FLUOROMETER PERFORMANCE EVALUATION REPORT 9/2002

Similar documents
An Introduction to Algae Measurements Using In Vivo Fluorescence

How to measure Chlorophyll-a

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

ALMOFRONT 2 cruise in Alboran sea : Chlorophyll fluorescence calibration

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy

DSM155 Digital LED Light Meter User s Manual DSM155

Phosphorus, Reactive (Orthophosphate)

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

EPA New England Regional Laboratory. EPA Project Team: Liz McCarthy. Matt Destino Tim Bridges Dick Siscanaw

Nederland België / Belgique

BACTERIAL ENUMERATION

Thermo Scientific NanoDrop Products. Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab

Method In Vitro Determination of Chlorophyll a and Pheophytin a in Marine and Freshwater Algae by Fluorescence

UVC LEDs for. Environmental Monitoring

Application Note: Absorbance

Using the Spectrophotometer

Experiment #5: Qualitative Absorption Spectroscopy

Real-Time Monitoring Buoy in Mystic River Watershed

UV-Vis spectrophotometers

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

About Coffee and Refractometers Voice Systems Technology, Inc. (VST)

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

A Beer s Law Experiment

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

The Determination of an Equilibrium Constant

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Agilent Cary 60 UV-Vis

A Complete Solution for Method Linearity in HPLC and UHPLC

Experiment 5. Lasers and laser mode structure

Pesticide Analysis by Mass Spectrometry

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter

Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

The photoionization detector (PID) utilizes ultraviolet

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Analytical Test Method Validation Report Template

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Quaternary Ammonium Compounds DOC

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Online Infrared Flue Gas Analyzer Gasboard 3000

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Drop Counter Sensor Product Number: ENDRP-AD100

Features Rich Detectors for Every Laboratory and Application

Quantification of Reactive Phosphorus in Lake Mendota Sediments

Conversion Between Analog and Digital Signals

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Appendix 5 Overview of requirements in English

AN Application Note: FCC Regulations for ISM Band Devices: MHz. FCC Regulations for ISM Band Devices: MHz

Data Communications Competence Center

Laboratory 5: Properties of Enzymes

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Application Requirement

The Determination of an Equilibrium Constant

REAL-TIME DUST AND AEROSOL MONITORING

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

the importance of ph in water purification the importance of ph in water purification Water Purification Systems UK

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS

Section 3. Sensor to ADC Design Example

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Beer's Law: Colorimetry of Copper(II) Solutions

Shimadzu UV-VIS User s Guide

Zeiss 780 Training Notes

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

The Measurement of Sensitivity in Fluorescence Spectroscopy

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

Duct Humidity Transmitter

DDX 7000 & Digital Partial Discharge Detectors FEATURES APPLICATIONS

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

PART 1 - INTRODUCTION...

Interference Identification Guide. Table of Contents

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL

Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry

2 Spectrophotometry and the Analysis of Riboflavin

Gas Custody Transfer Calibration

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

HITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE

Digital image processing

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

At the bottom of the screen you will also see some of our sister brands, all of which can also plug in to the Storm data logger.

Introduction to flow cytometry

NJWMC Continuous Monitoring Inventory & DEP s Continuous Monitoring

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Lab #11: Determination of a Chemical Equilibrium Constant

Hand-held thermometer Model CTH7000

Series DFMT Digital Paddlewheel Flowmeter

ph Alkalinity of Water

The Sievers 900 Series TOC Analyzers

Advanced Solutions for Gas Monitoring

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Measurement & Control. Features:

Neutralizing an Acid and a Base

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC

Measuring Laser Power and Energy Output

Characterizing Digital Cameras with the Photon Transfer Curve

Transcription:

FLUOROMETER PERFORMANCE EVALUATION REPORT 9/22 ABSTRACT A series of performance evaluation tests were conducted at three tests sites to investigate the overall performance of the Turner Designs Submersible Fluorometer and the chlorophyll probe. Performance was determined through results designed to determine minimum detection limit of chlorophyll a, dynamic range, accuracy of readings to actual chlorophyll a, ambient light rejection, and response time. The results indicate that the fluorometer outperforms the sensor but the performance of both sensors is adequate for most environmental conditions. A performance limitation of the is the response time. Using a moving average calculation, it responds relatively slowly to changing concentrations and would not be effective for vertical profiling applications. In some extreme conditions, such as chlorophyll concentrations less than 1µg/L and high ambient light and/or high concentrations of interfering compounds, the may also not provide accurate measures of algal biomass. INTRODUCTION A series of performance evaluation tests were conducted to investigate the overall performance of the Turner Designs Submersible Fluorometer and the chlorophyll probe. Tests were conducted at three sites; Moss Landing Marine Laboratory, Turner Designs Testing Laboratory and the USGS Menlo Park facility. Performance was evaluated through designing tests to investigate the following areas; minimum detection limit, instrument range, response time, light rejection and accuracy of fluorometer reading to actual chlorophyll a concentration. Due to different specification standards, comparing instruments on published specifications is not always straightforward. Below is the published performance data for each instrument. Through our testing we have attempted to compare the instruments in a meaningful and informative way for the customer. We have compared instruments identically and have attempted to present the data in as an objective manner as possible. In most cases the raw fluorescence signal was recorded in order to eliminate scaling issues related to calibration. The raw fluorescence was recorded as %FS which indicates the % of a sample s signal in relation to 1% full scale of the particular instrument. Calibrated data is referred to as in vivo fluorescence or. Published Performance Specifications Range -4µg/L 4 orders of magnitude Minimum Detection Limit N/A.2 µg/l Resolution.1 µg/l,.1%fs 12 bit Max Sampling Rate NA 5Hz Depth -61m -6m Instrument Overview: While both instruments are chlorophyll fluorometers there are several key distinctions between the two units. Firstly, the physical size and appearance is significantly different. The fluorometer is designed to operate as a stand-alone unit and can also be interfaced to multi-parameter systems. Being a stand-alone unit, it incorporates many additional features that are not required in the probe. The following is a list of unique features; 6m depth capability, temperature compensation, internal data logging, autoranging, turbidity option, digital and analog output, configurable analog (-5V) limits, and calibration software. The presence of these features has resulted in a much different looking sensor. However, when interfaced to a multi-parameter system, the operates in a very similar manner to the probe. The has been designed to interface directly onto multi-parameter sondes. The probe is also now available on an optical sonde that only offers fluorescence, conductivity, 1 FLUOROMETER COMPARISON REPORT

temperature and depth. The probe provides an analog signal to the sonde which integrates the data from all of the sonde sensors into one digital data stream. The provides both digital and analog data continuously. MATERIALS AND METHODS All tests were conducted under controlled settings in a laboratory. In most cases, monocultures on algae were used as the test media (see table below for algal culture details). In some cases natural algal samples were used. Chlorophyll a extractions were conducted on all samples using the Non-acidification fluorometric method (Welschmeyer, 1997) and the Acidification technique (E.P.A. Method 445.) in the case of the pheophytin testing. Algal samples used: Genus Algal group 1 Accessory Pigment Dunaliella sp. chlorophyte chlorophyll b Thallassiosira sp. diatom chlorophyll c Phaeodactylum sp. diatom chlorophyll c San Francisco Bay Delta natural assemblage RESULTS MINIMUM DETECTION LIMIT The Minimum Detection Limit is the lowest concentration of algae in clean culture media than can is distinguishable from a blank culture media solution. Three detection limit tests were run by mixing a dilution series of algal cultures. The readings were allowed to stabilize and at least 1 datapoints were recorded per sample. Dunaliella Phaeodactylum Chl A Chl A 1.36 1. 1..368 1. 1. 1.138 1. 1.125.638 1.263 1.378 2.54 1.26 1.792 1.36 1.447 2.585 3.321 1.385 2.167 1.61 1.737 3.835 Thallassiosira Chl A..14 1.46.34.93 5.92.7.89 8.33 2.6 4.76 15.42 Table 1: Dunaliella and Phaeodactylum test data from Moss Landing Marine Laboratory (MLML). Data from Thallassiosira test from Turner Designs. The Chloro column is extracted chlorophyll a data. The and column is the in vivo fluorescence data. From the data shown it is clear the difference in detection limit between algal groups. The data supports the fact that the was unable to distinguish some algal cultures at concentrations of 1µg/L chlorophyll a or less while the was capable of distinguishing all samples. The lowest concentration tested was.34µg/l of Thallassiosira culture. Because each species of algae has varying in vivo fluorescence ():extracted chlorophyll, the limits of detection can vary between species. The demonstrated the ability to detect concentration as low as.79µg/l of Thallassiosira culture while not detecting other cultures that were above 1µg/L. 2 FLUOROMETER COMPARISON REPORT

Minimum Detection Limit Test #1 Minimum Detection Limit Test #2 3.5 3 2.5 2 1.5 1.5 R 2 =.9737 R 2 =.923 1 2 3 4 5 chlorophyll a (ug/l) Graph 1 & 2: MDL test data from Moss Landing Marine Lab showing that the not differentiating between 2.5µg/L and 1.1µg/L Dunaliella samples. MDL Test#2 shows data from tests with Thallassiosira cultures indicate that the did not distinguish the blank and the first two culture samples. RANGE Both instruments provide an analog -5V output signal while the has a digital output as well. The -5V ranges are factory set for a given fluorescence range. The comes standard with a -1µg/L analog range while the comes standard with a -4µg/L range. As the analog range increases, the resolution of the data decreases and it is therefore advantageous to set the -5V limits to the environments in which you work. The digital outputs from the Sonde and the were used for all tests. Both instruments demonstrated good linearity for typical chlorophyll a ranges found in natural environments. The only difference of note for chlorophyll measurement is the ability to customize the -5V range on the. This abiltiy allows customers working in low chlorophyll environments to decrease the -5V range and therefore improve the resolution of the analog data. 3 25 2 15 1 5 R 2 =.9319 R 2 =.854 1 2 3 4 5 chlorophyll a (ug/l) Linearity Test 2 15 1 5 R 2 =.997 R 2 =.9941. 5. 1. 15. 2. 25. -5 chlorophyll a (ug/l) Graph 3: Data taken with calibrated values at MLML using Dunaliella cultures. 3 FLUOROMETER COMPARISON REPORT

ACCURACY In order to test the accuracy of measuring chlorophyll a fluorescence, the effects of various interfering compounds were investigated. In an ideal situation, an in situ fluorometer would measure chlorophyll a fluorescence only since chlorophyll a is the only photosynthetic pigment that all algae and photosynthetic bacteria have in common. By measuring chlorophyll a only you will obtain the best estimate of total, living algal biomass. Compounds that can interfere with fluorescent signals optimized to detect chlorophyll a include accessory pigments such as chlorophyll b and c, degraded chlorophyll a, pheophytin a, and dissolved organic matter. All of these interfering substances have fluorescence spectra that can overlap with chlorophyll a given the appropriate concentrations and fluorometer optics. When discussing an instrument s susceptibility to interfering compounds, the performance is determined by the optics. The optics in a fluorometer encompass three components; a light source, optical filters and a light detector. Both the and the instrument use solid-state components, LED light source and photodiode light detector, which makes them equivalent in this category except that the utilizes two LED lamps while the uses one. The difference between the optics lies in the optical filters, specifically the emission filters. The emission filter is placed in front of the light detector. It s purpose is to block all light except for the wavelengths of interest, in this case chlorophyll a fluorescence which peaks at 685nm. The more specific the filter is to the chlorophyll peak, the more accurate the fluorometer will be at measuring total algal biomass. The uses long-pass emission filters which allows in all light >6nm. This means that any material other than chlorophyll a that is excited by the blue LED light source (peak at 46nm) and fluorescing above 6nm will be detected. Likely candidates for this include pheophytin a, chlorophyll b, chlorophyll c and CDOM. The, on the other hand, utilizes a band-pass filter that allows in light from 66-7nm only, peaking at 685nm. Therefore, the fluorescence from compounds other than chlorophyll a are much less likely to be detected. The benefit of using a long-pass filter is the low material cost and higher transmission. The downside is greater susceptibility to interfering compounds. The tests were effective in demonstrating the effect of optical components in the pheophytin testing. One of the most likely interference s is pheophytin a. Pheophytin can accumulate in water column at density gradients, close to the bottom, and be increased due to active zooplankton feeding. A test was run to compare the fluorescence signal from a healthy algal culture and fluorescence from the same culture after exposed to a weak acid that converted much of the chlorophyll in the cells to pheophytin a. 1mL of 1N HCl was added to 35mL of Thallasisosira culture. The culture was stirred for 1 minute and then the fluorescence was read using both instruments. The results were as follows: Before After Before After FS%.45.163 2.91 2.14 Ratio 2.76 1.36 4 FLUOROMETER COMPARISON REPORT

The Difference in the ratios indicates a difference in the fluorometers ability to detect pheophytin fluorescence. The ratio of 2.76 indicates that it is less sensitive to pheophytin fluorescence than the sensor that had a ratio of 1.36. Regression Comprison 1.5 1 R2.95.9.85.8 TD USGS-Thal TD-Thal TD-Thal2 MLML-Pheo MLML-Pheo MLML-Dun MLML-Dun MLML-Dun TD Fresh USGS-Delta Graph 4: The bar graph displays the correlations (r 2 ) obtained from the minimum detection and linearity tests conducted at all three testing sites. In most cases the data correlated with actual chlorophyll a more closely. All tests returned an r 2 of greater than.85 or better. The samples market TD Fresh and USGS-Delta are natural water samples with an assemblage of algal species. All other samples were moncultures of algae. AMBIENT LIGHT REJECTION Ambient light rejection refers to the ability of the fluorometer to be unaffected by sunlight or other ambient light sources. Light rejection is typically performed by flashing the LED light source. When the light source is on, the instrument detects the fluorescence and any ambient light, when the light source is off, the instrument detects the ambient light. The instrument compares the two to determine the fluorescence. Ambient light rejection test results indicate that there is a difference in how the two instruments correct ambient light. The displayed no significant difference in fluorescence signal while the showed a decrease in signal as ambient light increased and became more variable. 7 6 5 4 3 2 1-1 Ambient Light Rejection Test Scufa no light light at 9 degrees light at 18 degrees Position of Light Source % Full Scale 1.5 1..5. -.5 Ambient Light Rejection Test #2 Lit Lab Dark Lab Light 12" Direct Light 12" Variable Graph 5 & 6: Graph 5 displays data from Moss Landing Marine Lab. The instruments were placed in a solution of rhodamine WT dye. A halogen light source was used to test ambient light. Graph 6 displays data from tests run at Turner Designs with a halogen light source. The units were not placed in a solution. 5 FLUOROMETER COMPARISON REPORT

RESPONSE TIME The graph below shows the response time of the two instruments as they were alternately placed into samples of natural water containing algae and blank deionized water. The readings were taken at one-second intervals. The difference in the graphs displays a significant difference between the two instruments. The system uses a moving average as opposed to the realtime sampling of the. The moving average allows the system to smooth the data as well as enables it to improve the limit of detection. The, designed for high-speed vertical profiling applications, samples at a rate of 1Hz and outputs data at 1Hz or 5Hz. The spikes at the beginning of the readings of the algal cultures is due to removing the instrument from the blank water and then placing it into an algal culture. The change from air to liquid creates some bubbles that cause spikes in the readings. The sensor is less sensitive to these changes due to the slow moving average (~1sec) of the data which smoothes the data but does not supply instantaneous data. RESPONSE TIME TEST IN VIVO FLUORESCENCE 45 4 35 3 25 2 15 1 5-5 13:55:55 13:56:38 13:57:22 13:58:5 13:58:48 13:59:31 14::14 14::58 14:1:41 TIME Graph 7 CONCLUSIONS The results of the performance tests indicate that both instruments are effective at estimating chlorophyll a concentrations in typical environmental conditions. Although accuracy tests favored the fluorometer, most correlations between the in vivo fluorescence and the actual chlorophyll concentration exceeded.9 for both instruments. It is in the presence of interfering compounds where the data from the two units may differ significantly. Results from testing with degraded chlorophyll, pheophytin, support this conclusion, with the not differentiating between chlorophyll and pheophytin as well as the. Other extreme environmental conditions could limit the effectiveness of the probe. Conditions such as high ambient light,from shallow waters with reflective sediment, and low productivity waters could be problematic. Also, due to the slow response time of the probe, vertical profiling would be a challenge. On the other hand, the probe seems well suited for deployed use in coastal or freshwater environments where chlorophyll levels exceed 1µg/L and ambient light and interfering compounds are not a significant issue. The performed well in all categories. It is a more versatile instrument because of ability to be used in self-contained mode. 6 FLUOROMETER COMPARISON REPORT

REFERENCES: Arar, E.J. and G.B. Collins 1996. In vitro determination of chlorophyll a and pheophytin a in marine and freshwater algae by fluorescence. E.P.A. Method 445., Revision 1.2 Welschmeyer, N.A. 1994. Fluorometeric analysis of chlorophyll a in the presence of chlorophyll b and pheopigments. Limnol. Oceanogr. 39: 1985-1992 TESTING FACILITIES: MOSS LANDING MARINE LABORATORY, MOSS LANDING, CA. Tests run under the supervision of Dr. Nick Welschmeyer. TURNER DESIGNS TESTING FACILITY, SUNNYVALE, CA. UNITED STATES GEOLOGICIAL SURVEY, MENLO PARK, CA Report written by Robert Ellison, Turner Designs, Inc., 22 7 FLUOROMETER COMPARISON REPORT