LSR II Users Guide Stanford Shared FACS Facility 06/10/13
|
|
|
- Carmella Gilbert
- 9 years ago
- Views:
Transcription
1 LSR II Users Guide Stanford Shared FACS Facility 06/10/13 A configuration sheet is located on the Website in the Instrument Overview area Do I need to run beads when I collect data using Diva Digital? For each data collection, you need to first collect data on 197 calibration beads. What samples do I need to bring? All users should always bring single stained compensation samples If using 3-color staining: User should bring 4 compensation tubes! Each single stained tube with one fluochrome AND one unstained tube. No PI or viability dye in the single color controls Don t have enough cells for compensation? User can purchase compensation beads from BD or Invitrogen BD CompBeads Anti-Rat/Hamster!for mouse samples BD CompBeads Anti-Mouse for almost all human samples PPE must be worn at all times when in our BSL-2 environment Lab coat, gloves, closed toed shoes, are mandatory, respirator and goggles are available
2 Starting Up the System Allow 30 minutes warm up time for the lasers 1) Turn on the Cytometer. The power switch is located on the lower right corner on the Cytometer. The fluid control button lights green and the Standby button lights amber. 2) SSFF computers are left on at all times. LSR-II 20L Pump System Instructions (Cytek) Fluid mangers are in use on all our LSR II s. Use offers the following advantages: 1) No need to open and fill the metal tank. 2) Stabilizing of sheath flow rates to allow a 6 usec window extension. To use: 1) The pump system turns on automatically with the LSR. However, it does have its own switch as well(on the.lsr II and LSR II.UV It is the blue box labeled Cytek under the LSR with the black lever switch, for the LSR II.2 it is the FACSFlow Supply System). Verify that this is on. The LSR will not work properly otherwise. 2) Make sure there is enough sheath in the cube for your experiment. The cube is on a scale, so you can look at the weight of the cube on the scale, remembering that 1 L weighs 2.2#. 3) If the sheath cube is empty replace with a new one (heavy so lift carefully) from the either the hall or the rear of the room. Note that the cube closure just screws on there is no pressurization of the cube. Be careful not to kink the tubing 4) Empty waste in sink in hallway when full, using the protective face shield, gloves and lab coat. Remember to add bleach to line marked on carboy. 4) Alarm sensors are connected to the cube and the waste, indicating when they need attention if you forget. The LSR won t run samples properly if the pump is turned off. The pump should never be turned off manually!! 5) Please [email protected] with any problems or questions Diva Software Instructions 1)Log into Diva software using BDFacs Diva software icon on desktop Username is Sunet ID/ password starts with a capitol letter 2) Watch for Instrument connected status in bottom of the Cytometer window 3) If CST mismatch dialog box appears select use CST settings Setting Up Experiment 1) Go to Experiment menu New Experiment Select the SSFF tab and choose the
3 newest date available select experiment Click! OK Rename - right-click!rename Use descriptive name and date 2)Edit menu!user Preferences!Select desired defaults. You can deselect save analysis and load data after acquisition to save time(buffer). Calibration 1) Run standard 197 calibration beads (located in rack or freezer) load sample on the SIP and hit RUN and LO buttons on Cytometer panel 2) Go to Beads Specimen (syringe icon) CLICK on + sign 3 ) Click on 197 Beads tube (green arrow! is pointing at the tube) This makes acquisition controls active In Acquisition Dashboard CLICK!Acquire Data 4) Set Events to Display to for calibration 5) Set flow rate to low 200 events per second. After beads appear on plots, use fine tune (black knob) to adjust flow rate to aprox 200 beads/sec 6) Check that the global work sheet is selected. (acquistion must be stopped in order to change the worksheets) 7 ) Follow the onscreen instructions and check the instruments laser delays, area scaling and MFI values are calibrated(located on second worksheet Standards). 8)If the delays are off first check the Current values, they are located on the desktop, use the latest performance shortcut for the most recent CST run. 9) On the first global worksheet check delay window extension a) Confirm the Forward scatter threshold is set to 50% of the forward scatter signal. B ) Check the laser delays by closing the window extension to 0. Open window extension to 6 when done. c) If properly set the area signal for the Green, Red and Violet lasers should drop no more than 50%.(you can use the height value for a quick reference to 100 %signal.) d) Check that Area Scaling values are correct. Area should be greater than or equal to the height within 10%. adjustments in the area scaling section in the laser tab.
4 e) Switch to the standards global worksheet and adjust the PMT voltages (cytometer window -> parameters tab) so the measured values are at or around displayed target values in bold(10%). f) Record data for beads in tube labeled 197 beads after calibration is complete. Experiment Definition 1) Under Experiment menu select New Specimen!Click on + to see tube 001 2) Highlight tube one in new specimen area >click on arrow on left 3) In the Cytometer pane, deselect parameters you will not be using. Do not delete Parameters that may be of interest. If you add back a deleted parameter to the experiment it will come in with different voltages not specified pmt settings by facility. Check cytometer settings for the 197 bead tube to see correct pmt voltage to assign. This may cause problems for compensation calculations In Parameters tab, click on small button to left of parameter name Click delete button (use shift key and highlight for multiple deletions) Repeat for each parameter you're not using 4) Keep forward and side scatter height and width selected 5) Lower forward scatter PMT for typical cells in the parameter tab 6)Lower the threshold for cell population in the threshold tab 7) The data may be viewed compensated or not, but the uncompensated data is written to the database along with the matrix so that other compensations may be used later in FlowJo. Note: For each tube collected in Diva software, uncompensated data is recorded and the compensation matrix is applied as a transform and not affect your data. Compensation All samples must be filtered for sorters 1. Select Experiment>Compensation Setup>Create compensation tubes. This will generate a new specimen in your experiment called Compensation Controls. Within the specimen it makes a tube and a worksheet for each compensation control, including one for unstained cells. 2. Run unstained cells first. Set Events to display to Adjust the FSC and SSC voltages to place the population of interest on scale. 4. Adjust the FSC threshold value to exclude most of the debris without excluding the population of interest.
5 5. Move p1 gate on FSC/SSC plot to include your cells of interest, then right-click and Apply to All Compensation Controls if all your controls are expected to fall in the same FSC v SSC gate. 6. Run each sample in the unstained cell tube area so we can define that all events are on scale, and that each stain is most dominant in its own channel (by a minimum ½ log). Keep in mind that if you lower the voltage below the baseline voltages defined for the instrument you may lose the ability to detect low positive events on that channel. Try to avoid lowering voltages below these target values. It is acceptable to have the negative peak shift above the first log decade in some detectors. 7. Put on your positively stained sample to verify that the positive peaks remain on scale. Lower voltages on those parameters that are off scale (place below 10 x 5). Remember, if you significantly lower the PMT voltage below the baseline settings in order to bring the positive population on scale, then the dim populations might not be easily resolved from the negative populations for that parameter. 8. Run the compensation samples and collect data for each in the appropriate tube. Use Next to move between tubes. For each tube, make sure the integral gate encompasses your positively stained cells. 9. Select Experiment>Compensation setup>calculate compensation. This calculates a compensation matrix. If the calculation is successful, a dialog appears for you to enter a name for the compensation setup. To keep track of compensation setups, include the experiment name, your initials, or the date in the setup name. Error messages that warn you of >100% compensation values indicate one of two things, the voltage settings used are suboptimal and a channel receiving less total photons has been over amplified or that the color combination at the settings you have chosen is not optimal. You can ignore these messages, but it would be best to try increasing or decreasing voltages on the problem channels and rerunning all your comp controls at new settings. You will have to recalculate compensation. If there are still problems, it may indicate that you do not have a good color combination. Cell analysis and Data Acquisition 1. Create a series of plots on the global worksheet that you will need to set up your acquisition. It is best to have at least one axis for each color minimum in case there is an issue with signal levels across samples. 2. Pull up a population hierarchy and make gates as needed for your sort or analysis. 3. High light and Select the first tube in specimen 2 and rename the tube with the proper sample ID 4. Load your tube and record the data 5. Save your analysis template for future use. Click on the analysis template and rename it, then export analysis template 6. Run bleach (blue top 10% bleach) for 2 minutes on high followed by water(black top)for 2 minutes on high. 7. Leave the machine in standby on the low setting with a tube of water on the SIP
6 You must check in your data to protect it Data is deleted monthly if it is not checked in it is gone USB and personal back up devises are not allowed on our machines Discs may be burned when alternatives are needed Data Transfer to FACSData Server When finished with data collection: 1)Select experiment to be exported for CHECKIN 2)Select File!Export!Experiments 3)Data Export Pathway C:\Export 4)Check that the proper experiment is being transferred to the folder area. Click OK. 5)On the Desktop, activate by double clicking!facs Data CHECKIN 6)The appropriate experiment should be displayed and selected. Click OK. The named of the person logged in needs to be selected on the list and the SUNet Id name should be displayed at the bottom. Click! OK. Add address to send the FlowJo summery file to additional people 7)A.jo file for MAC and a.wsp file for PC will be ed. 8)If it asks you something about delayed sessions, click! cancel. Note: The person s name in the first box becomes the owner 9)FlowJo PC users should install the WebStart Version of FlowJo PC. Start FlowJo PC and from the FileMenu select Open and then open your.wsp file. You will be prompted for Username: flowjo Password: Shutting Down the System Perform the following procedures at the end of each day or when required. fluidics cleaning system shutdown Always clean the BD LSR II Cytometer before it is turned off at the end of the day. Proper cleaning will ensure that your Cytometer will function consistently. Cleaning prevents the sample injection tube from becoming clogged and removes dyes that could remain in the tubing, causing carryover. Fluidics Cleaning Follow this procedure immediately after running viscous samples or nucleic acid dyes such as Hoechst, DAPI, propidium iodide (PI), acridine orange (AO), or thiazole orange (TO). It will take approximately 5-10 minutes to complete the shutdown procedure. NOTICE Make sure the sample fine adjust knob is at the middle flow rate position or greater for the entire cleaning procedure. 1 Set the fluid control to RUN. 2. Install a tube containing 1 ml of a bleach solution on the SIP,run for 3-5 minutes. Use our blue tube 1:10 dilution of bleach in DI water as the bleach solution. 3. Repeat steps 2 with DI water. 4. Leave a tube containing no more than 1 ml of DI water on the SIP. A tube with 1 ml of DI should remain on the SIP to prevent salt deposits from forming in the
7 injection tube. The tube also catches back drips from the flow cell. ~ CAUTION Do not leave more than 1 ml of water on the SIP. When the BD LSR II flow cytometer is turned off or left in standby mode, a small amount of fluid will drip back into the sample tube. If there is too much fluid in the tube, it could overflow and affect Cytometer performance. 5. Set the cytometer to standby mode. System Shutdown 1. Turn off the flow cytometer.
8
GETTING STARTED. 6. Click the New Specimen icon.
BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified
Analyzer Experiment setup guide for LSRII and Canto s
Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,
FACSCanto RUO Special Order QUICK REFERENCE GUIDE
FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW
How to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
UCHC Guide to FACS DiVa. LSR II instruments
UCHC Guide to FACS DiVa LSR II instruments Start Up: Turn on the instrument using the Green power button. Fluidics pump power, LSR II-B & C LSR II-B and LSRII-C-> Turn on the fluidics pump and ensure that
Institut für Experimentelle Immunologie und Bildgebung
Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done
Becton-Dickinson FACSCalibur Flow Cytometer Instructions
Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If
UCHC Guide to CellQuest Pro. FACS Calibur instruments
UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn
Attune Acoustic Focusing Cytometer
USER GUIDE Attune Acoustic Focusing Cytometer USER GUIDE Publication Part Number 4453328 Rev. F Revision Date June 2012 Copyright 2012, Life Technologies Corporation. All rights reserved. For Research
BD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin
BD LSR II and FACSDiVa Software, Basel Dr. Norbert Leclere,, Berlin Overview Electronics Covers Fluidics Connectors Controlpanel Sample Port On/Off The Control Panel Ease of Use The basic procedures to
Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835
Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that
Navios Quick Reference
Navios Quick Reference Purpose: The purpose of this Quick Reference is to provide a simple step by step outline of the information needed to perform various tasks on the system. We begin with basic tasks
After completing this module, you will be able to: Describe the BD FACScan system and how it works.
BD FACScan System After completing this module, you will be able to: Describe the BD FACScan system and how it works. Perform instrument startup and shutdown procedures. Describe instrument maintenance
CyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version 6.0.626)
CyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version 6.0.626) *Amended from Sampler Manual created by Zach Bjornson Because CyTOF computer is on a private network, you will be bombarded with
TROUBLESHOOTING FLOW CHART
TROUBLESHOOTING FLOW CHART Revised and updated on January 16, 2003, by Tony Leger and Automation Lab Technology. FACScan(FC)\FACSort(FST) FACScalibur (FCB) Fluidics/Pneumatics troubleshooting guide DCM
1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8
Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms
Attune NxT Acoustic Focusing Cytometer
USER GUIDE Attune NxT Acoustic Focusing Cytometer Catalog Number A24858, A24859, A24860, A24861, A24862, A24863, A24864 Publication Number 100024235 Revision B.0 For Research Use Only. Not for use in diagnostic
WebSphere Business Monitor V6.2 KPI history and prediction lab
Copyright IBM Corporation 2009 All rights reserved IBM WEBSPHERE BUSINESS MONITOR 6.2 LAB EXERCISE WebSphere Business Monitor V6.2 KPI history and prediction lab What this exercise is about... 1 Lab requirements...
FACS-CALIBUR. USER-Guide
FACS Calibur User guide Version 2.1-03/2006 AXS 1 FACS-CALIBUR Version 2.1 USER-Guide Note: This guide does note compensate for a personal introduction from an operator before first usage Draft mainly
COMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...
Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and
Science is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening
SA-9600 Surface Area Software Manual
SA-9600 Surface Area Software Manual Version 4.0 Introduction The operation and data Presentation of the SA-9600 Surface Area analyzer is performed using a Microsoft Windows based software package. The
LEGENDplex Data Analysis Software
LEGENDplex Data Analysis Software Version 7.0 User Guide Copyright 2013-2014 VigeneTech. All rights reserved. Contents Introduction... 1 Lesson 1 - The Workspace... 2 Lesson 2 Quantitative Wizard... 3
BIGPOND ONLINE STORAGE USER GUIDE Issue 1.1.0-18 August 2005
BIGPOND ONLINE STORAGE USER GUIDE Issue 1.1.0-18 August 2005 PLEASE NOTE: The contents of this publication, and any associated documentation provided to you, must not be disclosed to any third party without
Transitioning from TurningPoint 5 to TurningPoint Cloud - LMS 1
Transitioning from TurningPoint 5 to TurningPoint Cloud - LMS 1 A Turning Account is a unique identifier that is used to tie together all software accounts and response devices. A Turning Account is required
Hosting Users Guide 2011
Hosting Users Guide 2011 eofficemgr technology support for small business Celebrating a decade of providing innovative cloud computing services to small business. Table of Contents Overview... 3 Configure
Cell Sorting. S3e Cell Sorter Quick Guide
Cell Sorting S3e Cell Sorter Quick Guide STARTUP Checking Fluidics STARTUP There are three bulk fluid containers (Figure 1) that may require attention before starting the system. If fluidics need to be
1. Installing The Monitoring Software
SD7000 Digital Microphone Monitor Software manual Table of Contents 1. Installing The Monitor Software 1.1 Setting Up Receivers For Monitoring 1.2 Running The Application 1.3 Shutdown 2. The Detail Monitoring
Statgraphics Getting started
Statgraphics Getting started The aim of this exercise is to introduce you to some of the basic features of the Statgraphics software. Starting Statgraphics 1. Log in to your PC, using the usual procedure
MiraCosta College now offers two ways to access your student virtual desktop.
MiraCosta College now offers two ways to access your student virtual desktop. We now feature the new VMware Horizon View HTML access option available from https://view.miracosta.edu. MiraCosta recommends
BCSD WebMail Documentation
BCSD WebMail Documentation Outlook Web Access is available to all BCSD account holders! Outlook Web Access provides Webbased access to your e-mail, your calendar, your contacts, and the global address
Authorware Install Directions for IE in Windows Vista, Windows 7, and Windows 8
Authorware Install Directions for IE in Windows Vista, Windows 7, and Windows 8 1. Read entire document before continuing. 2. Close all browser windows. There should be no websites open. If you are using
Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract
arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents
User guide. Business Email
User guide Business Email June 2013 Contents Introduction 3 Logging on to the UC Management Centre User Interface 3 Exchange User Summary 4 Downloading Outlook 5 Outlook Configuration 6 Configuring Outlook
McAfee Endpoint Encryption Reporting Tool
McAfee Endpoint Encryption Reporting Tool User Guide Version 5.2.13 McAfee, Inc. McAfee, Inc. 3965 Freedom Circle, Santa Clara, CA 95054, USA Tel: (+1) 888.847.8766 For more information regarding local
Installing VPN for PC v1.3
Follow these instructions to access your Public/Private folders from home. Please Note: If you are having connectivity issues after installing or upgrading VPN, please refer to Troubleshooting Connectivity
You may have been given a download link on your trial software email. Use this link to download the software.
BackupVault / Attix5 Server Quickstart Guide This document takes about 5 minutes to read and will show you how to: Download the software Install the Attix5 Professional Backup software Backup your files
How to install and use the File Sharing Outlook Plugin
How to install and use the File Sharing Outlook Plugin Thank you for purchasing Green House Data File Sharing. This guide will show you how to install and configure the Outlook Plugin on your desktop.
Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.
1234 Hach Hall 515-294-5805 www.cif.iastate.edu Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V. Location: Contact: 1240 Hach Hall Steve Veysey, 1234 Hach Hall Safety All researchers working
Getting Started with BD FACSDiva Software
Getting Started with BD FACSDiva Software For In Vitro Diagnostic Use bdbiosciences.com Part No. 642213 Rev. A June 2007 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 USA Tel (877) 232-8995 Fax
CyAn : 11 Parameter Desktop Flow Cytometer
CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on
WebSphere Business Monitor V6.2 Business space dashboards
Copyright IBM Corporation 2009 All rights reserved IBM WEBSPHERE BUSINESS MONITOR 6.2 LAB EXERCISE WebSphere Business Monitor V6.2 What this exercise is about... 2 Lab requirements... 2 What you should
OPERATION MANUAL. MV-410RGB Layout Editor. Version 2.1- higher
OPERATION MANUAL MV-410RGB Layout Editor Version 2.1- higher Table of Contents 1. Setup... 1 1-1. Overview... 1 1-2. System Requirements... 1 1-3. Operation Flow... 1 1-4. Installing MV-410RGB Layout
To successfully initialize Microsoft Outlook (Outlook) the first time, email settings need to be verified.
TO: UAN CLIENTS FROM: UAN STAFF DATE: OCTOBER 8, 2008 SUBJECT: Steps for Initial Setup of Microsoft Outlook To successfully initialize Microsoft Outlook (Outlook) the first time, email settings need to
Passport installation Windows 8 + Firefox
Passport installation Windows 8 + Firefox OS Version Windows 8 System Type 32-bit and 64-bit Browser name & version Firefox 23 Before you begin Check you have Anti Virus software installed on your computer
Working with SQL Server Integration Services
SQL Server Integration Services (SSIS) is a set of tools that let you transfer data to and from SQL Server 2005. In this lab, you ll work with the SQL Server Business Intelligence Development Studio to
Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer
Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the
GIAC Introduction to Security Fundamentals. Laptop and External Drive Configuration Guide Version 1.2 SEC301
GIAC Introduction to Security Fundamentals Laptop and External Drive Configuration Guide Version 1.2 SEC301 SEC301 - Hands-on Exercises Addendum Page 1 Table of Contents Background... 3 Ensure that the
Training Manual. Version 6
Training Manual TABLE OF CONTENTS A. E-MAIL... 4 A.1 INBOX... 8 A.1.1 Create New Message... 8 A.1.1.1 Add Attachments to an E-mail Message... 11 A.1.1.2 Insert Picture into an E-mail Message... 12 A.1.1.3
Disabling Microsoft SharePoint in order to install the OneDrive for Business Client
Disabling Microsoft SharePoint in order to install the OneDrive for Business Client If you try to setup and sync your OneDrive online documents with the client software and Microsoft SharePoint opens,
BD CellQuest Pro Software Analysis Tutorial
BD CellQuest Pro Analysis Tutorial This tutorial guides you through an analysis example using BD CellQuest Pro software. If you are already familiar with BD CellQuest Pro software on Mac OS 9, refer to
Quick Start Guide. DVR DS-7200HWI-SH Series DVR. www.hikvision.com. First Choice For Security Professionals
Quick Start Guide DVR DS-7300HWI-SH Series DVR DS-7200HWI-SH Series DVR NOTE: For more detailed information, refer to the User s Manual on the CD-ROM. You must use your PC or MAC to access the files. www.hikvision.com
User s Manual. Management Software for ATS
ATS Monitor User s Manual Management Software for ATS Table of Contents 1. ATS Monitor Overview... 2 2. ATS Monitor Install and Uninstall... 2 2.1. System Requirement... 2 2.2. Software Install... 2 2.3.
Connecting to Remote Desktop Windows Users
Connecting to Remote Desktop Windows Users How to log into the College Network from Home 1. Start the Remote Desktop Connection For Windows XP, Vista and Windows 7 this is found at:- Star t > All Programs
TAMUS Terminal Server Setup BPP SQL/Alva
We have a new method of connecting to the databases that does not involve using the Texas A&M campus VPN. The new way of gaining access is via Remote Desktop software to a terminal server running here
UniFLOW Print and Copy Instructions for Students
UniFLOW Print and Copy Instructions for Students Getting Started, Setting up your PIN Page 2 Purchasing Credits for printing and Photocopying Page 4 How to use the MFD Page 8 Photocopying Page 10 Printing
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline
Virtual Office Remote Installation Guide
Virtual Office Remote Installation Guide Table of Contents VIRTUAL OFFICE REMOTE INSTALLATION GUIDE... 3 UNIVERSAL PRINTER CONFIGURATION INSTRUCTIONS... 12 CHANGING DEFAULT PRINTERS ON LOCAL SYSTEM...
USER MANUAL (PRO-CURO LITE, PRO & ENT) [SUPPLIED FOR VERSION 3]
Pro-curo Software Ltd USER MANUAL (PRO-CURO LITE, PRO & ENT) [SUPPLIED FOR VERSION 3] CONTENTS Everyday use... 3 Logging on... 4 Main Screen... 5 Adding locations... 6 Working with locations... 7 Duplicate...
Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab
Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Room 213 Beckman Coulter DTX 880 Multimode Detector Information The Beckman Coulter DTX 880 Multimode Detector is
ProjectWise Explorer V8i User Manual for Subconsultants & Team Members
ProjectWise Explorer V8i User Manual for Subconsultants & Team Members submitted to Michael Baker International Subconsultants & Team Members submitted by Michael Baker International ProjectWise Support
ANYWHERE POLLING - POLLING WITH A QUESTION LIST
Anywhere Polling - Polling with a Question List 1 ANYWHERE POLLING - POLLING WITH A QUESTION LIST Before Class This section covers question lists and participant lists. Question lists and participant lists
16.4.3 Lab: Data Backup and Recovery in Windows XP
16.4.3 Lab: Data Backup and Recovery in Windows XP Introduction Print and complete this lab. In this lab, you will back up data. You will also perform a recovery of the data. Recommended Equipment The
TM Online Storage: StorageSync
TM Online Storage: StorageSync 1 Part A: Backup Your Profile 1: How to download and install StorageSync? Where to download StorageSync? You may download StorageSync from your e-storage account. Please
Fleet Manager Quick Guide (Non Maintenance Mode)
Fleet Manager Quick Guide (Non Maintenance Mode) Launch Fleet Manager: Open the Fleet Manager Application by: 1. Double clicking the icon located on the desktop - or 2. Via Start > Programs > MobileView
Business Objects InfoView Quick-start Guide
Business Objects InfoView Quick-start Guide Last Modified: 10/28/2015 The latest PDF version of this document can be found at: http://www.calpolycorporation.com/docs/finance/boeinfoviewquickstart.pdf What
BD LSRFortessa Cell Analyzer User s Guide
BD LSRFortessa Cell Analyzer User s Guide For Research Use Only bdbiosciences.com 23-11093-00 Rev. A 3/2010 Becton, Dickinson and Company BD Biosciences San Jose, CA 95131 Tel 877.232.8995 Fax 800.325.9637
Laser Cutter User Manual
Laser Cutter User Manual frequently asked questions... the laser is cutting weird! it s cutting a thick line or not cutting through at all! Table Of Contents Section 1: Materials Guide 1.1 Can I Cut this?
Windows XP Pro: Basics 1
NORTHWEST MISSOURI STATE UNIVERSITY ONLINE USER S GUIDE 2004 Windows XP Pro: Basics 1 Getting on the Northwest Network Getting on the Northwest network is easy with a university-provided PC, which has
Quick Installation and Setup for WinPower Software
Quick Installation and Setup for WinPower Software A. On the WinPower Installation menu, simply follow on-screen instructions and key in the software serial number. The installation will be automatically
MICROSOFT OUTLOOK 2011 GETTING STARTED AND HELP RESOURCES
MICROSOFT OUTLOOK 2011 GETTING STARTED AND HELP RESOURCES Lasted Edited: 2012-07-10 1 Introduction... 4 Getting Started... 4 Tour of the Outlook 2011 Interface... 4 Start Outlook 2011... 5 Configure E-mail
Microsoft Excel Basics
COMMUNITY TECHNICAL SUPPORT Microsoft Excel Basics Introduction to Excel Click on the program icon in Launcher or the Microsoft Office Shortcut Bar. A worksheet is a grid, made up of columns, which are
Mariemont City Schools
Mariemont City Schools Citrix Virtual Desktop Environment Citrix is a virtual desktop system that allows users to access their Mariemont Windows 7 desktop from anywhere with an Internet connection. Once
Cell Cycle Tutorial. Contents
Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6
WebSphere Business Monitor V7.0 Business space dashboards
Copyright IBM Corporation 2010 All rights reserved IBM WEBSPHERE BUSINESS MONITOR 7.0 LAB EXERCISE WebSphere Business Monitor V7.0 What this exercise is about... 2 Lab requirements... 2 What you should
How to Connect to Berkeley College Virtual Lab Using Windows
How to Connect to Berkeley College Virtual Lab Using Windows Minimum Requirements Create and Save a Remote Desktop Services Connection Connecting to a Remote Desktop Services session Copy Files between
F9 Integration Manager
F9 Integration Manager User Guide for use with QuickBooks This guide outlines the integration steps and processes supported for the purposes of financial reporting with F9 Professional and F9 Integration
Steps for using MagicConnect
Getting Started Getting Started MagicConnect is a remote access service that allows you safe remote access to a target device from your remote device. What is a remote device? It is a Windows device, iphone/ipad,
First Time On-Campus Remote Desktop Connection ipad Edition
First Time On-Campus Remote Desktop Connection ipad Edition General: This guide is a step-by-step guide for configuring and using an ipad to remotely connect to a work computer from on campus. This guide
WAMLocal. Wireless Asset Monitoring - Local Food Safety Software. Software Installation and User Guide BA/WAM-L-F
Wireless Asset Monitoring - Local Food Safety Software BA/WAM-L-F Software Installation and User Guide System Overview The BAPI Wireless Asset Monitoring Local (WAM Local) Software receives temperature
RP Pocket PC Scanner Reference Manual For PPT8800
RP Pocket PC Scanner Reference Manual For PPT8800 Table of Contents Overview...2 Installation...2 Setting up your Pocket PC...3 Terminal Services Client...6 Scanning Barcodes...8 Loading Scanner Sessions
Introduction. POP and IMAP Servers. MAC1028 June 2007
MAC1028 June 2007 Getting Started with Thunderbird 2.0 For Macintosh OS X Author: John A. Montgomery Adapted to OS X by: Peter Lee Revised by Mitchell Ochi and Deanna Pasternak Introduction...1 POP and
Fleet Manager II. Operator Manual
Fleet Manager II Operator Manual Table of Contents Table of Contents Table of Contents 2 About this Publication 4 Trademarks 5 About Fleet Manager II 6 Contact BW Technologies by Honeywell 7 Getting Started
Task Force on Technology / EXCEL
Task Force on Technology EXCEL Basic terminology Spreadsheet A spreadsheet is an electronic document that stores various types of data. There are vertical columns and horizontal rows. A cell is where the
Hamline University Administrative Computing Page 1
User Guide Banner Handout: BUSINESS OBJECTS ENTERPRISE (InfoView) Document: boxi31sp3-infoview.docx Created: 5/11/2011 1:24 PM by Chris Berry; Last Modified: 8/31/2011 1:53 PM Purpose:... 2 Introduction:...
Mapping ITS s File Server Folder to Mosaic Windows to Publish a Website
Mapping ITS s File Server Folder to Mosaic Windows to Publish a Website April 16 2012 The following instructions are to show you how to map your Home drive using ITS s Network in order to publish a website
Using ELM Reports in WhatsUp Gold. This guide provides information about configuring ELM reports in WhatsUp Gold v15.0
Using ELM Reports in WhatsUp Gold This guide provides information about configuring ELM reports in WhatsUp Gold v15.0 Using Event Log Management (ELM) Reports in WhatsUp Gold In This Chapter Event Log
Using an Edline Gradebook. EGP Teacher Guide
Using an Edline Gradebook EGP Teacher Guide Table of Contents Introduction...3 Setup...3 Get the Gradebook Web Plugin... 3 Using Your Web Gradebook... 4 Using the Web Gradebook on a Shared Computer...
NextGen Setup Guide First-time Workstation Setup & Logging In
This guide will help you get setup on NextGen for the first time you log onto a City or SBC computer. It will also help if you are a first-time user and need to create your password. I. Setting up Desktop
BD FC Beads 8-Color Kit for BD OneFlow Assays
BD FC Beads 8-Color Kit for BD OneFlow Assays 7/2014 23-15785-00 IVD 5 tests per kit Catalog No. 658621 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2014 BD Becton,
Empower AutoArchive Software
Empower AutoArchive Software Installation and Configuration Guide 34 Maple Street Milford, MA 01757 71500044207, Revision A NOTICE The information in this document is subject to change without notice and
SOS SO S O n O lin n e lin e Bac Ba kup cku ck p u USER MANUAL
SOS Online Backup USER MANUAL HOW TO INSTALL THE SOFTWARE 1. Download the software from the website: http://www.sosonlinebackup.com/download_the_software.htm 2. Click Run to install when promoted, or alternatively,
Baylor Secure Messaging. For Non-Baylor Users
Baylor Secure Messaging For Non-Baylor Users TABLE OF CONTENTS SECTION ONE: GETTING STARTED...4 Receiving a Secure Message for the First Time...4 Password Configuration...5 Logging into Baylor Secure Messaging...7
Secure Global Desktop (SGD)
Secure Global Desktop (SGD) Table of Contents Checking your Java Version...3 Preparing Your Desktop Computer...3 Accessing SGD...5 Logging into SGD...6 Using SGD to Access Your Desktop...7 Using SGD to
ICP Data Entry Module Training document. HHC Data Entry Module Training Document
HHC Data Entry Module Training Document Contents 1. Introduction... 4 1.1 About this Guide... 4 1.2 Scope... 4 2. Step for testing HHC Data Entry Module.. Error! Bookmark not defined. STEP 1 : ICP HHC
Scan to PC Desktop: Image Retriever 5.2 for Xerox WorkCentre C2424
Scan to PC Desktop: Image Retriever 5.2 for Xerox WorkCentre C2424 Scan to PC Desktop includes Image Retriever, which is designed to monitor a specified folder on a networked file server or local drive
Microsoft Excel 2010 Linking Worksheets and Workbooks
Microsoft Excel 2010 Linking Worksheets and Workbooks Email: [email protected] Web Page: http://training.health.ufl.edu Microsoft Excel 2010: Linking Worksheets & Workbooks 1.5 hour Topics include