An LC-MS/MS Clinical Research Method for the Measurement of 25-OH Vitamin D 2 and D 3 Metabolites

Similar documents
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Oasis HLB Cartridges and 96-Well Plates

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Introduction. Methods. Sample Processing

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System

A Novel Approach to Low Volume Sample Preparation

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

How To Use An Acquity Qda Detector

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

Innovative vs Traditional

Salman Azimi, 1 Nayan S. Mistry, 2 and Michelle Wood 2. Drug Quality Control Laboratory, Supreme Council of Health, Doha, Qatar 2

How To Test For Contamination In Large Volume Water

Overview. Purpose. Methods. Results

[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS. Simplifed

United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science

Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns

Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )

Supplementary Materials and Methods (Metabolomics analysis)

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

AppNote 9/2013. Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides KEYWORDS ABSTRACT

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

How To Develop A Recipe Card

DEVELOPMENT OF PFCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

Development of a Systematic Toxicological Screening Method using an automated online-spe-lc-qqtof System (XLC-QqTOF)

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Purification of reaction mixtures using flash chromatography.

[ Care and Use Manual ]

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

CORESTA RECOMMENDED METHOD N 72

Determination of Tetracycline Antibiotics in Milk Using a Simple Strong Cation-Exchange SPE Cleanup Procedure and LC-MS/MS Analysis

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT

AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS

What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model?

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Chemical analysis service, Turner s Green Technology Group

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water

Protein Separation Technology

ApplicationNOTE Introduction

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

Patients with COPD were recruited according to the inclusion criteria; aged 40-75, current or

WATERS QUANTITATIVE ANALYSIS solutions

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

SCREENING FOR DRUGS IN SERUM AND URINE BY LC/ESI/CID-MS AND MS/MS WITH LIBRARY SEARCHING *

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

AppNote 6/2011. Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS KEYWORDS ABSTRACT

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Transcription:

An LC-MS/MS Clinical Research Method for the Measurement of 25-OH Vitamin D 2 and D 3 Metabolites Martijn van Faassen 1, Ido Kema 1, and Kendon Graham 2 1 University Medical Center Groningen, Groningen, The Netherlands; 2 Waters Corporation, Milford, MA, USA GOAL To demonstrate that the use of the ACQUITY UPLC Online SPE Manager (OSM) enables analytically sensitive measurement of important vitamin D metabolites in plasma. BACKGROUND Recent studies have shown that vitamin D deficiency is common in adults and children in many parts of the world. In addition to the well known effects of vitamin D deficiency such as calcium malabsorption, there is growing evidence that vitamin D deficiency may increase the risk of certain types of cancer, and play a role in many other diseases. As a consequence, the measurement of vitamin D metabolites is of increasing importance to many researchers. Vitamin D exists in two forms; vitamin D 3 (D 3 ), which is produced in the skin on exposure to sunlight, and vitamin D 2 (D 2 ), which is a plant derivative that is found in many dietary supplementation products. All forms of vitamin D are metabolized in the liver to form 25-hydroxyvitamin D (25-OH D), which is further metabolized in the kidneys to form the active metabolite 1,25-dihydroxyvitamin D that is responsible for calcium homeostasis. Improved workflows for the measurement of vitamin D metabolites in plasma using the ACQUITY UPLC (OSM). OH H OH H H 25-OH vitamin D 3 25-OH vitamin D 2 HO HO Figure 1. Structures of 25-OH D 2 and 25-OH D 3. In many laboratories, the measurement of vitamin D and its metabolites in plasma by LC-MS has gained popularity over other methods such as the competitive binding methods, immunoassays, and HPLC to quantify 25-OH D 2 and 25-OH D 3. The issues with these other types of measurement are that they may be complex and costly to perform, or they cannot discriminate between 25-OH D 2 and 25-OH D 3 and measure these vitamers accurately enough to provide the level of information that researchers are seeking. Even when using LC-MS, issues with preparing samples effectively and efficiently must be overcome, to allow for the development of an accurate method that can be used routinely for measuring vitamin D metabolites.

THE SOLUTION A clinical research method for the measurement of two vitamin D metabolites (25-OH D 2 and 25-OH D 3 ) from plasma has been developed. This method takes advantage of the unique capabilities of the ACQUITY UPLC OSM to enable analytically sensitive and reproducible measurements. The method enables direct measurement of these key vitamin D metabolites, and provides valuable information to clinical researchers as they investigate the role these molecules play in normal biological function and disease. Simultaneous measurement of these key metabolites with this method can also provide valuable information to clinical researchers as they investigate novel models of disease. Method details LC-MS system configuration: LC: ACQUITY UPLC MS: Xevo TQ-S Column: ACQUITY UPLC CSH Phenyl-Hexyl, 1.7 µm, 2.1 x 1 mm (p/n 186547) Sample preparation: ACQUITY UPLC OSM SPE cartridge: MassTrak C 8 Online SPE Cartridges, 1 μm, 1 x 1 mm (p/n 1865673IVD) Sample pre-treatment: 15 µl plasma 25 µl internal standard in methanol containing (isotopically labelled 25-OH D 2 and D 3 ) 8-µL of aqueous.4 M ZnSO 4 ; centrifuge to pellet protein Chromatography conditions Solvent A: 1 methanol + 2 mm ammonium acetate +.5 formic acid Solvent B: Methanol + 2 mm ammonium acetate +.5 formic acid Flow rate:.4 ml/min Column temp.: 35 C Gradient: Time A B. 35 65 3.75 1 9 4.25 1 5.25 1 5.5 35 65 6.25 35 65 Online SPE conditions Conditioning: Step 1:.5 ml acetonitrile at 4 ml/min Equilibration: Sample load: Step 2:.5 ml acetonitrile/methanol/ isopropyl alcohol/water +.2 formic at 3 ml/min.5 ml water at 4 ml/min.5 ml water at 2 ml/min Cartridge wash: Step 1:.5 ml 2 methanol +.1 ammonia at 2 ml/min Clamp flush: Elution time: Step 2:.5 ml 3 methanol +.1 ammonia at 2 ml/min Step 3: 1 ml 55 methanol +.1 formic acid at 2 ml/min Step 1:.5 ml acetonitrile at 3 ml/min Step 2: 1. ml acetonitrile/methanol/ isopropyl alcohol/water +.2 formic at 3 ml/min Step 3:.5 ml water at 3 ml/min.75 min Multiple reaction monitoring conditions Compound Precursor m/z Fragment m/z Cone (V) Collision (ev) 257.3 3 16 25(OH) D 3 383.4 245.3 3 17 d6-25(oh) D 3 389.4 263.3 3 16 269.3 3 17 25(OH) D 2 395.4 271.3 3 17 d6-25(oh) D 2 41.4 275.3 3 17

RESULTS The method described was successfully used to detect and measure two vitamin D metabolites in plasma. Figure 2 demonstrates the typical analysis of these analytes in plasma samples. 1 A) 55 nmol/l 25-OH D 3 in plasma 1 d6-25-oh D 3 internal standard 2.6 2.8 3. 3.2 3.4 3.6 3.8 4. 4.2 4.4 min 1 B) 14 nmol/l 25-OH D 2 in plasma 1 d6-25-oh D 2 internal standard 2.6 2.8 3. 3.2 3.4 3.6 3.8 4. 4.2 4.4 min Figure 2. Separation of vitamin D metabolites in plasma. Both the 25-OH D 2 and 25-OH D 3 metabolites were separated and analyzed from plasma samples. Extracted ion chromatograms shown in panels A and B demonstrate the measurement of typical plasma levels of these analytes.

Method assessment Intra-assay variability The reproducibility of the method for measuring the vitamin D metabolites was assessed for both intra- and inter-assay variability. For measuring intra-assay variability, 12 samples at three concentrations of 25-OH D 3 (27, 75, and 165 nmol/l) and two concentrations of 25-OH D 2 (4 and 185 nmol/l) in plasma were analyzed. The results presented in Table 1 indicate that the intra-assay variability of the method ranges from 1.5 to 3.8 for 25-OH D 3 and from 3.1 to 5.3 for 25-OH D 2. Table 1. Assessment of intra-assay variability. Twelve individual plasma samples containing a variety of concentrations representative of the low, medium, and high QC concentrations for of vitamin D metabolites for the method were measured. CVs ranged from 1.5 to 5.3 across all concentrations measured. 25-OH D 3 (nmol/l) 25-OH D 2 (nmol/l) n QC-low QC-medium QC-high QC-low QC-high 1 27.2 74.7 161 39.3 183 2 26.6 73.4 168 38. 194 3 25.3 74.6 167 43.5 181 4 26.6 75.3 164 42. 187 5 26.5 74. 163 41.1 192 6 26.7 75.2 164 42.9 183 7 26.3 75.6 166 43.2 189 8 29.3 75.1 163 44.6 189 9 28.2 74.7 162 41.5 177 1 26.8 73.1 158 39.9 179 11 26.6 77.1 163 46. 189 12 27.4 73.1 168 42.5 195 Average 27 74.7 164 42.1 187 Std. dev. 1. 1.1 3. 2.2 5.8 CV 3.8 1.5 1.8 5.3 3.1 Inter-assay variability Inter-assay variability of the method was assessed by performing the method daily over the course of six days using duplicate samples of the vitamin D metabolites in plasma at concentrations ranging from 25 to 18 nmol/l. The results presented in Table 2 indicate that the inter-assay variability range from CVs of 3.3 to 3.9 for 25-OH D 3 to CVs of 3.7 to 5 for 25-OH D 2. Table 2. Inter-assay variability determination. Duplicate plasma samples containing a range of concentrations representative of the low, medium, and high QC concentrations for of vitamin D metabolites were measured over the course of six days. CVs ranged from 3.3 to 5. across all concentrations measured. 25-OH D 3 (nmol/l) 25-OH D 2 (nmol/l) Day QC-low QC-medium QC-high QC-low QC-high 1 27.2 74.7 161 39.3 183 26.6 73.4 168 38. 181 2 27.4 72. 158 37.2 181 25.1 73.2 162 4.4 175 3 26. 67.3 157 4.6 173 24.9 69.9 16 4.3 178 4 25.9 68.2 154 39. 18 26.2 69.1 152 35.8 173 5 24.8 68.6 155 36.9 186 24.2 68.8 15 39.9 179 6 25.2 68.8 154 35.2 169 26.4 67.5 152 36.3 163 Average 25.8 7.1 157 38.2 177 Std. dev. 1.1 2.52 5.23 1.9 6.5 CV 3.9 3.6 3.3 5. 3.7

Method comparison The method described was compared to results obtained from an independent LC-MS/MS method, developed in the same laboratory. This method utilizes a different, less efficient workflow involving the use of an older online SPE system in conjunction with protein precipitation. The results indicate that the developed method displays excellent correlation with the existing LC-MS method. Scatter plot with Deming fit 12 1 ACQUITY UPLC OSM method 8 6 4 2 2 4 6 8 1 12 Online SPE-LC-MS/MS method Identity Deming fit (-1.25 + 1.8x) Figure 3. Comparison of the developed method using the ACQUITY UPLC OSM with an existing LC-MS/MS method using online SPE. The results obtained by the developed method correlated very well with the existing method, but provided a much more efficient workflow.

SUMMARY In many clinical research laboratories, the sensitive and reproducible measurement of vitamin D metabolites can be a challenge. Older methods involving the use of immunoassay or other technologies are unable to provide the level of information on these important vitamin D metabolites that many researchers are seeking to measure. While LC-MS is capable of making the specific and analytically sensitive measurement of these vitamin D metabolites, preparing samples adequately and efficiently has always been an issue that has not been easy to overcome. The method described here takes best advantage of the automated sample preparation capabilities of the ACQUITY UPLC OSM in the analysis of vitamin D metabolites. This highly automated method will reduce both the direct staff and instrument time required to perform these analyses when compared to other LLE or offline analysis techniques, and allow for rapid, simultaneous analysis of these vitamin D metabolites. The C 8 sorbent used for extraction and the ACQUITY UPLC CSH Hexyl-Phenyl Column used in this study also showed excellent performance in terms of sample clean-up and separation of these molecules (Figure 2). The method developed here provides: Simultaneous analysis of 25-OH D 2 and 25-OH D 3 from plasma at low nmol/l concentrations Effective separation of both vitamin D metabolites for easy measurement Efficient, online SPE sample preparation integrated with UPLC-MS/MS in approximately six minutes per sample Small volume required (1 µl injection of 15 µl plasma sample) Excellent reproducibility and robustness (intra- and inter-assay CV <5) Excellent correlation with an existing, online SPE LC-MS/MS method, but with a streamlined workflow resulting in greater efficiency Furthermore, both the intra- and inter-assay variability of the method were assessed to be excellent (Tables 1 and 2). Even at low nmol/l concentrations of either vitamin D metabolite, the calculated CVs were typically less than 5. The robustness of the method makes it a good option for routine measurement of vitamin D metabolites in a complex matrix such as plasma. In this study, a rapid clinical research method using the ACQUITY UPLC OSM for the measurement of vitamin D metabolites has been developed. The method is suitable for routine use and the small volume of serum used (15 µl) is desirable for clinical research laboratories that are sample limited. For Research Use Only. Not for use in diagnostic procedures. Waters, The Science of What s Possible, ACQUITY UPLC, and Xevo are registered trademarks of Waters Corporation. CSH is a trademark of Waters Corporation. All other trademarks are the property of their respective owners. 215 Waters Corporation. Produced in the U.S.A. May 215 725345EN KP-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 2 F: 1 58 872 199 www.waters.com