 Starting Chromeleon...

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

Determination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS

Determination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS Application Note No. 051 Determination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS by P Jüngel ¹), J Wittenzellner ²) and E Melcher ¹) ¹) Federal Research Centre

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics

More information

Analyzing Small Molecules by EI and GC-MS. July 2014

Analyzing Small Molecules by EI and GC-MS. July 2014 Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,

More information

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Carlton G. Slough TA Instruments, 109 Lukens Drive, New Castle DE 19720, USA ABSTRACT The use of

More information

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler

More information

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration ToF Mass Calibration Essentially, the relationship between the mass m of an ion and the time taken for the ion of a given charge to travel a fixed distance is quadratic in the flight time t. For an ideal

More information

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,

More information

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley

More information

MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis

MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis Overview MarkerView software is a novel program designed for metabolomics applications and biomarker profiling workflows 1. Using

More information

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird

More information

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,

More information

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas

More information

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,

More information

How To Analyze Pesticide Levels In Lemon Oil

How To Analyze Pesticide Levels In Lemon Oil AppNote 9/2006 Target GC-MS Analysis using Accelerated Column Heating and Interactive Deconvolution Software Albert Robbat Tufts University, Chemistry Department, Medford, MA 02155, USA Andreas Hoffmann

More information

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex

More information

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT

More information

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische

More information

AxION edoor. Web-Based, Open-Access Mass Spectrometry Software

AxION edoor. Web-Based, Open-Access Mass Spectrometry Software AxION edoor Web-Based, Open-Access Mass Spectrometry Software 2 AxION edoor is the open-access software solution for today s fast-paced laboratory. It is designed to operate and optimize the management

More information

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** mariadellarcopassaro@gmail.com

More information

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP and TripleTOF 5600 System Technology Targeted peptide quantification is a rapidly growing application

More information

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note. EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in

More information

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan

More information

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening Application Note Forensics Author Erin Shonsey Director of Research Alabama Department of Forensic Science Abstract

More information

Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD

Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai) Co., Ltd. 412 Ying Lun

More information

Process Instrumentation

Process Instrumentation Process Instrumentation Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Process Gas Chromatography Features of Process Chromatographs Agilent 7890A

More information

Thermo Scientific ExactFinder Software

Thermo Scientific ExactFinder Software mass spectrometry Thermo Scientific ExactFinder Software Workflow software for routine targeted and general unknown screening Unifies qualitative and quantitative high-resolution accurate mass workflows

More information

Mass Spectrometry. Overview

Mass Spectrometry. Overview Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2

More information

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH Cristiana C. Leandro 1, Peter Hancock 1, Christian Soulier 2, Françoise Aime 2 1 Waters Corporation, Manchester,

More information

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines

More information

Tuning & Mass Calibration

Tuning & Mass Calibration Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within

More information

Integrated Data Mining Strategy for Effective Metabolomic Data Analysis

Integrated Data Mining Strategy for Effective Metabolomic Data Analysis The First International Symposium on Optimization and Systems Biology (OSB 07) Beijing, China, August 8 10, 2007 Copyright 2007 ORSC & APORC pp. 45 51 Integrated Data Mining Strategy for Effective Metabolomic

More information

Tracer Detection Systems

Tracer Detection Systems Tracer Detection Systems WASSON ECE Engineered Solutions, Guaranteed Results. Wasson-ECE Tracer Analysis System The industry standard for low level perfluoronated tracer analysis. Tracer tagging technology

More information

Introduction. What Can an Offline Desktop Processing Tool Provide for a Chemist?

Introduction. What Can an Offline Desktop Processing Tool Provide for a Chemist? Increasing Chemist Productivity in an Open-Access Environment Ryan Sasaki, Graham A. McGibbon, Steve Hayward Featuring ACD/Spectrus Processor and Aldrich Library for ACD/Labs Advanced Chemistry Development,

More information

How To Build An Amd Quas 3 Ar Mass Spectrometer

How To Build An Amd Quas 3 Ar Mass Spectrometer Classical and special AMD Magnetic Sector mass spectrometers (extracted from power point presentations) High Resolution Mass Spectromter AMD 402/403 S 6 High Resolution Mass Spectrometer AMD 604 S 8 Version

More information

LC-MS/MS for Chromatographers

LC-MS/MS for Chromatographers LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,

More information

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples

More information

New Jersey Department of Environmental Protection Division of Remediation Management Response Site Remediation Program

New Jersey Department of Environmental Protection Division of Remediation Management Response Site Remediation Program New Jersey Department of Environmental Protection Division of Remediation Management Response Site Remediation Program NJDEP- SRP Low Level USEPA TO-15 Method (NJDEP-LLTO-15-3/2007) March 2007 March 2009

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,

More information

A commitment to quality and continuous improvement

A commitment to quality and continuous improvement Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous

More information

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10

More information

Multi-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible

Multi-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible Multi-Shot Pyrolyzer EGA/PY-3030D Flexible Versatile Reproducible Why Evolved Gas Analysis? Why Pyrolysis? Today, analytical pyrolysis encompasses much more than simple flash pyrolysis of polymeric materials.

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

1 Genzyme Corp., Framingham, MA, 2 Positive Probability Ltd, Isleham, U.K.

1 Genzyme Corp., Framingham, MA, 2 Positive Probability Ltd, Isleham, U.K. Overview Fast and Quantitative Analysis of Data for Investigating the Heterogeneity of Intact Glycoproteins by ESI-MS Kate Zhang 1, Robert Alecio 2, Stuart Ray 2, John Thomas 1 and Tony Ferrige 2. 1 Genzyme

More information