COMPENSATION MIT Flow Cytometry Core Facility

Similar documents
BD FACSComp Software Tutorial

APPLICATION INFORMATION

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...

Becton-Dickinson FACSCalibur Flow Cytometer Instructions

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

These particles have something in common

Immunophenotyping peripheral blood cells

123count ebeads Catalog Number: Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer

UCHC Guide to CellQuest Pro. FACS Calibur instruments

GETTING STARTED. 6. Click the New Specimen icon.

Three Rules for Compensation Controls

BD FACSDiva TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility

CyAn : 11 Parameter Desktop Flow Cytometer

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin

1. Instrument Layout pg Summit Overview pg Data Storage and Data Files pgs Saving Data pg Start Up pg 8

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.

Flow cytometry basics fluidics, optics, electronics...

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:

Fluorescence Compensation. Jennifer Wilshire, PhD

Introduction to Flow Cytometry

Spherotech, Inc Irma Lee Circle, Unit 101, Lake Forest, Illinois

Introduction to Flow Cytometry:

Principles of Flowcytometry

FACS-CALIBUR. USER-Guide

Instrument Characterization and Performance Tracking for Digital Flow Cytometers

Introduction to flow cytometry

Cell Cycle Tutorial. Contents

BD CellQuest Pro Software Analysis Tutorial

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

Compensation Controls Data Visualization Panel Development Gating Quantum Dots

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose

Attune Acoustic Focusing Cytometer

FACSCanto RUO Special Order QUICK REFERENCE GUIDE

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking

Science is hard. Flow cytometry should be easy.

LSR II Users Guide Stanford Shared FACS Facility 06/10/13

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T

Compensation in Flow Cytometry

Analyzer Experiment setup guide for LSRII and Canto s

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit G June Beckman Coulter, Inc N. Harbor Blvd. Fullerton, CA 92835

Introduction to Flow Cytometry

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction

Islet Viability Assessment by Single Cell Flow Cytometry

Navios Quick Reference

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

How to operate the BD FACSCanto flow cytometer

FACSCalibur System User s Guide

UCHC Guide to FACS DiVa. LSR II instruments

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc.

Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113

INSIDE THE BLACK BOX

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences

BD FC Beads 8-Color Kit for BD OneFlow Assays

CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles

Attune NxT Acoustic Focusing Cytometer

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and

Boundary-breaking acoustic focusing cytometry

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping

Standardization, Calibration and Quality Control

Zeiss 780 Training Notes

Fluorescent dyes for use with the

Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II

How To Read Flow Cytometry Data

Establishing Optimum Baseline PMT Gains to Maximize Resolution on BD Biosciences Digital Flow Cytometers

Flow Cytometry. flow cytometer DNA apoptosis ph

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee

Getting Started with BD FACSDiva Software

powerful. Intuitive. Customizable. The new star of benchtop flow cytometry

Institut für Experimentelle Immunologie und Bildgebung

Setting up and calibration of a flow cytometer for multicolor immunophenotyping

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab

MACSQuantify Software guide. Basics of flow cytometric analysis and software guide. Version 1 EN Original instructions

ArC Amine Reactive Compensation Bead Kit

Chapter 12 Filters for FISH Imaging

Deep profiling of multitube flow cytometry data Supplemental information

Zecotek S Light Projection Network Marketing

Compensation: An Instrumental Perspective

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size

MetaMorph Software Basic Analysis Guide The use of measurements and journals

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract

FACS Laboratory. BD LSRs Operator Training. Training Guide.

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Flow Cytometry A Basic Overview

Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH))

TROUBLESHOOTING FLOW CHART

FloMax Software for Cytometry. Operating Manual - Data Analysis. Version 2.3, FloMax Operating Manual Data Analysis 1

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.

Transcription:

COMPENSATION MIT Flow Cytometry Core Facility

Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE.

2) For sensitivity reasons, we must use broad bandpass and dichroic filters in the cytometer in order to separate fluorescence the emission spectra of dyes from the excitation light source.

THE END RESULT IS SPECTRAL OVERLAP where FITC bleeds into the PE channel and PE bleeds back into FITC.

To correct for spectral overlap during multicolor flow cytometry experiments, color compensation must be performed. The goal of color compensation is to correctly quantify each dye with which a particular cell is labeled. This is done by subtracting a portion of one detector's signal from another, leaving only the desired signal.

This example will show you how to compensate a FITC and PE experiment using a FacsCalibur flow cytometer. Open the Detectors/Amps and Threshold windows by selecting them from the Cytometer drop down menu in Cell Quest Pro software.

Loading your negative control Put the cytometer on run mode and put your negative control on the SIP While the negative control is running under setup mode press acquire and make the following changes Select a location to save your file Select a name for your file

Make sure the population of interest is clear visible by making adjustment to the FSC/ SSC. Adjust the forward and side scatter by sliding the bar to bring the population of interest toward the center of the plot.

Remove debris by using threshold. Adjust the threshold value by sliding the bar on the threshold window enough to remove most of the debris but not all of it. It is a good idea to have some debris on scale to make sure you are not thresholding out your population of interest. Debris

Select your population of interest by creating a gate around that population Draw a gate, which will allow you to remove unwanted debris and aggregates, around the population of interest by selecting a circle or a polygon from the tool template. REGION R1

Adjust the voltage on the FITC and PE detectors to put the negative cells in the center of the first log. Adjust the FL1 and FL2 voltage by sliding the bar up or down to bring the population in the middle of the first log. PE only PE + FITC NEGATIVE FITC only

You are now done with the negative control and ready to move on to the first single stained control Open the compensation window by selecting it from the Cytometer drop down menu. This bar allows you to remove spill over of PE into FITC This bar allows you to remove spill over of FITC into PE

Compensating the FITC Tube Still under the Setup Mode, load the FITC single color control in the cytometer then press Acquire. Your sample will look like the data on the left. Adjust the slide on the FL2- %FL1 to remove the % of FITC spill over in the PE channel. Try not to over compensate, pay attention to the mean of the negative and the FITC, they should be the same. uncompensated PE only PE + FITC We went from 0.0% to 26.0% Y mean = 65 Y mean = 5 compensated PE only PE + FITC NEGATIVE FITC only Y mean = 5 Y mean = 5 NEGATIVE FITC only

You are now done with the FITC tube and ready to compensate your PE control Still under Setup Mode, load the PE single control in the cytometer then press Acquire. Your sample will look like the data on the left. Adjust the slide on the FL1- %FL2 to remove the % of PE spill over in the FITC channel. Try not to over compensate, pay attention to the mean of the negative and the PE, they should be the same. We went from 0.0% to 0.6% uncompensated PE only PE + FITC X mean = 9 X mean = 5 compensated PE only PE + FITC NEGATIVE FITC only X mean = 5 X mean = 5 NEGATIVE FITC only

THE FITC AND PE COMPENSATION IS DONE. YOU ARE NOW READY TO RECORD DATA AND SAVE YOUR INSTRUMENT SETTING. TO SAVE YOUR INSTRUMENT SETTING: GO TO CYTOMETER MENU THEN SELECT INSTRUMENT SETTING. LOCATE YOUR DIRECTORY AND PRESS SAVE THEN PRESS DONE.