Carina Pocket Guide Software version 3.n. The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use.



Similar documents
Oxylog 3000 plus Emergency & Transport Ventilation

Vista 120 Central Monitoring System

Moving beyond telemetry INFINITY M300

Agila Ceiling Supply Unit Medical Supply Units

D Move life-critical patient data on your existing network INFINITY ONENET

Movita Ceiling Supply Unit Medical Supply Units

D A cost effective monitor that meets your clinical needs VISTA 120

Sekusept Cleaner Liquid Cleaning Agent Cleaning, Desinfecting and Drying

Your personal anesthesia assistant DRÄGER PRIMUS

Dräger Globe-Trotter GT5400 Neonatal Transport

Dräger X-dock Release Notes Software-Update and Manager Software 1.7.0

The bridge to more efficient care PONTA SUPPLY BEAMS

AutoFlow The Oxylog 3000 plus incorporates the benefits of pressure controlled ventilation into volume controlled ventilation

Polaris 5xx/7xx Medical Lights and Video

Dräger PSS 3000 Self Contained Breathing Apparatus

Dräger Communication Systems

Dräger Evita Infinity V500 ventilator ICU Ventilation and Respiratory Monitoring

D Non-Invasive Ventilation (NIV) with the Oxylog 2000 plus and Oxylog 3000 plus

The Anesthesia Ventilator

Dräger Babylog VN500 Neonatal Open Care

Reference Case: Continuous mobile patient monitoring

Dräger Alcotest 9510 Evidential Alcohol Measuring Device

D Gas supply secure at a glance DRÄGER ALARM MANAGEMENT SYSTEM

MT Monitoring and IT Solutions for supporting patient safety and care INFINITY IN-HOSPITAL TRANSPORT SOLUTIONS

Table of contents. Exercise 1 Passive exhalation port 13. Exercise 2 Active exhalation valve 15

The column to efficient care DRÄGER GEMINA DUO

Infinity Acute Care System Monitoring Solution

Impact Uni-Vent 754 Portable Ventilator

How To Program An Autodialer

Operator s Instruction Manual

Trilogy200. Proximal pressure line. Exhalation valve line. Quick start guide. Overview. Circuit set-up. Set-up screen. Prescription set-up

MEDICAL SYSTEMS. Monnal T60. Touch and Breathe.

Vivo 50/60 PC Software Help (english), version Introduction. 3. Dialogs. 7. Year data view. 8. Alarm view. 2. User interface. 9.

GSM AD05 Slave GSM Auto Dialer- Instruction Manual

EKOS Cart. Instructions for Use

The centrepiece of a complete anaesthesia workstation

FLIGHT MEDICAL INNOVATIONS LTD. FLIGHT 60 Ventilator. Operator's Manual

DATA SHEET. Capnography option. November 2013 V2.2

Technical Update TAA.TU Rev. 1

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

Stove Guard Kit User Manual

Installation Guide for Hive Active Heating

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

Puritan Bennett 840 Ventilator User s Pocket Guide. Puritan Bennett 840 Ventilator Pocket Guide

site monitoring Kit Site Monitoring Kit User Manual we prove it.

Fabius GS. Inhalation Anesthesia Machine Software 3.n Operating Instructions. Because you care

Vista 120 Patient Monitor

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss heating

Home Ventilator Breath of Life Home Medical Equipment and Respiratory Services

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss Heating Solutions

MEDUMAT Easy CPR The Guiding Emergency Ventilator

Technical Service Manual Babylog 8000/8000SC/8000 plus Intensive Care Ventilator

Proton therapy Process support with anesthesia

Monitoring and It solutions for supporting patient safety and care. InfInIty

Service manual. Website: CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL.

Please also see 10 golden rules for correct Saniflo installation on

GSM HOME SECURITY SYSTEM

Impact Uni-Vent 750 Portable Ventilator

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Wireless Home Security Alarm System AM 500

Quik Lining Systems Heater Owners Guide

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR

Dräger PARAT Escape Hoods

Ventilation modes in intensive care

Dräger Ventilation Mini Manual Brief explanation of ventilation modes and functions

Memcom Emergency Telephone

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

LS1024B / LS2024B/ LS3024B. Solar Charge Controller USER MANUAL

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

DC-8706K Auto Dial Alarm System

DIAMOND Retractable Rodding Robot Model SPRAYROD-R

1/11 CO2 DOSAGE SYSTEMS. CO2 dosage system consists of following parts;

Portable Air Conditioner

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS

Thermo Top - Troubleshooting Tree

Service Manual. Coffee maker Senseo Viva Café HD7825/30 10/05. Philips Consumer Lifestyle

English. Symbols used to mark instructions...3. Congratulations...5 Getting the best results...5. Warnings...6 Operating Procedure...

MAKING MODERN LIVING POSSIBLE. UniLynx Indoor User Manual. ULX 1800i ULX 3000i ULX 3600i ULX 5400i SOLAR INVERTERS

Fully Automatic Washing Machine User manual

SunMaxx Solar Filling Station Operating Instructions

Automatic Phone-Out Home Monitoring Systems

PORTABLE AIR CONDITIONER

Operating instructions IV / 2007

USER S MANUAL HSC-24A

2» 10» 18» 26» PD »

Datex Ohmeda. Centiva /5. Service Manual Software rev. 2.1n

VFS24/32HDIP. Public Display IP Monitor User Manual

Alarm Security Kit - NVR

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW

NESCAFÉ Solution 8/60 Beverage System. Operations Manual

AN500T, AN1000, AN1000T, AN1500, AN1500T AN2000, AN2000T

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W

For every breath he takes. Trilogy200 ventilator s added sensitivity lets you breathe easier knowing your patients are where they belong home.

Your Data Security Guardian. SafeTANK GR5630-WSB3+ User Manual. v.1.0 (Oct, 2012)

Important Safeguards


RRV934. Multi-controller. Synco living

SNMP Web card. User s Manual. Management Software for Uninterruptible Power Supply Systems

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

Transcription:

MT-0569-2008 Carina Pocket Guide Software version 3.n The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. The user must be familiar with the device according to the national and local laws and recommendations. There will be no exchange of the Pocket Guide when the product is updated/upgraded.

02 CARINA POCKET GUIDE

03 1 CONTENT 2 System Overview 4 2.1 Control panel, top 4 2.2 Carina underside 4 2.3 Front connection block 4 2.4 Rear connection block 5 2.5 Trolley 5 2.6 Screen layout 6 2.7 Status line for displaying current device settings 6 3 Preparation 7 3.1 Connecting to the power supply 7 3.2 Connecting the oxygen supply 7 3.3 Connecting a hose system 8 3.4 Switching on Carina 9 3.5 Setting ventilation modes 10 3.6 Setting alarm limits 11 3.7 Displaying set values and measured values 11 3.8 Configuration 12 3.9 Locking buttons 12 4 Alarm handling 13 4.1 In the event of an alarm 13 4.2 Suppressing the alarm tone 13 5 Stand-by and Switching off 14 5.1 Switching Carina to Standby 14 5.2 Switching off the device 14 6 Cleaning and maintenance 15 6.1 Cleaning intervals 15 6.2 Maintenance intervals 15 7 Alarm message Cause & Remedy 16 7.1 Important clinical related alarms 16 7.2 Important functional related alarms 17 7.3 Important battery related alarms 17 7.4 Important technical failure alarms 18 8 Accessoires 19 8.1 Accessories for Carina 19 9 Checklist 20

04 CARINA POCKET GUIDE 2 System Overview The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. 2.1 CONTROL PANEL, TOP 1 Carrying handle 2 Screen 3 Buttons for functions and parameters 4 Rotary knob 5 Audio paused 2 min. button 6 LED for indicating power supply: LED lit = External power supply (mains/external battery), internal battery is charged LED flashes = External power supply (mains/external battery), internal battery is being charged LED off = battery operation 7»Start/Standby«button 8»Select Menu«button MT-8229-2005 8 7 6 1 2 3 4 5 1 2.2 CARINA UNDERSIDE 1 Switch for hose system with leak valve or hose system with expiratory valve masked by sliding cover (not illustrated) 2 Future functions 3 Holes for positioning on the trolley (4) D-1493-2009 2 3 2.3 FRONT CONNECTION BLOCK 1 Ventilation hose connection 2 Emergency air inlet and oxygen overflow 3 Connection for control hose (for hose system with expiratory valve) 3 1 MT-8232-2005 2

05 2 System Overview 2.4 REAR CONNECTION BLOCK 1 Not assigned 2 Connection for nurse call system 3 RS 232 serial port (MEDIBUS) 4 Not assigned 5 Inlet filter (HEPA filter) 6 Connection for HPO oxygen supply 7 Connection for LPO oxygen supply 8 Mains switch 9 Connection for external battery 10 Power cable connection MT-8233-2005 5 1 2 3 4 10 9 6 7 8 2.5 TROLLEY 1 Carina bracket 2 Standard rail grip 3 Universal bracket with standard rail, optional 4 Trolley column 5 Pedestal 6 Red locking lever of the column mount closed (under pedestal) 7 Dual rollers with locking brakes, 4 each 1 2 3 4 5 7 MT-7683-2006 6

06 CARINA POCKET GUIDE 2 System Overview 2.6 SCREEN LAYOUT 1 Alarm line for alarm message 2 Status line for current device settings (refer to 2.7) 3 Display for real-time waveforms of flow and pressure 4 Display for 2 measured values, configurable 5 Information line 6 Display of ventilation parameters 2 1 3 4 D-1494-2009 5 6 2.7 STATUS LINE FOR DISPLAYING CURRENT DEVICE SETTINGS 1 LPO oxygen supply 2 Power supply 1 2 3 4 5 6 7 int Internal battery ext External battery Mains 3 Charge of internal battery 4 Hose system expiratory valve or leak valve 5 Ventilation method tube or mask 6 Set ventilation mode 7 Patient effort detection

07 3 Preparation The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. 3.1 CONNECTING TO THE POWER SUPPLY Plug the power cable into the device and secure with the wire clip A. Insert the mains plug into a power socket. The LED flashes or lights up according to the charge of the internal battery. MT-0488-2007 A B 3.2 CONNECTING THE OXYGEN SUPPLY Screw the O 2 compressed gas hose onto the HPO nozzle B of Carina. Supply with oxygen from the central supply or from the O 2 gas cylinder. C or: Connect the O 2 supply hose from the O 2 source to the LPO nozzle C. MT-0489-2007

08 CARINA POCKET GUIDE 3 Preparation 3.3 CONNECTING A HOSE SYSTEM Select hose system with leak valve or hose system with expiratory valve. To carry out the basic device setting correctly, please refer to the Instructions for Use! Hose system with leak valve 1 Connect the filter to socket. 2 Connect the ventilation hose: VentStar Carina LV (MP00312) to filter. Check that the hose system is displayed in the status line of the screen as follows:»leakv«. 1 or: MT-0487-2007 2 Hose system with expiratory valve: 1 Connect the filter to socket. 2 Connect the ventilation hose: VentStar Carina EV (MP00313) to filter 3 Connect the pilot line of the expiratory valve to pilot hose socket. Check that the hose system is displayed in the status line of the screen as follows:»expv«. 3 1 MT-0486-2007 2

power on power off 09 3 Preparation 3.4 SWITCHING ON CARINA 1 Set the mains switch to» «Carina conducts a self-test. During the self-test the screen indicates the progress of the self-test. The self test is followed by the screen showing the ventilation mode and parameters last set. The device is in standby mode. MT-8233-2005 1 Starting ventilation 2 Press the»start/standby«button. Carina ventilates at the ventilation parameters set. The screen shows the current ventilation waveforms and values. For configuration of the screen display, see Instructions for Use. D-1495-2009 2 MT-8229-2005

10 CARINA POCKET GUIDE 3 Preparation 3.5 SETTING VENTILATION MODES Push Menu key Vent. Settings Alarms Values Config. Unlocked VC-SIMV AF PC-BIPAP* PC-AC SPN-PS Volume guarantee SPN-CPAP Apnoea ventilation Volume Control Synchronized Intermittent Mandatory Ventilation AF - AutoFlow Pressure Control Biphasic Positive Airway Pressure Pressure-controlled, Assisted Ventilation Spontaneous - Pressure Support For patients with sufficient spontaneous breathing. VG can be switched on in ventilation mode SPN-PS. Tidal volume guarantee Spontaneous - Continuous Positive Airway Pressure For automatically switching to volume-controlled, mandatory ventilation (VC-SIMV AF) in the event of apnoea. Only on SPN-PS and SPN-CPAP * Trademark used under license

11 3 Preparation 3.6 SETTING ALARM LIMITS The current measured value is displayed between the upper and lower alarm limits. Upper Limit Vent. Settings Alarms Values Config. Unlocked D-1503-2009 Measured value Lower Limit Unit ALARM Paw high MV high MV low f high Tapnea Tdisconn SETTING RANGE 10 to 55 mbar 2 to 40 L/min 0.1 to 39 L/min 10 to 50 bpm 5 to 60 seconds Mask: 0, 15, 30, 60, 90, 120 sec. Tube: 0, 15, 30, 60 sec. 3.7 DISPLAYING SET VALUES AND MEASURED VALUES Measured values, set values and pressure curve are displayed. (Only available during operation). The screen displays: Settings Measured values Graphic representation of Paw Vent. Settings Alarms Values Config. Unlocked D-1503-2009

12 CARINA POCKET GUIDE 3 Preparation 3.8 CONFIGURATION The following settings for application can be made under»config.«: Vent. Settings Alarms Values Config. Unlocked 1st page Service: for system and contact information and log book (Password:12345) Volume: adjustable from 1 to 5 Tone: default tone or Dräger tone can be selected Screen: adjustable for displaying curves or values 2nd page Set value: for selecting measured values to display value 1 and value 2 Method: for setting mask or tube ventilation, only for LeakV LPO: for setting LPO oxygen supply Ti / I : E: for selecting the setting used for Ti or I : E ratio 3rd page Night: for switching on the screen in night screen mode after 3min. display 3.9 LOCKING BUTTONS All the buttons (apart from the»audio paused 2 min.«button) can be locked in order to prevent them from being pressed accidentally. Locking: Press the»un-locked«button and keep it pressed until the colour of the button switches to red and the button indicates»locked«. Vent. Settings Alarms Values Config. Unlocked Unlocking: Press the»locked«button to unlock again. Vent. Settings Alarms Values Config. Locked

13 4 Alarm handling The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. 4.1 IN THE EVENT OF AN ALARM If the alarm message, e.g.»mv high!!«, appears in the alarm line on the screen. MV high!! Carina assigns the message an appropriate level of priority. Red!!! Yellow!! Yellow! Alarm Message with the highest priority Important Message with medium priority Note Message with low priority 4.2 SUPPRESSING THE ALARM TONE For up to 2 minutes: 1 Press»Audio paused 2 min.«key. The alarm tone is suppressed for approx. 2 minutes. 2 The symbol of silent alarm is displayed on the screen. MT-8229-2005 1 If an alarm with a higher or equal priority occurs during the period when the alarm tone is suppressed, the alarm tone sounds again. D-1502-2009 2

power on power off 14 CARINA POCKET GUIDE 5 Stand-by and Switching off The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. 5.1 SWITCHING CARINA TO STANDBY Press»Start/Standby«button and confirm with rotary knob. Carina is in standby mode. LED in the button flashes. The screen displays»standby Patient not ventilated«. 5.2 SWITCHING OFF THE DEVICE Switch Carina to standby. Set the mains switch to» «Carina is switched off. Carina cannot be switched off during ventilation. Carina signals»set standby to shut-off!!«the internal battery is also charged when the device is switched off. MT-8233-2005 D-1496-2009 MT-8229-2005

15 6 Cleaning and maintenance The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. 6.1 CLEANING INTERVALS PART INTERVALS ACTIVITY Carina basic device, Per patient Wipe disinfection Trolley, hinged arm HPO hose Reusable ventilation hoses Per patient/ weekly Dismantle, Reusable water trap Sterilise hot steam 134 C Cleaning and disinfection machine 93 C / 10 min with detergents only. Disposable components: Please follow the manufacturers Instructions for Uses. These active ingredients for wipe disinfection are suitable: Aldehydes Quaternary ammonium compounds To avoid the possibility of damage to material do not use any disinfectants based on: Alkylamine-based compounds Phenol-based compounds Halogen-releasing compounds Strong organic acids Oxygen-releasing compounds 6.2 MAINTENANCE INTERVALS Clean and disinfect equipment and/or components before any maintenance procedures and before returning for repair. PART INTERVALS ACTIVITY Inlet filter (HEPA filter) Every 6 months Replace Internal Battery Every month (+35 to +55 C) Recharge Every 3 months (-20 to +35 C) Recharge After 3 years To be replaced by technicians. Carina basic device Annually Inspection by technicians Nurse call (if applied) Monthly Check

16 CARINA POCKET GUIDE 7 Alarm message Cause & Remedy The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. The following error messages are an extract from all possible alarms only. For more information, please refer to the Instructions for Use. 7.1 IMPORTANT CLINICAL RELATED ALARMS ERROR MESSAGE CAUSE REMEDY Airway pressure high!!! The excess pressure alarm can be Check condition of patient, caused by continuous coughing. check ventilation pattern, correct alarm limit if necessary. Airway pressure low!!! A connection has become Make sure the hoses, filters and detached in the hose system. other components are secure again. Ventilation not possible! Check the setting of Tdisconn. High frequency!!! Patient is breathing at a high Check condition of patient, spontaneous frequency, therefore check ventilation pattern, correct the monitored overall frequency alarm limit if necessary. is too high. The alarm is not active in ventilation mode PC-BIPAP and VC-SIMV. MV high!! The minute volume has exceeded Check condition of patient and the upper alarm limit. ventilation pattern, correct alarm Leak in breathing system. limit if necessary. Ensure that breathing system is leak-proof. MV low!!! The minute volume has exceeded Check condition of patient and the lower alarm limit. ventilation pattern, correct alarm Leak in breathing system. limit if necessary. Ensure that breathing system is leak-proof. Apnoea ventilation!!! The device has detected respiratory Apply controlled ventilation or arrest in SPN-PS or SPN-CPAP mode extend apnoea alarm time Tapnea. and will ventilate under control in VC-SIMV mode until Spontaneous breathing resumes. Pmax limit reached! The VT measured is lower than Check the range of Pmax. the VT set. The maximum permissible Check the setting for VT. pressure range has been reached.

17 7 Alarm message Cause & Remedy 7.2 IMPORTANT FUNCTIONAL RELATED ALARMS ERROR MESSAGE CAUSE REMEDY Disconnection/ Disconnection of the hose or large Check hoses, filters and other high leakage! leakage is detected. After the components to make sure the leakage alarm, Carina delivers a low connections are not leaking constant flow at PEEP level. When and are secure. the leak has been eliminated, Carina automatically resumes operation at the previous ventilation settings. Large leakage!!! The hose system is leaking. Check hoses, filters and other Note: This is only displayed after the components to make sure the 3rd ventilation stroke. Alarm is not connections are not leaking active in ventilation mode SPN-CPAP. and are secure. Insufficient O2/LPO HPO O2 supply is too low or Check the HPO and LPO settings set?! LPO source is used and does in the Configuration menu not correspond to the device In HPO mode: check central setting. gas supply. 7.3 IMPORTANT BATTERY RELATED ALARMS ERROR MESSAGE CAUSE REMEDY External power failure!! The device has no external Check the mains supply or DC power supply. power supply. The mains or DC plug is loose. Insert the mains plug or DC plug properly at both ends. The mains or DC cable has a break. Change the cable. Mains power failure. The external Connect a replacement battery. battery is completely discharged. Connect to the mains. Charge external battery. Int. battery activated! Lack of an external power supply. Reconnect external power supply. Int. battery low! Battery charge has dropped to Charge the battery: below 50 %. Connect external power supply. Int. battery low!!! Battery charge has dropped to Charge the battery: below 25 %. Connect external power supply. Int. battery empty!!! The operating time with supply Connect ventilator immediately to from the internal battery has expired. mains power supply or to a fully charged external battery.

18 CARINA POCKET GUIDE 7 Alarm message Cause & Remedy 7.4 IMPORTANT TECHNICAL FAILURE ALARMS ERROR MESSAGE CAUSE REMEDY Auto Reset!!! Internal system fault. The device performs a restart. Ventilation continues. After 3 attempts the device stops ventilation. Disconnect the patient from Carina and continue ventilation immediately using a different, separate ventilator. Call DrägerService. Technical alarm!! The device has a technical problem. Use a replacement device. This Technical Failure does not affect Call DrägerService. to the breathing function. e.g. Hard key defect Technical alarm!!! The device has a technical problem. Use a replacement device. This Technical Failure might affect to Call DrägerService. the breathing function. e.g. Temperature sensor defect Device failure!!! The device has a fatal technical Immediately disconnect the patient problem. Ventilation has failed. from Carina and continue ventilation using a different ventilator. If the device cannot be turned off, remove the O2 hose from the central supply and then turn off the device. Call DrägerService.

19 8 Accessoires The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. 8.1 ACCESSORIES FOR CARINA MT-7716-2006 VentStarLeakV MT-7718-2006 VentStarExpV MT-0970-2007 MT-4387-2007 MT-7704-2006 NovaStar CareStar HEPA filter MT-11178-2006 MT-0575-2007 Pump ball ClassicStar NR. PRODUCT NUMBER DESCRIPTION 1 Ventstar Carina LeakV MP00312 Dispo. hose with leakage valve Length: 150 cm, 5 pieces 2 Ventstar Carina ExpV MP00313 Dispo. hose with expiratory valve Length: 150 cm, 5 pieces 3 Filter CareStar 30 MP01770 For device protection 4 HEPA filter 5703105 Consumable parts for Carina Exchange: Every 6 month 5 NovaStar size L MP01581 Gel mask: Reusable 6 NovaStar size M MP01580 Gel mask: Reusable 7 NovaStar size S MP01579 Gel mask: Reusable 8 ClassicStar size L MP01575 Air cushion Mask 9 ClassicStar size M MP01574 Air cushion Mask 10 ClassicStar size S MP01573 Air cushion Mask 11 Two-side pump ball MP01590 To inflate ClassicStar Air cushion 12 Leak valve silent flow MP00224 Dispo. leakage valve, 5 pieces 13 Expiratory valve MP00220 Dispo. expiratory valve, 1 piece 14 Spiral hose E 150 cm 2165643 Reusable silicone hose: 150 cm

20 CARINA POCKET GUIDE 9 Checklist The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. This test must be conducted every time before the device is put into operation. Please use the blue Dräger testlung (partnumber: 84 03 201) ACTION OBSERVE CHECK 1 Connect all necessary equipment: Equipment is connected correctly, hoses, cable, filter, etc. refer to chapter preparation. 2 Turn the main switch on. Carina performs a self-test, LEDs light up and alarm tone is given. 1 Main switch LED LED 2 3 3 Check if the used hose system»leakv«for leak valve or»expv«for (leak valve or expiratory valve) expiratory valve is displayed in the conforms with the text in the status line. Example: status line. 4 Check if the used patient connection»mask«for mask or»tube«for tube (tube or mask) conforms with the application method is displayed in the text in the status line. status line. Example: 5 Connect the test lung to the hose system. 6 Make the following settings: Each time the lung is compressed, expiration should be initiated instantly. Mode PC-BIPAP Settings Pinsp 15 mbar PEEP 5 mbar FiO2 21% f 10 bpm Ti / I:E 2 s / 1:2 Alarms Ramp 0.2 s After the second compression on Paw high 20 mbar Pinsp level alarm tone starts and Tdisconn 0 s Airway pressure high!!! is displayed. Press»Start/ Standby«button to start the ventilation and squeeze the test lung several times strongly during inspiration.

21 9 Checklist ACTION OBSERVE CHECK 7 Confirm alarm, then disconnect the Alarm tone starts and Airway test lung from the hose system. pressure low!!! is displayed. 8 Reconnect test lung and hose system. 9 Change the following settings: Mode SPN-CPAP Settings AppVent on Alarm Tapn 5 s Without spontaneous breathing and after apnoea alarm time has elapsed, Steadily inflate and deflate the test alarm Apnoea ventilation!!! is lung to simulate spontaneous displayed. Apnoea ventilation starts. breathing, then stop simulation. 10 Set ventilation mode to PC-BIPAP. 11 Disconnect device from all external Alarm tone starts and display indicates power supplies. External power failure!!. 12 Reconnect the power supplies. Disconnect the test lung from the hose system.

22 CARINA POCKET GUIDE

23

CORPORATE HEADQUARTERS Drägerwerk AG & Co. KGaA Moislinger Allee 53 55 23558 Lübeck, Germany www.draeger.com Manufacturer: Drägerwerk AG & Co. KGaA Moislinger Allee 53 55 23558 Lübeck, Germany Locate your Regional Sales Representative at: www.draeger.com/contact REGION EUROPE CENTRAL AND EUROPE NORTH Drägerwerk AG & Co. KGaA Moislinger Allee 53 55 23558 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 2080 info@draeger.com REGION EUROPE SOUTH Dräger Médical S.A.S. Parc de Haute Technologie d Antony 2 25, rue Georges Besse 92182 Antony Cedex, France Tel +33 1 46 11 56 00 Fax +33 1 40 96 97 20 dlmfr-contact@draeger.com REGION MIDDLE EAST, AFRICA Drägerwerk AG & Co. KGaA Branch Office P.O. Box 505108 Dubai, United Arab Emirates Tel +971 4 4294 600 Fax +971 4 4294 699 contactuae@draeger.com REGION ASIA / PACIFIC Draeger Medical South East Asia Pte Ltd. 25 International Business Park #04-27/29 German Centre Singapore 609916, Singapore Tel +65 6572 4388 Fax +65 6572 4399 asia.pacific@draeger.com REGION CENTRAL AND SOUTH AMERICA Dräger Panama Comercial S. de R.L. Complejo Business Park, V tower, 10th floor Panama City Tel +507 377 9100 Fax +507 377 9130 contactcsa@draeger.com 90 51 312 15.06-5 Communications & Sales Marketing PP LE Printed in Germany Subject to modifications 2015 Drägerwerk AG & Co. KGaA