ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC)
|
|
- Margery Harrington
- 8 years ago
- Views:
Transcription
1 ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC) 1
2 Fast Analysis of C1 to C8 hydrocarbons for mud logging applications within 2 minutes using a temperature programmable Micro GC Fusion NEN GAS2015 Debbie Alcorn INFICON, Inc. 2 Technology Place East Syracuse, NY 13057, USA 2
3 Introduction Mud Logging Introduction Gas Sampling at the Well Site Gas Chromatography (GC) for Hydrocarbon Analysis Instrumentation Instrument Features Methods and Data C1-C5 in 34 seconds C1-C8 in 2 minutes On-Site Tips for Successful Analysis Future Development 3
4 Mud Logging Introduction Mud logging is the creation of reports of a borehole during active drilling A mud log provides geographical information, including (but not limited to): Lithology Rate of penetration Hydrocarbon content The concentration of C1-C8 compounds is critical Analysis must be performed on a continuous, 24-7 basis 4
5 Gas Sampling at the Well Site Mud is used to bring gas, formation fluids, and rock cuttings to the surface for analysis A gas trap separates the gas and liquid components The head space gas goes to the gas analyzer Gas chromatography (GC) is often used to analyze the gas sample 5
6 Gas Chromatography for Hydrocarbon Analysis GC has been used for mud logging for decades to characterize reservoir fluid composition Each reservoir is composed of different amounts of hydrocarbon species The ratios of compounds help identify zones of producible oil or gas 1 6
7 Gas Chromatography for Hydrocarbon Analysis Other calculations are used for wetness (W h ), balance (B h ), and character (C h ) 2 Example Interpretations: W h = C 2 + C 3 + ic 4 + nc 4 + C 5 C 1 + C C B h = C 1 + C 2 C 3 + ic 4 + nc 4 + C 5 C h = ic 4 + nc 4 + C 5 C 3 r o = gas density 7
8 Gas Chromatography for Hydrocarbon Analysis C1-C5 compounds are of main importance Analysis of heavier hydrocarbons is often desired Methylcyclohexane and C7 help indicate the presence of liquid phases 8
9 Instrumentation Small, transportable GC with MicroElectroMechanical Systems (MEMS) technology Successor to the 3000 Micro GC Gas only analysis MEMS Thermal Conductivity Detector (µtcd) Micro GC Fusion 9
10 Instrument Features Temperature programmable columns Minimizes carryover More compounds elute on a single column Increases sample throughput Resistively heated fused silica capillary columns 10
11 Instrument Features Micro GC utilizes a modular architecture Each module is its own GC: MEMs based injector Resistively heated fused silica capillary columns MEMs based µtcd 11
12 Instrument Features Micro GC Fusion utilizes a new User Interface (UI) Front panel display Direct Ethernet connection to a computer Direct Ethernet connection to a LAN hub for remote access Wi-Fi enabled License free Works with any platform (ex. Windows, Apple, Android) 12
13 Instrument Features Integrated sample conditioner (optional) Reduces sample pressures from up to 1000 psi down to ~ 5 psi (~34 kpa) Filters out particulates Heats the sample to 100 ºC Maintains sample temperature to prevent condensation 13
14 Isothermal Method C1-C5 2 module system Module A 8m RT-Q-Bond, variable volume injector Module B 10m Rxi-1ms, variable volume injector A method was designed to run continuously, using isothermal operation Total cycle time was 34 seconds 14
15 Calibration Standard Component Amount (%) Nitrogen Methane Ethane Propane ic nc ic nc
16 Chromatograms Module A: RT-Q-Bond Initial parameters used: 55 C, 37 psi (255 kpa), 15 ms injection time, 5 second pump, Helium carrier gas 16
17 Chromatograms Module B: Rxi-1ms Initial parameters used: 90 C, 33 psi (227 kpa), 15 ms injection time, 5 second pump, Helium carrier gas 17
18 Repeatability Exceptional repeatability Less than 0.1 %RSD for retention time Less than 0.6 %RSD for area 1/29/2014 Last 10 runs (10:09 a.m. to 10:16 a.m.) Area %RSD RT %RSD Nitrogen Methane Ethane Propane ic nc ic nc
19 10 Runs Overlaid RT-Q-Bond 19
20 10 Runs Overlaid Rxi-1ms 20
21 500 ppm Methane Module A: RT-Q-Bond - 37 psi (255 kpa) and 32 psi (220 kpa) Syringe dilution Reported concentration 0.051% 21
22 100 ppm Methane Module A: RT-Q-Bond - 37 psi (255 kpa) and 32 psi (220 kpa) Syringe dilution Reported concentration 0.014% 22
23 Temperature Ramped Method C1-C8 2 module system Module A 8m RT-Q-Bond, variable volume injector Module B 10m Rxi-1ms, variable volume injector A method was designed to run continuously, using temperature programming Total cycle time was approximately 2 minutes 23
24 Extended Calibration Standard Component Amount (%) Nitrogen Methane 0.5 CO Ethylene Ethane 0.2 Propylene Propane ic nc ic nc Component Amount (%) nc Benzene Cyclohexane nc Methylcyclohexane 0.01 Toluene nc
25 Chromatograms Module A: RT-Q-Bond 55 C(15s hold) 220 C(10s hold) 7 C/s, 35 psi (241 kpa), 30 ms injection time, Helium carrier gas 25
26 Chromatograms Module B: Rxi-1ms 75 C(9s hold) 220 C(12s hold) 7 C/s, 33 psi (227 kpa), 80ms injection time, Helium carrier gas 26
27 On-Site Tips for Successful Analysis Power Access to stable power Surge protector, UPS Heated transfer lines To keep heavier hydrocarbons from condensing Gas trap stability Water traps, filters 27
28 Challenges Method was optimized for speed Aggressive ramp Not optimized for instrument stability Carryover effect Without a backflush injector, heavier compounds can stick to the columns Peaks elute during subsequent runs 28
29 Future Method Developments Future method developments will include: Backflush injectors to prevent carryover Different combination of columns types/lengths Faster analysis with a less aggressive ramp Analysis of ethylbenzene, nc9 with the same cycle time 29
30 Conclusion Using isothermal operation, C1-C5 compounds can be analyzed within 34 seconds with excellent repeatability Using temperature programming, C1-C8 compounds can be analyzed within 2 minute cycle time Additional compounds, such as CO 2, ethylene, and propylene were also analyzed Future method developments will focus on: Optimizing the configuration of Micro GC Fusion Optimizing the corresponding methods 30
31 References 1. B.O. Pixler, Journal of Petroleum Technology, Vol 21, Number 6, 1969, pag J.H. Haworth, Interpretation of Hydrocarbon Shows Using Light (C1-C5) Hydrocarbon Gases from Mud-Log Data, The American Association of Petroleum Geologist Bulletin, Vol 69, Number 8, 1985, pag P. Ablard, D. Cook, K. Fielding, G. Haines, M. Herkommer, K. McCarthy, M. Radakovic, L. Umar, The Expanding Role of Mud Logging, Oilfield Review, Vol 24, Number 1, 2012, pag SPE International. Formation Evaluation During Mud logging. (accessed March ). 31
32 For more information, visit: Or Booth #7 32
33 Questions? 33
BTU ANALYSIS USING A GAS CHROMATOGRAPH
BTU ANALYSIS USING A GAS CHROMATOGRAPH By Joseph E. Landes Southern Petroleum Laboratories, Inc. Technical Director Hydrocarbon and Field Services 8820 Interchange Drive Houston, Texas 77054 Introduction
More informationChemical Analysis. Natural Gas Analyzer. A Family of optimized GC Solutions. Gas Chromatography. think forward
Chemical Analysis Natural Gas Analyzer A Family of optimized GC Solutions think forward Gas Chromatography Bruker Natural Gas Analyzers Natural gas is bought and sold as a bulk commodity with price based
More informationChromatograph Maintenance and Troubleshooting Class # 5050
Chromatograph Maintenance and Troubleshooting Class # 5050 Shane Hale Business Development Manager Emerson Process Management Gas Chromatographs Division 5650 Brittmoore Drive Houston, TX 77041 USA Introduction
More information1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
More informationEPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in
More informationPilot Gasifier Producer Gas Analysis and Measurement Arrangements in. University of Applied Sciences
Pilot Gasifier Producer Gas Analysis and Measurement Arrangements in Sievi Laboratory Kari Pieniniemi, Researcher, Centria University of Applied Sciences 1 2 MEASURING LINE AT SIEVI LABORATORY 3 MEASURING
More informationMicro GC Fusion Pre-Installation Checklist
Site Preparation Proper site preparation prior to installing is very important. Contact INFICON (www.inficon.com) with any questions regarding site preparation. NOTE: Checkboxes ( ) identify hardware that
More informationOn-line Dissolved Gas Analysis
On-line Dissolved Gas Analysis accord. to IEC 567/ASTM 3612 Online-Monitoring of transformers Automatic on-line analysis of 11 gases The analysis is done in 2 steps with high selectivity and accuracy fully
More informationTransformer Oil Gas Analysis
Get on the Analyzer FAST TRACK Transformer Oil Gas Analysis Helping customers protect the environment! 1 Why Analyzers Value to Customers Alleviates Resource Pressure Facilitates Configuration Use Advanced
More informationPLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative
PLOT Columns Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative PLOT Column Selection 93 Features & Benefits of Restek PLOT Columns Features
More informationGC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
More information23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
More informationGAS WELL/WATER WELL SUBSURFACE CONTAMINATION
GAS WELL/WATER WELL SUBSURFACE CONTAMINATION Rick Railsback Professional Geoscientist CURA Environmental & Emergency Services rick@curaes.com And ye shall know the truth and the truth shall make you free.
More informationGC/MS Considerations for Fracking Gases and Water. Dr. Harold McNair Virginia Tech Blacksburg, VA. hmcnair@vt.edu
GC/MS Considerations for Fracking Gases and Water Dr. Harold McNair Virginia Tech Blacksburg, VA. hmcnair@vt.edu 1 Outline What is fracking? Energy, Economics, Environmental Factors Chromatographic Considerations
More informationGet the Right GC for Your Industry
Get the Right GC for Your Industry Shimadzu was founded on the basis of "Technology for Science" and we continue to take the initiative in the analytical technology industry. From our absorption spectrophotometers
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationFlairFlex. Real-time fluid logging and analysis service
FlairFlex Real-time fluid logging and analysis service FlairFlex Real-time fluid logging and analysis service The FlairFlex* service from Geoservices provides early insight into reservoir fluid properties
More informationN O T E S. Environmental Forensics. Identification of Natural Gas Sources using Geochemical Forensic Tools. Dispute Scenarios
Environmental Forensics N O T E S V o l u m e 2 9 Identification of Natural Gas Sources using Geochemical Forensic Tools By Paul Boehm, Ph.D. and Tarek Saba, Ph.D. F o r m o r e i n f o r m a t i o n o
More informationCHROMOCTANE O N LINE ANALYZER
GECIL Process GT instruments CHROMOCTANE O N LINE ANALYZER ASTM D6733 DHA and RON and MON on line GC for Reformer unit Automatic Chromatographic system for analysis of naphta feeds and reforming gasolines
More informationTHERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles
THERMAL DESORPTION Introduction and Principles Thermal Desorption Thermal desorption is a simple extension of the technique of gas chromatography (GC) It involves the use of heat and a flow of inert gas
More informationProcess Instrumentation
Process Instrumentation Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Process Gas Chromatography Features of Process Chromatographs Agilent 7890A
More informationAn Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.
An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at
More informationGulf Coast Conference Prepared by: A. Coderre (Clearstone Engineering Ltd)
APPLICATIONS OF FAST TRANSPORTABLE CHROMATOGRAPHY Gulf Coast Conference Prepared by: A. Coderre (Clearstone Engineering Ltd) Introduction: Adam Coderre M.A.Sc., Mechanical Engineering Specializing in emissions
More informationGas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
More informationFundamentals of Gas Chromatography
Application Note Oil & Gas Fundamentals of Gas Chromatography Figure 1 - The Function Components of a Gas Chromatograph Process Line Sample System Chromatograph Oven GC Controller Probe Sample Return Slip
More informationOn-line Ozone Precursor Analyzers From the Leader in Thermal Desorption Technology
On-line Ozone Precursor Analyzers From the Leader in Thermal Desorption Technology GAS CHROMATOGRAPHY P R O D U C T N O T E PerkinElmer On-line Ozone Precursor System, comprising of a Clarus Gas Chromatograph
More informationAC DHA Solutions. Detailed Hydrocarbon Analysis of Light Petroleum Streams & Light End in Crude Oils. ASTM D 5134, D 6729, D 6730, D 6733 Compliance
AC DHA Solutions Detailed Hydrocarbon Analysis of Light Petroleum Streams & Light End in Crude Oils ASTM D 5134, D 6729, D 6730, D 6733 Compliance Signal Alignment Automated with LineUp TM Tool Compatible
More informationACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
More informationBackground Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
More informationA VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water
A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water Sjaak de Koning 1, Mark van Lieshout 2, Hans-Gerd Janssen 2 and Wil van Egmond 1 1 ATAS, De Run 4334, PO Box
More informationTABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ACKNOWLEDGEMENT ABSTRACT LIST OF TABLES LIST OF FIGURES LIST OF SYMBOLS LIST OF APPENDICES
vii TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ii ACKNOWLEDGEMENT iv ABSTRACT v TABLE OF CONTENTS vii LIST OF TABLES xii LIST OF FIGURES xiv LIST OF SYMBOLS xvii LIST OF APPENDICES xviii 1 INTRODUCTION
More informationTopics: The Petroleum System. Origin of Petroleum. Arguments for biogenic origin. Google search : Origin of Petroleum : Hugoton Gas Field (Kansas)
The Petroleum System Topics: Origin of Petroleum Controversy Carbon Cycle Global Implications Petroleum System Gas Oil Water Origin of Petroleum Long standing debate: Where does rock oil come from? Organic
More informationShimadzu Simulated Distillation Gas Chromatograph System C184-E030
Shimadzu Simulated Distillation Gas Chromatograph System C184-E030 Meets All Simulated Distillation Gas Chromatograph Standards - Provides Highly Accurate Analysis Results with Excellent Reproducibility-
More informationCHAPTER 7 THE DEHYDRATION AND SWEETENING OF NATURAL GAS
CHAPTER 7 THE DEHYDRATION AND SWEETENING OF NATURAL GAS Natural gases either from natural production or storage reservoirs contain water, which condense and form solid gas hydrates to block pipeline flow
More information83 to 87% carbon 11-15% hydrogen 0.1-7% sulphur 0.06-1.5% oxygen 0.1-0.5% nitrogen
Petroleum chemistry rude is the name for a liquid mixture of different hydrocarbon connections found in the earth. Petroleum, strictly speaking is crude oil, but it is used to describe all kinds of hydrocarbons.
More informationMLogger Series Logging Instruments
MLogger Series Logging Instruments Digital Data Acquisition and Logging Systems Model TG Total Gas Detector Model TGC Total Gas and Chromatograph Calibrated Logging Instruments Model TG Total Hydrocarbon
More informationDetermination of Caffeine in Beverages HPLC-1
Determination of in s HPLC-1 Introduction This experiment provides an introduction to the application of High Performance Liquid Chromatography (HPLC) to the solution of complex analytical problems. Cola
More informationFractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
More informationAN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler
AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY Andrew Tipler Chromatography Research and Technology Manager PerkinElmer, Inc. Table of Contents Introduction 3 Fundamental
More informationPurge and trap techniques using EPA 8260B will be used to determine the following:
The BP Gulf Oil Spill: Analysis of BTEX in Mississippi Canyon 252 Crude Oil Using Purge and Trap GC MS with a Unique Cyanopropylphenyl Stationary Phase Jack Cochran, Michelle Misselwitz, Chris English,
More informationß-CYCLODEXTRIN SYNONYMS
ß-CYCLODEXTRIN Prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995) superseding specifications prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications
More informationEnAlgae SOP: Biogas analysis
Table of content 1. PURPOSE... 2 2. PRINCIPLE... 2 3. REQUIREMENTS... 3 3.1. EQUIPMENT AND MATERIALS... 3 3.2. REAGENTS... 3 3.3. PREPARATION OF MEASUREMENT... 3 4. PROCEDURE... 4 4.1. SAMPLE MEASUREMENT...
More informationMicro Cryo-Trap Model 971
Micro Cryo-Trap Model 971 for Liquid CO2 Cooling Scientific Instrument Services, Inc. 1027 Old York Road Ringoes, NJ 08551 Phone (908) 788-5550 Part #951001M revised 10/ /98 Notice The information in this
More informationFiscal Measurement Natural Gas
White Paper FSG-WP-0012 February 2014 Fiscal Measurement Natural Gas Fiscal Measurement How much? Fiscal Measurement must not be confused with Custody Transfer; in fact, fiscal measurement is a more general
More informationLecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder
Lecture Chromo-3: Gas Chromatography CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Outline Introduction Instrument overview Carrier gas Sample
More informationFerda ORHUN Mineral Research and Exploration Institute of Turkey
GAS CHROMATOGRAPHY, EXPERIMENTAL STUDIES ON THE FLOW RATE EFFECTS OF CARRIER GAS AND APPLICATION OF THE METHOD TO HYDROCARBON ANALYSIS OF SOME NATURAL GASES IN TURKEY Ferda ORHUN Mineral Research and Exploration
More informationTitan: The Solar System s Abiotic Petroleum Factory
Titan: The Solar System s Abiotic Petroleum Factory J. Hunter Waite, Ph.D. Institute Scientist Space Science & Engineering Division Southwest Research Institute Titan: The Solar System s Abiotic Petroleum
More informationCRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT
CRYOCOOLER CONTAMINATION STUDY S.W.K. Yuan, D.T.Kuo, and A.S. Loc EI Technologies, Cryocooler Group Sylmar, CA 91342 ASTRACT Eliminating or reduction of contamination is essential for the success of cryocoolers
More informationThe Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
More informationSYNSPEC GC s ALPHA 115 and ALPHA 116
SYNSPEC GC s ALPHA 115 and ALPHA 116 Methane/TotalNonMethaneHydroCarbons Operator manual Version January 2010 Synspec b.v. De Deimten 1 9747 AV Groningen Nederland info@synspec.nl CONTENTS CONTENTS...
More informationHow To Run A Gas Plant
MATHESON Petrochemical Industry Program for Specialty Gas Supply MATHESON Petrochemical Industry Program for Specialty Gas Supply MATHESON is a single source for industrial, medical, specialty, and electronics
More informationTracer Detection Systems
Tracer Detection Systems WASSON ECE Engineered Solutions, Guaranteed Results. Wasson-ECE Tracer Analysis System The industry standard for low level perfluoronated tracer analysis. Tracer tagging technology
More informationExpectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
More informationAgilent 7697A Headspace Sampler Data Sheet
Agilent 7697A Headspace Sampler Data Sheet Overview The Agilent 7697A Headspace Sampler is part of Agilent s line of gas chromtography products. The 7697A based on the architecture of the market-leading
More informationUnconventional Challenges: Integrated Analysis for Unconventional Resource Development Robert Gales VP Resource Development
Unconventional Challenges: Integrated Analysis for Unconventional Resource Development Robert Gales VP Resource Development Opening Remarks The Obvious Fossil fuels will be the main energy supply, accounting
More informationFORMATION DAMAGE AND WELL TREATMENT SYSTEM
FORMATION DAMAGE AND WELL TREATMENT SYSTEM FDWTS-10K Includes ability to test Liquid Permeability & Gas Permeability Fines migration & critical velocity Static Filtration Drilling mud invasion evaluation
More informationSCION SQ Series GC-MS
SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative
More informationF321 MOLES. Example If 1 atom has a mass of 1.241 x 10-23 g 1 mole of atoms will have a mass of 1.241 x 10-23 g x 6.02 x 10 23 = 7.
Moles 1 MOLES The mole the standard unit of amount of a substance (mol) the number of particles in a mole is known as Avogadro s constant (N A ) Avogadro s constant has a value of 6.02 x 10 23 mol -1.
More informationApplication Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
More informationDetermination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*
Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU
More informationOil and Gas Terms. Anticline: An arch of stratified rock layers that may form a trap for hydrocarbons.
Oil and Gas Terms American Petroleum Institute (API): The API is the trade organization for the oil and gas industry, which establishes standards governing industry operations, safety and the manufacturing
More informationNatural Gas Processing Solutions Using the 700XA Gas Chromatograph
Application Note Oil & Gas Natural Gas Processing Solutions Using the 700XA Gas Chromatograph Optimizing the efficiency and profitability of a natural gas processing plant requires responding to the changing
More informationPennsylvania Marcellus Shale Short- Term Ambient Air Sampling Project
Pennsylvania Marcellus Shale Short- Term Ambient Air Sampling Project Marcellus Shale Conference November 16, 2011 Nick Lazor Chief, Air Quality Monitoring Division Bureau of Air Quality PA Department
More information1. Standard conditions are 29.92 in. Hg (760 mm Hg, 14.696 psia) and 68 F (20 C).
INTRODUCTION Emission Monitoring Inc. DETERMINING F-FACTORS FROM GAS CHROMATOGRAPHIC ANALYSES Roger T. Shigehara Emission Monitoring Incorporated Howard F. Schiff TRC Environmental Corporation EPA Method
More informationGas Separations: A Comparison of GasPro and Aluminum Oxide PLOT Columns for the Separation of Highly Volatile Compounds
Gas Separations: Comparison of GasPro and luminum Oxide PLOT Columns for the Separation of Highly Volatile Compounds Daniel W. rmstrong* Department of Chemistry, University of Missouri-Rolla, Rolla, MO
More informationFast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD
Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai) Co., Ltd. 412 Ying Lun
More informationMATHESON REFINERY PROGRAM for Specialty Gas Supply. Experience the MATHESON Commitment to Supply Chain Excellence
MATHESON REFINERY PROGRAM for Specialty Gas Supply MATHESON REFINERY PROGRAM for Specialty Gas Supply MATHESON is a single source for industrial, medical, specialty, and electronics gases as well as gas
More informationHolcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants
Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants Version 2004-01 / August 2004 (HGRS-JW-04-25s) HGRS-CTS/MT J. Waltisberg SUMMARY ( ) shall not
More informationGC-xt. The most flexible GC /GC-MS liquid sample injector
GC-xt Prep and Load Platform The most flexible GC /GC-MS liquid sample injector Environmental Foods/Beverages Consumer Products Forensics Petrochemicals/Polymers Pharmaceuticals GC-xt Liquid mode Prep
More informationHEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.
HEXANES Prepared at the 51st JECFA (1998), published in FNP 52 Add 6 (1998) superseding specifications prepared at the 14th JECFA (1970), published in NMRS 48B (1971) and in FNP 52 (1992). ADI "limited
More informationModelling and Simulation of the Freezing Systems and Heat Pumps Using Unisim Design
Modelling and Simulation of the Freezing Systems and Heat Pumps Using Unisim Design C. Patrascioiu Abstract The paper describes the modeling and simulation of the heat pumps domain processes. The main
More informationChapter 28: High-Performance Liquid Chromatography (HPLC)
Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion
More informationHands-on Workshop on Advanced Hydrocarbon Analysis July 23, 2015
Hands-on Workshop on Advanced Hydrocarbon Analysis July 23, 2015 Fast Analysis of Natural Gas and Fuel Gas Using Agilent 490 Micro GC - David Ramby, Quantum Analytics Quantum Analytics Value-added distributor
More informationRequest for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ)
Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) Closing Date and Time: Wednesday, April 13, 2016 at 5:00 p.m. (Pacific Time) Contact: Aaron Galloway Fax: 541-888-3250
More informationQuantitative Evaluation of Perfume Fleuressence Samples using the znose
Quantitative Evaluation of Perfume Fleuressence Samples using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses Conventional electronic noses (enoses) produce a recognizable response
More informationBakken Crude Oil & Response Considerations. Objectives
Bakken Crude Oil & Response Considerations UNITED STATES EPA Region 10 Emergency Management Program December 2014 NWEC Meeting Portland, Oregon Objectives Using EPA Bakken preparedness case study, objective
More informationOPTIONS FOR COMMERCIAL GAS DEVELOPMENT OF MARGINAL FIELDS
OPTIONS FOR COMMERCIAL GAS DEVELOPMENT OF MARGINAL FIELDS Gabriel N. Bujulu Principal Petroleum Engineer TANZANIA PETROLEUM DEVELOPMENT CORPORATION PRESENTATION OUTLINE! Introduction! Origin of Natural
More informationAMBIENT AIR QUALITY MONITORING SYSTEM. Air Quality Monitoring... Easier. Faster. Better. www.chemtrols.com
AMBIENT AIR QUALITY MONITORING SYSTEM Air Quality Monitoring... Easier. Faster. Better. www.chemtrols.com AMBIENT AIR QUALITY MONITORING FROM CHEMTROLS Chemtrols is a leading solution provider in the field
More informationIdentification of Arson Accelerants by Gas Chromatography
Identification of Arson Accelerants by Gas Chromatography Purpose Arson is an insidious crime that annually claims the lives of hundreds of Americans and costs billions of dollars. 1 Scientists and criminal
More informationSeparation of Amino Acids by Paper Chromatography
Separation of Amino Acids by Paper Chromatography Chromatography is a common technique for separating chemical substances. The prefix chroma, which suggests color, comes from the fact that some of the
More informationAnalysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.
Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class
More informationALLERGENS IN PERFUMES: GAS CHROMATOGRAPHY^ MASS SPECTROMETRY
1974 III / ALLERGENS IN PERFUMES: GAS CHROMATOGRAPHY^MASS SPECTROMETRY ALLERGENS IN PERFUMES: GAS CHROMATOGRAPHY^ MASS SPECTROMETRY S. C. Rastogi, National Environmental Research Institute, Rokilde, Denmark
More informationDGA - Method in the Past and for the Future
DGA - Method in the Past and for the Future Page 2 May 2012 1970 1973 Doerenburg introduced the differentiation between electrical and thermal failure mode and introduced ratios for fault gases with similar
More informationHow To Analyze Pesticide Levels In Lemon Oil
AppNote 9/2006 Target GC-MS Analysis using Accelerated Column Heating and Interactive Deconvolution Software Albert Robbat Tufts University, Chemistry Department, Medford, MA 02155, USA Andreas Hoffmann
More informationExtraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
More informationTheory and Instrumentation of GC Introduction
Theory and Instrumentation of GC Introduction i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
More informationDetermination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
More informationClarus 580 Gas Chromatograph
S P E C I F I C A T I O N S Gas Chromatography Clarus 580 Gas Chromatograph The PerkinElmer Clarus 580 Gas Chromatograph (GC) is a fully automated gas chromatograph. The system offers the capabilities
More informationThe Agilent 7890A Gas Chromatograph All the elements for perfect chemistry. Our measure is your success.
The Agilent 7890A Gas Chromatograph All the elements for perfect chemistry. Our measure is your success. Introducing the Agilent 7890A Gas Chromatograph. Step up to a higher level of GC reliability, productivity
More informationHow To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
More informationDew-Point Measurement Solutions
Dew-Point Measurement Solutions www.michell.com Dew-Point Measurement Solutions Special Features No downtime Michell Instruments offers a unique sensor exchange programme, which means that you will never
More informationValidation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana*
Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana* a Department of Pharmaceutical Analysis, Andhra University, Visakhapatnam. b Department of Pharmaceutics,
More informationIBP 2778_10 HIGH EFFICIENCY ON CO2 REMOVAL IN NATURAL GAS WITH UCARSOL SOLVENTS Thiago V. Alonso 1. Abstract. 1. Introduction
IBP 2778_10 HIGH EFFICIENCY ON CO2 REMOVAL IN NATURAL GAS WITH UCARSOL SOLVENTS Thiago V. Alonso 1 Copyright 2010, Brazilian Petroleum, Gas and Biofuels Institute - IBP This Technical Paper was prepared
More informationSeries 2. Air Server. Options for UNITY 2. A robust, cryogen-free solution for round-the-clock on-line air monitoring
Series 2 Air Server Options for UNITY 2 A robust, cryogen-free solution for round-the-clock on-line air monitoring Series 2 Air Server Three- and eight-channel options for UNITY offering round-the-clock
More informationSUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
More informationA Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas
A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas Paper #46 Heidi C. Hayes Air Toxics, Ltd., 180 Blue Ravine Rd. Ste. B, Folsom, CA 95630 Diane J.
More informationAMEC 6007: Linear Velocity and Acceleration Learning Outcome. CHEM 6012: Oil and Gas Compositions and Sales Specifications Learning Outcome
s/objectives Level F ADMN 6044: Post Job Management It is important to know the proper procedures for closing out a job so that rigging and equipment is maintained for future use and personnel are adequately
More informationMulti-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible
Multi-Shot Pyrolyzer EGA/PY-3030D Flexible Versatile Reproducible Why Evolved Gas Analysis? Why Pyrolysis? Today, analytical pyrolysis encompasses much more than simple flash pyrolysis of polymeric materials.
More information