ph and ORP Electrode Manual

Similar documents
ph Measurement and Control

SenTix 950 SenTix 980

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

REDOX AND REFERENCE ELECTRODE USER MANUAL

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

USEPA Electrode Method Method 8156

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Milwaukee USER MANUAL

ph Measurements of Common Substances

AIRMoN, ph and Conductivity Field Measurement

ELECTROCHEMISTRY. Theory and Practice

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Measurement by Ion Selective Electrodes

Professional Plus. Calibration Tips

WEL Series ph/orp ELECTRODES

Operator Quick Guide EC SENSOR

General Purpose ph/orp Sensor

DISSOLVED OXYGEN PROBE

6 H2O + 6 CO 2 (g) + energy

Features, Limitations & Potential Errors

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL

Recommended Procedures for Use of YSI Water Quality Monitoring Instruments during Oil Spills

General Purpose ph/orp Sensors

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

PART 1 - INTRODUCTION...

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Trace Dissolved Oxygen Sensor

PN /rev. D December Theory and Practice of ph Measurement

Ion Selective Electrodes

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

Global Water Instrumentation, Inc.

CONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

Compact ph Measuring Instruments

Operating Instructions

For additional information, please visit our website at

A Potentiometric Analysis of Fluoride Ion in Toothpaste

The Theory of ph Measurement

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter

After the reactor has been running and the air in the chamber has been purged the co2 can be introduced and the dial in process can begin.

UltraBasic Benchtop ph Meter Operation Manual

ExStik TM ph Waterproof Meters

Determining Equivalent Weight by Copper Electrolysis

Care and use guide. Contents. page

OxyGuard. Dissolved Oxygen Probe

CYCLE CHEMISTRY ph MEASUREMENT

Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Chemicals Needed:

Acid Dissociation Constants and the Titration of a Weak Acid

SDX Submersible Depth Transmitter User Manual

Handbook. Thermo Scientific. ph Electrode Selection Handbook

Chapter2. Installation

Ozone Sensor. Model 499A OZ CAUTION CAUTION. Instruction Sheet PN 51A-499AOZ/rev.F January 2007

Care and use guide. Contents. a handy solution made easy. page

Stainless Steel Dissolved Oxygen Electrodes

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation

Bluelab Soil ph Meter Instruction Manual

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Figure 1. A voltaic cell Cu,Cu 2+ Ag +, Ag. gas is, by convention, assigned a reduction potential of 0.00 V.

OXIGENO DISUELTO. Dissolved Oxygen. Install & forget. Sistema autolimpiable. Autoclean system

Chloride and Salinity

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance

SOLAR PV SYSTEM MAINTENANCE GUIDE

SINGLE RANGE DS METERS Operation Instructions

User Guide Dissolved Oxygen Probes

Pro1030 USER MANUAL. English

Combination ph/redox (ORP) Sensors

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

Neutralizing an Acid and a Base

BATTERY CHARGING GUIDELINES FOR 6-VOLT DEEP CYCLE BATTERIES

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets

Experiment 4 (Future - Lab needs an unknown)

MSFET MICROSENS Miniature ph Sensor Module

MIST DECORATIVE LAMP WITH ESSENTIAL OIL DIFFUSER

TEMPERATURE SENSOR USER MANUAL

If you accidentally get battery acid on your skin, flush it with lots of water

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600


white paper reviews why the chlor-alkali process is so damaging to ph electrodes and discusses a sensor design that has

EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

High Purity Water Resistivity/ Conductivity Measurement

Appendix C. Vernier Tutorial

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...

DeltaSpan Submersible Pressure Level Transmitter Manual LD31 Series 22 AUG 08 Rev A

Acid Base Titrations

JUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor

Experiment 8 - Double Displacement Reactions

EXPRESS STILL WARNING It is legal in New Zealand to own and operate a still for the purpose of producing alcohol for your own consumption.

Dissolved Oxygen Probe (Order Code DO-BTA or DO-DIN)

INSTALLATION & OPERATING INSTRUCTIONS

PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION

Electrical Conductivity of Aqueous Solutions

Transcription:

ph and ORP Electrode Manual Contents Page 2...... Electrode Applications Page 3...... Electrode Cable Page 4...... Installing Submersible Electrodes Page 5...... Electrode Care and Maintenance Page 6...... Calibration Page 7...... Electrode Problems

Electrode Applications ph and ORP (Redox) electrodes are sensitive measuring sensors and should be treated with care. The electrodes are replaceable items and the life span is dependent on the chemical and temperature conditions. A typical life span for an electrode is 2 to 3 years but this may be much less if the electrode is subjected to harsh conditions. The following manual is intended as a guide for maintaining common electrodes for laboratory, in-line and submersible applications. 1/ Laboratory Electrodes Laboratory electrodes usually have a body diameter of 12mm. The body material may be typically glass, epoxy or polymer. For specialised applications such as, measuring in a test tube or penetrating food products, the probe profile is specially designed. Laboratory probes are often adapted for in-line measurements, using a threaded adapter. 2/ In-line Electrodes For measuring ph or ORP in a pipe, a threaded in-line electrode is used. It is important to install the electrode so that an air cavity does not form around the sensing area of the electrode. Ideally install from the top of the pipe to maintain the fluid within the glass bulb. When installed the electrode should be at an angle of between 30 o and 90 o to the horizontal. 3/ Submersible Electrodes For ph and ORP measurements in an open tank or viaduct, a submersion electrode coupled with an immersion assembly is used. In this application it is important NOT to immerse the electrode cable. By using a submersion assembly, the electrode is supported and the cable contained within the immersion assembly. 1 of 6 PHORP02-1.0-1

Electrode Cable The cable used for ph and ORP electrodes is a special screened low noise cable. The electrode is a high impedance device and is coupled to a high impedance measuring instrument. The cable incorporates a special conductive rubber material known as an anti-microphonic layer. This prevents voltages being generated when the cable is moved or vibrated. Special Note: Most electrodes are supplied with a BNC connector fitted. It is often necessary to remove the BNC connector or shorten the cable to make the cable suitable for connecting to screw terminals. IMPORTANT When re-terminating the cable, the anti-microphonic layer must be stripped back from the centre conductor to prevent a short circuit in the cable. Techniques for re-terminating the cable 2 of 6 PHORP02-1.0-1

Installing Submersible Electrodes Electrodes may be submersed by using an immersion assembly. The electrode is fitted and thread sealed to the end of a pipe section. This provides a rigid assembly and protects and seals the cable. The assembly may be built from standard plumbing fittings, or purchased as a complete assembly. IMPORTANT - DO NOT IMMERSE THE CABLE Application Example In the example below additional protection is provided by the fixed open ended pipe. This is ideal for applications where turbulence is present. Example of immersion assembly The immersion assembly should be built to suit the application. In the example above a flange is incorporated to stabilise and support the assembly. The junction head makes electrode replacement easier, by allowing the cable to be accessed without undue twisting. In models with an inbuilt 4-20mA amplifier, the junction head contains the electronic circuit board, calibration adjustments and the 4-20mA terminals. 3 of 6 PHORP02-1.0-1

Electrode Care and Maintenance New Electrodes When you receive your electrode the sensor zone is protected by a cover which contains a soaker solution. Before you use the electrode, rinse off the soaker solution and leave in clean tap water or a weak acid for 15 minutes. This will re-activate the reference junction. Occasionally during shipment air bubbles may form in the glass bulb of ph electrodes. Before use, visually check the electrode for trapped air bubbles. Should bubbles be present, then they can be removed by shaking the electrode in a downward motion. Important The electrode membranes must be kept wet at all times. For short term storage the electrode may be stored in a tap water solution. When the electrode is not in use for long periods, it should be stored in a 3M KCl soaker solution adjusted to 4 ph. Note: When storing double junction electrodes, use a solution saturated with the same salt type as the reference junction. Cleaning the electrode If the electrode becomes sluggish in response it may be necessary to clean the electrode. Usually cleaning with a mild detergent will remove the deposit and restore the electrode performance. Take care not to touch the glass at any time. Clean the electrode by washing in the mild detergent and then rinsing with clean tap water. More difficult deposits may be removed by rinsing with methyl alcohol, followed by a rinse in clean tap water. If the methyl alcohol rinse does not restore the response, carry out the following procedure: 1/ Soak the electrode in 0.1M HCl for 1 hour and then rinse the electrode with tap water. 2/ Soak the electrode in 0.1M NaOH for 1 hour and then rinse with tap water. If the electrode performance is not restored repeat the steps 1 and 2, but increase the soaking times to 24 hours. If this does not restore the electrode performance, replace the electrode. Re-fillable Electrodes Most combination electrodes contain a gelled electrolyte which is never replaced. In some specialised applications the electrode may have a flowing electrolyte. The electrolyte is contained in a chamber which must be constantly replenished. A variety of refillable solutions are available and should be selected to suit the application. Note: single junction AgCl lab electrodes should be topped up with 3 molar KCl/sat with AgCl, as KCl only can shorten the electrode life and cause an offset in ph 7 buffer. Points to Remember By following these points, it is possible to significantly increase the expected life of an electrode, and also greatly improve the quality of measurement results. ph and ORP electrodes must always be stored wet. For short term storage soak the electrode in KCl. For long term storage fill a container with the soaker solution immerse the electrode sensor zone and seal to prevent evaporation. Never store the electrode in: de-ionised water, solvents, hydrofluoric acid, ph buffers containing mercury based preservatives. Sensing tips should always be rinsed after use (laboratory electrodes). Reference cells should be kept regularly topped up with electrolyte (refillable laboratory electrodes). Connectors should be kept clean and dry. If the electrode needs to be cleaned physically, always use a soft tissue soaked in a mild detergent or methyl alcohol. Important: do not wipe the glass - use a dabbing technique. Never touch the electrode glass bulb with your fingers or other oily or abrasive objects. 4 of 6 PHORP02-1.0-1

Calibration ph Calibration You will need: The equipment to be calibrated: comprising, ph electrode coupled to the ph meter. Buffer solutions: typically 1 x ph 7.00 and 1 x ph 4.00. DI rinsing water: de-ionised or distilled water (ideally in a fast flow wash bottle). Tissue to dab electrode dry - do not wipe electrode - this can damage the glass surface. Thermometer to measure temperature of buffer solutions. Before you start: Make sure the ph meter is powered up and stabilised. Check the temperature of the buffer solutions (they should both be at ambient temperature). If the meter does not have automatic temperature compensation, set manually to the solution temperature. Allow the ph electrode to stabilise to the same ambient temperature as the buffer solutions. Calibration procedure: Rinse the electrode in de-ionised or distilled water - shake off excess water - dab dry with tissue. Immerse the electrode in a buffer of 7.00 ph - allow to stabilise for 1 to 2 minutes. Adjust the calibration of ph meter to read 7.00. Remove electrode and rinse in DI water - shake off excess - dab dry with tissue. Immerse electrode in buffer of 4.00 ph - allow to stabilise for 1 to 2 minutes. Adjust ph meter to read 4.00. Repeat procedure to confirm calibration - if necessary repeat until calibration is acceptable. (note: this should not be necessary on microprocessor based instruments). ORP (Redox) Calibration You will need: The equipment to be calibrated: comprising, ORP electrode coupled to the ORP meter. Buffer solutions: typically 1 x ph 7.00 and 1 x ph 4.00, a small quantity of quinhydrone. DI rinsing water: de-ionised or distilled water (ideally in a fast flow wash bottle). Tissue to dab electrode dry - do not wipe electrode - this can damage the glass surface. Thermometer to measure temperature of buffer solutions. Before you start: Make sure the ORP meter is powered up and stabilised. Check the temperature of the buffer solutions (they should both be at ambient temperature). Allow the ORP electrode to stabilise to the same ambient temperature as the buffer solutions. Add a pinch of quinhydrone to each buffer, mix well, add more quinhydrone until the solutions are saturated and will not absorb more quinhydrone. Calibration procedure: Rinse the electrode in de-ionised or distilled water - shake off excess water - dab dry with tissue. Immerse the electrode in the ORP 7.00 ph buffer - allow to stabilise for 1 to 2 minutes. Adjust the calibration of ORP meter to read 86mV. Remove electrode and rinse in DI water - shake off excess - dab dry with tissue. Immerse electrode in the ORP 4.00 ph buffer - allow to stabilise for 1 to 2 minutes. Adjust ORP meter to read 263mV. Repeat procedure to confirm calibration - if necessary repeat until calibration is acceptable. (note: this should not be necessary on microprocessor based instruments). Note: the mv values quoted are for electrodes with a silver-silver chloride reference. For electrodes with a calomel reference subtract 42mV from the above values. 5 of 6 PHORP02-1.0-1

Electrode Problems ph and ORP electrodes continually age and will not last indefinitely. The life span cannot be accurately predicted and is dependent on the chemical and temperature conditions experienced by the electrode. As the electrode ages the slope (output) decreases. The electrode should be replaced when the electrode slope reduces below 90% of an ideal electrode. However it is often possible to rejuvenate a ph electrode by: 1/ Soak the electrode in 0.1M HCl for 5 minutes and then rinse the electrode with tap water. Soak the electrode in 0.1M NaOH for 5 minutes and then rinse with tap water. 2/ the process is detailed on the bottom of this page. Caution: this process uses hazardous chemicals and should only be carried out by a qualified chemist. See chapters Care and maintaining of your electrode for information on cleaning the electrodes and Electrode Cable for anti-microphonic layer. Slow Response Possible cause: Possible cause: Unstable Reading Possible cause: Electrode glass surface contaminated. Clean electrode. Reference junction or reference electrolyte contaminated. Clean electrode. Reference junction blocked or contaminated. Clean electrode. Low Slope (less than 90%) Possible cause: Electrode glass surface not clean. Clean electrode. Possible cause: Glass membrane aged. Try rejuvenating the electrode or replace electrode. Possible cause: Dampness or contamination around cable connections. Clean with methyl alcohol and allow to dry in a warm place. Displays approximately 7 ph (or 0mV for ORP) for all buffers Possible cause: Electrical short circuit. Check connector for short circuit. Check cable anti-microphonic layer. Displays 4 to 5 ph for all buffers Possible cause: Breakdown within electrode body. Replace electrode. Advanced information - Rejuvenating the glass electrode When the electrode is used at elevated temperatures the ph glass loses it s sensitivity. The following procedure may rejuvenate the glass surface and provide an extended electrode life. 1. Prepare a 20% solution of ammonium bifluoride. 2. Immerse the electrode in the solution for 15 to 20 seconds. 3. Rinse the electrode with tap water. 4. Immerse the electrode in 5-6M HCl for 5 minutes to remove the bifluoride. 5. Rinse electrode with tap water. 6. Immerse the electrode in ph buffer 4.00 for 1 to 2 hours. 6 of 6 PHORP02-1.0-1

This page intentionally left blank 7 of 6 PHORP02-1.0-1