Increasing Depth of Coverage in Data Independent Acquisition with Higher Sample Loads and Smaller Q1 Windows

Similar documents
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

ProteinPilot Report for ProteinPilot Software

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Tutorial 9: SWATH data analysis in Skyline

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

A Streamlined Workflow for Untargeted Metabolomics

MarkerView Software for Metabolomic and Biomarker Profiling Analysis

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Pesticide Analysis by Mass Spectrometry

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Alignment and Preprocessing for Data Analysis

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

Pinpointing phosphorylation sites using Selected Reaction Monitoring and Skyline

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Research-grade Targeted Proteomics Assay Development: PRMs for PTM Studies with Skyline or, How I learned to ditch the triple quad and love the QE

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Tutorial for Proteomics Data Submission. Katalin F. Medzihradszky Robert J. Chalkley UCSF

Quantitative proteomics background

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Mass Spectrometry Based Proteomics

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

using ms based proteomics

Retrospective Analysis of a Host Cell Protein Perfect Storm: Identifying Immunogenic Proteins and Fixing the Problem

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Guidance for Industry

Session 1. Course Presentation: Mass spectrometry-based proteomics for molecular and cellular biologists

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Analyst 1.6 Software. Software Reference Guide

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Signal, Noise, and Detection Limits in Mass Spectrometry

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

AMD Analysis & Technology AG

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Mass Spectrometry Signal Calibration for Protein Quantitation

Waters Core Chromatography Training (2 Days)

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

Quick Start Guide Chromeleon 7.2

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society

LC-MS/MS for Chromatographers

Overview. Purpose. Methods. Results

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Background Information

Setting up a Quantitative Analysis MS ChemStation

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Analyst Software

Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit

MASCOT Search Results Interpretation

Analyst 1.6 Software. Getting Started Guide

Introduction to Proteomics 1.0

Supplementary Materials and Methods (Metabolomics analysis)

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100)

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry

High-Throughput 3-D Chromatography Through Ion Exchange SPE

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Protein Prospector and Ways of Calculating Expectation Values

MassHunter for Agilent GC/MS & GC/MS/MS

Electrospray Ion Trap Mass Spectrometry. Introduction

Comparative LC-MS: A landscape of peaks and valleys

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

High Sensitivity MS Determination of Carryover in a New Autosampler Design

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Mass Frontier 7.0 Quick Start Guide

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Transcription:

Increasing Depth of Coverage in Data Independent Acquisition with Higher Sample Loads and Smaller Q1 Windows SWATH Acquisition on the TripleTOF 6600 Systems Christie Hunter 1, Ben Collins 2, Ludovic Gillet 2, Ruedi Aebersold 2 1 AB SCIEX, USA, 2 ETH Zurich, Switzerland In data independent acquisition (DIA), high resolution MS/MS is acquired across a mass range, using wide Q1 windows, which creates composite spectra from which fragment ion chromatograms (XICs) can be extracted for quantitation. The most widely used of these techniques is MS/MS ALL with SWATH Acquisition on the TripleTOF Systems. The depth of coverage and quality of quantitation obtained from these experiments will depend on the signal/noise of the generated XICs. Increasing specificity to decrease chemical noise can be achieved by using increasingly smaller Q1 windows, and varying the window width as a function of precursor density using the variable window functionality 1. Increasing signal can be obtained by using higher sample loads on optimized chromatography. Higher signal must be accompanied by an increase in dynamic range of the mass spectrometer detection system to take full advantage and dig deeper into the sample. Here, the combination of higher sample loads, smaller variable Q1 windows and expanded detector dynamic range of the TripleTOF 6600 System will be explored as a strategy for increasing the depth of coverage of complex proteomic samples. Key Advantages of the TripleTOF 6600 System for SWATH Acquisition Increased dynamic range on the TripleTOF 6600 System Greater than 5 orders linear dynamic range using an ADC detection system Mass accuracy is maintained intra-scan across a wide abundance range of peptides Increased workflow flexibility, enabling higher sample loads High sensitivity, resolution and speed of MS/MS acquisition allows a large number of Q1 isolation windows to be used in an LC time frame with no loss in quantitation quality Up to 200 Q1 windows per cycle on TripleTOF 6600 system, with upper mass limit 2250 m/z Narrower Q1 isolation windows provide improved data quality through increased specificity Variable window acquisition method building 1 can be used to optimize SWATH acquisition for the proteome of interest Figure 1. Importance of Dynamic Range. In SWATH Acquisition, fragment ion XICs are extracted from full scan MS/MS data postacquisition and used for quantitation of peptides. The intra-scan dynamic range of the spectrum will determine how broad an abundance range accurate quantitation can be obtained. The wide dynamic range of the TripleTOF 6600 System enables low abundant peptides and fragments (orange box) to be quantified in the presence of high abundant peptides, which further extends the power of SWATH acquisition for quantitation in complex biological samples. Automatically design methods that use of a greater number of smaller variable sized Q1 windows in combination with higher sample loads on the TripleTOF 6600 System provided a 90% increase in peptide coverage of the proteomic sample. p 1

Methods Sample Preparation: Yeast proteome samples were reduced, alkylated, and digested, providing a solution of ~1 µg/µl. irt peptides (Biognosis, Zurich) were spiked (1:20 dilution of stock) in for retention time calibration. Chromatography: The samples were analyzed using the ekspert nanolc 425 System with the chiplc System operating in serial column mode 2 (Eksigent). The samples were first loaded on the first chiplc column (75 µm x 15 cm ChromXP C18-CL, 3 µm, 300 Å) and washed for 30 mins at 0.5 µl/min. Then, elution gradients of 5-30% acetonitrile (0.1% formic acid) in 60 or 120 mins were used to elute peptides off the first column and through the second nano chiplc column. Both columns were maintained at 35 ºC for retention time stability. Mass Spectrometry: Eluent from the column was sprayed using the Nanospray Source into a TripleTOF 6600 system with Analyst Software TF 1.7 (AB SCIEX). The SWATH acquisition methods were built using the Variable Window Calculator 3 and the SWATH Acquisition method editor. A variety of window numbers and accumulation times were explored. Total cycle time was kept constant at 3.0 sec. Q1 mass range interrogated was 400-1250 m/z, and a TOF MS scan (200 msec) was acquired in every cycle. Data Processing: A spectral ion library for the yeast proteome was used to drive data processing 4. irt standard peptides were used for automatic retention time calibration of the different LC gradients with the ion library retention times. All SWATH acquisition data were processed using the MS/MS ALL with Figure 3. Higher Sample Loads and Longer Gradients Provide Significant Increases in Peptide Coverage. Five replicate analyses were performed with varied sample load and gradient length. Peptides with high detection confidence (1% FDR cutoff) were analyzed for reproducibility and plotted according to cumulative %CV. Using a high quantitation requirement (20% CV cutoff, gray line), a 40% increase in peptide coverage was obtained when the load was increased from 1 to 3 µg yeast on column. A further 12% gain was obtained when the gradient length was extended to 120min from 60mins. Figure 2. Higher Sample Loads And Longer Column Chromatography. Because the TripleTOF 6600 system has an expanded dynamic range and can accommodate a much higher signal level, higher sample loadings can be used during SWATH Acquisition. To take advantage of this, the serial column workflow was employed, where sample was first loaded onto a 15cm column (acting as a trap), desalted, and then eluted onto a second 15cm column to yield a 30 cm column for higher peak capacity. The load was increased from 1 µg (blue) to 3 µg (pink) and chromatographic peak shape and robustness was maintained. SWATH Acquisition MicroApp 2.0 in PeakView Software. Peptide peak group detections were filtered at a 1% global FDR and replicates were analyzed using the SWATH Replicates Analysis Template 5. Increasing Peptide Detection with Increasing Sample Load A spectral ion library with peptides representing proteins for the entire yeast proteome was used to interrogate the SWATH Acquisition experiments. irt peptides were added for retention time alignment during processing. Replicate analyses were also done for each experimental condition such that peak area variation could be measured (% CV). Six fragment ion XICs for each peptide were extracted and then scored. The peptides were then filtered at 1% FDR from the different SWATH acquisition experiments and reproducibility of the data assessed. The serial column workflow was used to increase the loading capacity of the experiment (Figure 2) and to also gain from an extended column length of 30cm column length. The baseline reference experiment was a typical 1 µg sample load and a 60 min gradient on the 30cm column, using 60 variable windows as explored in previous work 1. Increasing the sample load to 3 µg provided a 31% increase in peptide detections (< 1% FDR and <20% CV were used as the figures of merit). Next the gradient length was extended to 120 minutes and a further 22% increase in peptide detections was obtained (Figure 3). The extended dynamic range of the TripleTOF 6600 system allows for higher sample load providing this significant increase in sample coverage. p 2

Narrower Isolation Windows through Variable Window Acquisition Provides Increased Specificity Next, the impact of increasing the number of Q1 windows, and thereby decreasing the precursor ion isolation width, was explored for its impact on peptide detection and quantitation quality. This was done at the higher sample load to fully take advantage of the enhanced peptide signal. As the number of Q1 windows was increased from 60 to 80 to 100 windows and the width of the isolation windows was narrowed accordingly, the number of peptides that could be detected (<1% FDR and <20% CV) increased (Figure 4, top). This decrease in window size resulted in a further 20% increase in peptide coverage (Figure 4, bottom). This is mostly likely due to an improvement in specificity and signal/noise obtained from the reduced complexity of the MS/MS data from each SWATH window across the whole mass range. The higher specificity will improve the peak group scoring process as the XIC data will have much improved S/N and less chance of interferences that could confound peak group detection. Improved S/N will also improve consistency of peak integration and therefore quantitation quality. Further work will be done to explore how much farther we can refine this technique and utilize the 200 window maximum of the TripleTOF 6600 System. Figure 4. Further Increasing Window Sizes Provides More Quality Peptide Detection. (Top) The number of windows used to cover the mass range was increased from 60 to 100 windows and the reproducibility curves were plotted. The number of confident peptide detections (<1% FDR) with 20% CV or better further increased as the window size decreases (gray dotted line). (Bottom) The number of peptide detections with <1% FDR and <20% CV was plotted vs. increasing number of Q1 windows and 20% gain in peptides was observed. Maintaining High Quantitative Reproducibility Figure 5. Maintaining Quantitation Quality with Increasing Depth of Coverage. The accumulation time per MS/MS is reduced as the number of Q1 window steps is increased to ensure an optimal cycle time for LC/MS quantitation. Therefore, assessing the reproducibility of the XICs as a function of peak area for the confidently detected peptides is important to monitor. The median CV for the XIC peak areas in a particular peak area range was measured and plotted vs. peak area across 5 orders dynamic range and the CV curves for the two experimental conditions look very similar. Therefore, even though the acquisition rates are faster for the 100 variable window strategy, the quantitation has not degraded. As the number of windows increases, the accumulation time used per window must decrease to maintain the optimal cycle time. As this could have an impact on quantitation, especially on the lower abundant peptides, it is important to monitor the impact on %CV as a function of XIC peak area for the various acquisition strategies. As seen in Figure 5, the median CV for the XIC peak areas in a particular peak area range was measured and plotted vs. peak area across 5 orders dynamic range and the results show a very similar performance between the experimental strategies. Therefore, the accumulation time could be reduced from 37 msec per MS/MS for the 60 variable window experiment to 30 msec per MS/MS for the 100 variable window experiment with no major reduction in quantitative quality. p 3

Data Completeness One reason why researchers move from data dependent acquisition to targeted quantitation with MRM when doing larger study sizes is because of the missing values problem inherent in data dependent acquisition strategies. For the key peptides and proteins of interest, MRM provides a measurement for each targeted analyte in almost every sample across very large numbers of samples, across a broad intensity regime (Figure 6) providing data completeness in the study. However, the multiplexing capacity is limited to a few 100 proteins as opposed to the 1000s of proteins identified using conventional data dependent acquisition (DDA) strategies. But with SWATH Acquisition (DIA), MS/MS is collected on all masses all the time, and peptide data is analyzed in a targeted fashion from this comprehensive dataset. Therefore, much better data completeness, approaching that of MRM, is achieved but with much higher multiplexing. Conclusions MS/MS ALL with SWATH Acquisition combined with the expanded dynamic range of the TripleTOF 6600 system and the variable window acquisition strategy enables much deeper coverage of the proteomic sample. Here a 90% gain was achieved in peptides detected with high confidence and quantified with good reproducibility. Variable window SWATH Acquisition 1 combined with higher number of windows greatly improves the specificity of the SWATH acquisition experiment. The broad linear dynamic range (> 5 orders LDR) of the TripleTOF 6600 system enables higher sample loads during SWATH acquisition. Combining more optimized and smaller SWATH windows with increased signal due to higher sample loads provided a 90% gain in peptide coverage. In this study, we wanted to ensure that the additional peptide detections obtained as we expanded our coverage still provided the desired high data completeness (maintain a low number of missing values ). In Figure 6, the data completeness across the 5 replicates for the 1% FDR fragment ions and peptides were plotted vs. number of windows. Therefore more peptides are added to the dataset with high quantitative quality and high data completeness. Figure 6. Assessing Data Completeness. (Top) SWATH Acquisition provides the high depth of coverage of conventional data dependent (DDA) workflows but with the data completeness and quality of quantitation of the targeted MRM workflows. (Bottom) Analysis was done to confirm that the increased peptide coverage obtained with the increased sample load and decreased window size provided more fragment (brown) and more peptide (blue) coverage with 5 out of 5 measurements (complete rows, high data completeness). p 4

References 1. Improved Data Quality Using Variable Q1 Window Widths in SWATH Acquisition - Data Independent Acquisition on the TripleTOF 6600 and 5600+ Systems. AB SCIEX technical note 10090114-01. 2. Increasing Depth of Coverage using Serial Two- Column Workflows - Powerful Proteomics Workflows Enabled by the Eksigent chiplc with nanolc-ultra System. AB SCIEX Technical note 1870211-03. 3. Variable Window Calculator Excel tool. Download from www.absciex.com\variablewindowscalculator. 4. Yeast Spectral Library - Selevsek N, Chang CY, Gillet LC, Navarro P, Bernhardt O, Reiter L, Vitek O, Aebersold R. Reproducible and consistent quantification of the S. cerevisiae proteome by SWATH-MS (submitted manuscript). 5. SWATH Replicates Template. Download from www.absciex.com\swathreplicatestemplate. For Research Use Only. Not for use in diagnostic procedures. 2014 AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Publication number: 10220314-01 p 5