Introduction to Flow Cytometry:
|
|
|
- Roberta Gilmore
- 10 years ago
- Views:
Transcription
1 Introduction to Flow Cytometry: A Learning Guide Manual Part Number: April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA
2 Introduction to Flow Cytometry: A Learning Guide Copyright Disclaimer Trademarks 2000 Becton, Dickinson and Company. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means: electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without the prior written permission of BD Biosciences, 2350 Qume Drive, San Jose, CA 95131, United States of America. BD Biosciences reserves the right to change its products and services at any time to incorporate the latest technological developments. This guide is subject to change without notice. BD Biosciences welcomes customer input on corrections and suggestions for improvement. Although this guide has been prepared with every precaution to ensure accuracy, BD Biosciences assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information. Apple, the Apple logo, Mac, Macintosh, and Power Macintosh are trademarks of Apple Computer, Inc., registered in the U.S. and other countries. Finder is a trademark of Apple Computer, Inc. Modfit LT and QuantiCALC are trademarks of Verity Software House, Inc. CellQuest, FACS, FACSCalibur, FACScan, FACSort, FACStar PLUS, and FACS Vantage are trademarks of Becton, Dickinson and Company.
3 Contents Preface iii Chapter 1: Overview 5 Chapter 2: Fluidics 9 Chapter 3: Generation of Scatter and Fluorescence Light Scatter Fluorescence Chapter 4: Optical System Optical Bench Optical Filters Signal Detection Threshold Chapter 5: Data Analysis Data Collection and Display Gating Data Analysis for Subsetting Applications Data Analysis for Other Applications Chapter 6: Sorting Sorting Chapter 7: Lasers and Alignment 43 i
4 Introduction to Flow Cytometry: A Learning Guide 7.1 How Lasers Work Laser Alignment Chapter 8: Answer Key 47 ii
5 Preface Learning to operate a flow cytometer is best achieved by using the instrument. However, understanding the principles underlying this technology greatly facilitates the process. This document contains basic information on flow cytometry. Differences between flow cell based benchtop cytometers (FACScan, FACSort, FACSCalibur, and BD LSR) and stream-in-air cytometers (FACS Vantage, FACSVantage SE, and FACStar PLUS ) are described in relevant sections. Reading this material and answering the questions at the end of each section will enhance your hands-on training experience during Operator Training at BD Biosciences. This assignment will take approximately 2.5 hours to complete. Please review it before you attend the training session. An answer key is provided. If you have any questions or problems in the US, call , Option 4. In Europe, contact your local application specialist. For More Information For more information on general flow cytometry, review the following: Givan AL. Flow Cytometry: First Principles. New York, NY: Wiley-Liss; 1992 (ISBN ). Melamed MR. Flow Cytometry and Sorting. New York, NY: Wiley-Liss; 1990 (ISBN ). Shapiro H. Practical Flow Cytometry. 3rd ed. New York, NY: Alan R. Liss; 1994 (ISBN ). iii
6 Introduction to Flow Cytometry: A Learning Guide iv
7 1 Overview Flow cytometry is a technology that simultaneously measures and then analyzes multiple physical characteristics of single particles, usually cells, as they flow in a fluid stream through a beam of light. The properties measured include a particle s relative size, relative granularity or internal complexity, and relative fluorescence intensity. These characteristics are determined using an optical-to-electronic coupling system that records how the cell or particle scatters incident laser light and emits fluorescence. A flow cytometer is made up of three main systems: fluidics, optics, and electronics. The fluidics system transports particles in a stream to the laser beam for interrogation. The optics system consists of lasers to illuminate the particles in the sample stream and optical filters to direct the resulting light signals to the appropriate detectors. The electronics system converts the detected light signals into electronic signals that can be processed by the computer. For some instruments equipped with a sorting feature, the electronics system is also capable of initiating sorting decisions to charge and deflect particles. In the flow cytometer, particles are carried to the laser intercept in a fluid stream. Any suspended particle or cell from micrometers in size is suitable for analysis. Cells from solid tissue must be disaggregated before analysis. The portion of the fluid stream where particles are located is called the sample core. When particles pass through the laser intercept, they scatter laser light. Any fluorescent molecules present 5
8 Introduction to Flow Cytometry: A Learning Guide on the particle fluoresce. The scattered and fluorescent light is collected by appropriately positioned lenses. A combination of beam splitters and filters steers the scattered and fluorescent light to the appropriate detectors. The detectors produce electronic signals proportional to the optical signals striking them. List mode data are collected on each particle or event. The characteristics or parameters of each event are based on its light scattering and fluorescent properties. The data are collected and stored in the computer. This data can be analyzed to provide information about subpopulations within the sample (Figure 1-1). laser sample core data displays electronic pulses Figure 1-1 Scattered and emitted light signals are converted to electronic pulses that can be processed by the computer 6
9 Chapter 1: Overview Review Questions: Overview 1 What properties of a cell or particle can be measured by a flow cytometer? 2 What light source is used in most flow cytometers? 3 What are the three main systems in a flow cytometer? 4 What type of biological sample is best suited for flow cytometric analysis? 5 What is the name given to the portion of the fluid stream where the cells are located? 6 When cells labeled with fluorescent molecules pass through the focused laser beam, what two types of light signals are generated? 7 Light emitted from a particle is collected by: 8 The electronic signal produced by the detectors is proportional to: 9 All flow cytometric measurements can be made simultaneously on a single cell. T F 10 Particles must be in single-cell suspension before flow cytometric analysis. T F 7
10 Introduction to Flow Cytometry: A Learning Guide 8
11 2 Fluidics The purpose of the fluidics system is to transport particles in a fluid stream to the laser beam for interrogation. For optimal illumination, the stream transporting the particles should be positioned in the center of the laser beam. In addition, only one cell or particle should move through the laser beam at a given moment. To accomplish this, the sample is injected into a stream of sheath fluid within the flow chamber. The flow chamber in a benchtop cytometer is called a flow cell and the flow chamber in a stream-in-air cytometer is called a nozzle tip. The design of the flow chamber causes the sample core to be focused in the center of the sheath fluid where the laser beam will then interact with the particles. Based on principles relating to laminar flow, the sample core remains separate but coaxial within the sheath fluid. The flow of sheath fluid accelerates the particles and restricts them to the center of the sample core. This process is known as hydrodynamic focusing. For an illustration of hydrodynamic focusing in each type of flow cell, see Figure 2-1 and Figure
12 Introduction to Flow Cytometry: A Learning Guide low sample pressure (12 µl/min) laser beam high sample pressure (60 µl/min) laser beam sheath fluid sample sheath fluid sheath fluid sample sheath fluid Figure 2-1 Hydrodynamic focusing of the sample core through a flow cell sample sheath sheath sheath sheath laser laser low differential pressure high differential pressure Figure 2-2 Hydrodynamic focusing of the sample core through a nozzle tip The sample pressure and the sheath fluid pressure are different from each other. The sample pressure is always greater than the sheath fluid pressure. The sample pressure regulator controls the sample flow rate by changing the sample pressure relative to the sheath pressure. In BD benchtop cytometers, the sample stream is pressurized upward through an optically clear region of the flow cell or cuvette; particles pass through the laser beam while they are still within this flow cell (Figure 2-1). Most benchtop cytometers have fixed sample pressure settings (LO, MED, and HI). The benchtop BD LSR also has a fine adjustment knob for intermediate settings. 10
13 Chapter 2: Fluidics In stream-in-air cytometers, the sample stream passes through a small orifice in a nozzle tip before being intersected by the light beam in the open air (Figure 2-2). Sample pressure settings can be adjusted within a dynamic range. Increasing the sample pressure increases the flow rate by increasing the width of the sample core. This, in turn, allows more cells to enter the stream within a given moment. With a wider sample core, some cells could pass through the laser beam off-center and intercept the laser beam at a less optimal angle. However, this might be appropriate for your application. A higher flow rate is generally used for qualitative measurements such as immunophenotyping. The data are less resolved, since the cells are less in line in the wider core stream, but are acquired more quickly. A lower flow rate decreases the width of the sample core and restricts the position of the cells to a smaller area. The majority of cells passes through the center of the laser beam; thus the light illuminating the cells and emitted from the cells is more uniform. A lower rate is generally used in applications where greater resolution is critical, such as DNA analysis. Proper operation of fluidic components is critical for particles to properly intercept the laser beam. Therefore, the operator must always ensure that the fluidics system is free of air bubbles and debris and is properly pressurized at all times. 11
14 Introduction to Flow Cytometry: A Learning Guide Review Questions: Fluidics 1 The purpose of the fluidics system in a flow cytometer is: 2 What two factors can affect illumination of the particles within the laser beam? 3 How many cells or particles should pass through the laser beam at a given time? 4 The particle suspension is injected into within the. 5 The process of centering the sample core within the sheath fluid is known as: 6 Which regulator controls the diameter of the sample core? 7 What are the three possible pressure settings for a benchtop flow cytometer? 8 Increasing sample pressure the sample flow rate and the of the sample core. 9 Good data resolution is required for DNA studies. What flow rate is recommended? 10 A wider sample core decreases resolution. T F 11 A high flow rate can be used when performing qualitative measurements. T F 12
15 3 Generation of Scatter and Fluorescence In the last section, we learned how particles or cells are aligned to pass single file through the sample core. Before describing how the flow cytometer detects and processes signals, it is useful to understand what happens to the laser light as it strikes the single-file particles. 3.1 Light Scatter Light scattering occurs when a particle deflects incident laser light. The extent to which this occurs depends on the physical properties of a particle, namely its size and internal complexity. Factors that affect light scattering are the cell's membrane, nucleus, and any granular material inside the cell. Cell shape and surface topography also contribute to the total light scatter. Forward-scattered light (FSC) is proportional to cell-surface area or size. FSC is a measurement of mostly diffracted light and is detected just off the axis of the incident laser beam in the forward direction by a photodiode (Figure 3-1 on page 14). FSC provides a suitable method of detecting particles greater than a given size independent of their fluorescence and is therefore often used in immunophenotyping to trigger signal processing. 13
16 Introduction to Flow Cytometry: A Learning Guide Side-scattered light (SSC) is proportional to cell granularity or internal complexity. SSC is a measurement of mostly refracted and reflected light that occurs at any interface within the cell where there is a change in refractive index (Figure 3-1). SSC is collected at approximately 90 degrees to the laser beam by a collection lens and then redirected by a beam splitter to the appropriate detector. side scatter detector light source forward scatter detector Figure 3-1 Light-scattering properties of a cell Correlated measurements of FSC and SSC can allow for differentiation of cell types in a heterogeneous cell population. Major leucocyte subpopulations can be differentiated using FSC and SSC (Figure 3-2). neutrophils lysed whole blood monocytes lymphocytes Figure 3-2 Cell subpopulations based on FSC vs SSC 14
17 Chapter 3: Generation of Scatter and Fluorescence Review Questions: Light Scatter 1 When does light scattering occur? 2 Which key cell components contribute to light scatter? 3 Light scattered in the same direction as the laser beam is called: 4 FSC is proportional to: 5 Light scatter collected at 90 degrees to the laser beam is called: 6 SSC is proportional to the or of the cell. 7 Correlated measurements of both and can allow differentiation of cells types in a heterogeneous cell population. 15
18 Introduction to Flow Cytometry: A Learning Guide 3.2 Fluorescence A fluorescent compound absorbs light energy over a range of wavelengths that is characteristic for that compound. This absorption of light causes an electron in the fluorescent compound to be raised to a higher energy level. The excited electron quickly decays to its ground state, emitting the excess energy as a photon of light. This transition of energy is called fluorescence. The range over which a fluorescent compound can be excited is termed its absorption spectrum. As more energy is consumed in absorption transitions than is emitted in fluorescent transitions, emitted wavelengths will be longer than those absorbed. The range of emitted wavelengths for a particular compound is termed its emission spectrum. The argon ion laser is commonly used in flow cytometry because the 488-nm light that it emits excites more than one fluorochrome. (See Chapter 7 for more information about lasers.) One of these fluorochromes is fluorescein isothiocyanate (FITC). In the absorption spectrum of FITC (Figure 3-3 on page 17), the 488-nm line is close to the FITC absorption maximum. Excitation with this wavelength will result in a high FITC emission. If the fluorochrome were excited by another wavelength within its absorption spectrum, light emission of the same spectrum would occur but it would not be of the same intensity. More than one fluorochrome can be used simultaneously if each is excited at 488 nm and if the peak emission wavelengths are not extremely close to each other. The combination of FITC and phycoerythrin (PE) satisfies these criteria. The emission spectrum of each of these fluorochromes is shown in Figure 3-4 on page 17. Although the absorption maximum of PE is not at 488 nm, the fluorochrome is excited enough at this wavelength to provide adequate fluorescence emission for detection. More important, the peak emission wavelength is 530 nm for FITC and 570 nm for PE. These peak emission wavelengths are far enough apart so that each signal can be detected by a separate detector. The amount of fluorescent signal detected is proportional to the number of fluorochrome molecules on the particle. 16
19 Chapter 3: Generation of Scatter and Fluorescence FITC = fluorescein isothiocyanate PE = phycoerythrin PerCP = peridinin chlorophyll protein APC = allophycocyanin Figure 3-3 Absorption spectra of four common fluorochromes Figure 3-4 Emission spectra of the fluorochromes shown in Figure 3-3 When a fluorescent dye is conjugated to a monoclonal antibody, it can be used to identify a particular cell type based on the individual antigenic surface markers of the cell (Figure 3-5 on page 18). In a mixed population of cells, different fluorochromes can be used to distinguish separate subpopulations. The staining pattern of each subpopulation, combined with FSC and SSC data, can be used to identify which cells are present in a sample and to count their relative percentages. The cells can also be sorted if desired. 17
20 Introduction to Flow Cytometry: A Learning Guide fluorochrome-labeled antibodies antigenic surface marker Figure 3-5 Specific binding of fluorochrome-labeled antibodies to cell surface antigens Review Questions: Fluorescence 1 When fluorescent compounds absorb light energy and then release excess energy, they emit. 2 Characteristic wavelength ranges at which fluorescent compounds can be excited are called. 3 The longer wavelengths of light emitted by a fluorochrome make up its. 4 Which laser is most commonly used in flow cytometry? 5 The FITC and PE fluorochromes are excited by this emission wavelength of an argon-ion laser: 6 Two fluorescent dyes commonly used in flow cytometry are and. 7 Fluorochrome-labeled antibodies are used to detect. 18
21 4 Optical System The optical system consists of excitation optics and collection optics. The excitation optics consist of the laser and lenses that are used to shape and focus the laser beam. The collections optics consist of a collection lens to collect light emitted from the particle laser beam interaction and a system of optical mirrors and filters to route specified wavelengths of the collected light to designated optical detectors. The design of the optical bench allows for these functions to occur. 4.1 Optical Bench The optical bench in a flow cytometer provides a stable surface that holds the light source and the excitation and collection optics in fixed positions. The alignment of a benchtop analyzer is very stable because the flow cell is fixed in its alignment with the laser beam. This ensures that the laser intercepts the sample stream consistently from day to day. The optical bench for two benchtop cytometers is shown in Figure 4-1 and Figure 4-2 on page 20. In a stream-in-air cytometer, the laser beam is focused through an achromatic lens to intercept the fluid stream at an optimal angle and position after the stream exits the nozzle. Because of variations in focusing and in the position of the stream, the alignment of a stream-in-air cytometer is less stable than that of a benchtop analyzer and needs to be optimized on a daily basis. Misalignment of the laser beam with the stream can result in uneven illumination of the particles and increased variability in the signals emitted. This results in decreased precision. The optical bench for a stream-in-air cytometer is shown in Figure 4-3 on page
22 Introduction to Flow Cytometry: A Learning Guide Figure 4-1 Optical bench diagram of the FACSCalibur benchtop flow cytometer FL2 575/25 FL1 530/ LP 488/25 SSC FL5 380 LP 510 LP FL6 470 LP 620 SP 660/13 FL4 510/ LP 670 LP 488 nm laser fluorescence collection lens FL3 633 nm laser 488 pass/ 633 reflect 325 nm laser UV pass/ 488 & 633 reflect focusing lens flow cell FSC Figure 4-2 Optical bench diagram of the BD LSR benchtop flow cytometer 20
23 Chapter 4: Optical System detector option 5 (laser #1) FL3 PMT (laser #1) FL1 PMT (laser #1) filter filter FL2 PMT (laser #1) detector option 3 dichroic mirror prism 8 prism 7 lens filter dichroic mirror lens dichroic mirror filter lens filter iris intercept #3 FL4, 5, 6 (option) prism 5 laser #3 prism 6 relay lens iris dichroic mirror 3 beam splitter detector option 4 filter lens iris filter SSC PMT (laser #1) laser #2 prism 2 dichroic mirror filter laser #1 filter detector option 1 intercept #2 FL4, 5, 6 (option) detector option 2 beam splitter lens FL objective lens FL obscuration bar prism 4 FSC photodiode filter (488/10) lens iris 45 mirror FSC obscuration bar achromatic lens lens iris lens intercept point stream viewing video camera prism 1 prism 3 Figure 4-3 Optical bench diagram of the FACSVantage SE stream-in-air flow cytometer Review Questions: Optical Bench 1 The optical bench provides a stable surface for the interaction of the laser light with the. 2 Which two components of the flow cytometer must be perfectly aligned to illuminate cells uniformly? 3 The alignment of a stream-in-air cytometer requires daily optimization. T F 4 For benchtop flow cytometers, the fixed ensures that the laser intercepts the consistently from day to day. 21
24 Introduction to Flow Cytometry: A Learning Guide 4.2 Optical Filters Once a cell or particle passes through the laser light, emitted SSC and fluorescence signals are diverted to the photomultiplier tubes (PMTs) and a photodiode collects the FSC signals. All of the signals are routed to their detectors via a system of mirrors and optical filters. PMTs detect fluorescence signals, which are often weak. The specificity of a detector for a particular fluorescent dye is optimized by placing a filter in front of the PMT, which allows only a narrow range of wavelengths to reach the detector. This spectral band of light is close to the emission peak of the fluorescent dye. Such filters are called bandpass (BP) filters. For example, the filter used in front of the FITC detector is labeled 530/30. This number gives the characteristics of the spectral band transmitted: 530 ±15 nm, or wavelengths of light that are between 515 nm and 545 nm. Other filters used in the flow cytometer are shortpass (SP) filters, which transmit wavelengths of light equal to or shorter than a specified wavelength, and longpass (LP) filters, which transmit wavelengths of light equal to or longer than a specified wavelength (Figure 4-4 on page 23). Beam splitters are devices that direct light of different wavelengths in different directions. Dichroic mirrors are a type of beam splitter. The 560 SP dichroic mirror shown in Figure 4-1 transmits wavelengths of light 560 nm or shorter. Wavelengths of light longer than 560 nm are reflected at 45 degrees. 22
25 Chapter 4: Optical System Longpass Shortpass Bandpass LP500 SP /50 % transmittance % transmittance % transmittance wavelength (nm) wavelength (nm) wavelength (nm) Figure 4-4 Light transmittance through longpass, shortpass, and bandpass filters Review Questions: Optical Filters 1 Optical filters are placed in front of detectors to: 2 A 530/30 bandpass filter transmits wavelengths of light between and nm. 3 A is used to split light signals by wavelength and redirect the light signals to the appropriate detector. 4 filters transmit wavelengths equal to or shorter than a specified wavelength, while filters transmit wavelengths equal to or longer than a specified wavelength. 23
26 Introduction to Flow Cytometry: A Learning Guide 4.3 Signal Detection Light signals are generated as particles pass through the laser beam in a fluid stream. These light signals are converted to electronic signals (voltages) by photodetectors and then assigned a channel number on a data plot. There are two types of photodetectors in BD flow cytometers: photodiodes and photomultiplier tubes (PMTs). The photodiode is less sensitive to light signals than the PMTs and thus is used to detect the stronger FSC signal. PMTs are used to detect the weaker signals generated by SSC and fluorescence. A voltage pulse is created when a particle enters the laser beam and starts to scatter light or fluoresce. Once the light signals, or photons, strike one side of the PMT or the photodiode, they are converted into a proportional number of electrons that are multiplied, creating a greater electrical current. The electrical current travels to the amplifier and is converted to a voltage pulse. The highest point of the pulse occurs when the particle is in the center of the beam and the maximum amount of scatter or fluorescence is achieved. As the particle leaves the beam, the pulse comes back down to the baseline (Figure 4-5). Laser Voltage Time Laser Voltage Time Laser Voltage Time Figure 4-5 Creation of a voltage pulse The size of the voltage pulse depends on the number of photons detected, the PMT voltage or pre-amplifier gain, and the amplifier gain. Signals can be amplified by applying a voltage to the PMTs, thus creating a greater electrical current, or by increasing the amplification gain. Amplifier settings can be linear or logarithmic (Lin or Log). Log amplification is often used to separate negative from dim positive signals, 24
27 Chapter 4: Optical System whereas linear amplification is often used to amplify scatter and fluorescent parameters. The voltage pulse is assigned a digital value by the Analog-to-Digital Converter (ADC). The ADC converts a 0 1,000 mv pulse to a digital number representing 0 1,000 mv channels. The channel number is transferred to the computer via the General Purpose In/Out (GPIO) cable (Figure 4-6). The light signal is then displayed in an appropriate position on the data plot. Figure 4-6 Voltage pulses converted to channel values by ADC converter Review Questions: Signal Detection 1 Forward-scattered light is detected by a. 2 Side-scattered light and fluorescence are detected by highly sensitive 3 The light detectors generate current, which travels to an amplifier. It converts the current to voltages which are proportional to the intensity of light striking them. T F 4 What does the ADC do? 25
28 Introduction to Flow Cytometry: A Learning Guide 4.4 Threshold An electronic threshold can be used to limit the number of events that the flow cytometer acquires. A threshold is set on one parameter. When a threshold value is defined, only signals with an intensity greater than or equal to the threshold channel value will be processed and sent to the computer. For example, when running immunophenotyping samples, the threshold can be set on FSC to eliminate events such as debris that are smaller than the threshold channel number. Other parameters can be used to set the threshold depending on the application. A second threshold parameter is available on some benchtop analyzers equipped with the second-laser option. If two threshold parameters are chosen, then the particle must meet the values of both thresholds to be processed as an event. Review Questions: Threshold 1 FSC threshold can be used to eliminate signals generated from: 2 If two threshold parameters are chosen, the event must meet the values of one or the other parameter to be processed. T F 26
29 5 Data Analysis 5.1 Data Collection and Display Once light signals have been converted to electronic pulses and then converted to channel numbers by the ADC, the data must be stored by the computer system. Flow cytometric data is stored according to a standard format, the flow cytometry standard (FCS) format, developed by the Society for Analytical Cytology. * According to the FCS standard, a data storage file includes a description of the sample acquired, the instrument on which the data was collected, the data set, and the results of data analysis. A single cell analyzed for four parameters (FSC, SSC, FITC, and PE fluorescence) generates 8 bytes of data. When multiplied by the approximately 10,000 events collected for a single sample, an FCS data file typically contains 80 kb of data. Once a data file has been saved, cell populations can be displayed in several different formats. A single parameter such as FSC or FITC (FL1) can be displayed as a singleparameter histogram, where the horizontal axis represents the parameter s signal value in channel numbers and the vertical axis represents the number of events per channel number (Figure 5-1). Each event is placed in the channel that corresponds to its signal * Data file standard for flow cytometry. Data File Standards Committee of the Society for Analytical Cytology. Cytometry. 1990;11(3):
30 Introduction to Flow Cytometry: A Learning Guide value. Signals with identical intensities accumulate in the same channel. Brighter signals are displayed in channels to the right of the dimmer signals. Two parameters can be displayed simultaneously in a plot. One parameter is displayed on the x-axis and the other parameter is displayed on the y-axis. Three-dimensional data can also be viewed where the x- and y-axes represent parameters and the z-axis displays the number of events per channel (Figure 5-1). Histogram 2-D plot 3-D plot CellQuest 3-D plot Attractors Figure 5-1 Graphic representations of flow cytometric data Review Questions: Data Collection and Display 1 What does the horizontal axis in a histogram represent? 2 A dot plot can be used to display parameters. 3 What does the z-axis on a 3-D plot represent in CellQuest software? 28
31 Chapter 5: Data Analysis 5.2 Gating A subset of data can be defined through a gate. A gate is a numerical or graphical boundary that can be used to define the characteristics of particles to include for further analysis. For example, in a blood sample containing a mixed population of cells, you might want to restrict your analysis to only the lymphocytes. Based on FSC or cell size, a gate can be set on the FSC vs SSC plot to allow analysis only of cells the size of lymphocytes. The resulting display would reflect the fluorescence properties of only the lymphocytes (Figure 5-2). ungated data gated data Figure 5-2 Use of gating to restrict analysis to one population Review Questions: Gating 1 A gate can be used to restrict the analysis to a specific population within the sample. T F 29
32 Introduction to Flow Cytometry: A Learning Guide 5.3 Data Analysis for Subsetting Applications Data analysis consists of displaying the data from a list-mode file in a plot, then measuring the distribution of the events within the plots. As mentioned earlier, several types of plots can be used to present the data. Data can also be subdivided by gating upon specific populations. For example, in the dot plot shown in Figure 5-3, a gate is drawn around the population of interest, which in this case is the lymphocytes. A gate or a region is a boundary drawn around a subpopulation to isolate events for analysis or sorting. Figure 5-3 Dot plot with a gate encompassing the lymphocyte population Data for events within this gate can then be displayed in subsequent plots. In the examples that follow, you will see different ways to analyze fluorescence data from events in this gate to determine the percentages of various subpopulations (subsets) present. You can make a single-parameter histogram plot with histogram markers, a twoparameter dot plot with a quadrant marker, a two-parameter dot plot with regions, and three-dimensional plots. You can also create statistics and export the results that are associated with these plots to a spreadsheet. 30
33 Chapter 5: Data Analysis A histogram allows you to view a single parameter against the number of events. A subclass control is used to determine where the markers will be placed. Histogram markers are used to specify a range of events for a single parameter (Figure 5-4). On the first histogram, marker M1 is placed around the negative peak of the subclass control. Marker M2 is placed to the right of M1 to designate positive events. The second histogram shows events from a CD3 FITC sample. Figure 5-4 Histograms of subclass control (NORM001) and CD3 FITC (NORM002) with histogram markers M1 and M2 Figure 5-5 shows 619 events in M1 and 2272 in M2. To find out statistical percentages of the negatives and the positives, compare the event count with the gated events. There are 6000 events in the data file, but 2891 events found inside the lymphocyte gate. We want the percentage of lymphocytes that are CD3 positive, so we would look at the %Gated for M2: 2272/2891 = 78.59%. Figure 5-5 Histogram statistics 31
34 Introduction to Flow Cytometry: A Learning Guide A dot plot provides a two-parameter display of data. Each dot represents one or more events. The first dot plot in Figure 5-6 is the isotype or subclass control. A subclass control is used to determine where the quadrant markers will be placed. A quadrant marker divides two-parameter plots into four sections to distinguish populations that are considered negative, single positive, or double positive. The lower-left quadrant displays events that are negative for both parameters. The upper-left quadrant contains events that are positive for the y-axis parameter (CD19 PE) but negative for the x-axis (CD3 FITC) parameter. The lower-right quadrant contains events that are positive for the x-axis parameter (CD3 FITC) but negative for the y-axis (CD19 PE) parameter. The upper-right quadrant contains events that are positive for both parameters (CD19 + /CD3 + ), or double positive. Figure 5-6 Dot plots of subclass control (NORM001) and CD3 FITC/CD19 PE (NORM002) with quadrant markers To find out the percentages of CD19 + /CD3 - lymphocytes, look at the %Gated of the upper left (UL) quadrant divided by gated events (Figure 5-7): 296/2839 = 10.43%. Figure 5-7 Quadrant statistics 32
35 Chapter 5: Data Analysis An alternative way to get statistics is to create regions around the populations instead of using a quadrant marker. You can create differently shaped regions (Figure 5-8); then use region statistics to find out the percentages of specific populations. In Figure 5-9, the %Gated of R4 is the CD3 /CD4 + lymphocytes: 40/2866 = 1.40%. Figure 5-8 Dot plot of CD3 FITC/CD4 PE with four regions Figure 5-9 Region statistics There is a disadvantage to using both of these analysis methods if you have several files to analyze from different donor samples. If you had drawn the regions around populations or created quadrant markers from one data file and then read in another file, it is possible that the populations will fall outside the regions or markers due to sample variability. In this case, you would have to readjust the regions or markers for every file. There is a new analysis method available to avoid this situation. This innovative technology is called cluster analysis. BD MultiSET TM and Attractors TM software use cluster analysis to analyze data. In these software programs, regions shift their positions to encompass clusters from one data file to the next (Figure 5-10). 33
36 Introduction to Flow Cytometry: A Learning Guide Figure 5-10 Two-parameter plot in Attractors software with ellipsoid regions before and after analysis 5.4 Data Analysis for Other Applications The analysis methods illustrated so far are used for estimating percentages of discrete populations. This is why you only used the %Gated and %Total statistics. If you were analyzing a cloned cell line to determine if it were positive for a particular molecule, you would most likely not use percentages. Since a cloned cell line consists of a single population, in most cases it would either be 100% negative or 100% positive. It s also possible that it would express low amounts of the molecule in question. In that case, it would still be positive but dim. For these cases where you want to compare fluorescence intensities and measure the degree of positivity, you would compare the geometric means or medians of the subclass control data versus the sample data. If the sample data statistic is greater than that of the subclass control data by some limit set by the user, it would be considered positive. The greater the difference between the two, the more molecules are expressed per cell and the more positive, or brighter the population. The dot plot in Figure 5-11 shows a single population by scatter. The histogram overlays show data from a subclass control and staining with two different antibodies. The geometric mean for each histogram is respectively 2, 5, and 20 (left to right). In this example, the user would decide if sample data in the middle histogram was positive. 34
37 Chapter 5: Data Analysis Figure 5-11 Analysis of a cell line Besides being used to measure positivity, geometric means or medians can be used to estimate the quantity of molecules (ligands) expressed per cell. Specialized software programs such as QuantiCALC use the median in conjunction with data from a standard curve to calculate the number of antibodies bound per cell. An example is shown in Figure 5-12; circles on the y-axis indicate information from a standard curve. This information can be used to estimate the number of ligands per cell. information from a standard curve Figure 5-12 QuantiCALC estimate of number of antibodies bound per cell * * Data provided courtesy of Dr. David Miller, Community Blood Centers of South Florida. 35
38 Introduction to Flow Cytometry: A Learning Guide For DNA content analysis, another specialized software program such as ModFit LT can be used. Because the populations that represent a DNA histogram (G 0 /G 1, S, and G 2 +M) are not discrete (Figure 5-13), special algorithms are used. The area under the curve is integrated; then the percentages of each population present are calculated. Figure 5-13 DNA histogram Review Questions: Data Analysis 1 Why would you use a two-dimensional dot plot rather than a histogram? 2 Referring to Figure 5-4 and Figure 5-5: What percentage of the lymphocytes is NOT CD3 +? What percentage of the total events is CD3 +? 3 The population in the LL quadrant is the double-positive population. T F 4 Referring to Figure 5-6 and Figure 5-7, what is the percentage of lymphocytes that are CD3 + /CD19? 36
39 Chapter 5: Data Analysis 5 Regions can only be drawn as rectangles. T F 6 What is the disadvantage of using regions to analyze several files? 7 What kind of data analysis avoids the problem of shifting groups of cell populations? 8 What statistics are used to measure degree of positivity and for quantitation studies? 37
40 Introduction to Flow Cytometry: A Learning Guide 38
41 6 Sorting 6.1 Sorting In most applications, after a particle exits the laser beam, it is sent to waste. Sorting allows us to capture and collect cells of interest for further analysis. Once collected, the cells can be analyzed microscopically, biochemically, or functionally. Not all benchtop flow cytometers are equipped with a sorting feature; however, they can be upgraded to perform this function. To sort particles or cells, the cytometer first needs to identify the cells of interest, then separate out the individual cells. Once the population of interest has been identified on a data acquisition plot, a region is drawn around that population. A logical gate is created from the regions. This gate is then loaded into the cytometer s software as the sort gate. The sort gate identifies cells of interest to be sorted out of the stream. Different cytometers have different methods of capturing a particle of interest. The FACSCalibur system, a benchtop analyzer, uses a mechanical device called a catcher tube to sort cells. This catcher tube is located in the upper portion of the flow cell. It moves in and out of the sample stream to collect a population of desired cells at a rate of up to 300 cells per second. As a cell passes through the laser beam, the FACSCalibur electronics system, using the sort gate characteristics, quickly determines if the cell is a target. The target cell is captured according to the preselected sort mode. A sort mode is the criteria of capturing the target cells relative to accurate cell count or purity. Because laser 39
42 Introduction to Flow Cytometry: A Learning Guide alignment and stream velocity are fixed, the time it takes for desired cells to travel from the laser intercept to the catcher tube is constant. When the decision is made to capture the target cell, the electronics system waits for a fixed period of time to allow the cell to reach the catcher tube. It then triggers the catcher tube to swing into the sample stream to capture the cell. Figure 6-1(left) shows the catcher tube in its resting position in the sheath stream. Figure 6-1(right) shows the catcher tube positioned in the sample core stream ready to capture a target cell. waste waste sorted cell collection catcher s heath sheath Figure 6-1 (Left) catcher tube in sheath stream; (right) catcher tube in sample stream The FACSVantage SE, a stream-in-air flow cytometer, isolates a cell of interest by vibrating the entire stream. The sample stream vibrates along its axis and breaks up into drops. The distance between drops is fixed. When the sheath velocity and the vibration speed of the nozzle tip are constant, the pattern of drop formation is fixed. With the fixed drop formation, the FACSVantage SE is able to calculate the distance between the drops precisely, which allows for the isolation of individual cells. The FACSVantage SE applies a voltage charge to drops containing a cell that meets the predefined sorting criteria. Positively and negatively charged plates are present on either side of the vibrating stream. As the charged drops pass by the charged plates, the droplets are deflected to the collection tubes, depending on the droplet s charge polarity (Figure 6-2 on page 41). 40
43 Chapter 6: Sorting charging electrode sample injection tube sheath tube vent tube deflection plates waste collection collection tubes Figure 6-2 Sorting components on the FACSVantage SE Review Questions: Sorting 1 To separate individual cells, the liquid stream of the FACSVantage SE along its axis. 2 How many cells are contained in each sorted drop on the FACSVantage SE? 3 What is the name of the device used by the FACSCalibur to capture sorted cells? 4 What does the FACSVantage SE use to attract or repel charged particles? 41
44 Introduction to Flow Cytometry: A Learning Guide 42
45 7 Lasers and Alignment 7.1 How Lasers Work LASER is an acronym for light amplification by stimulated emission of radiation. Light is generated in the following manner. Gas lasers consist of cylinders, or plasma tubes, filled with an inert gas such as argon. The inert gas is ionized by a high-voltage electrical pulse, so that the electrons in the ionized gas atoms absorb energy and move to a higher energy state. As the excited electrons return to the ground state, all at once or in several rapid steps, they give off photons with wavelengths specific to each transition level. Optics at each end of the plasma tube reflect photons back and forth through the tube. These photons interact with other exited electrons, resulting in the release of more photons that are identical in wavelength, phase, and direction. As each photon is capable of stimulating more photons, light amplification or lasing occurs. Brewster windows seal each end of the plasma tube and transmit the light in one plane of polarization. An electromagnet placed around the plasma tube creates a magnetic field that compresses the electrons towards the center of the tube. This prevents the electrons from hitting the sides of the tube, increasing the current density and causing greater stimulated emission of light. 43
46 Introduction to Flow Cytometry: A Learning Guide A small percentage of this light is transmitted through the optic at the front (output coupler) of the tube. This transmitted light is the laser beam. The wavelength of the primary laser beam in benchtop flow cytometers is fixed at 488 nm by placing a filter in front of the output coupler (Figure 7-1). A secondary laser, the red-diode laser, is available for the benchtop analyzer. This is a solid-state laser that emits a 635-nm wavelength. The wavelength of the laser beam in stream-in-air cytometers is selected by adjusting the high reflector optic position to reflect just one of the refracted wavelengths of light traveling through the prism. Only the selected wavelength will be reflected between the output coupler and the high reflector. The output of the laser is monitored and maintained at a constant level through the use of feedback circuitry. Figure 7-1 Generation of laser light in a gas laser 7.2 Laser Alignment In the flow cytometer, the laser beam is focused on the sample core. This point of intercept must remain constant. To accomplish this, the laser head is held in a fixed position. The beam passes through an optic that causes the beam shape to be elliptical. The beam next passes through a focusing lens that focuses the beam at the point of intercept between the laser beam and the sample core. In benchtop flow cytometers, the laser is fixed in proper alignment within the cytometer and no operator adjustment is needed. Stream-in-air flow cytometers use removable lasers and laser optics; therefore, operator adjustment is necessary. 44
47 Chapter 7: Lasers and Alignment Review Questions: Lasers and Alignment 1 A(n) laser emits light that has a wavelength of 488 nm. 2 The laser beam shape is elliptical after it travels through the focusing optics. T F 3 What is the wavelength of the red-diode laser in the bench top analyzer? 4 The wavelength of the laser used in the stream-in-air flow cytometer can be selected by adjusting the optic position at the rear of the laser. 45
48 Introduction to Flow Cytometry: A Learning Guide 46
49 Answer Key Review Questions: Overview on page 7 1 Relative size, relative granularity or internal complexity, and relative fluorescence intensity 2 A laser 3 The fluidics, the optics, and the electronics 4 A single cell suspension 5 Sample core 6 Scattered light and fluorescence 7 Lenses 8 The amount of light striking them 9 True 10 True 47
50 Introduction to Flow Cytometry: A Learning Guide Review Questions: Fluidics on page 12 1 To position the sample core in the center of the laser beam 2 The width and the positioning of the stream 3 One 4 Sheath fluid; flow chamber 5 Hydrodynamic focusing 6 The sample pressure regulator 7 LO, MED, and HI 8 Increases, width 9 A lower flow rate 10 True 11 True Review Questions: Light Scatter on page 15 1 When a particle deflects incident laser light 2 The cell membrane, nucleus, and internal granular components 3 Forward-scattered light (FSC) 4 Cell surface area or size 5 Side-scattered light (SSC) 6 Granularity or internal complexity 7 FSC and SSC 48
51 Answer Key Review Questions: Fluorescence on page 18 1 Photons of light 2 Absorption spectra 3 Emission spectrum 4 The argon-ion laser nm 6 FITC and PE 7 Antigenic surface markers Review Questions: Optical Bench on page 21 1 The excitation and collection optics 2 The laser beam and the sample stream 3 True 4 Flow cell, sample stream Review Questions: Optical Filters on page 23 1 Limit the range of wavelengths reaching the detector and Beam splitter 4 Shortpass, longpass 49
52 Introduction to Flow Cytometry: A Learning Guide Review Questions: Signal Detection on page 25 1 Photodiode 2 Photomultiplier tubes 3 True 4 Converts voltage pulses to channel numbers Review Questions: Threshold on page 26 1 Dust, debris, or electronic noise 2 False Review Questions: Data Collection and Display on page 28 1 The parameter s signal value in channel numbers 2 Two 3 The number of events per channel Review Questions: Gating on page 29 1 True 50
53 Answer Key Review Questions: Data Analysis on page 36 1 A dot plot allows you to view a two-parameter display of data. Each event has a position in the plot according to its channel values for both parameters. Histograms can only display data of one parameter in relationship to the number of events % False % 5 False 6 As you read in one data file to the next in the plot, if the regions are created tightly around the population, the subsequent populations could fall outside of the regions because of sample variability. 7 Cluster analysis 8 Geometric mean or median Review Questions: Sorting on page 41 1 Vibrates 2 One 3 Catcher tube 4 Deflection plates Review Questions: Gating on page 29 1 True 51
54 Introduction to Flow Cytometry: A Learning Guide Review Questions: Lasers and Alignment on page 45 1 Argon 2 True nm 4 High reflector 52
Introduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
Introduction to Flow Cytometry
Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer
MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:
INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals
These particles have something in common
These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative
Flow cytometry basics fluidics, optics, electronics...
Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence
INSIDE THE BLACK BOX
FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben
Flow Cytometry A Basic Overview
Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,
Introduction to Flow Cytometry
Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: [email protected] BDBiosciences Application Examples
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
Flow Cytometry. flow cytometer DNA apoptosis ph
Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal
BD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement
What is Flow Cytometry? Flow Cytometry Basic Principle and Applications BD Biosciences Daisy Kuo ([email protected]) Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements are made on cells,
CyAn : 11 Parameter Desktop Flow Cytometer
CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
COMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
APPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
Flow Cytometry Protocols
METHODS IN MOLECULAR BIOLOGY TM TM Volume 263 Flow Cytometry Protocols SECOND EDITION Edited by Teresa S. Hawley Robert G. Hawley 1 Flow Cytometry An Introduction Alice L. Givan Summary A flow cytometer
UNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a
Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.
Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating
Principles of Flowcytometry
Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal
Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...
Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and
Three Rules for Compensation Controls
From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for
Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer
Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
FACSCalibur System User s Guide
11-10823-02 Rev. A FACSCalibur System User s Guide 02-61760-02 August, 1996 Becton Dickinson Immunocytometry Systems 2350 Qume Drive San Jose, CA 95131-1807 Ordering Information (800) 223-8226 Customer
FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed
FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows
Let the BD FACSAria II accompany your best performance ever
Let the BD FACSAria II accompany your best performance ever Meet the new BD FACSAria II high-speed, fixed-alignment cuvette cell sorter. The first generation BD FACSAria system brought the complex world
Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T
Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep
Instrument Characterization and Performance Tracking for Digital Flow Cytometers
Instrument Characterization and Performance Tracking for Digital Flow Cytometers BD Biosciences Cytometer Set-up & Tracking (CS&T) System For Research Use Only. Not for use in diagnostic or therapeutic
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor
No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes
BD CellQuest Pro Software Analysis Tutorial
BD CellQuest Pro Analysis Tutorial This tutorial guides you through an analysis example using BD CellQuest Pro software. If you are already familiar with BD CellQuest Pro software on Mac OS 9, refer to
Standardization, Calibration and Quality Control
Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic
Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size
Introduction to Flow Cytometry -- BD LSR II Daisy Kuo 郭 正 佼 Assistant Product Manager BD Biosciences [email protected] What is Flow Cytometry? Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements
Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113
Flow Cytometry 101 Jodi Moore, PhD Jan 2012 Goldenson 113 Origins:? Dye Chemistry 1940-1980 s Electronics Microscopy Automated Clinical Cytology ID Apps Computers Cytometry vs. Flow Cytometry Microspectrophotometry
Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction
Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline
Immunophenotyping peripheral blood cells
IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done
Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility
10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa
Becton-Dickinson FACSCalibur Flow Cytometer Instructions
Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If
Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835
Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that
Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and
Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry
MACSQuantify Software guide. Basics of flow cytometric analysis and software guide. Version 1 EN Original instructions
MACSQuantify Software guide Basics of flow cytometric analysis and software guide Version 1 EN Original instructions Miltenyi Biotec GmbH Friedrich-Ebert-Straße 68 51429 Bergisch Gladbach Germany Phone:
BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin
BD LSR II and FACSDiVa Software, Basel Dr. Norbert Leclere,, Berlin Overview Electronics Covers Fluidics Connectors Controlpanel Sample Port On/Off The Control Panel Ease of Use The basic procedures to
Experiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee
January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch
CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping
CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Molecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose
Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:
Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1
78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction
Boundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
Compensation Controls Data Visualization Panel Development Gating Quantum Dots
Compensation Controls Data Visualization Panel Development Gating Quantum Dots FITC Single Stain Control FITC PE Argon Laser FL1 FL2 450 500 550 600 FITC Compensation Control Unwanted signal detected in
Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc.
Compensation Basics C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. 2003 1 Intrinsic or Autofluorescence p2 ac 1,2 c 1 ac 1,1 p1 In order to describe how the general process of signal cross-over
Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract
arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.
Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Some limitations (disadvantages) of NSOM A trade-off
Compensation in Flow Cytometry
Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome
A Basic Guide. Institute of Medical Biology, University of Southern Denmark. Ungated Gate 1 (= R1) Gate 4 (=R4) Gate3 (=R3) Gate 2 (= R2) R3 R4
Flow Cytometry A Basic Guide R1 Ungated Gate 1 (= R1) R2 R3 R4 Gate 2 (= R2) Gate3 (=R3) Gate 4 (=R4) Graham Leslie, PhD Institute of Medical Biology, University of Southern Denmark Graham Leslie, PhD
Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES
BIODIVERSITY I BIOL1051 Microscopy Professor Marc C. Lavoie [email protected] MAJOR FUNCTIONS OF MICROSCOPES MAGNIFY RESOLVE: => INCREASE CONTRAST Microscopy 1. Eyepieces 2. Diopter adjustment
How to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences
FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data Monisha Sundarrajan, PhD BD Biosciences 23 13795 00 Agenda Introduction Overview of BD CBA bead based immunoassays FCAP
Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH))
Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH)) October 14, 2013 1 Licensing Under the Artistic License, you are
123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.
Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.
Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers
An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,
CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles
CyFlow SL Microbiology Detection and Analysis of Microorganisms and Small Particles 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
Cell Cycle Tutorial. Contents
Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6
Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:
Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T
powerful. Intuitive. Customizable. The new star of benchtop flow cytometry
powerful. Intuitive. Customizable. The new star of benchtop flow cytometry About ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
FACS Laboratory. BD LSRs Operator Training. Training Guide. http://london-research-institute.org.uk/technologies/facs
FACS Laboratory [email protected] Training Guide http://london-research-institute.org.uk/technologies/facs BD LSRs Operator Training LRI FACS lab 1 Components 1.1 Basic parts overview.pg3-6 Overview
Below is a list of things you should be aware of before you schedule your sort.
Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will
Science is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening
Zecotek S Light Projection Network Marketing
White Paper Zecotek Visible Fiber Laser Platform Enabling the future of laser technology Zecotek Photonics Inc. (TSX- V: ZMS; Frankfurt: W1I) www.zecotek.com is a Canadian photonics technology company
Institut für Experimentelle Immunologie und Bildgebung
Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check
Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry
Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry by Debby R. Walser-Kuntz Biology Department Carleton College, Northfield, MN Taylor goes in to see her doctor, Dr. Chavez,
2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
PUMPED Nd:YAG LASER. Last Revision: August 21, 2007
PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow
Tutorial 4.6 Gamma Spectrum Analysis
Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
Bacterial Detection and Live/Dead Discrimination by Flow Cytometry
A P P L I C A T I O N N O T E Bacterial Detection and Live/Dead Discrimination by Flow Cytometry M I C R O B I A L C Y T O M E T R Y BD FACSCalibur, BD FACSort Flow Cytometers Introduction Accurate determination
Lab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations
Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations Supervisors: Sissela Broos [email protected] tel: 222 96 78 Niclas Olsson [email protected]
Copyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
BD FACSAria Flow Cytometer. The first high-speed sorter with a fixed-alignment cuvette flow cell from BD Biosciences
BD FACSAria Flow Cytometer The first high-speed sorter with a fixed-alignment cuvette flow cell from BD Biosciences Header From Pioneers to Revolutionaries in Flow Cytometry Instrumentation BD Biosciences
TOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] [email protected] [e-mail] This
Setting up and calibration of a flow cytometer for multicolor immunophenotyping
Journal of Biological Regulators and Homeostatic Agents Setting up and calibration of a flow cytometer for multicolor immunophenotyping J. KRAAN 1, J.W. GRATAMA 1, M. KEENEY 2, J.L. D HAUTCOURT 3 1 Department
BD Trucount Tubes IVD
02/2015 23-3483-07 IVD BD Trucount Tubes For determining absolute counts of leucocytes in blood 25 Tubes Catalog No. 340334 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company.
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this
BD FACSCanto Flow Cytometer. Efficient Easy Powerful
BD FACSCanto Flow Cytometer Efficient Easy Powerful The BD FACSCanto System: Optimal Workflow and Time Management Automated compensation algorithms with optional off-line compensation, streamlined sample
UCHC Guide to CellQuest Pro. FACS Calibur instruments
UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn
Using CyTOF Data with FlowJo Version 10.0.7. Revised 2/3/14
Using CyTOF Data with FlowJo Version 10.0.7 Revised 2/3/14 Table of Contents 1. Background 2. Scaling and Display Preferences 2.1 Cytometer Based Preferences 2.2 Useful Display Preferences 3. Scale and
Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.
Helium-Neon Laser Experiment objectives: assemble and align a 3-mW HeNe laser from readily available optical components, record photographically the transverse mode structure of the laser output beam,
Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial
Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few
GETTING STARTED. 6. Click the New Specimen icon.
BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified