TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1



Similar documents
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level

DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR

IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE

PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS

DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION

NATURAL RADIOACTIVITY FROM THE SOIL OF SÃO BERNARDO DO CAMPO

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Radiation safety in dental radiography

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

X-Mind. Instinct for perfection

Calibração de medidores de P KA

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Evaluation of light filter of portable dark chamber and its influence on radiographic image quality

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

DEVELOPMENT OF A QUALITY MANAGEMENT SYSTEM FOR BRAZILIAN NUCLEAR INSTALLATIONS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control

portable x-ray survey report

Simone Alberigi and Brigitte Roxana Soreanu Pecequilo

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S.

THE AWARENESS OF THE FUNCTIONAL AND NEAR POPULATION WITH THE RELATION TO THE RESEARCH NUCLEAR REACTOR IEA-R1

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

Standards for accreditation of educational programs and approval of continuing education courses.

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Production of X-rays and Interactions of X-rays with Matter

How To Become An Xray Technician

Patient Exposure Doses During Diagnostic Radiography

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

Radiation safety in dental radiography

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

3rd International Symposium of the System of Radiological Protection Seoul, Korea, October 2015

CODE OF PRACTICE & SAFETY GUIDE

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Digital Dental X-ray. Midmark Animal Health Products

What Parents Should Know about the Safety of Dental Radiology.

Integrated Management & Educational Consultancy

VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

X-ray Imaging Systems

CORK UNIVERSITY DENTAL SCHOOL & HOSPITAL AND DIAGNOSTIC IMAGING PROGRAMMES DEPARTMENT SCHOOL OF MEDICINE & MEDICAL SCIENCE UNIVERSITY COLLEGE DUBLIN

Agency for Health Care Administration

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

TLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

DENTAL X-RAY TECHNICIAN SERIES

APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS

X-ray (Radiography), Chest

COMMUNICATING RADIATION RISKS IN PAEDIATRIC IMAGING

NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF DENTAL RADIOLOGIC TECHNOLOGY

Evaluation of the image quality in computed tomography: different phantoms

X-ray (Radiography) - Chest

THE PRINCIPLE OF JUSTIFICATION

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF DENTAL RADIOLOGIC TECHNOLOGY

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

EXTERNAL DOSE RATES IN COASTAL URBAN ENVIRONMENTS IN BRASIL

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Elements of required physical infrastructures: space, schielding, and patient flow..

There is no question: X Rays help dentists care for children. Select x rays for individual s needs, not merely as a routine

European Academy of DentoMaxilloFacial Radiology

X-RAY REGULATORY GUIDE

Radiologic Technology Training Program

Diagnostic X-ray facilities as per quality control performances in Tanzania

Froedtert Hospital School of Radiology Curriculum Analysis Grid. Clinical Practice

Periapical radiography

Radiation protection principles observance in Iranian dental schools

Development of a parallel plate ion chamber for radiation protection level

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT

Test Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor

3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Towards a National Dose Registry in Portugal

Radiation Safety Issues for Radiologic Technologists

Lasers and Specturbility of Light Scattering Liquids - A Review

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit

INTRODUCTION. A. Purpose

EFFECTS OF GAMMA IRRADIATION ON ELECTRIC CHARACTERISTICS OF POWER DIODES

The Nuclear Safety Council, in exercise of the functions conferred to

Bonn Call-for-Action Joint Position Statement by the IAEA and WHO

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

Radiation Dose Measurements in Routine X ray Examinations

REMOTE SYSTEM FOR MONITORING AND CONTROL OF CONTROLLED AREA OF NUCLEAR INSTALLATION

AN ACT RELATING TO HEALTH CARE; AMENDING THE MEDICAL IMAGING AND RADIATION THERAPY HEALTH AND SAFETY ACT TO EXEMPT EXPANDED

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

TRAINING PROGRAMMES IN RADIATION SAFETY CONDUCTED BY RADIOLOGICAL PHYSICS & ADVISORY DIVISION

R Recognized Certificate-granting Bodies R Approval of Radiologic Technology Schools

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Radiation Safety In Dental Practice

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Impact of a program of diagnostic imaging quality control in mammography centers of the Federal District, Brazil*

Transcription:

TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT In this work the field size of a dental X rays machine was evaluated considering the recommendation of the Brazilian Health Ministry Regulation 453 which established basic lines of radiological protection in medical and dental diagnostic radiology. The diameter of the field should not be superior to 6 cm in the localizer end point, limiting the radiated area and protecting the head-neck region. The measurements were carried out in a dental X rays machine, Dabi Atlante, model Spectro 70X Seletronic. For the field size or useful beam determination, the intra-oral films were positioned on a plain surface to be exposed in four stages and two focus-film distances (FFD), 20cm and 27.5 cm: 1) with spacer cone; 2) without spacer cone; 3) with spacer cone and film-holding device; 4) without spacer cone and film-holding device. The results show that the diameter of the field size is satisfactory only for FFD = 20cm. When the film-holding device is used, which is recommended by the Regulation 453, item 5.8d(ii), the diameter of the field size exceeds the maximum recommended value of 6 cm. INTRODUCTION Quality assurance programmes in diagnostic radiology have been increased in the last years due to the fact that the extensive use of X-rays in medicine for diagnosis of injuries and diseases represents the largest man-made source of public exposure to ionizing radiation (1). The assessment and control of the performance characteristics of X-ray generators and tubes is an essential part of a quality assurance programme. The Brazilian Health Ministry Regulation 453, published in June 1998, established basic lines of radiological protection in medical and dental diagnostic radiology, in order to guarantee the health of patients, workers and of the public (2). In dental area, specifically for intra-oral radiography, one of the recommendations of this regulation is the evaluation of the field size, which represents the exposed area of patient to the primary X rays beam. The diameter of the field should not be superior to 6 cm in the localizer end point, limiting the radiated area and protecting the head-neck region (3,4). Consequently the determination of the field size in dental diagnostic radiology systems is very important. The objective of this work is the elaboration of a reference procedure, following the recommendations of the regulation 453, for the radiation field size determination in dental diagnostic radiology systems. It would be applied in dental clinics, in order to achieve an improvement in the radiological image qualities and the patient dose reduction.

MATERIALS AND METHODS The measurements were carried out in a dental X rays machine, Dabi Atlante, model Spectro 70X Seletronic. Its nominal characteristics are: 70 kvp of tube voltage, 8 ma of tube current, inherent filtration of 0.51mmAl, additional filtration of 1.4 mm and focus-film distance of 20 cm.. The determination of the field size is part of this system quality control programme, other tests have already been performed (5,6). The X rays machine can be seen in the figure 1. Figure 1. Dental X rays machine, Dabi Atlante, model Spectro 70X Seletronic, Characteristics: 70kVp, 8mA and nominal filtration of 2.71mmAl. The intra-oral radiographic films used were Kodak, speed E (ektaspeed), 3 x 5 cm, lot 3103576, produced in February 2005. It was used two set-ups with four films, as showed in Figure 2(a), and two set-ups using nine films. The diameters of the field or useful beam size were measured using a calibrated ruler. Some measurements were taking using the film-holding device showed in Figure 2(b). This accessory is used to fix the radiographic films in the patient s mouth, leaving them parallel to teeth, making easy the rays-x beam alignment. The regulation 453 recommends its use. It reduces, significantly, the loss of radiographies due to the film movement, fold distortions or/and patient head positioning errors (7). a b Figure 2: a) Radiographic films for the field size determination of the X radiation system b) Film-holding device frequently used in dental practices to fix radiographic films and to line up them with the X radiation system spacer cone.

The irradiations were performed positioning the intra-oral radiographic films in four different conditions as described in Table 1. The exposition time was 0.5 seconds in all cases. Figure 4 shows the irradiation conditions A (FFD 20 cm, with cone spacer) and C (FFD 27,5 cm, with cone spacer and film-holding device). Table 1. Irradiations condition for the of the intra-oral radiographic films. Irradiation condition FFD Cone spacer Film-holding device A 20 with without B 20 without without C 27.5 with with D 27.5 without with a b Spacer Cone Film-holding device Figure 4. Set-up for the kvp measurement: a) DFF=20cm with spacer cone; b) DFF=27.5cm with spacer cone and film-holding device. For each irradiation condition it was irradiated five film groups and the revelation of the irradiated intra-oral radiographic films was made following the recommendations of the Kodak (producer of the films), taking into account the table time/temperature (8). The intra-oral radiographic films were mounted and analyzed in a negatoscopy, the diameter of the images were measured in 5 different positions.

RESULTS The obtained values for the X radiation field size in the four mounted arrangements are presented in Table 2. Each value represents the medium of the five individual lectures. The uncertainties associated were classified into type A (statistical methods as the measurement means) and type B (other than statistical, as the uncertainties associated to the measure instruments) (9). The expanded uncertainties were evaluated considering the confidence level of 95%. Table 2. X radiation field size using four different set-ups. Irradiation condition FFD Diameter of the field A 20 5.308±0.006 B 20 5.496±0.003 C 27.5 7.400±0.007 D 27.5 7.624±0.009 The results show that only for FFD of 20 cm the field diameter is less than 6 cm as recommended by regulation 453. As expected for FFD of 27.5 cm, when the film-holding device is used, the diameter of the field is superior to 6 cm over against the recommendations. Otherwise the use of the film-holding device is also recommended by the same regulation, item 5.8d(ii). The dental X radiation system evaluated in this work presents satisfactory results only when is used a FFD of 20cm. According to the regulation 453, values between 4 e 5 cm are allowed only when there is an alignment and positioning system for the film, as described in item 5.7d(ii). When the film-holding device is used, the field size exceeds the maximum recommended value. CONCLUSIONS The obtained results show that the use of film-holding device to fix the intra-oral radiographic films and to align the X radiation beans decreases the entrance skin dose, but increases the size of the field and exceeds the maximum recommended value of 6cm. It is necessary to extend the tests to other dental X radiation systems in use in Brazil to evaluate the real benefit for the use of the film-holding device. ACKNOWLEDGEMENTS The authors acknowledge the partial financial support of the Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP).

REFERENCES 1. S.C.BUSHONG, Radiologic Science for Technologists: Physics, Biology and Protection. Mosby-Year Book Incorporation, St. Louis, MO, USA (1997). 2. MINISTÉRIO DA SAÚDE, Portaria Federal 453. Diretrizes básicas de proteção radiológica em radiodiagnóstico médico e odontológico. Brasília: Diário Oficial da União. 02 Jun., 1998. 3. LAMBRECHT, J.T., ROTH, J., KIEFER, H.P. Dose exposition from intra and extraoral dental radiography. International Congress Series, 1268, 1147-1151, 2004. 4. OGUNDARE, F.O., ONI, O.M., BALOGUN, F.A. Measurements of x ray absorbed doses to dental patients in two dental x ray units in Nigeria. Radiat. Prot. Dosim., v. 104, n. 4, p. 355-358, 2002. 5. ANDRADE, P.S.S., POTIENS, M.P.A. Implementação de um procedimento para a avaliação de kv aplicado em tubos de raios X odontológicos. In : 2005 International Nuclear Atlantic Conference - INAC 2005, Santos, SP, Brazil, August 28 to September 2, 2005. Associação Brasileira de Energia Nuclear ABEN. 6. ANDRADE, P.S.S., POTIENS, M.P.A. Medidas de kerma no ar de um aparelho de raios X odontológico em diferentes angulações. In : X CongressoBrasileiro de Física Médica- BAHIA2005, Salvador, BA, Brazil, May 26 to 29, 2005. Associação Brasileira de Física Médica 7. LEONIDAS, F. Radiologia Bucal Técnicas e interpretações. Pancast editora, 2 a edition, 2000. 8. Tabela de tempo e temperatura para revelação de filmes periapicais http://wwwbr.kodak.com/br/pt/health/v5/dental/teguiatempo.jhtml, access in 09 april 2005. 9. International Organization for Standardization, Guide to the Expression of Uncertainty in Measurement, ISO, Genève (1993).